(103 days)
Not Found
No
The device description explicitly states it uses a "closed loop mechanism which is Pl controlled," which is a traditional control system, not AI/ML. There is no mention of AI, ML, or related concepts in the document.
No
The device is intended to measure and maintain cuff pressure of an endotracheal or tracheostomy tube, which supports the proper functioning of mechanical ventilation but does not directly treat a disease or condition itself.
No
The device continuously measures and automatically maintains cuff pressure, which is a therapeutic and maintenance function, not one designed to diagnose a condition. While it has a "Measure mode" to observe cuff pressure, this observation is for monitoring the effectiveness of its therapeutic function rather than for diagnosing a new patient condition.
No
The device description explicitly mentions hardware components such as a rechargeable Li-ion Battery, Mounting Rack, AC/DC Adaptor, and a sterile extension tube, indicating it is not solely software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body (like blood, urine, or tissue) to provide information about a person's health. This testing is done outside of the body (in vitro).
- Device Function: The Puritan Bennett™ Cuff Pressure Manager (CPM) directly interacts with a medical device (endotracheal or tracheostomy tube) that is inside the patient's body. It measures and maintains the pressure of the cuff on that tube.
- No Specimen Analysis: The device does not analyze any biological specimens from the patient. Its function is purely mechanical and related to managing the pressure of the airway tube cuff.
Therefore, the CPM falls under the category of a medical device used for patient care and monitoring, but not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The CPM is intended to continuously measure and automatically maintain the cuff pressure of an endotracheal tube or tracheostomy tube that has been set by the user/clinician during mechanical ventilation. The CPM is intended for use with adult and pediatric patients during mechanical ventilation in the listed areas:
Intensive care unit (ICU)
Recovery room
Emergency medical care
Medical emergency vehicles when the patient is being transported
During transport within and outside of the hospital
Product codes (comma separated list FDA assigned to the subject device)
BSK
Device Description
The Puritan Bennett™ Cuff Pressure Manager (CPM) continuously measures and automatically maintains cuff pressure during ventilation of patients using a cuffed endotracheal tube or tracheostomy tube. It is an integrated and continuous cuff pressure control solution that secures airway management in intensive care units, recovery room and during interhospital and intrahospital transport.
It operates in range of 5-50 cm H2O of user-set cuff pressures for various cuffed endotracheal tubes to provide suitable solutions for various clinical patient situations. The Puritan Bennett™ Cuff Pressure Manager introduces some new small features and convenient user interface to maximize safe use and usability. Both manual adjusting and one-click inflation function are provided alternatively for user’s preference to start inflation. Measure mode allows user to observe the cuff pressure without device intervening before taking actions. Auto Key-lock feature help to avoid mis-operation when there’s no user intervene for a long time. For inflation, room air is used and no contact to the respiratory gas system of a patient occurs. It is not intended to connect with ventilator.
The cuff pressure manager is designed with closed loop mechanism which is PI controlled, and the software level of concern is Class C/Major concern.
The Puritan Bennett™ Cuff Pressure Manager is provided non-sterile and no re-processing is required. A sterile extension tube is equipped with the Puritan Bennett™ Cuff Pressure Manager for inflating/deflating or measuring cuff pressure. The associated accessories include:
- Extension Tube with filter
- Rechargeable Li-ion Battery
- Mounting Rack
- AC/DC Adaptor
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adult and pediatric patients
Intended User / Care Setting
user/clinician during mechanical ventilation in: Intensive care unit (ICU), Recovery room, Emergency medical care, Medical emergency vehicles when the patient is being transported, During transport within and outside of the hospital.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
- Electrical Safety and Electromagnetic Compatibility Test: Device complies with ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012, IEC 60601-1-2:2014, IEC 60601-1-8:2006+AMD1:2012, IEC 60601-1-12:2014, and AIM 7351731 Rev. 2.00 2017-02-23.
- Function and performance Test-Cuff Pressure Manager: Evaluation per technical specification, battery performance study, transportation movement simulation test, performance test on ETT of different size and material, benchmark test to compare with predicate device. Closed loop design verified per IEC 60601-1-10.
- Function and performance Test-Extension tube: Evaluation per technical specification, Luer connector connectivity test per ISO 594-1:1986 and ISO 80369-7:2016, Stability Test (Accelerated aging test for 5 years shelf life per ASTM F1980:16, Packaging stability test per ASTM F88/F88M:2015, ASTM F1929:2015), Sterilization validation per ISO 11135:2014 and 10993-7: 2008 AC:2009.
- Software Test: Software Unit Test, Software Integration Test, User Interface Test, Software Verification & Validation Test.
- Cybersecurity: Cybersecurity assessment per FDA Guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", Penetration test.
- Usability Test: Per IEC 60601-1-6:2010+AMD1:2013 and IEC 62366-1:2015, validated performance meets user need and intended use, validated potential use error per risk and hazard analysis, validated Instructions for Use.
- Animal Studies: No animal data included.
- Clinical Studies: No clinical data included.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Pressure Accuracy (Display): ±1.5 cm H2O
Pressure Accuracy (Adjustment): ±1 cm H2O
Pressure monitoring resolution: ±0.01 cm H2O (0.01 mbar)
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 868.5750 Inflatable tracheal tube cuff.
(a)
Identification. An inflatable tracheal tube cuff is a device used to provide an airtight seal between a tracheal tube and a patient's trachea.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
Covidien Ilc Jenny Liu Regulatory Affairs Specialist 6F, Building 3, 2388 Chenhang Road, Minhang District Shanghai 201114 China
Re: K202874
Trade/Device Name: Puritan Bennett Cuff Pressure Manager Regulation Number: 21 CFR 868.5750 Regulation Name: Inflatable Tracheal Tube Cuff Regulatory Class: Class II Product Code: BSK Dated: December 8, 2020 Received: December 10, 2020
Dear Jenny Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
1
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K202874
Device Name
Puritan Bennett™ Cuff Pressure Manager
Indications for Use (Describe)
The CPM is intended to continuously measure and automatically maintain the cuff pressure of an endotracheal tube or
tracheostomy tube that has been set by the user/clinician during mechanical ventilation. The CPM is intended for use with
adult and pediatric patients during mechanical ventilation in the listed areas:
Intensive care unit (ICU) |
---|
Recovery room |
Emergency medical care |
Medical emergency vehicles when the patient is being transported |
During transport within and outside of the hospital |
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
General Information
Submitter Name: | Covidien IIc |
---|---|
Address: | 6F, Building 3, 2388 Chenhang Road, Minhang District |
Shanghai 201114, China | |
Contact Person: | Jenny Liu |
Regulatory Affairs Specialist | |
Telephone: +86 15618808588 | |
Fax: +86 21 3323-0131 | |
Email: jenny.liu@medtronic.com | |
Date Prepared: | January 9, 2021 |
Device Name | |
Proprietary Name | Puritan Bennett™ Cuff Pressure Manager |
Common Name | Cuff Pressure Manager |
Classification Name | Cuff, Tracheal Tube, Inflatable |
Device Class | Class II |
Product Code | BSK |
CFR Section | 21 CFR 868.5750 |
Catalog Numbers | 180-03, 180-05 |
Predicate Device
The Puritan Bennett™ Cuff Pressure Manager was compared to and found to be substantially equivalent to the following legally marketed products.
Proprietary Name: IntelliCuff
Common Name: Cuff Pressure Controller
4
510(k) Number: | K150893 (cleared January 29, 2016) |
---|---|
Manufacturer | Hamilton Medical AG |
Device Description
The Puritan Bennett™ Cuff Pressure Manager (CPM) continuously measures and automatically maintains cuff pressure during ventilation of patients using a cuffed endotracheal tube or tracheostomy tube. It is an integrated and continuous cuff pressure control solution that secures airway management in intensive care units, recovery room and during interhospital and intrahospital transport.
It operates in range of 5-50 cm H2O of user-set cuff pressures for various cuffed endotracheal tubes to provide suitable solutions for various clinical patient situations. The Puritan Bennett™ Cuff Pressure Manager introduces some new small features and convenient user interface to maximize safe use and usability. Both manual adjusting and one-click inflation function are provided alternatively for user's preference to start inflation. Measure mode allows user to observe the cuff pressure without device intervening before taking actions. Auto Key-lock feature help to avoid mis-operation when there's no user intervene for a long time. For inflation, room air is used and no contact to the respiratory gas system of a patient occurs. It is not intended to connect with ventilator.
The cuff pressure manager is designed with closed loop mechanism which is Pl controlled, and the software level of concern is Class C/Major concern.
The Puritan Bennett™ Cuff Pressure Manager is provided non-sterile and no re-processing is required. A sterile extension tube is equipped with the Puritan Bennett™ Cuff Pressure Manager for inflating/deflating or measuring cuff pressure. The associated accessories include:
- Extension Tube with filter Rechargeable Li-ion Battery Mounting Rack AC/DC Adaptor
Intended Use / Indications for Use
The CPM is intended to continuously measure and automatically maintain the cuff pressure of an endotracheal tube or tracheostomy tube that has been set by the user/clinician during mechanical ventilation. The CPM is intended for use with adult and pediatric patients during mechanical ventilation in the listed areas:
Intensive care unit (ICU) Recovery room Emergency medical care Medical emergency vehicles when the patient is being transported During transport within and outside of the hospital
5
Comparison of Technological Characteristics with Predicate Device
The intended use and indication for use for Puritan Bennett™ Cuff Pressure Manager is equivalent to those for the predicate devices, except for limitation of use in operation room, ship, jet and helicopter.
They are similar in fundamental scientific technology in that they are all multi-patient; controller is multiuse; extension tubes are sterile (EtO), single use; inflation/deflation of an endotracheal tube or tracheostomy tube; user set pressure; software controls pressure and automatic pressure maintenance; and AC and batter power supply.
The technical differences between the Puritan Bennett™ Cuff Pressure Manager and predicate device do not raise different issues of safety or effectiveness.
The following Table 10-1 represents a summary of the major technological characteristics of the proposed and predicate devices.
| Characteristic | Proposed Device Puritan Bennett™ Cuff
Pressure Manager | Predicate Device IntelliCuff Cuff
Pressure Controller (K150893) |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operation altitude | -411m to 4000m above sea level | -650 to 7,620 m above sea level |
| Operation temperature | 0° C to 40° C / 32° F to 104° F (operating),
0° C to 50° C / 32° C to 122° F (storage) | -15° C to 50° C / 5° F to 122° F
(operating),
-15° C to 70° C / 5° C to 158° F
(storage) |
| Operation relative humidity | 10% to 95% noncondensing (operating) / 10% to
95% noncondensing (storage) | 5% to 95% noncondensing (operating) /
5% to 95% noncondensing (storage) |
| Connection with ventilator | The device is not designed connected to any
mechanical ventilator. | The device can be connected to any
mechanical ventilator. |
| Sterilization | Cuff Pressure Manager is provided non-sterile.
Extension tube is provided sterile by EtO. | Controller is provided non-sterile.
Cuff Pressure tube is provided sterile by
EtO. |
| Electrical safety | Class-II Type BF in accordance with IEC 60601-1
and IEC 60601-1-12 | Class-II Type BF in accordance with IEC
60601-1 and IEC 60601-1-12 |
| Electromagnetic
Compatibility | In accordance with IEC 60601-1-2 | In accordance with IEC 60601-1-2 |
| Alarm | In accordance with IEC 60601-1-8 | In accordance with IEC 60601-1-8 |
| Sensor | Two Sensors | Two Sensors |
| AC power input | 85 to 264 VAC / 47 to 63 Hz ; 7.5 VA typical, 15 VA
maximum | 100 to 240 VAC / 50 to 60 Hz ± 10%;
1.25 VA typical, 3.25 VA maximum |
| Characteristic | Proposed Device Puritan Bennett™ Cuff
Pressure Manager | Predicate Device IntelliCuff Cuff
Pressure Controller (K150893) |
| Battery Requirement | 3.6 V /2300mAh Li-ion rechargeable battery.
A fully charged battery supports 4 hours operation. | AA (IEC-HR6) NiMH rechargeable 1.2 V,
1900 mA (two pieces to support energy
supply).
A fully charged battery supports 5 hours
operation. |
| Increase/Decrease Target
Pressure | Yes | Yes |
| Measure Mode | Yes | No |
| Power on Self Test | Yes | Yes |
| Time-limited Pressure Hold | Yes | Yes |
| One-Click Deflation | Yes | Yes |
| One-Click Inflation | Yes
Inflate to default target pressure when user press
"One-Click Inflation" Button. | No
Inflate to default target pressure the
moment pressing "Power on" button. |
| Manual Key Lock | Yes | Yes |
| Auto Key Lock | Yes | No |
| Default inflation pressure
settings | 25 cm H2O | 25 cm H2O |
| Pressure hold default
pressure increase | 5 cm H2O | 5 cm H2O |
| Pressure hold maximum
allowable pressure | 50 cm H2O | 55 cm H2O |
| Pressure hold time setting
range | 5 minutes to 30 minutes | 5, 10 minutes |
| Pressure hold time factory
default setting | 5 minutes | 5 minutes |
| Pressure setting range | 5 to 50 cm H2O | 5 to 50 cm H2O |
| Pressure unit switching | 1) cm H2O
- hPa
- mbar | 1) cm H2O
- hPa
- mbar |
| Resolution
(setting/display) | +/-1cm H2O | +/-1cm H2O |
| Characteristic | Proposed Device Puritan Bennett™ Cuff
Pressure Manager | Predicate Device IntelliCuff Cuff
Pressure Controller (K150893) |
| Pressure Accuracy | Pressure Accuracy (Display): ±1.5 cm H2O
Pressure Accuracy (Adjustment): ±1 cm H2O | Pressure Accuracy (Display): ±2 cm
H2O
Pressure Accuracy (Adjustment): ±1
cm H2O |
| Alarm Priority | High Priority Alarm:
Low Pressure (including cuff leakage), High
Pressure, Battery is depleted, Technical Errors, | High Priority Alarm:
Cuff system leakage, battery critically
low, Technical fault |
| | Medium Priority Alarm:
Battery is very low | Medium Priority Alarm:
Pressure above set limit, cuff deflated,
battery 10%, Deactivation impossible |
| Alarm Volume | 50 dB(A) ±5 dB(A)within 1 m distance | 56.5 dB(A) ± 6 dB(A) within 1 m distance |
| Applied pressure range in
deflation mode | -45 cm H2O | -100 cm H2O (-100 mBar) |
| Pressure monitoring
resolution | ±0.01 cm H2O (0.01 mbar) | ±0.1 mbar |
| Essential Performance | The applied cuff pressure must be maintained and
monitored. If it is higher or lower than the set limits
(±1cm H2O), this must be detected and the
operator informed through an alarm. | The applied cuff pressure must be
maintained and monitored. If it is higher
or lower than the set limits(±2mbar), this
must be detected and the operator
informed through an alarm. |
| Extension Tube with Filter | Yes | Yes |
| Connector compatibility | Compatible with endotracheal tube (ETT) or
tracheostomy tube (TT) | Compatible with endotracheal tube (ETT)
or tracheostomy tube (TT) |
Table 10-1 Comparison between proposed device and predicate device
6
7
Verification and validation activities were successfully completed. Evidence of safety and effectiveness in support of substantial equivalence were obtained from design verification and validation testing. The design differences were found to not affect safety or performance through applicable design verification activities that demonstrated conformance to applicable technical design specifications and performance requirements, applicable medical device performance standards and other non-clinical testing.
Performance Testing - Bench
1. Electrical Safety and Electromagnetic Compatibility Test
Electrical safety and EMC testing were conducted on the Puritan Bennett™ Cuff Pressure Manager. The device complies with the following standards:
- . ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
8
- . IEC 60601-1-8:2006+AMD1:2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- . IEC 60601-1-12:2014 Medical electrical equipment Part 1-12: General requirements for basic safety and essential performance — Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment
- . AIM 7351731 Rev. 2.00 2017-02-23 Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers
2. Function and performance Test-Cuff Pressure Manager
- . Cuff Pressure Manager function and performance evaluation per technical specification
- . Battery performance study on new and aged battery
- . Transportation movement simulation test (vibration test, bump test and broad-band random vibration test) per EN 1789:2007+A2:2014
- . Performance test on ETT of different size and material
- . Benchmark test to compare with predicate device
- . The performance of the closed loop design is verified per applicable clauses in IEC 60601-1-10.
3. Function and performance Test-Extension tube
- . Extension tube performance evaluation per technical specification
- . Luer connector connectivity test per ISO 594-1:1986 and ISO 80369-7:2016
- . Stability Test
- -Accelerated aging test for 5 years shelf life per ASTM F1980:16
- -Packaging stability test per ASTM F88/F88M:2015, ASTM F1929:2015
- . Sterilization validation per ISO 11135:2014 and 10993-7: 2008 AC:2009
4. Software Test
- . Software Unit Test
- . Software Integration Test
- . User Interface Test
- . Software Verification & Validation Test
Cybersecurity 5.
- . Cybersecurity assessment per FDA Guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".
- . Penetration test
9
6. Usability Test
Usability /human factor testing is performed per IEC 60601-1-6:2010+AMD1:2013 and IEC 62366-1:2015:
- . Validate performance meets user need and intended use identified in customer requirement.
- Validate potential use error per risk and hazard analysis .
- . Validate Instructions for Use
Consensus standards utilized
| Standards for Declaration of Conformity | FDA-
Recognized
Standard No. |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012, Medical electrical equipment - Part 1:
General requirements for basic safety and essential performance | 19-4 |
| IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic
safety and essential performance - Collateral Standard: Electromagnetic disturbances -
Requirements and tests | 19-8 |
| IEC 60601-1-8:2006+AMD1:2012 Medical electrical equipment - Part 1-8: General
requirements for basic safety and essential performance - Collateral Standard: General
requirements, tests and guidance for alarm systems in medical electrical equipment and
medical electrical systems | 5-76 |
| IEC 60601-1-12:2014 Medical electrical equipment Part 1-12: General requirements for basic
safety and essential performance - Collateral Standard: Requirements for medical electrical
equipment and medical electrical systems intended for use in the emergency medical services
environment | 19-15 |
| IEC 60601-1-6:2010+AMD1:2013 Medical electrical equipment - Part 1-6: General
requirements for basic safety and essential performance - Collateral standard: Usability | 5-89 |
| IEC 62366-1:2015 Medical devices - Application of usability engineering to medical devices | 5-114 |
| IEC 62304: 2015 Medical device software - Software life cycle processes | 13-79 |
| ISO 14971:2007 Medical devices: Application of risk management to medical devices | 5-40 |
| ANSI/AAMI ST67:2011(R2017) Sterilization of health care products-Requirements and
Guidance for selecting a sterility assurance level (SAL) for products labeled "STERILE" | 14-314 |
| ISO 11135:2014 Sterilization of health care products - Ethylene oxide - Requirements for
development, validation and routine control of a sterilization process for medical devices | 14-452 |
| ISO 10993-7: 2008 AC:2009 Biological evaluation of medical devices - Part 7: Ethylene oxide
sterilization residuals | 14-408 |
| ASTM F88/F88M:2015 Standard Test Method for Seal Strength of Flexible Barrier Materials | 14-482 |
| ASTM F1929:2015 Standard Test Method for Detecting Seal Leaks in Porous Medical
Packaging by Dye Penetration | 14-484 |
| ASTM F1980:16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical
Devices | 14-497 |
| ISO 15223-1:2016 Medical devices - Symbols to be used with medical device labels, labeling
and information to be supplied - Part 1: General requirement | 5-117 |
| ISO 594-1:1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other
medical equipment — Part 1: General requirements | 6-11 |
| ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications —
Part 7: Connectors for intravascular or hypodermic applications | 5-115 |
| AIM 7351731 Rev. 2.00 2017-02-23 Medical Electrical Equipment and System
Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers | 19-30 |
10
Performance Testing – Animal
No animal data are included in this submission.
Performance Testing - Clinical
No clinical data are included in this submission.
Conclusion
The evaluation testing concludes that the Puritan Bennett™ Cuff Pressure Manager was found to be substantially equivalent to the predicate device.