(29 days)
Mobius3D v3.0 (K192424)
No reference devices were used in this submission
No
The document describes the software as using a proprietary collapsed cone convolution superposition (CCCS) algorithm for dose calculation and image comparison for alignment analysis. There is no mention of AI, ML, or related terms, nor is there any description of training or test sets typically associated with ML models.
No
Explanation: This device is described as a "quality assurance tool" and an "analysis tool meant solely for quality assurance (QA) purposes" in radiation therapy. It does not directly provide therapy or treatment to a patient.
No
Explanation: The document states that Mobius3D is solely a quality assurance (QA) tool for radiation therapy, used for treatment plan verification, dose calculation verification, and patient alignment analysis. It is explicitly stated that it is "not a treatment planning system" and "never comes into contact with patients," indicating it does not diagnose patient conditions.
Yes
The device description explicitly states "Being a software-only QA tool, Mobius3D never comes into contact with patients." and describes its function as software-based analysis and calculation.
Based on the provided information, this device is NOT an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health.
- Mobius3D's Function: Mobius3D is a software tool used for quality assurance and verification in radiation therapy. It analyzes treatment plans, calculates radiation dose based on existing data (treatment plans, log files), and compares medical images (CT and CBCT) for patient alignment.
- Lack of Biological Sample Analysis: The description explicitly states that Mobius3D "never comes into contact with patients" and its inputs are treatment plans, log files, and medical images. It does not analyze any biological samples from a patient.
Therefore, Mobius3D falls outside the scope of an In Vitro Diagnostic device. It is a software tool used in the planning and quality control aspects of radiation therapy, not for diagnostic testing using biological samples.
N/A
Intended Use / Indications for Use
Mobius3D software is used for quality assurance, treatment plan verification, and patient alignment and anatomy analysis in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system and may additionally be based on external measurements of radiation from other sources such as linac delivery log data. Patient alignment and anatomy analysis is based on read-in treatment planning images (such as computed tomography) and read-in daily treatment images (such as registered cone beam computed tomography).
Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
Product codes (comma separated list FDA assigned to the subject device)
IYE
Device Description
Mobius3D (v4.0) is a software product used within a radiation therapy clinic for quality assurance and treatment plan verification. It is important to note that while Mobius3D operates in the field of radiation therapy, it is neither a radiation delivery device (e.g. a linear accelerator), nor is it a Treatment Planning System (TPS). Mobius3D cannot design or transmit instructions to a delivery device, nor does it control any other medical device. Mobius3D is an analysis tool meant solely for quality assurance (QA) purposes when used by trained medical professionals. Being a software-only QA tool, Mobius3D never comes into contact with patients.
Mobius3D performs dose calculation verifications for radiation treatment plans by doing an independent calculation of radiation dose is initially calculated by a Treatment Planning System (TPS), which is a software tool that develops a detailed set of instructions (i.e. a plan) for another system (e.g. a linear accelerator) to deliver radiation to a patient. The dose calculation performed by Mobius3D uses a proprietary collapsed cone convolution superposition (CCCS) algorithm.
Mobius3D also performs dose delivery quality assurance for radiation treatment plans by using the measured data recorded in a linear accelerator's delivery log files to calculate a delivered dose. This is presented to the end user in a software component of Mobius3D called MobiusFX. The MobiusFX component is available to users through licensing as an add-on to the core Mobius3D software features.
Mobius3D performs quality assurance of a patient's alignment and anatomy analysis is based on comparison of Cone Beam Computed Tomography (CBCT) images taken immediately before treatment to the images used for treatment planning, which are typically acquired using standard Computed Tomography (CT). This analysis is presented to the end user in an add-on software module within Mobius3D called CBCT Checks.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Computed Tomography, Cone Beam Computed Tomography
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
trained radiation oncology personnel / radiation therapy clinic
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Software Verification and Validation Testing
Software verification and validation were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern.
Test results demonstrate conformance to applicable requirements and specifications. No animal studies or clinical tests have been included in this pre-market submission.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Mobius3D v3.0 (K192424)
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
No reference devices were used in this submission
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.5050 Medical charged-particle radiation therapy system.
(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
January 14, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Varian Medical Systems, Inc. % Mr. Peter J. Coronado Senior Director, Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304
Re: K203669
Trade/Device Name: Mobius3D v4.0 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE Dated: December 14, 2020 Received: December 16, 2020
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
1
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K203669
Device Name Mobius3D v4.0
Indications for Use (Describe)
Mobius3D software is used for quality assurance, treatment plan verification, and patient alignment and anatomy analysis in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system and may additionally be based on external measurements of radiation from other sources such as linac delivery log data. Patient alignment and anatomy analysis is based on read-in treatment planning images (such as computed tomography) and read-in daily treatment images (such as registered cone beam computed tomography).
Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038 Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
510(k) Summary
The following information is provided as required by 21 CFR 807.92
SUBMITTER | |
---|---|
Name and Address: | Varian Medical Systems, Inc. |
3100 Hansen Way, m/s E110 | |
Palo Alto, CA 94304 | |
Contact Person: | Peter J. Coronado |
Senior Director, Regulatory Affairs | |
Phone: 650-424-6320 Fax: 650-646-9200 | |
submissions.support@varian.com | |
Date Prepared: | December 21, 2020 |
DEVICE | |
Subject Device Name: | Mobius3D v4.0 |
Common/Usual Name: | Mobius3D |
Product Code and Classification: | Medical charged-particle radiation therapy system |
IYE 21 CFR 892.5050 Class II | |
PREDICATE DEVICE | |
Predicate Device Name: | Mobius3D v3.0 (K192424) |
Reference Device(s): | No reference devices were used in this submission |
DEVICE DESCRIPTION
Mobius3D (v4.0) is a software product used within a radiation therapy clinic for quality assurance and treatment plan verification. It is important to note that while Mobius3D operates in the field of radiation therapy, it is neither a radiation delivery device (e.g. a linear accelerator), nor is it a Treatment Planning System (TPS). Mobius3D cannot design or transmit instructions to a delivery device, nor does it control any other medical device. Mobius3D is an analysis tool meant solely for quality assurance (QA) purposes when used by trained medical professionals. Being a softwareonly QA tool, Mobius3D never comes into contact with patients.
4
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038 Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
Mobius3D performs dose calculation verifications for radiation treatment plans by doing an independent calculation of radiation dose is initially calculated by a Treatment Planning System (TPS), which is a software tool that develops a detailed set of instructions (i.e. a plan) for another system (e.g. a linear accelerator) to deliver radiation to a patient. The dose calculation performed by Mobius3D uses a proprietary collapsed cone convolution superposition (CCCS) algorithm.
Mobius3D also performs dose delivery quality assurance for radiation treatment plans by using the measured data recorded in a linear accelerator's delivery log files to calculate a delivered dose. This is presented to the end user in a software component of Mobius3D called MobiusFX. The MobiusFX component is available to users through licensing as an add-on to the core Mobius3D software features.
Mobius3D performs quality assurance of a patient's alignment and anatomy analysis is based on comparison of Cone Beam Computed Tomography (CBCT) images taken immediately before treatment to the images used for treatment planning, which are typically acquired using standard Computed Tomography (CT). This analysis is presented to the end user in an add-on software module within Mobius3D called CBCT Checks.
INDICATIONS FOR USE
Mobius3D software is used for quality assurance, treatment plan verification, and patient alignment and anatomy analysis in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Patient alignment and anatomy analysis is based on read-in treatment planning images (such as computed tomography) and read-in daily treatment images (such as registered cone beam computed tomography).
Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The modified device, referred to as the "subject device" throughout this summary, is release version v4.0 (version 4.0) of the Mobius3D with additional software changes incorporated since the release version of the predicate device, version 3.0 (K192424).
5
The significant change in the subject device compared with the predicate device is as follows:
1. Dose Calculation on Cone-Beam CT (CBCT)
This is a change to the existing CBCT module. Besides the Gamma Comparison, the CBCT module will (optionally) calculate dose on the CBCT using the applied shifts and TPS (Treatment Planning System) plan and structure set. This allows the user to better understand the impact of a patient's anatomy change or a possible alignment issue because it will show the dose on the patient's 3D image of that day. Dose evaluation can be done in the Slice Viewer by showing calculated isodose lines on CBCT. The dose in the non-deformed targets will be compared with the TPS dose, no deformation of structures is performed. Dose calculation on CBCT is optional and can be configured per Plan Type.
The non-significant changes in the subject device since the predicate device include:
- a. Plan type specific calculation and tolerance settings, target objectives and required regions of interests (ROIs)
- b. Streamlined processing
- Multi-machine support for the MobiusAdapt patient session C.
- d. National Language Support (NLS)
- e. Active directory support
PERFORMANCE DATA
The following performance data was provided in support of the substantial equivalence determination.
Software Verification and Validation Testing
Software verification and validation were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern.
Test results demonstrate conformance to applicable requirements and specifications. No animal studies or clinical tests have been included in this pre-market submission.
Standards Conformance
The subject device conforms in whole or in part with the following standards:
- IEC 62304:2006+A1:2015 Medical device software - Software lifecycle processes
6
varian
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038 Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
- IEC 62366-1:2015 Medical device Part 1 Application of usability engineering to medical ● devices
- IEC 61217:2011 Radiotherapy Equipment Coordinates, Movements and Scales ●
CONCLUSION
The predicate device was cleared based only on non-clinical testing, and no animal or clinical studies were performed for the subject device. The non-clinical verification and validation data demonstrates that the subject device should perform as intended in the specified use conditions.
Therefore, Varian considers the subject device, Mobius3D v4.0, is substantially equivalent to the predicate device Mobius3D v3.0 (K192424).