K Number
K202379

Validate with FDA (Live)

Date Cleared
2021-03-19

(211 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Invacare AVIVA Storm RX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.

Device Description

The subject device is a rearward drive wheel version substantially equivalent to the previously cleared forward drive wheel Invacare® AVIVA FX Power Wheelchair (K192216) with LiNX Electronics and Ultra Low Maxx Seating System. The subject device version consists of the following changes:

  • . Rearward drive wheel placement
  • . Modification to suspension system
  • . Modification to battery box design and seat interface brackets
  • Minor changes to mechanical components such as new rims and colors .
    The Invacare® AVIVA Storm RX Power Wheelchair is a 24V DC system, motor-driven wheelchair, utilizing the predicate device LiNX® control system. The subject device consists of a rigid or "non-folding" type power wheelchair base with rearward drive wheel placement with two (2) casters in the front and two anti-tippers in the rear. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors.
AI/ML Overview

The provided text describes a 510(k) premarket notification for a powered wheelchair, the Invacare AVIVA Storm RX Power Wheelchair. It focuses on demonstrating substantial equivalence to a predicate device (Invacare AVIVA FX Power Wheelchair) through non-clinical testing against various international standards.

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally established by compliance with a comprehensive set of international (ISO), American National Standards Institute (ANSI), and European (EN) standards. The reported device performance is that the "subject device components met the performance requirements."

Acceptance Criteria (Test Standard)Reported Device Performance
ISO 7176-1:2014: Wheelchairs Part 1: Determination of Static StabilitySubject device met performance requirements.
ISO 7176-2:2017: Wheelchairs Part 2: Determination of Dynamic Stability of Electrically Powered WheelchairsSubject device met performance requirements.
ISO 7176-3:2012: Wheelchairs Part 3: Determination of Effectiveness of BrakesSubject device met performance requirements.
ISO 7176-4:2008: Wheelchairs Part 4: Energy Consumption of Electrical Wheelchairs and Scooters for Determination of Theoretical Distance RangeSubject device met performance requirements.
ISO 7176-5:2008: Wheelchairs Part 5: Determination of Dimensions, Mass and Maneuvering SpaceSubject device met performance requirements.
ISO 7176-6:2018: Wheelchairs Part 6: Determination of Maximum Speed, Acceleration and Deceleration of Electric WheelchairsSubject device met performance requirements.
ISO 7176-7:1998: Wheelchairs Part 7: Measurement of Seating and Wheel DimensionsSubject device met performance requirements.
ISO 7176-8:2014: Wheelchairs Part 8: Requirements and Test Methods for Static, Impact and Fatigue StrengthsSubject device met performance requirements.
ISO 7176-9:2009: Wheelchairs Part 9: Climatic Tests for Electric WheelchairsSubject device met performance requirements.
ISO 7176-10:2008: Wheelchairs Part 10: Determination of Obstacle Climbing Ability of Electrically Powered WheelchairsSubject device met performance requirements.
ISO 7176-11:2012: Wheelchairs Part 11: Test DummiesSubject device met performance requirements.
ISO 7176-13:1989: Wheelchairs Part 13: Determination of Coefficient of Friction of Test SurfaceSubject device met performance requirements.
ISO 7176-14:2008: Wheelchairs Part 14: Power and Control Systems for Electrically Powered Wheelchairs and Scooters – Requirements and Test MethodsSubject device met performance requirements.
ISO 7176-15:1996: Wheelchairs Part 15: Requirements for Information Disclosure, Documentation and LabelingSubject device met performance requirements.
ISO 7176-16:2012: Wheelchairs Part 16: Resistance to Ignition of Postural Support DevicesSubject device met performance requirements.
ISO 7176-19: Wheelchairs Part 19: Wheeled Mobility Devices for use as Seats in Motor VehiclesSubject device met performance requirements.
ISO 7176-21:2008: Wheelchairs Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargersSubject device met performance requirements.
ISO 7176-22:2014: Wheelchairs Part 22: Set-up ProceduresSubject device met performance requirements.
ISO 7176-25:2013: Wheelchairs Part 25: Batteries and chargers for powered wheelchairsSubject device met performance requirements.
IEC 62304:2006: Medical Device Software – Software Life CycleSubject device met performance requirements.
ANSI C63.27:2017: American National Standard for Evaluation of Wireless CoexistenceSubject device met performance requirements.
Other tests mentioned (Flammability, Ignitability of Upholstered Furniture)"Subject device components and features are the same or substantially equivalent to the predicate device."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a "test set" in terms of a patient cohort or a specific number of devices tested individually. Instead, it refers to "non-clinical test reports" and "design verification test reports" that support the finding of substantial equivalence. The testing is based on compliance with established international standards for wheelchairs. Therefore:

  • Sample size: Not explicitly stated as a numerical sample of devices or patients. The assumption is that sufficiently representative samples were used to conduct the required tests under each standard.
  • Data provenance: Not explicitly stated, as the focus is on compliance with standards rather than clinical data from a specific geographical region or retrospective/prospective study. The testing is implied to be laboratory-based non-clinical testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document does not describe the use of human experts to establish "ground truth" for a performance test set. The evaluation is based on objective, standardized non-clinical engineering and performance tests carried out by the manufacturer to demonstrate compliance with recognized standards.

4. Adjudication Method

There is no mention of an adjudication method as the evaluation relies on compliance with technical standards and objective measurements, not subjective expert reviews.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI or diagnostic devices where human reader performance is a key metric. This submission is for a physical medical device (power wheelchair) where performance is evaluated against engineering standards.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable. The device is a power wheelchair, not an algorithm, and its performance is evaluated against engineering and safety standards, not an algorithm's diagnostic output.

7. The Type of Ground Truth Used

The "ground truth" in this context is implicitly defined by the pass/fail criteria of the various international and national standards (ISO, IEC, ANSI) listed in the document. These standards define objective performance thresholds and test methods for aspects like stability, braking, speed, safety, and electromagnetic compatibility.

8. The Sample Size for the Training Set

This question is not applicable. There is no mention of a "training set" as this is a physical medical device, not a machine learning algorithm.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG" in a larger font size and "ADMINISTRATION" in a smaller font size below.

March 19, 2021

Invacare Corporation Elijah Wreh Regulatory Affairs Manager One Invacare Way Elyria, Ohio 44035

Re: K202379

Trade/Device Name: Invacare AVIVA Storm RX Power Chair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: February 10, 2021 Received: February 16, 2021

Dear Elijah Wreh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202379

Device Name Invacare AVIVA Storm RX Power Wheelchair

Indications for Use (Describe)

The Invacare AVIVA Storm RX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter Information per 21 CFR 807.92(a)(1)

SPONSOR:Invacare CorporationOne Invacare WayElyria, OH 44035
------------------------------------------------------------------------
PRIMARY CONTACT:Elijah Wreh
Regulatory Affairs Manager
Phone: (440) 329-6840
Email: ewreh@invacare.com
MANUFACTURER:Invacare Corporation1200 Taylor StreetElyria, Ohio 44035
---------------------------------------------------------------------------------
DATE PREPARED per 21 CFR 807.92(a)(1):March 19, 2021
--------------------------------------------------------

Device Information per 21 CFR 807.92(a)(2)

NAME OF SUBJECT DEVICE:Invacare® AVIVA Storm RX Power Wheelchair
COMMON/USUAL NAME:Power Wheelchair
CLASSIFICATION NAME:Powered Wheelchair [21 CFR §890.3860]
REGULATORY CLASS:2
PRODUCT CODE:ITI: Wheelchair, Mechanical
PREDICATE DEVICES:Primary: Invacare AVIVA FX Power Wheelchair(K192216)No reference devices were used in this submission.

{4}------------------------------------------------

Device Description per 21 CFR 808.92(a)(4)

The subject device is a rearward drive wheel version substantially equivalent to the previously cleared forward drive wheel Invacare® AVIVA FX Power Wheelchair (K192216) with LiNX Electronics and Ultra Low Maxx Seating System. The subject device version consists of the following changes:

  • . Rearward drive wheel placement
  • . Modification to suspension system
  • . Modification to battery box design and seat interface brackets
  • Minor changes to mechanical components such as new rims and colors .

The Invacare® AVIVA Storm RX Power Wheelchair is a 24V DC system, motor-driven wheelchair, utilizing the predicate device LiNX® control system. The subject device consists of a rigid or "non-folding" type power wheelchair base with rearward drive wheel placement with two (2) casters in the front and two anti-tippers in the rear. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors.

Intended Use per 21 CFR 807.92(A)(5)

The intended use of the device is to provide mobility and positioning to persons limited to a sitting position.

Indications for Use per FORM FDA 3881

The Invacare® AVIVA Storm RX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.

Indications for Use Characteristics Comparison

Both the subject and predicate device share the same Indications for Use and Intended use.

Technological Characteristics Comparison with the predicate device per 21 CFR 807.92(a)(6)

The technological characteristics comparison demonstrates that the subject device is substantially equivalent in intended use, design, materials, and operational principles to the previously cleared predicate device.

{5}------------------------------------------------

COMPARISON of TECHNOLOGICAL CHARACTERISTICS with the PREDICATE DEVICE per 21 CFR 807.92(a)(6)

The device comparison shows that the subject device is substantially equivalent in intended use, design, materials, and operational principles to the previously cleared predicate device. The subject and predicate devices are intended to provide mobility and positioning to persons limited to a sitting position. The subject device components met the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise different questions of safety and effectiveness. The different technological characteristics do not raise different questions of safety and effectiveness as compared to the predicate device.

Basis of Substantial Equivalence per 21 CFR 807.100(b)(2)(ii)(A)

The substantial equivalence of the subject device was determined as per the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" and the technological characteristics which include materials, design, energy source, and other device features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A). The design verification testing and device comparison demonstrates that the subject device is substantially equivalent to the predicate device. The data generated from the subject device design verification test reports support a finding of substantial equivalence regarding device comparison, device specifications, and design characteristics and to provide mobility and positioning to persons limited to a sitting position.

Summary of the Nonclinical Testing used for Substantial Equivalence

The performance testing, device comparison, and dimensional analysis demonstrate that the subject device components and features are the same or substantially equivalent to the predicate device regarding the following:

  • . Static Stability
  • Dynamic Stability of Electric Wheelchairs ●
  • . Effectiveness of Brakes
  • . Energy Consumption of Electric Wheelchairs and Scooters for Determination of Theoretical Distance Range
  • . Dimensions Mass and Maneuvering Space
  • . Maximum Speed Acceleration and Deceleration of Electric Wheelchairs
  • . Seating and Wheel Dimensions
  • Methods for Static Impact and Fatigue Strengths .

{6}------------------------------------------------

  • Climatic Tests for Electric Wheelchairs ●
  • Climbing Ability of Electrically Powered Wheelchairs ●
  • Power and Control Systems for Electrically Powered Wheelchairs and Scooters ●
  • Information Disclosure Documentation and Labeling ●
  • Wheeled Mobility Devices for Use as Seats in Motor Vehicles
  • Methods for Electromagnetic Compatibility of Electrically Powered Wheelchairs and ● Scooters and Battery Chargers
  • Batteries and chargers for powered wheelchairs ●
  • Software Life Cycle ●
  • Flammability Testing .
  • Assessment of the Ignitability of Upholstered Furniture Ignition Source Smoldering ● Cigarette
  • Wireless Coexistence

The data generated from the subject Invacare® AVIVA Storm RX Power Wheelchair design verification test reports support a finding of substantial equivalence regarding the device comparison, dimensional analysis, device specifications, design characteristics and to provide mobility and positioning to persons limited to a sitting position.

Indications for Use Comparison Table

DeviceSubject DevicePredicate Device
Invacare® AVIVA Storm RX Power WheelchairInvacare AVIVA FX Power Wheelchair(K192216)
Indications forUseTo provide mobility and positioning to personslimited to a sitting positionSame
Intended UseTo provide mobility and positioning to personslimited to a sitting positionSame
Type of UsePrescription (RX Only)Same

{7}------------------------------------------------

ComponentDescriptionInvacare® AVIVA FX Power Wheelchair (K192216) Predicate DeviceInvacare® AVIVA Storm RX Power Wheelchair Subject Device
SEATINGUltraLowMaxxPowered Positioning ConfigurationsFixed, Combinations of Tilt/Recline/ElevateCombinations of Tilt/Recline
Seat Widths16" to 24"Same
Seat Depths16" to 23"Same
Back Heights18" to 25" (tilt) or 20" to 27" (tilt and recline)Same
UpholsteryMeshtex, Startex, Spacetex, O-Vinyl, PolyesterSame
Tilt Range50°Same
Recline Range168°Same
Patient Support SurfacesRef subject device 'Essential Equivalence' sectionSame
System NameLiNXSame
CONTROLSYSTEMLiNXCablesVariable cable lengthsA range of standard cable lengths availableSame
System ArchitectureMicroprocessor ControlledSame
Non-Expandable OptionsYesSame
Expandable OptionsYesSame
Wireless DevicesBluetoothSame
Power Source24V nominalSame
Bus InterfaceCANSame
BASEBase ConfigurationForward Drive Wheel PlacementRearward Drive Wheel Placement
Base Width24.3"24.8"
Base Length42.7"33.5"
Ground Clearance2.6"2.8"
Batteries(2) 12V Sealed VRLA Gel BatteriesSame
Braking SystemElectro-mechanical Friction BrakeSame
Drive Wheel Size14" x 3"Same
Incline CapabilitySame
Maximum Speed5.8mphSame
Motors4-Pole SSDSame
Occupant WeightCapUp to 300lbs.Same
SuspensionFour Bar Linkage Suspension SystemFour Bar Linkage Suspension System
Four Bar Linkage Suspension System(Front Drive Wheel)Four Bar Linkage Suspension System(Rear Drive Wheel)

Design and Technological Characteristics Comparison – Finished Device

{8}------------------------------------------------

{9}------------------------------------------------

DescriptionSubject DeviceInvacare® A VIVA StormRX Power WheelchairPredicate DeviceInvacare A VIVA FX Power Wheelchair (K192216)
Base ConfigurationRear Wheel DriveFront Wheel Drive
Base Width24.8"24.3"
Base Length33.5"42.7"
Seat-to-Floor Heights16.75"17.25"18.25"19.25"16.75"17.75"18.75"19.75"
Ground Clearance2.8"2.6"
Batteries(2) 12V Sealed VRLA Gel Batteries(2) 12V Sealed VRLA Gel Batteries
Braking SystemElectro-mechanical Friction BrakeElectro-mechanical Friction Brake
Castor Size8" x 2.5"8" x 2.5"
Control SystemLiNXLiNX
Drive Wheel Size14" x 3"14" x 3"
Incline Capability
Maximum Speed5.8 mph5.8 mph
Motors4-Pole SSD4-Pole SSD
Maximum WeightBearing Capacity300lbs.300lbs.
Range>12 miles>12 miles
SuspensionFour Bar Linkage Independent Suspension System(Rear Drive Wheel)Four Bar Linkage Independent Suspension System(Front Drive Wheel)

Performance Characteristics Comparison – Finished Device

{10}------------------------------------------------

LiNX Control System- ModulesDescriptionAVIVA Storm RXAVIVA FXWireless TechnologyImpact to Safety and Effectiveness
Primary Driver ControlsDrive Only RemoteYesYesBluetooth LE Wireless: DiagnosticsThe versions are the same.
Remotes with LED DisplayYesYesBluetooth LE Wireless: DiagnosticsThe versions are the same.
Rehabilitation Remote with Colour LCD Display and TouchYesYesBluetooth LE Wireless: Mouse Mover and DiagnosticsThe versions are the same.
Alternative Driver Controls (LiNX)Compact RemoteYesYesBluetooth LE Wireless: Mouse Mover and DiagnosticsThe versions are the same.
Sip 'N' Puff (SNP)YesYesNo Wireless
Attendant Driver ControlYesYesNo Wireless
Power Modules120A, Dual Motor, Power Controller with Lights and 2 Actuator ChannelsYesYesNo WirelessThe versions are the same.
Actuator ModulesActuator Controls only modules: 4 ChannelYesYes
Stability ControlG-Trac / GyroYesYesNo WirelessThe versions are the same.
Programming toolLAK: LiNX Access KeyYesYesBluetooth LE Wireless: Remote Programming
Switch Input ModuleModule used to add separate switched for Power on/off when used with REM 215/216Yes - The difference is that the switch input module adds wired remote on/off capabilityNoNo Wireless*Differences
LiNX Control System- ModulesDescriptionAVIVA Storm RXAVIVA FXWireless TechnologyImpact to Safety andEffectiveness
Environmental Control ModuleEnvironmental Control module used foroff-chair control of various devicesYesYesNo WirelessThe versions are thesame.
Switch Interface ModuleSwitch Interface ModuleYes – Thedifference is theswitch module is ajunction box.NoNo Wireless*Differences
Light ModuleInterface to LED lights on StormRXchairYes – Thedifference is lowvoltage powersupply for LEDlight.NoNo Wireless*Differences
USB Charge PortAllows charging of Smart pone viaUSBYesYesNo WirelessThe versions are thesame.
Enhanced DLX-IN500 InputModuleSip 'N' Puff (SNP)Yes – Same as theregular DLX-IN500 inputmodule, but thedifference is thatthe firmware wasupdated.YesNo Wireless*Differences

LiNX® Components and Accessories Specifications

{11}------------------------------------------------

*Differences: Differences between the subject device features and the previously cleared predicate device do not raise new issues of safety or effectiveness.

{12}------------------------------------------------

Performance Data

Non-Clinical Test per 21 CFR 807.92(b)(1)

International Organization of Standardization (ISO) testing, California Technical (CAL), American National Standards Institute (ANSI) and European (EN) standards testing were performed to demonstrate that the subject Invacare® AVIVA Storm RX Power Wheelchair meet the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise questions of safety and effectiveness.

{13}------------------------------------------------

Test StandardTest Description
ISO 7176-1:2014Wheelchairs Part 1: Determination of Static Stability
ISO 7176-2:2017Wheelchairs Part 2: Determination of Dynamic Stability of Electrically Powered Wheelchairs
ISO 7176-3:2012Wheelchairs Part 3: Determination of Effectiveness of Brakes
ISO 7176-4:2008Wheelchairs Part 4: Energy Consumption of Electrical Wheelchairs and Scooters for Determination of Theoretical Distance Range
ISO 7176-5:2008Wheelchairs Part 5: Determination of Dimensions, Mass and Maneuvering Space
ISO 7176-6:2018Wheelchairs Part 6: Determination of Maximum Speed, Acceleration and Deceleration of Electric Wheelchairs
ISO 7176-7:1998Wheelchairs Part 7: Measurement of Seating and Wheel Dimensions
ISO 7176-8:2014Wheelchairs Part 8: Requirements and Test Methods for Static, Impact and Fatigue Strengths
ISO 7176-9:2009Wheelchairs Part 9: Climatic Tests for Electric Wheelchairs
ISO 7176-10:2008Wheelchairs Part 10: Determination of Obstacle Climbing Ability of Electrically Powered Wheelchairs
ISO 7176-11:2012Wheelchairs Part 11: Test Dummies
ISO 7176-13:1989Wheelchairs Part 13: Determination of Coefficient of Friction of Test Surface
ISO 7176-14:2008Wheelchairs Part 14: Power and Control Systems for Electrically Powered Wheelchairs and Scooters – Requirements and Test Methods
ISO 7176-15:1996Wheelchairs Part 15: Requirements for Information Disclosure, Documentation and Labeling
ISO 7176-16:2012Wheelchairs Part 16: Resistance to Ignition of Postural Support Devices
ISO 7176-19Wheelchairs Part 19: Wheeled Mobility Devices for use as Seats in Motor Vehicles
ISO 7176-21:2008Wheelchairs Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers
ISO 7176-22:2014Wheelchairs Part 22: Set-up Procedures
ISO 7176-25:2013Wheelchairs Part 25: Batteries and chargers for powered wheelchairs
IEC 62304:2006Medical Device Software – Software Life Cycle
ANSI C63.27:2017American National Standard for Evaluation of Wireless Coexistence

{14}------------------------------------------------

Performance Data Conclusions per 21 CFR 807.92(b)(3)

The subject device utilizes the same intended use, same material composition, and similar technological characteristics as the predicate device. The non-clinical laboratory data support a finding of substantial equivalence of the subject device and demonstrate that the subject device will perform as intended in the specified use conditions.

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).