K Number
K202601
Date Cleared
2020-12-15

(98 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is a fiber optic delivery device intended for use with the laser system for its FDA cleared indications for use. The product can be used for the surgical incision/excision, ablation, hemostasis and coagulation of soft tissue. All soft tissue is included, such as skin, cutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs and glands.

The product will deliver 532 nm laser energy from a compatible laser console (Surgical Green Laser System) to tissue during surgical procedures, including photo-selective vaporization of the prostatic hyperplasia (BPH).

Device Description

The single-use sterile medical laser fiber is designed, manufactured by Realton (Suzhou) Medical Technology Co., Ltd.. The product can be used for the surgical incision/excision, vaporization, ablation, hemostasis and coagulation of soft tissue. The optical fiber material is fused quartz, and its core diameter is 760 µm. One end of the optical fiber is a special connector, which can be connected to the medical laser surgery system. The output structure of the other end of the optical fiber can be divided into side-firing type and front-firing type: W760SF-30, B760SF-30, G760SF-30 and P760SF-30 are side-firing types, W760F-30, B760FF-30, G760FF-30 and P760FF-30 are front-firing types.

The product is sterilized by EO. It is a disposable medical device.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device: a "Single-use sterile medical laser fiber." This type of document is submitted to the FDA to demonstrate that a device is substantially equivalent to a legally marketed predicate device, rather than proving a specific level of clinical performance against pre-defined acceptance criteria in the manner of a novel AI/ML device.

Therefore, the information typically requested for AI/ML device studies (such as specific performance metrics, sample sizes for test and training sets, expert adjudication, MRMC studies, and ground truth establishment) is not present in this document. The document focuses on showing substantial equivalence through comparison of technical specifications, intended use, and non-clinical testing (biocompatibility and general performance testing to ensure the device meets its own specifications).

Here's what can be extracted from the document regarding acceptance criteria and performance, as well as the types of studies performed:

1. A table of acceptance criteria and the reported device performance

The document mentions "predefined acceptance criteria" for performance testing, but the specific acceptance criteria and the detailed reported device performance values are not explicitly listed in a table format in the provided text.

However, the following specifications are given for the device, and the document states that performance testing confirmed these were met:

ItemSpecification
Overall Length300 cm
Core Diameter0.76 mm
Applicable wave lengths532 nm
Minimum transmission efficiency80 %
Maximum transmission power180 W
Tensile strength10 N
Minimum bending working radius500 mm
Light output angle (side-firing types)70°± 20°
ConnectorSMA-905

The document states: "All of the tested parameters met the predefined acceptance criteria."

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document refers to "Performance testing" and "Biocompatibility testing" but does not detail the sample sizes for these tests or the data provenance. These are likely engineering and material tests, not clinical studies with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This information is not applicable/provided. This device is a physical medical instrument (laser fiber), not an AI/ML algorithm that requires expert labeling for ground truth. Technical specifications are likely verified by engineers or lab technicians using calibrated equipment.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This information is not applicable/provided. Adjudication methods are relevant for subjective interpretations of data, typically in AI/ML or clinical imaging studies. The tests mentioned are objective performance and biocompatibility assessments.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable/provided. This is a physical medical device, not an AI-assisted diagnostic or treatment planning system. Therefore, MRMC studies are not relevant to its 510(k) submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable/provided. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is not applicable/provided in the context of clinical ground truth. For the performance and biocompatibility tests, "ground truth" would be established by standard engineering and chemical/biological testing methodologies that define material properties, optical performance, and biological safety. The document implies that industry standards (e.g., ISO 10993-1:2018 for biocompatibility) and internal test protocols were used to define and verify these "truths."

8. The sample size for the training set

This information is not applicable/provided. This is a physical medical device, not an AI/ML system that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable/provided. As above, this is not an AI/ML system.


Summary of Device Performance Study as Described in the Document:

The document describes non-clinical testing to support the substantial equivalence determination. These tests include:

  • Biocompatibility testing: Conducted according to ISO 10993-1:2018. This included tests for Cytotoxicity, Sensitization, Irritation, and Systemic toxicity. The document states these tests were performed and implies they met acceptance criteria (though specific results are not provided).
  • Performance testing: This involved evaluating "technical parameters about dimension, optical performance and mechanical strength." Additionally, "shelf life was evaluated." The document explicitly states: "All of the tested parameters met the predefined acceptance criteria."

The "study" in this context is a series of engineering and laboratory tests to confirm the physical and biological characteristics of the laser fiber, not a clinical study involving patients or AI model validation. The purpose is to demonstrate that the device performs according to its specifications and is safe for its intended use, thereby showing substantial equivalence to a predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

December 15, 2020

Realton (Suzhou) Medical Technology Co., Ltd % Olivia Meng Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Service Co., Ltd 8-9th Floor, R&D Building, No. 26 Oinglan Street, Panvu District Guangzhou, Guangdong 510006, China

Re: K202601

Trade/Device Name: Single-use sterile medical laser fiber Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 26, 2020 Received: November 30, 2020

Dear Olivia Meng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202601

Device Name Single-use sterile medical laser fiber

Indications for Use (Describe)

The device is a fiber optic delivery device intended for use with the laser system for its FDA cleared indications for use. The product can be used for the surgical incision/excision, ablation, hemostasis and coagulation of soft tissue. All soft tissue is included, such as skin, cutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs and glands.

The product will deliver 532 nm laser energy from a compatible laser console (Surgical Green Laser System) to tissue during surgical procedures, including photo-selective vaporization of the prostatic hyperplasia (BPH).

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for "REALTON Real Estate Corporation". The logo is in white text on a green background. To the left of the word REALTON is a white graphic that looks like a stylized letter R.

Page 1 of 5

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

    1. SUBMITTER
      Realton (Suzhou) Medical Technology Co., Ltd.

601 Room, B8 Building, 218 Xinghu Road, Suzhou Industrial Park, Suzhou, P.R.China Phone: +86-512-62868599

Primary ContactOlivia Meng
Person:Regulatory Affairs Manager
Guangzhou Osmunda Medical Device Technical Service
Co., Ltd.
8-9th Floor, R&D Building, No. 26 Qinglan Street, Panyu
District, Guangzhou, CHN 510006 Guangdong
Tel: (+86)-20-6231 6262
Fax: (+86) -20-8633 0253
Secondary ContactMingzhu Liu
Person:Quality Manager
Realton (Suzhou) Medical Technology Co., Ltd.
Tel: (+86)-512-62868599
Date preparedDecember 11, 2020
2. DEVICE
Device Name:Single-use sterile medical laser fiber
Common/Usual Name:Single-use sterile medical laser fiber
Model:W760SF-30, B760SF-30, G760SF-30, P760SF-30, W760F-30, B760FF-30, G760FF-30 and P760FF-30
Regulation number21 CFR 878.4810
Regulation Class:II
Product Code:GEX (Powered Laser Surgical Instrument)
Regulation Name:Laser Surgical Instrument for Use in General and Plastic Surgery and InDermatology

PREDICATE DEVICE 3.

K120870 GreenLight MoXyTM Fiber Optic

    1. DEVICE DESCRIPTION

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Realton Realtime Corporation. The logo is white text on a green background. The logo consists of a stylized "R" symbol to the left of the word "REALTON" in all caps. Below the word "REALTON" is the text "Realtime Corporation" in a smaller font.

The single-use sterile medical laser fiber is designed, manufactured by Realton (Suzhou) Medical Technology Co., Ltd.. The product can be used for the surgical incision/excision, vaporization, ablation, hemostasis and coagulation of soft tissue. The optical fiber material is fused quartz, and its core diameter is 760 µm. One end of the optical fiber is a special connector, which can be connected to the medical laser surgery system. The output structure of the other end of the optical fiber can be divided into side-firing type and front-firing type: W760SF-30, B760SF-30, G760SF-30 and P760SF-30 are side-firing types, W760F-30, B760FF-30, G760FF-30 and P760FF-30 are front-firing types.

The product is sterilized by EO. It is a disposable medical device.

ItemSpecification
Overall Length300 cm
Core Diameter0.76 mm
Applicable wave lengths532 nm
Minimum transmission efficiency80 %
Maximum transmission power180 W
Tensile strength10 N
Minimum bending working radius500 mm
Light output angle (side-firing types)70°± 20°
ConnectorSMA-905

The specification of the single- use sterile medical laser fiber is as followed:

5. INDICATIONS FOR USE

The device is a fiber optic delivery device intended for use with the laser system for its FDA cleared indications for use.

The product can be used for the surgical incision/excision, vaporization, ablation, hemostasis and coagulation of soft tissue. All soft tissue is included, such as skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs and glands.

The product will deliver 532 nm laser energy from a compatible laser console (Surgical Green Laser System) to tissue during surgical procedures, including photo-selective vaporization of the prostate for benign prostatic hyperplasia (BPH).

The indications for use statement for the subject device is same as the predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for EALTON Realtime Corporation. The logo is white text on a green background. The logo has a stylized letter "R" on the left side of the word EALTON. Below the word EALTON is the text "Realtime Corporation" in a smaller font.

Page 3 of 5

SpecificationSubject devicePredicate deviceDiscussionofdifference
Device nameSingle-use Sterile Medical Laser FiberGreenLight MoXy™ Fiber OpticNA
ManufacturerRealton (Suzhou) Medical TechnologyCo., Ltd.American Medical SystemsNA
510(k)numberNAK120870NA
Product codeGEXGEXSame
Intended useThe device is a fiber optic delivery deviceintended for use with the laser systemfor its FDA cleared indications for use.The product can be used for the surgicalincision/excision, vaporization, ablation,hemostasis and coagulation of softtissue. All soft tissue is included, such asskin, cutaneous tissue, subcutaneoustissue, striated and smooth tissue,muscle, cartilage meniscus, mucousmembrane, lymph vessels and nodes,organs and glands.The product will deliver 532 nm laserenergy from a compatible laser console(Surgical Green Laser System) to tissueduring surgical procedures, includingphoto-selective vaporization of theprostate for benign prostatic hyperplasia(BPH).Model Number 0010-2400 is a fiberoptic delivery device intended for usewith the GreenLight™ XPS LaserSystem for its FDA cleared indicationsfor use.The Model Number 0010-2400features a side firing mechanismdelivering up to 180W of 532nm lightto tissue. Model Number 0010-2400can be used for the surgicalincision/excision, vaporization,ablation, hemostasis and coagulationof soft tissue. All soft tissue is included,such as skin, cutaneous tissue,subcutaneous tissue, striated andsmooth tissue, muscle, cartilagemeniscus, mucous membrane, lymphvessels and nodes, organs and glands.Model Number 0010-2400 will deliver532 nm laser energy from a compatiblelaser console (GreenLight™ XPS LaserSystem) to tissue during surgicalprocedures, including photoselectivevaporization of the prostate for benignprostatic hyperplasia (BPH).Same
Contraindica-tionsThe fiber is contraindicated in thepresence of severe urethral strictures;however, the system can be used in thetreatment of urethral strictures withproper cautions. A severe stricture is anystricture with visible narrowing viaurethrography or ultrasonography, withnear total obstruction that make passageof instruments difficult or dangerous.The fiber is contraindicated in thepresence of severe urethral strictures;however, the system can be used inthe treatment of urethral strictureswith proper cautions. A severestricture is any stricture with visiblenarrowing via urethrography orultrasonography, with near totalobstruction that make passage ofinstruments difficult or dangerous.Same
Prescription orOTCPrescriptionPrescriptionSame
ComponentsThe device is mainly composed of fourparts: fiber connect part (cap forconnector, connector), fibertransmission part, fiber handle andtreatment part (Side-firing type: bareThe device is mainly composed ofmetal cap, laser firing point, outerflow tubing, control knob, fiberhandle, inlet flow tubing andconnector, etc.Similar
indicator and fiber cap; front-firing type:bare fiber and laser firing point).
TypeFront-firing and side-firingSide-firingSimilar
MaterialoffiberFused quartzFused quartzSame
Fiber core size0.76 mm0.75 mmSimilar
Catheterlength300 cm305 cmSimilar
Operationlength34.5 cm36 cmSimilar
Power densityat 2 mm@180W=326 W/mm2@120W=218 W/mm2@180W=326 W/mm2@120W=218 W/mm2Same
Applicablewave lengths532 nm532nmSame
Maximumtransmissionpower180 W180 WSame
Total amountof energytransferred643 KJ643 KJSame
ConnectorSMA-905 standard optical connectorOptical ConnectorDifferent
Laser systemscompatibilitySurgical Green Laser SystemGreenLight™ XPS Laser SystemDifferent
Endoscope/Cystoscopecompatibility24 Fr continuous flow22 to 24 Fr continuous flowSimilar
SterilizationEthylene oxideEthylene oxideSame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for EATON. The logo is white text on a green background. The text "EATON" is in a sans-serif font, and below it is the text "Realiza Corporation" in a smaller font. To the left of the text is a stylized symbol that looks like a stylized letter "R".

Page 4 of 5

The indications for use and technological characteristics of the subject and predicate devices are comparable. The subject device and the predicate device differ in some technical specifications such as components, connector and compatible laser systems. However, these differences do not raise different questions of safety and effectiveness.

7. SUMMARY OF NON-CLINICAL TESTING

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for single-use sterile medical laser fiber was conducted in accordance with the International Standard ISO 10993-1:2018, "Biological evaluation

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for EALTON Realtime Corporation. The logo is white text on a green background. To the left of the word EALTON is a white symbol that looks like a stylized letter R. Below the word EALTON is the text Realtime Corporation in a smaller font.

of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:

  • Cytotoxicity
  • Sensitization ●
  • Irritation
  • Systemic toxicity

Performance testing

Performance testing was conducted on single-use sterile medical laser fiber. Technical parameters about dimension, optical performance and mechanical strength were evaluated in the performance testing. In addition, the shelf life was evaluated. All of the tested parameters met the predefined acceptance criteria.

8. CONCLUSION

The indications for use statement for the subject device is the same as that of the predicate. The differences between the subject device and its predicate device do not raise different questions of safety and effectiveness. The non-clinical data support the safety of the device and the performance testing report demonstrate that single-use sterile medical laser fiber should perform as intended in the specified application conditions.

From the results of non-clinical data including the performance testing described, Realton (Suzhou) concludes that the subject device single-use sterile medical laser fiber is substantially equivalent to the predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.