(256 days)
Not Found
No
The description explicitly states the software uses "patented shape-recognition processes" and "deterministic algorithms" to sort candidate shapes. There is no mention of training data, learning, or adaptive behavior characteristic of AI/ML. The user interaction involves selecting from pre-calculated options or manually adjusting contours, not interacting with a system that learns or improves over time.
No.
Explanation: This device is a software tool that assists in measuring objects in medical images, but it explicitly states it does not make clinical decisions, is not a decision-support tool, and the information it provides must not be used in isolation for patient management, indicating it's not directly therapeutic.
Yes
Explanation: The device is intended to assist trained professionals in measuring dimensions of objects within a region of interest (ROI) in DICOM images, generating information such as estimated greatest long-axis and short-axis dimensions, area, volume, and margin of objects. While it does not make clinical decisions, its function of providing precise measurements of pathological objects from medical images is a clear diagnostic aid.
Yes
The device is explicitly described as a "software device" and a "software application" that integrates with existing hardware (medical Image Viewers and PACS workstations) but does not include or require specific hardware components of its own beyond the standard computing environment.
Based on the provided information, DeepLook PRECISE is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening.
- DeepLook PRECISE Function: DeepLook PRECISE operates on medical images (DICOM images) and assists in measuring objects within those images. It does not analyze biological specimens.
- Intended Use: The intended use clearly states that it assists trained professionals in measuring dimensions of objects within a region of interest in DICOM images. It explicitly states it does not make clinical decisions or is a decision-support tool.
- Device Description: The device description reinforces that it works with digital medical imaging and provides automated measurement, replacing manual digital calipers.
Therefore, DeepLook PRECISE falls under the category of medical image processing software, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
DeepLook PRECISE is a software device that is integrated into medical Image Viewers and PACS workstations to assist trained professionals in measuring dimensions of objects within a region of interest (ROI) that is identified by the user in DICOM images. The generated information consists of an estimated greatest long-axis and greatest short-axis dimensions, area, volume, and margin of the objects. For illustration purposes, DeepLook PRECISE can optionally provide a colorization of the interior area defined by a margin. DeepLook PRECISE does not make clinical decisions nor is a decision-support tool. The information provided by the software and must not be used in isolation when making patient management decisions.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using and FDA cleared monitor that offers at least 5 Megapixel resolutions and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained physicians and radiologists.
DeepLook PRECISE, is not intended for use on mobile devices.
Product codes (comma separated list FDA assigned to the subject device)
LLZ
Device Description
DeepLook PRECISE, is a software application that works embedded in PACS/EIS or OEM viewers. It provides automated measurement, replacing manual digital calipers currently used to measure objects in digital medical imaging.
It is a Windows OS service that uses XML messaging to receive Commands from the viewer application: it processes results independently and returns requested data to the viewer application via the same XML messaging. The software is not compiled within the viewer's application. As a result, integration is simplified and limited to establishing XML protocols for reciprocal Commands.
DeepLook PRECISE can operate on an individual workstation, local servers or in Cloudbased applications.
The software uses patented shape-recognition processes to analyze Regions of Interest (ROIs) identified by a trained medical user (i.e., imaging technologist, radiologist, phvsician or researcher). When requesting a measurement, the user triggers a set of XML Commands via the viewer interface. The primary Command is a request for a measurement of an object located within an ROI designated by the user using a mouse to click on the location of a suspected object.
DeepLook receives the Command, processes the area within the ROI and assembles the candidate shapes (and all relevant metrics) and returns a full set of displays (bundled and prioritized) that depict the possible boundaries of the object. To facilitate initial viewing, DeepLook PRECISE designates a default shape; this shape is recommended as the best depiction of the targeted margin of the object. The selection of the default shape from the display stack is determined by a set of deterministic algorithms that sort for the best candidate shape based on shape recognition ratios developed by DeepLook. Based on training and professional skill, a user can use simple keystroke Commands or a track ball or mouse wheel to move through the entire stack of alternative margins (shapes) and select the one that they conclude best represents the targeted object.
Once a candidate shape has been chosen, the user has the option to extend or contract any specific section of the margin in order to include or exclude a feature they deem relevant. Any alteration of the contours of the displayed margin will recalculate the overall measurement metrics (i.e., calculate two new axis measurements, and the area and estimated volume). The results will instantly appear with the modified margin.
The Commands to accept the default shape-display or select alternative shape-displays and any modification of selected shape-displays are all executed using keystroke or track ball functions that are initiated by the user through the viewer interface. Each resulting shape includes all relevant measurement metrics when displayed, allowing for quick comparison and selection.
For illustrative purposes, the software offers a colorization display of the internal shapes within a candidate margin: each of the shapes within the outer margin of a targeted object or anatomical structure is assigned a calibrated color. This is offered solely to assist the user in distinguishing the shape components when making a final selection. The colorization is not a decision-support or diagnostic tool.
The measurements and graphical display can be saved. Depending on the viewer configuration, this data can be saved 1) to the hard drive of the workstation; 2) to a server on the premises or in the Cloud; 3) in a structured report; or 4) configured to comply with DICOM data fields and saved to the PACS. The location of the saved display will be determined by the viewer manufacturer and/or the user.
The displays of DeepLook PRECISE described above offer the user consistent measurement of each shape. The modification functions provide the user with maximum flexibility to adjust the default shape or any other shapes in the display stack. The user can also decline all suggested shapes generated by DeepLook PRECISE and use standard mouse-operated digital calipers to measure the target object.
The final configuration of the user interface (i.e., keystrokes, track ball and hot-key Commands) will be determined by each vendor that integrates DeepLook PRECISE. To assist the vendor during integration, each XML Command and display option is illustrated separately in this manual. This provides the integrator with the option of selecting some or all the functions and to determine the best Commands to incorporate, consistent with its own user interface.
DeepLook PRECISE is stand-alone Windows OS service designed to permit maintenance and performance improvements without the need to modify established XML protocols used to send and receive Commands. This eliminates the need for additional vendor integrations when patches and/or updates are required to DeepLook PRECISE's Windows OS service.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
DICOM images
Anatomical Site
No restrictions
Indicated Patient Age Range
Not Found
Intended User / Care Setting
trained professionals, trained physicians and radiologists
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Nonclinical Testing:
The subject device has been assessed and tested at the factory to assess all functions and has passed all predetermined testing criteria.
Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.
Conclusion:
The 510(k) Pre-Market Notification for DeepLook PRECISE, software device contains adequate information, data, and nonclinical test results to enable FDA -CDRH to determine substantial equivalence to the predicate device.
The subject device will be manufactured in accordance with the voluntary standards listed in the voluntary standard survey. The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
Nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
April 9, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
DeepLook Inc. % Mr. Carl Alletto Regulatory Consultant OTech Inc. 8317 Belew Drive MCKINNEY TX 75071
Re: K202084
Trade/Device Name: DeepLook PRECISE Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: March 5. 2021 Received: March 15, 2021
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
1
devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K202084
Device Name DeepLook PRECISE
Indications for Use (Describe)
DeepLook PRECISE is a software device that is integrated into medical Image Viewers and PACS workstations to assist trained professionals in measuring dimensions of objects within a region of interest (ROI) that is identified by the user in DICOM images. The generated information consists of an estimated greatest long-axis and greatest short-axis dimensions, area, volume, and margin of the objects. For illustration purposes, DeepLook PRECISE can optionally provide a colorization of the interior area defined by a margin. DeepLook PRECISE does not make clinical decisions nor is a decision-support tool. The information provided by the software and must not be used in isolation when making patient management decisions.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using and FDA cleared monitor that offers at least 5 Megapixel resolutions and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained physicians and radiologists.
DeepLook PRECISE, is not intended for use on mobile devices.
Type of Use (Select one or both, as applicable) | |
---|---|
------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared: March 29, 2011
510(k) Number: K202084
I. SUBMITTER
DeepLook Inc. 1220 Burton Street Silver Spring, MD 20910 TEL: 202.306.0808 Email: Steve.schwadron@DeepLookMedical.com
Contact Person: Mr. Steven Schwadron, Chief Operating Officer
II. DEVICE
Name of Device: | DeepLook PRECISE |
---|---|
Common or Usual Name: | Picture Archive and Communications System |
Regulation Name: | System, Image Processing, Radiological |
Regulation: | 21 CFR 892.2050 |
Product Code: | LLZ |
Regulatory Class: | II |
III. PREDICATE DEVICE
Device Classification Name | System, Image Processing, Radiological |
---|---|
510(K) Number | K191530 |
Device Name | StoneChecker |
Applicant | Imaging Biometrics, LLC |
13416 Watertown Plank Road, Suite 260 | |
Elm Grove, WI 53122 | |
Regulation Number | 892.2050 |
Regulation Medical Specialty | Radiology |
IV. DEVICE DESCRIPTION
DeepLook PRECISE, is a software application that works embedded in PACS/EIS or OEM viewers. It provides automated measurement, replacing manual digital calipers currently used to measure objects in digital medical imaging.
It is a Windows OS service that uses XML messaging to receive Commands from the viewer application: it processes results independently and returns requested data to the viewer application via the same XML messaging. The software is not compiled within the viewer's application. As a result, integration is simplified and limited to establishing XML protocols for reciprocal Commands.
DeepLook PRECISE can operate on an individual workstation, local servers or in Cloudbased applications.
The software uses patented shape-recognition processes to analyze Regions of Interest (ROIs) identified by a trained medical user (i.e., imaging technologist, radiologist, phvsician or researcher). When requesting a measurement, the user triggers a set of XML Commands via the viewer interface. The primary Command is a request for a
4
measurement of an object located within an ROI designated by the user using a mouse to click on the location of a suspected object.
DeepLook receives the Command, processes the area within the ROI and assembles the candidate shapes (and all relevant metrics) and returns a full set of displays (bundled and prioritized) that depict the possible boundaries of the object. To facilitate initial viewing, DeepLook PRECISE designates a default shape; this shape is recommended as the best depiction of the targeted margin of the object. The selection of the default shape from the display stack is determined by a set of deterministic algorithms that sort for the best candidate shape based on shape recognition ratios developed by DeepLook. Based on training and professional skill, a user can use simple keystroke Commands or a track ball or mouse wheel to move through the entire stack of alternative margins (shapes) and select the one that they conclude best represents the targeted object.
Once a candidate shape has been chosen, the user has the option to extend or contract any specific section of the margin in order to include or exclude a feature they deem relevant. Any alteration of the contours of the displayed margin will recalculate the overall measurement metrics (i.e., calculate two new axis measurements, and the area and estimated volume). The results will instantly appear with the modified margin.
The Commands to accept the default shape-display or select alternative shape-displays and any modification of selected shape-displays are all executed using keystroke or track ball functions that are initiated by the user through the viewer interface. Each resulting shape includes all relevant measurement metrics when displayed, allowing for quick comparison and selection.
For illustrative purposes, the software offers a colorization display of the internal shapes within a candidate margin: each of the shapes within the outer margin of a targeted object or anatomical structure is assigned a calibrated color. This is offered solely to assist the user in distinguishing the shape components when making a final selection. The colorization is not a decision-support or diagnostic tool.
The measurements and graphical display can be saved. Depending on the viewer configuration, this data can be saved 1) to the hard drive of the workstation; 2) to a server on the premises or in the Cloud; 3) in a structured report; or 4) configured to comply with DICOM data fields and saved to the PACS. The location of the saved display will be determined by the viewer manufacturer and/or the user.
The displays of DeepLook PRECISE described above offer the user consistent measurement of each shape. The modification functions provide the user with maximum flexibility to adjust the default shape or any other shapes in the display stack. The user can also decline all suggested shapes generated by DeepLook PRECISE and use standard mouse-operated digital calipers to measure the target object.
The final configuration of the user interface (i.e., keystrokes, track ball and hot-key Commands) will be determined by each vendor that integrates DeepLook PRECISE. To assist the vendor during integration, each XML Command and display option is illustrated separately in this manual. This provides the integrator with the option of selecting some or all the functions and to determine the best Commands to incorporate, consistent with its own user interface.
DeepLook PRECISE is stand-alone Windows OS service designed to permit maintenance and performance improvements without the need to modify established XML protocols used to send and receive Commands. This eliminates the need for additional vendor integrations when patches and/or updates are required to DeepLook PRECISE's Windows OS service.
V. INDICATIONS FOR USE
DeepLook PRECISE is a software device that is integrated into medical Image Viewers and PACS workstations to assist trained professionals in measuring dimensions of objects within a region of interest (ROI) that is identified by the user in DICOM
5
images. The generated information consists of an estimated greatest long-axis and greatest short-axis dimensions, area, volume, and margin of the objects. For illustration purposes, DeepLook PRECISE can optionally provide a colorization of the interior area defined by a margin. DeepLook PRECISE does not make clinical decisions nor is a decision-support tool. The information provided by the software is an initial estimate and must not be used in isolation when making patient management decisions.
Lossy compressed mammoqraphic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using and FDA cleared monitor that offers at least 5 Mega-pixel resolutions and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained physicians and radiologists. DeepLook PRECISE, is not intended for use on mobile devices.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
| Ref
| Technical
Characteristic | Subject Device -
DeepLook
PRECISE | Predicate -
StoneChecker
K191530 | Difference/Comments |
|--------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------------|-------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Indications for Use | | | |
| 1 | Statement references
a software only
device. | Yes | Yes | No significant difference |
| 2 | Statement references
DICOM Standard. | Yes | Yes | No significant difference |
| 3 | Statement references
the
display / view of
medical images. | Yes | Yes | No significant difference |
| 4 | Statement references
the output of
information | Yes | Yes | No significant difference |
| 5 | Statement references
the user selection of
ROIs | Yes | Yes | No significant difference |
| 6 | Statement references
providing the clinician
with information
about an ROI. | Yes | Yes | No significant difference |
| 7 | Statement references
providing the user
with information
related to kidney
stones. | No | Yes | Yes, there is a difference. The
subject device can measure any
object within a ROI identified by the
User and is not intended
specifically for kidney stones. The
device Intended Use is clearly
detailed in the device labeling and
does not modify existing risks or
raise any new potential safety risks.
Therefore, we believe there is no
impact on safety or efficacy of the
subject device. |
| 8 | Intended Users | Trained
physicians,
radiologists. | Trained physicians,
radiologists. | No Difference |
| 9 | Target population | No restrictions | Patients diagnosed
with | Yes, there is a difference. There
are no restrictions with the subject device. |
| Ref
| Technical
Characteristic | Subject Device -
DeepLook
PRECISE | Predicate -
StoneChecker
K191530 | Difference/Comments |
| | | | kidney stones that
require medical
intervention. | device on the target population and
is not intended for a specific
population as detailed in the
labeling. The subject device is
applied to DICOM Images and not
specific target populations. Target
populations are the responsibility of
the modalities. This difference does
not modify existing risks or raise
any new potential safety risks.
Therefore, we believe there is no
impact on safety or efficacy of the
subject device. |
| 10 | Anatomical sites | No restrictions | Kidneys, ureters,
and bladder (KUB) | Yes, there is a difference. No
anatomical sites are specified in
the subject device labeling. This
does not modify existing risks or
raise any new potential safety risks.
Therefore, we believe there is no
impact on safety or efficacy of the
subject device. |
| 11 | Where used | No restrictions | No restrictions | No significant difference. |
| | Design | | | |
| 1 | Software device that
operates on off-the-
shelf hardware. | Yes | Yes | No significant difference. |
| 2 | Software device uses
a standard windowing
user interface. | Yes | Yes | No significant difference. |
| 3 | Software device uses
software algorithms
for image post
processing analysis. | Yes | Yes | No significant difference. The
subject device software is used for
measurement. |
| 4 | Conforms to DICOM
standards (PS 3.10) | Yes | Yes | No significant difference |
| | Features and Capabilities | | | |
| 5 | Data loading of CT
image series using
DICOM standard | Yes | Yes | DeepLook PRECISE can handle
any DICOM image and does not
restrictive to a specific modality. |
| 6 | 2D image review | Yes | Yes | No significant difference |
| 7 | Image navigation
tools (pan, zoom,
scroll, window/level) | No | Yes | DeepLook PRECISE functions are
measurement tools. Other image
management functions are left to
the PACS OEM software which is
not part of DeepLook PRECISE.
Therefore, it is our determination
that this difference does not have a
negative impact on safety or
efficacy and there are no new |
| Ref
| Technical
Characteristic | Subject Device -
DeepLook
PRECISE | Predicate -
StoneChecker
K191530 | Difference/Comments |
| 8 | Measurement tools
(ruler, ROI) | Yes | Yes | No significant difference |
| 9 | Size calculations
(area, volume) | Yes | Yes | No significant difference |
| 10 | Statistical
calculations | No | Mean, standard
deviation (SD),
mean of positive
pixels, skewness,
kurtosis, entropy | Yes, there is a difference.
DeepLook PRECISE functions are
measurement tool related. The
generated information consists of
an initial estimated greatest long-
axis and greatest short-axis
dimensions, area, volume, and
positions of objects. Other image
management functions are left to
the PACS OEM software which is
not part of DeepLook PRECISE.
This does not modify existing risks
or raise any new potential safety
risks. Therefore, we believe there is
no impact on safety or efficacy of
the subject device. |
| 11 | CT texture analysis
calculations | No | Yes | Yes. There is a difference.
DeepLook PRECISE functions are
measurement tools and is
integrated with an OEM PACS
system which is not part of
DeepLook PRECISE. The subject
device is applied to any DICOM
image and is not modality specific.
This is explained in the device
Manual. This difference does not
modify existing risks or raise any
new potential safety risks.
Therefore, we believe there is no
impact on safety or efficacy of the
subject device. |
| 12 | Display output of
measurements and
information. | Yes | Yes | No significant difference |
| 13 | Report generation | No | Excel | Yes, there is a difference. Reports
are not generated by DeepLook
PRECISE but are generated by the
OEM PACS application which is
not part of the subject device. This
difference does not modify existing
risks or raise any new potential
safety risks. Therefore, we believe
there is no impact on safety or
efficacy of the subject device. |
| Physical Characteristics | | | | |
| Ref
| Technical
Characteristic | Subject Device -
DeepLook
PRECISE | Predicate -
StoneChecker
K191530 | Difference/Comments |
| 14 | Post-processing
(non-real-time, non-
contacting, non-life
supporting and not
life sustaining). | Yes | Yes | No significant difference |
The following is a comparison of the subject device and the predicate device:
6
7
8
VII. PERFORMANCE DATA
Nonclinical Testing:
The subject device has been assessed and tested at the factory to assess all functions and has passed all predetermined testing criteria.
Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.
The following Standards were used to test the system and DeepLook PRECISE, has met the applicable requirements:
- ISO 14971 Third Edition 2019-12 Medical devices Application of risk management to . medical devices: FDA FR Recognition 5-125.
- . NEMA PS 3.1 - 3.20 (2016, Digital Imaging and Communications in Medicine (DICOM) Set, FDA FR Recognition # 12-300.
- . IEC 62304:2006/A1:2016, Medical device software - Software life cycle processes, FDA FR Recognition # 13-79.
- FDA Guidance on Cyber Security: Content of Premarket Submissions for Management of . Cybersecurity in Medical Devices Document Issued on: October 2, 2014.
- . FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005.
Conclusion:
The 510(k) Pre-Market Notification for DeepLook PRECISE, software device contains adequate information, data, and nonclinical test results to enable FDA -CDRH to determine substantial equivalence to the predicate device.
The subject device will be manufactured in accordance with the voluntary standards listed in the voluntary standard survey. The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
Nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
Therefore, DeepLook PRECISE, is substantially equivalent to the predicate device.