(765 days)
The Eonis™ SCID-SMA kit is intended for the semi-quantitative determination of TREC (T-cell receptor excision circle) as an aid in screening newborns for Severe Combined Immunodeficiency (SCID) and for the semi-quantitative determination of KREC (Kappa-deleting recombination excision circle) as an aid in screening newborns for X-linked agammaglobulinemia (XLA). The test is intended for DNA from blood specimens dried on a filter paper and for use on the QuantStudio™ Dx Real-Time PCR instrument.
This test is not intended for screening of SCID-like Syndromes, such as DiGeorge Syndrome, or Omenn Syndrome. lt is also not intended to screen for less acute SCID syndromes such as leaky-SCID or variant SCID. The test is not indicated for screening B-cell deficiency disorders other than XLA, such as atypical XLA, or for screening of XLA carriers.
This test is not intended for use as a diagnostic test and a positive screening result should be followed by confirmatory testing.
The Eonis SCID-SMA kit is a multiplex real-time PCR-based assay. It uses target sequence-specific primers and TaqMan™ probes to amplify and detect three targets: TREC, and RPP30, in the DNA extracted from newborn dried blood spot (DBS) using Eonis DNA Extraction kit in a single PCR reaction.
Each Eonis SCID-SMA kit contains reagents for up to 384 reactions (for 3241-001U) or 1152 reactions (for 3242-001U) including kit controls.
The document describes the Eonis SCID-SMA kit, a real-time PCR-based assay for newborn screening of Severe Combined Immunodeficiency (SCID) and X-linked agammaglobulinemia (XLA). The study provided demonstrates the device's analytical and screening performance to support its substantial equivalency to a predicate device.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" as a separate table. However, it presents Sensitivity and Specificity for both TREC and KREC analytes, which serve as key performance metrics. These values are compared to the predicate device.
Reported Device Performance of Eonis SCID-SMA Kit:
| Analyte | Metric | Percent | Confidence Limits |
|---|---|---|---|
| TREC | Sensitivity | 100 % | 80.5 % - NA |
| False-negative rate | 0 % | NA - 19.5 % | |
| Specificity | 99.7 % | 99.4 % - 99.9 % | |
| False-positive rate | 0.3 % | 0.1 % - 0.6 % | |
| KREC | Sensitivity | 100 % | 54.1 % - NA |
| False-negative rate | 0 % | NA - 45.9 % | |
| Specificity | 99.7 % | 99.4 % - 99.9 % | |
| False-positive rate | 0.3 % | 0.1 % - 0.6 % |
Comparison to Predicate Device (PerkinElmer EnLite Neonatal TREC Kit) for TREC:
| Analyte | Metric | Percent | Confidence Limits |
|---|---|---|---|
| TREC | Sensitivity | 100 % | 79.4 % - NA |
| False-negative rate | 0 % | NA - 20.6 % | |
| Specificity | 99.7 % | 99.4 % - 99.8 % | |
| False-positive rate | 0.3 % | 0.2 % - 0.6 % |
The reported performance clearly aims to meet or exceed the performance of the predicate device, demonstrating 100% sensitivity and high specificity for both analytes.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Screening Performance Study (Test Set):
- Total DBS specimens: 3090
- Confirmed SCID positive: 17
- Confirmed XLA positive: 6
- Normal newborn screening specimens: 3018 (retrospective archived)
- Data Provenance: Retrospective archived dried blood spot specimens.
- Country of Origin: US and Denmark.
- Study conducted in Denmark.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: Not explicitly stated as a specific number. The document mentions "clinical experts" were used.
- Qualifications of Experts: The document states "clinical experts" retrospectively reviewed the clinical status of routine subjects to confirm they were from unaffected individuals. Further specific qualifications (e.g., specific medical specialty, years of experience) are not provided in this document.
4. Adjudication Method for the Test Set
- Adjudication Method: The document describes a retesting protocol for initial "screen positive" results.
- "The specimens having TREC and KREC levels below the cut-off values in the initial round of testing were re-tested in duplicate."
- "The final results (presumptive positive, invalid result) were classified after the second round of testing."
- This implies a form of internal re-adjudication based on duplicate retesting for samples near the cut-off. There is no mention of external expert adjudication for discordant results or a specific "X+Y" type of adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- MRMC Study: No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic (IVD) kit for semi-quantitative determination of biomarkers, not an AI assisting human readers of medical images. Therefore, the concept of human readers improving with AI assistance is not applicable to this type of device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Standalone Performance: Yes, the entire performance data regarding sensitivity, specificity, reproducibility, precision, limit of detection, and linearity are based on the standalone performance of the Eonis SCID-SMA kit (the algorithm of the kit combined with the instrument) on dried blood spot samples. This device operates as an automated assay, therefore, its performance is inherently "standalone" in terms of its analytic and clinical validity.
7. The Type of Ground Truth Used
- Ground Truth Type:
- Confirmatory testing: For SCID and XLA positive cases, "Confirmatory test results were used as the comparator." This implies clinical diagnosis or gold standard laboratory tests.
- Clinical expert retrospective review: For normal newborn screening specimens, "The clinical status of the routine subjects was determined through a retrospective review by clinical experts to confirm the routine subject cohort samples were from unaffected individuals." This indicates clinical outcomes or medical records adjudicated by experts.
8. The Sample Size for the Training Set
- Training Set Sample Size: The document does not explicitly state the sample size of a separate "training set" for the assay. The study described is a clinical validation (test set). For assay development (which would include "training" for establishing parameters like cut-offs), the document mentions:
- Cut-off values were established using "an independent dataset." The size of this independent dataset is not specified.
- Reproducibility and precision studies used panels of dried blood spots at different TREC/KREC levels, but these are for analytical validation rather than establishing classification criteria.
9. How the Ground Truth for the Training Set Was Established
- Training Set Ground Truth Establishment: As no specific "training set" is detailed, the method for establishing ground truth for any data used during the assay's development or cut-off determination (the "independent dataset" mentioned for cut-off study) is not explicitly described. However, it's reasonable to infer that similar methods to the test set ground truth (confirmatory testing for affected individuals and clinical review for unaffected individuals) would have been applied during the development phase.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
November 9, 2022
PerkinElmer Inc Casey Fox, Ph.D. Sr. Manager, Regulatory Affairs 940 Winter St Waltham, MA 02451
Re: K203035
Trade/Device Name: Eonis SCID-SMA kit Regulation Number: 21 CFR 866.5930 Regulation Name: Newborn screening test for severe combined immunodeficiency disorder (SCID) Regulatory Class: Class II Product Code: PJI Dated: April 28, 2021 Received: April 30, 2021
Dear Dr. Fox:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Image /page/1/Picture/6 description: The image contains the name "Pamela S. Gallagher" in a large font. Next to the name is "-S" in a smaller font. On the right side of the image, there is the word "Digitally" followed by "S. Gallagher" and "Date: 20" and "-05'00'".
Digitally signed by Pamela S. Gallagher -S Date: 2022.11.09 09:36:15
Pamela Ebrahimi, Ph.D. Deputy Branch Chief Division of Molecular Genetics and Pathology OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K203035
Device Name Eonis™ SCID-SMA kit
Indications for Use (Describe)
The Eonis™ SCID-SMA kit is intended for the semi-quantitative determination of TREC (T-cell receptor excision circle) as an aid in screening newborns for Severe Combined Immunodeficiency (SCID) and for the semi-quantitative determination of KREC (Kappa-deleting recombination excision circle) as an aid in screeming newborns for X-linked agammaglobulinemia (XLA). The test is intended for DNA from blood specimens dried on a filter paper and for use on the QuantStudio™ Dx Real-Time PCR instrument.
This test is not intended for screening of SCID-like Syndromes, such as DiGeorge Syndrome, It is also not intended to screen for less acute SCID syndromes such as leaky-SCID or variant SCID. The test is not indicated for screening B-cell deficiency disorders other than XLA, such as atypical XLA, or for screening of XLA carriers.
This test is not intended for use as a diagnostic test and a positive screening result should be followed by confirmatory testing.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for PerkinElmer. The logo consists of the company name in black, with a blue stylized "P" above it. The "P" is formed by a vertical bar and a triangle pointing to the right. Below the company name is the tagline "For the Better" in a smaller font.
510(k) Summary
This summary of safety and effectiveness information is supplied in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned number is _k203035
| Submitted by: | PerkinElmer, Inc.940 Winter StreetWaltham MA 02451 |
|---|---|
| Contact Person: | Casey FoxTel: 408-823-5561 |
| Trade Name: | Eonis SCID-SMA kit |
| Common Name: | Eonis SCID-SMA kit |
| Regulation: | 21 CFR 866.5930 |
| Classification: | II |
| Panel: | 75 Chemistry |
| Product Code: | PJI |
| Predicate device: | PerkinElmer ENLITE™ Neonatal TREC Kit (DEN140010) |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for PerkinElmer. The logo consists of a blue abstract symbol above the company name. The symbol is a blue shape resembling a stylized arrow or a forward-pointing chevron. The company name "PerkinElmer" is written in a combination of black and gray text, with "Perkin" in black and "Elmer" in gray. Below the company name is the tagline "For the Better" in blue.
1. Device Description
The Eonis SCID-SMA kit is a multiplex real-time PCR-based assay. It uses target sequence-specific primers and TaqMan™ probes to amplify and detect three targets: TREC, and RPP30, in the DNA extracted from newborn dried blood spot (DBS) using Eonis DNA Extraction kit in a single PCR reaction.
Each Eonis SCID-SMA kit contains reagents for up to 384 reactions (for 3241-001U) or 1152 reactions (for 3242-001U) including kit controls.
Table 1 Eonis SCID-SMA Kit content
| Component | Quantity |
|---|---|
| SCID-SMA Kit Controls | 2 filter paper cassettes containing 4 sets of dried bloodspots for 384 reaction kit4 filter paper cassettes containing 8 sets of dried bloodspots for 1152 reaction kit |
| C1 Analyte-negative (TREC/KREC) control | |
| C2 Low TREC/KREC control | |
| C3 High TREC/KREC control | |
| PCR Reagent 1 | 1 vial, 2.7 mL for 384 reaction kit3 vials, 2.7 mL for 1152 reaction kit |
| PCR Reagent 2 | 1 vial, 2.7 mL for 384 reaction kit3 vials, 2.7 mL for 1152 reaction kit |
| Lot-specificquality control certificate | 1 pc |
2. Intended Use
The Eonis™ SCID-SMA kit is intended for the semi-quantitative determination of TREC (T-cell receptor excision circle) as an aid in screening newborns for Severe Combined Immunodeficiency (SCID) and for the semiquantitative determination of KREC (Kappa-deleting recombination excision circle) as an aid in screening newborns for X-linked agammaglobulinemia (XLA). The test is intended for DNA from blood specimens dried on a filter paper and for use on the QuantStudio™ Dx Real-Time PCR instrument.
This test is not intended for screening of SCID-like Syndromes, such as DiGeorge Syndrome, or Omenn Syndrome. lt is also not intended to screen for less acute SCID syndromes such as leaky-SCID or variant SCID. The test is not indicated for screening B-cell deficiency disorders other than XLA, such as atypical XLA, or for screening of XLA carriers.
TaqMan is a trademark of Roche Molecular Systems, Inc. QuantStudio is a trademark of Thermo Fisher Scientific.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for PerkinElmer, a company focused on diagnostics, life sciences, and environmental solutions. The logo features the company name in a bold, sans-serif font, with the "Perkin" part in a darker shade and the "Elmer" part in a lighter shade. Above the name is a stylized blue graphic element that resembles an abstract shape. Below the company name is the tagline "For the Better" in a smaller, italicized font.
This test is not intended for use as a diagnostic test and a positive screening result should be followed by confirmatory testing.
3. Substantial Equivalency
The PerkinElmer Eonis SCID-SMA kit claims substantial equivalency to the PerkinElmer ENLITE™ Neonatal TREC Kit (DEN140010). Both devices are test systems intended for screening newborns for inherited immunodeficiency disorders.
The EnLite™ Neonatal TREC Kit is an in vitro diagnostic device intended for the semi-quantitative determination of TREC (T-cell receptor excision circle) DNA in blood specimens dried on filter paper. The test is indicated for use as an aid in screening newborns for severe combined immunodeficiency disorder (SCID).
Both devices are test systems intended for the semi-quantitative determination of TREC (T-cell receptor excision circle) DNA in blood specimens dried on filter paper on PCR instruments. In addition, Eonis SCID-SMA kit is intended for the semi-quantitation of KREC (Kappa-deleting recombination excision circle) as an aid in screening newborns for X-linked agammaglobulinemia (XLA).
The EnLite™ Neonatal TREC Kit uses a polymerase chain reaction (PCR) based nucleic acid amplification and time-resolved fluorescent-based detection to semi-quantify concentration. The PerkinElmer Eonis SCID-SMA Kit also uses fluorescent-based PCR nucleic acid amplification and detection to semi-quantify the same concentration. The use of PCR in newborn screening is well-established, and introduction of well-established real-time detection (qPCR) does not raise new questions of safety or effectiveness due to its use in this device.
Both devices are not intended for use as a diagnostic test or for screening of SCID-like Syndromes, such as DiGeorge Syndrome, or Omenn Syndrome. It is also not intended to screen for less acute SCD syndromes such as leaky-SCID or variant SCID.
In a study of 3018 clinical newborn samples with known medical status, 17 of which were confirmed positive for SCID and 6 were confirmed positive for XLA, the PerkinElmer Eonis SCID-SMA kit demonstrated the following sensitivity and specificity:
| Analyte | Sensitivity | False-negativerate | Specificity | False-positiverate | |
|---|---|---|---|---|---|
| TREC | Percent | 100 % | 0 % | 99.7 % | 0.3 % |
| Confidence Limits | 80.5 % - NA | NA - 19.5 % | 99.4 % - 99.9 % | 0.1 % - 0.6 % | |
| KREC | Percent | 100 % | 0 % | 99.7 % | 0.3 % |
| Confidence Limits | 54.1 % - NA | NA - 45.9 % | 99.4 % - 99.9 % | 0.1 % - 0.6 % |
In a study of 5437 clinical newborn samples with known medical status, 17 of which were confirmed positive for SCID, the PerkinElmer EnLite Neonatal TREC kit (excluding invalid results) demonstrated the following
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is the logo for PerkinElmer. The logo consists of a blue icon above the company name. The icon is a stylized arrow pointing to the right, with a vertical line on the left side. The company name "PerkinElmer" is written in a bold, sans-serif font, with the "Elmer" part in a slightly lighter shade of gray. Below the company name is the tagline "For the Better" in a smaller, italicized font.
sensitivity and specificity (DEN140010):
| Analyte | Sensitivity | False-negative rate | Specificity | False-positiverate | |
|---|---|---|---|---|---|
| TREC | Percent | 100 % | 0 % | 99.7 % | 0.3 % |
| Confidence Limits | 79.4 % - NA | NA - 20.6 % | 99.4 % - 99.8 % | 0.2 % - 0.6 % |
Table below provides the similarities and differences of the proposed device and the predicate. The analytical and clinical tests performed with the EONIS SCID-SMA kit have demonstrated that the two products are substantially equivalent.
| Characteristics | Proposed DeviceEONIS SCID-SMA kit | PredicateENLITE™ Neonatal TREC Kit(DEN140010). |
|---|---|---|
| Intended Use/Indications forUse | The Eonis™ SCID-SMA kit is intended for the semi-quantitative determination of TREC (T-cell receptorexcision circle) as an aid in screening newborns forSevere Combined Immunodeficiency (SCID) and forthe semi-quantitative determination of KREC(Kappa-deleting recombination excision circle) asan aid in screening newborns for X-linkedagammaglobulinemia (XLA). The test is intended forDNA from blood specimens dried on a filter paperand for use on the QuantStudio™ Dx Real-Time PCRinstrument.This test is not intended for screening of SCID-likeSyndromes, such as DiGeorge Syndrome, or OmennSyndrome. It is also not intended to screen for lessacute SCID syndromes such as leaky-SCID or variantSCID. The test is not indicated for screening B-celldeficiency disorders other than XLA, such asatypical XLA, or for screening of XLA carriers.This test is not intended for use as a diagnostic testand a positive screening result should be followedby confirmatory testing. | The EnLite™ Neonatal TREC Kit is anin vitro diagnostic device intendedfor the semi-quantitativedetermination of TREC (T-cellreceptor excision circle) DNA in bloodspecimens dried on filter paper. Thetest is for use on the VICTOR™ EnLiteinstrument. The test is indicated foruse as an aid in screening newbornsfor severe combinedimmunodeficiency disorder (SCID).This test is not intended for use as adiagnostic test or for screening ofSCID-like Syndromes, such asDiGeorge Syndrome, or OmennSyndrome. It is also not intended toscreen for less acute SCID syndromessuch as leaky-SCID or variant SCID. |
| TestMethodology | Semi-quantitative, multiplex real-time fluorescent-based polymerase chain reaction (PCR) basednucleic acid amplification and detection | Semi-quantitative, polymerase chainreaction (PCR) based nucleic acidamplification and time-resolvedfluorescence resonance energytransfer (TR-FRET) based detection |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for PerkinElmer. The logo consists of the company name in bold, black font, with the tagline "For the Better" underneath. Above the name is a blue graphic consisting of a vertical rectangle and a right-pointing triangle.
| Instrument /SoftwarePlatform | QuantStudio™ Dx Real-Time PCR instrument(K123955) and Eonis Analysis Software | VICTOR™ EnLite instrument and theEnLite™ workstation software. | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sample Type | Punch from dried blood spot (DBS) specimen | Same | ||||||||||||
| ReportableRange | TREC 242 – 4320 copies/105 cellsKREC 459 - 24300 copies/105 cells | TREC 29-473 copies/µL blood | ||||||||||||
| Lower Limitsof Measure | TREC LoB=0 copies/105 cells, LoD=LoQ=242copies/105 cellsKREC LoB=0 copies/105 cells, LoD=LoQ=459copies/105 cells | TREC LoB=3 copies/ µL blood,LoD=20 copies/ µL blood, LoQ=29copies/ µL blood | ||||||||||||
| Calibrators /Standards | Calibration is based on internal reference (RPP30)in each well and manufacturer calibration for eachkit lot. | 3 levels of DBS calibrators preparedfrom porcine whole blood spikedwith TREC and beta-actin (reference)plasmids, and a no-template blank | ||||||||||||
| Controls | 3 levels of DBS controls prepared from leucocyte-depleted human red blood cells and TREC, KREC,SMN1 and RPP30 plasmids spiked in, plus Notemplate control (NTC) | 3 levels of DBS controls preparedfrom porcine whole blood, TREC andbeta-actin (reference) plasmidsspiked in. | ||||||||||||
| ExpectedValues | Analytecopies/105cells | N | Median | Min | Max | 0.3% | 0.5% | 1.0% | Analytecopies/µLblood | N | Median | 2.0% | 2.5% | 5.0% |
| TREC | 3341 | 2520 | 117 | 9990 | 262 | 413 | 563 | TREC | 2846 | 150 | 34 | 36 | 46 | |
| KREC | 3341 | 3590 | 12 | 20100 | 129 | 261 | 484 | |||||||
| Reproducibility | ||||||||||||||
| SampleRange (copies/105cells) | Lognormal %CVRange | SampleRange(copies/µLblood) | Lognormal%CV Range | |||||||||||
| TREC | 299 - 3867 | 27%-65% | TREC 56 - 545 | 49%-87% | ||||||||||
| KREC | 763 - 9648 | 26%-50% |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for PerkinElmer. The logo consists of a blue abstract shape resembling a stylized "P" with an arrow pointing to the right. Below the symbol is the company name "PerkinElmer" in a bold, sans-serif font. Underneath the company name is the tagline "For the Better" in a smaller, italicized font.
4. Summary of the Studies
SITE-TO-SITE REPRODUCIBILITY
The reproducibility of the Eonis SCID-SMA assay was determined using a panel of dried blood spots at different TREC, KREC levels, 1 reagent kit lot, 2 external newborn screening laboratories and 1 internal site. In each laboratory, 2 operators performed 5 runs each during 5 operating days. Each run consisted of 1 plate with 5 replicates per sample. Total number of measurements was 150 per sample in each laboratory). The analysis of variance approach was used to calculate the following:
The values in the mean TREC and KREC copies/10° cells column are transformed from logarithmic (Ln) mean values, and therefore they represent geometric means in the copies/105 cells scale. The analysis of variance approach was used to calculate the results presented as SDs in the logarithmic (Ln) scale complemented with total %CV in lognormal scale. Summary mean, min and max copies/105 cells and %CV results without log transformation for total imprecision are also shown.
| Calculations with logarithmic transformation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Sample | GeometricMean(Copies105/cells) | Mean Ln(Copies105/cells) | Repeatability Ln SD | BetweenRun Ln SD | BetweenOperatorLn SD | BetweenSite Ln SD | TotalLn SD | Log-normalCV% |
| 6 | 23 | 3.14 | 0.73 | 0.60 | 0.01 | 0.53 | 1.09 | 151 |
| 13 | 34 | 3.53 | 1.12 | 0.31 | 0.00 | 0.32 | 1.20 | 180 |
| 10 | 180 | 5.19 | 0.77 | 0.01 | 0.00 | 0.03 | 0.77 | 90 |
| 5 | 299 | 5.70 | 0.55 | 0.16 | 0.01 | 0.15 | 0.59 | 65 |
| 3 | 912 | 6.82 | 0.29 | 0.06 | 0.00 | 0.10 | 0.31 | 32 |
| 12 | 936 | 6.84 | 0.34 | 0.00 | 0.00 | 0.18 | 0.38 | 40 |
| 4 | 1026 | 6.93 | 0.28 | 0.10 | 0.01 | 0.13 | 0.33 | 34 |
| 2 | 1478 | 7.30 | 0.26 | 0.01 | 0.04 | 0.04 | 0.27 | 27 |
| 8 | 3013 | 8.01 | 0.48 | 0.20 | 0.01 | 0.16 | 0.55 | 59 |
| 11 | 3290 | 8.10 | 0.25 | 0.05 | 0.00 | 0.09 | 0.27 | 28 |
| 7 | 3867 | 8.26 | 0.29 | 0.08 | 0.00 | 0.11 | 0.32 | 33 |
| 9 | 8225 | 9.01 | 0.24 | 0.08 | 0.02 | 0.18 | 0.31 | 32 |
| 1 | 9000 | 9.11 | 0.21 | 0.08 | 0.00 | 0.14 | 0.26 | 27 |
TREC Reproducibility data pooled across three laboratories.
TREC total variation results without logarithmic transformation.
| Calculations without logarithmic transformation | ||||||
|---|---|---|---|---|---|---|
| Sample | N | Mean(Copies/105 cells) | Min (Copies/105 cells) | Max (Copies/105 cells) | SD(Copies/105 cells) | CV% |
| 6 | 54 | 40 | 4 | 308 | 55.3 | 138 |
| 13 | 71 | 58 | 3 | 239 | 56 | 97 |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for PerkinElmer. The logo consists of a blue stylized "P" shape that resembles a forward-pointing arrow or play button. Below the symbol, the word "PerkinElmer" is written in a bold, sans-serif font. Underneath "PerkinElmer", the tagline "For the Better" is written in a smaller, italicized font.
| Calculations without logarithmic transformation | ||||||
|---|---|---|---|---|---|---|
| Sample | N | Mean(Copies/105 cells) | Min (Copies/105 cells) | Max (Copies/105 cells) | SD(Copies/105 cells) | CV% |
| 10 | 148 | 223 | 5 | 796 | 126 | 56 |
| 5 | 150 | 346 | 41 | 971 | 174 | 50 |
| 3 | 150 | 952 | 220 | 1792 | 272 | 29 |
| 12 | 150 | 997 | 239 | 2563 | 349 | 35 |
| 4 | 150 | 1076 | 497 | 2490 | 335 | 31 |
| 2 | 150 | 1526 | 654 | 2411 | 370 | 24 |
| 11 | 150 | 3388 | 513 | 6401 | 767 | 23 |
| 7 | 150 | 4076 | 2043 | 16934 | 1590 | 39 |
| 8 | 150 | 4601 | 797 | 218936 | 17700 | 384* |
| 9 | 150 | 8571 | 3233 | 14974 | 2420 | 28 |
| 1 | 150 | 9277 | 4236 | 16270 | 2320 | 25 |
*Dataset for sample 8 has one high outlier (max value) affecting the variability estimate. Without the outlier, the estimated CV is 41% and similar to other samples within the measuring range.
| KREC Reproducibility data pooled across three laboratories. | ||
|---|---|---|
| ------------------------------------------------------------- | -- | -- |
| Calculations with logarithmic transformation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Sample | GeometricMean (Copies105/cells) | Mean Ln(Copies105/cells) | Repeata-bilityLn SD | BetweenRunLn SD | BetweenOperatorLn SD | BetweenSiteLn SD | Total LnSD | Log-normalCV% |
| 10 | 348 | 5.85 | 0.72 | 0.01 | 0.13 | 0.22 | 0.76 | 89 |
| 8 | 763 | 6.64 | 0.46 | 0.05 | 0.00 | 0.13 | 0.48 | 50 |
| 13 | 792 | 6.67 | 0.43 | 0.09 | 0.06 | 0.09 | 0.45 | 47 |
| 12 | 2770 | 7.93 | 0.24 | 0.00 | 0.00 | 0.17 | 0.30 | 31 |
| 3 | 3849 | 8.26 | 0.28 | 0.07 | 0.00 | 0.09 | 0.31 | 31 |
| 7 | 4037 | 8.30 | 0.23 | 0.02 | 0.05 | 0.11 | 0.26 | 27 |
| 1 | 4359 | 8.38 | 0.19 | 0.08 | 0.03 | 0.06 | 0.22 | 22 |
| 6 | 5007 | 8.52 | 0.38 | 0.01 | 0.00 | 0.06 | 0.38 | 40 |
| 4 | 9313 | 9.14 | 0.20 | 0.05 | 0.00 | 0.02 | 0.21 | 21 |
| 11 | 9648 | 9.17 | 0.25 | 0.05 | 0.00 | 0.05 | 0.26 | 26 |
| 5 | 16531 | 9.71 | 0.19 | 0.07 | 0.06 | 0.13 | 0.24 | 25 |
| 2 | 34982 | 10.5 | 0.27 | 0.00 | 0.01 | 0.01 | 0.27 | 28 |
| 10 | 348 | 5.85 | 0.72 | 0.01 | 0.13 | 0.22 | 0.76 | 89 |
KREC total variation results without logarithmic transformation.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for PerkinElmer. The logo consists of a blue abstract shape above the company name, "PerkinElmer," which is written in a bold, sans-serif font. Below the company name is the tagline "For the Better" in a smaller, italicized, blue font. The abstract shape above the name is a blue triangle with a curved shape on the right side.
| Sample | N | Mean(Copies/105 cells) | Min (Copies/105 cells) | Max (Copies/105 cells) | SD(Copies/105 cells) | CV% |
|---|---|---|---|---|---|---|
| 10 | 148 | 431 | 40 | 1071 | 242 | 56 |
| 8 | 150 | 841 | 186 | 1925 | 349 | 41 |
| 13 | 150 | 862 | 183 | 2003 | 330 | 38 |
| 12 | 150 | 2877 | 1097 | 5232 | 792 | 28 |
| 3 | 150 | 4025 | 1075 | 13447 | 1300 | 32 |
| 7 | 150 | 4162 | 2241 | 6788 | 1030 | 25 |
| 1 | 150 | 4457 | 2104 | 7746 | 960 | 22 |
| 6 | 150 | 5457 | 1567 | 36630 | 3230 | 59 |
| 4 | 150 | 9508 | 4735 | 15856 | 1910 | 20 |
| 11 | 150 | 9915 | 1564 | 16967 | 2130 | 21 |
| 5 | 150 | 16955 | 8813 | 29034 | 3880 | 23 |
| 2 | 150 | 36370 | 16841 | 118413 | 11200 | 31 |
PRECISION
The quantitative precision was determined in accordance with CLSI document EP05-A3.
The variation of the Eonis SCID-SMA assay was determined using dried blood spot samples, 3 kit lots, 3 sets of Eonis test systems (including three JANUS Extraction Instruments, three JANUS PCR Mastermix Instruments and three QuantStudio™ Dx Real-Time PCR Instruments), 2 operators, and 54 runs over 23 calendar days. Each run consisted of 1 plate with 2 replicates per sample in a randomized plate map. Total number of measurements was 108 per sample. The analysis of variance approach was used to calculate the following:
The values in the mean TREC and KREC copies/10° cells column are transformed from logarithmic (Ln) mean values, and therefore they represent geometric means in the copies/105 cells scale. The analysis of variance approach was used to calculate the results presented as SDs in the logarithmic (Ln) scale complemented with total %CVs in lognormal scale. Summary mean, min and max copies/10 cells and SD and %CV results without log transformation for total imprecision are also shown.
| Calculations with logarithmic transformation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Sample | GeometricMean (Copies105/cells) | Mean Ln(Copies105/cells) | RepeatabilityLn SD | BetweenRunLn SD | BetweenInstrumentLn SD | BetweenLotLn SD | Total LnSD | Log-normalCV% |
| 6 | 20 | 3.00 | 0.92 | 0.00 | 0.00 | 0.33 | 0.97 | 125 |
| 13 | 34 | 3.54 | 0.89 | 0.00 | 0.00 | 0.23 | 0.92 | 116 |
| 10 | 159 | 5.07 | 0.76 | 0.17 | 0.00 | 0.14 | 0.79 | 92 |
| 5 | 464 | 6.14 | 0.42 | 0.06 | 0.14 | 0.13 | 0.47 | 49 |
TREC Precision data across three kit lots and three instruments.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for PerkinElmer. The logo consists of a blue stylized "P" shape with an arrow-like extension, followed by the text "PerkinElmer" in a sans-serif font. Below the company name, the tagline "For the Better" is written in a smaller, italicized font.
| 12 | 1022 | 6.93 | 0.34 | 0.16 | 0.00 | 0.08 | 0.39 | 40 |
|---|---|---|---|---|---|---|---|---|
| 2 | 1130 | 7.03 | 0.35 | 0.20 | 0.00 | 0.02 | 0.40 | 42 |
| 4 | 1176 | 7.07 | 0.33 | 0.13 | 0.13 | 0.15 | 0.41 | 42 |
| 3 | 2165 | 7.68 | 0.37 | 0.00 | 0.00 | 0.05 | 0.37 | 38 |
| 11 | 4105 | 8.32 | 0.24 | 0.08 | 0.16 | 0.04 | 0.30 | 30 |
| 7 | 4146 | 8.33 | 0.32 | 0.15 | 0.10 | 0.01 | 0.37 | 38 |
| 8 | 4866 | 8.49 | 0.35 | 0.31 | 0.00 | 0.07 | 0.47 | 50 |
| 9 | 8604 | 9.06 | 0.30 | 0.35 | 0.19 | 0.18 | 0.54 | 58 |
| 1 | 11048 | 9.31 | 0.21 | 0.14 | 0.16 | 0.03 | 0.30 | 31 |
TREC total variation results without logarithmic transformation.
| Sample | N | Mean (Copies/ 105 cells) | Min (Copies/ 105 cells) | Max (Copies/ 105 cells) | SD (Copies/ 105 cells) | CV% |
|---|---|---|---|---|---|---|
| 6 | 29 | 32 | 5 | 150 | 34.9 | 109 |
| 13 | 43 | 49 | 4 | 173 | 41.4 | 84 |
| 10 | 105 | 200 | 11 | 479 | 114 | 57 |
| 5 | 107 | 508 | 99 | 1183 | 216 | 43 |
| 12 | 106 | 1094 | 189 | 2247 | 383 | 35 |
| 2 | 107 | 1206 | 230 | 2576 | 414 | 34 |
| 4 | 108 | 1259 | 422 | 3011 | 479 | 38 |
| 3 | 107 | 2263 | 107 | 3084 | 510 | 23 |
| 11 | 106 | 4263 | 2031 | 7418 | 1220 | 29 |
| 7 | 107 | 4407 | 1014 | 8206 | 1530 | 35 |
| 8 | 107 | 5754 | 2011 | 74368 | 7120 | 124 |
| 9 | 107 | 9540 | 471 | 19530 | 3760 | 39 |
| 1 | 107 | 11501 | 3218 | 19982 | 3170 | 28 |
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for PerkinElmer. The logo features a stylized blue "P" shape, with a vertical bar and an arrow-like shape pointing to the right. Below the symbol is the company name "PerkinElmer" in a bold, sans-serif font, with the "E" in "Elmer" being in blue. Underneath the company name is the tagline "For the Better" in a smaller, italicized font, also in blue.
| Calculations with logarithmic transformation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Sample | GeometricMean(Copies105/cells) | Mean Ln(Copies105/cells) | Repeata-bilityLn SD | BetweenRunLn SD | BetweenInstrumentSD | BetweenLotLn SD | Total LnSD | Log-normalCV% |
| 9 | 85 | 4.44 | 1.14 | 0.00 | 0.00 | 0.37 | 1.20 | 178 |
| 10 | 478 | 6.17 | 0.53 | 0.36 | 0.00 | 0.16 | 0.66 | 74 |
| 13 | 1033 | 6.94 | 0.40 | 0.13 | 0.15 | 0.09 | 0.45 | 48 |
| 8 | 1075 | 6.98 | 0.47 | 0.00 | 0.00 | 0.13 | 0.49 | 52 |
| 7 | 2416 | 7.79 | 0.43 | 0.00 | 0.08 | 0.01 | 0.44 | 46 |
| 12 | 3361 | 8.12 | 0.36 | 0.10 | 0.00 | 0.07 | 0.38 | 39 |
| 1 | 5271 | 8.57 | 0.35 | 0.00 | 0.09 | 0.04 | 0.37 | 38 |
| 3 | 5541 | 8.62 | 0.30 | 0.05 | 0.00 | 0.03 | 0.31 | 31 |
| 6 | 8691 | 9.07 | 0.26 | 0.26 | 0.08 | 0.10 | 0.39 | 40 |
| 4 | 11499 | 9.35 | 0.29 | 0.18 | 0.12 | 0.06 | 0.36 | 38 |
| 11 | 13494 | 9.51 | 0.24 | 0.08 | 0.17 | 0.09 | 0.32 | 32 |
| 5 | 13767 | 9.53 | 0.22 | 0.21 | 0.13 | 0.07 | 0.34 | 35 |
| 2 | 39735 | 10.6 | 0.20 | 0.21 | 0.12 | 0.07 | 0.32 | 33 |
KREC Precision data across three kit lots and three instruments.
KREC total variation results without logarithmic transformation.
| Calculations without logarithmic transformation | ||||||
|---|---|---|---|---|---|---|
| Sample | N | Mean(Copies/105 cells) | Min (Copies/105 cells) | Max (Copies/105 cells) | SD(Copies/105 cells) | CV% |
| 9 | 10 | 146 | 17 | 435 | 139 | 95 |
| 10 | 104 | 570 | 43 | 1396 | 306 | 54 |
| 13 | 106 | 1139 | 212 | 4188 | 546 | 48 |
| 8 | 107 | 1195 | 356 | 3180 | 538 | 45 |
| 7 | 107 | 2605 | 203 | 5513 | 898 | 34 |
| 12 | 107 | 3568 | 686 | 7096 | 1150 | 32 |
| 1 | 107 | 5560 | 474 | 10461 | 1620 | 29 |
| 3 | 106 | 5774 | 863 | 10031 | 1500 | 26 |
| 6 | 107 | 9308 | 3333 | 19079 | 3420 | 37 |
| 4 | 108 | 12212 | 2707 | 23846 | 4180 | 34 |
| 11 | 107 | 14103 | 5330 | 23372 | 4030 | 29 |
| 5 | 106 | 14492 | 5673 | 35070 | 4960 | 34 |
| 2 | 108 | 41717 | 15231 | 74089 | 12400 | 30 |
Qualitative imprecision was determined in accordance with CLSI document EP12-A2 to classify the results, using the screening performance study cut-offs (262 copies/105 cells for TREC, 484 copies/108 cells for KREC). The C5-C95 interval was determined to be 79–626 copies/105 cells for TREC, and 189–1064 copies/105 cells for KREC.
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the logo for PerkinElmer. The logo consists of a blue stylized "P" shape, with a vertical line and an arrow-like shape pointing to the right. Below the symbol, the name "PerkinElmer" is written in a dark gray, sans-serif font. Underneath the name, the tagline "For the Better" is written in a smaller, blue, sans-serif font.
Concentrations outside of these intervals were consistently analyte negative (concentrations <C5, equals to screen positive) or consistently analyte positive (concentrations >C95, equals to screen negative).
LIMIT OF DETECTION
The Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) were determined in accordance with CLSI document EP17-A2. The data for LoB was analyzed using "Assign LoB = Zero and Confirm" approach. The data for LoD was analyzed with probit approach.
Based on 300 determinations of blank samples (150 for each kit lot) the limit of blank (LoB) for TREC, KREC, SMN1 is 0 copies/μL blood and 0 copies/105 cells.
Based on total of 960 determinations, 20 replicates per dilution, the limit of detection (LoD) for TREC is 242 copies/10° cells with 95% probability. The Limit of Quantitation (LoQ) for TREC is 242 copies/10° cells, which is equal to LoD. The limit of detection (LoD) for KREC is 459 copies/105 cells) with 95% probability. The Limit of Quantitation (LoQ) for KREC is 459 copies/105 cells, which is equal to LoD.
LINEARITY
Linearity was determined in accordance with CLSI document EP06 ED2:2020 with one kit lot. Three (3) sets of contrived samples were used for this evaluation. Two sample sets were diluted to 9 levels and tested with 4 replicates for each level. For KREC the linearity panel was amended with results from LoD-study to reach lower concentrations. The allowable maximum deviation from linearity in the study was 25%.
The Eonis SCID-SMA assay is demonstrated to be linear for TREC from 94 copies/105 cells with observed maximum deviation of -14.3%. The KREC analyte is demonstrated to be linear from 117 to 24343 copies/10° cells with observed maximum deviation of 17.3%.
INTERFERENCE
The Eonis SCID-SMA kit was evaluated for interference from potential endogenous and exogenous sources in accordance with CLSI document EP07-A3.
The following potentially interfering substances were added to whole blood spiked with at three different TREC plasmid and KREC plasmid concentrations and were found not to interfere at the concentration indicated.
Interference substances tested and their concentrations
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image is the logo for PerkinElmer. The logo consists of a blue abstract shape resembling a stylized "P" with an arrow pointing to the right. Below the shape is the company name "PerkinElmer" in a bold, sans-serif font. Underneath the company name is the tagline "For the Better" in a smaller, italicized font.
| Tested substance | Added concentration of tested substance |
|---|---|
| Conjugated bilirubin | 16.6 mg/dL in blood |
| Hemoglobin | 200 g/L in blood |
| Unconjugated bilirubin | 10 mg/dL in blood |
| Intralipid® | 1500 mg/dL in blood |
| Li-heparin | 7500 USP /dL in blood |
| EDTA | 9.8 mg/mL in blood |
| Na-citrate | 0.0645 mol/L in blood |
SCREENING PERFORMANCE
The screening performance of the Eonis SCID-SMA kit was determined in a clinical study conducted in Denmark. Retrospective archived dried blood spot specimens (collected from US and Denmark) from subjects confirmed positive for SCID, XLA were included to enrich the cohort of routine newborn screening specimens obtained from the Danish Newborn Screening Biobank.
Confirmatory test results were used as the comparator for the confirmed positive SCID, XLA cases. The clinical status of the routine subjects was determined through a retrospective review by clinical experts to confirm the routine subject cohort samples were from unaffected individuals.
Using the data collected to establish the expected values of the Eonis SCID-SMA kit were determined by calculating the TREC concentrations corresponding to the 0.3" and 1.0" population percentiles (262 copies/10 cells for TREC, 484 copies/10 cells for KREC) established in a cut-off study with an independent dataset. The specimens having TREC and KREC levels below the cut-off values in the initial round of testing were re-tested in duplicate. The final results (presumptive positive positive, invalid result) were classified after the second round of testing. Summary of the retest rate and final results for routine screening specimens is provided below.
Intralipid is a registered trademark of Fresenius Kabi AB.
{15}------------------------------------------------
Image /page/15/Picture/1 description: The image shows the logo for PerkinElmer. The logo features a blue abstract design above the company name. The design includes a vertical blue bar and a blue triangle pointing to the right. Below the company name, the tagline "For the Better" is written in a smaller, italicized font.
| Routine screening samples | SMN1 | TREC | KREC |
|---|---|---|---|
| Screened samples | 3018 | 3018 | 3018 |
| Initial screen positive | 0 | 10 | 18 |
| Initial screen negative | 3018 | 3008 | 3000 |
| Retest rate | 0% | 0.3% | 0.6% |
| Final screen positive | 0 | 9 | 9 |
| Final screen negative | 3018 | 3008 | 3004 |
| False-positive rate | 0% | 0.3% | 0.3% |
Summary of the samples tested in the pivotal study.
The screening performance of the Eonis SCID-SMA kit was established by measuring TREC, KREC (and RPP30) in 3090 DBS specimens.
In total, 17 SCID and 6 XLA retrospective case specimens and 3018 normal newborn screening specimens were available for the study. The screening performance of the Eonis SCID-SMA kit was established by measuring TREC, and KREC (and RPP30) and the final results after retesting are presented below.
Screening performance of Eonis SCID-SMA kit.
| TREC | Clinical Status | Total (%) | ||
|---|---|---|---|---|
| SCID affected (%) | Normal (%) | |||
| Screeningresult | Presumptive positive (%) | 17 (100 %) | 9 (0.3 %) | 26 (0.9 %) |
| Presumptive normal (%) | 0 (0.0 %) | 3008 (99.7 %) | 3008 (99.1 %) | |
| Total (%) | 17 (100 %) | 3017 (100 %) | 3034 (100 %) |
| KREC | Clinical Status | |||
|---|---|---|---|---|
| XLA affected (%) | Normal (%) | Total (%) | ||
| Screeningresult | Presumptive positive (%) | 6 (100 %) | 9* (0.3 %) | 15 (0.5 %) |
| Presumptive normal (%) | 0 (0.0 %) | 3004 (99.7 %) | 3004 (99.5 %) | |
| Total (%) | 6 (100 %) | 3013 (100 %) | 3019 (100 %) |
*Includes one false positive result from a female population the false positive rate is estimated to be 0.1 %
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for PerkinElmer, a company specializing in detection, imaging, informatics, and services. The logo features a stylized blue symbol resembling a forward-pointing arrow or a stylized 'P' above the company name. The name 'PerkinElmer' is written in a combination of black and gray, with 'Perkin' in black and 'Elmer' in gray, and the tagline 'For the Better' is written in blue below the company name.
| KREC(Male only) | Clinical Status | |||
|---|---|---|---|---|
| XLA affected (%) | Normal (%) | Total (%) | ||
| Screeningresult | Presumptive positive (%) | 6 (100 %) | 8 (0.5 %) | 14 (0.9 %) |
| Presumptive normal (%) | 0 (0.0 %) | 1515 (99.5 %) | 1515 (99.1 %) | |
| Total (%) | 6 (100 %) | 1523 (100 %) | 1529 (100 %) |
Conclusion
The Eonis SCID-SMA kit demonstrates analytical and screening performance that supports its substantial equivalency with the predicate device, PerkinElmer ENLITE™ Neonatal TREC Kit (DEN140010).
§ 866.5930 Newborn screening test for severe combined immunodeficiency disorder (SCID).
(a)
Identification. A newborn screening test for SCID is a prescription device intended to measure T-cell receptor excision circle (TREC) DNA obtained from dried blood spot specimens on filter paper using a polymerase chain reaction based test as an aid in screening newborns for SCID. Presumptive positive results must be followed up by diagnostic confirmatory testing. This test is not intended for use as a diagnostic test, or for screening of SCID-like syndromes, such as DiGeorge syndrome or Omenn syndrome. It is also not intended to screen for less acute SCID syndromes, such as leaky SCID or variant SCID.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The intended use must indicate:
(A) The test is not intended for diagnostic use, or for screening of SCID-like syndromes, such as DiGeorge syndrome or Omenn syndrome; and
(B) The test is not intended to screen for less acute SCID syndromes, such as leaky SCID or variant SCID.
(ii) A detailed description of all components in the test that includes:
(A) A detailed description of the test components, all required reagents, instrumentation and equipment, including illustrations or photographs of nonstandard equipment or methods;
(B) Detailed documentation of the device software including, but not limited to, standalone software applications and hardware-based devices that incorporate software;
(C) Specifications for the filter paper, which must be appropriately labeled for in vitro diagnostic use, to be used in specimen collection and how it will be used in specimen collection validation. These specifications must include: descriptive characteristics of the filter paper, instructions on how a lab should choose the appropriate filter paper, chemical properties of the filter paper, interference concerns associated with the chemicals in the filter paper, absorption properties of the filter paper, punch size, absorption capacity, testing for homogeneity of punches, diameter of the circle for the dried blood spot aliquot, absorption time, physical composition, and number and size of punches to be tested;
(D) Methodology and protocols for detection of T-cell receptor excision circles and methods for determination of results. The cutoff must be selected before conducting clinical and analytical studies;
(E) A description of the result outputs along with sample reports. Sample reports must include the scale used in reporting of results (
e.g., TREC copies/µL) and the range of values that will be reported out; and(F) A description of appropriate internal and external controls that are recommended or provided. The description must identify those control elements that are incorporated into the testing procedure.
(iii) Information that demonstrates the performance characteristics of the test, including:
(A) Data that demonstrates the clinical validity of the device, using well characterized prospectively or retrospectively obtained clinical specimens representative of the intended use population. A minimum of 10 to 15 confirmed positive specimens must be obtained from more than 1 site, including relevant annotation, and, at 1 year or beyond, a SCID diagnosis by flow cytometry or clinically meaningful information regarding the status of the subject must be obtained. Additional specimens should have been obtained that are characterized by other disorders that can be found by screening specimens that have low or absent TREC (
e.g., other T-cell lymphopenic disorders) to supplement the range of results. The clinical validation study must have a pre-specified clinical decision point (i.e., cutoff to distinguish positive and negative results). Results must be summarized in tabular format comparing interpretation of results to the reference method. Point estimates together with two-sided 95 percent confidence intervals must be provided for the positive percent agreement, negative percent agreement, and overall percent agreement. Data must include the retest rate, the false positive rate before retest, the final false positive rate, and the false negative rate;(B) Device reproducibility data generated, using a minimum of three sites of which at least two must be external sites, with two operators at each site. Each site must conduct a minimum of five runs per operator over five nonconsecutive days evaluating a minimum of six different relevant TREC concentrations that span and are well distributed over the measuring range and include the clinical cutoff. Specimens must include cord blood and cord blood diluted with ABO matched adult blood specimens. Identical specimens from the same sample panel must be tested at each site. Each specimen must be run in triplicate and include controls run in triplicate. Results must be reported as the standard deviation and percentage coefficient of variation for each level tested. Results must also be displayed as a dichotomous variable around the cutoff. Total variation must be partitioned into the sum of within-lab and between-lab variations with pre-specified acceptance criteria and 95 percent confidence intervals for all data. Pre-specified acceptance criteria must be provided and followed;
(C) Device precision data using clinical samples to evaluate the within-lot, between-lot, within-run, between run, and total variation. A range of TREC levels of the specimen must include samples within the measuring range, samples above and below the measuring range, as well as with samples very near above and below the cutoff value. At least three replicates of each specimen must be tested with controls and calibrator(s) according to the device instructions for use. The precision study must use well characterized samples using different lots, instruments, and operators. Results must be summarized in tabular format. Pre-specified acceptance criteria must be provided and followed;
(D) Linearity of the test must be demonstrated using a dilution panel from clinical samples. The range of dilution samples must include samples within the measuring range, samples above and below the measuring range, as well as with samples very near above and below the cutoff value. Results of the regression analysis must be summarized in tabular format and fitted into a linear regression model with the individual measurement results against the dilution factors. Pre-specified acceptance criteria must be provided and followed;
(E) Device analytic sensitivity data, including limit of blank, limit of detection, and limit of quantification;
(F) Device specificity data, including interference, carryover, cross-contamination, and in silico analysis of potential off-target genomic sequences;
(G) Device stability data, including real-time stability of samples under various storage times, temperatures, and freeze-thaw conditions. A separate shipping stability study must be performed;
(H) Lot-to-lot reproducibility study of each filter paper that will be validated with the test. The lot-to-lot study must include a minimum of three lots of each blood spot card that will be validated with the test and be conducted over five nonconsecutive days. The sample panel must consist of specimens with a range of TREC levels and include samples within the measuring range, samples above and below the measuring range, and samples very near above and below the cutoff value. Multiple punches must be obtained from each card for demonstration of homogeneity of the analyte across the dried blood spot. Comparability of the test performance for each filter paper must be demonstrated. Stability and storage of TREC DNA on each blood spot card must be demonstrated. Results of the lot-to-lot study must be summarized providing the mean, standard deviation, and percentage coefficient of variation in a tabular format. Data must be calculated for within-run, between-run, within-lot, and between-lot. Data demonstrating the concordance between results across different filter papers must be provided. Study acceptance criteria must be provided and followed; and
(I) If applicable, a thermocycler reproducibility study must be performed using thermocyclers from three independent thermocyler manufacturers. The sample panel must consist of specimens with a range of TREC levels and must include samples within the measuring range, samples above and below the measuring range, and samples very near above and below the cutoff value. The study must be done using three filter paper lots and conducted over five nonconsecutive days. Results of the thermocycler reproducibility study must be summarized providing the mean, standard deviation, and percentage coefficient of variance in a tabular format. Data must be calculated for the within-run, between-run, within-lot, between-lot, and between thermocycler manufacturer study results. Study acceptance criteria must be provided and followed.
(iv) Identification of risk mitigation elements used by your device, including a description of all additional procedures, methods, and practices incorporated into the directions for use that mitigate risks associated with testing.
(2) Your § 809.10 compliant labeling must include:
(i) A warning statement that reads “This test is not intended for diagnostic use, preimplantation or prenatal testing, or for screening of SCID-like syndromes, such as DiGeorge syndrome or Omenn syndrome. It is also not intended to screen for less acute SCID syndromes, such as leaky SCID or variant SCID.”;
(ii) A warning statement that reads “Test results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods and clinical evaluation, as appropriate.”;
(iii) A description of the performance studies listed in paragraph (b)(1)(iii) and a summary of the results; and
(iv) A description of the filter paper specifications required for the test.