(107 days)
The Dermatological Carbon Dioxide Laser System is used for human tissue vaporization, coagulation in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.
The Dermatological Carbon Dioxide Laser Systems, and it is also a gas laser. With its own wavelength of 10600 nm, the laser energy density is high and the divergence angle is very small. The focus light beam was used to vaporization, coagulation the human tissue.
The provided document is a 510(k) summary for a Dermatological Carbon Dioxide Laser System. This document does not describe a study involving an AI/Machine Learning device or a diagnostic device. Instead, it focuses on demonstrating substantial equivalence of a physical medical device (a laser system) to a predicate device through non-clinical testing and comparison of technical specifications.
Therefore, many of the requested criteria related to AI/ML study design (ground truth, expert readers, MRMC studies, training/test sets, etc.) are not applicable to this submission.
However, I can extract the acceptance criteria and performance as presented for this specific device:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are implied by the comparison to a predicate device and compliance with established safety and performance standards. The "performance" in this context refers to the device's technical specifications matching or being acceptably different from the predicate, and its compliance with safety standards.
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance/Compliance |
|---|---|---|
| General Comparison (Substantial Equivalence) | ||
| Product Code | Must be GEX | GEX |
| Regulation Number | Must be 21 CFR 878.4810 | 21 CFR 878.4810 |
| Class | Must be Class 2 | Class 2 |
| Where used | Must be hospital | hospital |
| Intended Use | Vaporization, coagulation in dermatology, plastic surgery, general surgery, gynecology, podiatry, dental, otorhinolaryngology | Same as predicate device (human tissue vaporization, coagulation in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology) |
| Performance Comparison (Technical Specifications) | ||
| Maximum Power | 30W | 30W |
| Work Mode | CW Mode (Continuous Mode), P Mode (Multi-Pulse) | Surgery (Single Pulse, Continuous, Multi-Pulse) (SAME, despite different naming convention) |
| Wavelength | 10.6 um | 10.6 um |
| Beam Delivery | 7 joint Light guide arm | 7 knucklearmkey joints light arm (SAME) |
| Light Arm Length | Compared to predicate (1.36m) | 1.576m (Analysis: "different not affect the therapeutic effect") |
| Aiming Beam | 650nm | 630-650nm red diode laser (≤5 mW) (SAME) |
| Spot Size | Compared to predicate (0.5 mm ±10%) | 0.4 mm (Analysis: "only affects the area of treatment, not affect the therapeutic effect") |
| Pulse Setting | 0.1-1000ms | 1-1000ms (SAME) |
| Power Calibration | Period of 1 year | Period of 1 year |
| Control System | Touch screen, footswitch | Touch screen, footswitch |
| Laser Operation | Footswitch | Footswitch |
| Laser Medium/Energy Source | CO2 | CO2 |
| Cooling System | Air cooling | Air cooling |
| Clean Method | 70% alcohol | 70% medical alcohol (SAME) |
| Patient Contacted Part | Skin | Skin |
| Dimension | Compared to predicate (6642125cm without light arm) | 52cm40cm125cm (Analysis: "complying with IEC 60601-1, the mechanical performance... accepted") |
| Weight | Compared to predicate (80 kg) | 65Kg (Analysis: "complying with IEC 60601-1, the mechanical performance... accepted") |
| Power input | AC 110V/60Hz | AC 100-240v~, 50/60Hz (SAME, implying compatibility) |
| Safety Comparison (Standards Compliance) | ||
| Electrical Safety | Comply with IEC 60601-1, IEC 60601-2-22 | Complied |
| EMC | Comply with IEC 60601-1-2 | Complied |
| Laser Safety | Comply with IEC 60601-2-22, IEC 60825 | Complied |
| Patient Contact Materials | Must be safe | Aluminum (Analysis: "meet the requirements of ISO 10993-1") |
| Cytotoxicity | Comply with ISO 10993-1 | Complied |
| Sensitization | N/A - Covered by ISO 10993-1 compliance | N/A - Covered by ISO 10993-1 compliance |
| Irritation | N/A - Covered by ISO 10993-1 compliance | N/A - Covered by ISO 10993-1 compliance |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not applicable. This is a physical device submission, not an AI/ML diagnostic or predictive model that uses a "test set" of patient data. The "testing" involved non-clinical performance and safety evaluations of the device itself.
- Data Provenance: Not applicable. The "data" are technical specifications and test results from the device manufacturer, not patient data from a specific country or collected retrospectively/prospectively.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. There is no "ground truth" in the context of an AI/ML model for this device. The assessment is based on engineering specifications, safety standards, and comparison to a legally marketed predicate device.
4. Adjudication method for the test set:
- Not applicable. There is no test set or expert adjudication process for this type of device submission.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI-based device, and therefore no MRMC study was conducted or relevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm or software-only device.
7. The type of ground truth used:
- Not applicable. The "ground truth" or basis for acceptance in this submission relies on:
- Predicate Device Specifications: The technical and performance characteristics of a previously cleared device (K192528).
- International Standards Compliance: Adherence to recognized medical device safety and performance standards (e.g., IEC 60601-1, IEC 60601-2-22, IEC 60825, IEC 60601-1-2, ISO 10993 series).
- Non-Clinical Testing: Results of tests performed on the device to confirm its specifications and safety.
8. The sample size for the training set:
- Not applicable. There is no "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established:
- Not applicable. There is no training set for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 25, 2020
Zhuolu Jontelaser Manufacturing Technology Co., Ltd % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District BeiJing, BeiJing 102401 China
Re: K202250
Trade/Device Name: Dermatological Carbon Dioxide Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: August 27, 2020 Received: August 31, 2020
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K202250
Device Name
Dermatological Carbon Dioxide Laser System
Indications for Use (Describe)
The Dermatological Carbon Dioxide Laser System is used for human tissue vaporization, coagulation in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| Over-The-Counter Use (21 CFR 801.109) |
| X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
-
Date of Preparation 1.
11/24/2020 -
- Applicant Name and Address
Zhuolu Jontelaser Manufacturing Technology Co., Ltd. No. 31, Sanguanmiao Alley, Zhuolu Town, Zhuolu County, Zhangjiakou City, Hebei, China, 072750
- Applicant Name and Address
-
Contact Person Information 3. Karen Liu General Manager Tel: +86-13911459627 Fax: +86-10- 57212057 Email: 919309354@qq.com
-
- Submission Correspondent
Mr. Ray Wang Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing City, China, 102401 Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com
{4}------------------------------------------------
-
Identification of Proposed Device 5.
Trade Name: Dermatological Carbon Dioxide Laser Systems Common Name: Powered Laser Surgical Instrument Model(s): L300 Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery; -
Identification of Primary Predicate 6. 510(k) Number: K192528 Product Name: CO2 Laser Therapy Machine Manufacturer: Shangdong Huamei Technology Co., Ltd.
-
Device Description 7.
The Dermatological Carbon Dioxide Laser Systems, and it is also a gas laser. With its own wavelength of 10600 nm, the laser energy density is high and the divergence angle is very small. The focus light beam was used to vaporization, coagulation the human tissue.
-
- Indication For Use
The Dermatological Carbon Dioxide Laser Systems is used for human tissue vaporization, coagulation in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.
- Indication For Use
-
Substantially Equivalent (SE) Comparison 9.
Table 1 General Comparison
| ITEM | Proposed Device | Predicate Device | Remark |
|---|---|---|---|
| Product Code | GEX | GEX | SAME |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | SAME |
| Class | 2 | 2 | SAME |
| Where used | hospital | hospital | SAME |
| Intended Use | The Dermatological Carbon Dioxide LaserSystems is used for human tissuevaporization, coagulation in dermatology | The CO2 Laser Therapy Machine is usedfor human tissue vaporization, coagulationin dermatoloty and plastic surgery, general | SAME |
{5}------------------------------------------------
| and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology. | surgery, gynecology, podiatry, dental and otorhinolaryngology. |
|---|---|
| --------------------------------------------------------------------------------------------- | ---------------------------------------------------------------- |
Table 2 Performance Comparison
| ITEM | Proposed Device | Predicate Device | Remark |
|---|---|---|---|
| Maximum Power | 30W | 30W | SAME |
| work mode | CW Mode (Continuous Mode)P Mode (Muti-Pulse) | Surgery (Single Pulse, Continuous, Muti-Pulse) | SAME |
| Wavelength | 10.6 um | 10.6 um | SAME |
| Beam delivery | 7 joint Light guide arm | 7 knucklearmkey joints light arm | SAME |
| Light arm | 1.576m | 1.36m | Analysis |
| Aiming Beam | 650nm | 630-650nm red diode laser (≤5 mW) | SAME |
| Spot size | 0.4 mm | 0.5 mm (±10%) | Analysis |
| Pulse Setting | 1-1000ms | 0.1-1000ms | SAME |
| Powercalibration | Period of 1 year | Period of 1 year | SAME |
| Control System | Touch screen, footswitch | Touch screen, footswitch | SAME |
| Laser operation | Footswitch | Footswitch | SAME |
| Laser medium/energy source | CO2 | CO2 | SAME |
| Cooling System | Air cooling | Air cooling | SAME |
| Clean Method | 70% alcohol | 70% medical alcohol | SAME |
| PatientContacted Part | Skin | Skin | SAME |
| Dimension | 52cm40cm125cm | 6642125cm (without light arm) | Analysis |
| Weight | 65Kg | 80 kg | Analysis |
| Power input | AC 100-240v~, 50/60Hz | AC 110V/60Hz | SAME |
Analysis 1
The proposed device is different in light arm length from the predicate device, the different not affect the therapeutic effect. Therefore, this difference will not affect the substantially equivalency.
Analysis 2
The proposed device is different in Spot Size from the predicate device, Spot size only affects the
{6}------------------------------------------------
area of treatment, not affect the therapeutic effect. Therefore, this difference will not affect the substantially equivalency.
Analysis 3
The proposed device is different in dimension and weight from the predicate device. By complying with IEC 60601-1, the mechanical performance of the proposed device is determined to be accepted, therefore, this difference will not affect the substantially equivalency.
Table 3 Safety Comparison
| Item | Proposed Device | Predicate Device | Remark |
|---|---|---|---|
| EMC, Electrical and Laser Safety | |||
| Electrical Safety | Comply with IEC 60601-1, IEC60601-2-22 | Comply with IEC 60601-1, IEC60601-2-22 | SE |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE |
| Laser Safety | Comply with IEC 60601-2-22, IEC 60825 | Comply with IEC 60601-2-22, IEC 60825 | SE |
| Patient Contact Materials and Biocompatibility | |||
| Patient ContactMaterials | Aluminum | / | Analysis |
| Cytotoxicity | Comply with ISO 10993-1 | Comply with ISO 10993-1 | SE |
| Sensitization | |||
| Irritation |
Analysis
Although the materials in contact with human skin are different, they all meet the requirements of ISO 10993-1 and will not affect product safety.
10. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
-
IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance:
-
IEC 60601-2-22:2007, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment;
-
IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements.
-
IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic
{7}------------------------------------------------
safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.
-
ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity
-
ISO 10993-10:2010, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity
-
Software Validation & Verification Test
-
- Clinical Testing
No clinical study is included in this submission.
-
- Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
- Conclusion
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.