K Number
K202394
Device Name
EOSedge
Manufacturer
Date Cleared
2020-09-16

(26 days)

Product Code
Regulation Number
892.1680
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EOSedge is intended for use in general radiographic exams and applications, excluding the evaluation of lung nodules and exams involving fluoroscopy, and mammography. EOSedge allows the radiographic acquisition of either one or two orthogonal X-ray images, for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient, in the upright or seated position.

The Micro Dose feature is indicated for assessing global skeletal deformities in follow-up pediatric exams.

Device Description

The EOSedge system is a digital radiography system comprised of an acquisition workstation, a gantry including an electrical cabinet housing the system power and communication controls, and an acquisition software to obtain diagnostic images. Two identical sets of detectors and X-ray tubes are positioned orthogonally to generate frontal and lateral images simultaneously by scanning the patient over the area of interest. If desired, the Micro Dose feature enables image acquisition for assessing global skeletal deformities in follow-up pediatric exams. The two sets of detectors and X-ray tubes are identical between the predicate device and the subject device. The image acquisition can be performed with a MANUAL mode or an AUTO mode of patient examination. To select the area of interest (acquisition area), a vertical collimation is set using green lasers and to correctly position the patient in the EOSedge, a centering system is used based on red lasers. The diagnostic images are stored in a local database and are displayed on a high-resolution medical-quality non-diagnostic monitor. The diagnostic image can be transmitted through a DICOM compatible digital network for printing and archiving.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the EOSedge™ system, an updated digital radiography system. The document focuses on demonstrating that the updated device is substantially equivalent to a previously cleared predicate device (EOS imaging's EOSedge System K192079).

Based on the provided text, the acceptance criteria and study proving the device meets these criteria are primarily based on bench testing and comparative analysis with a predicate device, rather than a clinical study evaluating diagnostic performance.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly tied to demonstrating substantial equivalence to the predicate device, implying that the updated device must perform at least as safely and effectively. The performance data section refers to conformity to various IEC standards and internal functional testing.

Feature / StandardAcceptance Criteria (Implicit: Equivalence to Predicate, Conformity to Standards)Reported Device Performance
Safety (Electrical, Radiation, Usability)Conformity to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-54, IEC 60825-1"EOSedge is designed and has been certified to conform to IEC 60601-1 and collateral standards."
Image QualityNo explicit quantitative criteria mentioned, but implied to be equivalent to predicate. Bench testing conducted."Bench testing to confirm appropriate dosing and image quality." "Bench performance testing were conducted based on FDA's Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices (September 1, 2016), to verify that EOSedge performs according to specifications and is as safe and effective as the predicate device."
DosingNo explicit quantitative criteria mentioned, but implied to be appropriate and equivalent to predicate."Bench testing to confirm appropriate dosing..."
Software FunctionalityVerified and validated software performance."Software verification and validation testing was also conducted."
Technical Specifications (Detectors)As per predicate (K192079)From Table 1 and Table 3 (for both current and predicate): - Quantity: 2 - Type: Hybrid CdTe-CMOS dual energy photons counting X-ray detector - Dimensions: Width 585 mm x 98 mm, Thickness 131 mm - Weight: <10 kg - X-ray conversion material: CdTe - X-ray voltage range: < 160 kVp - Active area: 514 mm x 6 mm - Pixel size: 100 x 100 µm², elementary - Tile gap: <=100 µm - Overall fill factor: 99.63% - Number of pixels/lines: 5,139 pixels - Imaging speed: Up to 5,000 lines/s, in TDS mode - Counting output range (Pixel depth): 17 bits (> 131 000 gray levels) - Nominal input voltage: 12 VDC – 20 A - Temperature control: Internal check, Peltier + PWM, Ventilator system checked - DQE type: RQA5 spectrum ~ 80% - MTF type: ~70% @ 2lp / mm, ~25% @ 5lp / mm - Pixel Depth: 17 bits (> 131 000 grey levels) - Pixel Size: 100 µm - Resolution: 3.7 lp/mm - Typical Dynamic Range: > 100 dB
Average Acquisition TimeMust be comparable or improved compared to predicate.Current Device (EOSedge™): 7 seconds for a spine and 13 seconds for an entire body Predicate Device (EOSedge System K192079): 8 seconds for a spine and 15 seconds for an entire body (Improved performance stated for current device)

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a clinical test set with patient data for evaluating diagnostic performance. The testing described is primarily bench testing and software verification/validation. Therefore, there are no details on sample size for a test set of patient images or their provenance (country of origin, retrospective/prospective). The assessment is based on technical specifications and comparison to the predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

As there is no clinical test set for diagnostic performance described, there is no information on experts establishing ground truth for such a set. The ground truth for the engineering/technical tests would be the established specifications and accepted performance metrics for medical imaging devices, as defined by relevant IEC and FDA guidance.

4. Adjudication Method for the Test Set

Not applicable, as no clinical test set requiring expert adjudication is described in the provided text.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not explicitly done or described in the provided text. The submission focuses on demonstrating substantial equivalence through technical performance data and comparison to a predicate device, not on improved human reader performance with AI assistance. The device itself is an X-ray imaging system, not an AI diagnostic assistant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not explicitly applicable in the context of an X-ray imaging system. The performance evaluated here is the image acquisition and processing capabilities of the X-ray device itself, which operates as a standalone system to produce images for human interpretation.

7. The Type of Ground Truth Used

The ground truth used for these tests is based on established engineering and performance standards for X-ray imaging devices (e.g., IEC standards, FDA guidance for solid-state X-ray imaging devices). For comparative purposes, the predicate device's cleared performance serves as a benchmark for substantial equivalence.

8. The Sample Size for the Training Set

Not applicable. The document describes a medical imaging device (hardware and software for image acquisition), not a machine learning algorithm that requires a training set of data for inference.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for a machine learning algorithm described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

EOS imaging % Mr. Bernard Ismael Quality and Regulatory Affairs Director 10 rue Mercoeur Paris, F-75011 FRANCE

Re: K202394

Trade/Device Name: EOSedge™ Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: August 19, 2020 Received: August 21, 2020

Dear Mr. Ismael:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

September 16, 2020

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K202394

Device Name

EOSedge™

Indications for Use (Describe)

EOSedge is intended for use in general radiographic exams and applications, excluding the evaluation of lung nodules and exams involving fluoroscopy, and mammography. EOSedge allows the radiographic acquisition of either one or two orthogonal X-ray images, for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient, in the upright or seated position.

The Micro Dose feature is indicated for assessing global skeletal deformities in follow-up pediatric exams.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K202394 510(k) SUMMARY

EOS imaging's EOSedge System

EOS imaging 10 Rue Mercoeur F-75011, Paris FRANCE

Phone: + 33 1 55 25 60 60 Facsimile: + 33 1 55 25 60 61

Contact Person: Bernard ISMAEL, Quality and Regulatory Affairs Director

Date Prepared: August 19, 2020

Name of Device:EOSedge™
Common or Usual Name:Digital Radiography System
Classification Name:21 C.F.R §892.1680; Stationary X-ray System
Regulatory Class:Class II
Product Code:KPR - System, X-ray, Stationary
Predicate Device:EOS imaging's EOSedge System (K192079)
Common or Usual Name:Digital Radiography System
Classification Name:21 C.F.R §892.1680; Stationary X-ray System
Regulatory Class:Class II
Product Code:KPR - System, X-ray, Stationary

Device Description

The EOSedge system is a digital radiography system comprised of an acquisition workstation, a gantry including an electrical cabinet housing the system power and communication controls, and an acquisition software to obtain diagnostic images. Two identical sets of detectors and X-ray tubes are positioned orthogonally to generate frontal and lateral images simultaneously by scanning the patient over the area of interest. If desired, the Micro Dose feature enables image acquisition for assessing global skeletal deformities in follow-up pediatric exams. The two sets of detectors and X-ray tubes are identical between the predicate device and the subject device. The image acquisition can be performed with a MANUAL mode or an AUTO mode of patient examination. To select the area of interest (acquisition area), a vertical collimation is set using green lasers and to correctly position the patient in the EOSedge, a centering system is used based on red lasers. The diagnostic images are stored in a local database and are displayed on a high-resolution medical-quality non-diagnostic monitor. The diagnostic image can be transmitted through a DICOM compatible digital network for printing and archiving.

{4}------------------------------------------------

The frontal and lateral detectors are identical. The technical specifications of the detectors are presented in the Table 1 below:

AttributeTechnical Specifications
Quantity2
TypeHybrid CdTe-CMOS dual energy photons counting X-ray detector
DimensionsWidth 585 mm x 98 mmThickness 131 mm
Weight<10 kg
X-ray conversion materialCdTe
X-ray voltage range< 160 kVp
Active area514 mm x 6 mm
Pixel size100 x 100 $\mu m^2$ , elementary
Tile gap<=100 $\mu m$
Overall fill factor99.63%
Number of pixels/lines5,139 pixels
Imaging speedUp to 5,000 lines/s, in TDS mode
Counting output range (Pixel depth)17 bits (TDS mode) (> 131 000 gray levels)
Nominal input voltage12 VDC – 20 A
Temperature controlInternal check, Peltier + PWMVentilator system checked
DQE typeRQA5 spectrum ~ 80%
MTF type~70% @ 2lp / mm~25% @ 5lp / mm
Table 1: DETECTORS SPECIFICATIONS

Intended Use / Indications for Use

EOSedge is intended for use in general radiographic exams and applications, excluding the evaluation of lung nodules and exams involving fluoroscopy, and mammography. EOSedge allows the radiographic acquisition of either one or two orthogonal X-ray images, for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient, in the upright or seated position.

The Micro Dose feature is indicated for assessing global skeletal deformities in follow-up pediatric exams.

Performance Data

EOSedge is designed and has been certified to conform to IEC 60601-1 and collateral standards. Software verification and validation testing was also conducted. Additional performance and functional testing have confirmed the equivalent performance of EOSedge compared to the cleared predicate EOS System. This included bench testing to confirm appropriate dosing and image quality. Bench performance testing were conducted based on FDA's Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices (September 1, 2016), to verify that EOSedge performs according to specifications and is as safe and effective as the predicate device.

All the testing standards used for bench testing of the EOSedge system are listed in Table 2 below:

{5}------------------------------------------------

Standards N°.Standards TitleDate
AAMI/ANSIES60601-1:2005/(R)2012 and A1:2012Medical electrical equipment. Part 1: general requirements for basicsafety and essential performance2012
IEC 60601-1-2Medical electrical equipment. Part 1-2: general requirements forsafety; electromagnetic compatibility-requirements and tests.2014
IEC 60601-1-3Medical electrical equipment. Part 1-3: general requirements forsafety; general requirements for radiation protection in diagnostic X-ray equipment.2013
IEC 60601-1-6Medical electrical equipment. Part 1-6: General requirements for basicsafety and essential performance -Collateral standard: Usability2013
IEC 60601-2-54Medical electrical equipment. Part 2-54: Particular requirements forthe basic safety and essential performance of X-ray equipment forradiography and radioscopy2015
IEC 60825-1Safety of laser products - Part 1: Equipment classification andrequirements2007
IEC 62220-1Medical Electrical Equipment-Characteristics Of Digital X-RayImaging Devices, Part 1-1: Determination Of The Detective QuantumEfficiency, Detectors Used In Radiographic Imaging2015

Table 2: Bench Testing Standards Used for the EOSedge system

Substantial Equivalence

EOSedge has the same intended use, indications, principles of operation, and technological characteristics as the cleared predicate device. EOSedge is an updated version of the cleared EOSedge System (K192079) with the main modifications being optimization of some acquisition protocol parameters, improvement of image processing and change in system data processing.

A substantial equivalence table summarizing the similarities and differences between EOSedge and its predicate device is provided in Table 3 below.

EOS imaging's EOSedgeEOS imaging's cleared EOSedgeSystem (K192079)
Intended UseGeneral X-ray imaging systemGeneral X-ray imaging system
Indications forUseEOSedge is intended for use in generalradiographic exams and applications,excluding the evaluation of lung nodulesand exams involving fluoroscopy,angiography, and mammography.EOSedge allows the radiographicacquisition of either one or two orthogonalX-ray images, for diagnostic purposes, inone single scan, of the whole body or areduced area of investigation of a patient,in the upright or seated position.The Micro Dose feature is indicated forassessing global skeletal deformities infollow-up pediatric examinations.EOSedge is intended for use in generalradiographic exams and applications,excluding the evaluation of lung nodulesand exams involving fluoroscopy,angiography, and mammography.EOSedge allows the radiographicacquisition of either one or two orthogonalX-ray images, for diagnostic purposes, inone single scan, of the whole body or areduced area of investigation of a patient,in the upright or seated position.The Micro Dose feature is indicated forassessing global skeletal deformities infollow-up pediatric examinations.
ContraindicationsEOSedge is not designed to performfluoroscopy, angiography ormammography exams.The Micro Dose feature is not designed formorphological analysis of bone structuresand lesions.EOSedge is not designed to performfluoroscopy, angiography ormammography exams.The Micro Dose feature is not designed formorphological analysis of bone structuresand lesions.
EOS imaging's EOSedgeEOS imaging's cleared EOSedgeSystem (K192079)
The Micro Dose feature is not indicated:The Micro Dose feature is not indicated:
for analysis of spine static in the presence of fusion material or for analysis of the bone-implant interface in the case of focal skeletal anomalies and/or other pediatric abnormalities for use in patients with a Body Mass Index over 30for analysis of spine static in the presence of fusion material or for analysis of the bone-implant interface in the case of focal skeletal anomalies and/or other pediatric abnormalities for use in patients with a Body Mass Index over 30
User PopulationTrained medical personnelTrained medical personnel
TechnologicalCharacteristicsDigital radiography system in which twosets of detectors and X-ray tubes arepositioned orthogonally to generate frontaland lateral images simultaneously byscanning the patient over the area ofinterest. The 2 orthogonal acquisitionchains consist of HV generators, X-raytubes, collimators and detectors,positioned on C-shaped arms translatingDigital radiography system in which twosets of detectors and X-ray tubes arepositioned orthogonally to generate frontaland lateral images simultaneously byscanning the patient over the area ofinterest. The 2 orthogonal acquisitionchains consist of HV generators, X-raytubes, collimators and detectors,positioned on C-shaped arms translating
along a vertical axis.along a vertical axis.
Dimensions (I x w2.58 m x 2.58 m x 2.706 m2.58 m x 2.58 m x 2.706 m
x h)(8.5 ft x 8.5 ft x 8.9 ft)(8.5 ft x 8.5 ft x 8.9 ft)
Weight2 005 kg (4 420 lb)2 005 kg (4 420 lb)
AccessoriesLaser safety barriersMotorized lifting platformLaser safety barriersMotorized lifting platform
Platform consolePlatform console
Laser positioning systemLaser positioning system
Access and stabilization barsAccess and stabilization bars
Bar code readerBar code reader
Principles ofOperationTube preheatingSelection of the patient, the acquisitionplanes, and the anatomical area, as wellas patient positioningSelection of either automatic mode ormanual modeSelection of patient morphotypeScout ViewDisplay of the exposure area,configuration of the reference planes, andverification and validation of theacquisition parametersImage analysis and export of the examDose ReportPrintingTube preheatingSelection of the patient, the acquisitionplanes, and the anatomical area, as wellas patient positioningSelection of either automatic mode ormanual modeSelection of patient morphotypeScout ViewDisplay of the exposure area,configuration of the reference planes, andverification and validation of theacquisition parametersImage analysis and export of the exam
Permanentminimum totalfiltration1.7 mm Al at 75 kV1.7 mm Al at 75 kV
(Al equivalent)Additional filtrations:0.1 mm copper thickness (used for images with or without Micro-Dose) 0.5 mm copper thickness (used for the Scout View)Additional filtrations:0.1 mm copper thickness (used for images with or without Micro-Dose) 0.5 mm copper thickness (used for the Scout View)
DetectorsDirect conversion device - solid statedetectorDirect conversion device - solid statedetector
17 bits (> 131 000 grey levels)17 bits (> 131 000 grey levels)
Pixel Depth
Pixel Size100 μm100 μm
Resolution3.7 lp/mm3.7 lp/mm
Typical Dynamic> 100 dB> 100 dB
Range
EOS imaging's EOSedgeEOS imaging's cleared EOSedgeSystem (K192079)
Values of the kVparameters40 to 140 kV, bi-polar40 to 140 kV, bi-polar
Values of the mAparameters10 to 500 mA10 to 500 mA
SoftwareSpecificationsPatient information managementImage acquisitionDisplay imagesGeneration of dose reportSend exam to the PACS or USBMaintenance managementAccess management to the systemSystem acquisition configurationPatient information managementImage acquisitionDisplay imagesSend exam to the PACSMaintenance managementAccess management to the systemSystem acquisition configuration.
Focal Spot0.6 x 1.3 at 120 kV - 100 mA0.6 x 1.3 at 120 kV - 100 mA
Linear ScanningSpeed (cm/s)From 4.1 to 32.5From 4.1 to 32.5
AverageAcquisition Time7 seconds for a spine and 13 seconds foran entire body8 seconds for a spine and 15 seconds foran entire body
Laser PositioningSystem
Maximum averageradiant power/wavelength0.136 mW / 520 nm (green)0.130 mW / 635 nm (red)0.136 mW / 520 nm (green)0.130 mW / 635 nm (red)
Class according toIEC 60825-1Class1Class1

Table 3: EOS IMAGING'S EOSEDGE - SUBSTANTIAL EQUIVALENCE CHART

{6}------------------------------------------------

{7}------------------------------------------------

Conclusion

Modified EOSedge has the same intended use, indications for use, technological characteristics, and similar principles of operation as its predicate device, the cleared EOSedge System (K192079). The minor differences in principles of operation and software specifications between EOSedge and its predicate device raise no new questions of safety or effectiveness. Performance data further demonstrates that modified EOSedge is as safe and effective as the cleared EOSedge System (K192079). Thus, modified EOSedge is substantially equivalent.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.