K Number
K201952

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-08-05

(22 days)

Product Code
Regulation Number
884.1175
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CrossGlide™ ETS Plus is indicated for use for sonohysterography and to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction in order to obtain tissue for histological biopsy.

Device Description

The CrossGlide™ ETS Plus is a sterile, disposable, single-use device which enables saline infusion sonohysteroraphy and the removal of mucosal tissue from the uterus for histological biopsy. The CrossGlide™ ETS Plus contains a Delivery Catheter with an everting Membrane, an Inner Aspiration Catheter that contains a lumen for saline infusion and aspiration. The everting Membrane places the Inner Aspiration Catheter into the uterine cavity. A proximal luer connector allows a 10cc luer lock syringe to infuse saline into the uterine cavity for sonohysterography. The 10cc syringe can be used to remove the saline from the uterine cavity and create negative pressure. Movement of the CrossGlide™ ETS Plus device, after creating negative pressure with the 10cc syringe, removes material and tissue from the uterine cavity.

AI/ML Overview

This document is a 510(k) summary for the CrossGlide™ ETS Plus, an endometrial suction curette and accessories. It focuses on demonstrating substantial equivalence to a predicate device, the CrossBay™ Endometrial Tissue Sampler (K192534), and references another device, the CrossBay SonoSure™ Sonohysterography and Endometrial Sampling Device (K133144).

The provided text does not include a study describing the acceptance criteria and performance of an AI-powered device. Instead, it discusses a medical device and its performance through various physical and functional tests, biocompatibility testing, sterilization validation, and packaging/shipping validation.

Therefore, I cannot fulfill your request to provide information about the acceptance criteria and study proving an AI device meets those criteria, or details regarding sample sizes, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance, as these items are not present in the provided document.

The document describes the following for the CrossGlide™ ETS Plus:

1. A table of acceptance criteria and the reported device performance:
There is no specific table of acceptance criteria for a new AI device's performance. Instead, the document discusses performance data for the CrossGlide™ ETS Plus physical device.

Test TypeDescription of Performance
Bench TestingConfirmed the device performs according to product specifications. Mechanical (tensile, pressurization) testing was conducted to determine if pinch clamps operate under expected functional conditions and if the 10cc syringe maintains adequate negative pressure for sample collection. The subject device successfully passed all functional bench testing (i.e., met predefined acceptance criteria).
BiocompatibilityUtilizes identical materials and manufacturing as the predicate device. Biocompatibility testing on the predicate device (according to ISO 10993-1 and FDA guidance) showed passing results for Cytotoxicity, Vaginal Irritation, and Sensitization.
SterilizationUtilizes the same product design and packaging as the predicate device. Evaluated for adoption into the predicate's ethylene oxide (EO) sterilization cycle (validated per ANSI/AAMI/ISO 11135:2014 via overkill method). Evaluation confirmed the subject device could be adopted into the previously validated cycle.
Packaging, Shipping, and Shelf-LifeUtilizes the same product design and packaging as the predicate device. Studies conducted per ASTM F88/F88M - 15, ASTM F 2096-11, and ASTM D4169-16 (Distribution Cycle 13). Shelf-life supported by packaging and performance tests on samples exposed to accelerated aging per ASTM F1980 - 16.

The remaining requested information (2-9 pertaining to AI studies) is not available in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 5, 2020

CrossBay Medical % Cindy Domecus Principal Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, CA 94010

Re: K201952

Trade/Device Name: CrossGlide™ ETS Plus Regulation Number: 21 CFR 884.1175 Regulation Name: Endometrial Suction Curette and Accessories Regulatory Class: II Product Code: HHK Dated: July 13, 2020 Received: July 14, 2020

Dear Cindy Domecus:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jason Roberts Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201952

Device Name CrossGlide™ ETS Plus

Indications for Use (Describe)

The CrossGlide™ ETS Plus is indicated for use for sonohysterography and to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction in order to obtain tissue for histological biopsy.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (21 CFR § 807.92(c))

I. SUBMITTER INFORMATION

Submitter:CrossBay Medical Inc.13240 Evening Creek Drive, Suite 304San Diego, CA 92128
Submission Correspondent:Cindy Domecus, R.A.C. (US & EU)Principal, Domecus Consulting Services LLCRegulatory Consultant to CrossBay Medical Inc.Phone: 650.343.4813Fax: 650.343.7822Email: domecusconsulting@comcast.net
Date Summary Prepared:August 3, 2020

II. SUBJECT DEVICE INFORMATION

Device Trade Name:CrossGlide™ ETS Plus
Common Name:Curette, Suction, Endometrial (And Accessories)
Regulation Number:21 CFR §884.1175
Regulation Name:Endometrial suction curette and accessories
Regulatory Class:II
Product Code:HHK

III. PREDICATE DEVICE INFORMATION

The predicate device is the CrossBay™ Endometrial Tissue Sampler, K192534. The reference device is the CrossBay SonoSure™ Sonohysterography and Endometrial Sampling Device, K133144.

The predicate device and the reference device have not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

The CrossGlide™ ETS Plus is a sterile, disposable, single-use device which enables saline infusion sonohysteroraphy and the removal of mucosal tissue from the uterus for histological biopsy. The CrossGlide™ ETS Plus contains a Delivery Catheter with an everting Membrane, an Inner Aspiration Catheter that contains a lumen for saline infusion and aspiration. The everting Membrane places the Inner Aspiration Catheter into the uterine cavity. A proximal luer

{4}------------------------------------------------

connector allows a 10cc luer lock syringe to infuse saline into the uterine cavity for sonohysterography. The 10cc syringe can be used to remove the saline from the uterine cavity and create negative pressure. Movement of the CrossGlide™ ETS Plus device, after creating negative pressure with the 10cc syringe, removes material and tissue from the uterine cavity.

V. INDICATIONS FOR USE

The CrossGlide™ ETS Plus is indicated for use for sonohysterography and to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction in order to obtain tissue for histological biopsy.

VI. COMPARISON OF INDICATIONS FOR USE WITH PREDICATE DEVICE

Table 1 below presents the Indications for Use for the subject and predicate devices, with the differences noted in bold font. The differences between the subject and predicate Indications for Use do not alter the intended use of the device. As a reference device is used in this submission, the Indications for Use for the reference device are also provided.

Subject DeviceCrossGlide™ ETS PlusPredicate Device,K192534CrossGlide™ ETSReference DeviceK133144CrossBay SonoSure™Sonohysterography andEndometrial Sampling DeviceSubject DeviceCrossGlide™ETS PlusPredicateDeviceK192534CrossGlide™ETSReference DeviceK133144CrossBaySonoSure™Sonohysterographyand EndometrialSampling DeviceComparison toPredicate Device
The CrossGlide™ ETS Plus isindicated for use forsonohysterography and toremove material from theuterus and from the mucosallining of the uterus by scrapingand vacuum suction in order toobtain tissue for histologicalbiopsy.The CrossGlide™ ETS,Endometrial TissueSampler, is indicated foruse to remove materialfrom the uterus and fromthe mucosal lining of theuterus by scraping andvacuum suction in orderto obtain tissue forhistological biopsy.The CrossBay SonoSure™Sonohysterography andEndometrial Sampling Device isindicated for use to access theuterine cavity for saline infusionsonohysterography and toobtain endometrial biopsy, ifindicated, utilizing the samedevice.DeviceComponentsInner catheter,evertingmembrane,aspiration withhole, blue andwhite pinchclamp,compliant tube,syringeInner catheter,evertingmembrane,aspiration withhole, blue andwhite stopcock,compliant tube,syringeDelivery catheter,cytology brush, salineinfusion bag, handle,acorn tipDifferent: Thesubject device hasthe same componentsas compared to thepredicate deviceexcept the blue andwhite stopcocks havebeen substituted byblue and white pinchclamps, whichperform the samefunction. Thisdifference does notraise differentquestions of safetyand effectiveness(S&E).
InflationMethodManual, with3cc syringeManual, with3cc syringeN/ASame
AspirationMethodManual, with10cc syringeManual, with3cc syringeN/ASame aspirationmethod usingdifferent size ofsyringe
OverallCatheterLength (cm)43-48 cm43-48 cm28 cmSame
InnerCatheterLength: 315mm(includesproximalconnector andpinch clamp)Outer diameter:2mmInternalLength: 315mm(includesStopcock)Outer diameter:2mmInternaldiameter: 1.3mmN/ASame

TABLE 1. COMPARISON OF INDICATIONS FOR USE

VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The technological characteristics of the subject and predicate device are compared below in Table 2.

{5}------------------------------------------------

{6}------------------------------------------------

Subject DeviceCrossGlide™ETS PlusPredicateDeviceK192534CrossGlide™ETSReference DeviceK133144CrossBaySonoSure™Sonohysterographyand EndometrialSampling DeviceComparison toPredicate Device
diameter: 1.3mm
DeliveryCatheterLength: 156mm(includes AcornTip)Outer diameter:4.1mmInternal diameter3.3.mmLength: 156mm(includes AcornTip)Outer diameter:4.1mmInternal diameter3.3.mmLength: 100mm(includes Acorn Tip)Same
EvertingMembraneLength: 65mm atfull deploymentOuter diameter:3.5mm @ 3atmospheres ofpressureWall thickness:0.04mmLength: 65mmat fulldeploymentOuter diameter:3.5mm @ 3atmospheres ofpressureWall thickness:0.04mmN/ASame
AspirationDevice SideHole &InternalLumenSide holeellipticalopeningInternal lumen1.3mmSide holeellipticalopeningInternal lumen1.3mmN/ASame
Acorn TipLength: 16mmOuter diameter:12.4mmLength: 16mmOuter diameter:12.4mmLength: 22.5mmOutside diameter:20.5mmSame
SterilizationMethodEthylene OxideEthylene OxideEthylene OxideSame
SalineInfusionMethodManual(syringe)NASameDifferent: Thepredicate device isnot indicated forsaline infusion. Assuch the reference
Subject DeviceCrossGlideTMETS PlusPredicateDeviceK192534CrossGlideTMETSReference DeviceK133144CrossBaySonoSureTMSonohysterographyand EndometrialSampling DeviceComparison toPredicate Device
device is included toprovide scientific andtechnical informationto address the safetyand effectiveness ofthe saline infusioncapability of thesubject device.

{7}------------------------------------------------

At a high level, the subject and predicate devices comprise the following same technological elements:

  • . Both the subject and predicate device use a delivery catheter to place the aspiration device into the uterine cavity to remove material.
  • Both the subject and predicate device employ a syringe to apply vacuum on the aspiration . device.
  • . Both the subject and predicate device use similar materials in the construction of the catheter.
  • . Both the subject device and predicate device utilize everting Membrane technology using the same materials, configuration, dimensions, and hydraulic principle for crossing the cervical canal.
  • . Both the subject device and the predicate device have the same sterilization method, packaging construction and materials.

The following technological differences exist between the subject and predicate device:

  • The subject device uses a 10cc syringe to apply vacuum suction to the Aspiration Device. The predicate device uses a 3cc syringe to apply vacuum suction.
  • . The subject device utilizes a Blue Pinch Clamp and a White Pinch Clamp as one-way valves (on/off). The predicate device utilizes a Blue Stopcock and White Stopcock as one-way valves in the same locations on the catheter. The stopcock and pinch clamp share the same function and performance characteristics.
  • . The subject device utilizes a 10cc syringe to infuse saline trough the catheter and into the uterine cavity.

{8}------------------------------------------------

The differences in technology do not raise different questions of safety and effectiveness, as both the subject and predicate device operate to transfer fluid/tissue through the catheter via manual pressure.

VIII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Bench Testing

Physical bench testing confirmed that the CrossGlide™ ETS Plus performs according to the product specifications. Device evaluation consisted of physical and functional testing performed pursuant to test protocols that were used to support clearance of the predicate or reference device. Mechanical testing, including tensile and pressurization testing, was conducted to determine if the pinch clamps could operate during expected functional conditions. Additional pressurization testing was conducted to determine if the use of a 10cc syringe can maintain adequate negative pressure required for sample collection. The subject device successfully passed all functional bench testing (i.e., met predefined acceptance criteria).

Biocompatibility testing

The subject device utilizes identical materials and manufacturing as the predicate device. Therefore, biocompatibility testing on the predicate device, which was conducted according to ISO 10993-1 "Biological Evaluation of Medical Devices" and FDA's guidance "Use of International Standard ISO 10993-1", final document issued on June 16, 2016 can be leveraged to support the biocompatibility of the subject device. Testing included the following with passing results: 1) Cytotoxicity, 2) Vaginal Irritation, and 3) Sensitization.

Sterilization Validation

The subject device utilizes the same product design and packaging as the predicate device. The predicate device was evaluated for adoption into the current ethylene oxide (EO) sterilization cycle of the contract sterilizer. The evaluation process was performed utilizing the AAMI Guidance contained in TIR28:2016 Product adoption and process equivalence for ethylene oxide sterilization. The EO Sterilization Cycle was originally validated in 2016, using the overkill method, in accordance with ANSI/AAMI/ISO 11135:2014. The results and conclusion of this evaluation confirmed that the subject device could be adopted into the previously validated cycle.

Packaging, Shipping Validation, and Shelf-Life

The subject device utilizes the same product design and packaging as the predicate device. Packaging and shipping validation studies were conducted pursuant to the applicable ASTM guidelines (ASTM F88/F88M - 15 "Standard Test Method for Seal Strength of Flexible Barrier

{9}------------------------------------------------

Materials"; and, ASTM F 2096-11 "Standard Test Methods for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble leaks)"; ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems, using Distribution Cycle 13. The shelf-life is supported by packaging and performance tests conducted on samples exposed to accelerated aging conditions pursuant to ASTM F1980 - 16 "Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices".

IX: CONCLUSIONS

The CrossGlide™ ETS Plus Sonohysterography and Endometrial Tissue Sampler has the same intended use as the predicate device. There are differences in the indications for use statements for the subject and predicate devices. The differences between the subject and predicate Indications for Use do not alter the intended use of the device. In addition, the subject device has the similar technological characteristics as the predicate device and the differences do not raise different questions of safety and effectiveness. Finally, the submitted testing demonstrates that the subject device is as safe, as effective, and performs as well as the predicate device. Therefore, the CrossGlide™ ETS Plus is substantially equivalent to the cleared predicate device.

§ 884.1175 Endometrial suction curette and accessories.

(a)
Identification. An endometrial suction curette is a device used to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction. This device is used to obtain tissue for biopsy or for menstrual extraction. This generic type of device may include catheters, syringes, and tissue filters or traps.(b)
Classification. Class II (performance standards).