Intended Use

The Nisus Touch Negative Pressure Wound Therapy Pump is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material, and tissue debris.

Device Description

Cork Medical Products has developed a negative pressure wound therapy pump with the same intended use as its predicate device (K140022) but has the addition of an integrated touch screen. The Nisus Touch and Nisus pumps are nearly identical in appearance and utilize the same buttons to power the device and adjust pump settings; however, the Nisus Touch offers the alternative touch screen to easily maneuver between settings based on operator preference. Visual and audible alarms are consistent with the predicate device K140022 and alert the user when critical battery, pressure leakage, system blockage, and full canister occurs. Additionally, the Nisus Touch mechanical components and therapy application are identical to the Nisus pump. The intent of the new model is to appeal to users who encounter touch screens in daily technology and might feel navigation by touch screen is more simplistic.

The components included within the Cork Nisus Touch NPWT Pump are:

  • Nisus Touch Negative Pressure Wound Therapy Pump
  • Nisus Touch Pump Battery Charger

Accessory components are required to operate the device. Iniection molded components are sonically welded to form a canister designed to mate with the pump and collect excess exudates, infectious material, and tissue debris. The canister design was previously provided clearance within the Nisus NPWT system 510k application (K140022).

The Wound Kit comes in multiple iterations containing a minimum of one port pad, wound foam and peel and stick drape. These components are used in coniunction with the pump to ensure safe transfer of the wound debris to the canister. The NPWT Accessories were previously provided clearance on K132004. Our application for the Nisus Touch NPWT system includes no revisions to these previously cleared components.

AI/ML Overview

This document describes the Nisus Touch Negative Pressure Wound Therapy Pump, a device that facilitates wound healing by removing exudates, infectious material, and tissue debris.

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly present a table of acceptance criteria with numerical targets. Instead, it states that the device underwent "bench performance testing to establish basic functionality" and that "The testing results show the Nisus Touch NPWT pump functioned as expected."

Based on the "NONCLINICAL TESTS" section, here are the tests conducted and the general performance reported:

Test/Criteria CategorySpecific Tests ConductedReported Performance / Conclusion
Functional PerformanceContinuous Mode Low Pressure (40-mmHg) TestFunctioned as expected
Continuous Mode Typical Pressure (125-mmHg) TestFunctioned as expected
Continuous Mode High Pressure (200-mmHg) TestFunctioned as expected
Intermittent Mode TestFunctioned as expected
Low Battery TestFunctioned as expected
Leakage Alarm TestFunctioned as expected
Blockage Alarm TestFunctioned as expected
Canister Full Alarm TestFunctioned as expected
Suction Capacity and Maximum Vacuum PressuresFunctioned as expected (while the predicate has 4 liters/minute suction and 220 mmHg max vacuum, the subject device has ~8-10 liters/minute suction and 200 mmHg max vacuum. The report implies these are acceptable).
Software Verification & ValidationSoftware V&V testing according to FDA guidance for "moderate" level of concern.Documentation provided and aligned with guidance.
Electrical Safety & EMCTesting against IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11, and IEC 60601-1-2.System complies with the respective standards.
Usability EngineeringTesting against IEC 62366:2007 (Application of usability engineering to medical devices).Not explicitly stated "complies" but listed as "Additional non-clinical testing conducted," implying meeting requirements.

2. Sample Size Used for the Test Set and Data Provenance:

The document refers to "bench performance testing" using "simulated wound exudate" and a "wound test bed fixture." No specific sample size (e.g., number of devices tested, number of simulated scenarios) is mentioned for this testing. The data provenance is from non-clinical bench testing, not from human subjects or clinical data. Therefore, there is no country of origin or retrospective/prospective classification in the typical sense.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

Not applicable. The ground truth for this device's performance was established through engineering specifications and bench testing against those specifications. No human expert consensus was used to establish ground truth for this type of performance testing.

4. Adjudication Method for the Test Set:

Not applicable. As the testing was bench performance testing against engineering specifications, there was no need for an adjudication method by human experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical testing required to support this 510(k) submission. No clinical testing has been performed."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

Given that this is a physical medical device (a negative pressure wound therapy pump) and not an AI/software algorithm intended for diagnostic or prognostic purposes, the concept of "standalone (algorithm only)" performance does not directly apply. The performance evaluated was the device's physical and software-controlled functionality.

7. The Type of Ground Truth Used:

The ground truth used for these non-clinical tests was based on engineering specifications and expected functional behavior of a negative pressure wound therapy pump, including adherence to established electrical safety, EMC, and usability standards. For example, alarms for leakage or blockage are expected to activate under specific simulated conditions according to design specifications.

8. The Sample Size for the Training Set:

Not applicable. This device is a physical medical device with integrated software for control, not a machine learning or AI algorithm that requires a "training set" in the context of learning data. The software within the device operates based on pre-programmed logic.

9. How the Ground Truth for the Training Set was Established:

Not applicable, as there is no "training set" for a machine learning algorithm. The software functionality is based on established engineering principles and medical device requirements.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

August 13, 2021

Cork Medical Products, LLC Patrick McGinley Chief Executive Officer 8000 Castleway Drive Indianapolis, Indiana 46250

Re: K203693

Trade/Device Name: Nisus Touch Negative Pressure Wound Therapy Pump Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: June 16, 2021 Received: June 21, 2021

Dear Patrick McGinley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (0S) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203693

Device Name

Nisus Touch Negative Pressure Wound Therapy Pump

Indications for Use (Describe)

The Nisus Touch Negative Pressure Wound Therapy Pump is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material, and tissue debris.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

807.92(c)

SPONSOR

807.92(a)(1)

Company Name:Cork Medical Products LLC
Company Address:8000 Castleway DriveIndianapolis, IN 46250

Telephone: 866-551-2580 Fax: 844-269-8439 Contact Person: Patrick McGinley Date Prepared: August 6th, 2021

DEVICE NAME

807.92(a)(2)

Trade Name:Nisus Touch Negative Pressure Wound Therapy Pump
System Common/Usual Name:Negative Pressure Wound Therapy Powered Suction Pump
Classification Name:Powered Suction Pump
Regulation Number:21 CFR 878.4780
Product Code:OMP
Device Class:Class II

PREDICATE DEVICE

807.92(a)(3)

CompanyBrand Name510(k) Number
Cork Medical ProductsNisus Negative Pressure Wound Therapy SystemK140022

DEVICE DESCRIPTION

807.92(a)(4)

Cork Medical Products has developed a negative pressure wound therapy pump with the same intended use as its predicate device (K140022) but has the addition of an integrated touch screen. The Nisus Touch and Nisus pumps are nearly identical in appearance and utilize the same buttons to power the device and adjust pump settings; however, the Nisus Touch offers the alternative touch screen to easily maneuver between settings based on operator preference. Visual and audible alarms are consistent with the predicate device K140022 and alert the user when critical battery, pressure leakage, system blockage, and full canister occurs. Additionally, the Nisus Touch mechanical components and therapy application are identical to the Nisus pump. The intent of the new model is to appeal to users who encounter touch screens in daily technology and might feel navigation by touch screen is more simplistic.

The components included within the Cork Nisus Touch NPWT Pump are:

  • Nisus Touch Negative Pressure Wound Therapy Pump
  • Nisus Touch Pump Battery Charger

Accessory components are required to operate the device. Iniection molded components are sonically welded to form a canister designed to mate with the pump and collect excess exudates, infectious material, and tissue debris. The canister design was previously provided clearance within the Nisus NPWT system 510k application (K140022).

{4}------------------------------------------------

The Wound Kit comes in multiple iterations containing a minimum of one port pad, wound foam and peel and stick drape. These components are used in coniunction with the pump to ensure safe transfer of the wound debris to the canister. The NPWT Accessories were previously provided clearance on K132004. Our application for the Nisus Touch NPWT system includes no revisions to these previously cleared components.

DEALCE TATEMDED OSE

807.92(a)(5)

Indications for Use

The Nisus Touch Neqative Pressure Wound Therapy Pump is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material, and tissue debris.

Physician Orders

Caution: Federal Law (USA) restricts this device to sale by or on the order of a licensed physician.

Use of the Nisus Touch NPWT Pump must be prescribed by a physician per the stated indications for use. As a condition of use, the Nisus Touch NPWT Pump should only be used by qualified and authorized personnel.

The user must have the necessary knowledge of the specific medical application for which negative pressure wound therapy is being used.

Prior to placement of the Nisus Touch NPWT Pump, the medical professional treating the wound must assess how to best use the system for an individual wound. It is important to carefully assess the wound and patient to ensure clinical indications for negative pressure wound therapy are met.

User

The Nisus Touch NPWT Pump is designed for use by licensed healthcare professionals only. The keypad and device menus are locked by the healthcare professional to prevent patient from changing the setting prescribed by the physician.

NOTE: Patient functions are limited to power on/off and respond to any alarm conditions.

Use Environment

Cork Medical Products Nisus Touch NPWT Pump is designed for the following environmental conditions:

Operating Temperature: 18°C to 34°C (65°F to 94°F) Operating Relative Humidity: 10% - 95% Operating Pressure: 700-hPA - 1060-hPA (10.15-atm - 15.37-atm) atmospheric pressure

{5}------------------------------------------------

PREDICATE PRODUCT COMPARISON TABLE

807.92(a)(6)

Subject DevicePredicate Device
CompanyCork Medical ProductsCork Medical Products
Device NameNisus Touch Negative PressureWound Therapy PumpNisus Negative PressureWound Therapy System
510(k) NumberK203693K140022
Regulation Number /Product Code21 CFR 878.4780 / OMP21 CFR 878.4780 / OMP
Indications for UseThe Nisus Touch NegativePressure Wound TherapyPump is indicated for use inpatients who would benefit fromnegative pressure woundtherapy particularly as thedevice may promote woundhealing by the removal ofexcess exudates, infectiousmaterial, and tissue debris.The Cork Medical ProductsNisus Negative PressureWound Therapy System isindicated for use in patientswho would benefit fromnegative pressure woundtherapy particularly as thedevice may promote woundhealing by the removal ofexcess exudates, infectiousmaterial, and tissue debris.
Features• Multiple keys for navigationincluding: power, menu/select,exit, up arrow, down arrow, leftarrow, right arrow• Touch screen capability tomaneuver between therapymodes and device settings• Multiple keys for navigationincluding: power, menu/select,exit, up arrow, down arrow, leftarrow, right arrow
Suction Capacity~8-10 liters / minute4 liters / minute
Maximum VacuumPressure200-mmHg220-mmHg
Power Requirements18 VDC, 2A18 VDC, 2A
Battery TypeLi-ionLi-ion
Dimensions151 x 117 x 84-mm(~6 x 4.6 x 3.3-inches)151 x 108 x 71-mm(~6 x 4.3 x 2.8-inches)
Weight0.616-kg (~1.36-lb)0.575-kg (~1.27-lb)
ReusableYesYes
SterileNon-sterileNon-sterile
ComplianceIEC 60601-1, 4th Edition (AAMIES 60601-1, CAN/CSA C22.2No. 60601-1-08, EN 60601-1)IEC 60601-1-2IEC 60601-1-6/IEC 62366IEC 60601-1-11IEC 60601-1, 3rd Edition (AAMIES 60601-1, CAN/CSA C22.2No. 60601-1-08, EN 60601-1)IEC 60601-1-2IEC 60601-1-6/IEC 62366IEC 60601-1-11
Storage & ShippingConditions-25°C (-13°F) without relativehumidity control to 70°C (158°F)up to 93% relative humidity(non-condensing)-25°C (-13°F) without relativehumidity control to 70°C (158°F)up to 93% relative humidity(non-condensing)
Environmental ConditionsOperating Temperature: 18°C -34°C (65°F - 94°F)Operating Relative Humidity:10% - 95% non-condensingOperating Pressure: 700-hPA –1060-hPA (10.15-atm - 15.37-atm) atmospheric pressureOperating Temperature: 5°C -40°C (41°F - 104°F)Operating Relative Humidity:15% - 93% non-condensingOperating Pressure: 700-hPA –1060-hPA (10.15-atm - 15.37-atm) atmospheric pressure
AccessoriesCanisters:Two Sizes: 250mL, 500mLFeatures: hydrophobicmembrane filter, liquid solidifier(cleared on K14022)Wound Dressing Kit:Wound FoamWound DrapePort Pad AssemblyPreviously cleared on K132004Provided sterileBench Testing completedutilized accessories previouslycleared on separate 510kapplications to mimic thepredicate device setting.Canisters:Two Sizes: 250mL, 500mLFeatures: hydrophobicmembrane filter, liquid solidifier(cleared with application)Wound Dressing Kit:Wound FoamWound DrapePort Pad AssemblyPreviously cleared on K132004Provided sterileBench Testing utilized canistersin application, and accessoriescleared on separate application.

{6}------------------------------------------------

NONCLINICAL TESTS

807.92(b)(1)

The Cork Medical Products Nisus Touch Negative Pressure Wound Therapy Pump underwent bench performance testing to establish basic functionality. The bench performance tests conducted are:

  • . Continuous Mode Low Pressure (40-mmHg) Test
  • . Continuous Mode Typical Pressure (125-mmHg) Test
  • . Continuous Mode High Pressure (200-mmHg) Test
  • . Intermittent Mode Test
  • . Low Battery Test
  • Leakage Alarm Test .
  • Blockage Alarm Test .
  • . Canister Full Alarm Test
  • . Suction Capacity and Maximum Vacuum Pressures

The testing results show the Nisus Touch NPWT pump functioned as expected. Performance tests used simulated wound exudate. Pressure measurements were taken using a wound test bed fixture.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. The corresponding software documentation required for a 510(k) submission is included per "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" as published by the FDA.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Nisus Touch Negative Pressure Wound Therapy Pump. The system complies with the IEC 60601-1: 2005, IEC 60601-1-6: 2010, and IEC 60601-1-11: 2010 standards for safety and the IEC 60601-1-2: ed 4.0 (2014-02) standard for EMC.

Additional non-clinical testing conducted

62366: 2007 - Medical devices - Application of usability engineering to medical devices. IEC

CLINICAL TESTS

No clinical testing required to support this 510(k) submission. No clinical testing has been performed.

CONCLUSIONS

Cork Medical concludes, based on nonclinical testing, that the Nisus Touch Negative Pressure Wound Therapy Pump is as safe, as effective, and performs as well as the predicate device, Cork Medical Products Nisus Neqative Pressure Wound Therapy System (K140022).

807.92(b)(3)

807.92(b)(2)

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.