K Number
K202159
Device Name
actiTENS
Manufacturer
Date Cleared
2020-12-22

(141 days)

Product Code
Regulation Number
882.5890
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
actiTENS is intended to be used as: - Transcutaneous Electrical Nerve Stimulator (TENS), used for the following indications: - Symptomatic relief and management of chronic, intractable pain - Adjunctive treatment for post-surgical and post-trauma acute pain - Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities - Relief of pain associated with arthritis Programs P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12 and P13 correspond to TENS mode. - Electrical Muscle Stimulation (EMS), used for the following indications: - Temporary relaxation of muscle spasms - Prevent or retard disuse atrophy - Increase of local blood flow in the treatment area - Re-educate muscles - Maintain or increase the range of motion - Prevention of venous thrombosis of the calf muscles immediately after surgery Program P9 corresponds to EMS mode.
Device Description
actiTENS is a transcutaneous electrical nerve stimulation (TENS) medical device designed to manage chronic pain in people ages 22 and older. The compact design and flexible shape of the actiTENS electrical impulse generator allows to fix it for daily use directly to the body of the patient. The impulse generator is delivered with a separate cradle which allows safely recharging its nonremovable battery. The actiTENS is compatible with a selection of disposable electrodes and compatible connector cables transmit the electrical stimulation impulses to the targeted nerves according to the selected therapy. The electrical stimulation waveforms are biphasic and asymmetric. The actiTENS is controlled via a downloadable mobile app for iPhone which allows the comfortable and simple selection of the adequate stimulation program. The actiTENS is intended to be used by the patient at home and also for therapeutic application by medical professionals.
More Information

No
The summary does not mention AI or ML, and the device description and performance studies focus on standard TENS/EMS technology and electrical/software safety.

Yes

The device is a Transcutaneous Electrical Nerve Stimulator (TENS) and Electrical Muscle Stimulator (EMS), which are explicitly described as being used for symptomatic relief, management of chronic pain, temporary relief of pain, relaxation of muscle spasms, and other therapeutic indications.

No

The device is described as a transcutaneous electrical nerve stimulator (TENS) and electrical muscle stimulation (EMS) device used for pain relief and muscle conditioning, not for diagnosing medical conditions.

No

The device description explicitly states it includes an "electrical impulse generator," a "cradle" for charging, and is delivered with "disposable electrodes and compatible connector cables." These are hardware components, making it a hardware device controlled by software, not a software-only device.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that actiTENS is a Transcutaneous Electrical Nerve Stimulator (TENS) and Electrical Muscle Stimulation (EMS) device. These are therapeutic devices used for pain relief and muscle stimulation.
  • IVD Definition: In Vitro Diagnostics are medical devices used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information for diagnosis, monitoring, or screening. actiTENS does not perform any such analysis of bodily specimens.
  • Device Description: The description details a device that delivers electrical impulses to the body via electrodes, which is consistent with TENS/EMS technology, not IVD technology.
  • Performance Studies: The performance studies listed focus on electrical safety, software validation, and benchtop performance, which are relevant for a therapeutic electrical stimulation device, not an IVD. There are no studies related to analyzing biological samples or diagnostic accuracy.
  • Key Metrics: The "Not Applicable" for Key Metrics like Sensitivity, Specificity, PPV, and NPV further indicates it's not a diagnostic device, as these metrics are crucial for evaluating the performance of IVDs.

In summary, the actiTENS is a therapeutic medical device, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

actiTENS is intended to be used as:

  • Transcutaneous Electrical Nerve Stimulator (TENS), used for the following indications:
    • Symptomatic relief and management of chronic, intractable pain
    • Adjunctive treatment for post-surgical and post-trauma acute pain
    • Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities
    • Relief of pain associated with arthritis

Programs P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12 and P13 correspond to TENS mode.

  • Electrical Muscle Stimulation (EMS), used for the following indications:
    • Temporary relaxation of muscle spasms
    • Prevent or retard disuse atrophy
    • Increase of local blood flow in the treatment area
    • Re-educate muscles
    • Maintain or increase the range of motion
    • Prevention of venous thrombosis of the calf muscles immediately after surgery

Program P9 corresponds to EMS mode.

Product codes

GZJ, NGX, NYN, IPF

Device Description

actiTENS is a transcutaneous electrical nerve stimulation (TENS) medical device designed to manage chronic pain in people ages 22 and older.

The compact design and flexible shape of the actiTENS electrical impulse generator allows to fix it for daily use directly to the body of the patient.

The impulse generator is delivered with a separate cradle which allows safely recharging its nonremovable battery.

The actiTENS is compatible with a selection of disposable electrodes and compatible connector cables transmit the electrical stimulation impulses to the targeted nerves according to the selected therapy. The electrical stimulation waveforms are biphasic and asymmetric.

The actiTENS is controlled via a downloadable mobile app for iPhone which allows the comfortable and simple selection of the adequate stimulation program.

The actiTENS is intended to be used by the patient at home and also for therapeutic application by medical professionals.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

22 and older.

Intended User / Care Setting

The actiTENS is intended to be used by the patient at home and also for therapeutic application by medical professionals.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The following performance data were provided in support of the substantial equivalence determination:

  • Sterilization: N/A - Device is non-sterile.
  • Shelf-life Testing: The electrodes are already cleared under K070807, and shelf-life testing for the electrodes is provided in that submission. The battery is the only component that is likely to be impacted by time-dependent product degradation for the controller. The balance of the hardware is not expected to be negatively impacted by time-dependent product degradation. Information regarding the battery shelf-life was provided by the battery manufacturer and the battery shelf-life is 12 months.
  • Biocompatibility Testing: Biocompatibility of the electrodes was provided in a previous 510(k) submission (K070807).
  • Electrical safety and electromagnetic compatibility (EMC): The following electrical safety and EMC testing reports were provided:
    • Electrical Safety Testing IEC 60601-1, plus national deviations to address the differences between IEC 60601-1:2005 and ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012
    • EMC Testing IEC 60601-1-2
    • Usability IEC 60601-1-6
    • Home Use Environment IEC 60601-1-11
    • Nerve and Muscle Stimulators IEC 60601-2-10
  • Software Verification and Validation Testing: Software verification and validation testing was provided in accordance with IEC 62304 and FDA guidance documents.
  • Benchtop Performance Testing: The following benchtop performance testing was provided:
    • Transportation Testing ASTM D4169-16
    • Lithium Battery UL 1642
    • Usability IEC 62366-1
    • Lithium Battery IEC 62133-2
  • Animal Performance Testing: Animal performance testing was not required to demonstrate safety and effectiveness of the device.
  • Human Clinical Performance Testing: Clinical testing was not required to demonstrate the safety and effectiveness of the device.

Key results: The tests and comparison performed demonstrate the subject device is substantially equivalent to the predicate device. Therefore, the subject device is as safe and effective as the predicate device that has been legally marketed in the United States.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K182203

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 22, 2020

Sublimed % Robert Packard President Medical Device Academy, Inc. 345 Lincoln Hill Road Shrewsbury, Vermont 05738

Re: K202159

Trade/Device Name: actiTENS Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ, NGX, NYN, IPF Dated: July 30, 2020 Received: August 3, 2020

Dear Robert Packard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

1

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K202159

Device Name actiTENS

Indications for Use (Describe)

actiTENS is intended to be used as:

  • · Transcutaneous Electrical Nerve Stimulator (TENS), used for the following indications:

  • Symptomatic relief and management of chronic, intractable pain

  • Adjunctive treatment for post-surgical and post-trauma acute pain

  • Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities

  • Relief of pain associated with arthritis

Programs P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12 and P13 correspond to TENS mode.

· Electrical Muscle Stimulation (EMS), used for the following indications:

  • Temporary relaxation of muscle spasms
  • Prevent or retard disuse atrophy
  • Increase of local blood flow in the treatment area
  • Re-educate muscles
  • Maintain or increase the range of motion
  • Prevention of venous thrombosis of the calf muscles immediately after surgery

Program P9 corresponds to EMS mode.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:

  • -SUBMITTER SUBLIMED 137 rue de Mayoussard Moirans, F-38430 France Tel: +33(0)4-76-93-37-15
    Contact Person: Corinne Bulteau Date Prepared: December 22, 2020
II. DEVICE
Trade/Device Name:actiTENS
Classification Name:Transcutaneous Electrical Nerve Stimulator for Pain Relief
Regulation:21 CFR § 882.5890
Regulatory Class:Class II
Product Classification Code:GZJ, NGX, NYN, IPF
III. PREDICATE DEVICE
Predicate Manufacturer:JKH USA, LLC
Predicate Trade Name:JKH Stimulator Plus
Predicate 510(k):K182203

No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

actiTENS is a transcutaneous electrical nerve stimulation (TENS) medical device designed to manage chronic pain in people ages 22 and older.

The compact design and flexible shape of the actiTENS electrical impulse generator allows to fix it for daily use directly to the body of the patient.

The impulse generator is delivered with a separate cradle which allows safely recharging its nonremovable battery.

The actiTENS is compatible with a selection of disposable electrodes and compatible connector cables transmit the electrical stimulation impulses to the targeted nerves according to the selected therapy. The electrical stimulation waveforms are biphasic and asymmetric.

The actiTENS is controlled via a downloadable mobile app for iPhone which allows the comfortable and simple selection of the adequate stimulation program.

The actiTENS is intended to be used by the patient at home and also for therapeutic application by medical professionals.

4

It is available for prescription only.

V. INDICATIONS FOR USE

actiTENS is intended to be used as:

· Transcutaneous Electrical Nerve Stimulator (TENS), used for the following indications:

  • Symptomatic relief and management of chronic, intractable pain

  • Adjunctive treatment for post-surgical and post-trauma acute pain

  • Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities

  • Relief of pain associated with arthritis

Programs P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12 and P13 correspond to TENS mode.

· Electrical Muscle Stimulation (EMS), used for the following indications:

  • Temporary relaxation of muscle spasms
  • Prevent or retard disuse atrophy
  • Increase of local blood flow in the treatment area
  • Re-educate muscles
  • Maintain or increase the range of motion
  • Prevention of venous thrombosis of the calf muscles immediately after surgery

Program P9 corresponds to EMS mode.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVCE

The table below provides a comparison of indications for use and technological characteristics between the subject device and the predicate device in order to demonstrate substantial equivalence:

5

Comparison with Predicate Device:

Table 1: Comparison of actiTENS with JKH Stimulator Plus (K182203).

| Feature | actiTENS | JKH Stimulator Plus -
K182203 | Comparison |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | actiTENS is intended to be
used as:
• Transcutaneous Electrical
Nerve Stimulator (TENS),
used for the following
indications:

  • Symptomatic relief and
    management of chronic,
    intractable pain
  • Adjunctive treatment for
    post-surgical and post-trauma
    acute pain
  • Temporary relief of pain
    associated with sore and
    aching muscles due to strain
    from exercise or normal
    household and work activities
  • Relief of pain associated
    with arthritis
    Programs P1, P2, P3, P4, P5,
    P6, P7, P8, P10, P11, P12 and
    P13 correspond to TENS
    mode.
    • Electrical Muscle
    Stimulation (EMS), used for
    the following indications: | Over-the-Counter Use:
    TENS (Modes 1, 2, 4, 6,
    8):
    PL-029K12 and PL-
    029K13 are used for
    temporary relief of pain
    associated with sore and
    aching muscles in the
    shoulder, waist, back, arm,
    and leg, due to strain from
    exercise or normal household
    work activities. PL-0219K12
    and PL-
    029K13 are also intended
    for symptomatic relief an
    management of chronic,
    intractable pain and relief of
    pain associated with
    arthritis.
    The device of PL-029K12
    may be used during sleep.
    The device of PL-029K12 is
    labeled for use only with its
    own compatible electrodes.
    PMS (also called EMS,
    Modes 1, 3, 7):
    PL-029K12 and PL- | The subject device is not
    intended for OTC use, and
    the prescription use of the
    subject device is identical to
    the indications of the
    predicate model PL-
    029K12. |

6

| - Temporary relaxation of
muscle spasms | 029K13 are also intended to
temporarily increase local
blood circulation in the
healthy muscles of lower
extremities. |
|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| - Prevent or retard disuse
atrophy | Heating:
The device of PL-029K13
is intended for temporary
relief of minor aches and
pains. |
| - Increase of local blood flow
in the treatment area | Prescription Use:
PL-029K12 and PL-
029K13 are intended for
the following use: |
| - Re-educate muscles | -Symptomatic relief and
management of chronic,
interactable pain |
| - Maintain or increase the
range of motion | -Adjunctive treatment for
post-surgical and post-
trauma acute pain |
| - Prevention of venous
thrombosis of the calf
muscles immediately after
surgery | -Relief of pain associated
with arthritis |
| Program P9 corresponds to
EMS mode. | -Temporary relaxation of
muscle spasm |
| | -Prevention or retardation of
disuse atrophy |
| | -Muscle re-education |
| | -Maintaining or increasing
range of motion |
| | -Increase of local blood
flow in the treatment area |
| | -Prevention of post-
surgical venous
thrombosis through |

7

| | | immediate stimulation of
calf muscles | |
|-----------------------------------------|------------------------|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription or OTC | Prescription | OTC and Prescription | The subject device does not
include OTC indications,
but this eliminates the risks
associated with OTC use. |
| Power Source(s) | Rechargeable | Rechargeable or non-
rechargeable battery | The subject device is
limited to use with the
rechargeable lithium-ion
battery, but this does not
introduce any new risks. |
| Compliance with Voluntary
Standards? | Yes | Yes | Same |
| Compliance With 21 CFR
898? | Yes | Yes | Same |
| Functions and design | Electrical stimulation | Electrical stimulation and
heat | The subject device does not
include the IRT product
classification, and this
eliminates the risks
associated with that
function. |
| Maximum skin temperature | N/A | 43°C | The subject device does not
include the IRT product
classification and does not
provide heat |

8

actiTENSJKH Stimulator Plus – K182203Comparison
Maximum output
voltage (Volts+/- 20%) at
500ΩP130.2 PL- 029K12: 57.6Maximum output voltage of the subject
device is in the range or below this of the
predicate device. These technological
differences between the subject and
predicate device do not raise new safety
and effectiveness questions
P230.2 PL- 029K13: 46.0
P330
P430.2
P530.2
P630
P730.2
P830.2
P929.5
P1029.1
P1129.3
P1230.2
P1330.2
Maximum output
voltage (Volts+/- 20%) at
2KΩP157.9 PL- 029K12: 96.0Maximum output voltage of the subject
device is in the range or below this of the
predicate device. These technological
differences between the subject and
predicate device do not raise new safety
and effectiveness questions
P257.9 PL- 029K13: 90.4
P357.9
P457.9
P557.9
P657.9
P757.9
P857.9
P957.9
P1057.9
P1157.9
P1257.9
P1357.9
Maximum output
voltage (Volts+/- 20%) at
$10KΩ$P1
P2
P3
P4
P5
P6
P7
P8
P9
P10
P11
P12
P13N/APL-029K12: 134
PL-029K13: 124The EIG monitors the impedance of each
channel and stop any running program (or
refuse to launch one) if the measured
impedance is out of a specified range. This
feature ensures a proper installation of the
EIG, the cables and electrodes. If an
electrode peels off, stopping the stimulation
program quickly enough prevents the
patient from feeling an electric discharge as
the attached surface diminishes. The 10
$kOhm$ is at the upper range and the EIG
refused to launch a stimulation program,
thus the measurement could not be
performed
Maximum output
current(mA+/- 20%) at
$500Ω$P1
P2
P3
P4
P5
P6
P7
P8
P9
P10
P11
P12
P1360.4
60.4
60
60.4
60.4
60
60.4
60.4
59
58.2
58.6
60.4
60.460.4 PL-029K12: 115.2
60.4 PL-029K13: 92.0Maximum output current of the subject
device is in the range or below this of the
predicate device. These technological
differences between the subject and
predicate device do not raise new safety
and effectiveness questions
Maximum output
current(mA+/- 20%) at
2KΩP128.95PL-029K12: 48.0Maximum output current of the subject
device is in the range or below this of the
predicate device. These technological
differences between the subject and
predicate device do not raise new safety
and effectiveness questions
P228.95PL-029K13: 45.2
P328.95
P428.95
P528.95
P628.95
P728.95
P828.95
P928.95
P1028.95
P1128.95
P1228.95
P1328.95
Maximum output
current(mA+/- 20%) at
10kΩP1N/APL-029K12: 13.4The EIG monitors the impedanceof each
channel and stop any running program (or
refuse to launch one) if the measured
impedance is out of a specified range. This
feature ensures a proper installation of the
EIG, the cables and electrodes. If an
electrode peels off, stopping the stimulation
program quickly enough prevents the
patient from feeling an electric discharge as
the attached surface diminishes. The 10
kOhm is at the upper range and the EIG
refused to launch a stimulation program,
thus the measurement could not be
performed
P2PL- 029K13: 12.4
P3
P4
P5
P6
P7
P8
P9
P10
P11
P12
P13
P1197.5
Pulse Width (µSec)P2147 PL-029K12: 100These technological differences between
the subject and predicate device do not
raise new safety and effectiveness
questions
P3247.5 PL-029K13: 92
P4197.5
P5147.5
P6200 or 147
P7147.5
P8modulated 100-200
P9247.5
P10147.5
P11147.5
P1257.5
P13177.5
Frequency (Hz)P1100 PL-029K12: 156.2Frequencies of the subject device is in the
range or below this of the predicate
device. These technological differences
between the subject and predicate device
do not raise new safety and effectiveness
questions
P280 PL-029K13: 178.5
P32
P4100
P5100
P62 or 100
P7100
P8modulated 2-80
P950
P1080
P1180
P1280
P1310

Table 2: Comparison of actiTENS with JKH Stimulator Plus (K182203) with regards to TENS and EMS performance.

9

10

11

12

| Maximum Phase

charge (µC) at 500Ω
Maximum Phase
charge (µC) at 500ΩP111.9PL-029K12: 23.0The maximum phase charge is in the same
range than the maximum phase charge of
the predicate. Furthermore the highest
value for the predicate is over than the
intended device. This raises no new safety
or effectiveness questions.
P28.9PL-029K13: 16.9
P314.9
P411.9
P58.9
P612.0
P78.9
P812.1
P914.6
P108.6
P118.6
P123.5
P1310.7
Maximum current
density (mA/cm2) at
500ΩP13.0PL-029K12: 3.26The maximum current density is in the
same range than the maximum current
density of the predicate. Furthermore the
highest value for the predicate is over than
the intended device. This raises no new
safety or effectiveness questions.
P23.0PL-029K13: 3.29
P33.0
P43.0
P53.0
P63.0
P73.0
P81.7
P92.9
P102.9
P112.9
P123.0
P133.0

13

14

The subject device and predicate device have equivalent indications for use and technological characteristics, with the exception that the subject device does not include over-the-counter-use (NUH product classification) and the subject device does not include heat as a modality (IRT product classification).

The predicate device claims two main functions: electrical stimulation and heat. The predicate device modalities work independently. Each modality has its own independent therapeutic effect. The subject device claims only the electrical stimulation effect as many devices cleared by the FDA, while the heating modality is exempt from 510(k) clearance. This technological difference between the subject and predicate device does not raise new safety and effectiveness questions. Concerning the maximum output voltage and the maximum output current, the subject device has been developed as a current generator with a maximum setpoint current of 60mA and has been limited by hardware to deliver output voltage of 60V. Thus it delivers a current of 60mA at 60V for 1KΩ. Above this charge the maximum output power will decrease. In each case the maximum output current and maximum output voltage are lower than the predicate. The subject device presents a maximum phase charge lower than the predicate device. In addition, the output voltage, phase charge, current density and average power density depends on the output current that is set by the user. The output current always starts at 0mA and is increased manually by the user with a 0.5mA increment, to achieve a strong but comfortable sensation, so the output voltage, output current, phase charge, current density and average power density delivered are therapeutically effective with the subject device. These technological differences between the subject and predicate device do not raise new safety and effectiveness questions.

Maximum output current and maximum output voltage at 10k Ω are not defined in the subject device because 10k Ω is the upper limit of the range of charge at which the subject device stops delivering currents. For the comfort and safety of the patient a limited range of charge has been defined in the subject device, to forbid electric current generation in case of short circuit or if an electrode is peeling off of the patient skin.

The difference of frequencies and pulse width between the subject device and predicate device does not raise new types of safety or effectiveness questions because the user chooses the frequency and the pulse width through different kind of programs using the mobile application and programs are using standard TENS stimulation frequencies and pulse width. The maximum current density and maximum average power density are directly linked to the pads surfaces used with the device. The only recommended pads with the subject device are 4.5cmX4.5cm then the density is calculating using the voltage and current generator capacities. The subject device presents maximum current density and maximum average power density higher than the predicate but the maximum power density is less than 250 mW/cm² as required by the guidance for Powered Muscle Stimulator the maximum power density is also less than 250 mW/cm². These technological differences between the subject and predicate device do not raise new safety and effectiveness questions.

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

15

Sterilization

N/A - Device is non-sterile.

Shelf-life Testing

The electrodes are already cleared under K070807, and shelf-life testing for the electrodes is provided in that submission.

The battery is the only component that is likely to be impacted by time-dependent product degradation for the controller. The balance of the hardware is not expected to be negatively impacted by time-dependent product degradation.

Information regarding the battery shelf-life was provided by the battery manufacturer and the battery shelf-life is 12 months.

Biocompatibility Testing

Biocompatibility of the electrodes was provided in a previous 510(k) submission (K070807).

Electrical safety and electromagnetic compatibility (EMC)

The following electrical safety and EMC testing reports were provided:

  • Electrical Safety Testing IEC 60601-1, plus national deviations to address the differences . between IEC 60601-1:2005 and ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012
  • EMC Testing IEC 60601-1-2
  • Usability IEC 60601-1-6 ●
  • Home Use Environment IEC 60601-1-11 ●
  • . Nerve and Muscle Stimulators IEC 60601-2-10

Software Verification and Validation Testing

Software verification and validation testing was provided in accordance with IEC 62304 and FDA guidance documents.

Benchtop Performance Testing

The following benchtop performance testing was provided:

  • Transportation Testing ASTM D4169-16 ●
  • Lithium Battery UL 1642 ●
  • Usability IEC 62366-1 ●
  • Lithium Battery IEC 62133-2 .

Animal Performance Testing

Animal performance testing was not required to demonstrate safety and effectiveness of the

16

device.

Human Clinical Performance Testing

Clinical testing was not required to demonstrate the safety and effectiveness of the device.

VIII. CONCLUSIONS

The tests and comparison performed demonstrate the subject device is substantially equivalent to the predicate device. Therefore, the subject device is as safe and effective as the predicate device that has been legally marketed in the United States.