K Number
K192528

Validate with FDA (Live)

Date Cleared
2019-12-03

(81 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CO2 Laser Therapy System is used for human tissue vaporization in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.

Device Description

The CO2 Laser Therapy System is a carbon dioxide laser used in medical and aesthetic industry. Due to the CO2 laser's high absorption of water, its high-energy beam of laser light interacts with the skin's surface causing the upper layer to peel off and and then achieve the indications. The proposed device is mainly used for human tissue vaporization, coagulation and exposure to achieve the purpose of treatment.

AI/ML Overview

It appears there might be a misunderstanding regarding your request. The provided document is an FDA 510(k) summary for a CO2 Laser Therapy System, which is a medical device for surgical applications. This document does NOT contain information about an AI/ML-based device or a study involving acceptance criteria for such a device, human readers, ground truth establishment, or training sets.

My analysis of the provided text reveals that the medical device in question is a CO2 Laser Therapy System. This is a physical, hardware-based medical equipment used for tissue vaporization and coagulation. The FDA submission focuses on demonstrating its substantial equivalence to a predicate device, primarily through non-clinical testing against established safety and performance standards for laser equipment.

Therefore, I cannot fulfill your request for information about:

  • Acceptance criteria and reported device performance for an AI/ML device: The document describes a physical laser device, not an algorithm.
  • Sample size and data provenance for a test set (related to AI/ML): There is no mention of an AI/ML test set. The "test results" refer to engineering and safety tests on the hardware.
  • Number of experts and qualifications for ground truth (related to AI/ML): No ground truth establishment for an AI/ML model is discussed.
  • Adjudication method (related to AI/ML): Not applicable to this type of device submission.
  • Multi-reader multi-case (MRMC) comparative effectiveness study with AI assistance: No AI component is mentioned.
  • Standalone (algorithm only) performance: Not applicable.
  • Type of ground truth used (pathology, outcomes data, etc. for AI/ML): Not applicable.
  • Sample size for the training set (for AI/ML): Not applicable.
  • How ground truth for the training set was established (for AI/ML): Not applicable.

The document explicitly states in Section 9: "No clinical study is included in this submission." and Section 8, "Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device."

The "acceptance criteria" discussed in this document are primarily related to compliance with international electrical, EMC, and laser safety standards (e.g., IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2) which are applied to the physical device. The "performance" refers to the device's physical specifications (e.g., maximum power, wavelength, work mode) and demonstrating they are equivalent to the predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 3, 2019

Shangdong Huamei Technology Co., Ltd. % Ray Wang Official Correspondent Beijing Believe-Med Technology Service Co., Ltd. Rm. 912, Building #15, XiYueHui, No.5, YiHe North Rd., Fangshan District Beijing, 102401 CN

Re: K192528

Trade/Device Name: CO2 Laser Therapy System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 11, 2019 Received: September 13, 2019

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K192528

Device Name CO2 Laser Therapy System

Indications for Use (Describe)

The CO2 Laser Therapy System is used for human tissue vaporization in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K192528

    1. The Submitter Information

Shangdong Huamei Technology Co., Ltd. No. 588, Changning Street, High-tech District, WeiFang, ShanDong, China, 261205 Contact Person: Xu QingHua Position: General Manager Tel: +86-536-2110001 Fax: +86-536-2109823 Email: xuqh(@yeah.net Submission Correspondent Mr. Ray Wang Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing City, China, 102401 Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang(@believe-med.com

2. Date of Preparation

11/25/2019

    1. Identification of Proposed Device
      Trade Name: CO2 Laser Therapy System Common Name: Powered Laser Surgical Instrument Model(s): HM-CL200 Regulatory Information: Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX; Regulation Number: 21 CFR 878.4810;

{4}------------------------------------------------

Review Panel: General& Plastic Surgery;

  • Identification of Predicate Device 4.
    510(k) Number: K161925 Product Name: CO2 Laser Therapy Machine Manufacturer: Beijing ADSS Development Co., Ltd

    1. Device Description
      The CO2 Laser Therapy System is a carbon dioxide laser used in medical and aesthetic industry. Due to the CO2 laser's high absorption of water, its high-energy beam of laser light interacts with the skin's surface causing the upper layer to peel off and and then achieve the indications. The proposed device is mainly used for human tissue vaporization, coagulation and exposure to achieve the purpose of treatment.
    1. Indication For Use Statement
      The CO2 Laser Therapy Machine is used for human tissue vaporization, coagulation in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.
    1. Substantially Equivalent (SE) Comparison

Table 1 General Comparison

ITEMProposed DevicePredicate Device K161925Remark
Product CodeGEXGEXSE
Regulation No.21 CFR 878.481021 CFR 878.4810SE
Class22SE
Where usedhospitalhospitalSE
Intended UseThe equipment is used for human tissue vaporization, coagulation in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.The equipment is used for human tissue vaporization, coagulation in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.SE

{5}------------------------------------------------

ITEMProposed DevicePredicate Device K161925Remark
Maximum Power30W30WSAME
work modeSurgery (Single Pulse, Continuous,Muti-Pulse)Surgery (Single Pulse, Continuous,Muti-Pulse)SAME
Wavelength10.6 um10.6 umSAME
Mode StructureTEM00TEM00SAME
Beam delivery7 knucklearmkey joints light arm7 knucklearmkey joints light armSAME
Light arm1.36m1.32mSIMILAR
Handpiece TypeBe part of light armBe part of light armSAME
Aiming Beam630-650nm red diode laser (<5 mW)650nm red diode laser(0.5 mW)SAME
Spot size0.5 mm0.5mmSAME
Pulse SettingSingle Pulse0.1-1000msPulse10-1000msSIMILAR
Muti-PulseTime On0.1-1000msTime Off0.1-1000msRepeatTime On10-1000 msTime Off10-1000 msSIMILAR
Continuous1-30WCW0-30WSIMILAR
Control SystemTouch screen, footswitchTouch screen, footswitchSAME
Laser operationFootswitchFootswitchSAME
Lasermedium/energysourceCO2CO2SAME
Cooling SystemAir coolingAir coolingSAME
Clean Method70% medical alcohol70% medical alcoholSAME
Dimension6642125cm(without light arm)FG900: 56X54X112 cm;FG900-B: 660X54X32 cm;FG900-C: 46X42X125 cmAnalysis
Weight80 kgFG900: 49 kg;FG900-B: 28kg;FG900-C: 34kgAnalysis
Power inputAC 110V/60Hz ;AC 110V/50Hz-60HzSAME

Table 2 Performance Comparison

{6}------------------------------------------------

ItemProposed DevicePredicate Device K161925Remark
EMC, Electrical and Laser Safety
Electrical SafetyComply with IEC 60601-1, IEC60601-2-22Comply with IEC 60601-1, IEC60601-2-22SE
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2SE
Laser SafetyComply with IEC 60601-2-22, IEC 60825Comply with IEC 60601-2-22, IEC 60825SE

Table 3 Safety Comparison

Analysis

The only difference between proposed device and predicate device is appearance (dimension, weight), which are not affect the safety and effectiveness of proposed device. Based on the nonclinical tests performed, the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device.

8. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1:2005/A1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;

  • IEC 60601-2-22:2012, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment;

  • IEC 60825-1: 2014, Safety of laser products - Part 1: Equipment classification and requirements.

  • IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.

  • Software Validation & Verification Test

    1. Clinical Test Conclusion

No clinical study is included in this submission.

    1. Conclusion drawn from clinical and non-clinical testing
      Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.