K Number
K202483
Device Name
MyHomeDoc
Manufacturer
Date Cleared
2021-03-23

(207 days)

Product Code
Regulation Number
870.1875
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MyHomeDoc device, combined with a smartphone application, is a multiple-sensor device that is intended to measure, record, and transmit the recorded data of auscultation sound of human body, human body temperature, oxygen saturation (SpO2) and pulse rate, and images of an examined body part.

MyHomeDoc device is intended for use by adult lay users in a non-clinical environment. MyHomeDoc application enables counseling with the physician and transmission of the information over an IP network.

Device Description

The MyHomeDoc device is a home use device is comprised of a handheld unit that connects with a smartphone and a dedicated smartphone software application that runs on the user's personal smartphone An API is defined to enable healthcare providers to communicate with the device.

The Device enables user's examination at home with or without the guidance of a remote physician. It also enables guidance and data examination by a remote physician over an IP network. The Smartphone Application controls the Handheld Unit's functionality and processes and displays the collected data.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the MyHomeDoc device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document provides acceptance criteria implicitly through the successful completion of tests and explicit statements of meeting criteria in specific cases.

Test NameAcceptance Criteria (Implicit/Explicit from description)Reported Device Performance
Thermometer Performance & AccuracyExplicit: Difference between IR readings and Black body less than 0.3°C. Compliance with ASTM E1965-98.The MyHomeDoc device met all acceptance criteria. The results demonstrate that the difference between the IR readings ("Direct mode") and the Black body is less than 0.3°C. The test tables demonstrate compliance of the MyHomeDoc Thermometer component with ASTM E1965-98.
MyHomeDoc Device vs. Braun Predicate for Oral EquivalenceExplicit: Difference between MyHomeDoc and Brown NTF3000 predicate device less than or equal to 0.4°C.The difference between MyHomeDoc and the Brown NTF3000 predicate device was less than or equal to 0.4°C in 81 users. Therefore, the MyHomeDoc device met the acceptance criteria.
SpO2 VerificationImplicit: Met all acceptance criteria, including displayed SpO2 value for each simulator value, and MyHomeDoc measurements of monotonic increasing/decreasing values are monotonic and align within the 20% range.The MyHomeDoc device met all acceptance criteria, including meeting specified criteria for displayed SpO2 values and monotonic behavior of measurements within a 20% range.
Pulse Rate VerificationImplicit: Met acceptance criteria, with measured pulse results between 30BPM to 250BPM.MyHomeDoc device measured pulse results between 30BPM to 250BPM. MyHomeDoc device met the acceptance criteria.
Pulse Oximeter Hypoxia ProtocolImplicit: Met all acceptance criteria for validating SpO2 measurements.The MyHomeDoc device met all acceptance criteria.
Stethoscope Test ReportImplicit: Met the acceptance criteria. (Specific criteria are not detailed in the provided text).MyHomeDoc devices used in the test met the acceptance criteria.
Otoscope Test ReportImplicit: Met the acceptance criteria for line-pair separation, color differentiation, and illumination flux.Line-pair separation test, Color differentiation and Illumination flux were evaluated. MyHomeDoc devices used in the test met the acceptance criteria.
Otoscope-Gold Standard vs. MyHomeDoc Ear Test ComparisonImplicit: MyHomeDoc images were clear, allowing clinical evaluation; landmarks of the ear were easily identified; physician subjective assessment was positive; images equivalent to those of the gold standard.MyHomeDoc images were cleared allowing clinical evaluation, Landmarks of the ear were easily identified and the physician subjective assessment was positive. Therefore, the MyHomeDoc device met the acceptance criteria and MyHomeDoc images are equivalent to those of the gold standard.
Biocompatibility EvaluationImplicit: Compliance with ISO 10993-1.In compliance with ISO 10993-1.
Cleaning, PackagingImplicit: Compliance with relevant standards for cleaning validation, transportation, and use life calculation.All tests were successfully completed.
Software ValidationImplicit: As recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."Software verification and validation testing were conducted and documentation was provided as recommended. The software level of concern is moderate.
Electrical Safety and EMCImplicit: Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 62471 First edition, ISO 80601-2-61 and IEC 60529, wireless coexistence testing.Conducted on the MyHomeDoc device. In addition, the system complies with the listed standards.
Usability TestingImplicit: Demonstrate that MyHomeDoc device can be used by the intended users for the intended uses and under the expected use conditions without serious use errors or problems.A usability study was conducted to evaluate the use-related safety and effectiveness of the MyHomeDoc when used by lay-user via observational data, knowledge task data, and interview data, and demonstrated the device can be used without serious use errors or problems.

2. Sample Sizes and Data Provenance

  • Thermometer Oral Equivalence Test: 81 users.
  • Hypoxia Protocol (Pulse Oximeter): The sample size is not explicitly stated in the provided text, but it was a "hypoxia test" performed for validation.
  • Other Performance Tests (Stethoscope, Otoscope, SpO2 verification, Pulse Rate verification): The specific sample sizes for these tests are not provided in the document.
  • Usability Study: The sample size is not explicitly stated, but it involved "lay-users" via observational data, knowledge task data, and interview data.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical or performance data. It is a submission by "MyHomeDoc Ltd." which has an address in Ra'anana, Israel. The study type is not explicitly labeled as retrospective or prospective, but the description of "testing performed" and "study conducted" implies prospective data collection for the performance and usability evaluations.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

  • Thermometer Oral Equivalence Test: Not applicable, as ground truth was established by comparison to a predicate device (Brown NTF3000 thermometer) and a quantitative difference threshold.
  • SpO2 Verification / Hypoxia Protocol: Not applicable, as these tests are based on device measurements against a simulator or established physiological conditions rather than expert interpretation.
  • Stethoscope Test Report: Not explicitly stated how ground truth was established or if experts were involved in a comparative clinical sense. It was evaluated against a predicate stethoscope (Littman).
  • Otoscope-Gold Standard vs. MyHomeDoc Ear Test Comparison Report:
    • Number of Experts: Not explicitly stated. The text mentions "the physician subjective assessment was positive," implying at least one physician.
    • Qualifications of Experts: Mentioned as "physician," but no specific details on experience or specialty (e.g., ENT, general practitioner) are provided.
    • Type of Ground Truth: "Gold standard" (implicitly, in comparison to other otoscope images) and "physician subjective assessment."

4. Adjudication Method

The document does not describe a formal adjudication method (e.g., 2+1, 3+1) for establishing ground truth from multiple experts for any of the performance studies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study focusing on human readers with vs. without AI assistance is described. The usability study evaluates lay-user interaction, but not the comparative diagnostic effectiveness of humans aided by the MyHomeDoc device.

6. Standalone Performance Study (Algorithm Only)

The performance data presented primarily focuses on the standalone performance of the device's sensors (thermometer, pulse oximeter, stethoscope, otoscope) against established standards, predicate devices, or physical measurements. For example, the thermometer's accuracy against a black body, SpO2 against a simulator or hypoxia protocol, and otoscope image clarity. This aligns with the "algorithm only" aspect for each specific measurement component. The overall device is a multi-sensor device, where each sensor's performance is independently evaluated.

7. Type of Ground Truth Used

  • Thermometer (Performance & Accuracy): Black body (physical standard) and comparison to a legally marketed predicate device (Brown NTF3000).
  • Pulse Oximeter (SpO2, Pulse Rate, Hypoxia): Simulator values, established physiological conditions (hypoxia protocol), and performance within specified ranges (30-250 BPM).
  • Stethoscope: Comparison to a legally marketed predicate stethoscope (Littman).
  • Otoscope (Images): "Gold standard" (implicitly, other high-quality otoscope images) and "physician subjective assessment" (expert opinion on clarity and identification of landmarks).
  • Biocompatibility, Cleaning, Packaging, Electrical Safety, EMC: Compliance with international standards (ISO, IEC).
  • Software Validation: Compliance with FDA guidance.
  • Usability Testing: Observational data, knowledge task data, and interview data from lay users against the criterion of "no serious use errors or problems."

8. Sample Size for the Training Set

The document does not provide information about a training set or its sample size. This is a 510(k) submission for a device that primarily measures and transmits data, not a device with a complex machine learning algorithm that requires a separate training set for development. The "software" validation mentioned refers to the general software engineering practices rather than AI/ML model training.

9. How Ground Truth for the Training Set Was Established

As no training set is described for a complex AI/ML algorithm, this question is not applicable based on the provided text. The device's functionality leans more towards traditional instrumentation and data acquisition rather than advanced AI-driven diagnostics requiring extensive training data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

March 23, 2021

MyHomeDoc Ltd. % Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, District of Columbia 20004

Re: K202483

Trade/Device Name: MyHomeDoc Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD, FLL, DQA, ERA Dated: March 23, 2021 Received: March 22, 2021

Dear Jonathan Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for LT Stephen Browning Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number ( if known )K202483
Device Name

MyHomeDoc device

Indications for Use (Describe)

The MyHomeDoc device, combined with a smartphone application, is a multiple-sensor device that is intended to measure, record, and transmit the recorded data of auscultation sound of human body, human body temperature, oxygen saturation (SpO2) and pulse rate, and images of an examined body part.

MyHomeDoc device is intended for use by adult lay users in a non-clinical environment. MyHomeDoc application enables counseling with the physician and transmission of the information over an IP network.

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human

Services Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA)

Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20) Page 1 of 1 FDA PSC Publishing Services (301) 443-6740 EF

Page 1

{3}------------------------------------------------

Traditional Premarket Notification Submission - 510(k) MyHomeDoc device 510(k) Number K202483

Date Prepared: August 28, 2020

I. SUBMITTER

MyHomeDoc Ltd. 13 Zarchin St. Ra'anana, 4366241 Tel: +972-(0)4- 9874202

Contact Person

Jonathan Kahan, Partner Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington, DC 20004 Tel:+1 202 637 5794 jonathan.kahan@hoganlovells.com

Back-up Company Contact:

Orly Maor 25 Sirkin Street Kfar Saba 44421 Israel Tel: +972-9-7453607 oram.ma@gmail.com

II. DEVICE

Name of Device: MyHomeDoc device Common or Usual Name: MyHomeDoc device Classification Name: Oximeter, Clinical Electronic Thermometer, Oximeter and Otoscope Regulatory Class: II Product Code: DQD, FLL, DQA, ERA

PREDICATE DEVICE III.

MyHomeDoc Ltd. believes that the MyHomeDoc device is substantially equivalent to the following predicate devices:

  • Pulse Oximeter- PulseOx 5500 Finger Device, manufactured by SPO Medical Ltd cleared under K040178, Classification name: Oximeter, Product code: DQA, Regulation: 21 CFR 870.2700

{4}------------------------------------------------

  • Stethoscope- Tyto Stethoscope manufactured by Tyto care Ltd. cleared under ● K160401, Classification name Electronic stethoscope, Product code: DQD, Regulation: 21 CFR 870.1875
  • . Thermometer- Braun No Touch + Forehead NTF3000 Thermometer, manufactured by Kaz USA, Inc, cleared under K163516, Classification: Clinical Electronic Thermometer, Product code: FLL, Regulation: 21 CFR 880.2910

In addition, the following reference device is used in this 510(k):

  • Electronic Stethoscope Model 3200, manufactured by 3M Littman, cleared under ● K101834, Classification name Electronic stethoscope, Product code: DRG, Regulation: 21 CFR 870.2910

IV. DEVICE DESCRIPTION

The MyHomeDoc device is a home use device is comprised of a handheld unit that connects with a smartphone and a dedicated smartphone software application that runs on the user's personal smartphone An API is defined to enable healthcare providers to communicate with the device.

The Device enables user's examination at home with or without the guidance of a remote physician. It also enables guidance and data examination by a remote physician over an IP network. The Smartphone Application controls the Handheld Unit's functionality and processes and displays the collected data.

V. INDICATIONS FOR USE

The MyHomeDoc device, combined with a smartphone application, is a multiple-sensor device that is intended to measure, record, and transmit the recorded data of auscultation sound of human body, human body temperature, oxygen saturation (SpO2) and pulse rate, and images of an examined body part.

MyHomeDoc device is intended for use by adult lay users in a non-clinical environment. MyHomeDoc application enables counseling with the physician and transmission of the information over an IP network.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE AND REFERENCE DEVICE

The MyHomeDoc device is substantially equivalent to its predicate devices.

MyHomeDoc device intended use combines the indications of the cleared predicate devices.

{5}------------------------------------------------

The MyHomeDoc device, like the Tyto Stethoscope, is intended for adults to record sounds from adult and pediatric patients' heart and lungs and abdomen in a home environment. Like the PulseOx 5500 predicate, the MyHomeDoc device is intended for saturation and pulse rate measurements. Similar to the Braun Forehead NTFF3000 thermometer, MyHomeDoc device is intended for body temperature determination.

The MyHomeDoc device has similar technological characteristics as the predicate Tyto Stethoscope in that it records and wirelessly transmits the patient's data to the care provider at a different location. Both devices contain software guidance that instructs the user where to place the sensor to measure the auscultation signal. An additional similarity is that the Tyto includes a class I otoscope as well.

The temperature sensor in the MyHomeDoc device is equivalent to its predicate in that both are contact/non-contact IR sensors that measure temperature from the center of the forehead. Both devices then convert this temperature to the oral equivalent and display this temperature in Fahrenheit or Celsius. While the thermometers have the same accuracy, there are slight differences in regards to the measurement range, but as the subject device has a greater range, this does not raise any questions. The company has performed testing demonstrating the performance of the thermometer over this temperature range.

The pulse oximeter in the MyHomeDoc device and the predicate are equivalent in that they both measure change in light intensity at 660 and 880 nm. Both devices perform spot measurement of oxygen saturation and pulse rate at the patient's fingertip. The measurement methods slightly differ in that the predicate measures the transmitted signal. while the subject device assesses the reflected signal.

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination:

  • Biocompatibility evaluation in compliance with ISO 10993-1. -

Cleaning, Packaging -

Cleaning validation testing, transportation and use life calculation of the MyHomeDoc device was performed to demonstrate compliance with the relevant standards. All tests were successfully completed.

Performance Testing -

Performance testing included the following:

{6}------------------------------------------------

Name of testTest description
Thermometer
Thermometer performance & accuracytestingTest in compliance to ASTM E1965-98.TheMyHomeDoc device met all acceptance criteria.The results of this test demonstrate that thedifference between the IR readings (the "Directmode") and the Black body is less than 0.3°C
ASTM E1965-98 ThermometerThe test tables demonstrate compliance of theMyHomeDoc Thermometer component withASTM E1965-98, Standard Specification forInfrared Thermometers for IntermittentDetermination of Patient Temperature
MyHomeDoc Device vs. BraunPredicate Device for Oral EquivalenceTest ReportThe difference between MyHomeDoc and theBrown NTF3000 predicate device was less thanor equal to 0.4°C in a 81users. Therefore, theMyHomeDoc device met the acceptance criteria.
Pulse oximeter
SpO2 verification protocol and reportThe MyHomeDoc device met all acceptancecriteria.- displayed SpO2 value for each simulator value.- MyHomeDoc measurements of monotonicincreasing/decreasing values are monotonic andalign within the 20% range.
Pulse rate verificationMyHomeDoc device measured pulse resultsbetween 30BPM to 250BPM.MyHomeDoc device met the acceptance criteria.
Pulse Oximeter Hypoxia Protocol andReportThe pulse oximeter hypoxia test performed inorder to validate the MyHomeDoc device SpO2measurements. The MyHomeDoc device met allacceptance criteria.
Stethoscope
Stethoscope Test ReportThe MyHomeDoc device and the Littmanpredicate stethoscope where evaluated.MyHomeDoc devices used in the test met theacceptance criteria.
Otoscope
Otoscope Test ReportLine-pair separation test, Color differentiation andIllumination flux were evaluated. MyHomeDocdevices used in the test met the acceptance criteria
Otoscope-gold standard vs. MyHomeDocear test comparison reportMyHomeDoc images were cleared allowingclinical evaluation, Landmarks of the ear wereeasily identified and the physician subjectiveassessment was positive. Therefore, theMyHomeDoc device met the acceptance criteriaand MyHomeDoc images are equivalent to those
Name of testTest description
of the gold standard.

{7}------------------------------------------------

All tests met the predefined acceptance criteria.

Software Validation -

The MyHomeDoc software level of concern is moderate. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

Electrical Safely and EMC -

Electrical Safety per IEC 60601-1 and Electromagnetic compatibility (EMC) per IEC 60601-1-2 and IEC 60601-1-6 were conducted on the MyHomeDoc device. In addition, the system complies with IEC 60601-1-11, IEC 62471 First edition, ISO 80601-2-61 and IEC 60529, wireless coexistence testing.

Usability testing -

Usability study was conducted to evaluate the use-related safety and effectiveness of the MyHomeDoc when used by lay-user via observational data, knowledge task data, and interview data. The purpose of the current study is to demonstrate that MyHomeDoc device can be used by the intended users for the intended uses and under the expected use conditions without serious use errors or problems.

VII. CONCLUSIONS

The MyHomeDoc device was determined to be substantially equivalent to the predicate devices.

§ 870.1875 Stethoscope.

(a)
Manual stethoscope —(1)Identification. A manual stethoscope is a mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.(b)
Electronic stethoscope —(1)Identification. An electronic stethoscope is an electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.