K Number
K201867

Validate with FDA (Live)

Date Cleared
2020-11-19

(135 days)

Product Code
Regulation Number
876.5630
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Homechoice Claria APD System:

Baxter's Homechoice Claria APD system is intended for automatic control of dialysis solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis in the HOME HEALTHCARE ENVIRONMENT including comparable use in professional healthcare facilities.

Sharesource Connectivity Platform for use with the Homechoice Claria APD System:

The Sharesource portal is intended for use by healthcare professionals to remotely communicate new or modified treatment parameters with compatible dialysis instruments and transfer completed treatment database to aid in the review, analysis, and evaluation of patients' historical treatment results. This system is not intended to be a substitute for good clinical management practices, nor does its operation create decisions or treatment pathways.

Device Description

The Homechoice Claria APD System (Claria) is intended for automatic control of dialysis solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis. The system automatically cycles peritoneal dialysis solution in the amounts and at the times prescribed by a clinician familiar with and well informed about peritoneal dialysis. The clinician may use the Sharesource Connectivity Platform for Use with the Homechoice Claria APD System (Sharesource) to remotely communicate with the Claria cycler. Sharesource allows the transfer of treatment data originating from the treatment device to the clinician for review of historical treatment results. It also allows the clinician to adjust the device settings of the Claria cycler remotely. Changes to the device program by the clinician require the patient to review and accept the changes prior to the implementation of those changes. If the patient does not accept the changes, the Claria cycler will not accept the modified program. Sharesource does not include any real-time monitoring or real-time programming capabilities.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study information for the Baxter Homechoice Claria APD System and Sharesource Connectivity Platform:

It's important to note that the provided text is a 510(k) summary and not a detailed clinical study report. Therefore, some information, especially detailed performance metrics, sample sizes for specific tests, and expert qualifications, is either summarized at a high level or not explicitly stated in the document.


Acceptance Criteria and Study for Baxter Homechoice Claria APD System & Sharesource Connectivity Platform

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary does not explicitly present a table of numerical acceptance criteria with corresponding performance metrics. Instead, it describes a high-level approach to demonstrating substantial equivalence. The "acceptance criteria" are implied to be achieving equivalence to the predicate devices through successful performance testing and risk mitigation.

Acceptance Criteria (Implied from Summary)Reported Device Performance (Summary)
Meets user needs and intended useDesign validation confirms user needs and intended use are met, and the device is substantially equivalent to the predicate.
Electrical safetyElectrical safety testing was successfully completed for Claria.
Electromagnetic compatibility (EMC)Electromagnetic compatibility testing was successfully completed for Claria.
AIM testing (unspecified)AIM testing was successfully completed for Claria.
Risk assessment and controlHazard analyses confirm that Claria and Sharesource do not perform in an unexpected or unsafe manner; all potential risks were deemed acceptable after mitigation.
Labeling requirementsLabeling has been successfully implemented.
Software (including cybersecurity)Software including cybersecurity has been successfully implemented.
Human FactorsHuman Factors have been successfully implemented.
Functional performance equivalent to predicate devicesPerformance testing confirms that Homechoice Claria is as safe and effective as the HomeChoice/HomeChoice PRO cycler, and that the Sharesource Platform for Claria is as safe and effective as the Sharesource Platform for Amia. The differences do not affect intended use, fundamental technology, or operating principles, nor do they raise safety/effectiveness issues.

2. Sample Sizes and Data Provenance (Test Set)

  • Sample Size Used for the Test Set: Not explicitly stated in the provided summary. The summary refers to "performance testing" but does not give specific numbers of cases, cycles, or patient data points used for these tests.
  • Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not explicitly stated. The document refers to "design validation" and "performance testing," which typically involve internal testing and potentially external validation, but specific details about the type or origin of data used for these tests are not provided.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts:

This information is not provided in the summary. The nature of this device (a peritoneal dialysis system and connectivity platform) makes "ground truth" establishment less about expert interpretation of images or clinical findings, and more about functional verification and validation against requirements and predicate performance. Therefore, "experts" would likely be engineers, clinicians (involved in setting requirements), and quality assurance personnel, rather than ground truth adjudicators in the typical sense of diagnostic AI.

4. Adjudication Method for the Test Set:

This information is not provided and is generally not applicable in the context of device functional performance testing for a 510(k) submission. Adjudication methods like 2+1 or 3+1 are typically used in studies where clinical diagnosis or interpretation of data (e.g., images) requires independent expert review to establish a "ground truth" that a device's output is compared against. For a physical device and software platform, the assessment would be against pre-defined functional specifications, safety standards, and equivalence to a predicate.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this device. This type of study is relevant for diagnostic AI tools that assist human readers in tasks like image interpretation. The Baxter Homechoice Claria APD System and Sharesource Connectivity Platform are for managing peritoneal dialysis and remote monitoring/programming, not for diagnostic interpretation by human readers. The "AI" (Sharesource) here functions as a data transfer and remote programming tool, not as a diagnostic aid that enhances human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, in spirit, but not in the diagnostic AI sense. The performance testing described suggests that the system's functionalities (e.g., automatic solution exchanges, data transfer, remote programming capabilities) were validated as standalone functions according to their specifications. The safety and effectiveness claims are based on the device operating independently (the cycler itself) and the Sharesource platform facilitating data/program exchange, rather than its performance in conjunction with a human diagnostic interpreter. The Sharesource explicitly states it "is not intended to be a substitute for good clinical management practices, nor does its operation create decisions or treatment pathways," emphasizing that human clinical judgment remains central.

7. The Type of Ground Truth Used:

The "ground truth" for this type of device is established through:

  • Engineering specifications and requirements: The device functions according to its design.
  • Predicate device performance: The new device performs equivalently to the legally marketed predicate devices.
  • Risk analysis and mitigation: All identified risks are addressed, and the device operates safely.
  • Compliance with standards: Adherence to electrical safety, EMC, software, and human factors standards.
  • Verification and Validation (V&V) testing: Demonstrating that the device meets its design inputs and user needs.

This is a functional and safety "ground truth" derived from design control processes, rather than a clinical ground truth like pathology or outcome data typically used for diagnostic or predictive AI.

8. The Sample Size for the Training Set:

This information is not provided and is not applicable in the typical sense. This device is a medical device system (hardware and software) and a connectivity platform, not a machine learning or AI model that undergoes a "training phase" with a specific dataset. The "software" mentioned would likely be traditional embedded software and web-based applications, developed through standard software engineering practices (coding, testing, debugging), rather than being "trained" on a dataset like a deep learning model.

9. How the Ground Truth for the Training Set was Established:

This information is not provided and is not applicable for the reasons stated above (not an AI model with a training set). The "ground truth" for developing the software and hardware would be the design requirements, functional specifications, and relevant regulatory standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 19, 2020

Baxter Healthcare Corporation Brad Roynon Sr. Manager, Regulatory Affairs 25212 W. Illinois Route 120 Round Lake, IL 60073

Re: K201867

Trade/Device Name: Homechoice Claria Automated Peritoneal Dialysis (APD) System, Sharesource Connectivity Platform for Use with the Homechoice Claria APD System Regulation Number: 21 CFR 876.5630 Regulation Name: Peritoneal dialysis system and accessories Regulatory Class: Class II Product Code: FKX Dated: October 8, 2020 Received: October 19, 2020

Dear Brad Roynon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K201867

Device Name Homechoice Claria APD System &

Sharesource Connectivity Platform for use with the Homechoice Claria APD System

Indications for Use (Describe)

Homechoice Claria APD System:

Baxter's Homechoice Claria APD system is intended for automatic control of dialysis solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis in the HOME HEALTHCARE ENVIRONMENT including comparable use in professional healthcare facilities.

Sharesource Connectivity Platform for use with the Homechoice Claria APD System:

The Sharesource portal is intended for use by healthcare professionals to remotely communicate new or modified treatment parameters with compatible dialysis instruments and transfer completed treatment database to aid in the review, analysis, and evaluation of patients' historical treatment results. This system is not intended to be a substitute for good clinical management practices, nor does its operation create decisions or treatment pathways.

Type of Use (Select one or both, as applicable)
and the comments of the comments of the comments of the first of the first of

× | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right.

Section 5. 510(k) Summary

DATE:

October 14, 2020

OWNER:

Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015

CONTACT PERSON:

Brad Roynon Sr. Manager, Global Regulatory Affairs 25212 W. Illinois Route 120 Round Lake, IL 60073 Telephone: (262) 581-7157 Fax: (224) 270-4119

IDENTIFICATION OF THE DEVICE:

Common Name: Automated Peritoneal Dialysis System Trade/Device Name: Homechoice Claria Automated Peritoneal Dialysis (APD) SystemClassification Panel: 78 Gastroenterology and Urology Regulation Number: 21 CFR 876.5630 Regulation Name: Peritoneal dialysis sytem and accessories Regulatory Class: Class II Product Code: FKX

Common Name: Peritoneal Dialysis Software Trade/Device Name: Sharesource Connectivity Platform for Use with Homechoice Claria APD SystemClassification Panel: 78 Gastroenterology and Urology Regulation Number: 21 CFR 876.5630 Regulation Name: Peritoneal dialysis sytem and accessories Regulatory Class: Class II Product Code: FKX

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is the only element in the image.

Product CodeName
5C6M40Homechoice Claria APD System
5CGM01Sharesource Connectivity Platform for Use withthe Homechoice Claria APD System

Table 1. Product Codes in this Submission

PREDICATE DEVICE:

The Homechoice Claria APD System is substantially equivalent to the HomeChoice/HomeChoice PRO Automated Peritoneal Dialysis personal cyclers, which were previously cleared under 510(k) premarket notification K102936. The Sharesource Connectivity Platform for Use with the Homechoice Claria APD System is substantially equivalent to the Sharesource Connectivity Platform for Use with the AMIA APD System, which was previously cleared under 510(k) premarket notification K151525.

DESCRIPTION OF THE DEVICE:

The Homechoice Claria APD System (Claria) is intended for automatic control of dialysis solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis. The system automatically cycles peritoneal dialysis solution in the amounts and at the times prescribed by a clinician familiar with and well informed about peritoneal dialysis. The clinician may use the Sharesource Connectivity Platform for Use with the Homechoice Claria APD System (Sharesource) to remotely communicate with the Claria cycler. Sharesource allows the transfer of treatment data originating from the treatment device to the clinician for review of historical treatment results. It also allows the clinician to adjust the device settings of the Claria cycler remotely. Changes to the device program by the clinician require the patient to review and accept the changes prior to the implementation of those changes. If the patient does not accept the changes, the Claria cycler will not accept the modified program. Sharesource does not include any real-time monitoring or real-time programming capabilities. A more detailed device description is included in Section 10 of this submission.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are closely spaced together. The word is slightly slanted to the right, giving it a dynamic appearance.

INDICATIONS FOR USE: Homechoice Claria APD System

Baxter's Homechoice Claria APD System is intended for automatic control of dialysis solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis in the HOME HEALTHCARE ENVIRONMENT including comparable use in professional healthcare facilities.

Sharesource Connectivity Platform for Use with the Homechoice Claria APD System

The Sharesource portal is intended for use by healthcare professionals to remotely communicate new or modified treatment parameters with compatible dialysis instruments and transfer completed treatment data to a central database to aid in the review, analysis, and evaluation of patients' historical treatment results. This system is not intended to be a substitute for good clinical management practices, nor does its operation create decisions or treatment pathways.

TECHNOLOGICAL CHARACTERISTICS

The Homechoice Claria APD System has similar technological characteristics as compared to its' predicate device. The Sharesource Connectivity Platform for Use with the Homechoice Claria APD System has similar technological characteristics as compared to its' predicate device. Risk analysis has been completed and potential hazards associated with the differences in the proposed devices from their predicates have been identified and mitigated. All potential risks were deemed acceptable after mitigation.

DISCUSSION OF NONCLINICAL TESTS:

Performance testing was conducted on the Homechoice Claria APD System with Sharesource to evaluate the functional performance of the device. The performance testing confirms that Homechoice Claria is as safe and effective as the HomeChoice/HomeChoice PRO cycler, and that the Sharesource Platform for Claria is as safe and effective as the Sharesource Platform for Amia.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly slanted to the right, giving it a dynamic appearance. The word is the only element in the image, and it is centered.

In summary, Claria and Sharesource have successfully implemented performance requirements and subsequent outputs verifying and validating:

  • . The design validation meets the user needs and intended use and is substantially equivalent to the predicate.
  • . Electrical safety, electromagnetic compatibility, and AIM testing was successfully completed for Claria.
  • Risk assessment and risk control measures. Hazard analyses confirm that Claria and Sharesource do not perform in an unexpected or unsafe manner.
  • . Labeling, software including cybersecurity, and Human Factors have been successfully implemented.

CONCLUSION:

Based on demonstrable evidence, the differences between the proposed devices and their predicates described within this submission do not affect the intended use, the fundamental technology or the operating principles of the device, nor do any changes raise safety or effectiveness issues with regard to the Homechoice Claria APD System and the Sharesource Connectivity Platform for Use with the Homechoice Claria APD System.

§ 876.5630 Peritoneal dialysis system and accessories.

(a)
Identification. (1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments.
(3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles.
(4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
(b)
Classification. Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.