(535 days)
The Xpert GBS LB XC test, performed on the GeneXpert® Instrument Systems, is an automated qualitative in vitro diagnostic test for the detection of Group B Streptococcus (GBS) DNA from enriched vaginal/rectal swab specimens, using real- time polymerase chain reaction (PCR).
Xpert GBS LB XC testing is indicated as an aid in determining the GBS colonization status of antepartum women.
· The Xpert GBS LB XC test is intended for antepartum testing on enriched Lim broth cultures of vaginal/rectal swabs after 18-24 hours of incubation
· The Xpert GBS LB XC test does not provide antimicrobial susceptibility test results. Culture is necessary to obtain isolates to perform susceptibility testing as recommended for penicillin-allergic women
The Xpert GBS LB XC test is an automated in vitro diagnostic test for qualitative detection of DNA from Group B Streptococcus (GBS) from vaginal-rectal swab specimens obtained from pregnant women that are transported to the laboratory following enrichment in Lim broth.
The primers and probes in the Xpert GBS LB XC test are designed to simultaneously amplify and detect two unique GBS chromosomal targets: the first is a target within a coding region for a glycosyl transferase family protein and the second is within a coding region for a LysR family transcriptional regulator of Streptococcus agalactiae DNA.
The Xpert GBS LB XC test includes reagents for the detection of DNA from GBS in Lim broth-enriched vaginal/rectal swabs. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge. The SPC is present to control for adequate extraction and processing of the target sequences and to monitor for the presence of inhibitors in the PCR reaction. The PCC verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dve stability.
The single-use, multi-chambered fluidic cartridges are designed to complete sample preparation and real-time PCR for the detection of GBS genomic DNA in as little as 27 minutes with high titer specimens; GBS negative specimens generate results in approximated 43 minutes. The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems and the GeneXpert Infinity Systems, have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR and RT-PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and RT-PCR and detection.
The Xpert GBS LB XC test is performed on the Cepheid GeneXpert® Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48s, and GeneXpert Infinity-80 systems). The GeneXpert Instrument System platform automates sample preparation, amplification and real-time detection. The GeneXpert systems consist of an instrument, computer, and preloaded software for running tests and viewing the results.
The GeneXpert Instrument Systems require the use of single-use, disposable cartridges (the Xpert GBS LB XC cartridges) that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained and specimens never come into contact with working parts of the instrument modules, cross-contamination between samples is minimized.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
The document describes the performance of the Xpert GBS LB XC test, a qualitative in vitro diagnostic test for the detection of Group B Streptococcus (GBS) DNA. The acceptance criteria are implicitly derived from the reported performance characteristics.
| Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Sensitivity | High (e.g., > 95%) | 99.3% (95%CI: 96.1-99.9) |
| Specificity | High (e.g., > 95%) | 98.7% (95%CI: 97.3-99.4) |
| PPV (Positive Predictive Value) | High | 95.9% (95%CI: 91.4-98.1) |
| NPV (Negative Predictive Value) | High | 99.8% (95%CI: 98.8-100.0) |
| Limit of Detection (LoD) | Consistent detection across GBS serotypes at low concentrations | Varies by serotype, e.g., Serotype Ia: 663 CFU/mL (50 CFU/swab); Serotype Ib: 40 CFU/mL (3 CFU/swab) |
| Analytical Reactivity (Inclusivity) | 100% detection of all GBS strains, including cfb mutants | 100% positivity for all tested cfb mutant GBS strains |
| Analytical Specificity (Exclusivity) & Microbial Interference | No cross-reactivity or interference with clinically relevant pathogens/flora | No cross-reactivity or microbial interference observed with a panel of 128 strains |
| Interfering Substances | No interference in the presence of common vaginal/rectal substances | All positive and negative samples correctly identified in the presence of various interfering substances |
| Carry-Over Contamination | No false positives from highly positive samples | All 20 positive samples correctly reported, all 22 negative samples correctly reported following high GBS positive samples |
| Reproducibility | High agreement across sites, operators, and lots | Overall percent agreement for individual samples: 97.2% - 100.0% |
Study Details
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: 621 specimens were included in the Xpert GBS LB XC versus Composite Comparator analysis.
- Data Provenance: The study was a multi-site clinical study conducted in the United States. It used vaginal/rectal swab specimens collected from pregnant females as part of routine care. The nature of the specimen collection ("as a part of routine care") suggests it was likely a prospective collection for this study, though it might have utilized leftover samples. The text specifies "aliquots of leftover Lim broth samples were obtained for testing."
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not specify the number of experts or their qualifications for establishing the ground truth. The ground truth was established using a "composite comparator method" which involved enriched bacterial culture and an FDA cleared NAAT. For the culture component, species identification was performed via Matrix Assisted Laser Desorption Ionization - Time of Flight Mass Spectroscopy (MALDI-TOF MS). While MALDI-TOF MS requires skilled lab personnel, the term "expert" in the sense of a medical professional making a diagnostic determination is not used for this ground truth establishment.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- The adjudication method for the test set was: "a specimen was considered positive if either enriched bacterial culture or the FDA cleared NAAT was positive and negative when both enriched bacterial culture and the FDA cleared NAAT were negative." This is a rule-based composite comparator rather than an expert adjudication method like 2+1.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study focusing on human readers improving with AI assistance was not done. This device is an automated in vitro diagnostic test (NAAT) for detecting GBS DNA. It does not involve human "readers" in the diagnostic process beyond laboratory technicians operating the equipment and interpreting the automated qualitative results.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, the clinical "Performance of Xpert GBS LB XC v Composite Comparator" (Table 9) is a standalone performance assessment. The Xpert GBS LB XC test is an automated qualitative test where the instrument system performs sample preparation, amplification, and real-time detection, with the software running tests and viewing results. This is inherently an "algorithm only" performance, as the final output is a qualitative (Positive/Negative) result generated by the device itself.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- A composite comparator method was used as the ground truth. This method combined:
- Enriched bacterial culture with species identification via Matrix Assisted Laser Desorption Ionization - Time of Flight Mass Spectroscopy (MALDI-TOF MS).
- An FDA cleared Nucleic Acid Amplification Test (NAAT).
- A composite comparator method was used as the ground truth. This method combined:
-
The sample size for the training set:
- The document describes a clinical validation study for regulatory submission. It does not provide information about the sample size used for the training set of the algorithm or device. Regulatory submissions typically focus on validation performance rather than development/training data.
-
How the ground truth for the training set was established:
- Since information on the training set (including its sample size) is not provided, the method for establishing its ground truth is also not detailed in this document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it, also in blue.
May 9, 2022
Cepheid Robert Resnick Principal Regulatory Affairs Specialist 904 Caribbean Drive Sunnyvale, California 94089
Re: K203429
Trade/Device Name: Xpert GBS LB XC. GeneXpert Dx System. GeneXpert Infinity System Regulation Number: 21 CFR 866.3740 Regulation Name: Streptococcus Spp. Serological Reagents Regulatory Class: Class I Product Code: NJR, OOI Dated: November 18, 2020 Received: November 20, 2020
Dear Robert Resnick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Ribhi Shawar, Ph.D. (ABMM) Branch Chief, General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K203429
Device Name Xpert GBS LB XC
Indications for Use (Describe)
The Xpert GBS LB XC test, performed on the GeneXpert® Instrument Systems, is an automated qualitative in vitro diagnostic test for the detection of Group B Streptococcus (GBS) DNA from enriched vaginal/rectal swab specimens, using real- time polymerase chain reaction (PCR).
Xpert GBS LB XC testing is indicated as an aid in determining the GBS colonization status of antepartum women.
· The Xpert GBS LB XC test is intended for antepartum testing on enriched Lim broth cultures of vaginal/rectal swabs after 18-24 hours of incubation
· The Xpert GBS LB XC test does not provide antimicrobial susceptibility test results. Culture is necessary to obtain isolates to perform susceptibility testing as recommended for penicillin-allergic women
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
× Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
As required by 21 CFR Section 807.92(c).
I. SUBMITTER
| Submitted by: | Cepheid 904 Caribbean Drive Sunnyvale, CA 90489 Phone number: (408) 541-4191 Fax number: (408) 541-4192 |
|---|---|
| Contact: | Robert Resnick. M.S. |
| Date of Preparation: | December 06, 2021 |
| II. DEVICE | |
| Trade name: | Xpert® GBS LB XC |
| Common name: | Xpert® GBS LB XC |
| Type of Test: | Qualitative real-time polymerase chain reaction (PCR) and detection test |
| Regulation number,Classification name,Product code: Definition | 21 CFR 866.3740, Streptococcus spp. serological reagents. NJR;Definition: A nucleic acid amplification assay system (including probes,other reagents, and instrumentation) is an aid in the identification ofgroup b streptococci from pre-partum and intra-partum women toestablish colonization status. |
| 21 CFR 862.2570, Instrumentation for clinical multiplex test systems,OOI; Definition: The system is a clinical multiplex instrument intendedto measure and sort multiple signals generated my multiple probes,intercalating dyes, or other ligands in an assay from a clinical sample.Signals may be generated by fluorescence or other phenomena and maybe measured using filters on a photodiode or other detector. It mayintegrate sample and/or reagent handling, amplification, dedicatedinstrument control, data acquisition software, raw data storagemechanisms and other essential hardware components along with thesignal reader unit. The system is used with specific assays to comprisean assay test system. | |
| Classification AdvisoryPanel | Microbiology (83) |
| Prescription Use | Yes |
III. PREDICATE DEVICE
Predicate Device Assay: Xpert GBS LB (K121539)
{4}------------------------------------------------
IV. DEVICE DESCRIPTION
The Xpert GBS LB XC test is an automated in vitro diagnostic test for qualitative detection of DNA from Group B Streptococcus (GBS) from vaginal-rectal swab specimens obtained from pregnant women that are transported to the laboratory following enrichment in Lim broth.
The primers and probes in the Xpert GBS LB XC test are designed to simultaneously amplify and detect two unique GBS chromosomal targets: the first is a target within a coding region for a glycosyl transferase family protein and the second is within a coding region for a LysR family transcriptional regulator of Streptococcus agalactiae DNA.
The Xpert GBS LB XC test includes reagents for the detection of DNA from GBS in Lim broth-enriched vaginal/rectal swabs. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge. The SPC is present to control for adequate extraction and processing of the target sequences and to monitor for the presence of inhibitors in the PCR reaction. The PCC verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dve stability.
The single-use, multi-chambered fluidic cartridges are designed to complete sample preparation and real-time PCR for the detection of GBS genomic DNA in as little as 27 minutes with high titer specimens; GBS negative specimens generate results in approximated 43 minutes. The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems and the GeneXpert Infinity Systems, have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR and RT-PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and RT-PCR and detection.
The Xpert GBS LB XC test is performed on the Cepheid GeneXpert® Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48s, and GeneXpert Infinity-80 systems). The GeneXpert Instrument System platform automates sample preparation, amplification and real-time detection. The GeneXpert systems consist of an instrument, computer, and preloaded software for running tests and viewing the results.
The GeneXpert Instrument Systems require the use of single-use, disposable cartridges (the Xpert GBS LB XC cartridges) that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained and specimens never come into contact with working parts of the instrument modules, cross-contamination between samples is minimized.
{5}------------------------------------------------
V. DEVICE INTENDED USE:
The Xpert® GBS LB XC test, performed on the GeneXpert® Instrument Systems, is an automated qualitative in vitro diagnostic test for the detection of Group B Streptoccus (GBS) DNA from enriched vaginal/rectal swab specimens, using real-time polymerase chain reaction (PCR).
Xpert GBS LB XC testing is indicated as an aid in determining the GBS colonization status of antepartum women.
- . The Xpert GBS LB XC test is intended for antepartum testing on enriched Lim broth cultures of vaginal/rectal swabs after 18-24 hours of incubation.
- . The Xpert GBS LB XC test does not provide antimicrobial susceptibility test results. Culture is necessary to obtain isolates to perform susceptibility testing as recommended for penicillin-allergic women.
This device is intended for prescription use only.
{6}------------------------------------------------
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Substantial Equivalence:
The Xpert GBS LB XC test is substantially equivalent to the Xpert GBS LB Assay [510(k) # K121539].
The performance of the Xpert GBS LB XC test was evaluated in a multi-site clinical study in which the performance of the Xpert GBS LB XC test was determined relative to a composite comparator reference method using both results from culture and a FDA cleared nucleic acid test. The results of the study demonstrated that the performance of the Xpert GBS LB XC is substantially equivalent to the predicate device.
The following tables compare Xpert® GBS LB XC test to Xpert® GBS LB Assay (K121539). Table 1 shows similarities between the new device and the predicate, while Table 2 shows the differences.
| Comparison | ||
|---|---|---|
| Attribute | New DeviceXpert® GBS LB XC | Predicate DeviceXpert GBS LB (K121539) |
| Regulation | Same | 21CFR 866.3740Streptococcus spp. serologicalreagents |
| Product Code | Same | NJRNucleic acid amplification assaysystem, group b streptococcus, directspecimen test |
| Device Class | Same | I (non-exempt) |
| Intended Use | The Xpert GBS LB XC test,performed on the GeneXpert®Instrument Systems, is an automatedqualitative in vitro diagnostic test forthe detection of Group BStreptococcus (GBS) DNA fromenriched vaginal/rectal swabspecimens, using real-timepolymerase chain reaction (PCR).Xpert GBS LB XC testing isindicated as an aid in determining theGBS colonization status ofantepartum women.• The Xpert GBS LB XC test isintended for antepartum testing onenriched Lim broth cultures ofvaginal/rectal swabs after 18–24hours of incubation | The Cepheid Xpert GBS LB Assay,performed on the GeneXpert®Instrument Systems, is a qualitative in vitro diagnostic test designed to detectGroup B Streptococcus (GBS) DNAfrom enriched vaginal/rectal swabspecimens, using fully automatedreal-time polymerase chain reaction(PCR) with fluorogenic detection ofthe amplified DNA.Xpert GBS LB Assay testing isindicated as an aid in determiningGBS colonization status inantepartum women.• The Xpert GBS LB Assay is usedfor antepartum testing on enrichedLim broth cultures of vaginal/rectalswabs after 18–24 hours of incubation• The Xpert GBS LB Assay does notprovide susceptibility results. Culture |
Table 1. Similarities between New Device and Predicate Device
{7}------------------------------------------------
| Comparison | ||
|---|---|---|
| Attribute | New Device | Predicate Device |
| Xpert® GBS LB XC | Xpert GBS LB (K121539) | |
| • The Xpert GBS LB XC test doesnot provide antimicrobialsusceptibility test results. Culture isnecessary to obtain isolates toperform susceptibility testing asrecommended for penicillin-allergicwomen, | isolates are needed for performingsusceptibility testing as recommendedfor penicillin-allergic women. | |
| Instrument Systems | Same | Cepheid GeneXpert InstrumentSystems |
| Laboratory Users | Same | CLIA Moderate Complexity |
| Specimen Type | Same | Enriched Lim broth cultures preparedfrom vaginal/rectal swabs after 18-24hours of incubation |
| Collection and TransportMedia | Same | Cepheid Swab Collection Device orequivalent swab in non-nutritivetransport medium |
| Assay Technology | Same | Real-Time PCR |
| Self-Contained SystemAssay | Same | Yes |
| Single Use | Same | Yes |
| Automated Extraction,detection and resultinterpretation | Same | Yes |
| Fluidics | Same | Self-contained |
| External Assay Controls | Same | Materials available but not required |
| Built in Lysis Control | Same | Yes |
| Assay Results | Same | Qualitative |
| Time to Result(after enrichment) | Same | <60 minutes |
Table 2. Differences between New Device and Predicate
| Comparison | ||
|---|---|---|
| Attribute | New Device | Predicate Device |
| Xpert® GBS LB XC | Xpert GBS LB (K121539) | |
| Analyte Target | Dual target assay design:The test targets two conservedchromosomal sequences in S.agalactiae: 1) a member of theglycosysl transferase gene family and2) a LysR transcriptional regulator. | Single target assay design:The test targets the 3' DNA regionadjacent to the cfb CAMP-factorhemolysin gene of S. agalactiae. |
| Assay Internal Controls | Same except that the IPC waseliminated since SPC is a fullyprocess control monitoring bothsample processing and PCR inhibition(monitored by the IPC). | Integrated internal controls to monitorsample processing, reagent hydrationand PCR inhibition.- Specimen Processing Control(SPC)- Internal Process Control (IPC)- Probe Check Control (PPC) |
{8}------------------------------------------------
| Comparison | ||
|---|---|---|
| Attribute | New Device | Predicate Device |
| Xpert® GBS LB XC | Xpert GBS LB (K121539) | |
| Early Assay Termination(EAT) Feature | YesEarly Assay Termination can reducethe test time for positive results asearly as 27 minutes. With GBSnegative samples, the test returnsresults in approximately 43 minutes. | No |
The Xpert GBS LB XC test has the same general intended use as the predicate device and has the same technological characteristics as the predicate device. The differences between the Xpert GBS LB XC test and the predicate device do not raise different questions of safety and effectiveness.
{9}------------------------------------------------
VII. PERFORMANCE DATA
Non-Clinical Studies:
Analytical Sensitivity (Limit of Detection) and Analytical Reactivity (Inclusivity)
The analytical reactivity and limit of detection (LoD) of the Xpert GBS LB XC test were determined for 12 different strains representing 12 known serotypes of GBS, of which 2 were characterized as non-hemolytic (Table 3). Serial dilutions of each serotype were prepared in a Lim broth negative clinical sample matrix or in a simulated sample matrix. Serotypes Ia. III and V were tested with 24 replicates per dilution level for each of two reagent lots across three days. Serotypes Ib, Ic, II, IV and VI-X were tested with one reagent lot for a total of 24 replicates of each dilution level across three days. The LoD was established for each serotype and reagent lot by probit logistic regression analysis.
The LoD for each serotype was verified by testing 20 replicates at the 95% confidence interval upper limit with one reagent lot across three days. The results for all serotypes except serotype V and VI were >95% (>19/20) detected. The result for serotype V and VI was 85% (17/20) detected and the claimed LoD is based on the upper level of 95% confidence interval.
| Serotype | LoD (CFU/mL) Probit Result | 95% CI Probit Result | Percent Detected | LoD (CFU/mL) Verified | LoD (CFU/swab) Verified |
|---|---|---|---|---|---|
| Ia | 663 | 492-835 | 100% | 663 | 50 |
| Ib | 40 | 32-49 | 95% | 40 | 3 |
| Ica | 301 | 231-370 | 100% | 301 | 23 |
| IIa | 173 | 132-213 | 100% | 173 | 13 |
| III | 540 | 409-670 | 100% | 540 | 41 |
| IV | 429 | 324-533 | 95% | 429 | 32 |
| V | 618 | 384-618 | 85% | 618b | 46 |
| VI | 544 | 353-544 | 85% | 544b | 41 |
| VII | 620 | 512-728 | 100% | 620 | 47 |
| VIII | 682 | 509-855 | 100% | 682 | 51 |
| IX | 465 | 354-575 | 100% | 465 | 35 |
| X | 677 | 525-829 | 95% | 677 | 51 |
Table 3. GBS Limit of Detection (LoD)
ª Non-hemolytic strain
b Claimed LoD corresponds to upper 95% upper CI
{10}------------------------------------------------
Analytical Reactivity with GBS cfb Mutants
A study was performed to evaluate the analytical reactivity of Xpert GBS LB XC test using GBS strains containing deletions in or adjacent to the chromosome that encodes the CAMP factor hemolysis gene ctb. Ten unique, well characterized GBS clinical isolates representing different c/o mutations were tested at 833 CFU/mL. All strains with c/b mutations were detected with a positivity rate of 100%.
Analytical Specificity (Exclusivity) and Microbial Interference
The analytical specificity of the Xpert GBS LB XC test was evaluated by testing a panel of 128 strains, representing bacterial, viral, parasite and yeast strains commonly found in vaginal/rectal flora or phylogenetically related to GBS (Table 4) in the Xpert GBS LB XC test in the absence (exclusivity) or presence (microbial interference) of GBS. Bacteria were tested at ≥ 1x106 CFU/ml, except as noted, and viruses and parasites were tested at a level of > 1x105 organisms, yeast, IU or copies/ml. Microorganisms with potential to grow to high titers in Lim broth during enrichment (Candida albicans, Enterococcus faecalis, Enterococcus faecium, Enterococcus gallinarum, Streptococcus anginosus, Streptococcus parasanguinis, Corynebacterium accolens) were tested at > 1x10° CFU/ml. Of the 128 strains, 121 were tested in Lim broth clinical sample matrix or in simulated sample matrix, both in presence of GBS at 3x LoD and in absence of GBS.
Seven of 128 strains (Finegoldia magna, Mobiluncus curtisii subsp. curtisii, Peptoniphilus asaccharolyticus. Fusobacterium nucleatum. Peptostreptococcus anaerobius, Anaerococcus tetradius and Anaerococcus prevotii ) were not available for in vitro testing and were evaluated by in silico analysis using the Xpert GBS LB XC primer and probe sequences as queries for organism- specific BLAST (Basic Local Alignment Search Tool) analysis of the NCBI (National Center for Biotechnology Information) Nucleotide collection (nr/nt) database.
No cross-reactivity or microbial interference of GBS detection was observed, both in silico and in vitro, with any clinically relevant pathogens.
| Organism | ||
|---|---|---|
| Arcanobacterium (Trueperella) pyogenes | Haemophilus influenzae | Serratia marcescens |
| Atopobium (Fannyhessea) vaginae | Hafnia alvei | Shigella flexneri |
| Abiotrophia defectiva | Hepatitis B virus | Shigella sonnei |
| Acinetobacter baumannii | Hepatitis C virus | Staphylococcus aureusc |
| Acinetobacter lwoffii | Human immunodeficiencyvirus | Staphylococcus epidermidis |
| Actinobacillus pleuropneumoniae | Human Papillomavirus 18b | Staphylococcus haemolyticus |
| Organism | Organism | Organism |
| Aeromonas hydrophila | Klebsiella (Enterobacter)aerogenes | Staphylococcus intermedius |
| Alcaligenes faecalis | Klebsiella oxytoca | Staphylococcus lugdunensis |
| Anaerococcus lactolyticus | Klebsiella pneumoniae | Staphylococcus saprophyticus |
| Anaerococcus prevotiia | Lactobacillus acidophilus | Staphylococcus simulans |
| Anaerococcus tetradiusa | Lactobacillus casei | Stenotrophomonas maltophilia |
| Bacillus cereus | Lactobacillus delbrueckii lactis | Streptococcus acidominimus |
| Bacillus coagulans | Lactobacillus gasseri | Streptococcus anginosus |
| Bacteroides fragilis | Lactobacillus plantarum | Streptococcus bovis |
| Bifidobacterium adolescentis Reuter | Lactobacillus reuteri | Streptococcus canis |
| Bifidobacterium brevis | Listeria monocytogenes | Streptococcus constellatus |
| BK virus | Micrococcus luteus | Streptococcus criceti |
| Blastocystis hominisb | Mobiluncus curtisii subsp.Curtisiia | Streptococcus cristatus |
| Bordetella pertussis | Moraxella atlantae | Streptococcus downei |
| Burkholderia cepacia | Moraxella catarrhalis | Streptococcus dysgalactiaesubsp. dysgalactiae |
| Campylobacter jejuni | Morganella morganii | Streptococcus dysgalactiaesubsp. equisimilis |
| Candida albicans | Mycoplasma genitaliumb | Streptococcus equi subsp. equi |
| Candida glabrata | Neisseria gonorrhoeae | Streptococcus gordonii |
| Candida tropicalis | Norovirus | Streptococcus intermedius |
| Chlamydia trachomatis | Pantoea agglomerans | Streptococcus mitis |
| Citrobacter freundii | Pasteurella aerogenes | Streptococcus mutans |
| Clostridium difficile | Peptoniphilusasaccharolyticusa | Streptococcus oralis |
| Cytomegalovirus | Peptostreptococcusanaerobiusa | Streptococcus parasanguinis |
| Corynebacterium accolens | Porphyromonasasaccharolytica | Streptococcus pneumoniae |
| Corynebacterium sp. (genitalium) | Prevotella bivia | Streptococcus pseudoporcinus |
| Corynebacterium urealyticum | Prevotella melaninogenica | Streptococcus pyogenesb |
| Cryptococcus neoformans | Prevotella oralis | Streptococcus ratti |
| Enterobacter cloacae | Propionibacterium acnes | Streptococcus salivarius |
| Organism | ||
| Enterococcus durans | Proteus mirabilis | Streptococcus sanguinis |
| Enterococcus faecalis | Proteus vulgaris | Streptococcus sobrinus |
| Enterococcus faecium | Providencia stuartiib | Streptococcus suis |
| Enterococcus gallinarum | Pseudomonas aeruginosa | Streptococcus uberis |
| Epstein-Barr virus | Pseudomonas fluorescens | Streptococcus vestibularis |
| Escherichia coli | Rhodococcus equi | Toxoplasma gondii |
| Finegoldia magnaa | Rubella virus | Trichomonas vaginalis |
| Fusobacterium nucleatuma | Salmonella enterica subsp. enterica ser. Dublin (group D) | Vibrio cholerae |
| Gardnerella vaginalis | Salmonella enterica subsp. typhimurium | Yersinia enterocolitica |
| Giardia lambliab | Serratia liquefaciens | |
| a Evaluated in silico | ||
| b Evaluated with DNA |
Table 4. Analytical Specificity of the Xpert GBS LB XC Test
{11}------------------------------------------------
{12}------------------------------------------------
Interfering Substances Study
Substances that may be present in vaginal/rectal specimens with the potential to interfere with the Xpert GBS LB XC test were evaluated. Potentially interfering endogenous and exogenous substances include human amniotic fluid, meconium, urine, fecal material, human blood, lubricating gel, vaginal anti-itch medications, vaginal antifungal medications, anti- diarrheal medications, laxatives, stool softeners, topical hemorrhoid ointments, body oil, body powder, deodorant sprays, enema solutions, and spermicidal foam.
These substances are listed in Table 5. All liquid substances were tested by adding 100% of the substance to the swab, solid substances by covering swab head to 75% and tablets were dissolved to their highest soluble concentration in simulate sample matrix and added to the swab. Five exogenous substances (Aquasonic® gel, Floraplus, Pepto Bismol®, Skin oil and Xyloproct) were tested at lower concentration to determine the highest tolerated amount on swab (Table 5). The interferents were tested in the absence of GBS or in the presence of GBS at 3x LoD.
There was no interference in the presence of the substances at the concentrations tested in this study. All positive and negative samples were correctly identified using the Xpert GBS LB XC test.
{13}------------------------------------------------
| Substance (Source) | Substance Form | Concentration on Swab |
|---|---|---|
| Human Amniotic Fluid (Lee Biosolutions Inc.) | Liquid | 60% (v/v) |
| Human Urine (Novakemi) | Liquid | 60% (v/v) |
| Human Whole Blood - EDTA (BioChemed) | Liquid | 80%(v/v) |
| Human Whole Blood - Na Citrate (BioChemed) | Liquid | 80% (v/v) |
| Leukocytes, Buffy coat, 2x107 WBCs/ml(Karolinska University Hospital) | Liquid | 80% (v/v) |
| Meconium (Lee Biosolutions) | Solid | 100% (w/v) |
| Mucus (Sigma-Aldrich/Merck) | Solid | 30% (w/v) |
| Human Feces - Pool of 10 donors | Solid | 100 % (w/v) |
| Anti-Diarrheal Medication (Pepto Bismol) | Liquida | 40 % (v/v) |
| Anti-Diarrheal Medication (Dimor Comp[Dimeticone] from Nordic Drugs) | Tablet | 0.03% loperamid +2% dimetkon (w/v) |
| Lubricant (RFSU Klick Ultra Glide) | Solid | 100% (w/v) |
| Lubricant (Sense Me Aqua Glide) | Solid | 100% (w/v) |
| Lubricant (KY-Jelly) | Solid | 100% (w/v) |
| Body Oil (ACO Repairing Skin Oil) | Liquidb | 100% (w/v) |
| Dialon Baby (Dialon Baby powder) | Solid | 100% (w/v) |
| Deodorant Powder (Vagisil® Deodorant Powder) | Solid | 100% (w/v) |
| Deodorant Spray (LN Intimate Deo) | Liquid | 60% (v/v) |
| Deodorant Suppositories (Norforms FeminineDeodorant Suppositories) | Solid | 28.5% (w/v) |
| Enema solution (Microlax mikrolavemang –McNeil) | Solid | 100% (w/v) |
| Oral Laxative (Inolaxol – Mylan) | Solid | 25% (w/v) |
| Oral Laxative (Phillips Milk of Magnesia - Bayer) | Liquid | 60% (w/v) |
| Oral Laxative (Pursennid Ex-Lax - GSK) | Tablet | 0.64% (w/v) |
| Spermicidal Foam (Caya preventivgel) | Solid | 100% (v/v) |
| Stool Softener (Laktulos - Meda | Liquid | 60% (v/v) |
| Stool Softener (Movicol – Norgine) | Solid | 11% (w/v) |
| Topical Hemorrhoid Ointment (Xyloproct RectalOintment - Aspen) | Solidc | 10% (w/v) |
| Topical Hemorrhoid Ointment (Scheriproctrektalsalva / Prednisolone Ointment- Bayer) | Solid | 100% (v/v) |
| Ultrasound Transmission Gel (Aquasonic Gel -Parker) | Solidc | 25% /w/v) |
| Vaginal Antifungal Gel (Multi-Gyn Actigel) | Solid | 100% (v/v) |
| Vaginal Antifungal Gel (Multi-Gyn Floraplus) | Solidc | 75% (w/v) |
| Vaginal Anti-itch Cream (Ellen Probiotisk UtvärtesIntim Creme) | Solid | 100% (v/v) |
| Vaginal Antifungal Cream (Canesten - Bayer) | Solid | 100% (v/v) |
| Vaginal Antifungal Cream (Daktar – McNeil) | Solid | 100% (v/v) |
Table 5. Potentially Interfering Substances Tested
Pepto Bismol® diluted to 50% in background simulated background matrix and no interference observed.
b Skin oil tolerated when 67% of swab head covered (tested as solid substance).
b Skin oil tolerated when 67% of swab head covered (tested as solid substance).
c Substances were diluted into a simulated background matrix prior to testing: Xyloproot Rectal at 10%, Aquasonic Gel was tested at 25% and MultiGyn Floraplus was tested at 75%. No interference was detection after dilution.
{14}------------------------------------------------
Carry-Over Contamination
A study was conducted to demonstrate that no carry-over contamination occurs when testing these single-use, self-contained GeneXpert cartridges in the same GeneXpert module. The study consisted of a negative sample processed in the same GeneXpert module immediately following a high GBS positive sample. Twenty-one runs alternating high titer GBS positive and GBS negative samples were performed consecutively on two GeneXpert modules, thus a total of 42 runs were executed for the study. All 20 positive samples were correctly reported as GBS POSITIVE. All 22 negative samples were correctly reported as GBS NEGATIVE.
Reproducibility Study
A panel of ten samples with varying concentrations of four different GBS strains were tested by two operators each in triplicate on six different days at three sites using 3 different lots of Xpert GBS LB XC (2 operators × 3 replicates/day × 6 days × 3 sites) including negative panel members. Three lots of Xpert GBS LB XC were used at each of the three testing sites. The source of the three strains (serotype Ia, III, IV) representing a hemolytic phenotype and one strain (serotype Ic) representing a non-hemolytic phenotype is presented in Table 6. The three levels were ~3 × LOD and ~1 × LOD and negative.
| Strain | Phenotype | Panel Member |
|---|---|---|
| Not applicable | Not applicable | Negative |
| GBS Serotype Ia - ATCC12386 | Hemolytic | Low Positive (~1X LoD) |
| Moderate Positive (~3X LoD) | ||
| GBS Serotype Ic -ATCC13813 | Non-emolytic | Low Positive (~1X LoD) |
| Moderate Positive (~3X LoD) | ||
| GBS Serotype III - ATCC12403 | Hemolytic | Low Positive (~1X LoD)Moderate Positive (~3X LoD) |
| GBS Serotype IV -ATCC49446 | Hemolytic | Low Positive (~1X LoD)Moderate Positive (~3X LoD) |
Table 6. Reproducibility Panel
Xpert GBS LB XC testing was performed on the GeneXpert Instrument Systems according to the Xpert GBS LB XC test procedure.
The percent agreement of the qualitative results for GBS detection for each sample calculated for each of the six operators and for each site is shown in Table 7. In addition, the overall percent agreement for each sample (total agreement) and the 95% two-sided Wilson Score confidence interval are shown in the last column.
{15}------------------------------------------------
| Sample | Site 1 | Site 2 | Site 3 | Total Agreement bySample with95% CI | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| OP 1 | OP 2 | Subtotal | OP 1 | OP 2 | Subtotal | OP 1 | OP 2 | Subtotal | ||||
| 01 | Negative 1ª | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(108/108)96.6-100.0 | |
| 02 | Negative 2ª | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(108/108)96.6-100.0 | |
| 03 | GBSserotypeIa~1xLoD | 100.0%(18/18) | 94.4%(17/18) | 97.2%(35/36) | 100.0%(18/18) | 94.4%(17/18) | 97.2%(35/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 98.1%(106/108)93.5-99.5 | |
| 04 | GBSserotypeIcb~1xLoD | 100%(18/18) | 100%(18/18) | 100%(36/36) | 100%(18/18) | 100%(18/18) | 100%(36/36) | 100%(18/18) | 100%(18/18) | 100%(36/36) | 100.0%(108/108)96.6-100.0 | |
| 05 | GBSserotypeIII~1xLoD | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(108/108)96.6-100.0 | |
| 06 | GBSserotype IV~1xLoD | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 94.0%(17/18) | 94.0%(17/18) | 94.0%(34/36) | 100.0%(18/18) | 94.0%(17/18) | 97.0%(35/36) | 97.2%(105/108)92.1-99.4 | |
| 07 | GBSserotype Ia~3xLoD | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(108/108)96.6-100.0 | |
| 08 | GBSserotypeIcb~3xLoD | 100.0%(17/17) | 100.0%(18/18) | 100.0%(35/35) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(107/107)96.5-100.0 | |
| 09 | GBSserotype III~3xLoD | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(108/108)96.6-100.0 | |
| 10 | GBSserotypeIV~3xLoD | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(18/18) | 100.0%(18/18) | 100.0%(36/36) | 100.0%(108/108)96.6-100.0 |
Table 7: Summary of Reproducibility and Precision Results
Testing with Serotype Ic was performed separately from testing by the other panel members. Negative panel a. members were included in both rounds of testing and are represented separately in the performance table above.
b. Serotype Ic is characterized as non-hemolytic.
No statistically significant difference in performance (p-value of <0.01) was observed between the study sites, operators, or the lots used; p-values ranged from 0.7715 to 1.
ANOVA Analysis
The study also enabled evaluation of repeatability (within-run variance/assay precision) as well as the within-laboratory precision of the underlying quantitative result (Ct values) obtained from the Xpert test, on which the qualitative reportable results are based.
{16}------------------------------------------------
The mean, standard deviation (SD), and coefficient of variation (CV) between-sites, between-lots, between-days, between-operators and within-assay for each panel member are presented in Table 8.
| N | MeanCT | Variance Source | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Panel Member | Site | Operator | Lot | Day | Within-Run | Total | ||||||||
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | CV(%) | |||
| Mod PosGBS serotype Ia~3xLoD | 108 | 32.7 | 0.1 | 0.6 | 0 | 0 | 0.3 | 11.0 | 0 | 0 | 0.9 | 88.4 | 0.9 | 2.9 |
| Mod PosGBS serotype III~3xLoD | 108 | 33.9 | 0 | 0 | 0.2 | 5.6 | 0.3 | 10.7 | 0 | 0 | 0.9 | 83.6 | 0.9 | 2.8 |
| Mod PosGBS serotype IV~3xLoD | 108 | 33.3 | 0 | 0 | 0.2 | 4.1 | 0.4 | 21.0 | 0.1 | 1.0 | 0.8 | 73.9 | 1.0 | 2.9 |
| Low PosGBS serotype Ia~1xLoD | 106a | 34.6 | 0.6 | 16.0 | 0 | 0 | 0.7 | 23.2 | 0 | 0 | 1.1 | 60.8 | 1.5 | 4.3 |
| Low PosGBS serotype III~1xLoD | 108 | 35.0 | 0.3 | 4.0 | 0 | 0 | 0.2 | 2.6 | 0.5 | 14.9 | 1.1 | 78.5 | 1.3 | 3.6 |
| Low PosGBS serotype IV~1xLoD | 105b | 34.9 | 0 | 0 | 0.5 | 19.1 | 0.2 | 3.1 | 0.2 | 3.9 | 1.1 | 73.9 | 1.2 | 3.6 |
| Negative 1º | 108 | 32.1 | 0.1 | 0.7 | 0 | 0 | 0.3 | 18.2 | 0.2 | 10.7 | 0.6 | 70.4 | 0.7 | 2.2 |
| Mod PosGBS serotype Ic~3xLoD | 107d | 33.5 | 0 | 0 | 0 | 0 | 0.3 | 11.0 | 0.5 | 34.9 | 0.6 | 54.1 | 0.8 | 2.4 |
| Low PosGBS serotype Ic~1xLoD | 108 | 35.2 | 0 | 0 | 0.2 | 2.7 | 0 | 0 | 0.6 | 21.5 | 1.0 | 75.8 | 1.2 | 3.4 |
| Negative 2º | 108 | 31.9 | 0 | 0 | 0 | 0 | 0.4 | 26.8 | 0.2 | 5.6 | 0.6 | 67.6 | 0.7 | 2.1 |
Table 8. ANOVA Summary of Reproducibility Data
a. Two panel members with GBS negative results were excluded from ANOVA analysis.
b. Three panel members with GBS negative results were excluded from ANOVA analysis.
c. ANOVA calculations are based on the SPC Ct values for the negative panel members.
d. One sample with a GBS negative result was excluded from ANOVA analysis.
The total SDs ranged from 0.7 to 1.5. The SDs did not exceed 50% of the total SD for the following factors: Site, Operator-to-Operator, Lot-to-Lot and Day-to-Day. The majority of the total variance came from within-assay random variance (assay noise), and not from the other evaluated factors.
{17}------------------------------------------------
Clinical Studies
Clinical Performance
Clinical performance of the Xpert GBS LB XC test was evaluated in a multisite study conducted in the United States using the GeneXpert Dx instrument system. Vaginal/rectal swab specimens were collected at three (3) geographically diverse sites from pregnant females for GBS testing as a part of routine care. Specimens were inoculated in Lim broth per institutional policy. For eligible specimens, aliquots of leftover Lim broth samples were obtained for testing with the Xpert GBS LB XC test. The results of the Xpert GBS LB XC test were compared to a composite comparator method. The composite comparator method comprised of enriched bacterial culture with species identification via Matrix Assisted Laser Desorption Ionization - Time of Flight Mass Spectroscopy (MALDI-TOF MS) and an FDA cleared NAAT. For the composite comparator, a specimen was considered positive if either enriched bacterial culture or the FDA cleared NAAT was positive and negative when both enriched bacterial culture and the FDA cleared NAAT were negative. Additionally, the Xpert GBS LB XC test was compared directly to the FDA cleared NAAT test.
Performance of Xpert GBS LB XC v Composite Comparator
A total of 621 specimens with results from enriched bacterial culture and the FDA cleared NAAT were included in the analyses of Xpert GBS LB XC versus the composite comparator method. The results are shown in the table below.
| Composite Comparator | ||||
|---|---|---|---|---|
| Positive | Negative | Total | ||
| Positive | 142 | 6 | 148 | |
| Xpert GBS LB XC | Negative | 1 | 472 | 473 |
| Total | 143 | 478 | 621 | |
| Sensitivity | 99.3% (95%CI: 96.1-99.9) | |||
| Specificity | 98.7% (95%CI: 97.3-99.4) | |||
| PPV | 95.9% (95%CI: 91.4-98.1) | |||
| NPV | 99.8% (95%CI: 98.8-100.0) | |||
| Prevalence | 23.0% (95%CI: 19.9-26.5) |
Table 9. Xpert GBS LB XC Performance vs. Composite Comparator
As shown in Table 9, the sensitivity and specificity of the Xpert GBS LB XC test compared to the composite comparator method were 99.3% and 98.7%, respectively.
{18}------------------------------------------------
Performance of the Xpert GBS LB XC Test vs. FDA Cleared NAAT
A direct comparison of the Xpert GBS LB XC test to the FDA cleared NAAT was also performed. The results are shown in the table below.
Of 622 samples tested with the Xpert GBS LB XC test during this study, nine yielded nondeterminate results on the initial test. These nine samples were retested and eight returned valid results. The initial non-determinate rate was 1.4% (9/622) and the final non-determinate rate was 0.2% (1/622).
{19}------------------------------------------------
VIII. CONCLUSIONS
The results of the nonclinical analytical and clinical performance studies summarized above demonstrate that the Xpert GBS LB XC test is substantially equivalent to the predicate device.
§ 866.3740
Streptococcus spp. serological reagents.(a)
Identification. Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identifyStreptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genusStreptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.