vatech A9 (Model : PHT-30CSS) is intended to produce panoramic, cone beam computed tomography, or cephalometric digital x-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
vatech A9 (Model : PHT-30CSS) is an advanced 3-in-1 digital X-ray imaging system that incorporates PANO, CEPH (Optional), and CBCT scan imaging capabilities into a single system. vatech A9 (Model : PHT-30CSS), a digital radiography imaging system, is specially designed to take X-ray images of patients on the chair and assist dentists. Designed explicitly for dental radiography, vatech A9 (Model : PHT-30CSS) is a complete digital X-ray system equipped with imaging viewers, an X-ray generator, and a dedicated SSXI detector. The digital CBCT system is based on a CMOS digital X-ray detector. The CMOS CT detector is used to capture 3D radiographic images of the head, neck, oral surgery, implant, and orthodontic treatment.
Here's an analysis of the acceptance criteria and study information for the Vatech A9 (Model: PHT-30CSS) based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list numerical acceptance criteria in a table format. Instead, it states that the device's performance was compared to a predicate device and international standards. The general acceptance criterion appears to be "equivalent or better than the predicate device" in terms of image quality and meeting relevant international standards for X-ray systems.
| Acceptance Criteria (Inferred) | Reported Device Performance (as stated) |
|---|---|
| Image Quality: Equivalent or better than the predicate device (Green16/Green18, K170066) in terms of Contrast, Noise, CNR, and MTF, for CBCT, PANO, and CEPH images. | "The results demonstrated that the general image quality of the subject device is equivalent or better than the predicate device." (Applies to Contrast, Noise, CNR, MTF in CT; also stated for PANO/CEPH/CBCT images generically) |
| Dosimetry Performance (DAP): For Panoramic mode, dose to be in line with the predicate device. For Cephalometric mode, DAP measurements to be the same as the predicate device under identical conditions. For CBCT mode, similar performance to the predicate device considering FOV differences. | Panoramic Mode: "The mA setting for the subjective device was increased to be in line with the DAP of the predicate device in the Normal Panoramic mode." CEPH Mode: "The CEPH mode for the subject device and the predicate device has the same FDD... the same DAP measurement under the same X-ray exposure conditions." CBCT Mode: "the outcome result confirmed that the CBCT mode for both devices performed similarly." |
| Compliance with International Standards: Meeting requirements of 21 CFR Part 1020.30, 1020.33, IEC 61223-3-5,IEC 60601-1:2005+AMD1:2012(Edition 3.1), IEC 60601-1-3:2008+AMD1:2013 (Edition 2.1), IEC 60601-2-63:2012+AMD1:2017 (Edition 1.1), IEC 60601-1-2:2014 (Edition 4), NEMA PS 3.1-3.18. | "The acceptance test was performed according to the requirements of 21 CFR Part 1020.30. 1020.33 and IEC 61223-3-5..." "Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1:2005+AMD1:2012(Edition 3.1), IEC 60601-1-3:2008+AMD1:2013 (Edition 2.1), IEC 60601-2-63:2012+AMD1:2017 (Edition 1.1) were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2:2014 (Edition 4)." "The vatech A9 (Model : PHT-30CSS) conforms to the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set." |
| Software: "Moderate" level of concern, with existing cleared viewing programs. | "Software verification and validation were conducted and documented as recommended by FDA's Guidance... The software for this device was considered as a 'moderate' level of concern..." "vatech A9 (Model: PHT-30CSS) provides the following imaging viewer programs; -2D Image viewing program: EzDent-i(K202116) -3D Image viewing program: Ez3D-i(K200178)" |
| X-Ray Source (D-054SB): Specifications (max rating, emission & filament characteristics) equivalent to the predicate device's D-052SB. | "The specification for both D-054SB and D-052SB x-ray source (tube) is the same as confirmed by the maximum rating charts, emission & filament characteristics." |
| Detector (Xmaru1404CF-PLUS): Previously cleared. | "The subject device is equipped with the Xmaru1404CF-PLUS detector which has been cleared with previous 510k submissions, PCH-30CS (K170731)." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the numerical sample size (number of images or cases) used for the performance testing or image quality evaluations. It mentions that "the same test protocol was used to test the performance of the subject and the predicate device for comparison."
- Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given that the testing involved comparing the subject device with a predicate device and was conducted in a laboratory, it appears to be bench testing/non-clinical performance testing rather than testing on patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document mentions that "PANO/CEPH/CBCT images from the subject and predicate device are evaluated in the Image Quality Evaluation Report." However, it does not specify the number of experts who performed this evaluation, nor does it provide their qualifications (e.g., "radiologist with 10 years of experience").
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for the image quality evaluation or performance testing. It simply states the images "are evaluated."
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described or indicated. The evaluation mentioned is an "Image Quality Evaluation Report" comparing the subject and predicate device, but it doesn't detail a study involving multiple human readers to assess improvement with or without AI assistance. The device is an X-ray imaging system, not an AI diagnostic aid.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This question is not directly applicable in the typical sense for this device. The Vatech A9 is itself an imaging device, an X-ray system, not an AI algorithm that generates a diagnosis or interpretation in a standalone manner. Its performance (image quality, dose) is evaluated as a standalone system. The software components (viewing programs EzDent-i and Ez3D-i) are also cleared, indicating their standalone functionality in displaying images.
7. The Type of Ground Truth Used
The "ground truth" for the performance evaluation appears to be based on:
- Physical Measurements and Standards: For Contrast, Noise, CNR, MTF, and Dosimetry (DAP), these are objective physical measurements taken with phantoms or test protocols.
- Comparison to a Predicate Device: The performance of the subject device was directly compared to the performance of the legally marketed predicate device (Green16/Green18, K170066) using the "same test protocol."
- Expert Evaluation: For the "Image Quality Evaluation Report," the "ground truth" implicitly relies on expert subjective assessment of the images, although details are missing.
8. The Sample Size for the Training Set
The document does not describe a training set. This is because the device is a medical imaging hardware system (CT X-ray system), not an AI algorithm that requires a training set for machine learning. The viewing software (EzDent-i, Ez3D-i) is separate and was cleared through previous 510(k) submissions.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no mention of a training set for an AI algorithm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "ADMINISTRATION" in blue text. The word is in all caps and is written in a sans-serif font. To the left of the word is a blue square. The background of the image is white.
January 25, 2021
Vatech Co., Ltd. % Dave Kim Medical Device Regulatory Affairs Mtech Group 7707 Fannin St., Ste. 200-V111 HOUSTON TX 77054
Re: K203797
Trade/Device Name: vatech A9 (Model: PHT-30CSS) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: December 18, 2020 Received: December 28, 2020
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K203797
Device Name vatech A9(Model : PHT-30CSS)
Indications for Use (Describe)
vatech A9 (PHT-30CSS) is intended to produce panoramic, cone beam computed tomography, or cephalometric digital xray images. It provides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
Type of Use (Select one or both, as applicable):
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K203797
1. Special 510(k) Summary
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
2. Date 510K Summary prepared: January 11, 2021
3. Administrative Information
| Official Correspondent: | Dave Kim / Mtech GroupAddress: 7707 Fannin St. Ste 200, V-111, Houston, TX 77054Tel: +713-467-2607Contact person: Mr. Dave Kim (davekim@mtech-inc.net) |
|---|---|
| 510(k) Submitter: | VATECH Co., Ltd.Address: 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, 18449, KoreaTel: +82-31-379-9492Fax: +82-31-379-9400Contact person: Daniel Kim / Manager (daniel.kim@vatech.co.kr) |
4. Device Information
| Type of 510(k) Submission: | Special |
|---|---|
| Trade or Proprietary Name: | vatech A9(Model : PHT-30CSS) |
| Common or Usual Name: | System, X-ray, Computed tomography, Dental |
| Regulation Classification: | Computed tomography x-ray system(21 CFR 892.1750) |
| Product Code: | OAS |
| Class of Device: | Class II |
| Panel: | Radiology |
5. Predicate Device Information
| Manufacturer: | VATECH Co., Ltd. |
|---|---|
| Trade or Proprietary Name: | Green16/Green18 (Model : PHT-65LHS) |
| Common or Usual Name: | System, X-ray, Computed tomography, Dental |
| Regulation Classification: | Computed tomography x-ray system(21 CFR 892.1750) |
| Product Code: | OAS |
| Class of Device: | Class II |
| Panel: | Radiology |
| 510(k) Number: | K170066 |
{4}------------------------------------------------
6. Device Description
vatech A9 (Model : PHT-30CSS) is an advanced 3-in-1 digital X-ray imaging system that incorporates PANO, CEPH (Optional), and CBCT scan imaging capabilities into a single system.
vatech A9 (Model : PHT-30CSS), a digital radiography imaging system, is specially designed to take X-ray images of patients on the chair and assist dentists.
Designed explicitly for dental radiography, vatech A9 (Model : PHT-30CSS) is a complete digital X-ray system equipped with imaging viewers, an X-ray generator, and a dedicated SSXI detector.
The digital CBCT system is based on a CMOS digital X-ray detector. The CMOS CT detector is used to capture 3D radiographic images of the head, neck, oral surgery, implant, and orthodontic treatment.
The materials and safety characteristics of the subject device are similar to the predicate device (PHT-65LHS(K170066)) and the reference device (PCH-30CS(K170731)).
The subject device has the same modalities and identical indications for use compared to the predicate device. In addition, the subject device uses the same image reconstruction algorithm and has similar FOV and DAP performance in CBCT mode. The subject device and the reference device use the same pano and ceph detector. Both devices have similar DAP measurements.
7. Indication for use
vatech A9 (Model : PHT-30CSS) is intended to produce panoramic, cone beam computed tomography, or cephalometric digital x-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
{5}------------------------------------------------
VATECH Co., Ltd.
Premarket Notification 510(k)=
8. Substantial Equivalence Chart
| Subject Device | Predicate Device | Reference Device | |
|---|---|---|---|
| vatech A9 (Model : PHT-30CSS) | Green16/Green18 (Model : PHT-65LHS) | PaX-i Plus / PaX-i Insight(Model : PCH-30CS) | |
| Device Name | |||
| Applicant Name | VATECH Co., Ltd. | VATECH Co., Ltd. | VATECH Co., Ltd. |
| 510(k) Number | N/A | K170066 | K170731 |
| Device Classification Name | X-Ray, Tomography, Computed, Dental | X-Ray, Tomography, Computed, Dental | System, X-Ray, Extra oral Source, Digital |
| Classification Product Code | OAS | OAS | MUH |
| Regulation Number | 21 CFR 892.1750 | 21 CFR 892.1750 | 21 CFR 872.1800 |
| Indications for Use | vatech A9 (Model : PHT-30CSS) isintended to produce panoramic, conebeam computed tomography, orcephalometric digital x-ray images. Itprovides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult andpediatric patients. The system also utilizescarpal images for orthodontic treatment.The device is to be operated byhealthcare professionals. | PHT-65LHS is intended to producepanoramic, cephalometric or 3D digital x-ray images. It provides diagnostic detailsof the dento-maxillofacial, ENT, sinus andTMJ for adult and pediatric patients. Thesystem also utilizes carpal images fororthodontic treatment. The device is tobe operated by healthcare professionals. | PCH-30CS is intended to producepanoramic or cephalometric digital x-rayimages. It provides diagnostic details ofthe dento-maxillofacial, sinus and TMJ foradult and pediatric patients. The systemalso utilizes carpal images for orthodontictreatment. The device is to be operatedby physicians, dentists, and x-raytechnicians. |
| Performance Specification | Panoramic, Cephalometric and computedtomography | Panoramic, Cephalometric and computedtomography | Panoramic, Cephalometric |
| Input Voltage | AC 100 - 240 V | AC 100 - 240 V | AC 100 - 240 V |
| X-Ray source | D-054SB | D-052SB | D-052SB |
| Tube Voltage | 60 - 99 kV | 60 - 99 kV | 60 - 99 kV |
| Tube Current | 4 - 10 mA (for 60 - 99 kV)4- 12 mA (for 60 - 80 kV) | 4 - 16 mA | 4 - 10 mA |
| Focal Spot Size | 0.5 x 0.5 mm | 0.5 x 0.5 mm | 0.5 x 0.5 mm |
| Exposure Time | Max. 15.5 s | Max. 13.5 s | Max. 21 s |
| Slice Width | Min. 0.1 mm | Min. 0.1 mm | Min. 0.1 mm |
| Total Filtration | Min. 2.5 mm Al | Min. 2.5 mm Al | Min. 2.5 mm Al |
| Mechanical | Compact design | Compact design | Compact design |
| Electrical | LDCP logic circuit | LDCP logic circuit | LDCP logic circuit |
| Software | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible |
| 2D Image Viewing Program | EzDent-i (K202116) | EzDent-i (K202116) | EasyDent(Cleared under K122155)EzDent-i (K202116) |
| 3D Image Viewing Program | Ez3D-i (K200178) | Ez3D-i (K200178) | - |
| Anatomical Sites | Maxillofacial | Maxillofacial | Maxillofacial |
{6}------------------------------------------------
VATECH Co., Ltd.
| Premarket Notification 510(k) | Subject Device | Predicate Device | Reference Device | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| ImageReceptor | CT&PANO | Xmaru1404CF-PLUS | Xmaru1515CF | Xmaru1404CF-PLUS (PANO) | ||||||
| Xmaru1314CF | Xmaru1501CF-PLUS (PANO) | |||||||||
| CEPH | Xmaru2602CF | Xmaru2602CF | Xmaru2602CF | |||||||
| Size of Imaging Volume | Xmaru1404CF-PLUS | Max. 80 x 80 mm | Xmaru1515CF | Max. 180 x 100 mm | ||||||
| Xmaru1314CF | Max. 160 x 90 mm | |||||||||
| PixelResolution | CT&PANO | Xmaru1404CF-PLUS | 5 lp/mm -2x2 binning(detector spec) | Xmaru1515CF | 5 lp/mm -2x2binning (detectorspec)2.5 lp/mm -4x4binning (systemspec) | Xmaru1501CF-PLUS | 5 lp/mm | |||
| CEPH | 5 lp/mm-Non binning (detector spec)2.5 lp/mm -2x2 binning (system spec) | 5 lp/mm-Non binning (detector spec)2.5 lp/mm -2x2 binning (system spec) | 5 lp/mm-Non binning (detector spec)2.5 lp/mm -2x2 binning (system spec) | |||||||
| Pixel Size | CT&PANO | Xmaru1404CF-PLUS | 99 $ μ $ m -2X2 binning(detector spec)198 $ μ $ m - 4X4 binning(system spec) | Xmaru1515CF | 99 $ μ $ m -2X2 binning(detector spec)198 $ μ $ m - 4X4binning (systemspec) | Xmaru1501CF-PLUS | 100 $ μ $ m | |||
| CEPH | 100 $ μ $ m- Non binning (detector spec)200 $ μ $ m -2X2 binning (system spec) | 100 $ μ $ m- Non binning (detector spec)200 $ μ $ m -2X2 binning (system spec) | 100 $ μ $ m- Non binning (detector spec)200 $ μ $ m -2X2 binning (system spec) | |||||||
| Dimension(Length xWidth x Height)With Base | With CEPH | 1938.9 (L) x 1173.3 (W) x 1716.0 (H) (mm) | 1488.7 (L) x 1874.1 (W) x 2335.5 (H) (mm) | 1910 (L) x 1200 (W) x 2300 (H) (mm) | ||||||
| Without CEPH | 983.3 (L) x 1173.3 (W) x 1716.0 (H) (mm) | 1488.7 (L) x1125 (W) x2335.5 (H) (mm) | 990 (L) x 1200 (W) x 2300 (H) (mm) | |||||||
| Weight(With Base) | With CEPH | 130 kg (286.6 lbs.) | 187 kg (412.3 lbs.) | 135 kg (297.6 lbs.) | ||||||
| Without CEPH | 156 kg (343.9 lbs.) | 212 kg (467.4 lbs.) | 160 kg (352.7 lbs.) |
{7}------------------------------------------------
9. Performance Data
- Summary of Performance Testing
The vatech A9 (Model : PHT-30CSS) digital X-ray system described in this 510(k) is similar to the predicate device in terms of indications for use, materials and safety characteristics.
The following information further substantial equivalence between the subject device and the predicate device : The fundamental technological characteristics of the subject and predicate device are similar. The imaging modes are identical; CBCT, PANO and CEPH (Optional). All viewing software programs have been cleared with previous 510k submissions; EzDent-i (K202116) and Ez3D-i (K200178).
The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the performance of the subject and the predicate device for comparison. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device.
For both devices, the differences are as follows.
- a. The subject device is equipped with the Xmaru1404CF-PLUS detector which has been cleared with previous 510k submissions, PCH-30CS (K170731).
- b. The specification for both D-054SB and D-052SB x-ray source (tube) is the same as confirmed by the maximum rating charts, emission & filament characteristics. The model name, D-054SB, is designated to indicate its NMPA approval according to the company's business strategic plan.
- ﻥ The physical appearance of the subject device is different in comparison with the predicate device. Both the height and the weight of the subject device have been reduced by about 600 mm and about 50 kg, respectively.
The acceptance test was performed according to the requirements of 21 CFR Part 1020.30. 1020.33 and IEC 61223-3-5, international performance standard for computed tomography X-ray system. Contrast, Noise, CNR, and MTF, the representative indicators for CT image quality were measured with CS(iterative) reconstruction algorithm for the new X-ray equipment. The results demonstrated that the general image quality of the subject device is equivalent or better than the predicate device.
In addition, the dosimetry performance of the subject device, reference device and the predicate device was compared in terms of DAP. The subject device's Panoramic mode uses the same detector and exposure area compared to the reference device, but there is a difference in FDD (Focal Spot to Detector Distance). The FDD of the reference device (PCH-30CS) is 490.3 mm whereas the FDD for PHT-30CSS is 584.6 mm. The dose decreases by about 25°30% as the FDD increases under the same exposure conditions. The mA setting for the subjective device was increased to be in line with the DAP of the predicate device in the Normal Panoramic mode.
The CEPH mode for the subject device and the predicate device has the same FDD(Focal Spot to Detector Distance), the same detector specifications, the same DAP measurement under the same X-ray exposure conditions (exposure time, tube voltage, tube current).
In CBCT mode, the direct comparison of the dosimetry performance for each mode available in the subject and predicate device is difficult due to different exposure conditions such as the FDD, exposure time. Considering these differences, the FOV 80x80mm of PHT-30CSS and the FOV 80 x 90 mm of PHT-65LHS have been compared. The FOV size of PHT-65LHS was transposed from FOV 80x90 mm to FOV 80x80 mm, the outcome result confirmed that the CBCT mode for both devices performed similarly.
{8}------------------------------------------------
Moreover, PANO/CEPH/CBCT images from the subject and predicate device are evaluated in the Image Quality Evaluation Report. The results demonstrated that the general image quality of the subject device is equivalent or better than the predicate device.
- Software Verification and Validation Testing
Software verification and validation were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator.
vatech A9 (Model: PHT-30CSS) provides the following imaging viewer programs;
- Safety, EMC and Performance Data
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1:2005+AMD1:2012(Edition 3.1), IEC 60601-1-3:2008+AMD1:2013 (Edition 2.1), IEC 60601-2-63:2012+AMD1:2017 (Edition 1.1) were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2:2014 (Edition 4). The manufacturing facility is in conformance with the relevant EPRC standards as specified in 21 CFR 1020.30, 31, and 33 and the records are available for review. The vatech A9 (Model : PHT-30CSS) conforms to the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.
Non-clinical consideration report according to FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" was provided. Bench testing according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, Performance Testing – Bench' were performed. Acceptance test and Image evaluation report according to IEC 61223-3-4 and IEC 61223-3-5 were also performed.
All test results were satisfactory.
10. Conclusions
Safety and effectiveness of the hardware modifications and additional software functions have been clarified through each verification. Each software function has already been incorporated in previous devices which obtained premarket clearance by FDA and widely used in the market. Additional performance testing has been conducted for the subject device and all testing outcomes met the acceptance performance criteria according to the international standard and U.S. code, and proved to be equivalent to the predicate device.
The subject device and the predicate device have similar indications for use and demonstrated similar technical characteristics. As demonstrated in the performance test, 1) Xmaru1404CF-PLUS was previously cleared by the reference device, PCH-30CS (K170731), 2) The specification of D-054SB tube has the same maximum rating charts, emission & filament characteristics as D-052SB of the predicate device. Furthermore, image quality of new X-ray detector and X-ray source has been evaluated in compliance with IEC 61223-3-4 and IEC 61223-3-5. Both standard requirements were satisfied.
Quality assurance procedures are adhered to, and the specifications and functional requirements have been verified.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, VATECH Co., Ltd. concludes that vatech A9(Model: PHT-30CSS) is
{9}------------------------------------------------
substantially equivalent to the predicate device as described herein.
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.