K Number
K201648
Device Name
Sterispine™ PS
Manufacturer
Date Cleared
2020-07-30

(42 days)

Product Code
Regulation Number
888.3070
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterispine™ PS system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct of fusion. Sterispine™ PS system is intended for posterior non-cervical, pedicle and nonpedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation) ; spinal stenosis ; tumor ; pseudoarthrosis and failed previous fusion.

Device Description

The cleared range of Sterispine™ PS system includes multiaxial screws and cannulated multiaxial screws with or without extended head (Ø5.5, 6.5 and 7.5mm, lengths from 25 to 85mm) and straight and prebent rods (Ø5.5, 6.5 and 7.5mm, lengths from 30 to 380mm). Sterispine™ PS implants are supplied sterile with a sterile single-use set of surgical instruments supplied in kits. Bacterial endotoxin testing as specified in USP standard is used for pyrogenicity testing to achieve the Endotoxin limit of 20 EU / device.

AI/ML Overview

Here's an analysis of the provided text regarding acceptance criteria and device performance.

Important Note: The provided document is a 510(k) summary for a medical device (Sterispine™ PS) and focuses on demonstrating substantial equivalence to a predicate device. It does not present a study proving an AI device meets acceptance criteria. The document pertains to a traditional orthopedic implant system, not an AI-powered diagnostic or therapeutic device. Therefore, many of the requested elements for an AI device (like sample size for test/training sets, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable to this document.

The document primarily discusses non-clinical mechanical testing of a physical implant (pedicle screw system).


Based on the provided text, here's what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

The document mentions specific non-clinical mechanical tests and states the general outcome. It does not provide a quantitative table of specific acceptance criteria values and corresponding performance results.

Acceptance Criteria CategoryReported Device Performance (Summary)
Mechanical StrengthMeets or exceeds functional requirements and suitability for use.
Performs equivalent or better than predicates systems.
Cross-connector & Surgical Instrument FunctionalityAll tests met the acceptance criteria.
Single-use Surgical Instruments FunctionalityNew added surgical instruments can be assembled with each component of the Sterispine PS instrument/implant kit.
No deterioration or blocking observed during functional testing.
Suitable and fulfill requirements for single-use.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified quantitatively. The tests are mechanical in nature, likely on a set number of implants/components per test (e.g., n=5 or n=10 per ASTM standard).
  • Data Provenance: Not explicitly stated regarding country of origin or retrospective/prospective. This refers to lab testing of physical devices, not clinical data from patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not applicable. This document describes mechanical device testing against engineering standards, not clinical diagnosis or AI model evaluation requiring human expert ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable. Adjudication methods are relevant for subjective human interpretations, typically in clinical studies or AI model ground truth establishment. This is mechanical testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This relates to AI-assisted human performance in diagnostic tasks. The device in question is a surgical implant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This relates to the performance of an AI algorithm as a standalone diagnostic tool. The device is a surgical implant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for this device's performance is adherence to established mechanical testing standards (e.g., ASTM F1798, ASTM F1717) and functional engineering requirements. It's engineering/mechanical performance standards, not clinical ground truth.

8. The sample size for the training set

  • Not applicable. This refers to AI model training data.

9. How the ground truth for the training set was established

  • Not applicable. This refers to AI model training data.

Summary regarding the provided document:

The provided K201648 510(k) summary is for a traditional surgical implant (pedicle screw system). The "acceptance criteria" and "study" described are focused on non-clinical mechanical and functional testing to demonstrate the device's physical properties and performance against established engineering standards, and its substantial equivalence to previously cleared predicate devices. It does not involve AI technology or clinical performance studies in the way envisioned by the posed questions.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure embracing a globe. The FDA acronym is in a blue square, and the full name, "U.S. Food & Drug Administration," is in blue text to the right of the square.

Safe Orthopaedics Pierre Dumouchel CEO Parc des Bellevues, Le Californie - Allee Rosa Luxemburg Eragny Sur Oise, 95610 France

Re: K201648

Trade/Device Name: Sterispine™ PS Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWQ Dated: June 16, 2020 Received: June 18, 2020

Dear Pierre Dumouchel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

July 30, 2020

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Colin O'Neill, M.B.E. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201648

Device Name Sterispine™ PS

Indications for Use (Describe)

The Sterispine™ PS system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct of fusion. Sterispine™ PS system is intended for posterior non-cervical, pedicle and nonpedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation) ; spinal stenosis ; tumor ; pseudoarthrosis and failed previous fusion.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Traditional 510k STERISPINE™ PS

Image /page/3/Picture/2 description: The image shows the logo for Safe Orthopaedics. The logo features a stylized letter "S" in a light green color, with a small dot to the right of the "S". Below the "S", the words "Safe Orthopaedics" are written, with "Safe" in green and "Orthopaedics" in gray.

510(k) SUMMARY

510kTraditional
Basis for submissionExtension of the product line
Submitted bySafe OrthopaedicsParc des BellevuesLe Californie - Allée Rosa Luxemburg95610 Eragny-sur-Oise, France
ContactsPierre DUMOUCHEL (CEO)Phone number 33 (0) 1 84 28 01 79e-mail p.dumouchel@safeorthopaedics.comRegulatory contact: Dr Isabelle DRUBAIX (Idée Consulting) idrubaix@nordnet.fr
Date PreparedJuly 23, 2020
Common NamePedicle screw spinal system
Trade NameSterispine™ PS
Classification NameThoracolumbosacral pedicle screw system
ClassII
Product CodeNKB, KWQ
CFR section888.3070
Device panelORTHOPEDIC
Legally marketedpredicate devicesPrimary predicate: Sterispine™ PS manufactured by SAFE ORTHOPAEDICS (K170528)Additional predicates: CD HORIZON™ Fenestrated Screw Set manufactured byMedtronic Sofamor Danek (K170347), Synergy 4.0 mm, 4.5 mm VLS screwsmanufactured by Interpore Cross Intl (K011437), Sterispine™ Ps manufactured bySAFE ORTHOPAEDICS (K151921, K151747, K150092, K140802, K130632, K121299,K112453)

{4}------------------------------------------------

Indications for useThe Sterispine™ PS system is intended to provide immobilization and stabilizationof spinal segments in skeletally mature patients as an adjunct of fusion.Sterispine™ PS system is intended for posterior non-cervical, pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD)(defined as back pain of discogenic origin with degeneration of the disc confirmedby history and radiographic studies); spondylolisthesis; trauma (i.e, fracture ordislocation) ; spinal stenosis ; tumor ; pseudoarthrosis and failed previous fusion.
Description of thedeviceThe cleared range of Sterispine™ PS system includes multiaxial screws andcannulated multiaxial screws with or without extended head (Ø5.5, 6.5 and7.5mm, lengths from 25 to 85mm) and straight and prebent rods (Ø5.5, 6.5 and7.5mm, lengths from 30 to 380mm). Sterispine™ PS implants are supplied sterilewith a sterile single-use set of surgical instruments supplied in kits. Bacterialendotoxin testing as specified in USP standard is used for pyrogenicity testing toachieve the Endotoxin limit of 20 EU / device.
Technologicalcharacteristicscompared to thepredicate devicesThe present submission is an extension of the product line that includes- Modified and new sterile single-use surgical instruments- Add of a cross-connector kit- Add of multi-axial fenestrated screws (with and without extended head)- Add of Ø4.5 mm for all STERISPINE™ PS screwsAs established in this submission, the modified or added components aresubstantially equivalent to predicate devices in areas including indications for use,function materials in contact with patients and technological and mechanicalcharacteristics.The safety and effectiveness of Sterispine™ PS fenestrated screws have not beenestablished when used in conjunction with bone cement or for use in patients withpoor bone quality (e.g., osteoporosis, osteopenia). This device is intended only tobe used with saline or radiopaque dye.
Discussion of TestingThe following non-clinical tests were conducted on the STERISPINE™ PS systemMechanical testing of fenestrated screw φ4.5 mm / with cross -connector whereapplicable):- Axial Grip, Flexural Grip, Torsional Grip and Dynamic Flexion per ASTM F1798- Static Compression Bending, Static Tension Bending, Static Torsion andDynamic compression per ASTM F1717.Results demonstrate that the Sterispine™ PS system meets or exceeds functionalrequirements and suitability for use. Additionally, the Sterispine™ PS systemperforms equivalent or better than predicates systems.- Functional evaluation of the cross-connector and associated surgicalinstrumentationAll tests met the acceptance criteria- Functional testing of the single-use surgical instruments

{5}------------------------------------------------

Overall functional testing demonstrates that the new added surgical instruments can be assembled with each component of the Sterispine PS instrument/implant kit. No deterioration or blocking were observed during the functional testing. Additional mechanical and / or functional testings conducted on Jamshidi, threaded rod introducer, spatula, screw extender arm, fenestrated screwdriver, persuader and handle show that surgical instruments are suitable and fulfill requirements for single-use.
ConclusionBased on the design features, technological characteristics, feature comparisons, indications for use, and non-clinical performance testing, the added surgical implants and the added or modified sterile single-use surgical instruments have demonstrated substantial equivalence to the identified predicate devices.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.