K Number
K203670
Manufacturer
Date Cleared
2021-02-16

(62 days)

Product Code
Regulation Number
870.1340
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Extra Large Check-Flo introducers and guiding sheaths are intended to introduce therapeutic devices into the vasculature.

Device Description

Cook Incorporated's Extra Large Check-Flo® Introducer is a single-use, sterile, disposable product that is used to introduce therapeutic or diagnostic devices into the vasculature. The Extra Large Check-Flo® Introducer consists of an introducer sheath, Check-Flo® hemostatic valve assembly, and dilator. The sheath is constructed with a hemostatic valve that accepts a large range of French sizes while preventing blood reflux and air aspiration. The subject device, the Extra Large Check-Flo® Introducer, is manufactured in diameters of 20.0, 22.0 and 24.0 French with sheath lengths of 25, 40, or 65 centimeters. The dilator is manufactured in lengths 16 centimeters longer than the sheath (41, 56, and 81 centimeters, respectively). The diameter of the distal endhole of the dilator measures 0.035 inch. The introducer sheath and dilator are matched as a unit to provide a smooth transition.

AI/ML Overview

Based on the provided text, the document is a 510(k) Premarket Notification for a medical device called the "Extra Large Check-Flo® Introducer." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials as would be required for a PMA (Premarket Approval) application.

Therefore, the information typically requested in your prompt regarding acceptance criteria for AI/ML device performance (e.g., sample sizes for test/training sets, expert adjudication, MRMC studies, standalone performance, ground truth establishment) is not applicable to this document. This document describes the acceptance criteria and study that proves a physical medical device (a catheter introducer) meets performance specifications, not an AI/ML-based diagnostic or therapeutic device.

However, I can extract the acceptance criteria and the performance results for the physical device as described in the provided text.

Here's the relevant information from the document:

1. A table of acceptance criteria and the reported device performance:

The document lists various tests performed and states that the "Test results met predetermined criteria" or "The predetermined acceptance criterion was met." Specific numerical acceptance criteria values are generally not provided in this summary, but the success of meeting them is stated.

Test PerformedAcceptance Criteria (as described)Reported Device Performance (as described)
BiocompatibilityPer ANSI AAMI ISO 10993-1All materials are biocompatible.
Tensile TestingPer BS EN ISO 11070Test results met predetermined criteria.
Compatibility TestingPer BS EN ISO 11070Test results met predetermined criteria.
Integrity TestingSheath tip no visible damage (cracks, splits, rollback); Sheath patent and unwrinkled upon dilator withdrawalTest results met predetermined criteria.
Rollback TestingPer BS EN ISO 11070, Annex AThe predetermined acceptance criterion was met.
Radiopacity EvaluationPer ASTM F640-12Test results met predetermined criteria.
Dilator Removal Force TestingMet specification and comparable to reference device (K171866)The force to remove the dilator from the sheath met the specification and was comparable to the reference device.
Dimensional TestingEffective dimensions derived from manufacturing specificationsEffective product dimensions derived from the manufacturing specifications were confirmed.
Simulated Use TestingReliability in a clinical simulated use scenarioSubject device tested through an anatomical model to test the subject device's reliability in a clinical simulated use scenario. (Implied successful performance)
Valve LeakagePer BS EN ISO 11070Tested in accordance with the acceptance criterion established in BS EN ISO 11070. (Implied successful performance)

Regarding the other points of your prompt (2-9), these are not applicable to the information provided in this 510(k) summary for a physical medical device:

  • 2. Sample sized used for the test set and the data provenance: Not applicable. This refers to a physical device, not an AI/ML model with test data. The number of physical units tested per criterion is not specified.
  • 3. Number of experts used to establish the ground truth... and the qualifications of those experts: Not applicable. Ground truth as typically understood in AI/ML (e.g., expert labels on images) is not relevant here. Device performance is measured by physical tests according to standards.
  • 4. Adjudication method: Not applicable. Device performance is determined by adherence to engineering specifications and international standards, not expert adjudication of diagnostic outcomes.
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done...: Not applicable. This is for AI/ML diagnostic or assistive devices, not for a catheter introducer.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. No algorithm involved.
  • 7. The type of ground truth used: Not applicable. Device performance is measured against engineering specifications and test standards (e.g., tensile strength, dimensions, biocompatibility), not clinical "ground truth" derived from patient outcomes or expert consensus in the diagnostic sense.
  • 8. The sample size for the training set: Not applicable. This is a physical device, not an AI/ML model that undergoes training.
  • 9. How the ground truth for the training set was established: Not applicable. See point 8.

In summary: The provided document is a 510(k) summary for a physical medical device. It demonstrates performance by stating that the device passed a series of engineering and biocompatibility tests in accordance with established industry standards and internal specifications, thus supporting its substantial equivalence to a predicate device. It does not contain information related to AI/ML model performance or associated clinical study designs.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Cook Incorporated Reuben Lidster Regulatory Affairs Specialist P.O. Box 489, 750 Daniels Way Bloomington, Indiana 47402

February 16, 2021

Re: K203670

Trade/Device Name: Extra Large Check-Flo Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: December 17, 2020 Received: December 18, 2020

Dear Reuben Lidster:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements,

{1}------------------------------------------------

including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH and Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See DICE (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Finn Donaldson Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203670

Device Name Extra Large Check-Flo® Introducer

Indications for Use (Describe)

Extra Large Check-Flo introducers and guiding sheaths are intended to introduce therapeutic devices into the vasculature.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Extra Large Check-Flo® Introducer

Traditional 510(k) Summary

21 CFR §807.92(c)

Date Prepared: December 15, 2020

Submitter Information:

Applicant:Cook Incorporated
Applicant Address:750 Daniels WayBloomington, IN 47404
Contact:Reuben G. Lidster
Email:RegSubmissions@cookmedical.com
Contact Phone Number:812-335-3575 x 104866 or 812-325-4172
Contact Fax Number:(812) 332-0281

Subject Device Information:

Trade Name:Extra Large Check-Flo® Introducer
Common Name:Introducer, Catheter
Classification Name:Catheter Introducer
Classification Regulation:21 CFR §870.1340, Product Code DYB
Device Class/Classification Panel:Class II, Cardiovascular

Predicate Device:

The GORE® DrySeal Sheath (K093791) was cleared for commercial distribution on March 22, 2010.

{4}------------------------------------------------

Reference Devices:

The GORE® DrySeal Flex Introducer Sheath (K160254) was cleared for commercial distribution on May 12, 2016. The reference device offers additional working lengths (33, 45 and 65 cm), compared to the predicate (28 cm only).

The Medtronic Sentrant Introducer Sheath with Hydrophilic Coating (K171866) was cleared for commercial distribution on December 20, 2017. The reference device offers comparative characterization testing for the dilator removal force.

Subject Device Description:

Cook Incorporated's Extra Large Check-Flo® Introducer is a single-use, sterile, disposable product that is used to introduce therapeutic or diagnostic devices into the vasculature. The Extra Large Check-Flo® Introducer consists of an introducer sheath, Check-Flo® hemostatic valve assembly, and dilator. The sheath is constructed with a hemostatic valve that accepts a large range of French sizes while preventing blood reflux and air aspiration. The subject device, the Extra Large Check-Flo® Introducer, is manufactured in diameters of 20.0, 22.0 and 24.0 French with sheath lengths of 25, 40, or 65 centimeters. The dilator is manufactured in lengths 16 centimeters longer than the sheath (41, 56, and 81 centimeters, respectively). The diameter of the distal endhole of the dilator measures 0.035 inch. The introducer sheath and dilator are matched as a unit to provide a smooth transition.

Intended Use:

Extra Large Check-Flo introducers and guiding sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature.

Comparison to the Predicate Device:

The Extra Large Check-Flo® Introducer, subject of this submission, is substantially equivalent to the predicate device, the GORE® DrySeal Sheath (K093791), in that these devices have the same general intended use, design, fundamental technological characteristics, and method of operation. The following table offers a comparison of different aspects of the subject and predicate device:

{5}------------------------------------------------

PREDICATE DEVICESUBJECT DEVICE
GORE DrySeal SheathExtra Large Check-Flo Introducer
RegulationNumber870.1340870.1340
Product CodeDYBDYB
ClassificationClass IIClass II
Indications forUseThe GORE® DrySeal Sheath is intended tobe inserted in the vasculature to provide aconduit for the insertion of endovasculardevices while minimizing blood lossassociated with such insertions.Extra Large Check-Flo introducers and guidingsheaths are intended to introduce therapeutic ordiagnostic devices into the vasculature.
ImageImage: GORE DrySeal SheathImage: Extra Large Check-Flo Introducer
Introducer Sheath
TubingPolyethyleneThick Wall Sheathing RadiopaqueTetrafluoroethylene (VRTS)
IntroducerSheath Size (Fr)12.0, 14.0, 16.0, 18.0, 20.0, 22.0, 24.0, and26.020.0, 22.0, and 24.0
IntroducerSheath Length(WorkingLength) (cm)2825, 40, and 65
Tip ShapeStraightIdentical
Check-Flo Assembly
ConnectingTube (Material)UnknownNon-di(2-ethylhexyl) phthalate (DEHP) PolyvinylChloride (PVC) & PVC
StopcockUnknownPolycarbonate and Polyethylene
Check-Flo BodyUnknownPolycarbonate
Check-FloValveSiliconeSilicone
Check-Flo CapUnknownWhite Polyacetal
Dilator
ShaftUnknownVinyl Radiopaque Dilator Tubing (VRDT)
HubUnknownPolyacetal
IntroducerDilator Size (Fr)Unknown20.0, 22.0, 24.0
IntroducerDilator Length(cm)Unknown41, 56, and 81
Dilator EndholeDiameter (in)0.035Identical

{6}------------------------------------------------

Test Data:

The subject device, the Extra Large Check-Flo® Introducer, was subjected to applicable testing to assure reliable design and performance under the testing parameters. The tests performed are listed below:

  • . Biocompatibility - Per ANSI AAMI ISO 10993-1, testing for cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, material-mediated pyrogenicity, and hemocompatibility were performed to ensure the biocompatibility of the subject device. Test results indicated that all materials are biocompatible.
  • . Tensile Testing - Tested in accordance with the requirements of BS EN ISO 11070. Test results met predetermined criteria.
  • . Compatibility Testing – Tested in accordance with the requirements of BS EN ISO 11070. Test results met predetermined criteria.
  • . Integrity Testing - Testing verified that, under proper clinical use of the subject device, the sheath tip exhibited no visible damage, including cracks, splits, and rollback. Upon withdrawal of the dilator, the sheath remained patent and unwrinkled. Test results met predetermined criteria.
  • . Rollback Testing - Tested in accordance with BS EN ISO 11070, Annex A. The predetermined acceptance criterion was met.
  • . Radiopacity Evaluation - Tested in accordance with ASTM F640-12. Test results met predetermined criteria.
  • . Dilator Removal Force Testing – The force to remove the dilator from the sheath met the specification and was comparable to the reference device (K171866).
  • . Dimensional Testing – Effective product dimensions derived from the manufacturing specifications were confirmed.
  • . Simulated Use Testing - Subject device tested through an anatomical model to test the subject device's reliability in a clinical simulated use scenario.
  • . Valve Leakage – Tested in accordance with the acceptance criterion established in BS EN ISO 11070.

{7}------------------------------------------------

Conclusion:

The results of the tests support the conclusion that, the Extra Large Check-Flo® Introducer met the design input requirements based on the intended use. Further, these results also support the conclusion that the design and material differences for the subject device, when compared to the predicate device, the GORE® DrySeal Sheath (K093791), support a conclusion that the Extra Large Check-Flo Introducer, subject of this submission, met the design input requirements based on the device's intended use and that it is substantially equivalent to the predicate device.

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).