(188 days)
The NaviCam Xpress Stomach System is intended for visualization of the stomach of adults (≥22 years) with BMI
The NaviCam Xpress Stomach System is a endoscopic capsule imaging system intended to obtain images of the stomach. It differs from passive capsule endoscopy systems in that it uses magnetic fields to allow the position of the capsule within the stomach to be controlled by an operator. The NaviCam Xpress Stomach System includes the following key components: Ingestible capsule (AKEM-11SW) for obtaining images, Data recorder (AKR-1) for logging image data, Locator (AKS-1) for turning on the capsule and for determining if the capsule is still in the body, Controller (NaviEC-2000) with the NaviCtrl software that allows the navigation of the capsule within the stomach, ESView software for review of the images obtained by the capsule and generating reports.
The provided text is a 510(k) summary for the NaviCam Xpress Stomach Capsule Endoscope System. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain the acceptance criteria or a study proving the device meets those criteria, especially in terms of diagnostic performance or clinical accuracy. The "Performance Data" section solely discusses bench/in-vitro testing related to the controller's functionality and compliance with electrical safety standards, not the diagnostic performance of the capsule endoscope system itself.
Therefore, most of the requested information regarding acceptance criteria, performance, sample sizes, expert involvement, and ground truth for a diagnostic study cannot be extracted from this document.
Here's what can be inferred or explicitly stated based on the provided text, alongside what is missing:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance |
---|---|
Diagnostic Performance: (e.g., Sensitivity, Specificity, Accuracy for detecting gastric abnormalities) | NOT PROVIDED IN THIS DOCUMENT. The document focuses on the controller's performance and system compliance, not the diagnostic capabilities of the capsule endoscope. |
NaviCam Xpress Controller meets performance requirements in specifications and implements software functions. | Passed: Verified that the NaviCam Xpress Controller meets its specifications and implements software functions as stated in its User Manual. |
Magnetic field intensity range meets specifications with various examination beds. | Passed: Verified that the magnetic field intensity range meets specifications when used with examination beds/tables provided by users. |
Mechanical motion range meets specifications. | Passed: Verified that the mechanical motion range meets specifications. |
Titanium clips do not affect data communication, and clips do not move during magnetic head rotation. | Passed: Data communication not affected, and clip position does not change. |
Compliance with IEC 60601-1, ANSI/AAMI ES60601-1, CAN/CSAC22.2 No. 60601-1:14, EN 60601-1:2006, AAMIIEC 60601-1-2:2014, EN 60601-1-2:2015. | Passed: The system was found to comply with all applicable requirements of these standards. |
The following information cannot be extracted from the provided text:
- 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- 8. The sample size for the training set
- 9. How the ground truth for the training set was established
Reasoning for Missing Information:
This 510(k) summary focuses on demonstrating substantial equivalence for a modification to an already cleared device (the NaviCam Stomach System). The "Performance Data" section details bench/in-vitro testing and compliance with electrical safety and EMC standards. It explicitly states, "The NaviCam Xpress Stomach System successfully passed the following test," and then lists functional and safety-related tests for the controller. It does not discuss any clinical diagnostic performance studies comparing the NaviCam Xpress Stomach System's ability to detect stomach abnormalities against a ground truth, which would involve patient data, expert review, and metrics like sensitivity or specificity.
§ 876.1310 Magnetically maneuvered capsule endoscopy system.
(a)
Identification. A magnetically maneuvered capsule endoscopy system consists of an ingestible capsule and magnetic controller and is used for visualization of the stomach and duodenum. The ingestible capsule contains a camera that wirelessly captures images of the mucosa. The magnetic controller is used outside of the patient and is magnetically coupled with the capsule to control its location and viewing direction.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing with the device under anticipated conditions of use must evaluate visualization of the intended region and document the adverse event profile.
(2) Non-clinical testing data must demonstrate the optical, mechanical, and functional integrity of the device under physically stressed conditions. The following performance characteristics must be tested, and detailed protocols must be provided for each test:
(i) A bite test must be performed to ensure that the capsule can withstand extreme cases of biting;
(ii) A pH resistance test must be performed to evaluate integrity of the capsule when exposed to a physiological relevant range of pH values;
(iii) A battery life test must be performed to demonstrate that the capsule's operating time is not constrained by the battery capacity;
(iv) A shelf life test must be performed to demonstrate that the device performs as intended at the proposed shelf life date;
(v) Optical testing must be performed to evaluate fundamental image quality characteristics such as resolution, field of view, depth of field, geometric distortion, signal to noise ratio, dynamic range, and image intensity uniformity;
(vi) A color performance test must be performed to compare the color differences between the input scene and output image;
(vii) A photobiological safety analysis must be performed based on maximum (worst-case) light exposure to internal gastrointestinal mucosa, and covering ultraviolet, visible, and near-infrared ranges, as appropriate. A mitigation analysis must be provided;
(viii) Performance testing must demonstrate that the viewing software clearly presents the current frame rate, which is either adjustable manually by the user or automatically by the device. Testing must demonstrate that the viewing software alerts the user when the video quality is reduced from nominal due to imaging data communication or computation problems;
(ix) A data transmission test must be performed to verify the robustness of the data transmission between the capsule and the receiver. This test must include controlled signal attenuation for simulating a non-ideal environment; and
(x) Magnetic field strength testing characterization must be performed to identify the distances from the magnet that are safe for patients and users with ferromagnetic implants, devices, or objects.
(3) Software validation, verification, and hazard analysis must be provided.
(4) Electrical safety, thermal safety, mechanical safety, and electromagnetic compatibility testing must be performed.
(5) The patient-contacting components of the device must be demonstrated to be biocompatible.
(6) Performance data must validate the reprocessing instructions for the reusable components of the device.
(7) Performance data must demonstrate the sterility of any device components labeled sterile.
(8) Human factors testing must demonstrate that the intended users can safely and correctly use the device, based solely on reading the instructions for use.
(9) Clinician labeling must include:
(i) Specific instructions and the clinical and technical expertise needed for the safe use of the device;
(ii) A detailed summary of the clinical testing pertinent to use of the device, including information on effectiveness and device- and procedure-related complications;
(iii) The patient preparation procedure;
(iv) A detailed summary of the device technical parameters;
(v) Magnetic field safe zones;
(vi) A screening checklist to ensure that all patients and operating staff are screened from bringing ferromagnetic implants, devices, or objects near the external magnet;
(vii) Reprocessing instructions for reusable components;
(viii) Shelf life for single use components; and
(ix) Use life for reusable components.
(10) Patient labeling must include:
(i) An explanation of the device and the mechanism of operation;
(ii) The patient preparation procedure;
(iii) A brief summary of the clinical study; and
(iv) A summary of the device- and procedure-related complications pertinent to use of the device.