K Number
K202440
Date Cleared
2021-06-17

(296 days)

Product Code
Regulation Number
880.2910
Panel
HO
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The infrared thermometer is intended for the intermittent measurement of body temperature from the auditory canal or central forehead skin surface on people of all ages. It can be used by consumers in the household environment and by healthcare providers.

Device Description

HT series infrared thermometers (includes model HT-101, HT-102, HT-103 and HT-104) measure the body temperature through receiving infrared energy radiation via the ear or forehead. These thermometers have the capability to measure temperature via forehead temperature mode or ear temperature mode, and the temperature is directly shown on the LCD display. These thermometers have the following features:

  • The device is intended to be used by consumers in the household environments and by healthcare providers;
  • The device is mainly composed of infrared sensor, signal receiving processor, buttons, buzzer, LCD display. It is powered by 2*1.5 AAA batteries;
  • It focuses the infrared energy received from the human's forehead or ear by using the Fresnel lens of the thermometer;
  • Switching between Celsius and Fahrenheit;
  • Multi-functional, can measure ear and forehead temperature;
  • Automatic range selection; resolution is 0.1 Celsius (0.1 Fahrenheit);
  • The latest 35 measurement data can be memorized and stored; the user can view the previous measurement results;
  • High temperature alarm function, displayed in green, orange and red backlight;
  • Switching between mute and un-mute mode;
  • Low battery indication, and auto shut-down.
AI/ML Overview

The document describes the acceptance criteria and the study conducted for the Infrared Thermometer (Models HT-101, HT-102, HT-103, HT-104).

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Infrared Thermometer are primarily based on established medical device standards and comparison to a predicate device (K191570).

Feature / CriterionAcceptance Criteria (from predicate or standards)Reported Device Performance (Subject Devices)Comments
Product CodeFLLFLLSame
Regulation Number880.2910880.2910Same
Indications for UseIntermittent measurement of body temperature from auditory canal or central forehead skin surface on people of all ages; for consumers and healthcare providers.Intermittent measurement of body temperature from auditory canal or central forehead skin surface on people of all ages; for consumers and healthcare providers.Same
Thermometer typeDigital thermometerDigital thermometerSame
SensorThermopileThermopileSame
Operational PrincipleInfrared radiation detectionInfrared radiation detectionSame
Display typeLCDLCDSame
Measurement siteForehead and earForehead and earSame
Measurement RangeEar & Forehead: 32.0°C43.0°C (89.6°F109.4°F)Ear & Forehead: 32.0°C43.0°C (89.6°F109.4°F)Same
Measurement AccuracyPredicate: ±0.4°F (0.2°C) within 95107.6°F (35.042.0°C), ±0.5°F (0.3°C) for other range.
Standard: ISO80601-2-56 and ASTM E1965-98 requirements.34.0°C43.0°C (89.6°F109.4°F): ±0.2°C(0.4°F)
32.0 °C34.0°C(89.6°F93.2°F): ±0.3°C(0.5°F)Comparable (meets ISO80601-2-56 and ASTM E1965-98)
Display Resolution0.1°C (0.1°F)0.1°C (0.1°F)Same
Power Supplyd.c. 3.0V (2pcs AAA batteries)d.c. 3.0V (2pcs AAA batteries)Same
Measurement Time1 second1 secondSame
Measurement data memories20 sets35 setsComparable (software verified, no safety/effectiveness issues)
Beeper settingYesYesSame
Date and time settingNoNoSame
BacklightYesYesSame
Auto-off time30s15s±1sSimilar (meets IEC 60601-1:2005+AMD 1:2012, no safety/effectiveness issues)
Operation Condition (Temperature)10°C~40°C10°C~40°CSame
Operation Condition (Relative Humidity)≤85%≤95%Comparable (tested per IEC 60601-1-2 EMC, IEC 60601-1, ISO 80601-2-56 and ASTM E1965-98)
Operation Condition (Atmospheric Pressure)70kPa~106kPa70kPa~106kPaSame
Storage & Transportation Temp.-20°C~55°C-20°C~50°CComparable (no safety/effectiveness issues)
Storage & Transportation Humidity15%~93%≤95%Comparable (tested per IEC 60601-1-2 EMC, IEC 60601-1, ISO 80601-2-56 and ASTM E1965-98)
Storage & Transportation Pressure70kPa~106kPa70kPa~106kPaSame
Protection against electric shockExternally power supply, Type BFExternally power supply, Type BFSame
IP ClassIP22IP22Same
Mode of operationContinuous OperationContinuous OperationSame
Materials of skin-contacting componentsABSABS/ PA757(+); PC/PC-110Comparable (tested per IEC 60601-1-2 EMC, IEC 60601-1, ISO 80601-2-56, ASTM E1965-98 and ISO 10993-1)
Physical DimensionsNot available (for predicate)HT-101: 1653850 mm; HT-102: 1563744mm; HT-103: 1555238 mm; HT-104: 1543950mmComparable (no new safety/effectiveness issues)
WeightNot available (for predicate)HT-101: 62.5g (Battery excluded); HT-102: 58g (Battery excluded); HT-103: 64.5g (Battery excluded); HT-104: 63g (Battery excluded)Comparable (no new safety/effectiveness issues)
Electromagnetic CompatibilityMeets IEC 60601-1-2 requirementsPassed testing according to IEC 60601-1-2:2014, IEC 60601-1:2005+AMD 1:2012, IEC 60601-1-11:2015Compliance demonstrated
Electrical SafetyMeets IEC 60601-1 requirementsPassed testing according to IEC 60601-1:2005+AMD 1:2012Compliance demonstrated
BiocompatibilityMeets ISO 10993 requirementsPassed testing according to ISO 10993-5:2009 (cytotoxicity) and ISO 10993-10:2010 (irritation and skin sensitization)Compliance demonstrated
Performance (Bench)Meets ISO 80601-2-56 and ASTM E1965-98 requirementsPassed testing according to ISO 80601-2-56:2017+AMD2018 and ASTM E1965-98 (Reapproved 2016)Compliance demonstrated
Software Verification & ValidationModerate level of concern, all software requirements met, hazards mitigated to acceptable risk levels.Software documentation consistent with moderate level of concern, system validation testing demonstrates all software requirement specifications met and all software hazards mitigated to acceptable risk levels.Compliance demonstrated
Clinical AccuracyMeet requirements for clinical bias and clinical repeatability per ASTM E1965-98 (Reapproved 2016)Clinical accuracy validation test report demonstrates that the subject devices met the requirements for clinical bias and clinical repeatability per ASTM E1965-98(Reapproved 2016).Compliance demonstrated

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: A total of 150 subjects were included in the clinical accuracy validation test.
    • Infants: 50 subjects (newborn to one year)
    • Children: 50 subjects (greater than one to five years)
    • Adults: 50 subjects (greater than five years old)
  • Data Provenance: The document does not explicitly state the country of origin or whether the study was retrospective or prospective. It is a clinical accuracy validation test, which implies prospective data collection for the purpose of validating the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not provide details on the number of experts or their qualifications for establishing ground truth during the clinical accuracy validation. It only states that the clinical accuracy validation testing was conducted according to ASTM E1965-98 (Reapproved 2016), which implies standard clinical measurement procedures and potentially comparison to a reference thermometer.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method (such as 2+1 or 3+1). The testing focused on comparing the device's readings against established clinical accuracy standards (ASTM E1965-98), which inherently define acceptable differences from a reference method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an infrared thermometer, not an AI-assisted diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable in the context of an infrared thermometer. The device itself is the standalone measurement tool. Its performance is evaluated both on the bench (bench testing) and in clinical settings (clinical accuracy validation), where it is operated by a human user to take a measurement. However, there is no separate "algorithm only" performance that needs to be assessed, as the device's function is direct temperature measurement.

7. The Type of Ground Truth Used

For the clinical accuracy validation, the ground truth was implied by the reference measurements taken in accordance with ASTM E1965-98 (Reapproved 2016). This standard typically defines the reference method for temperature measurement against which the test device's accuracy (bias and repeatability) is assessed. It's not expert consensus, pathology, or outcomes data in the traditional sense, but rather a standard measurement reference.

8. The Sample Size for the Training Set

The document does not mention a training set as this device is a hardware medical device (infrared thermometer), not an AI/machine learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.