K Number
K202652
Manufacturer
Date Cleared
2021-08-19

(339 days)

Product Code
Regulation Number
876.5880
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Servator P SALF Solution with THAM is indicated for flushing, storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.

Device Description

Servator P SALF Solution with THAM is a clear to light yellow, single use only, sterile (by steam sterilization) is indicated for the flushing, static cold storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into recipient. The solution is used for organ perfusion and hypothermic preservation. It is an ideal solution for the preservation of lungs.

The solution cooled to 4° to 8° C (39° to 46° F) is used to spray the organ isolated immediately after removal from the donor. The colloidal component Dextran 40 protects in particular microvessels from a possible damage caused by the post-ischemic reperfusion, since it prevents pathological leukocyte-endothelial interactions. In addition, Dextran 40 prevents edema and the formation of thrombi.

The primary containers used for the device are:

    1. PVC free bags 1000ml, therefore they are free of phthalates. The solution may be used without any point of use filtration.
    1. Clear glass bottles 25ml, therefore they are free of phthalates. The solution may be used without any point of use filtration.
AI/ML Overview

The provided FDA 510(k) summary for "Servator P SALF Solution with THAM" does not describe an AI/ML-based medical device. Instead, it details a solution used for organ preservation (specifically lungs) during transplantation.

Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, MRMC studies, standalone performance, and ground truth establishment for an AI/ML device is not applicable to this submission.

The document primarily focuses on demonstrating substantial equivalence to a predicate device (Perfadex® with THAM) based on:

  • Identical Indications for Use and Intended Use.
  • Identical Chemical Composition: This is explicitly stated multiple times ("The composition list is identical for the subject and predicate device composition and therefore, the subject and predicate devices are identical in chemical composition.").
  • Similar Technological Characteristics: Such as mode of operation (cold storage), container type, and sterilization method (steam).
  • Non-Clinical Performance Data:
    • Biocompatibility: Tests performed according to ISO 10993 series. The subject device passed all biocompatibility test standards.
    • Sterilization and Shelf Life Validation: Sterility validated according to ISO 17655-1. Shelf life is 24 months, identical to the predicate.
    • Performance Testing: Chemical comparisons and leachable performance testing were conducted to demonstrate substantial equivalence to the predicate.

In summary, this 510(k) submission establishes substantial equivalence through direct comparison of the product's attributes and non-clinical testing, rather than through a study proving an AI/ML device meets specific performance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 19, 2021

S.A.L.F. S.p.A % Joyce St. Germain Regulatory Consultant The 510k Consulting, LLC 1449 Springleaf Drive Ormond Beach, FL 32174

Re: K202652

Trade/Device Name: Servator P SALF Solution with THAM Regulation Number: 21 CFR§ 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: II Product Code: KDN Dated: July 15, 2021 Received: July 19, 2021

Dear Joyce St. Germain:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202652

Device Name Servator P SALF Solution with THAM

Indications for Use (Describe)

Servator P SALF Solution with THAM is indicated for flushing, storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a stylized key design. The key's head is formed by two intertwined lines, one in gold and the other in black, creating a diamond shape. The key's shaft is also made of intertwined gold and black lines, giving it a modern and abstract appearance. The background is plain white, which helps the key design stand out.

The 510k Consulting LLC

to success in obtaining your medical device clearance ingleaf Dr., Ormond Beach, FL 32174 da@gmail.com

510(k) Summary

Submitter/Applicant

S.A.L.F. S.p.A Via Marconi, 2, Cenate Sotto, BG, Italy Phone: +39-035-940097 Contact: Dr. Carmelo Gagliano (Quality Manager) carmelo.gagliano@salfspa.it

Date Prepared: September 4, 2020

Preparer/Consultant

The 510k Consulting, LLC 1449 Springleaf Drive Ormond Beach, FL 32174

Phone:904-477-3203
Contact:Joyce St. Germain, Regulatory Consultant, joyce510kfda@gmail.com

Device Classification

Trade/Device Name:Servator P SALF Solution with THAM
Common Name:Solution for lung preservation
Classification Name:System, Perfusion, Kidney
Regulation Name:Isolated kidney perfusion and transport system and accessories
Regulation Number:21 CFR 876.5880
Product Code:KDN
Regulatory Class:II
510k Review Panel:Gastroenterology/Urology Panel

Predicate Device

The subject device claims equivalence to the following legally marketed predicate:

510(k) Number:K091989
Date ClearedOctober 1, 2010
Submitter:XVIVO Perfusion AB
Trade Name:Perfadex® with THAM

{4}------------------------------------------------

Common Name:Solution for lung preservation
Classification Name:System, Perfusion, Kidney
Regulation Name:Isolated kidney perfusion and transport system and accessories
Regulation Number:21 CFR 876.5880
Product Code:KDN
Regulatory Class:II
Medical Specialty:Gastroenterology/Urology Panel

Indications for Use

Servator P SALF Solution with THAM is indicated for flushing, storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.

Intended Use

Organ storage and preservation for transplantation of lungs.

Device Description

Servator P SALF Solution with THAM is a clear to light yellow, single use only, sterile (by steam sterilization) is indicated for the flushing, static cold storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into recipient. The solution is used for organ perfusion and hypothermic preservation. It is an ideal solution for the preservation of lungs.

The solution cooled to 4° to 8° C (39° to 46° F) is used to spray the organ isolated immediately after removal from the donor. The colloidal component Dextran 40 protects in particular microvessels from a possible damage caused by the post-ischemic reperfusion, since it prevents pathological leukocyte-endothelial interactions. In addition, Dextran 40 prevents edema and the formation of thrombi.

The primary containers used for the device are:

    1. PVC free bags 1000ml, therefore they are free of phthalates. The solution may be used without any point of use filtration.
    1. Clear glass bottles 25ml, therefore they are free of phthalates. The solution may be used without any point of use filtration.

Comparison of Technological Characteristics with Predicate

  • The indications for use and intended use of the subject and predicate devices are identical.

{5}------------------------------------------------

  • The technologies are substantially equivalent as the composition of both solutions are . identical.
  • The subject and predicate devices are both supplied in bags with overbags for single . use.
  • The subject and predicate devices are both supplied sterile. ●
  • Tests were performed in order to confirm the equivalence between the subject and . predicate devices. The following table compares technological and other characteristics of the subject and predicate device.

Table of Comparison

Technological Comparison

Subject DevicePredicate DeviceComparison
DeviceServator P SALF Solutionwith THAMPerfadex and Perfadexwith THAMNA
ManufacturerSALF spa, ItalyXVIVO Perfusion AB,SwedenNA
510(k) NumberNAK091989NA
Classification &Product Code876.5880; KDN876.5880; KDNSame
RegulationNameIsolated Kidney Perfusionand Transport System andAccessoriesIsolated Kidney Perfusionand Transport System andAccessoriesSame
DeviceClassificationNameSystem, Perfusion, KidneySystem, Perfusion, KidneySame
Common NameSolution for lungpreservationSolution for lungpreservationSame
DeviceDescriptionServator P SALF Solutionwith THAM is a clear to lightyellow, single use only,sterile (by steam) is indicatedfor the flushing, static coldstorage and transportation ofisolated lungs after removal fromthe donor inpreparation of eventualtransplantation intorecipient. The solutionis used for organ perfusionand hypothermicpreservation. It is an idealsolution for the preservationof lungs.Perfadex® and Perfadex®with TRAM is a clear,sterile, non-pyrogenic,colloidbased, lightly buffered socalled "extracellular" lowpotassium dextran solutionprimarily for rapid cooling,perfusion and storage oflungs in connection withtransplantation.Same
Indication forUseServator P SALF Solutionwith THAM is indicated forflushing, storage andtransportation of isolatedlungs after removal fromthe donor in preparation foreventual transplantationinto a recipient.Perfadex® Solution forLung Perfusion is intendedfor the flushing, storageand transportation ofisolated lungs afterremoval from the donor inpreparation for eventualtransplantation into arecipient.Same
Intended UseOrgan storage andpreservation fortransplantationOrgan storage andpreservation fortransplantationSame
Mode ofoperationCold storageCold storageSame
Container/BagPVC free bagsPVC free bagsSame
Solutionqualitative andquantitativecompositionDextran 4050gGlucose monohydrate 1gPotassium chloride 0.4gSodium chloride8gMagnesium sulfateheptahydrate 0.201gPotassium dihydrogenphosphate0.063gDisodium phosphatedihydrate0.0576gSodium 138 mmol/lPotassium 6 mmol/lGlucose 5 mmol/lChlorides 142 mmol/lDextran 4050gGlucose monohydrate 1gPotassium chloride0.4gSodium chloride8gMagnesium sulfateheptahydrate 0.201gPotassium dihydrogenphosphate0.063gDisodium phosphatedihydrate0.0576gSodium 138 mmol/lPotassium 6 mmol/lGlucose 5 mmol/lChlorides 142 mmol/lSame
Total Phosphates 0.8 mmol/lWater for injections q.s. to1000mlTotal Phosphates 0.8mmol/lWater for injections q.s. to1000ml
Meets UNOSPolicyYesYesSame
PhysicalPropertiesThe solution has acalculated osmolarity ofabout 295 mOsmol/l. pH:5.3 to 5.5. Sterile pyrogen-free solution for organpreservation of Class II. Thesolution is clear, colorless orslightly yellow.Before use the pH of thesolution should be adjustedto 7.4 by adding a suitablebuffer 25 ml of THAM1mmol/l.The solution has acalculated osmolarity ofabout 295 mOsmol/l. pH:5.3 to 5.5. Sterile pyrogen-free solution for organpreservation of Class II.The solution is clear,colorless or slightlyyellow.Before use the pH of thesolution should be adjustedto 7.4 by adding a suitablebuffer 25 ml of THAM1mmol/l.Same
ContainerPVC free bagsPVC free bagsSame
ParticulateMatter andBiocompatibilityParticle Counts less thanlimits for Large VolumeInjections per USP <788>;Biocompatible per ISO10993-1 battery of tests forExternally CommunicatingBlood Path Indirect Contactfor prolonged periods >24hours.Particle Counts less thanlimits for Large VolumeInjections per USP <788>;Biocompatible per ISO10993-1 battery of testsfor ExternallyCommunicating BloodPath Indirect Contact forprolonged periods >24hours.Same
DeviceStandards ofConformityISO 10993-4ISO 10993-5ISO 10993-10ISO 10993-11ISO 11607-1ISO 11607-2ISO 11737-1ISO 17665ISO 10993 SeriesUSP <788>USP <1211>Exact test series ofpredicate device areunknownSubjectdevicepassedaccordingto ISOStandards
ISO 14971ISO 15223-1ISO 17025ISO 17665-1USP 39 <71>USP 39 <85>USP 41 <151>USP <1211>
ProtectingOverwrap bagYesYesSame
Bag connections1 flip off, 1 needle point1 flip off, 1 needle pointSame
Single use onlyYesYesSame
SterilizationSterilization processesvalidated accordingto ISO 17665 or USP<1211>Sterilization processesvalidated accordingto ISO 17665 or USP<1211>Same
SterilizationmethodSteamSteamSame
Nominal value1000 mL bags + THAM1000 mL bags + THAMSame
Shelf Life24 months24 monthsSame
StorageTemperatureStore below 30°C. Donot freeze.Store in its originalcontainer. Do not removethe overwrap untilimmediately before use.The device is sterileand disposable. Thesolution must be usedfor one singleuninterruptedadministration and anyresidue must bediscarded to avoidrisk of contaminationdue to loss of sterility.Do not use if thesolution is frozen.Store below 30°C. Donot freeze.Store in its originalcontainer. Do not removethe overwrap untilimmediately before use.The device is sterileand disposable. Thesolution must be usedfor one singleuninterruptedadministration and anyresidue must bediscarded to avoidrisk of contaminationdue to loss of sterility.Do not use if thesolution is frozen.Same
Interaction withother medicaltechnologyNot intended for continuousperfusion. Standardtransplantation surgicalexpertise and techniques arerequired.Not intended forcontinuous perfusion.Standard transplantationsurgical expertise andtechniques are required.Same

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

TestSALFPredicate
Temperature of storage ofproduct<30°C*; Do not freeze2°-25°C
Temperature of transport4-8°C4-8°C
Shelf life2 years2 years
Shelf life afterreconstitution24 hours24 hours

*It is calculated on basis of stability study. The product is stable at 25℃ for 2 years and at least 6months at 40°C.

About the storage condition please let me know if we have to restrict the rage from <30°C*; Do notfreeze to 2°-25°C.

The above comparison shows the subject and predicate devices are identical in eachattribute/characteristic.

The above comparison shows the subject and predicate devices are identical in each attribute/characteristic.

NOTE. . . The composition list is identical for the subject and predicate device composition and therefore, the subject and predicate devices are identical in chemical composition.

The above comparison shows the subject and predicate devices are substantially equivalent in technology characteristics.

The Servator P SALF Solution with THAM have the same indication for use, intended use, design, materials, packaging and other technological characteristics to the predicate device.

{10}------------------------------------------------

Non-Clinical Performance Data

The following performance data is provided in support of the substantial equivalence determination. All tests performed are included in this submission.

Biocompatibility ... is required for this device. The tests were all performed according to the ISO 10993 series that are listed in the Table of Comparison above. The subject device passed all biocompatibility test standards.

Sterilization and Shelf Life . . . is required for the subject device. The Validation of Sterility was performed, and the results passed according to ISO 17655-1. Steam sterilization and storage conditions are the subject and predicate device. Shelf life for the subject and predicate device is the same at 24 months.

Electrical Safety and EMC . . . testing was not applicable for this device.

Software ... was not applicable for this device.

Performance Testing...was completed as a direct comparison between the subject and predicate device. The chemical comparisons and leachable performance testing demonstrated the substantial equivalence of this device to the predicate.

Conclusion

The subject and predicate devices have the same indications for use and the same intended use.

Both devices are substantially equivalent in design, materials, packaging and other technological characteristics and performance (since they have, in fact, the same chemical composition).

The Servator P SALF Solution with THAM does not raise any questions regarding safety and effectiveness and is equivalent to the predicate device. The non-clinical data supports and demonstrates the safety of the device.

The conclusion is that Servator P SALF Solution with THAM warrants a finding of substantial equivalence to the legally marketed Perfadex® with THAM solution, and therefore, should have clearance for premarket activities in the United States.

§ 876.5880 Isolated kidney perfusion and transport system and accessories.

(a)
Identification. An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.(b)
Classification. Class II (performance standards).