K Number
K203292
Date Cleared
2021-03-22

(133 days)

Product Code
Regulation Number
866.3830
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit is intended as a qualitative test for the detection of IgG and IgM antibodies to Borrelia burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. When used as the first-tier screening test, positive and equivocal results must be confirmed through additional testing by one of the following methods:

• Standard two-tier test methodology (STTT) using an IgG and/or IgM blot testing following current interpretation guidelines, OR

• Modified two-tier test methodology (MTTT) using the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/ IgM ELISA Test.

The assay can also be used as a second-tier confirmation test using the MTTT methodology when used with the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test as the first-tier screening test.

Positive test results by either the STTT or MTTT methodology are supportive evidence for the presence of antibodies and exposure to Borrelia burgdorferi, the cause of Lyme disease. A diagnosis of Lyme disease should be made based on the presence of Borrelia burgdorferi antibodies, history, symptoms, and other laboratory findings.

Device Description

The kit includes 12 x 8 well Antigen Coated strips, Conjugate, Substrate, Stop Solution, Wash Buffer, Diluent, Negative Control, Positive Control, and Cutoff Control. The controls are provided to determine if the assay is functioning properly and to determine the antibody level. The reagents are sufficient for 96 determinations.

During the test procedure, antibodies to B. burgdorferi (sensu stricto) if present in the human serum sample will bind to the antigens coated onto the wells forming antigen-antibody complexes. Excess antibodies are removed by washing. A conjugate of goat anti-human IgG/IgM antibodies conjugated with horseradish peroxidase are then added, which binds to the antigen-antibody complexes. Excess conjugate is removed by washing. This is followed by the addition of a chromogenic substrate, tetramethylbenzidine (TMB). If specific antibodies to the antigen are present in the patients' serum, a blue color will develop. The enzymatic reaction is then stopped with a stopping solution causing the contents of the well to turn yellow. The wells are read photometrically with a microplate reader at 450nm.

The antigens used in the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test kit is a combination of B. burgdorferi sensu stricto strain B31 lysate, B. burgdorferi sensu stricto strain 2591 lysate, and a recombinant VlsE from B. burgdorferi sensu stricto strain B31. The lysates use spirochetes growing in BSK-H complete medium until mid-exponential phase. The recombinant VlsE protein is produced in E. coli SURE2 cells and purified by affinity chromatography. The purity of each antigen is assayed by SDS-PAGE followed by Coomassie staining and/or Western blotting.

AI/ML Overview

The provided document describes the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit and its performance in various studies. Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance (STTT comparison):

The document does not explicitly state pre-defined acceptance criteria for the comparative studies. Instead, it presents observed performance metrics (Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA)) against a predicate device. For the purpose of this response, I will highlight the reported performance metrics from the comparative study.

MetricReported Device Performance (STTT)Basis of Comparison
Positive Percent Agreement (PPA)96.0% (72/75) with 95% CI (88.8% - 99.2%)Predicate IgG/IgM ELISA
Negative Percent Agreement (NPA)97.5% (434/445) with 95% CI (95.6% - 98.8%)Predicate IgG/IgM ELISA

1. Table of Acceptance Criteria and Reported Device Performance (MTTT Comparison):

Similar to the STTT comparison, the document reports observed performance metrics for the MTTT protocol against the STTT predicate.

MetricReported Device Performance (MTTT) (Second-tier test of MTTT vs. STTT for first-tier positive/equivocal samples)Reported Device Performance (MTTT) (MTTT vs. STTT for all prospective study samples)Basis of Comparison
Positive Percent Agreement (PPA)100.0% (95% CI (90.3% - 100.0%))100.0% (95% CI (90.3% - 100.0%))Predicate Immunoblots (STTT)
Negative Percent Agreement (NPA)27.8% (95% CI (9.7% - 53.5%))97.1% (95% CI (95.1% - 98.4%))Predicate Immunoblots (STTT)

2. Sample Size Used for the Test Set and Data Provenance:

For Comparison with Predicate Device (STTT):

  • Sample Size: 520 serum samples.
  • Data Provenance: Prospective samples submitted for Lyme serology testing. Collected from three sites (one internal and two external reference laboratories).

For MTTT Comparison (Prospective Study):

  • Sample Size: 481 serum samples for the initial screening. Of these, 54 positive or equivocal samples were further tested.
  • Data Provenance: Prospective samples submitted for Lyme serology testing. Collected from three sites (one internal and two external reference laboratories).

For Sensitivity Study (STTT and MTTT):

  • Sample Size: 89 clinically characterized samples for STTT and 125 clinically characterized samples for MTTT.
  • Data Provenance: Clinically characterized samples encompassing early, disseminated, and late stages of Lyme disease. The document doesn't specify geographical origin but implies a clinical context.

For CDC Panel (STTT and MTTT):

  • Sample Size: 40 samples for STTT and 280 samples for MTTT.
  • Data Provenance: Acquired from the Centers for Disease Control (CDC).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

The document does not provide specific details on the number or qualifications of experts used to establish the ground truth for the test sets. The ground truth appears to be based on:

  • "Clinically characterized samples" (for sensitivity studies).
  • "Patients diagnosed with Lyme disease" and "healthy individuals" from the CDC panel.
  • Comparison with predicate devices (for comparative studies).

4. Adjudication Method for the Test Set:

The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for resolving discrepancies or establishing ground truth within the test sets. For comparative studies, the predicate device results largely serve as the reference, and for sensitivity studies, samples are clinically characterized, implying pre-existing diagnoses.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This device is an in-vitro diagnostic (IVD) ELISA test kit, not an AI-powered image analysis or diagnostic tool that involves human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The device is an ELISA test kit designed to provide a qualitative result (positive, equivocal, negative) based on optical density readings. It functions as a standalone diagnostic assay without a human-in-the-loop component in its primary function. The interpretation of the results by laboratory personnel is part of standard lab practice, but the "standalone performance" in this context refers to the assay's accuracy in detecting antibodies. The reported PPA, NPA, and sensitivity studies reflect this standalone performance.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc):

The ground truth used for different studies varies:

  • Comparison Studies with Predicate Device: The results of the legally marketed predicate devices (Trinity Biotech Captia™ Borrelia burgdorferi IgG/IgM ELISA Test Kit and Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM Blot Tests) served as the reference standard.
  • Sensitivity Studies: "Clinically characterized samples" and "patients diagnosed with Lyme disease" from the CDC panel. This implies a combination of clinical assessment, symptomology, and potentially other laboratory findings, which can be seen as a form of expert consensus or aggregated clinical data.
  • Analytical Specificity and Cross-Reactivity: For analytical specificity, asymptomatic individuals from endemic and non-endemic regions were used. For cross-reactivity, samples confirmed positive for specific disease markers (from serum vendors) were used.

8. The Sample Size for the Training Set:

The document describes the determination of the assay cutoff but does not explicitly mention a "training set" in the context of machine learning or AI. For the cutoff determination:

  • Assay Cutoff: 200 normal sera (100 from endemic, 100 from non-endemic regions).
  • Verification: 125 characterized Lyme disease samples were used with the 200 normal samples for ROC analysis to verify the chosen cutoff.

9. How the Ground Truth for the Training Set Was Established:

As noted above, there isn't a "training set" in the typical AI sense. However, for the determination of the assay cutoff:

  • Normal Sera: These were identified as "normal" based on their origin from non-diseased individuals in endemic and non-endemic regions.
  • Characterized Lyme Disease Samples: These 125 samples were "characterized" for Lyme disease, implying a pre-established clinical diagnosis or a well-defined status as positive Lyme disease samples.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

March 22, 2021

Gold Standard Diagnostics Jennifer Roth Vice President, Product Development 2851 Spafford St. Davis, California 95618

Re: K203292

Trade/Device Name: Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema Pallidum Treponemal Test Reagents Regulatory Class: Class II Product Code: LSR Dated: November 4, 2020 Received: November 9, 2020

Dear Jennifer Roth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Maria Ines Garcia, Ph.D. Branch Chief Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203292

Device Name

Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit

Indications for Use (Describe)

The Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit is intended as a qualitative test for the detection of IgG and IgM antibodies to Borrelia burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. When used as the first-tier screening test, positive and equivocal results must be confirmed through additional testing by one of the following methods:

• Standard two-tier test methodology (STTT) using an IgG and/or IgM blot testing following current interpretation guidelines, OR

• Modified two-tier test methodology (MTT) using the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/ IgM ELISA Test.

The assay can also be used as a second-tier confirmation test using the MTTT methodology when used with the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test as the first-tier screening test.

Positive test results by either the STTT or MTTT methodology are supportive evidence for the presence of antibodies and exposure to Borrelia burgdorferi, the cause of Lyme disease. A diagnosis of Lyme disease should be made based on the presence of Borrelia burgdorferi antibodies, history, symptoms, and other laboratory findings.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Gold Standard Diagnostics. The logo features the company name in bold, dark blue text, stacked in three lines. To the right of the text is a graphic of three overlapping, stylized hexagons in a gold color. The hexagons are arranged in a way that they appear to be interconnected.

510(k) Summary

This 510(k) summary is being submitted in accordance with the requirement of SMDA 1990 and 21 CFR 807.92.

  1. Submitter's Name: Gold Standard Diagnostics Address: 620 Cantrill Drive Davis, CA. 95618 Phone Number: 530-759-8000 Contact Person: Jennifer Roth November 4, 2020 Date:
    1. Product and Trade Name: Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit
      Common Name: Lyme ELISA Test

Regulation Section:

(21 CFR 866.3830) Treponema pallidum treponemal test reagents.

Classification: Class II

Product Code: LSR; Reagent, Borrelia Serological Reagent

  • Note: This clearance is for a modified use (Modified Two-tier Testing or MTTT use) for the previously cleared IVD test, the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit (K180264). The information and study data for the modified use is presented under the heading of "MTTT Comparison" below. With the exception of the intended use, all other information and data remain the same.

3) Legally Marketed Device to Which the Submitter Claims Equivalence:

  • a. STTT Trinity Biotech Captia™ Borrelia burgdorferi IgG/IgM ELISA Test Kit (K033070).
  • b. MTTT Gold Standard Diagnostics Borrelia burgdorferi IgG Blot Test Kit (K113847) and the Gold Standard Diagnostics Borrelia burgdorferi IgM Blot Test Kit (K113846).

4) Description of the Device:

The kit includes 12 x 8 well Antigen Coated strips, Conjugate, Substrate, Stop Solution, Wash Buffer, Diluent, Negative Control, Positive Control, and Cutoff Control. The controls are provided to determine if the assay is functioning properly and to determine the antibody level. The reagents are sufficient for 96 determinations.

During the test procedure, antibodies to B. burgdorferi (sensu stricto) if present in the human serum sample will bind to the antigens coated onto the wells forming antigen-antibody

{4}------------------------------------------------

complexes. Excess antibodies are removed by washing. A conjugate of goat anti-human IgG/IgM antibodies conjugated with horseradish peroxidase are then added, which binds to the antigen-antibody complexes. Excess conjugate is removed by washing. This is followed by the addition of a chromogenic substrate, tetramethylbenzidine (TMB). If specific antibodies to the antigen are present in the patients' serum, a blue color will develop. The enzymatic reaction is then stopped with a stopping solution causing the contents of the well to turn yellow. The wells are read photometrically with a microplate reader at 450nm.

The antigens used in the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test kit is a combination of B. burgdorferi sensu stricto strain B31 lysate, B. burgdorferi sensu stricto strain 2591 lysate, and a recombinant VlsE from B. burgdorferi sensu stricto strain B31. The lysates use spirochetes growing in BSK-H complete medium until mid-exponential phase. The recombinant VlsE protein is produced in E. coli SURE2 cells and purified by affinity chromatography. The purity of each antigen is assayed by SDS-PAGE followed by Coomassie staining and/or Western blotting.

5) Intended Use of the Device:

The Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit is intended as a qualitative test for the detection of IgG and IgM antibodies to Borrelia burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. When used as the first-tier screening test, positive and equivocal results must be confirmed through additional testing by one of the following methods:

  • . Standard two-tier test methodology (STTT) using an IgG and/or IgM blot testing following current interpretation guidelines, OR
  • Modified two-tier test methodology (MTTT) using the Gold Standard Diagnostics . Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test.

The assay can also be used as a second-tier confirmation test using the MTTT methodology when used with the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IzM ELISA Test as the first-tier screening test.

Positive test results by either the STTT or MTTT methodology are supportive evidence for the presence of antibodies and exposure to Borrelia burgdorferi, the cause of Lyme disease. A diagnosis of Lyme disease should be made based on the presence of Borrelia burgdorferi antibodies, history, symptoms, and other laboratory findings.

6) Comparison with the Predicate Device:

The tables below provide a comparison of the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test kit with the Trinity Biotech Captia™ Borrelia burgdorferi IgG/IgM ELISA Test kit (predicate device: K033070) when used as the first-step for the detection of IgG and IgM antibodies to B. burgdorferi in human serum. Positive and equivocal results must be supplemented by testing with a second-step Immunoblot assay.

{5}------------------------------------------------

Similarities
ItemSubject Device: Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test KitPredicate Device: Trinity Biotech Captia™ Borrelia burgdorferi IgG/IgM ELISA Test Kit
Intended UseThe Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test kit is intended as a qualitative presumptive (first step) test for the detection of IgG and IgM antibodies to B. burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. Positive and equivocal results must be supplemented by testing with a second-step Western blot assay.Trinity Biotech Captia™ Borrelia burgdorferi IgG/IgM ELISA Test kit is intended for the qualitative presumptive (first-step) detection of total (IgG/IgM) antibodies to B. burgdorferi in human serum. This ELISA should only be used for patients with signs and symptoms that are consistent with Lyme disease. Equivocal or positive results must be supplemented by testing with a standardized Western-blot (second step) procedure. Positive supplemental (second-step) results are supportive evidence of exposure to B. burgdorferi and can be used to support a clinical diagnosis of Lyme disease. The diagnosis of Lyme disease must be made based on history, signs (such as erythema migrans), symptoms, and other laboratory data, in addition to the presence of antibodies to B. burgdorferi. Negative results (either first or second step) should not be used to exclude Lyme disease.
Assay FormatAntigen coated microtiter plate – 96 wells.Same
TechnologyELISASame
Sample MatrixHuman serumSame
Controls ProvidedPositive, Cutoff, NegativeSame
Reagents ProvidedDiluent, Wash, Conjugate, Substrate, Stop SolutionSame
Reported ResultsPositive, Equivocal, NegativeSame
InterpretationOptical density readings from SpectrophotometerSame
Differences
ItemSubject Device: Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test KitPredicate Device: Trinity Biotech Captia™ Borrelia burgdorferi IgG/IgM ELISA Test Kit

{6}------------------------------------------------

Sample ProcessingDilute Samples 1:100 in DiluentDilute Samples 1:20 in Diluent
Volumes100ul sample, 50ul substrate, 50ul stop solution100ul sample, 100ul substrate, 100ul stop solution
Incubation15/15/15 minutes at room temperature25/25/10-15 minutes at room temperature
AntigensB. burgdorferi B31 strain,B. burgdorferi 2591 strain,B. burgdorferi recombinant VlsEB31 strainB. burgdorferi B31 strain
Results InterpretationConvert to units.Negative <9Equivocal 9.0-11.0Positive >11.0Convert to units.Negative ≤0.90Equivocal 0.91-1.09Positive ≥1.10

6(b1): Nonclinical Studies:

Determination of the Assay Cutoff

The cutoff was determined by testing a total of 200 normal sera which consisted of 100 sera from an endemic region of Lyme disease and 100 sera from a non-endemic region of Lyme disease. The mean plus two standard deviations was used to determine the assay cutoff. A known positive sample was then diluted to produce a ready to use cutoff control. After the cutoff was determined 125 characterized Lyme disease samples were tested. An ROC analysis was then generated with the 325 samples (200 normal and 125 characterized samples) to verify the chosen cutoff. The analysis confirmed that the assay cutoff provided an optimal level of sensitivity and specificity.

Precision

To determine the precision of the Borrelia burgdorferi IgG/IgM ELISA Test, a within-lab precision study was conducted. A precision panel consisting of a negative sample, a high negative sample, a low positive sample, and a moderate positive sample, along with the kit controls, was tested in-house. The sample panel was masked and randomized. Each of the panel members was tested in duplicate, twice per day, for 12 days. The sample panel was masked and randomized. The results are summarized in the following table:

SampleNMeanUnitsWithin-RunBetween-RunBetween-DayTotal
ModeratePositive4819.6SD0.8151.5341.4721.737
CV4.2%7.8%7.5%8.9%
LowPositive4812.1SD0.2671.4171.2481.442
CV2.2%11.7%10.3%11.9%
HighNegative486.1SD0.2110.6620.6420.695
CV3.4%10.8%10.5%11.4%
Negative481.7SD0.1130.1640.1510.199
CV6.6%9.6%8.8%11.6%

{7}------------------------------------------------

Reproducibilitv

A reproducibility panel consisting of a negative sample, a high negative sample, a low positive sample, and a moderate positive sample, along with the kit controls, was tested at three different sites. The sample panel was masked and randomized. Each of the panel members was tested in triplicate, twice per day, for five days. The Within-Run, Between-Run, Between-Days, and Between-Sites Standard Deviation and Coefficients of Variation (CV) were calculated. The sample panel was masked and randomized. The results are summarized in the following table:

SampleNMeanUnitsWithin-RunBetween-RunBetween-DayBetween-SitesTotal
ModeratePositive9021.1SD1.1171.5431.4340.3861.905
CV5.3%7.3%6.8%1.8%9.0%
LowPositive9013.8SD0.5911.1551.0260.4041.297
CV4.3%8.4%7.5%2.9%9.4%
HighNegative906.4SD0.3230.7560.7150.2370.822
CV5.0%11.7%11.1%3.7%12.8%
Negative901.6SD0.1450.3350.3120.3470.365
CV9.1%21.1%19.6%21.8%23.0%

Analytical Specificity

The analytical specificity was determined by testing 210 asymptomatic individuals' samples from endemic (Pennsylvania) and non-endemic (Arizona) regions. The Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test results are summarized in the following table:

Number of SamplesNumberPositive/EquivocalAnalytical Specificity
Endemic Region110397.3%
Non-endemic Region100298.0%

Cross Reactivity

A study using 221 samples was conducted to evaluate potential cross reactivity from different disease conditions. The samples were obtained from serum vendors who confirmed their positivity for each respective marker. The samples were tested on the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test. The results are summarized in the following table:

Infection / DiagnosisNumber of SeraTested# Positive / (%)
Tick-borne Relapsing Fever262 / (7.7%)
Treponemal Infections232* (8.7%)
Rickettsia100 / (0%)
Ehrlichiosis100 / (0%)
Babesiosis110 / (0%)

{8}------------------------------------------------

Leptospirosis110 / (0%)
Parvovirus B19120 / (0%)
Influenza A&B100 / (0%)
Epstein-Barr Virus100 / (0%)
Cytomegalovirus190 / (0%)
H. pylori110 / (0%)
Fibromyalgia101/ (10%)
Rheumatoid Arthritis110 / (0%)
Herpes Simplex Virus130 / (0%)
Varicella Zoster Virus120 / (0%)
Autoimmune Disease220 / (0%)

*Also positive on the predicate device

Interfering Substances

The effect of potential interfering substances on samples using the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test was evaluated. Three samples, a negative, a low positive and a moderate positive were spiked with high levels of interferants and were tested along with serum without spiked interferants. The recommended concentrations from the guideline "Interference Testing in Clinical Chemistry EP7-A2" from the Clinical and Laboratory Standards Institute were used (see table below). The tested substances did not affect the performance of the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test.

SubstanceConcentrationInterference
Albumin120 g/LNone detected
Bilirubin342 μmol/LNone detected
Cholesterol13 mmol/LNone detected
Hemoglobin2 g/LNone detected
Triglycerides37 mmol/LNone detected

6(b2): Clinical Studies:

Comparison with Predicate Device

Comparison studies were conducted at three sites (one internal and two external reference laboratories) using prospective samples submitted for Lyme serology testing. Five hundred and twenty (520) serum samples were tested on both the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test and on the predicate B. burgdorferi IgG/IgM ELISA Test. The results are summarized in the following table:

Predicate IgG/IgM ELISA
PositiveEquivocal*NegativeTotal
Gold StandardDiagnostics Borreliaburgdorferi IgG/IgMELISA Test KitPositive557264
Equivocal*82919
Negative21434437
Total6510445520

*Equivocal samples counted as positive

{9}------------------------------------------------

Positive percent agreement = 96.0% (72/75)95% CI (88.8% - 99.2%)
Negative percent agreement = 97.5% (434/445)95% CI (95.6% - 98.8%)

Clinical Sensitivity

Sensitivity Study

A sensitivity study was performed on 89 clinically characterized samples. The samples encompass early, disseminated, and late stages of Lyme disease. The samples were tested on both the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test and on the predicate B. burgdorferi IgG/IgM ELISA Test. The results are summarized in the following table:

DiseaseStagenGold StandardDiagnostics Borreliaburgdorferi IgG/IgMELISA Test KitPredicateIgG/IgM ELISA
Early3876.3% (29/38)78.9% (30/38)
Disseminated15100.0% (15/15)100.0% (15/15)
Late3697.2% (35/36)94.4% (34/36)

CDC Panel

Forty samples of various reactivity were acquired from the Centers for Disease Control (CDC). Of the 40 samples, five were from healthy individuals and 35 were from patients diagnose with Lyme disease and stratified by disease stage. All samples were tested on the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test and on the predicate Borrelia burgdorferi IgG/IgM ELISA Test. The results are summarized in the following table:

DiseaseStagenGold Standard DiagnosticsBorrelia burgdorferiIgG/IgM ELISA Test KitPredicateIgG/IgM ELISA
PositiveorEquivocal% Agreementwith ClinicalDiagnosisPositiveorEquivocal% Agreementwith ClinicalDiagnosis
Healthy50100.0%0100.0%
Early(0-2 months)151386.7%1280.0%
Convalescent(3-12 months)13753.8%753.8%
Late(>1 year)77100.0%7100.0%

Expected Values

The range of values and positivity rate among different studies and population for the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test are as follows:

{10}------------------------------------------------

Population# SamplesUnit ResultsQualitative Results
MeanRangeStd.Dev.#Positive/Equivocal%Positive/Equivocal
NormalEndemic1105.80.6 - 11.32.07532.7%
NormalNon-Endemic1004.20.9 - 12.32.07722.0%
ProspectiveStudy5206.30.70 - 32.35.4078316.0%
SensitivityStudy8925.03.6 - 34.98.4497988.8%

Note: It is recommended that each laboratory determine its own normal range based on the population.

7) Conclusion:

From the comparison data, we conclude that the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test is substantially equivalent to the Trinity Biotech Captia™ Borrelia burgdorferi IgG/IgM ELISA Test kit (predicate device: K033070).

MTTT Comparison:

Comparison with the Predicate Device - MTTT:

The Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit, when used as the first-step or second-step test in combination with another Gold Standard Diagnostics Borrelia burgdorferi ELISA Test Kit in the Modified Two-tier Testing (MTTT) method, was compared to the Standard Two-tier testing (STTT) method using the predicates Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA (K180264) Test Kit as the first-step test followed by testing all the positive and equivocal results on the Gold Standard Diagnostics Borrelia burgdorferi IgG Blot Test (K113847) and the Gold Standard Diagnostics Borrelia burgdorferi IgM Blot Test (K113846). Below are tables comparing the two devices.

Similarities
ItemSubject Device: GoldStandard DiagnosticsBorrelia burgdorferiIgG/IgM ELISA Test KitPredicate Device: GoldStandard DiagnosticsBorrelia burgdorferi IgG Blot(K113847)Predicate Device: GoldStandard Diagnostics Borreliaburgdorferi IgM Blot(K113846)
Intended UseThe Gold StandardDiagnostics Borreliaburgdorferi IgG/IgM ELISATest Kit is intended as aqualitative test for thedetection of IgG and IgMantibodies to Borreliaburgdorferi sensu stricto inhuman serum fromsymptomatic patients orpeople suspected of infection.The Gold Standard DiagnosticsBorrelia burgdorferi B31 IgGLine Blot Test Kit is intendedfor the qualitative detection ofIgG antibodies to B.burgdorferi sensu stricto (B31)in human serum. This test isintended for use in testinghuman serum samples whichhave been found positive orThe Gold Standard DiagnosticsBorrelia burgdorferi B31 IgMLine Blot Test Kit is intendedfor the qualitative detection ofIgM antibodies to B.burgdorferi sensu stricto (B31)in human serum. This test isintended for use in testinghuman serum samples whichhave been found positive or

{11}------------------------------------------------

When used as the first-tierscreening test, positive andequivocal results must beconfirmed through additionaltesting by one of thefollowing methods:equivocal using an ELISA orIFA test procedure to providesupportive evidence ofinfection with B. burgdorferi .equivocal using an ELISA orIFA test procedure to providesupportive evidence of infectionwith B. burgdorferi .
Standard two-tier testmethodology (STTT) usingan IgG and/or IgM blottesting following currentinterpretation guidelines, ORModified two-tier testmethodology (MTTT) usingone or more of the followingthree ELISA based assays:• Gold Standard DiagnosticsBorrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test• Gold Standard DiagnosticsBorrelia burgdorferi IgGELISA Test• Gold Standard DiagnosticsBorrelia burgdorferi IgMELISA Test
The assay can also be used asa second-tier confirmationtest using the MTTTmethodology when used withone or more of the followingthree ELISA based assays:• Gold Standard DiagnosticsBorrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test• Gold Standard DiagnosticsBorrelia burgdorferi IgGELISA Test• Gold Standard DiagnosticsBorrelia burgdorferi IgMELISA Test
Positive test results by eitherthe STTT or MTTTmethodology are supportiveevidence for the presence ofantibodies and exposure toBorrelia burgdorferi , thecause of Lyme disease. Adiagnosis of Lyme diseaseshould be made based on the
presence of Borreliaburgdorferi antibodies,history, symptoms, and otherlaboratory findings.
AntigensB. burgdorferi B31 strain,SameSame
Sample MatrixHuman serumSameSame
Controls ProvidedPositive, Cutoff, NegativeSameSame
Sample ProcessingDilute Samples 1:100SameSame
Assay TypeQualitativeSameSame

{12}------------------------------------------------

Differences
ItemSubject Device: GoldStandard Diagnostics Borreliaburgdorferi IgG/IgM ELISATest KitPredicate Device: GoldStandard Diagnostics Borreliaburgdorferi IgG Blot(K113847)Predicate Device: GoldStandard DiagnosticsBorrelia burgdorferi IgM Blot(K113846)
Assay FormatAntigen coated microtiter plate- 96 wells.Nitrocellulose StripsNitrocellulose Strips
ReagentsProvidedDiluent, Wash, Conjugate,Substrate, Stop SolutionDiluent/Wash, Conjugate,SubstrateDiluent/Wash, Conjugate,Substrate
Volumes100ul sample, 50ul substrate,50ul stop solution1500ul sample, 1500ulsubstrate1500ul sample, 1500ulsubstrate,
Incubation15/15/15 minutes at roomtemperature30/30/10-13 minutes at roomtemperature30/30/10-13 minutes at roomtemperature
InterpretationOptical density readings fromSpectrophotometerVisualVisual
ResultsInterpretationConvert to units.Negative <9Equivocal 9.0-11.0Positive >11.0Compare to cutoff bandCompare to cutoff band
ReportedResultsPositive, Equivocal, NegativePositive, NegativePositive, Negative

Method Comparison MTTT - IgG/IgM

The following studies were conducted to determine the performance of the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test as a first-tier or second-tier assay in the modified two-tier testing (MTTT) methodology.

Gold Standard Diagnostics MTTT-IgG/IgM ELISA Method Comparison: The Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test was utilized in a MTTT (2-ELISA) protocol with the Gold Standard Diagnostics Borrelia burgdorferi VISE-OspC IgG/IgM ELISA Test. The MTTT (2-ELISA) results were compared to the standard two-tier testing (STTT) using the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA followed by testing all positive and equivocal results on the predicate Gold Standard Diagnostics Borrelia burgdorferi IgG blot test and Gold Standard Diagnostics Borrelia burgdorferi IgM blot test.

{13}------------------------------------------------

Prospective Study

Comparison studies were conducted at three sites (one internal and two external reference laboratories) using prospective samples submitted for Lyme serology testing. Four hundred eighty-one (481) serum samples were tested on the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test. A total of 54 positive and equivocal samples were obtained.

In the STTT protocol the samples that were positive or equivocal (n=54) were tested with B. burgdorferi IgG and IgM blot tests. In the MTTT protocol the samples (n=54) were tested on a second ELISA, the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test. In the second-tier ELISA test, positive and equivocal results were considered positive.

The results of the second-tier test of the STTT when compared to the second-tier test of the MTTT, including only the samples that were positive in the first tier, are summarized in the following table:

Predicate IgG & IgMImmunoblots
PositiveNegativeTotal
Gold Standard DiagnosticsBorrelia burgdorferi VlsE -OspC IgG/IgM ELISAPositive361349
Gold Standard DiagnosticsBorrelia burgdorferi VlsE -OspC IgG/IgM ELISANegative055
Gold Standard DiagnosticsBorrelia burgdorferi VlsE -OspC IgG/IgM ELISATotal361854

Positive percent agreement = 100.0% Negative percent agreement = 27.8% 95% CI (90.3% - 100.0%) 95% CI (9.7% - 53.5%)

The results of the MTTT when compared to the STTT, including all samples that were part of the prospective study (n=481), are summarized in the following table:

Predicate STTT- IgG/IgM
PositiveNegativeTotal
Gold Standard DiagnosticsMTTT- IgG/IgMPositive361349
Negative0432432
Total36445481

Positive percent agreement = 100.0% Negative percent agreement = 97.1%

95% CI (90.3% - 100.0%) 95% CI (95.1% - 98.4%)

Sensitivity Study

A sensitivity study was performed on 125 clinically characterized samples encompass early, disseminated, and late stages of Lyme disease. The samples were tested on both the Gold Standard Diagnostics MTTT-IgG/IgM and on the predicate STTT-IgG/IgM. The results are summarized in the following table:

{14}------------------------------------------------

DiseaseStagenGold Standard DiagnosticsMTTT – IgG/IgMPredicateSTTT - IgG/IgM
PositiveorEquivocal% Agreementwith ClinicalDiagnosisPositiveorEquivocal% Agreementwith ClinicalDiagnosis
Early623861.3%3962.9%
Disseminated222090.9%2090.9%
Late414097.6%4097.6%

CDC Reference Panel

A panel of 280 positive and negative specimens from the Centers of Disease Control (CDC) for Lyme disease detection was tested on both the Gold Standard Diagnostics MTTT-IgG/IgM and on the predicate STTT-IgG/IgM. The results are summarized in the following table:

Disease StagenGold Standard Diagnostics MTTT-IgG/IgMPredicate STTT- IgG/IgM
Positive or Equivocal% Agreement with Clinical DiagnosisPositive or Equivocal% Agreement with Clinical Diagnosis
Healthy1000100.0%0100.0%
Early Lyme604676.7%3761.7%
Cardiac Lyme3266.7%266.7%
Neurological Lyme7685.7%685.7%
Late2020100.0%20100.0%
Look-alike Disease*90198.9%495.6%

*infectious mononucleosis, fibromyalgia, multiple sclerosis, rheumatoid arthritis, syphilis and severe periodontitis

8) Conclusion:

From the comparison data, we conclude that the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test is substantially equivalent to the Gold Standard Diagnostics Borrelia burgdorferi IgG Blot Test Kit (K113847) and the Gold Standard Diagnostics Borrelia burgdorferi IgM Blot Test Kit (K113846) when used for the Modified Two-tier Testing (MTTT) Lyme testing.

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).