K Number
K203549

Validate with FDA (Live)

Date Cleared
2022-10-18

(683 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nova Primary Glucose Analyzer System is indicated for in vitro diagnostic use by healthcare professionals in clinical laboratory setting for the quantitative determination of Glucose in lithium heparinized venous whole blood and plasma.

The measurement of Glucose is used in the diagnosis and treatment of carbohydrate metabolism distuding diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

Device Description

The Nova Primary Glucose Analyzer System is a small, portable laboratory glucose analyzer that measures blood glucose levels in lithium heparinized whole blood or plasma utilizing a glucose oxidase based sensor and membrane/cap assembly.

The Nova Primary Glucose Analyzer accepts samples from syringes, blood collection tubes, microcentrifuge tubes, and sample cups. The sample size for analysis is 25 microliters (aspirated volume).

Nova Primary System Components:
The Nova Primary Glucose Analyzer System is comprised of the following components.

  • Nova Primary Glucose Analyzer
  • Nova Primary Glucose Sensor
  • Nova Primary Glucose Membrane
  • Nova Primary Calibrator Cartridge
  • Optional Barcode Scanner
  • IFU/Labeling

Sample Types:
The Nova Primary Glucose Analyzer System accepts lithium heparinized venous whole blood and plasma.

AI/ML Overview

The Nova Primary Glucose Analyzer System by Nova Biomedical Corporation is indicated for in vitro diagnostic use by healthcare professionals in clinical laboratory settings for the quantitative determination of Glucose in lithium heparinized venous whole blood and plasma.

The study presented does not explicitly list acceptance criteria in a tabular format with corresponding performance results. Instead, it describes performance testing conducted to demonstrate substantial equivalence to the predicate device, the YSI 2300 Stat Glucose and L-Lactate Analyzer. The performance tests included:

  • Method Comparison Studies
  • Precision/Reproducibility Studies
  • Linearity Testing
  • Specificity / Interference Testing
  • Detection Limit
  • Shelf Life Stability Testing

The "Summary of Performance Testing" section states, "The results of the performance testing confirmed that the Nova Primary Glucose Analyzer demonstrates substantial equivalence to the YSI 2300 Stat Glucose and L-Lactate Analyzer." This implies that the device met the criteria necessary to be considered substantially equivalent to the predicate device in terms of these performance characteristics. However, specific numerical acceptance criteria and the exact reported device performance metrics are not detailed in the provided text.

Here's an attempt to structure the information based on the provided text, while acknowledging the lack of explicit numerical acceptance criteria:

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance (Implied)
Method ComparisonPerformance comparable to the YSI 2300 Stat Glucose and L-Lactate AnalyzerConfirmed substantial equivalence to predicate device
Precision/ReproducibilityPerformance comparable to the YSI 2300 Stat Glucose and L-Lactate AnalyzerConfirmed substantial equivalence to predicate device
LinearityPerformance comparable to the YSI 2300 Stat Glucose and L-Lactate AnalyzerConfirmed substantial equivalence to predicate device
Specificity / InterferencePerformance comparable to the YSI 2300 Stat Glucose and L-Lactate AnalyzerConfirmed substantial equivalence to predicate device
Detection LimitPerformance comparable to the YSI 2300 Stat Glucose and L-Lactate AnalyzerConfirmed substantial equivalence to predicate device
Shelf Life StabilityPerformance comparable to the YSI 2300 Stat Glucose and L-Lactate AnalyzerConfirmed substantial equivalence to predicate device
Software ValidationSoftware functions as intended and safelyConfirmed substantial equivalence to predicate device
Plasma Equivalent Glucose ValueAlgorithms provide plasma equivalent glucose values traceable to isotope dilution mass spectrometryAutomatic adjustment for whole blood specimens based on conductivity detector

2. Sample Size for Test Set and Data Provenance

The document does not explicitly state the sample sizes used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature) for the performance testing. It only mentions that "Testing was completed to show that the Nova Primary Glucose Analyzer demonstrates substantial equivalence to the YSI 2300 Stat Glucose and L-Lactate Analyzer."

3. Number of Experts for Ground Truth and Qualifications

This information is not provided in the document. The device is a laboratory instrument for quantitative determination of glucose, and ground truth would typically be established by recognized reference methods or predicate devices, not by human experts in the diagnostic sense.

4. Adjudication Method for the Test Set

This information is not applicable and not provided. As noted above, the ground truth is based on objective laboratory measurements and reference methods, not subjective assessments requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically relevant for interpretative diagnostic devices where human reader performance is a key metric, often in image-based diagnostics. The Nova Primary Glucose Analyzer System is a quantitative measurement device.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

Yes, the performance testing described (Method Comparison, Precision/Reproducibility, Linearity, Specificity/Interference, Detection Limit, Shelf Life Stability) would inherently be a standalone evaluation of the algorithm and device's accuracy and reliability in measuring glucose concentration. The device is designed for automated quantitative measurement.

7. Type of Ground Truth Used

The ground truth for the device's performance is established by comparison to a legally marketed predicate device (YSI 2300 Stat Glucose and L-Lactate Analyzer) and traceability to isotope dilution mass spectrometry (for plasma equivalent glucose results). This suggests established laboratory reference methods and comparison to a 'gold standard' device.

8. Sample Size for the Training Set

The document does not specify a separate "training set" sample size. For an analytical device like this, the "training" (calibration and algorithm development) is typically an internal process, and the performance testing is done on independent samples.

9. How Ground Truth for the Training Set Was Established

The document states that "All glucose measurement algorithms are intended to report plasma equivalent glucose results that are substantially equivalent to the predicate device plasma glucose result, and are traceable to isotope dilution mass spectrometry." This implies that the algorithms were developed or "trained" using data and methods that align with these established analytical principles and reference standards. However, the specific details of how this "ground truth" was established for internal algorithm development (training) are not elaborated.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 18, 2022

Nova Biomedical Corporation Robert Zinck Senior Manager Regulatory Affairs 200 Prospect Street Waltham, MA 02454

Re: K203549

Trade/Device Name: Nova Primary Glucose Analyzer System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CGA Dated: July 29, 2022 Received: August 2, 2022

Dear Robert Zinck:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203549

Device Name

Nova Primary Glucose Analyzer System

Indications for Use (Describe)

The Nova Primary Glucose Analyzer System is indicated for in vitro diagnostic use by healthcare professionals in clinical laboratory setting for the quantitative determination of Glucose in lithium heparinized venous whole blood and plasma.

The measurement of Glucose is used in the diagnosis and treatment of carbohydrate metabolism distuding diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K203549

510(K) Owner:Nova Biomedical Corporation
Registration Number:1219029
Address:200 Prospect St.Waltham, MA 02454
Phone:781-894-0800
Fax Number:784-891-4806
Contact Person:Robert Zinck, Senior Manager Regulatory Affairs
Date Prepared:October 14, 2022

Proprietary Name: Nova Primary Glucose Analyzer System

Common or Usual Name: Blood Glucose Analyzer

Classification Name:

Regulation sectionClassificationProductcode
21 CFR § 864.1345 Glucose Test SystemClass IICGA

Product Code:

CGA

Predicate Device:

K891480 -YSI 2300 Stat Glucose and L-Lactate Analyzer

Device Description:

The Nova Primary Glucose Analyzer System is a small, portable laboratory glucose analyzer that measures blood glucose levels in lithium heparinized whole blood or plasma utilizing a glucose oxidase based sensor and membrane/cap assembly.

The Nova Primary Glucose Analyzer accepts samples from syringes, blood collection tubes, microcentrifuge tubes, and sample cups. The sample size for analysis is 25 microliters (aspirated volume).

Nova Primary System Components:

The Nova Primary Glucose Analyzer System is comprised of the following components.

  • Nova Primary Glucose Analyzer ●
  • . Nova Primary Glucose Sensor
  • . Nova Primary Glucose Membrane
  • Nova Primary Calibrator Cartridge ●
  • Optional Barcode Scanner .
  • . IFU/Labeling

Sample Types:

The Nova Primary Glucose Analyzer System accepts lithium heparinized venous whole blood and plasma.

{4}------------------------------------------------

Intended Use:

Please see Indication for Use.

Indication for Use:

The Nova Primary Glucose Analyzer System is indicated for in vitro diagnostic use by healthcare professionals in clinical laboratory settings for the quantitative determination of Glucose in lithium heparinized venous whole blood and plasma.

The measurement of Glucose is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypodycemia, and idiopathic hypodlycemia, and pancreatic islet cell carcinoma.

Summary of the Technological Characteristics:

The Nova Primary Glucose Analyzer was designed to be substantially equivalent in performance to the predicate device, incorporating modern technology updates in the analyzer platform.

The Nova Primary Glucose Analyzer utilizes a discrete, easily replaceable glucose sensor and membrane/cap assembly. Glucose measurement is conducted on diluted venous whole blood or plasma samples and is based on the enzymatic reaction between glucose and oxygen molecules in the presence of the glucose oxidase enzyme. All glucose measurement algorithms are intended to report plasma equivalent glucose results that are substantially equivalent to the predicate device plasma glucose result, and are traceable to isotope dilution mass spectrometry. For whole blood specimens, the Nova Primary Glucose Analyzer utilizes an impedance-based conductivity detector to automatically adjust glucose measurement and provide the plasma equivalent glucose value based on the conductivity of the sample.

Principle of Measurement:

Glucose measurement is based on the level of H2O2 produced during the enzymatic reaction between glucose and oxygen molecules in the presence of the dlucose enzyme. At a constant potential of 0.70 volts, electroactive HzO2 is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the surface of the platinum electrode is proportional to the glucose concentration of the sample.

Summary of Performance Testing:

Testing was completed to show that the Nova Primary Glucose Analyzer demonstrates substantial equivalence to the YSI 2300 Stat Glucose and L-Lactate Analyzer. The performance testing included:

  • Method Comparison Studies ●
  • Precision/Reproducibility Studies .
  • Linearity Testing
  • Specificity / Interference Testing ●
  • Detection Limit .
  • Shelf Life Stability Testing .

The results of the performance testing confirmed that the Nova Primary Glucose Analyzer demonstrates substantial equivalence to the YSI 2300 Stat Glucose and L-Lactate Analyzer.

Conclusion:

The results of software validation and performance verification testing confirmed that the Nova Primary Glucose Analyzer System demonstrates substantial equivalence to the YSI 2300 Stat Glucose and L-Lactate Analyzer (K891480 - predicate device).

{5}------------------------------------------------

Table 1: Comparison of Predicate and Proposed Devices
---------------------------------------------------------
CharacteristicPredicate:Proposed:
Intended/Indicationsfor UseThe YSI 2300 STAT PLUS Glucose andLactate Analyzer is a laboratory instrumentintended for use in clinical care and sportsmedicine applications. It provides quickmeasurements of glucose in whole blood,plasma or serum; and of L-lactate in wholeblood, plasma, or cerebrospinal fluid (CSF).In whole blood or plasma, glucose and L-lactate can be measured simultaneously.The Nova Primary Glucose Analyzer Systemis indicated for in vitro diagnostic use byhealthcare professionals in clinical laboratorysettings for the quantitative determination ofGlucose in lithium heparinized venous wholeblood and plasma.The measurement of Glucose is used in thediagnosis and treatment of carbohydratemetabolism disturbances including diabetesmellitus, neonatal hypoglycemia, andidiopathic hypoglycemia, and pancreatic isletcell carcinoma.
Intended UsersTrained laboratory personnelSame
Measured TestsGlucose and LactateGlucose
Glucose EnzymeGlucose OxidaseSame
Acceptable SampleTypesWhole Blood, Plasma, SerumLithium Heparin Whole Blood or Plasma
Sample SizeRequirement25 microliters (aspirated volume)Same
Physical Dimensions10.0 x 14.0 x 14.0 inches19.0 x 19.0 x 6.0 inches
Instrument Weight25 poundsLess than 30 pounds including the calibratorcartridge
Sample TrayAvailable OptionNot available
Bar Code ScannerNot availableOptional, wireless bar code scanner viainternal USB dongle
KeyboardNot availableOptional, wireless keyboard via internal USBdongle
Power Supply110-120 VAC, 220-240 VAC50–60 Hz50 Watts nominal100-240 VAC47-63 HzLess than 50 Watts
Automatic WholeBlood HematocritCorrectionNo - Hematocrit is independently measuredon a whole blood specimen then manuallyentered into YSI 2300 to calculate equivalentplasma glucose value.Yes - Contains an impedance-basedConductivity Detector. Whole blood samplesare automatically adjusted to provide plasmaequivalent glucose values based on theconductivity of the sample.
User InterfaceDisplayAlpha Numeric Liquid Crystal Display with 20membrane keypadPCAP Color Touchscreen Display with iconbased Graphical User Interface
PrinterYes onboard 2 ¼" thermal printerYes, onboard 2" thermal printer
CommunicationProtocolRS-232 serial portNot available

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.