(191 days)
Nanum Syringe is the plastic syringe for Single use, with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
A sterile device consisting of a calibrated barrel (cylinder) with plunger intended to be used for injection/withdrawal of fluids/gas (e.g., medication) to/from a medical device or the body (i.e., capable of both); a needle is not included. It is intended for various medical applications and is not dedication administration. At the distal end of the barrel is a male connector (typically Luer-lock/slip type) for the attachment to a hypodermic needle or an administration set. It is typically made of plastic and silicone materials and may have anti-stick plunger allowing smooth plunger movement, either manually or by a syringe pump. This is a single-use device.
Here's an analysis of the provided text, focusing on the acceptance criteria and the study used to demonstrate the device meets those criteria:
Device: Nanum Syringe (Piston Syringe)
Summary of Study Type: This documentation describes a bench testing and biocompatibility testing study conducted to demonstrate substantial equivalence of the Nanum Syringe to a legally marketed predicate device (K190002, Shanghai Kohope Medical Devices Co., Ltd. Sterile Hypodermic Syringe for Single use). The study's purpose is to show that the Nanum Syringe performs in a substantially equivalent manner in terms of functionality and biological safety.
1. Table of Acceptance Criteria and Reported Device Performance
| Test Item | Acceptance Criteria (Requirement) | Reported Device Performance (Result) | Details / Standard Referenced |
|---|---|---|---|
| Mechanical/Physical | |||
| Appearance and Structure | MFDA Notification No. 2018-72, The standards for the medical device | Pass | MFDA Notification No. 2018-72 |
| Nozzle | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | Pass | ISO 7886-1:2017 (Report: MTK-2019-000829) |
| Scale | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | Pass | ISO 7886-1:2017 |
| Piston/Plunger Assembly | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | Pass | ISO 7886-1:2017 |
| Tightness | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | Pass | ISO 7886-1:2017 |
| Extraction Test : pH | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | Pass | ISO 7886-1:2017 |
| Extraction Test : Heavy Metal | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | Pass | ISO 7886-1:2017 |
| Extraction Test : Potassium Permanganate-Reducible Substances | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | Pass | ISO 7886-1:2017 |
| Extraction Test : Residue on Evaporation | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | Pass | ISO 7886-1:2017 |
| Quantity of Silicone Oil | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | Pass | ISO 7886-1:2017 |
| Shelf Life | ASTM1980-07: 2011 Standard guide for Accelerated Aging of Sterile Barrier System for Medical Device & ISO11607-1:2006 Packaging for terminally sterilized medical devices | Pass | ASTM1980-07: 2011, ISO11607-1:2006 (Reports: NC-SLR-1902, NC-060-F0729, NC-060-F0421) |
| Biocompatibility | |||
| Cytotoxicity | ISO 10993-5(2009) Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | Pass | ISO 10993-5(2009) |
| Acute Systemic Toxicity Test | ISO 10993-11(2009) Biological evaluation of medical devices - Part 11: Tests for systemic toxicity | Pass | ISO 10993-11(2009) |
| Pyrogen Test | ISO 10993-11 Test for systemic toxicity, pyrogen test | Pass | ISO 10993-11 |
| Sensitization Test | ISO 10993-10 (2010) Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | Pass | ISO 10993-10 (2010) |
| Hemolytic Test | ISO 10993-4:2017, ASTM F756-17 Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood | Pass | ISO 10993-4:2017, ASTM F756-17 |
| Intracutaneous Reactivity Test | ISO 10993-10 (2010) Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | Pass | ISO 10993-10 (2010) |
| LAL Test (Endotoxin) | ISO 10993-11 Test for systemic toxicity, pyrogen test | Pass | ISO 10993-11 (Report: MSK-2020-002259) |
2. Sample Size Used for the Test Set and Data Provenance
The provided text does not explicitly state the sample sizes used for each individual test within the bench and biocompatibility studies. It only states that "Bench test were performed" and lists the tests.
- Sample Size (Test Set): Not explicitly stated for each test.
- Data Provenance: The tests are described as "Bench test were performed." This implies the data were generated in a laboratory setting, likely by Nanum Company Co., Ltd or a testing facility they contracted. The company is located in Daegu, Korea, suggesting the testing was performed there or outsourced. The studies are prospective in nature, as they involve newly manufactured devices being subjected to specific tests to determine their performance against established standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: This type of testing (bench and biocompatibility) generally relies on standardized protocols and objective measurements, not on human expert consensus to establish "ground truth" in the diagnostic sense. Therefore, the concept of "experts" establishing a ground truth as one would for image review or clinical outcomes is not applicable here.
- Qualifications of Experts: Not applicable. The results are typically interpreted by qualified laboratory personnel following the specified test standards.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. For bench and biocompatibility testing, results are typically objectively measured and compared against predefined pass/fail criteria outlined in the relevant ISO or ASTM standards. There is no mention of a subjective "adjudication" process akin to what would be used for clinical trial endpoints or image interpretation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, an MRMC comparative effectiveness study was not conducted. This is a medical device (syringe), not an AI-powered diagnostic or assistive technology. The evaluation focuses on the physical and biological properties of the syringe.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: No, a standalone performance evaluation (as defined for an algorithm) was not done. This device is a physical medical instrument, not a software algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device's evaluation is based on established international and national standards for medical device performance and biocompatibility.
- For mechanical/physical tests, the ground truth is defined by the objective pass/fail criteria specified in standards like ISO 7886-1:2017 and MFDA Notification No. 2018-72.
- For biocompatibility tests, the ground truth is defined by the objective pass/fail criteria specified in standards like ISO 10993 series and ASTM F756-17, which evaluate the biological response to the device materials.
8. The Sample Size for the Training Set
- Sample Size (Training Set): Not applicable. This is not an AI/machine learning device that requires a training set. The device is a physical product designed and manufactured to meet specific specifications.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable, as there is no training set for this type of device evaluation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 17, 2021
Nanum Company Co., Ltd % Peter Chung President Plus Global 300 Atwood St. Pittsburgh, Pennsylvania 15213
Re: K201930
Trade/Device Name: Nanum Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: June 24, 2020 Received: July 10, 2020
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K201930
Device Name Nanum Syringe
Indications for Use (Describe)
Nanum Syringe is the plastic syringe for Single use, with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
1. Applicant
- Company : NANUM COMPANY Co., Ltd 1)
- Address : 4th floor, 40, Siji-ro, Suseong-gu, Daegu, Korea 2)
- Tel : 82-53-795-7076 3)
-
- Fax : 82-53-795-7078
-
- Prepared date : June 24, 2020
- ୧) Contact person : Peter Chung, 412-512-8802
-
- Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
-
- Submission date : Nov. 20, 2020
2. Device Information
-
- Trade name : Nanum Syringe
-
- Common name : Piston Syringe
- Regulation name : Syringe, Piston 2)
-
- Product code : FMF
-
- Regulation number : 880.5860
- Class of device : Class II 5)
-
- Panel : General Hospital
3. The legally marketed device to which we are claiming equivalence
K190002, Shanghai Kohope Medical Devices Co., Ltd. / Sterile Hypodermic Syringe for Single use, with/without needle
4. Device description
A sterile device consisting of a calibrated barrel (cylinder) with plunger intended to be used for injection/withdrawal of fluids/gas (e.g., medication) to/from a medical device or the body (i.e., capable of both); a needle is not included. It is intended for various medical applications and is not dedication administration. At the distal end of the barrel is a male connector (typically Luer-lock/slip type) for the attachment to a hypodermic needle or an administration set. It is typically made of plastic and silicone materials and may have anti-stick plunger allowing smooth plunger movement, either manually or by a syringe pump. This is a single-use device.
5. Intended Use
Nanum Syringe is the plastic syringe for Single use, with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
6. Performance data:
-
- Bench test were performed. Bench testing included biocompatibility and mechanical testing. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.
| Test item | Requirements | Results |
|---|---|---|
| Appearance and Structure | MFDA Notification No. 2018-72, The standards for the medical device | |
| Nozzle | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringesfor manual use | Pass — MTK-2019-000829 |
| Scale | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringesfor manual use | |
| Piston/Plunger Assembly | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringesfor manual use |
{4}------------------------------------------------
| Tightness | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | |
|---|---|---|
| Extraction test : pH | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | |
| Extraction test : Heavy metal | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | |
| Extraction test : Potassiumpermanganate-reduciblesubstances | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | |
| Extraction test : Residue onevaporation | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | |
| Quantity of silicone oil | ISO 7886-1:2017 Sterile hypodermic syringes for single use — Part 1: Syringes for manual use | |
| ASTM1980-07: 2011 Standard guide for Accelerated Aging of SterileBarrier System for Medical DeviceISO11607-1:2006 Packaging for terminally sterilized medical devices | NC-SLR-1902 | |
| Shelf life test | Performance test of Real time and accelerated aged devices | NC-060-F0729NC-060-F0421 |
2) Biocompatibility
| # | Test item | Test method / Test criteria | Test result |
|---|---|---|---|
| 1 | Cytotoxicity | ISO 10993-5(2009) Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | Pass |
| 2 | Acute systemic toxicity test | ISO 10993-11(2009) Biological evaluation of medical devices - Part 11: Tests for systemic toxicity | Pass |
| 3 | Pyrogen test | ISO 10993-11 Test for systemic toxicity, pyrogen test | Pass |
| 4 | Sensitization test | ISO 10993-10 (2010) | Pass |
| 5 | Hemolytic test | ISO 10993-4:2017, ASTM F756-17 | Pass |
| 6 | Intracutaneous reactivity test | ISO 10993-10 (2010) | Pass |
| 7 | LAL Test | ISO 10993-11 Test for systemic toxicity, pyrogen test | PassMSK-2020-002259 |
The performance tests demonstrated that the subject device performs in a substantially equivalent manner to the predicate device.
7. Predicate device comparison table
| Manufacturer | NANUM COMPANY Co., Ltd | Shanghai Kohope Medical Devices Co., Ltd. | Remark |
|---|---|---|---|
| 510(k) No. | N/A | K190002 | N/A |
| Indication for use | Nanum Syringe is the plastic syringe forSingle use, with/without needle is intendedto be used for medical purposes to injectfluid into or withdraw fluid from body. | The sterile Hypodermic Syringe for Singleuse, with/without needle is intended to beused for medical purposes to inject fluid intoor withdraw fluid from body. | Similar |
| Classificationname | Syringe, Piston | Syringe, Piston | Same |
{5}------------------------------------------------
| Manufacturer | NANUM COMPANY Co., Ltd | Shanghai Kohope Medical Devices Co., Ltd. | Remark |
|---|---|---|---|
| Trade name | Nanum Syringe | Sterile Hypodermic Syringe for Single use,with/without needle, Sterile Insulin Syringefor Single use, with needle, SterileHypodermic needle for Single use | N/A |
| Model/type | NS-10, NS-20, NS-50 | N/A | N/A |
| Components | Cylinder, Plunger, Gasket, Clip | Barrel, Plunger, Piston | Similar |
| Nozzle type | Lock type, Lock clip type | Lock and Slip type | Similar |
| MaterialsBarrelPlungerPiston | PPPPRubber | PPPPRubber | Same |
| Capacity | 10mL, 20mL, 50mL | 1mL, 2mL, 3mL, 5mL, 10mL, 20mL, 30mL,35mL, 50mL, 60mL | Similar |
| Principle ofoperation | Manual | Manual | Same |
| Performances | Complies with ISO 7886-1 : 2017 Sterilehypodermic syringes for single use – Part 1 :Syringes for manual use | Complies with ISO 7886-1 : 2017 Sterilehypodermic syringes for single use - Part 1 :Syringes for manual use | Same |
| Biocompatibility | CytotoxicityAcute systemic toxicityPyrogenicitySensitizationHemolysisIntracutaneous testEndotoxin (LAL) test | CytotoxicityAcute systemic toxicityPyrogenicitySensitizationIrritationSubacute toxicity | Similar |
| Principle ofOperation | The plunger of syringe can be pulled andpushed along inside the barrel, allowing thesyringe to take in and expel the fluidsthrough the connector to the patient. | The plunger of syringe can be pulled andpushed along inside the barrel, allowing thesyringe to take in and expel the fluidsthrough the connector to the patient. | Same |
8. Conclusion
The device is investigated for function to compare the operation of function between Nanum Syringe and predicate devices.
Comparison results demonstrate that the specifications and performance of the device are substantially equivalent to the legally marketed predicate device.
Therefore, it is concluded that Nanum Syringe are substantially equivalent to the legally marketed predicate device.
§ 880.5860 Piston syringe.
(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).