K Number
K202372
Date Cleared
2020-12-30

(133 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Upper Arm Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia).

Device Description

The Upper Arm Electronic Blood Pressure Monitor, Model BM100-EN operation is based on the oscillometric method. This method takes advantage of the pressure pulsations taken during measurements. An occluding cuff is placed on the left arm and is connected to an air pump and a pressure sensor. The cuff is inflated until a pressure higher than the typical systolic value is reached, then the cuff is slowly deflated. As the cuff deflates, when systolic pressure value approaches, pulsations start to appear. These pulsations represent the pressure changes due to heart ventricle contraction and can be used to calculate the heartbeat rate. Pulsations grow in amplitude until mean arterial pressure (MAP) is reached, then decrease until they disappear. The blood pressure information is displayed as digital information on the device's screen.

AI/ML Overview

The provided text is related to a 510(k) premarket notification for an Upper Arm Electronic Blood Pressure Monitor, Model BM100-EN. The document describes the device, compares it to a predicate device, and outlines the testing conducted to support its substantial equivalence.

However, the request asks for specific details about acceptance criteria and a study that proves the device meets the acceptance criteria, particularly in the context of an AI/ML medical device. The provided document is for a traditional medical device (blood pressure monitor) and does not contain information about AI/ML acceptance criteria or an AI-based study. It focuses on performance standards (like IEC 80601-2-30) for blood pressure measurement accuracy, not AI performance metrics.

Therefore, I cannot fulfill the request as it asks for information (AI acceptance criteria, AI study details, expert ground truth, MRMC study, etc.) that is not present in the provided document. The document describes non-clinical and performance testing for a standard blood pressure monitor.

To directly answer the prompt and explain why I cannot provide the requested information, here's a breakdown of what is and is not in the text:

Information Present in the Document:

  • Device Name: Upper Arm Electronic Blood Pressure Monitor, Model BM100-EN
  • Predicate Device: Jiangsu Yuyue Medical Equipment & Supply Co., Ltd Upper Arm Type Electronic Blood Pressure Monitor Series, Electronic Blood Pressure Monitor: YE670D (K170605)
  • Indication for Use: Measuring blood pressure and pulse rate of adults at household or medical center (not for neonate, pregnancy, or pre-eclampsia).
  • Measurement Method: Non-invasive, oscillometric.
  • Performance Standards Met: IEC 80601-2-30 (Medical electrical equipment Part 2-30: Particular Performance requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers) and ISO 81060-2 (Non-invasive sphygmomanometers Part 2: Clinical validation of automated measurement type).
  • Accuracy Claims (from characteristics table): Blood Pressure Range Accuracy: ±3 mmHg (±0.4kPa); Pulse Rate Measurement Accuracy: ±5% of reading value.
  • Non-Clinical Testing: Safety and EMC (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11), Software Verification and Validation (following FDA Guidance for moderate level of concern software), Cleaning Validation, Biocompatibility (ISO 10993-1, ISO 10993-5, ISO 10993-10).
  • Clinical Study: "a comparative clinical study to verify the performance of the subject device and predicate device as well as a mercury sphygmomanometer according to ISO 81060-2:2013." This is a study to validate the accuracy of the blood pressure readings against a reference, not an AI performance study.

Information Not Present in the Document (and thus, cannot be provided to answer your specific questions about AI/ML acceptance criteria and studies):

  • Table of AI acceptance criteria and reported device performance: The document lists performance criteria for a blood pressure monitor (e.g., accuracy ±3 mmHg) but not AI-specific metrics like sensitivity, specificity, AUC, F1-score, etc.
  • Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): While a "comparative clinical study" is mentioned, no details about sample size (number of subjects), data provenance, or whether it was retrospective/prospective are provided.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as this is for a blood pressure monitor, not an AI image interpretation device. Ground truth for blood pressure is typically established by direct measurement with a reference standard (like a mercury sphygmomanometer) in a clinical setting.
  • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for this type of device.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable for this device.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical measurement device, not an AI algorithm.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For blood pressure, the ground truth is often the reading from a validated reference device (e.g., mercury sphygmomanometer) or invasive arterial pressure. The document mentions comparison to a mercury sphygmomanometer.
  • The sample size for the training set: Not applicable, as this is not an AI/ML device that requires a training set in that context.
  • How the ground truth for the training set was established: Not applicable.

In conclusion, the provided FDA 510(k) clearance document is for a basic medical device (blood pressure monitor) and does not describe an AI/ML device or its associated acceptance criteria and validation studies as detailed in your query.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 30, 2020

Conmo Electronic Company Limited % Charlie Mack Principal Engineer International Regulatory Consultants 2950 E Lindrick Drive Chandler, Arizona 85249

Re: K202372

Trade/Device Name: Upper Arm Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 23, 2020 Received: December 1, 2020

Dear Charlie Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for LT Stephen Browning Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K202372

Device Name

Upper Arm Electronic Blood Pressure Monitor, BM100-EN

Indications for Use (Describe)

Upper Arm Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia).

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "CONMO" in blue letters. The word is surrounded by two curved lines, one above and one below, that form an oval shape. The background is white.

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92.

Date: August 13, 2020

    1. Company and Correspondent submitting this 510(k):
      Name – Conmo Electronic Company Limited Address – Room 1402, Building No.10 Shenzhen Biomedicine Innovations Industrial Park Jinhui Road No.14, Kengzi Street Pingshan District Shenzhen City, Guangdong, China 518126 Tel: +86-755-28397469 Fax: +86-755-82922769 Contact - Mrs. Alice Qi General Manager

US Agent and Correspondent Mr. Charles Mack Principal Engineer IRC USA 2950 E Lindrick Drive, Chandler, Arizona 85249 USA Tel: 931-6254938 Email: charliemack@irc-us.com

    1. Device:
Trade/proprietary name:Upper Arm Electronic Blood Pressure MonitorModel BM100-EN
Common Name:Non-invasive blood pressure measurement system
Classification Name:System, measurement, blood-pressure, non-invasive
Product Code:DXN
Regulation Number:21CFR870.1130
Device Class:2

{4}------------------------------------------------

    1. Predicate Device :
ManufacturerPredicate Device510(k)Number
Jiangsu YuyueMedicalEquipment&Supply Co., LtdUpper Arm TypeElectronic BloodPressure MonitorSeries, ElectronicBlood PressureMonitor: YE670DK170605

4. Reason for Submission: New Device

5. Device Description:

The Upper Arm Electronic Blood Pressure Monitor, Model BM100-EN operation is based on the oscillometric method. This method takes advantage of the pressure pulsations taken during measurements. An occluding cuff is placed on the left arm and is connected to an air pump and a pressure sensor. The cuff is inflated until a pressure higher than the typical systolic value is reached, then the cuff is slowly deflated. As the cuff deflates, when systolic pressure value approaches, pulsations start to appear. These pulsations represent the pressure changes due to heart ventricle contraction and can be used to calculate the heartbeat rate. Pulsations grow in amplitude until mean arterial pressure (MAP) is reached, then decrease until they disappear.

The blood pressure information is displayed as digital information on the device's screen.

{5}------------------------------------------------

    1. Indication for use:
      The Upper Arm Electronic Blood Pressure Monitor is intended to measure the blood pressure and pulse rate of adults at a household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia).
    1. Comparison with the predicate device:
      Conmo Electronic Company Limited believes that the Upper Arm Electronic Blood Pressure Monitor, BM100-EN is substantially equivalent to the Jiangsu Yuyue Medical Equipment& Supply Co., Ltd Upper Arm Type Electronic Blood Pressure Monitor Series, Electronic Blood Pressure Monitor: YE670D, FDA 510(k) clearance K170605.

{6}------------------------------------------------

CharacteristicsSubmitted DevicePredicate DeviceDifference Discussion
Device Name and modelUpper Arm Electronic Blood PressureMonitor, BM100-ENUpper Arm Type Electronic Blood PressureMonitor YE670DN/A
ManufacturerCONMO ELECTRONIC COMPANYLIMITEDJiangsu Yuyue Medical Equipment & SupplyCo., LtdN/A
510(K) NumberPendingK170605N/A
Indication for UseThe Upper Arm Electronic Blood PressureMonitor is intended to measure the bloodpressure and pulse rate of adults at ahousehold or medical center. (Not suitablefor neonate, pregnancy or pre-eclampsia).Electronic blood pressure monitor isintended to measure the blood pressure andpulse rate of adult at household or medicalcenter. (Not suitable for neonate, pregnancyor pre-eclampsia).Identical
Clinical Usehousehold or medical centerhousehold or medical centerIdentical
Patient PopulationAdultAdultIdentical
Measurement TypeUpper armUpper armIdentical
Measurement MethodNon-invasive, SociometricNon-invasive, SociometricIdentical
Cuff Circumference220mm ~ 320mm220mm ~ 320mmIdentical
Power Source4x1.5V4x1.5VIdentical
Measurement PressureRange0~295mmHg (0-39.3kPa)0~280mmHg (0-37.3kPa)The Pressure Range of BM100-EN is widerthan Predicate Devices, but still complieswith IEC 80601-2-30
Range Accuracy±3 mmHg(±0.4kPa)±3 mmHg(±0.4kPa)Identical
Measurement PulseRange40 ~ 180 beats/min40 ~ 200 beats/minThe Pulse Range of BM100-EN is narrowerthan Predicate Devices, but still complies withIEC 80601-2-30
Pulse Rate MeasurementAccuracy±5% of readingvalue±5% of readingvalueIdentical
Number of Users22Identical
Memory Space9960The difference does not raise any newquestions of safety and effectiveness.
Number of buttons33Identical
CharacteristicsSubmitted DevicePredicate DeviceDifference Discussion
ControlsUser 1/ ConfirmGuest/SettingUser 2/PlusSET Button,Plus/Minus Button,START/PULSE ButtonThe control difference does not raise any newquestions of safety and effectiveness.
Display TypeLCDLCDIdentical
Pressurization ModeAutomatic InflationAutomatic InflationIdentical
Deflation ModeAutomatic Pressure Release ValveAutomatic Pressure Release ValveIdentical
Display ContentCuff Pressure, Pulse,Date,Time,Systolic/DiastolicPressure, error message,Measurements result in memory,Irregular Heart BeatFeature, Bodymovement detectionCuff Wrapping DetectionCuff Pressure, Pulse,Date,Time,Systolic/DiastolicPressure, error message,Measurements result in memory,Irregular Heart BeatFeature, Bodymovement detectionCuff Wrapping DetectionIdentical
Operation EnvironmentsTemperature: +10 °C~ +40 °CHumidity: 15% RH~ 85% RH(nocondensation)Temperature: +10 °C~ +40 °CHumidity: 15% RH~ 90% RH(nocondensation)The operating environment difference doesnot raise any new questions of safety andeffectiveness.
Storage EnvironmentsTemperature: -20 °C~ +55 °CHumidity: 15% RH~ 85% RH(nocondensation)Temperature: -20 °C~ +55 °CHumidity: 15% RH~ 90% RH(nocondensation)The storage environment difference does notraise any new questions of safety andeffectiveness.
PerformanceANSI/AAMI/ISO 81060-2IEC 80601-2-30ANSI/AAMI/ISO 81060-2IEC 80601-2-30Identical
BiocompatibilityISO 10993-1, FDA Guidance, Testsincluded Cytotoxicity, Sensitization andIntracutaneous ReactivityISO 10993-1, FDA Guidance, Tests includedCytotoxicity, Sensitization andIntracutaneous ReactivityIdentical
Electrical SafetyIEC 60601-1IEC 60601-1Identical
EMCIEC 60601-1-2IEC 60601-1-2Identical
Home UseIEC 60601-1-11IEC 60601-1-11Identical
Prescription or OTCOTCOTCIdentical

{7}------------------------------------------------

{8}------------------------------------------------

Testing Summary:

The following performance data is provided in support of the substantial equivalence determination.

Non-Clinical Study:

Non-clinical tests were conducted to verify that the proposed device meets the same design specifications as the predicate Yuyue Medical Equipment device. The test results demonstrate that the proposed device complies with the following standards:

Safety and EMC

Testing was performed to verify the basic safety and essential performance of the Upper Arm Electronic Blood Pressure Monitor. BM100-EN. The following tests were performed:

  • IEC 60601-1: 2005+A1: 2012 Medical electrical equipment Part 1: - General requirements for basic safety and essential performance
  • IEC 60601-1-2: 2014 Medical devices part 1-2: General requirements for basic safety and essential performance – Collateral standards: electromagnetic compatibility - Test and requirements
  • IEC 60601-1-11 Edition 2.0 2015-01, medical electrical equipment part 1--11: General requirements for basic safety and essential performance collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

The test results demonstrate the device complies with the necessary safety and essential performance standards requirements.

{9}------------------------------------------------

Performance Data:

The subject Upper Arm Electronic Blood Pressure Monitor, BM100-EN was subjected to the following tests and passed all test criteria:

  • IEC 80601-2-30 Edition 2.0 Medical electrical equipment Part 2-30: Particular Performance 2018-03 requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
  • ISO 81060-2 Non-invasive sphygmomanometers Part 2: Clinical validation NIBP Performance Second edition 2013-05-01 of automated measurement type
  • FDA Guidance Non-Invasive Blood Pressure (NIBP) Monitor Guidance -

Software Verification and Validation

Software documentation, including verification & validation, was provided following FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern.

Cleaning Validation

The device was cleaned following the procedure defined in the User's Manual. The device was checked for performance following the cleaning and met all performance requirements following the cleaning.

Shelf Life:

The Upper Arm Blood Pressure Monitor BM100-EN, is not subject to the shelf life, as the device doesn't contain any sterile or degradable components.

Biocompatibility

The subject Upper Arm Electronic Blood Pressure Monitor, BM100-EN, uses similar material to the predicate Yuyue Medical Equipment device. Biocompatibility testing was performed to demonstrate compliance with the same biocompatibility standards as performed by the predicate device.

{10}------------------------------------------------

The Upper Arm Blood Pressure Monitor, Model BM100-EN device is classified per ISO 10993-1: 2009 Annex 1 Biological evaluation tests as follows: Surface Device - Contact Skin - Contact Duration <24h

Based on this classification, the following cytotoxicity, skin irritation and sensitization tests were conducted per the following standards:

  • ISO 10993-5: 2009 Biological evaluation of medical devices Part 5 Tests । for In Vitro Cytotoxicity
  • ISO 10993-10: 2010 Biological evaluation of Medical Devices Part 10: । Tests for Irritation and Delayed-Type Hypersensitivity

The test results confirm compliance with the requirements of the standards.

Sterility Information:

The device is not delivered sterile, nor is it required to be sterilized for use.

Package and Shelf Life:

The Upper Arm Blood Pressure Monitor, Model BM100-EN, is not subject to the shelf life, as the device contains no sterile or degradable components.

Clinical Study:

We conducted a comparative clinical study to verify the performance of the subject device and predicate device as well as a mercury sphygmomanometer according to ISO 81060-2:2013. The study demonstrates a correlation in the performance of subject device and predicate device. was significantly correlated

The design of the submitted device specifications is substantially the same as the predicate.

All the labeling and characteristics of the submitted Upper Arm Electronic Blood Pressure Monitor, BM100-EN, is the same as the predicate device, and most normal blood pressure monitors currently on the market. The submitted device and predicate both use similar measuring methodologies and components to achieve the measurements.

{11}------------------------------------------------

    1. Conclusions:
      Following the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Conmo Electronic Company Limited concludes that the Upper Arm Electronic Blood Pressure Monitor, BM100-EN is substantially equivalent to predicate Yuyue Medical Equipment, Inc's device as described herein.

Conmo Electronic Company Limited will update and include in a summary any other information deemed reasonably necessary by the FDA.

END

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).