K Number
K210073
Manufacturer
Date Cleared
2022-03-01

(413 days)

Product Code
Regulation Number
880.5725
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Agilia VP Infusion System is intended for adult and pediatric care for the intermittent or continuous delivery of parenteral fluids, medications, blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, subcutaneous and intraosseous using dedicated administration sets.

The Agilia VP Infusion System is also intended for neonatal care for the intermittent or continuous delivery of parenteral fluids for hydration and nutrition, blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, and subcutaneous using dedicated administration sets.

It is intended for use by trained healthcare professionals in healthcare facilities.

Device Description

The Agilia VP Infusion System includes the Agilia VP MC WiFi Volumetric Infusion Pump, which is a programmable electronic medical system dedicated to administering a pre-determined volume of an infusion product at a programmed rate, in combination with Volumat Line administration sets and optional accessories. The optional accessories are identified as follows:

  • Agilia Link 4, 6 and 8 Stacking rack system intended to power and organize 4, 6 or 8 pumps at the patient bedside.
  • Agilia Duo two-channel accessory designed to power two Agilia infusion pumps.
  • Agilia USB Cable - intended to connect the Agilia VP infusion pump to a PC for serial communication.
AI/ML Overview

The provided document is a 510(k) summary for the Fresenius Kabi Agilia VP Infusion System. This document focuses on demonstrating substantial equivalence to a predicate device (Agilia Infusion System (K121613)), rather than proving the device meets specific acceptance criteria based on a clinical study evaluating its diagnostic or therapeutic performance (like an AI/CADe device).

Therefore, the requested details specific to a study that proves the device meets acceptance criteria (especially for AI/CADe devices, such as sample size, ground truth, expert adjudication, MRMC studies, standalone performance, training set details) are not applicable (N/A) in this context.

This submission relies on non-clinical testing and comparison to a predicate device to establish safety and effectiveness.

Here's an analysis of the document regarding acceptance criteria and the methods used to prove substantial equivalence:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not present a formal table of quantitative acceptance criteria with reported numerical performance values in the way one would for a diagnostic AI device (e.g., sensitivity, specificity thresholds). Instead, it relies on demonstrating similarity to a predicate device and showing conformance to industry standards and internal design verification/validation testing.

The "acceptance criteria" are implied by:

  • Substantial Equivalence: The primary "acceptance criterion" for a 510(k) submission is to demonstrate that the new device is as safe and effective as a legally marketed predicate device.
  • Conformance to Standards: Meeting the requirements of relevant electrical safety, EMC, and usability standards (e.g., IEC 60601-1, IEC 62366-1).
  • Verification and Validation Testing: This implicitly means the device performs according to its design specifications.
  • Reliability Goals: The device met its internal reliability goals.

Implied Acceptance Criteria and Reported Performance (from comparison table and non-clinical testing section):

Acceptance Criteria (Implied)Reported Device Performance
Infusion AccuracyAgilia VP Infusion System: ± 5% under most conditions. (Stated as "Similar" to predicate which also had ± 5%). Detailed flow rate accuracy disclosed in labeling. Flow rate and bolus accuracy testing conducted per AAMI TIR101 2021.
Safety (Electrical & EMC)Agilia VP Infusion System: Complies with IEC 60601-1: 2005 +A1:2012 (3rd edition) for Electrical Safety and IEC 60601-1-2: 2014 for EMC testing. ("Similar" due to application of state-of-the-art standards).
Usability/ Human FactorsAgilia VP Infusion System: Human factors evaluation conducted to validate effectiveness of use-related features/functionality and use error-related mitigations, following IEC 62366-1 Edition 1.0 2015. ("Passing results").
Software Verification & ValidationAgilia VP Infusion System: Performed per FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005).
CybersecurityAgilia VP Infusion System: Cybersecurity testing performed confirmed the system is effective in addressing cybersecurity threats, referencing FDA guidances (Management of Cybersecurity, Oct 2014; Postmarket Management of Cybersecurity, Dec 2016).
ReliabilityAgilia VP Infusion System: Device reliability activities, testing and statistical analysis confirmed the system met its reliability goal at the system, device subsystem, and subsystem component levels.
Risk AssessmentAgilia VP Infusion System: Residual risks analyzed and determined to be acceptably low using industry-standard risk analysis practices and regulatory guidance.
Performance of New Features (Ramp-up/down, Sequential modes, WIFI, auto-restart)Agilia VP Infusion System: Performance testing demonstrates these additional modes/features do not introduce any new issues of safety or effectiveness. ("Similar" or "Similar" in comments of Table 2). The document states: "Results of verification and validation activities demonstrate that the Agilia VP Infusion System is substantially equivalent to the predicate Agilia Infusion System (K121613)." and "Design verification and validation testing confirmed the Agilia VP Infusion System met user needs and design inputs. Testing results conformed with acceptance criteria."

The following points are N/A for this type of 510(k) submission which is for an infusion pump, not an AI/CADe device.

2. Sample size used for the test set and the data provenance: N/A (No clinical test set for diagnostic performance)

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (No ground truth establishment for diagnostic performance)

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: N/A

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: N/A

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): N/A (Ground truth in this context would refer to engineering specifications and performance measurements against those specifications, rather than clinical ground truth for diagnostic accuracy)

8. The sample size for the training set: N/A (Not an AI/ML device in the context of diagnostic algorithms; "training set" would relate to internal engineering development and testing, not a formal clinical data training set for an AI model)

9. How the ground truth for the training set was established: N/A

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 1, 2022

Fresenis Kabi AG % Keith Dunn Director Regulatory Affairs Fresenius Kabi LLC, USA 3 Corporate Dr Suite 300 Lake Zurich, Illinois 60047

Re: K210073

Trade/Device Name: Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat Lines Administration Sets, Agilia Link, Agilia Duo Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: FRN, MRZ, FPA Dated: January 8, 2021 Received: January 12, 2021

Dear Keith Dunn:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Courtney H. Lias, Ph.D. Director OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210073

Device Name Agilia VP Infusion System

Indications for Use (Describe)

The Agilia VP Infusion System is intended for adult and pediatric care for the intermittent or continuous delivery of parenteral fluids, medications, blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, subcutaneous and intraosseous using dedicated administration sets.

The Agilia VP Infusion System is also intended for neonatal care for the intermittent or continuous delivery of parenteral fluids for hydration and nutrition, blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, and subcutaneous using dedicated administration sets.

It is intended for use by trained healthcare professionals in healthcare facilities.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Fresenius Kabi. The logo consists of a blue abstract symbol on the left, resembling three vertical wavy lines. To the right of the symbol, the words "FRESENIUS" and "KABI" are stacked on top of each other, also in blue. The text is in a bold, sans-serif font.

510(k) SUMMARY

Date Prepared

February 28, 2022

Owner/Operator

Fresenius Kabi AG 61346 Bad Homburg Germany

Owner/Operator Number: 9027285

Contact Person

Keith Dunn

Director, Regulatory Affairs Fresenius Kabi USA, LLC Three Corporate Drive, 2nd Floor Lake Zurich, IL 60047 USA

Phone: 224-817-2430 Fax: 847 550 2960 E-mail: keith.dunn(@fresenius-kabi.com

Secondary Contact Person

Jason Ma

Sr. Manager, Regulatory Affairs Fresenius Kabi USA, LLC Three Corporate Drive, 2nd Floor Lake Zurich, IL 60047 USA

Phone: 224-817-4100 847 550 2960 Fax: E-mail: Jason.ma01(@fresenius-kabi.com

Device Trade Name

Agilia VP Infusion System

Common Name/Usual Name:

Infusion Pump and Accessories

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the Fresenius Kabi logo. The logo consists of a blue graphic on the left and the words "FRESENIUS KABI" on the right. The graphic is made up of three blue, curved lines. The words "FRESENIUS KABI" are in blue, with "FRESENIUS" on top and "KABI" on the bottom.

Classification Name

21 CR 880.5725 Infusion Pump

Product Code and Classification Panel

The Agilia VP Infusion System (pump, set and accessories) has been classified as Class II under 21 CFR 880.5725, and reviewed by the General Hospital Devices Panel (80). For reference, the product code and classification of each device subsystem is identified in Table 1.

Device NameRegulation NumberClassDescriptionPanelProductCode
Agilia VP MC WiFi VolumetricInfusion Pump21 CFR 880.5725IIInfusion Pump80FRN
Agilia Link21 CFR 880.5725IIInfusion pump accessories80MRZ
Agilia Duo21 CFR 880.5725IIInfusion pump accessories80MRZ
Agilia USB Cable21 CFR 880.5725IIInfusion pump accessories80MRZ
Volumat Line administrationsets21 CFR 880.5725IIInfusion pump accessories80FPA

Table 1: Device Classification

Legally Marketed Device Under Which Substantial Equivalence is Being Claimed

Agilia Infusion System (K121613)

Device Description

The Agilia VP Infusion System includes the Agilia VP MC WiFi Volumetric Infusion Pump, which is a programmable electronic medical system dedicated to administering a pre-determined volume of an infusion product at a programmed rate, in combination with Volumat Line administration sets and optional accessories. The optional accessories are identified as follows:

  • Agilia Link 4, 6 and 8 Stacking rack system intended to power and organize 4, 6 or ● 8 pumps at the patient bedside.
  • Agilia Duo two-channel accessory designed to power two Agilia infusion pumps. ●
  • Agilia USB Cable - intended to connect the Agilia VP infusion pump to a PC for serial communication.

Statement of Intended Use/Indications For Use

The Agilia VP Infusion System is intended for adult and pediatric care for the intermittent or continuous delivery of parenteral fluids. medications, blood and blood derivatives through clinically accepted parenteral routes of administration. These routes

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for Fresenius Kabi. The logo consists of a blue graphic on the left, which appears to be three stylized vertical lines. To the right of the graphic are the words "FRESENIUS KABI" in blue, with "FRESENIUS" on top of "KABI". There is a small subscript number 1 next to the letter S in Fresenius.

of administration include intravenous, intra-arterial, subcutaneous and intraosseous using dedicated administration sets.

The Agilia VP Infusion System is also intended for neonatal care for the intermittent or continuous delivery of parenteral fluids for hydration and nutrition, blood and blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, and subcutaneous using dedicated administration sets.

It is intended for use by trained healthcare professionals in healthcare facilities.

Technological Comparison as Compared to the Predicate Device

A comparison between the predicate device and the subject device is provided in the Table 2.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the Fresenius Kabi logo. The logo consists of a blue abstract symbol on the left and the words "FRESENIUS KABI" in blue on the right. The abstract symbol is made up of three vertical wavy lines.

CharacteristicPredicate DeviceSubject DeviceComment
K121613K210073
Intended Use /Indications forUseThe Agilia Infusion System isa transportable equipmentintended for use by trainedhealthcare professionals inhealthcare facilities andhomecare environments onadults, pediatrics andneonate's human patients toadminister via a singlechannel or mounted on amultiple channels rackaccessory:• Intermittent or continuousdelivery of parenteral fluids(solutions, colloids,parenteral nutrition) andmedications (including butnot limited to diluted drugs,chemotherapy) throughclinically accepted IV routesof administration.• Transfusion of blood andblood derivatives products.The Agilia VP InfusionSystem is intended for adultand pediatric care for theintermittent or continuousdelivery of parenteral fluids,medications, blood and bloodderivatives through clinicallyaccepted parenteral routes ofadministration. These routesof administration includeintravenous, intra-arterial,subcutaneous andintraosseous using dedicatedadministration sets.The Agilia VP InfusionSystem is also intended forneonatal care for theintermittent or continuousdelivery of parenteral fluidsfor hydration and nutrition,blood and blood derivativesthrough clinically acceptedparenteral routes ofadministration. These routesof administration includeintravenous, intra-arterial,and subcutaneous usingdedicated administration sets.It is intended for use bytrained healthcareprofessionals in healthcarefacilities.SimilarThe intended usebetween the predicateand proposed deviceare the same in termsof intended patient,route ofadministrations, andtreatment. Per FDAguidance, thestatement was revisedto specifically namethe intended routes ofadministration and todifferentiate theintended patientpopulation .
Type of PumpVolumetric, Linear PeristalticPumpVolumetric, Linear PeristalticPumpSame
PatientPopulationAdults, pediatrics, neonatesAdults, pediatrics, neonatesSame
DeliveryModesContinuous, Intermittent,Primary, Secondary, KVO,Bolus (Manual andProgrammable), LoadingDoseContinuous, Intermittent,Primary, Secondary, KVO,Bolus (Manual andProgrammable), LoadingDose, Ramp-up / Rampdown, SequentialSimilarPerformance testingdemonstrates theadditional modes donot introduce any newissues of safety oreffectiveness.
CharacteristicPredicate DeviceSubject DeviceComment
K121613K210073
InfusionAccuracy$\pm$ 5%$\pm$ 5% under most conditionsSimilarDetailed flow rateaccuracy disclosed inthe labeling
WIFINoYesSimilarPerformance testingdemonstrates theaddition of WIFIfunctionality does notintroduce any newissues of safety oreffectiveness.
Pump ControlMicrocontroller controlledMicrocontroller controlledSame
Number ofChannels11Same
MechanicalSensorsair, occlusion, temperature,pressure, door, set installationair, occlusion, temperature,pressure, door, set installationSame
BarcodingNoNoSame
AutomatedProgrammingNoNoSame
ManualProgrammingYesYesSame
Piggy back /secondaryYesYesSame
AlarmsAir, occlusion, temperature,system malfunction,set installation, door,low drug reservoir volume,infusion complete, battery,technicalAir, occlusion, temperature,system malfunction,set installation, door,low drug reservoir volume,infusion complete, battery,technicalSame
ClinicalAdvisoriesYesYesSame
InfusionModesBasic, continuous,intermittent, primary /secondary, KVO, Bolus(direct and programmed),Loading DoseBasic, continuous,intermittent, primary /secondary, KVO, Bolus(direct and programmed),Loading Dose, Ramp-up /Ramp Down, SequentialSimilarPerformance testingdemonstrates theadditional infusionmodes do notintroduce any newissues of safety oreffectiveness.
CharacteristicPredicate DeviceSubject DeviceComment
K121613K210073
FeaturesKeypad lockKeypad lock/automatic lock,occlusion auto restartSimilarPerformance testingdemonstrates theadditional keypadfeatures do notintroduce any newissues of safety oreffectiveness.
Dose ErrorReductionSoftware /Drug LibrarySoftwareCompatibilityBasic Profile (within pump)Basic Profile (within pump)Same
WirelessServerSoftwareNoYesSimilarPerformance testingdemonstrates theaddition of wirelessserver software doesnot introduce any newissues of safety oreffectiveness.
AdministrationSetsVolumat Line administrationsetsVolumat Line administrationsetsSame
Racking /Organization /PowerLink+ AgiliaAgilia LinkSimilarSimpler, non-communicatingaccessory for proposeddevice
Agilia DuoAgilia DuoSame
ElectricalSafetyIEC 60601-1: 1998 +A1:1995(2nd edition)IEC 60601-1-8: 2003+A1:2006IEC 60601-1-2: 2007ANSI/AAMI ID26:2004/(R)2009IEC 60601-1: 2005+A1:2012 (3rd edition)IEC 60601-1-8: 2012IEC 60601-1-2: 2014ANSI/AAMI ID26:2004/(R)2009SimilarPerformance testingdemonstrates theapplication of state-of-the-art standards doesnot introduce any newissues of safety oreffectiveness.
Mechanical /PowerSpecificationsPump Size: 5.3 x 7.5 x 6.7inchesPump Weight: App. 4.4 lbs.Line Power: 100-240 VACPump Size: 5.3 x 7.5 x 6.7inchesPump Weight: App. 4.4 lbs.Line Power: 100-240 VACSame
OperatingEnvironmentTemperature: 41° F to 104° FRelative Humidity: 20-90%Temperature: 41° F to 104° FRelative Humidity: 20-90%Same
CharacteristicPredicate DeviceK121613Subject DeviceK210073Comment
StorageEnvironmentTemperature: 14° F to 140° FRelative Humidity: 10-90%w/o condensationTemperature: 14° F to 140° FRelative Humidity: 10-90%w/o condensationSame

Table 2: Predicate Comparison

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the Fresenius Kabi logo. The logo consists of a blue abstract symbol on the left and the words "FRESENIUS KABI" in blue on the right. The abstract symbol is made up of three curved lines.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the Fresenius Kabi logo. The logo consists of a blue abstract symbol on the left and the words "FRESENIUS KABI" in blue on the right. The word "FRESENIUS" is above the word "KABI".

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the Fresenius Kabi logo. The logo consists of three blue wavy lines on the left side. To the right of the wavy lines is the text "FRESENIUS KABI" in blue, with "FRESENIUS" on top and "KABI" on the bottom.

To demonstrate substantial equivalence between the subject and predicate device the following non-clinical tests were performed:

  • Verification testing of product requirements.
  • Human factors engineering testing of product requirements associated with critical tasks.
  • Testing for the reliability goals of the device.

Non-Clinical Testing

A safety assurance case was provided for the Agilia VP Infusion System as recommended in the FDA Guidance Document, Infusion Pumps Total Product Life Cycle issued December 2, 2014.

The stated goal of the safety assurance case is:

  • . The system design is acceptably safe for its intended use by its intended users and in its intended use environment.
    The assurance case defines the device system, including the operational description, system definition, indications for use, patient population, intended users and use environments. The supporting assurance arguments covered the following attributes:

  • . Residual risks are analyzed and determined to be acceptably low using industrystandard risk analysis practices and regulatory guidance

  • The Agilia VP Infusion System is at least as safe as an equivalent infusion system.

  • The Agilia VP Infusion System is Verified and Validated for its intended use as it relates to safety

  • The Agilia VP Infusion System is reliable over the system's expected lifetime. ●

The following evidence was included in the safety assurance case:

  • Risk Management File ●

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image contains the logo for Fresenius Kabi. The logo consists of a blue graphic on the left, which is three curved lines stacked on top of each other. To the right of the graphic is the text "FRESENIUS" in blue, with "KABI" in blue below it.

  • Design verification and validation testing confirmed the Agilia VP Infusion System . met user needs and design inputs. Testing results conformed with acceptance criteria. Flow rate and bolus accuracy testing were conducted by following AAMI TIR101 2021.
  • Device reliability activities, testing and statistical analysis confirmed the Agilia VP Infusion System met its reliability goal at the system, device subsystem, and subsystem component level.
  • Software verification and validation were performed per FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices issued May 11, 2005.
  • Human factors evaluation conducted to validate the effectiveness of use-related features/functionality and use error-related mitigations in the associated use environments. IEC 62366-1 Edition 1.0 2015 Medical Devices Part 1: Application of usability engineering to medical devices was followed
  • Electrical and Electromagnetic Compatibility testing were conducted. The Infusion System complies with the following standards:
    • Electrical Safety per IEC 60601-1 O
    • o EMC testing per IEC 60601-1-2
  • Cybersecurity testing performed confirmed the system is effective in addressing ● cybersecurity threats.
    • Content of Premarket Submissions for Management of Cybersecurity, October 2, O 2014
    • Postmarket Management of Cybersecurity in Medical Devices, December 28, O 2016

Clinical Testing

Clinical evaluation is not required for this submission to support substantial equivalence. Human Factors studies have been conducted on the subject device demonstrating passing results.

Conclusion

The results of software, electrical safety, system verification and validation testing conclude that the Agilia VP Infusion System is safe and effective for the intended users, uses and use environments, and that no further clinical investigation or testing is needed. The methods and results described in the verification and human factors evaluation (HFE/UE) reports support this conclusion.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image contains the logo for Fresenius Kabi. The logo consists of a blue abstract symbol on the left, resembling three interconnected wave-like shapes. To the right of the symbol, the words "FRESENIUS" and "KABI" are stacked on top of each other, also in blue. The overall design is clean and corporate, suggesting a company in the healthcare or pharmaceutical industry.

Results of verification and validation activities demonstrate that the Agilia VP Infusion System is substantially equivalent to the predicate Agilia Infusion System (K121613).

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).