(294 days)
The COR-KNOT MICRO™ device with COR-KNOT MICRO™ fastener when used in conjunction with USP 6-0, 7-0, or 8-0 polypropylene surgical suture, is indicated for use in the approximation of soft tissue.
The COR-KNOT MICRO™ Device is a sterile, single patient use device provided to the user preloaded with a single COR-KNOT MICRO™ fastener. Made from medical grade titanium, the COR-KNOT MICRO™ fastener is a hollow sleeve with a rounded base. A white plastic target holds the loop shape of a wire snare. The wire snare passes through the COR-KNOT MICRO™ fastener and is attached to a snare puller knob. A suture slot in the device shaft lies under the opening in the snare puller. The ends of a USP 6-0, 7-0, or 8-0 polypropylene suture are passed through the wire snare and subsequently threaded into the titanium fastener. The snare puller attached to its wire snare is pulled up along the device shaft until it snaps onto the puller retainer feature of the purple knob, which also has an integrated indicator fin. The suture slot and the indicator fin are located on the same side of the device shaft. The subsequently crimped fastener and remnant trimmed suture tails bend slightly in the direction away from or opposite from the suture slot and indicator fin. By rotating the purple knob and the device's white handle, the surgeon can ergonomically orient the direction of the suture tails, if desired. A yellow lever stop is located behind the purple lever to restrict inadvertent squeezing of the lever during device handling before crimping. The lever stop is removed by pinching its sides together and pulling it out of the handle. By squeezing the purple lever, the COR-KNOT MICRO™ Device crimps the COR-KNOT MICRO™ fastener to fasten together segments of suture and trims away excess suture.
One COR-KNOT MICRO™ Device is provided per package, and the COR-KNOT MICRO™ Device is not intended to be reloaded or reused.
The COR-KNOT MICRO™ device is intended for use in the approximation of soft tissue.
The acceptance criteria and device performance information for the COR-KNOT MICRO™ device are derived from the provided FDA 510(k) summary.
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance (COR-KNOT MICRO™) |
|---|---|---|
| Functional Verification & Validation | ||
| Knot Pull Apart (KPA) Force | A Fastener must secure 6-0, 7-0, and 8-0 polypropylene suture with knot pull apart (KPA) forces exceeding USP standards. Both the predicate and subject devices exceed USP KPA specifications for their compatible suture. | The COR-KNOT MICRO™ device successfully secured 6-0, 7-0, and 8-0 polypropylene suture, exceeding USP KPA standards. This was demonstrated through bench testing and a simulated "malignant hypertension pressure simulator" study. In the latter, 60 fasteners (20 for each suture size) held without compromise through 5,591,970 distinct hypertensive pulses each, totaling over 335 million pulses cumulatively. |
| Fastener Integrity | Fastener shall not be damaged prior to deployment. | All functional verification and validation tests were met, including ensuring the fastener was not damaged prior to deployment. |
| Device Retention of Fastener | Device shall retain fastener prior to crimping/cutting without the fastener dislodging from the tip of the device. | All functional verification and validation tests were met. |
| Fastener Release | Releasing the crimped fastener from the device shall not damage the fastener. | All functional verification and validation tests were met. |
| Suture Cutting | The device shall cut the suture after a completed crimp. | All functional verification and validation tests were met. |
| Tissue Damage | Crimped fastener shall not cause more damage to anatomy than hand tied suture/knots. | An acute animal study compared COR-KNOT MICRO™ fasteners to hand-tied knots. The COR-KNOT MICRO™ titanium fastener provided superior or equivalent – and never inferior – results for ease of use, wound approximation, hemostasis, burst strength, and tensile strength compared to controls. |
| Device Compatibility | The device shall be compatible with only COR-KNOT MICRO™ fasteners. | All functional verification and validation tests were met. |
| Suture Compatibility | System compatible with 6-0, 7-0, and 8-0 sized polypropylene sutures. | The device is specifically designed and indicated for use with USP 6-0, 7-0, or 8-0 polypropylene surgical suture. Performance testing confirmed compatibility and efficacy with these suture sizes. |
| Reloading Prevention | The device will prevent reloading in the field. | The device is explicitly stated as sterile, single-patient use, and not intended to be reloaded or reused. Functional testing verified this. |
| Shaft Length | Inserted shaft length shall be comparable to other related surgical devices demonstrated to reach superficial surgical access sites and most open and minimally invasive surgical site locations. | All functional verification and validation tests were met. |
| Surgical Site Visibility | Device shall allow for surgical site visibility. | All functional verification and validation tests were met. |
| Tissue Damage Mitigation | Inserted portion of device and entirety of fastener shall be shaped to reduce risk of tissue damage or other iatrogenic injury. | All functional verification and validation tests were met, supported by the animal study showing no inferior results for tissue damage compared to hand-tied knots. |
| Function in Surgical Conditions | The device and fastener must function with exposure to typical surgical conditions. | All functional verification and validation tests were met. The malignant hypertension simulation specifically tested extreme conditions. |
| Single Fastener Firing | Device must be capable of firing the single fastener provided with the device. If multiple fasteners are required within a case, multiple devices must be used. | All functional verification and validation tests were met. |
| Suture Loading Ease | The device shall allow suture to be easily loaded through the titanium fastener and distal end of the device. | All functional verification and validation tests were met. |
| Lever Stop Removal | The user must be able to remove Lever Stop prior to use of the device. | All functional verification and validation tests were met. |
| One-Handed Crimping | The user must be able to crimp a fastener with one hand. | All functional verification and validation tests were met. |
| User Safety | The device shall not cause harm to the user. | All functional verification and validation tests were met. |
| Biocompatibility | Per ISO 10993-1:2018. The COR-KNOT MICRO™ device is made from identical materials as the predicate device (stainless steel and polymer delivery device with a titanium fastener), which conforms to ISO 10993. | The device successfully met biocompatibility requirements per ISO 10993-1:2018. |
| Packaging/Shelf-life | Per ISO 11607-1 demonstrating a 3 year shelf life. The device is comparably packaged to the predicate (rigid thermoformed blister tray with Tyvek cover vs. PETG tray and snap retainer lid sealed in Tyvek/nylon pouch). | Packaging/Shelf-life testing per ISO 11607-1 demonstrated a 3-year shelf life. |
| Durability/Worst-Case Scenario | Secure and maintain closure under extreme physiological conditions (e.g., supraphysiologic pressure, prolonged healing time, high heart rate, worst-case wound repair). | A custom malignant hypertension pressure simulator study successfully demonstrated the strength and stability of the fasteners. 60 fasteners (20 of each suture size: 6-0, 7-0, 8-0) held without compromise for 6 weeks under >360/240 mmHg arterial pressure at 90 pulses/minute, totaling 5,591,970 distinct hypertensive pulses per closure. Burst strengths in an acute animal study also exceeded supraphysiologic worst-case anticipated conditions. |
| Ease of Use & Efficacy in Vivo | Comparable or superior performance to hand-tied knots regarding ease of use, wound approximation, hemostasis, burst strength, and tensile strength in an acute animal model. | An acute animal study confirmed that the device is easy to use and provides excellent wound approximation resulting in hemostasis, with superior or equivalent (never inferior) results compared to hand-tied knots for the specified endpoints. Burst strength of harvested specimens exceeded supraphysiologic worst-case anticipated conditions. |
Study Details:
-
A table of acceptance criteria and the reported device performance: (Provided above)
-
Sample sizes used for the test set and the data provenance:
- Bench Testing (Functional Verification & Validation): The specific sample sizes for each of the functional verification and validation tests are not explicitly stated as numerical values in the provided text. The text states "all results met the performance and risk-based acceptance criteria," implying that adequate samples were tested for each criterion.
- Malignant Hypertension Pressure Simulator (Durability):
- Test Set: 60 COR-KNOT MICRO™ titanium fasteners (20 each for 6-0, 7-0, and 8-0 polypropylene suture).
- Data Provenance: Retrospective, conducted in a laboratory setting (bench testing). No geographical data provenance is specified, but it's an in-vitro simulation.
- Acute Animal Study (Safety & Effectiveness):
- Test Set: Specific number of animal subjects is not given, but the study compared the COR-KNOT MICRO™ Device in worst-case size blood vessels to hand-tied knots under the same conditions.
- Data Provenance: Prospective, conducted in an animal model (in vivo pig model). No geographical data provenance is specified beyond it being an "acute animal study."
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The provided text does not indicate the use of experts to establish a "ground truth" in the traditional sense of human readers/interpreters adjudicating medical images or clinical outcomes.
- The "ground truth" for this device appears to be established through objective performance metrics from standardized bench testing protocols, validated simulation models, and comparison to a known standard (e.g., USP KPA standards) or predicate device performance.
- For the animal study, the "ground truth" for clinical endpoints (ease of use, wound approximation, hemostasis, burst strength, tensile strength) would have been established by the study investigators/veterinary surgeons performing the procedures and analyses, but their specific number or qualifications beyond being part of the "LSI" team are not detailed.
-
Adjudication method for the test set:
- Not applicable in the context of human reader/AI adjudication. The testing involved objective measurements and comparisons against predetermined criteria and a predicate device's performance.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done. This is a medical device, not an AI software intended for image interpretation assistance to human readers.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- This is a standalone medical device (a knot-tying device), not an algorithm. Therefore, "standalone performance" refers to the device's inherent mechanical and functional performance, which was extensively tested without human "interpretation" in the loop. The device's performance is tested directly (e.g., crimping strength, cutting ability, durability).
-
The type of ground truth used:
- Objective Performance Metrics / Engineering Specifications: For functional verification, validation, and durability testing (e.g., KPA force, proper crimping, suture cutting, structural integrity under pressure). These are derived from established engineering standards and the device's design inputs.
- Comparison to USP Standards: For Knot Pull Apart (KPA) force.
- Comparison to Predicate Device Performance: The underlying performance of the predicate device serves as a benchmark for equivalence.
- Animal Model Outcomes: For in-vivo safety and effectiveness, including measures like ease of use, wound approximation, hemostasis, burst strength, and tensile strength, compared to hand-tied knots. These are physiological outcomes observed and measured in the animal subjects.
-
The sample size for the training set:
- Not applicable. This is a physical, mechanical medical device, not an AI/ML algorithm that requires a training set.
-
How the ground truth for the training set was established:
- Not applicable, as no training set was used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 24, 2021
LSI Solutions, Inc. Christopher Miller Executive Director of Regulatory Affairs and Quality 7796 Victor-Mendon Road Victor, New York 14564
Re: K202551
Trade/Device Name: Cor-Knot Micro Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: September 2, 2020 Received: September 3, 2020
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K202551
Device Name COR-KNOT MICRO™
Indications for Use (Describe)
The COR-KNOT MICRO™ device with COR-KNOT MICRO™ fastener when used in conjunction with USP 6-0, 7-0, or 8-0 polypropylene surgical suture, is indicated for use in the approximation of soft tissue.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 5: 510(k) Summary
| Submitted By: | LSI SOLUTIONS, Inc. |
|---|---|
| 7796 Victor-Mendon Road | |
| Victor, NY 14564 | |
| Contact Person: | Christopher B. Miller |
| Executive Director of Regulatory Affairs and Quality | |
| Phone: (585) 869-6665 | |
| Fax: (585) 742-8086 | |
| Email: cmiller@lsisolutions.com | |
| Date Prepared: | September 2, 2020 (updated June 23, 2021) |
| Trade Name: | COR-KNOT MICRO™ |
| Common Name: | Knot tying device |
| Classification Name: | Accessory to an endoscope (per 21 CFR 876.1500); and |
| Manual Surgical Instrument for General Use (per 21 CFR 878.4800) | |
| Classification Regulation: | 21 CFR 876.1500 (Per the higher classification) |
{4}------------------------------------------------
K202551 Page 2 of 9
Product Code: GCJ (which is reviewed by the General and Plastic Surgery Panel)
Device Classification: Class II
Predicate Device: TK® Ti-KNOT® Device
Device Description
The COR-KNOT MICRO™ Device is a sterile, single patient use device provided to the user preloaded with a single COR-KNOT MICRO™ fastener. Made from medical grade titanium, the COR-KNOT MICRO™ fastener is a hollow sleeve with a rounded base. A white plastic target holds the loop shape of a wire snare. The wire snare passes through the COR-KNOT MICRO™ fastener and is attached to a snare puller knob. A suture slot in the device shaft lies under the opening in the snare puller. The ends of a USP 6-0, 7-0, or 8-0 polypropylene suture are passed through the wire snare and subsequently threaded into the titanium fastener. The snare puller attached to its wire snare is pulled up along the device shaft until it snaps onto the puller retainer feature of the purple knob, which also has an integrated indicator fin. The suture slot and the indicator fin are located on the same side of the device shaft. The subsequently crimped fastener and remnant trimmed suture tails bend slightly in the direction away from or opposite from the suture slot and indicator fin. By rotating the purple knob and the device's white handle, the surgeon can ergonomically orient the direction of the suture tails, if desired. A yellow lever stop is located behind the purple lever to restrict inadvertent squeezing of the lever during device handling before crimping. The lever stop is removed by pinching its sides together and pulling it out of the handle. By squeezing the purple lever, the COR-KNOT MICRO™ Device crimps the
LSI SOLUTIONS, Inc.
Page 2 of 9
{5}------------------------------------------------
COR-KNOT MICRO™ fastener to fasten together segments of suture and trims away excess suture.
One COR-KNOT MICRO™ Device is provided per package, and the COR-KNOT MICRO™ Device is not intended to be reloaded or reused.
The COR-KNOT MICRO™ device is intended for use in the approximation of soft tissue.
Indications for Use
The COR-KNOT MICRO™ device with COR-KNOT MICRO™ fastener when used in conjunction with USP 6-0, 7-0, or 8-0 polypropylene surgical suture, is indicated for use in the approximation of soft tissue.
Technological characteristics (comparison to Predicate Device)
Both the predicate Ti-KNOT® device and the subject COR-KNOT MICRO™ device have the same intended use: "Intended for use in the approximation of soft tissue." The predicate Ti-KNOT® device is "indicated for use in the approximation of soft tissue", and LSI recommends its use with 2-0 polypropylene suture, among other suture types. The subject COR-KNOT MICRO™ device has very similar indications. The COR-KNOT MICRO™ device, when used with USP 6-0, 7-0, and 8-0 polypropylene suture is also "indicated for use in the approximation of soft tissue." While the predicate and subject devices are each optimized for securing different types of suture, they operate on identical technical principles. Both devices are single use, nonpowered, hand operated devices which incorporate a hammer-anvil mechanism that crimps a
LSI SOLUTIONS, Inc.
Page 3 of 9
Summary K202551 23 JUN 2021
{6}------------------------------------------------
K202551 Page 4 of 9
titanium fastener held in the device tip onto suture previously snared through the fastener. Both devices also trim away excess suture tails.
Knot pull apart (KPA) force is a key performance requirement for both subject and predicate devices. Both the predicate Ti-KNOT® device and the subject COR-KNOT MICRO™ device exceed the USP KPA specifications for their compatible suture. Both the predicate and subject devices are provided sterile and are sterilized with Ethylene Oxide such that a minimum lethality of 10 ° is achieved. The predicate Ti-KNOT® device is a stainless steel and polymer delivery device with a titanium fastener. The COR-KNOT MICRO™ device is made from identical materials for the delivery device and the fastener. Both devices conform to the requirements of ISO 10993. The predicate device is packaged in a rigid thermoformed blister tray with a Tyvek cover. The subject device is comparably packaged in a PETG tray and snap retainer lid that is sealed in a Tyvek/nylon pouch.
The COR-KNOT MICRO™ device and the predicate Ti-KNOT® device are substantially equivalent. The differences between the subject and predicate devices are largely related to size, their function is comparable, and the differences do not introduce any new risks and have no negative impact on the safety and efficacy of the device.
Performance Testing Summary
Bench top performance testing was conducted to verify that the COR-KNOT MICRO™ device will perform as intended and to ensure the device will perform equivalently to the predicate
Page 4 of 9
{7}------------------------------------------------
device. The following non-clinical tests were conducted, and all results met the performance and risk-based acceptance criteria:
● Functional Verification and Validation Testing to ensure the following Design Input Requirements were met:
- A Fastener must secure 6-0. 7-0 and 8-0 polypropylene suture with knot pull apart (KPA) forces exceeding USP standards.
- Fastener shall not be damaged prior to deployment. A
-
Device shall retain fastener prior to crimping/cutting without the fastener dislodging from the tip of the device.
- Releasing the crimped fastener from the device, shall not damage the fastener. >
-
The device shall cut the suture after a completed crimp.
-
Crimped fastener shall not cause more damage to anatomy than hand tied suture/knots.
-
The device shall be compatible with only COR-KNOT MICRO™ fasteners.
-
System compatible with 6-0, 7-0, and 8-0 sized polypropylene sutures.
-
The device will prevent reloading in the field.
-
Inserted shaft length shall be comparable to other related surgical devices demonstrated to reach superficial surgical access sites and most open and minimally invasive surgical site locations.
-
Device shall allow for surgical site visibility.
- Inserted portion of device and entirety of fastener shall be shaped to reduce risk of > tissue damage or other iatrogenic injury.
- The device and fastener must function with exposure to typical surgical conditions. A
{8}------------------------------------------------
- A Device must be capable of firing the single fastener provided with the device. If multiple fasteners are required within a case, multiple devices must be used.
- A The device shall allow suture to be easily loaded through the titanium fastener and distal end of the device.
- A The user must be able to remove Lever Stop prior to use of the device.
- A The user must be able to crimp a fastener with one hand.
-
The device shall not cause harm to the user.
- Biocompatibility per ISO 10993-1:2018. ●
- Packaging/Shelf-life Testing per ISO 11607-1 demonstrating a 3 year shelf life.
Additional bench testing involving a custom malignant hypertension pressure simulator successfully demonstrated the excellent strength and stability of COR-KNOT MICRO™ titanium fasteners on large diameter pulsatile tubular structures, mimicking extreme, supraphysiologic conditions during a simulated protracted healing period. This worst-case durability experiment incorporated supraphysiologic pressure waves mimicking systolic and diastolic pressure conditions more than 3x greater (>360/240 mmHg arterial blood pressure) than normal blood pressure in a human patient (120/80 mmHg). The pulsatile rate was also 25% faster, at 90 pulses per minute, than a normal rate, which is typical cited at 72 beats per minute. The experiment was conducted over a 6 week period to exceed the normal healing duration by a factor of 2–3x. All 60 COR-KNOT MICRO™ titanium fasteners used to secure 6-0, 7-0, and 8-0 polypropylene suture held without compromise throughout this protracted simulated healing period, despite being exposed to conditions beyond the physiologic worst case. For the duration of the six week test, all of the COR-KNOT MICRO™ titanium fasteners successfully secured simple interrupted
LSI SOLUTIONS, Inc.
Page 6 of 9
{9}------------------------------------------------
K202551 Page 7 of 9
6-0, 7-0, and 8-0 sutures to close 8 Fr dilated wounds (consistent with the predicate device in vivo study) in simulation chambers utilizing pulsatile segments of latex tubing, which has an unpressurized diameter of 2.22 cm (7/8 in) and a wall thickness of 1.59 mm (1/16 in), to simulate aortic tissue within the human circulatory system.
The worst-case physiologic scenario in this test included several elements. The worst-case wound repair method was a simple interrupted suture, which used the lowest minimum number of strands through tissue to hold wound edges together. Thus, the force exerted on this single loop of suture, and the titanium fastener securing it, is higher than when multiple suture strands are incorporated at the same wound closure site. The worst-case suture size regarding titanium fastener holding force was the smallest diameter suture tested. The size of tissue structures routinely sutured with 6-0, 7-0, and 8-0 polypropylene is generally smaller than the human femoral artery, which is typically less than 10 mm in diameter. The worst-case tissue type was a large artery in a hypertensive patient who is tachycardic and slow to heal. The diameter of the tissue simulated in this experiment was ~30 mm. This simulation experiment also provided a sustained physiologic worst-case blood pressure of >360/240 mmHg over 6 weeks, which would be nonsurvivable in a patient. A worst-case healing time of 6 weeks was selected, which is several times greater than the 2–3 week period during which wounds would be expected to heal. Additionally, the pulsatile pattern simulating heart rate in this experiment (90 pulses per minute, ppm) was 25% more ppm than the typical human heart rate of 72 beats per minute (bpm). At a pulse rate of 90 bpm and an experimental duration of 6 weeks, 1 day, 3 hours, and 33 minutes (62,133 minutes), each wound closure experienced 5,591,970 distinct hypertensive pulses. The 60 COR-KNOT MICRO™ titanium fasteners--20 each for 6-0, 7-0, and 8-0 polypropylene
LSI SOLUTIONS, Inc.
{10}------------------------------------------------
K202551 Page 8 of 9
suture-experienced a total cumulative 335,518,200 distinct hypertensive pulses without any compromise.
This study confirmed that the COR-KNOT MICRO™ Device and its titanium fastener have the ability to secure and maintain closure of 6-0, 7-0, and 8-0 polypropylene suture under physiologic worst-case conditions over protracted healing time while being exposed to continuous pulsatile forces—384.6 mmHg average systolic pressure and 259.1 mmHg average diastolic pressure—beyond the stresses anticipated on the product under conditions of the clinically relevant worst-case scenario. The COR-KNOT MICRO™ Device and its titanium fastener successfully met all predetermined requirements and acceptance criteria defined for retention testing of the titanium fastener.
LSI also conducted an acute animal study to evaluate the safety and effectiveness of the proposed COR-KNOT MICRO™ Device and preloaded micro titanium fastener relative to a comparative study conducted in 1995 for the predicate device. The test report for the animal study compares the COR-KNOT MICRO™ Device in the worst-case size blood vessels to hand tied knots under the same conditions. The COR-KNOT MICRO™ titanium fastener provided superior or equivalent—and never inferior—results for ease of use, wound approximation, hemostasis, burst strength, and tensile strength as compared to the controls. Results and observations from the animal procedure tests demonstrate that the COR-KNOT MICRO™ Device with its titanium fastener is easy to use and provides excellent wound approximation resulting in hemostasis in conjunction with 6-0, 7-0, and 8-0 polypropylene suture. Overall, this study confirmed that the COR-KNOT MICRO™ Device and its titanium fastener have the ability
LSI SOLUTIONS, Inc.
Page 8 of 9
{11}------------------------------------------------
K202551 Page 9 of 9
to hemostatically secure and maintain closure of wounds using 6-0, 7-0, and 8-0 polypropylene suture in an acute in vivo pig model; harvested specimens were tested and demonstrated burst strengths beyond supraphysiologic worst case anticipated conditions.
Clinical testing
Not applicable. Neither the predicate product nor the subject product require clinical testing since the performance of this product is to hold the suture that holds tissue (but not directly the tissue itself). The reliability and excellent performance of this product has been sufficiently demonstrated in multiple experiments that do not require clinical testing in patients.
Substantial Equivalence
The subject device and predicate device have the same intended use and technological characteristics. Non-clinical performance data has demonstrated the subject device is substantially equivalent to the predicate device.
In Conclusion
Extensive testing demonstrates the proposed COR-KNOT MICRO™ product consistently provides excellent suture holding security for 6-0. 7-0. and 8-0 polypropylene suture and can be dependably manufactured. The subject COR-KNOT MICRO™ device and titanium fastener are substantially equivalent to the predicate Ti-KNOT® device, K981531.
LSI SOLUTIONS, Inc.
Page 9 of 9
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.