K Number
K203255
Date Cleared
2021-02-24

(112 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 4K UHD Laparoscope is intended to be used for endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The device is also indication of transanal and transvaginal applications.

Device Description

The 4K UHD Laparoscope is a rigid endoscope, which consists of lighting system and optical imaging system. The 4K UHD Laparoscope has 12 models which are available in two insertion portion widths (5.5 mm and 10 mm), four working lengths (290mm, 300mm, 320mm) and three different directions of view (0°, 30°, 45°). The 12 models are all connected with adaptor 2 for different brand source light. The 4K UHD Laparoscope can be connected to Scivita 4K UHD Camera System, Scivita LED Light Source or Wolf light source and SONY Monitor. The 4K UHD Laparoscope is a reusable device that is cleaned and steam sterilized before first use and each subsequent use.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device meets them, specifically for the "4K UHD Laparoscope" (K203255).

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The document doesn't explicitly present a table of acceptance criteria with reported device performance in the typical sense of numerical thresholds and corresponding results for each. Instead, it describes adherence to recognized standards and qualitative comparisons to a predicate device. The performance is implied by the successful completion of these tests and the determination of substantial equivalence.

However, based on the non-clinical tests mentioned, we can infer some performance aspects and how "acceptance" was framed:

Acceptance Criteria (Inferred from standards and comparisons)Reported Device Performance
Biocompatibility
ISO 10993-5:2009 (Tests for in vitro cytotoxicity)Complies with ISO 10993-5
ISO 10993-10:2010 (Tests for irritation and skin sensitization)Complies with ISO 10993-10 for Sensitization
ISO 10993-10:2010 (Tests for intracutaneous reactivity)Complies with ISO 10993-10 for Intracutaneous Reactivity
Electrical Safety & Essential Performance
IEC 60601-2-18:2009 (Basic safety and essential performance of endoscopic equipment)Complies with IEC 60601-2-18
Endoscope Standards
ISO 8600-1:2015 (General requirements for medical endoscopes)Complies with ISO 8600-1; Field of view meets requirements within 15% deviation.
ISO 8600-5:2005 (Determination of optical resolution)Complies with ISO 8600-5
Image Quality Performance
Depth of FieldRange of 3-200 mm (Proposed device has a wider range than predicate device (6-200 mm), which "gives physicians more options." This is presented as an advantage meeting clinical needs.)
Geometric DistortionTests conducted (results not numerically detailed, but implies acceptable performance for SE)
Color ReproductionTests conducted (results not numerically detailed, but implies acceptable performance for SE)
SNR (Signal-to-Noise Ratio)Tests conducted (results not numerically detailed, but implies acceptable performance for SE)
Intensity UniformityTests conducted (results not numerically detailed, but implies acceptable performance for SE)
Image Frame FrequencyTests conducted (results not numerically detailed, but implies acceptable performance for SE)
System DelayTests conducted (results not numerically detailed, but implies acceptable performance for SE)
Sterilization Efficacy
Steam sterilization / Low-temperature plasma sterilizationSterility Assurance Level (SAL) reached 10^-6 after steam sterilization and low-temperature plasma sterilization.
Labeling
21 CFR Part 801Complied with 21 CFR Part 801

The study described is a non-clinical test program designed to compare the proposed device's performance against recognized standards and a legally marketed predicate device ("ULTRA" Telescopes, K150633, from Olympus Winter & Ibe GmbH). The goal was to establish Substantial Equivalence (SE).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified in the document. The non-clinical tests typically involve a sample of the manufactured device models.
  • Data Provenance: The tests were conducted by Scivita Medical Technology Co., Ltd. (based in China, as per the sponsor identification) as part of their 510(k) submission. The data is prospective from these internal tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This device is a hardware laparoscope, and the performance assessment is based on physical and optical measurements against engineering standards and comparison with a predicate device, not on expert interpretation of output data like medical imaging analysis. Therefore, there's no "ground truth" derived from medical experts in the context of diagnostic performance.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Since the "test set" refers to physical and optical performance characteristics of the device itself (e.g., cytotoxicity, electrical safety, optical resolution, depth of field), expert adjudication methods typically used for interpreting diagnostic outputs (like imaging) are not relevant here. The evaluation involves compliance with established standards and direct measurement/comparison.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This document pertains to the clearance of a 4K UHD Laparoscope, which is a physical endoscopic device, not an AI-powered diagnostic tool. Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study focusing on human reader improvement with AI assistance is not relevant and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This device is a hardware laparoscope. There is no algorithm to be tested in a standalone fashion.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests:

  • Biocompatibility: Ground truth is defined by the absence of toxicity, irritation, or sensitization as per the criteria outlined in ISO 10993 standards.
  • Electrical Safety & Performance: Ground truth is compliance with the requirements of IEC 60601-2-18.
  • Endoscope Standards: Ground truth is compliance with the requirements of ISO 8600-1 and ISO 8600-5.
  • Image Quality Performance: Ground truth is established by quantitative measurements (e.g., depth of field, geometric distortion) and comparison to the predicate device's performance, ensuring similar or improved characteristics that meet clinical needs.
  • Sterilization Efficacy: Ground truth is achieving a Sterility Assurance Level (SAL) of 10^-6.

The overall "ground truth" for the submission is the determination of Substantial Equivalence to the predicate device based on these engineering and performance standards.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device; therefore, there is no "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 24, 2021

Scivita Medical Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.Box 120-119 Shanghai, 200120 China

Re: K203255

Trade/Device Name: 4K UHD Laparoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: December 25, 2020 Received: December 30, 2020

Dear Diana Hong:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen. Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203255

Device Name 4K UHD Laparoscope

Indications for Use (Describe)

The 4K UHD Laparoscope is intended to be used for endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The device is also indication of transanal and transvaginal applications.

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

Tab #6 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: _K203255

    1. Date of Preparation: 12/24/2020
    1. Sponsor Identification

Scivita Medical Technology Co., Ltd.

No. 8, Zhong Tian Xiang, Suzhou Industrial Park, Suzhou, Jiangsu Prov., China.

Establishment Registration Number: Not yet registered.

Contact Person: Ruqin Wu Position: Quality Manager Tel: +86-512-81877788 Fax: +86-512-85187285 Email: wuruqin@scivitamedical.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)

Mid-Link Consulting Co., Ltd.

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: +1-360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: 4K UHD Laparoscope Common Name: Laparoscopes Model: 4K5500, 4K5500R, 4K5530, 4K5530R, 4K5545, 4K5545R, 4K1000, 4K1000R, 4K1030, 4K1030R, 4K1045 and 4K1045R

Regulatory Information Classification Name: Laparoscope, General & Plastic Surgery Classification: II; Product Code: GCJ / NMH / HET; Regulation Number: 21 CFR 876.1500 & 21 CFR 884.1720 Review Panel: General & Plastic Surgery

Indication for use:

The 4K UHD Laparoscope is intended to be used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The device is also indicated for visualization of transanal and transvaginal applications.

Device Description:

The 4K UHD Laparoscope is a rigid endoscope, which consists of lighting system and optical imaging system.

The 4K UHD Laparoscope has 12 models which are available in two insertion portion widths (5.5 mm and 10 mm), four working lengths (290mm, 300mm, 320mm) and three different directions of view (0°, 30°, 45°). The 12 models are all connected with adaptor 2 for different brand source light. The specifications of the proposed device are listed in Table 1.

The 4K UHD Laparoscope can be connected to Scivita 4K UHD Camera System, Scivita LED Light Source or Wolf light source and SONY Monitor. The associated compatible device information is as follows:

Endoscope camera system4K UHD Camera System, Model: ES-CS4K100/100C and ES-CS4K200/100C (K200216)
Scivita light sourceLed Light Source manufactured by Scivita, Model: ES-LS110D (510K Exempt, product code: NTN)
Wolf light sourceVisera 4k UHD Xenon Light Source manufactured by Richard Wolf Medical Instruments Corp., Model: 5123
MonitorMonitor manufactured by SONY, Model: LMD-X2705MD,

{5}------------------------------------------------

LMD-X550MD, LMD-X310MD and LMD-X310MT (K150377)

The 4K UHD Laparoscope is a reusable device that is cleaned and steam sterilized before first use and each subsequent use.

No.ModelDiameterWorking LengthDirections of View
14K55005.5 mm300 mm
24K5500R5.5 mm290 mm
34K55305.5 mm300 mm30°
44K5530R5.5 mm290 mm30°
54K55455.5 mm300 mm45°
64K5545R5.5 mm290 mm45°
74K100010 mm330 mm
84K1000R10 mm320 mm
94K103010 mm330 mm30°
104K1030R10 mm320 mm30°
114K104510 mm330 mm45°
124K1045R10 mm320 mm45°

Table 1 Specifications

    1. Identification of Predicate Devices
      Predicate Device 510(k) Number: K150633 Product Name: "ULTRA" Telescopes Manufacturer: Olympus Winter & Ibe GmbH Models: WA4KL100, WA4KL130, WA4KL145, WA4KL500, WA4KL530 and WA4KL545
    1. Non-Clinical Test Conclusion
      Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
  • ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity;

  • ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization;

  • IEC 60601-2-18:2009 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment

  • ISO 8600-1:2015 Endoscopes - Medical endoscopes and endotherapy devices -- Part 1: General

{6}------------------------------------------------

requirements

  • ISO 8600-5:2005 Optics and photonics - Medical endoscopes and endotherapy devices - Part 5: Determination of optical resolution of rigid endoscopes with optics
    In addition, the image quality performance tests were conducted to quantitatively compare the proposed device and predicate device in terms of depth of field, geometric distortion, color reproduction and SNR, intensity uniformity, image frame frequency and system delay.

    1. Clinical Test Conclusion
      No clinical study is included in this submission.
    1. Substantially Equivalent (SE) Comparison
ITEMProposed DevicePredicate Device K150633Remark
Product CodeGCJ / NMH / HETGCJ / NMH / HETSame
Regulation21 CFR 876.150021 CFR 876.1500Same
No.21 CFR 884.172021 CFR 884.1720Same
ClassIIIISame
Indication forUseThe 4K UHD Laparoscope is intended tobe used for endoscopy and endoscopicsurgery within the thoracic andperitoneal cavities including the femalereproductive organs. The device is alsoindicated for visualization of transanaland transvaginal applications.These endoscopes are intended tobe used for endoscopy andendoscopic surgery within thethoracic and peritoneal cavitiesincluding the female reproductiveorgans. The device is alsoindicated for visualization oftransanal and transvaginalapplications.Same
Single use /ReusableReusableReusableSame
FeatureRigid endoscopeRigid endoscopeSame
MechanismViewing OpticsViewing OpticsSame
Principle ofoperationDuring operation, the external lightsource is transmitted to the front end ofthe laparoscope through the lightingsystem of the laparoscope to illuminatethe area to be observed. The object isimaged on the imaging system throughthe objective lens, and then transmittedDuring operation, the externallight source is transmitted to thefront end of the laparoscopethrough the lighting system ofthe laparoscope to illuminate thearea to be observed. The objectis imaged on the imaging systemSame
to the eyepiece for direct observationby the human eye, or magnified by anexternal camera system and observedon the monitor.through the objective lens, andthen transmitted to the eyepiecefor direct observation by thehuman eye, or magnified by anexternal camera system andobserved on the monitor.
OuterDiameter5.5 mm and 10 mm5.4 mm and 10 mmDifferent
WorkingLength4K5500, 4K5530, 4K5545: 300mm4K5500R, 4K5530R, 4K5545R: 290mm4K1000, 4K1030, 4K1045: 330mm4K1000R, 4K1030R, 4K1045R: 320mmWA4KL500: 315.3±0.5mmWA4KL530: 315.5±1.5mmWA4KL545: 315.5±1.5mmWA4KL100: 315.9±0.5mmWA4KL130: 317.1±1.5mmWA4KL145: 317.7±1.5mmDifferent
Direction ofView0°, 30°, 45°0°, 30°, 45°Same
Field of View75°88°Different
Depth of Field3-200 mm6-200 mmDifferent
Patient Contact Material
Objectiveglass coverSapphireSapphire
Optical fiberemissionsurfaceLight Guide fibreLight Guide fibreSame
Metal fillerAISI 304 Medical stainless steelMedical stainless steel
AdhesivesEpoxy resin fiber optic adhesive (Epotek353ND)Epoxy resin fiber optic adhesive
Main partAISI 304 Medical stainless steelMedical stainless steel
ThermalSafetyIEC 60601-2-18IEC 60601-2-18Same
Biocompatibility
CytotoxicityComply with ISO 10993-5Comply with ISO 10993-5Same
SensitizationComply with ISO 10993-10UnknownDifferent
IntracutaneousReactivityComply with ISO 10993-10UnknownDifferent
Sterilization
MethodSteam sterilization / Low-temperatureplasma sterilizationAutoclaved sterilizationDifferent
LabelingComplied with 21 CFR Part 801Complied with 21 CFR Part 801Same

Table 2 General Comparison

{7}------------------------------------------------

K203255

{8}------------------------------------------------

Different - Outer Diameter

Both the proposed device and the predicate device provide two options for outer diameter. Although the declared size of the outer diameter of the proposed device is not exactly the same as that of the predicate device, the difference is too slight to cause safety and performance issues. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

Different - Working Length

Although the working length of the proposed device is not the same as that of the predicate device, the proposed device offers a wider range of options, allowing physicians to make choices based on the patient's condition. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

Different - Field of View

Field of view, namely the view of an endoscope with optics, refers to the maximum range that the endoscope can observe at the same time, that is, the cone angle with the vertex at the end of the endoscope head. Although the field of view of the proposed device is different from that of predicate device, it can meet the needs of clinical diagnosis and treatment. In addition, the field of view of the proposed device meets the requirements of ISO 8600-1 with a deviation of not greater than 15%. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

Different - Depth of Field

The proposed device has a wider range of depth of field than the predicate device to give physicians more options for diagnosis and treatment based on the patient's condition. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

Different - Sensitization and Intracutaneous Reactivity

Although the results of sensitization and intracutaneous reactivity of the predicate device are unknown, the proposed device did both biocompatibility tests. The results showed that the proposed device posed no risk to humans. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

Different - Sterilization

{9}------------------------------------------------

Although the sterilization method of the proposed device is different from that of the predicate device, the sterilization verification is performed for the proposed device. The results showed that the Sterility Assurance Level (SAL) reached 10 after steam sterilization and low-temperature plasma sterilization. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

    1. Substantially Equivalent (SE) Conclusion
      Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.