The SteriSpine" PS system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion.
SteriSpine"PS System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion.
system includes Pedicle Screws and Rods. Components SteriSpine™PS of SteriSpine™PS system are made of Titanium Ta6V Eli grade conforming to ASTM F136. SteriSpine™PS components are supplied sterile with a single-use set of surgical instruments. Components added within this submission include multi-axial pedicle screws with extended head and associated instruments.
This document describes a 510(k) premarket notification for a medical device, the "SteriSpine™ PS (Pedicle Screw) System." It is a class III device, which means it carries a higher risk than class I or II devices. The submission concerns a modification to an existing device (K112453 and K121299). The information provided is not for an AI/ML powered device, but rather a traditional medical device (pedicle screw spinal system). Therefore, several of the requested sections related to AI/ML device performance criteria, data provenance, expert ground truthing, and MRMC studies are not applicable.
Here's the breakdown of the available information:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Standards) | Reported Device Performance |
|---|---|
| Conforms to special control established for Pedicle screw spinal system | Met |
| Conforms to «Spinal System 510(k)s - Guidance for Industry and FDA Staff Document» issued on: May 3, 2004 | Met |
| Mechanical testing per ASTM F1717 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model): Static Compression | Results demonstrate comparable mechanical properties to the predicate device. |
| Mechanical testing per ASTM F1717: Static Torsion | Results demonstrate comparable mechanical properties to the predicate device. |
| Mechanical testing per ASTM F1717: Dynamic Compression | Results demonstrate comparable mechanical properties to the predicate device. |
| Mechanical testing per ASTM F1798 (Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants): Static slipping | Results demonstrate comparable mechanical properties to the predicate device. |
| Mechanical testing per ASTM F1798: Static bending | Results demonstrate comparable mechanical properties to the predicate device. |
| Mechanical testing per ASTM F1798: Static rotation | Results demonstrate comparable mechanical properties to the predicate device. |
| Validation of instrumentation via Cadaver testing | Performed and presented |
| Verification Activity and Validation Activity for added components | Demonstrated that added components are as safe, effective, and perform at least as safely and effectively as predicate device. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not specified for mechanical or cadaver testing. For the clinical data, it states "Clinical data from a review of the literature has been presented in the class III summary," but no specific sample size from this literature review is provided in the extract.
- Data Provenance:
- Mechanical Testing: Not specified (country of origin). This is typically performed in a lab setting.
- Cadaver Testing: Not specified (country of origin).
- Clinical Data: "a review of the literature" implies retrospective data, but the specific origins are not detailed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is a traditional mechanical device, not one that relies on expert interpretation for ground truth. The "ground truth" here is adherence to mechanical standards and performance comparable to a predicate device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a study requiring adjudication of expert interpretations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This is not an AI-powered device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm. The performance is the mechanical integrity of the device itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance is established through:
- Mechanical Standards: Adherence to ASTM F1717 and ASTM F1798 standards.
- Predicate Device Performance: Direct comparison of mechanical properties to the legally marketed predicate devices (SteriSpine™ PS K112453 and K121299).
- Cadaver Testing: Validation of instrumentation.
8. The sample size for the training set
Not applicable. This device does not use a "training set" in the context of machine learning.
9. How the ground truth for the training set was established
Not applicable. There is no training set for this type of device.
{0}------------------------------------------------
K130632 :: page 1 of 2
MAY 0 3 2013
510(k) SUMMARY
| Submitter | Safe OrthopaedicsParc des BellevuesAllée R. Luxembourg - Le Californie95610 Eragny sur Oise - France | |
|---|---|---|
| Contacts | Pierre DUMOUCHEL p.dumouchel@safeorthopaedics.comRegulatory contact :Isabelle Drubaix idee-consulting@nordnet.fr+33 (0) 3 21 05 64 23 | |
| Trade Name | Sterispine™ PS Pedicle Screw | |
| Common Name | Pedicle screw spinal system | |
| Classification Name | ||
| Product code | MNI, MNH, KWP, NKB | |
| CFR section | 888.3070 | |
| Class | III | |
| Classification Panel | Orthopedic | |
| Legally marketedpredicate device | Sterispine™ PS Pedicle Screw K112453 and K121299Manufactured by SAFE ORTHOPAEDICS | |
| SPECIAL 510k | Modification to Sterispine™ PS Pedicle Screw system | |
| e-copy Statement | The eCopy is an exact duplicate of the paper copy |
Description :
system includes Pedicle Screws and Rods. Components SteriSpine™PS of SteriSpine™PS system are made of Titanium Ta6V Eli grade conforming to ASTM F136. SteriSpine™PS components are supplied sterile with a single-use set of surgical instruments. Components added within this submission include multi-axial pedicle screws with extended head and associated instruments.
Indications for use:
The SteriSpine" PS system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion.
SteriSpine" PS System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion.
Performance data:
SteriSpine™PS conforms to special control established for Pedicle screw spinal system and to « Spinal System 510(k)s - Guidance for Industry and FDA Staff Document » issued on: May 3, 2004.
Mechanical testing was conducted per ASTM F1717 Standard Test Methods for Spinal Implant Constructs in a Vertebrectorny Model and ASTM F1798 Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants. Testing according to ASTM F1717 includes Static Compression, Static Torsion and Dynamic Compression. Testing according to ASTM F1798 includes Static slipping, Static bending and Static rotation.
{1}------------------------------------------------
K130632 :: page 2 of 2
Results demonstrate comparable mechanical properties to the predicate device. Cadayer testing performed to validate the instrumentation have been presented.
No additional testing has been performed for the added components.
Clinical data from a review of the literature has been presented in the class III summary.
Substantial equivalence:
The extended range of SteriSpine"PS system is substantially equivalent to its predicate devices SteriSpine™PS system (K112453, K121299) in terms of intended use, material, design, mechanical properties and function.
Verification Activity and Validation Activity demonstrate that components added to SteriSpine"PS system are as safe, as effective, and performs at least as safely and effectively as predicate SteriSpine™™PS system.
2013, February 28th
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three curved lines representing its wings or feathers. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 3, 2013
Safe Orthopaedics % Mr. Pierre Dumouchel Parc des Bellevues Allée R. Luxembourg - Le Californie 95610 Eragny sur Oise - FRANCE
Re: K130632
Trade/Device Name: Sterispine" PS (Pedicle Screw) System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWP Dated: April 3, 2013 Received: April 5, 2013
Dear Mr. Dumouchel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898: In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Pierre Dumouchel
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Erin DKeith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE
510(k) Number (if known): Device Name: SteriSpine "PS Indications for Use:
The SteriSpine" PS system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion.
SteriSpine"PS System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion.
| Prescription Use ✓ | |
|---|---|
| (Part 21 CFR 801 Subpart D) | AND/OR |
| Over-The-Counter Use | |
| (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ronald#. Jean -S
(Division Sign-Off)
Division of Orthopedic Devices
510(k) Number: K130632