(218 days)
The Straumann Cassettes are used to organize, sterilize and transport surgical instruments and auxiliaries between and during surgical uses. They are indicated to be used in healthcare facilities by healthcare professionals.
The Straumann Cassettes are to be enclosed in FDA cleared sterilization pouches in two layers to maintain the sterility of the enclosed devices using the following sterilization parameters: pre-vacuum steam exposure at 132°C (270° F) for 4 minutes, 30 minutes drying time.
The Straumann Cassettes have been validated for a maximum load of 1005 grams for the Surgical Cassettes, 2007 grams for the Osteotome Cassettes, 283 grams for the Bone Block Fixation Cassette and 65 grams for the Screw Container, including cassette and instruments.
Straumann® Surgical Cassettes are designed to hold various dental surgical drills and tools in order to organize, sterilize and protect the instruments that are sterilized by the healthcare provider. The subject devices consist of reusable rigid containers, comprising of a case bottom (base), a removable inner tray base (tray), and tray lid (lid). The case base, tray base, and tray lid are designed to be integrated into a single unit which contains and protects the interior components during sterilization and subsequent storage. The tray lid latches to the case bottom is designed to secure the kit into a single unit. The kits are perforated to allow for penetration of the moist heat (steam) sterilant and require the use of an FDA-cleared wrap to maintain sterility. To facilitate the surgical procedure and the correct use and positioning of the instruments, the trays and modules have instrument pictograms and/ or color-coded workflows printed on the surface. Surgical and prosthetic instruments of the Straumann® Dental Implant System intended to be placed in the Straumann® Surgical Cassettes are used for bed preparation, placement, maintenance and explanation of the implants from Straumann® Dental Implant System. These devices are all Class I exempt as described in 21 CFR 872.3980 (Endosseous dental implant accessories) and are not subject devices of this submission.
In addition, the Straumann Screw Container is included in this submission. The Straumann Screw Container is used for storing and sterilizing closure screws and healing caps. All the closure screws and healing caps that may be contained in the Screw Container are cleared Class II devices according to 21 CFR 872.3630 (Endosseous dental implant abutment) and are not subject devices of this submission.
The provided document describes the acceptance criteria and the performance testing for the Straumann® Surgical Cassettes, which are used to organize, sterilize, and transport surgical instruments.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
| Type of Testing | Purpose (Acceptance Criteria) | Reported Device Performance |
|---|---|---|
| Manual Cleaning Validation (AAMI TIR 30:2011) | Evaluate and validate the manual cleaning of the subject devices: • No visible soil • Amount of protein per sample <100 µg • Amount of protein/cm² <3.0 µg/cm² • Amount of hemoglobin/cm² <2.2 µg/cm² | Pass |
| Automated Cleaning Validation (AAMI TIR 30:2011) | Evaluate and validate the automated cleaning of the subject devices: • No visible soil • Amount of protein per sample <100 µg • Amount of protein/cm² <3.0 µg/cm² • Amount of hemoglobin/cm² <2.2 µg/cm² | Pass |
| Sterilization Validation (ISO 17665-1:2006/(R)2013) | Validate a sterilization cycle and drying time of the subject devices: • SAL ≤10⁻⁶ using the biological indicator (BI) overkill method | Pass |
| Reprocessing of Medical Devices in a Health Care Setting (Life Cycle) | Life Cycle (simulated use) testing after 100 cleaning and sterilization cycles: • Test samples must withstand 100 cycles of use (cleaning, sterilization, and functional tests) without compromising function | Pass |
| Biocompatibility (ISO 10993-5) | Cytotoxicity Testing: • Viability of cultures treated with test extracts >70% | Pass |
2. Sample size used for the test set and the data provenance:
- The document states: "The worst-case cassettes were tested for performance based on critical impact factors including weight, vent to volume ratio and worst-case instruments to be contained."
- It also mentions "Test samples must withstand 100 cycles of use". This indicates that at least one "worst-case" sample was subjected to repeated testing for the life cycle validation.
- Specific numerical sample sizes for each test (e.g., number of cassettes for cleaning validation, or number of sterilization runs) are not provided in the excerpt.
- Data provenance (country of origin, retrospective/prospective) is not specified in the provided text.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable to this type of device and study. The "ground truth" for these performance tests is based on established scientific and regulatory standards (e.g., ISO, AAMI standards) and objective measurements of sterilization, cleaning efficacy, and material biocompatibility, rather than expert consensus on interpretive tasks.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This information is not applicable as the evaluation of device performance in these tests is based on objective measurements against defined criteria, not on subjective interpretations requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This information is not applicable. The device is not an AI-powered diagnostic or assistive tool for human readers. It is a physical medical device (sterilization cassette).
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- This information is not applicable as the device is not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for the performance testing is based on objective measurements against established scientific and regulatory standards. For example:
- Cleaning efficacy: Measured levels of protein and hemoglobin, and visual inspection for soil, compared against quantitative thresholds.
- Sterilization: Sterility Assurance Level (SAL) of ≤10⁻⁶ using biological indicators, which is a quantitative measure of microbial inactivation.
- Biocompatibility: Cell viability percentage (>70%) compared against a quantitative threshold in cytotoxicity testing.
- Life cycle: Functional integrity after a specified number of cycles, assessed objectively.
8. The sample size for the training set:
- This information is not applicable as the device is not an AI algorithm that requires a training set. The performance testing validates the physical characteristics and reprocessing effectiveness of the cassettes.
9. How the ground truth for the training set was established:
- This information is not applicable as the device is not an AI algorithm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 29, 2021
Straumann USA, LLC Jennifer Jackson Director, Regulatory Affairs 60 Minuteman Rd Andover, Massachusetts 01810
Re: K203753
Trade/Device Name: Straumann Surgical Cassettes Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: June 25, 2021 Received: June 28, 2021
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K203753
Device Name
Straumann® Surgical Cassettes
Indications for Use (Describe)
The Straumann Cassettes are used to organize, sterilize and transport surgical instruments and auxiliaries between and during surgical uses. They are indicated to be used in healthcare facilities by healthcare professionals.
The Straumann Cassettes are to be enclosed in FDA cleared sterilization pouches in two layers to maintain the sterility of the enclosed devices using the following sterilization parameters: pre-vacuum steam exposure at 132°C (270° F) for 4 minutes, 30 minutes drying time.
The Straumann Cassettes have been validated for a maximum load of 1005 grams for the Surgical Cassettes, 2007 grams for the Osteotome Cassettes, 283 grams for the Bone Block Fixation Cassette and 65 grams for the Screw Container, including cassette and instruments.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
Residential Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Straumann® Surgical Cassettes
510(k) Summary
510(k) Summary
Submitter's Contact Information
| Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)60 Minuteman RoadAndover, MA 01810Registration No.: 1222315 Owner/Operator No.: 9005052 |
|---|---|
| Contact Person: | Jennifer M. Jackson, MSDirector of Regulatory Affairs and QualityPhone Number: +1 978 747-2509Fax Number: +1 978 747-0023 |
| Prepared By &Alternate Contact: | Chanrasmey White, MSRegulatory Affairs SpecialistStraumann, USA, LLC |
| Date of Submission: | July 29, 2021 |
Name of the Device
| Trade Names: | Straumann® Surgical Cassettes |
|---|---|
| Common Name: | Sterilization Wrap Containers, Trays, Cassettes & OtherAccessories |
| Classification Name: | Sterilization Wrap Containers, Trays, Cassettes & OtherAccessories |
| Regulation Number: | §880.6850 |
| Device Classification: | II |
| Product Code(s): | KCT |
| Classification Panel: | General Hospital |
Predicate Device(s)
Primary Predicate:
- . K180791 – Straumann BLX Surgical Cassette (Institut Straumann AG)
{4}------------------------------------------------
Straumann® Surgical Cassettes
510(k) Summary
Reference Devices:
- K191522 Straumann Modular Surgical Cassette (Institut Straumann AG) .
Device Description
Straumann® Surgical Cassettes are designed to hold various dental surgical drills and tools in order to organize, sterilize and protect the instruments that are sterilized by the healthcare provider. The subject devices consist of reusable rigid containers, comprising of a case bottom (base), a removable inner tray base (tray), and tray lid (lid). The case base, tray base, and tray lid are designed to be integrated into a single unit which contains and protects the interior components during sterilization and subsequent storage. The tray lid latches to the case bottom is designed to secure the kit into a single unit. The kits are perforated to allow for penetration of the moist heat (steam) sterilant and require the use of an FDA-cleared wrap to maintain sterility. To facilitate the surgical procedure and the correct use and positioning of the instruments, the trays and modules have instrument pictograms and/ or color-coded workflows printed on the surface. Surgical and prosthetic instruments of the Straumann® Dental Implant System intended to be placed in the Straumann® Surgical Cassettes are used for bed preparation, placement, maintenance and explanation of the implants from Straumann® Dental Implant System. These devices are all Class I exempt as described in 21 CFR 872.3980 (Endosseous dental implant accessories) and are not subject devices of this submission.
In addition, the Straumann Screw Container is included in this submission. The Straumann Screw Container is used for storing and sterilizing closure screws and healing caps. All the closure screws and healing caps that may be contained in the Screw Container are cleared Class II devices according to 21 CFR 872.3630 (Endosseous dental implant abutment) and are not subject devices of this submission.
Intended Use
The Straumann Cassettes are intended to store, organize and reprocess surgical instruments and auxiliaries of the Straumann® Dental Implant System during implant/prosthetic treatment and sterilization.
{5}------------------------------------------------
Straumann® Surgical Cassettes
510(k) Summary
Indications for Use
The Straumann Cassettes are used to organize, sterilize and transport surgical instruments and auxiliaries between and during surgical uses. They are indicated to be used in healthcare facilities by healthcare professionals.
The Straumann Cassettes are to be enclosed in FDA cleared sterilization pouches in two layers to maintain the sterility of the enclosed devices using the following sterilization parameters: prevacuum steam exposure at 132°C (270° F) for 4 minutes, 30 minutes drying time.
The Straumann Cassettes have been validated for a maximum load of 1005 grams for the Surgical Cassettes, 2007 grams for the Osteotome Cassettes, 283 grams for the Bone Block Fixation Cassette and 65 grams for the Screw Container, including cassette and instruments.
Technological Characteristic Comparison
The subject and predicate device share the following characteristics:
- Equivalent indications for use .
- . Equivalent design
- ldentical sterilization method .
- . Equivalent sterilization parameters
- . Reusable
The subject device is technologically different from the predicate device as follows:
- . Materials (the only part of the cassette made out of stainless steel is the lid hinge)
- Vent-to-Volume ratio .
- . Drying time
The technological characteristics of the subject devices are compared to the predicate devices in Table 1.
{6}------------------------------------------------
Straumann® Surgical Cassettes
510(k) Summary
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE PREDICATE DEVICE | COMPARISON |
|---|---|---|---|---|
| K Number | K203753 | K180791 | K191522 | |
| Indications forUse | The Straumann Cassettes are usedto organize, store, sterilize andtransport surgical instruments andauxiliaries between and duringsurgical uses. They are indicated tobe used in healthcare facilities byhealthcare professionals.The Straumann Cassettes are to beenclosed in FDA cleared sterilizationpouches in two layers to maintainthe sterility of the enclosed devicesusing the following sterilizationparameters: pre-vacuum steamexposure at 132°C (270° F) for 4minutes, 30 minutes drying time.The Straumann Cassettes havebeen validated for a maximum loadof 1005 grams for the SurgicalCassettes, 2007 grams for theOsteotome Cassettes, 283 gramsfor the Bone Block Fixation Cassetteand 65 grams for the ScrewContainer, including cassette andinstruments. | The Straumann BLX Cassette isused in healthcare facilities toorganize, enclose, cleaning,sterilize, transport, and storemedical devices between surgicaluses. The BLX Cassette is notintended to maintain sterility; it isintended to be used in conjunctionwith a legally marketed, validatedsterilization wrap.The BLX Cassette has beenvalidated for a maximum load of 300grams, including cassette andinstruments.Sterilization parameters:Pre-vacuum steam: 132°C (270° F)for 4 minutes with 20 minutes dryingtime. | Straumann® Modular Cassette is used inhealthcare facilities to organize, enclose,sterilize, transport and store Straumanninstruments between surgical uses.Straumann® Modular Cassette is notintended to maintain sterility on its own,but is intended to be used in conjunctionwith a legally marketed, validatedsterilization double pouch to maintain thesterility of the enclosed devices.The Straumann® Modular Cassette hasbeen validated for the followingmaximum loads:Module A 400gBCC Maximum permissible stackBCC 611gThe A module is intended to be sterilizedindividually, without stacking with othermodules.The B and C module are intended to besterilized individually, or by stacking theB module on top of C module bases. Themaximum permissible stack forsterilization is one B module on top oftwo C module bases.The B module lid and C module lid couldbe used to enclose an ultrasonic mat(Art. No. 041.774) for ultrasonic bathcleaning used instruments.Only use the following sterilizationparameters:Fractionated vacuum: 132°C (270 °F) for4 minutes with 30 minutes drying time | Similar |
| Product Code | KCT | KCT | KCT | Same |
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE PREDICATE DEVICE | COMPARISON |
| K Number | K203753 | K180791 | K191522 | |
| Design | Reusable rigid container, casebottom (base), a removable innertray base (tray), and tray lid (lid). | Plastic tray and lid | Plastic modules, trays and lids | Similar |
| Materials | Polyphenylsulfone (Radel 5000)Tecapro MT, Propylux SiliconeStainless SteelEUROPLEX® PPSU99050 GT, greyPP HS2 WHV200 WM, white | Polyphenylsulfone (Radel R5000)Silicone | PolyphenylsulfoneSiliconeStainless Steel | Similar |
| Materialscompatible withsterilizationMethod | Yes | Yes | Yes | Same |
| Perforated | Yes; allows moist heat (steam)penetration to achieve sterilization | Yes; allows moist heat (steam)penetration to achieve sterilization | Yes; allows moist heat (steam)penetration to achieve sterilization | Same |
| Reusable | Yes, up to 100x | Yes, up to 100x | Yes, up to 100x | Same |
| Sterilizationmethod | Moist heat (steam) | Moist heat (steam) | Moist heat (steam) | Same |
| Cycles | Pre-vacuum | Pre-vacuum | Fractionated vacuum | Same asK180791 |
| Parameters | Pre-Vacuum:132º C (270º F) for 4 minutes;30 minutes drying time | Pre-Vacuum:132º C (270º F) for 4 minutes;20 minutes drying time | Fractionated vacuum:132°C (270 °F) for 4 minutes;30 minutes drying time | Same asK180791 |
| Sterile barrier | FDA cleared sterilization pouch | FDA cleared sterilization pouch | FDA cleared sterilization pouch | Same |
| Biocompatibility | The Biocompatibility assessmentwas performed per ISO 10993-1 andtesting was performed usingmethods described in ISO 10993-5.The results indicate that the subjectdevices are non-cytotoxic. | The Biocompatibility assessmentwas performed per ISO 10993-1 andtesting was performed usingmethods described in ISO 10993-5.The results indicate that the subjectdevices are non-cytotoxic. | The Biocompatibility assessment wasperformed per ISO 10993-1 and testingwas performed using methods describedin ISO 10993-5. The results indicate thatthe subject devices are non-cytotoxic. | Same |
{7}------------------------------------------------
Straumann® Surgical Cassettes
510(k) Summary
Table 1 – Comparison Matrix
{8}------------------------------------------------
Traditional 510(k) Submission
Straumann® Surgical Cassettes
510(k) Summary
Performance Testing
The performance during multiple reprocessing steps for the Straumann® Surgical Cassettes, as recommended in the labeling, was validated according to applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015". The worst-case cassettes were tested for performance based on critical impact factors including weight, vent to volume ratio and worstcase instruments to be contained. The summary of testing performed is provided in Table 2.
| Type of Testing | Purpose | Acceptance Criteria | Result |
|---|---|---|---|
| Manual CleaningValidation(AAMI TIR 30:2011) | Evaluate and validatethe manual cleaning ofthe subject devices | • No visible soil• Amount of protein persample <100 µg• Amount of protein/cm²<3.0 µg/cm²• Amount ofhemoglobin/cm² <2.2µg/cm² | Pass |
| Automated CleaningValidation(AAMI TIR 30:2011) | Evaluate and validatethe automated cleaningof the subject devices | • No visible soil• Amount of protein persample <100 µg• Amount of protein/cm²<3.0 µg/cm²• Amount ofhemoglobin/cm² <2.2µg/cm² | Pass |
| Sterilization Validation(ISO 17665-1:2006/(R)2013) | Validate a sterilizationcycle and drying time ofthe subject devices | • SAL ≤10⁻⁶ using thebiological indicator(BI) overkill method | Pass |
| Reprocessing ofMedical Devices in aHealth Care Setting | Life Cycle (simulateduse) testing after 100cleaning andsterilization cycles | • Test samples mustwithstand 100 cyclesof use (cleaning,sterilization, andfunctional tests)without compromisingfunction | Pass |
{9}------------------------------------------------
Traditional 510(k) Submission
Straumann® Surgical Cassettes
510(k) Summary
| Biocompatibility(ISO 10993-5) | Cytotoxicity Testing | • Viability of culturestreated with testextracts >70% | Pass |
|---|---|---|---|
| ----------------------------------- | ---------------------- | --------------------------------------------------------------- | ------ |
Table 2 – Performance testing summary
Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission K203753, Straumann Surgical Cassettes, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K180791 and K191522.
§ 880.6850 Sterilization wrap.
(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).