K Number
K202647

Validate with FDA (Live)

Device Name
Surgical Mask
Date Cleared
2021-04-26

(227 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Surgical Masks are single use, three-layer, flat-pleated style with ear loops and nose piece. The Surgical Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of Polyvinyl chloride coated iron wire. The surqical masks will be provided in blue. The surqical masks are sold non-sterile and are intended to be single use, disposable devices.

AI/ML Overview

The provided document is a 510(k) Pre-Market Notification for a Surgical Mask (K202647). It does not include information about AI/ML device performance or clinical studies in humans.

This document describes the performance of a physical medical device (a surgical mask) against established standards. The questions you've asked are typically relevant for AI/ML-powered medical devices that analyze medical images or data.

Therefore, many of your questions, such as those related to "AI," "human readers," "effect size," "training set," and "ground truth for training set," are not applicable to the content of this document.

However, I can extract the relevant information regarding the surgical mask's acceptance criteria and the study that proves it meets those criteria.


Acceptance Criteria and Device Performance for Surgical Mask (K202647)

1. A table of acceptance criteria and the reported device performance

Test Method (Standard)Acceptance CriteriaReported Device PerformanceResult
Fluid Resistance (ASTM F1862)29 out of 32 pass at 120 mmHg (16.0 kPa)31 out of 32 pass at 120 mmHg (16.0 kPa)Pass
Particulate Filtration Efficiency (ASTM F2299)≥ 98%Average 99.6%Pass
Bacterial Filtration Efficiency (ASTM F2101)≥ 98%Average 99.75%Pass
Differential Pressure (Delta P) (MILM-36954C)< 6.0 mmH2O/cm²Average 2.2 mmH2O/cm²Pass
Flammability (16 CFR 1610)Class 1Class 1Pass
Cytotoxicity (ISO 10993-5)Not cytotoxicity effectDevice is noncytotoxicPass
Irritation (ISO 10993-10)Not an irritantDevice is nonirritatingPass
Sensitization (ISO 10993-10)Not an sensitizationDevice is nonsensitizingPass

2. Sample size used for the test set and the data provenance

The document mentions "31 out of 32" for the Fluid Resistance test, indicating a sample size of 32 for that specific test. Sample sizes for other performance tests (Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability) are not explicitly stated in the summary, though "Average" values imply multiple samples were tested.

For biocompatibility tests (Cytotoxicity, Irritation, Sensitization), standard ISO methods were used, but the specific number of test items or biological replicates is not provided in this summary.

Data Provenance: The tests were conducted on the "Subject Device" (Ningbo Green Textile Co., Ltd. Surgical Mask, K202647). The location of testing is not explicitly stated, but the manufacturer is based in China. The data would be considered prospective as it was generated specifically for this submission to demonstrate compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable as the device is a physical surgical mask, not an AI/ML system requiring expert interpretation of data. The "ground truth" for the performance tests consists of objective measurements against established engineering and biological standards.

4. Adjudication method for the test set

This question is not applicable. The performance testing involves objective measurements rather than human interpretation that would require adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. This document describes a physical medical device (surgical mask), not an AI/ML algorithm intended to assist human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. This document describes a physical medical device, not an algorithm.

7. The type of ground truth used

The ground truth used for proving the device meets the acceptance criteria are established performance standards and laboratory test results for surgical masks, as specified by ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) guidelines, and other regulatory standards (e.g., 16 CFR 1610 for flammability, MILM-36954C for differential pressure). These are objective, measurable criteria.

8. The sample size for the training set

This question is not applicable. There is no "training set" as this is not an AI/ML device.

9. How the ground truth for the training set was established

This question is not applicable. There is no "training set" or corresponding ground truth establishment process for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 26, 2021

Ningbo Green Textile Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K202647

Trade/Device Name: Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX. Dated: March 19, 2021 Received: March 23, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202647

Device Name Surgical Mask

Indications for Use (Describe)

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K202647

This 510(k) summary is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's Information

Name: NINGBO GREEN TEXTILE CO.,LTD. Address:NO.498,3RD XINGCI ROAD,HANGZHOU BAY NEW ZONE, NINGBO,ZHEJIANG,CHINA 315336 Tel: 86-574-63541998 Fax: 86- 574-63528444 Contact: Carol Date of Preparation: Mar.19,2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Surqical Mask Common name: Surgical Face Mask Classification name: Surgical Face Mask Green2020-32 Model:

3.0 Classification

Production code: FXX Regulation number: 21CFR 878.4040 Classification: Class II Surgical Apparel Panel:

4.0 Predicate Device Information

Manufacturer: WUHAN DYMEX HEALTHCARE CO., LTD. Device: SURGICAL FACE MASK

{4}------------------------------------------------

510(k) number: K182515

5.0 Device Description

The Surgical Masks are single use, three-layer, flat-pleated style with ear loops and nose piece. The Surgical Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of Polyvinyl chloride coated iron wire. The surqical masks will be provided in blue. The surqical masks are sold non-sterile and are intended to be single use, disposable devices.

6.0 Indication for Use Statement

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

7.0 Comparison to the Predicate Device

ItemSubject DeviceK202647Predicate DeviceK182515Remark
Product NameSurgical MaskSurgical Face Mask--
Product CodeFXXFXXSame
Regulation No.21 CFR 878.404021 CFR 878.4040Same
ClassIIIISame
510(k) Summary
Intended Use&Indications for useThe Surgical Mask isintended to be worn toprotect both the patient andhealthcare personnel fromtransfer of microorganisms,body fluids and particulatematerial. These face masksare intended for use ininfection control practices toreduce the potentialexposure to bloodand body fluids. This is asingle use, disposabledevice(s),provided non-sterile.The Surgical FaceMasks are intended to beworn to protect both thepatient and healthcarepersonnel from transferofmicroorganisms,bodyfluids and particulatematerial. These facemasks are intended foruse in infection controlpractices to reduce thepotential exposure toblood and body fluids.This is a single use,disposable device(s),provided non-sterile.Same
Design featuresEar Loops, 3 layersEar Loops, 3 layersSame
Mask StylesFlat pleatedFlat pleatedSame
MaterialOuterfacinglayerSpun-bond polypropyleneSpun-bondpolypropyleneSame
MiddlelayerMelt blown polypropylenefilterMelt blownpolypropylene filterSame
InnerFacinglayerSpun-bond polypropyleneSpun-bondpolypropyleneSame
NosepiecePolyvinyl chloride coatediron wireMalleable polyethylenewireDifferent
Earloopsnot made with naturalrubber latexnot made with naturalrubber latexSame
ColorBlueBlueSame
Dimension(Length)17.5 cm +/- 1cm17.5 cm +/- 1cmSame
Dimension (Width)9.5 cm +/- 1cm9.5 cm +/- 1cmSame
OTC useYesYesSame
SterilityNon-SterileNon-SterileSame
Single UseYesYesSame
ASTM F2100 LevelLevel 2Level 2Same
510(k) Summary
Biocompatibility
CytotoxicityComply with ISO 10993-5;Under the conditions of thestudy, not cytotoxicityeffect.Comply with ISO10993-5Same
SensitizationComply with ISO10993-10;Under the conditions of thestudy, not an sensitizationComply with ISO10993-10Same
IrritationComply with ISO10993-10;Under the conditions of thestudy, not an irritant.Comply with ISO10993-10Same

Table 1 General Comparison

{5}------------------------------------------------

{6}------------------------------------------------

8.0 Non-Clinical Test Conclusion

The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004.

Table 2 - Performance Testing

Test MethodSubject DeviceAcceptance CriteriaResult
FluidResistancePerformanceASTM F186231 Out of 32 passat 120mmHg(16.0 kPa)29 Out of 32 pass at120 mmHg (16.0kPa)Pass
ParticulateFiltrationEfficiencyASTM F2299Average 99.6%≥ 98%Pass
BacterialFiltrationEfficiencyASTM F2101Average 99.75%≥ 98%Pass

{7}------------------------------------------------

510(k) Summary

DifferentialPressure(Delta P) MILM-36954CAverage 2.2mmH20/cm²< 6.0 mmH20/cm²Pass
Flammability16 CFR 1610Class 1Class 1Pass

Table 3 - Biocompatibility Testing

Test MethodSubject DeviceResult
CytotoxicityUnder the conditions of the study, the deviceis noncytotoxic.Pass
IrritationUnder the conditions of the study, the deviceis nonirritating.Pass
SensitizationUnder the conditions of the study, the deviceis nonsensitizingPass

9.0 Clinical Test Conclusion

No clinical study is included in this submission.

10.0 Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device in K182515.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.