K Number
K202647
Device Name
Surgical Mask
Date Cleared
2021-04-26

(227 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Surgical Masks are single use, three-layer, flat-pleated style with ear loops and nose piece. The Surgical Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of Polyvinyl chloride coated iron wire. The surqical masks will be provided in blue. The surqical masks are sold non-sterile and are intended to be single use, disposable devices.

AI/ML Overview

The provided document is a 510(k) Pre-Market Notification for a Surgical Mask (K202647). It does not include information about AI/ML device performance or clinical studies in humans.

This document describes the performance of a physical medical device (a surgical mask) against established standards. The questions you've asked are typically relevant for AI/ML-powered medical devices that analyze medical images or data.

Therefore, many of your questions, such as those related to "AI," "human readers," "effect size," "training set," and "ground truth for training set," are not applicable to the content of this document.

However, I can extract the relevant information regarding the surgical mask's acceptance criteria and the study that proves it meets those criteria.


Acceptance Criteria and Device Performance for Surgical Mask (K202647)

1. A table of acceptance criteria and the reported device performance

Test Method (Standard)Acceptance CriteriaReported Device PerformanceResult
Fluid Resistance (ASTM F1862)29 out of 32 pass at 120 mmHg (16.0 kPa)31 out of 32 pass at 120 mmHg (16.0 kPa)Pass
Particulate Filtration Efficiency (ASTM F2299)≥ 98%Average 99.6%Pass
Bacterial Filtration Efficiency (ASTM F2101)≥ 98%Average 99.75%Pass
Differential Pressure (Delta P) (MILM-36954C)

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.