K Number
K203216
Date Cleared
2020-11-25

(23 days)

Product Code
Regulation Number
892.1750
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Multimodality Advanced Vessels Analysis (MM AVA) application is intended for visualization, assessment and quantification of vascular datasets.

Device Description

Philips Medical Systems' Multimodality Advanced Vessels Analysis (MM AVA) application is intended for visualization, assessment, and quantification of vessels in CTA and MRA data with a unified workflow for both modalities. For CTA data, it provides both automatic and manual bone removal and vessels segmentation including extraction of vessel centerlines, lumen contours and vessel contours. For both modalities, it provides tools for extracting and editing centerlines. MM AVA offers inspection views for selected vessels centerlines and local analysis. It allows creating, capturing, and reviewing of basic user selected endovascular measurements (and calculations when applicable), as well as predefined measurements sets and measurements correlations. The physician retains the ultimate responsibility for making the final diagnosis. Philips Medical Systems' MM AVA is launched from Philips Medical Systems' IntelliSpace Portal (ISP) Platform (K162025).

AI/ML Overview

The provided document outlines the substantial equivalence of Philips Medical Systems' Multimodality Advanced Vessel Analysis (MM AVA) application to existing predicate devices. However, this submission does not contain details about specific acceptance criteria, study methodologies (like sample size for test and training sets, data provenance, expert qualifications, or adjudication methods), or performance metrics demonstrating the device meets quantitative acceptance criteria.

The document states:

  • "The test results in this 510(k) premarket notification demonstrate that Multimodality Advanced Vessels Analysis (MM AVA) application: ... Meets the acceptance criteria and is adequate for its intended use and specifications." (Page 17)
  • It also mentions that "Philips Medical Systems' Multimodality Advanced Vessels Analysis (MM AVA) application was tested in accordance with Philips verification and validation processes. Verification and Validation tests have been performed to address intended use, the technological characteristics, requirement specifications and the risk management results." (Page 17)

However, the specifics of these "acceptance criteria" and the "test results" demonstrating they were met are not provided in the publicly available sections of this 510(k) summary. This summary focuses on establishing substantial equivalence based on features, intended use, and existing standards rather than detailed performance study results against predefined objective metrics.

Therefore, I cannot populate the requested table or answer the specific questions about the study design that proves the device meets the acceptance criteria, as that information is not present in the provided text.

Based on the available text, here's what can be stated:

1. Table of acceptance criteria and the reported device performance:

Acceptance Criteria (e.g., Specificity, Sensitivity, Accuracy, etc.)Reported Device Performance
Not specified in the provided document.Not specified in the provided document.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not specified in the provided document.
  • Data Provenance: Not specified in the provided document (e.g., country of origin, retrospective or prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: Not specified in the provided document.
  • Qualifications of Experts: Not specified in the provided document.

4. Adjudication method for the test set:

  • Adjudication Method: Not specified in the provided document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC Study Done: Not specified whether an MRMC study was performed in the context of human-in-the-loop performance. The document focuses on the software's features and its substantial equivalence to predicate devices, not on human reader improvement with AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Standalone Performance Study: The document states "Non-clinical performance tests has been performed on Multimodality Advanced Vessels Analysis (MM AVA) application" and "Philips Medical Systems Multimodality Advanced Vessels Analysis (MM AVA) application was tested in accordance with Philips verification and validation processes." This implies standalone testing, but specific performance metrics (e.g., accuracy of automatic measurements) are not detailed.

7. The type of ground truth used:

  • Type of Ground Truth: Not specified in the provided document.

8. The sample size for the training set:

  • Sample Size for Training Set: Not specified in the provided document.

9. How the ground truth for the training set was established:

  • Ground Truth Establishment for Training Set: Not specified in the provided document.

In summary, the provided 510(k) summary focuses on demonstrating substantial equivalence through feature comparison and compliance with general standards and internal validation processes, rather than presenting detailed performance study results that would typically include the requested information.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".

November 25, 2020

Philips Medical Systems Nederland B.V. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114

Re: K203216

Trade/Device Name: The Multimodality Advanced Vessel Analysis (MM AVA) application Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: October 28, 2020 Received: November 2, 2020

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203216

Device Name

The Multimodality Advanced Vessel Analysis (MM AVA) application

Indications for Use (Describe)

The Multimodality Advanced Vessels Analysis (MM AVA) application is intended for visualization, assessment and quantification of vascular datasets.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. Above the word "PHILIPS" is the alphanumeric code "K203216" in a smaller, black font. The background of the image is white. The text is centered.

510(k) Summary

The Multimodality Advanced Vessel Analysis (MM AVA) Application

  • Date Prepared: October 6, 2020

I. Submitter's name and address

Manufacturer:Philips Medical Systems Nederland B.V.Veenpluis 4-65684 PC BestThe NetherlandsEstablishment Registration Number: 3003768277
------------------------------------------------------------------------------------------------------------------------------------------------------------
Primary Person:Contact Vered NitzanRegulatory Affairs LeadPhone: +972 54-9797047E-mail: vered.nitzan@Philips.com
--------------------------------------------------------------------------------------------------------------------------------

II. Device information

Device:
Trade Name:The Multimodality Advanced Vessel Analysis(MM AVA) application
Classification Name:Computed tomography x-ray system
Classification Regulation:21 CFR 892.1750
Classification Panel:Radiology
Device Class:II
Primary Product Code:JAK

III. Device Description

Philips Medical Systems' Multimodality Advanced Vessels Analysis (MM AVA) application is intended for visualization, assessment, and quantification of vessels in CTA and MRA data with a unified workflow for both modalities. For CTA data, it provides both automatic and manual bone

{4}------------------------------------------------

removal and vessels segmentation including extraction of vessel centerlines, lumen contours and vessel contours. For both modalities, it provides tools for extracting and editing centerlines.

MM AVA offers inspection views for selected vessels centerlines and local analysis. It allows creating, capturing, and reviewing of basic user selected endovascular measurements (and calculations when applicable), as well as predefined measurements sets and measurements correlations.

The physician retains the ultimate responsibility for making the final diagnosis.

Philips Medical Systems' MM AVA is launched from Philips Medical Systems' IntelliSpace Portal (ISP) Platform (K162025).

Kev Features:

    1. Provides automatic bone removal for CTA data
    1. Provides automatic vessel centerline extraction and labeling for the main vessels (CTA data)
    1. Provides a set of centerline creation and editing tools (manual and semi-automatic)
    1. Provides dedicated views for vessels review: Curved MPRS, Straighten MPR, Crosssectional and Longitudinal MPR
    1. Provides 3D Volume rendering capabilities for CTA and MRA, including different volumetric presentation options such as Volume rendering, MIP (Maximum Intensity Projection) etc.
    1. Allows measurements along a vessel centerline, based on a user selected location/s such as Max Diameter, Min Diameter and Area
    1. Allows the calculation of measurements such as: Length, Angle, Tortuosity, Stenosis, Aneurysm etc. based on user selected locations and their intra-correlations.
    1. Allows navigation and local inspection of images, including capturing measurements such as diameters (Quick Inspection)
    1. Allows Batch (series of images processed by user) generation for any of the selected user views.

IV. Intended Use and Indications for Use:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.

The Multimodality Advanced Vessels Analysis (MM AVA) application is intended for visualization, assessment, and quantification of vascular datasets.

V. Predicate Devices:

The following table shows the predicate devices of the proposed Philips Medical Systems Multimodality Advanced Vessels Analysis (MM AVA) application:

Device NameManufacturer510k No
Primary PredicateBrilliance VolumePhilips Medical Systems (Cleveland), Inc.K060937
Secondary Predicatesyngo.CT Vascular AnalysisSiemens Medical Solutions, Inc.K112020
Secondary Predicatesyngo.MR VascularSiemens Medical Solutions, Inc.K130749

The proposed Philips Medical Systems Multimodality Advanced Vessels Analysis (MM AVA) application and its predicate devices, Brilliance Volume (K060937), syngo.CT Vascular Analysis (K112020) and syngo.MR Vascular (K130749) are equivalent regarding their intended uses, clinical indications, principle of operation and fundamental technology.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. The background is plain white, which makes the blue text stand out.

VI. Substantial Equivalence to Predicate Devices:

FeatureThe proposed device:MM AVA ApplicationPrimary Predicate:Brilliance Volume(K060937)Secondary Predicate:Siemens syngo.CTVascular Analysis(K112020)Secondary Predicate:Siemens syngo.MRVascular(K130749)
Device ClassificationNameSystem, X-Ray,Tomography, ComputedSystem, X-Ray,Tomography, ComputedSystem, X-Ray,Tomography, ComputedSystem, Image Processing,Radiological
Device ClassClass IIClass IIClass IIClass II
Classification PanelRadiologyRadiologyRadiologyRadiology
Product CodeJAKJAKJAKLLZ, LNH
Regulation DescriptionComputed tomography x-ray systemComputed tomography x-ray systemComputed tomography x-ray systemPicture Archiving andCommunication System(PACS)
Regulation Number21 CFR 892.175021 CFR 892.175021 CFR 892.175021 CFR 892.2050
Indication for UseThe MultimodalityAdvanced Vessels Analysis(MM AVA) application isThe "Brilliance Volume" isa Computed TomographyX-Ray System intended tosyngo CT VascularAnalysis is an imageanalysis software packageThe software comprisingthe syngo.MR post-processing applications are
The proposed device:Primary Predicate:Secondary Predicate:Secondary Predicate:
FeatureMM AVA ApplicationBrilliance Volume(K060937)Siemens syngo.CTVascular Analysis(K112020)Siemens syngo.MRVascular(K130749)
intended for visualization,assessment andquantification ofvascular datasets.producecross-sectional images ofthe body by computerreconstruction of x-raytransmission datataken at different anglesand planes. This devicemay include signal analysisand displayequipment, patient, andequipment supports,components andaccessories.for evaluating enhanced CTimages. Combining digitalimage processing andvisualization tools(multiplanar reconstruction(MPR) thin/thick,maximum intensityprojection (MIP) thin/thick,inverted MIP thin/thick,volume renderingtechnique (VRT), curvedplanar reformation (CPR),processing tools (boneremoval (based both onsingle energy and DualEnergy), table removal)and evaluation tools (vesselcenterline calculation,lumen calculation, stenosiscalculation) and reportingtools (lesion location,lesion characteristics andpost-processing software /applications to be used forviewing and evaluating thedesignated imagesprovided by a magneticresonance diagnosticdevice.syngo.MR Vascular is asyngo based post-processing software forviewing, manipulating, andevaluating MR vascularimages.
FeatureThe proposed device:MM AVA ApplicationPrimary Predicate:Brilliance Volume(K060937)Secondary Predicate:Siemens syngo.CTVascular Analysis(K112020)Secondary Predicate:Siemens syngo.MRVascular(K130749)
key images), the software package is designed to support the physician in confirming the presence or absence of physician-identified lesions in blood vessels and evaluation, documentation and follow-up of any such lesion. These visualization/processing/evaluation tools allow for characterization of vascular lesions and lesion size over time, helping the physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue.
The proposed device:Primary Predicate:Secondary Predicate:Secondary Predicate:
FeatureMM AVA ApplicationBrilliance Volume(K060937)Siemens syngo.CTVascular Analysis(K112020)Siemens syngo.MRVascular(K130749)
Intended body partHead and neck, body,peripheralsHead and neck, body,peripheralsHead and neck, body,peripheralsHead and neck, body,peripherals
Type of scansCT Angiography and MRAngiographyCT AngiographyCT AngiographyMR Angiography
Automatic bone removalfor CTA dataYesYesYes(Including Dual Energy)NA
FeatureThe proposed device:MM AVA ApplicationPrimary Predicate:Brilliance Volume(K060937)Secondary Predicate:Siemens syngo.CTVascular Analysis(K112020)Secondary Predicate:Siemens syngo.MRVascular(K130749)
Semi-automatic boneremoval for CTA dataYesYesYesNo
Bone Removal Edit forCTA dataManual interactions forcorrection of boneremovalresults.YesYesYesNA
Automatic vesselcenterline extractionYes (CT only)YesYesNo
Automatic vessel labelingYes (CT only)YesYesNo
Manual Vessel TrackingManual drawing andediting of vascular pathsfor curved MPRYes (manual and semi-automatic)YesYesYes
FeatureThe proposed device:MM AVA ApplicationPrimary Predicate:Brilliance Volume(K060937)Secondary Predicate:Siemens syngo.CTVascular Analysis(K112020)Secondary Predicate:Siemens syngo.MRVascular(K130749)
visualization
Vessel TrackingTools for semi-automatictracking and editing ofvascular structures.YesYesYesYes
Vessel centerline andlumen contoursextractionYesYesYesYes
(Semi) automaticsegmentation of vesselanatomyYesYesYesYes
Curved MPR (CPR) andCross Section MPRVisualization of vessels incurved MPR withcorresponding cross-sectional Images.YesYesYesYes
FeatureThe proposed device:MM AVA ApplicationPrimary Predicate:Brilliance Volume(K060937)Secondary Predicate:Siemens syngo.CTVascular Analysis(K112020)Secondary Predicate:Siemens syngo.MRVascular(K130749)
Curved reformat vesselviewsYesYesYesYes
Vessel aligned MPR views(Straighten MPR)YesYesYesYes
3D volumetric vesselviewsYesYesYesYes
Diameter measurementYesYesYesYes
Stenosis MeasurementStenosis values based onlumen contouring.YesYesYesYes
Aneurysm measurementYesYesNot specifiedNot specified
3D vascularmeasurements (diameter,stenosis, aneurysm)YesYesYesYes
FeatureThe proposed device:MM AVA ApplicationPrimary Predicate:Brilliance Volume(K060937)Secondary Predicate:Siemens syngo.CTVascular Analysis(K112020)Secondary Predicate:Siemens syngo.MRVascular(K130749)
MeasurementseditingFunctionality for editingvalues of clinical findingssuch as location,pathology, etc.YesYesYesYes
Lumen evaluationDelineation of vesselcontours for vascularanalysisYesYesYesYes
Navigate Along VesselSemi-automatic alignmentof MPR views to localvascular anatomy withinteractive navigationalongvessel course.YesYesYesYes
FeatureThe proposed device:MM AVA ApplicationPrimary Predicate:Brilliance Volume(K060937)Secondary Predicate:Siemens syngo.CTVascular Analysis(K112020)Secondary Predicate:Siemens syngo.MRVascular(K130749)
Review MarkerFunctionality for settinglocation of a clinicalfinding on any view.YesYesYesYes
Curved and Cross-Sectional RangesCreation of MPR seriesaround and along vesselspath.YesYesYesYes
Basic ReadingFunctionalityConventional navigationon 2D and 3D views,change of layouts, adaptwindow values.YesYesYesYes
Bone Opacity for CTAdataYesYesYesNA
FeatureThe proposed device:MM AVA ApplicationPrimary Predicate:Brilliance Volume(K060937)Secondary Predicate:Siemens syngo.CTVascular Analysis(K112020)Secondary Predicate:Siemens syngo.MRVascular(K130749)
Table (couch) RemovalSegmentation of patienttable for clear VRTvisualization of anatomy.YesYesYesNA
Angio ViewInverted MIP visualizationof vessels filled withcontrast agent.YesYesYesYes
Calcification Removal forCTA dataMasking of high intensitystructures along vesselsfor true lumenvisualization.YesYesYes(also for dual energy)NA
DICOM compatibleYesYesYesYes

Premarket Notification [510(K)] Submission
Philips Medical Systems Nederland B.V.

Section 5-1 Page 5 of 16

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Premarket Notification [510(k)] Submission
Philips Medical Systems Nederland B.V.

Section 5-1
Page 9 of 16

{11}------------------------------------------------

Premarket Notification [510(k)] Submission
Philips Medical Systems Nederland B.V.

Section 5-1
Page 10 of 16

{12}------------------------------------------------

{13}------------------------------------------------

Premarket Notification [510(k)] Submission
Philips Medical Systems Nederland B.V.

Section 5-1 Page 12 of 16

{14}------------------------------------------------

Premarket Notification [510(k)] Submission
Philips Medical Systems Nederland B.V.

Section 5-1 Page 13 of 16

{15}------------------------------------------------

Premarket Notification [510(k)] Submission
Philips Medical Systems Nederland B.V.

Section 5-1
Page 14 of 16

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white.

The proposed Philips Medical Systems Multimodality Advanced Vessels Analysis (MM AVA) application and its predicate devices, Brilliance Volume (K060937), syngo.CT Vascular Analysis (K112020) and syngo.MR Vascular (K130749) are substantially equivalent in regard to their intended use, clinical indications, principle of operation and fundamental technology.

In conclusion, Philips believes that the Multimodality Advanced Vessels Analysis (MM AVA) application does not introduce any new potential safety and/or effectiveness issues and is substantially equivalent to the identified predicate devices, Brilliance Volume (K060937), syngo.CT Vascular Analysis (K112020) and syngo.MR Vascular (K130749).

VII. Nonclinical Tests performed:

Non-clinical performance tests has been performed on Multimodality Advanced Vessels Analysis (MM AVA) application and demonstrates compliance with the following International and FDArecognized consensus standards:

  • ISO 14971 Medical devices Application of risk management to medical devices
  • IEC 62304 Medical device software Software life cycle processes
  • NEMA PS 3.1-3.21 Digital Imaging and Communications in Medicine (DICOM) Standard
  • IEC 62366-1 Medical devices Part 1: Application of usability engineering to medical devices

Philips Medical Systems Multimodality Advanced Vessels Analysis (MM AVA) application was tested in accordance with Philips verification and validation processes. Verification and Validation tests have been performed to address intended use, the technological characteristics, requirement specifications and the risk management results.

The test results in this 510(k) premarket notification demonstrate that Multimodality Advanced Vessels Analysis (MM AVA) application:

  • . Complies with the aforementioned international and FDA-recognized consensus standards and FDA guidance documents, and
  • Meets the acceptance criteria and is adequate for its intended use and specifications. .

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.

VIII. Overall Conclusion:

The Multimodality Advanced Vessels Analysis (MM AVA) application is substantially equivalent to the identified predicate devices, Brilliance Volume (K060937), syngo.CT Vascular Analysis (K112020) and syngo.MR Vascular (K130749) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, verification and validation tests demonstrate the safety and efficacy of the device to meet its intended use and specifications.

Philips Medical believes that the proposed device, Multimodality Advanced Vessels Analysis (MM AVA) application, is substantially equivalent to its identified predicate devices and is as safe and effective as its predicate devices without raising any new safety and/or effectiveness concerns.

N/A