K Number
K201594

Validate with FDA (Live)

Date Cleared
2020-12-17

(188 days)

Product Code
Regulation Number
878.4810
Age Range
18 - 65
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Intended Use: PRIMELASE Excellence is intended for use in dermatologic and general surgical procedures.

Indications for Use:

PRIMELASE Excellence System with 810nm Laser applicator is intended for:

· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

  • · Treatment of Pseudofolliculitis barbae (PFB).
  • · Use on all skin types (Fitzpatrick I-VI).
  • · Treatment of benign pigmented lesions

PRIMELASE Excellence System with 755nm Laser applicator is intended for:

· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

  • · Treatment of Pseudofolliculitis barbae (PFB).
  • · Use on all skin types (Fitzpatrick I-VI).
  • · Treatment of benign pigmented lesions

PRIMELASE Excellence System with 1060nm Laser applicator is intended for:

· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

  • Treatment of Pseudofolliculitis barbae (PFB).
  • · Use on all skin types (Fitzpatrick I-VI).

· Treatment of benign vascular lesions, including angiomas, telangiectasia, and other benign vascular lesions and leg veins.

  • Treatment of wrinkles
    PRIMELASE Excellence System with 810-1060nm Laser applicator is intended for:

· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

  • Treatment of Pseudofolliculitis barbae (PFB).
    · Use on all skin types (Fitzpatrick I-VI).
Device Description

PRIMELASE Excellence is a diode laser device which consists of a central unit and external applicators.

The PRIMELASE equipment emits laser radiation (beam infrared light with a wavelength range of 755 nm to 1060 nm, typically and most used 810 nm.), pulsed through the laser aperture situated at the tip of the applicator.

The device applicator contains the diode which emits the laser energy whereas the power delivered, and the working frequency being controlled by the machine's central unit.

The emission of energy is activated in the form of continuous pulses when pressing the applicator button. The applicator sapphire tip is cooled to a constant temperature to cool the skin, so that it partially anaesthetizes the tissue reducing the risk of damage to the epidermis during treatment.

AI/ML Overview

The provided text does not contain information about acceptance criteria and a study that proves the device meets them in the format requested.

The document is a 510(k) premarket notification for the "Primelase Excellence" laser device. It focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than presenting acceptance criteria and a study proving their fulfillment as typically seen for novel devices or significant modifications.

Here's a breakdown of why the requested information is not available in the provided text:

  • Acceptance Criteria and Reported Device Performance Table: The document provides a "Comparison of Technological Characteristics" table (Table 2) comparing the subject device to predicate devices. This table lists technical specifications (e.g., wavelength, spot size, fluence) but does not present specific acceptance criteria (e.g., a minimum percentage of hair reduction) or reported device performance against such criteria. Instead, it aims to show the subject device's specifications are comparable to those of already cleared devices.

  • Sample Size, Data Provenance, Expert Qualifications, Adjudication Method, MRMC Study, Standalone Performance, Ground Truth Type, Training Set Sample Size, and Training Set Ground Truth: All these items are related to a clinical study and its methodology. The document explicitly states: "No clinical study is included in this submission." Therefore, none of this information is present. The substantial equivalence argument relies on the similarity of the device's technological characteristics and intended use to predicate devices for which efficacy and safety were established.

In summary, the document demonstrates substantial equivalence through technical comparisons and adherence to recognized standards for safety and performance (non-clinical testing), rather than through a dedicated clinical study with specific acceptance criteria and performance metrics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

December 17, 2020

High Technology Products SLU % Vardhini Kirthivas Vice President - Regulatory Services Freyr Global Regulatory Solutions & Services Level 4 Building No. H-08 Phoenix SEZ Phase 2 Gachibowli, Hyderabad, Telangana 500081 India

Re: K201594

Trade/Device Name: Primelase Excellence Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 8, 2020 Received: June 12, 2020

Dear Vardhini Kirthivas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201594

Device Name Primelase Excellence

Indications for Use (Describe) Intended Use: PRIMELASE Excellence is intended for use in dermatologic and general surgical procedures.

Indications for Use:

PRIMELASE Excellence System with 810nm Laser applicator is intended for:

· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

  • · Treatment of Pseudofolliculitis barbae (PFB).
  • · Use on all skin types (Fitzpatrick I-VI).
  • · Treatment of benign pigmented lesions

PRIMELASE Excellence System with 755nm Laser applicator is intended for:

· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

  • · Treatment of Pseudofolliculitis barbae (PFB).
  • · Use on all skin types (Fitzpatrick I-VI).
  • · Treatment of benign pigmented lesions

PRIMELASE Excellence System with 1060nm Laser applicator is intended for:

· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

  • Treatment of Pseudofolliculitis barbae (PFB).
  • · Use on all skin types (Fitzpatrick I-VI).

· Treatment of benign vascular lesions, including angiomas, telangiectasia, and other benign vascular lesions and leg veins.

  • Treatment of wrinkles
    PRIMELASE Excellence System with 810-1060nm Laser applicator is intended for:

· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

  • Treatment of Pseudofolliculitis barbae (PFB).
    · Use on all skin types (Fitzpatrick I-VI).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

High Technology Products SLU, Spain

Traditional 510(k) Primelase Excellence

5. 510(k) Summary

5.1. Submitter Information:Application Correspondent:Vardhini KirthivasFreyr Global Regulatory Solutions and ServicesLevel 4, Building No. H-08, Phoenix SEZ,Phase 2, Gachibowli, Hyderabad,Telangana, India, 500081
Phone:+91 - 9940445965
E-mail:vardhini.kirthivas@FreyrSolutions.com
Legal Manufacturer:High Technology Products, S.L.UPassatge Masoliver, 24-2808005 Barcelona, Catalonia Spain
Phone:+34 93 458 85 66
Contact Person:Mr. Sergi Lozano
E-mail:slozano@cocoonmedical.com
Date Prepared:14 December 2020

5.2. Device Identification:

Device Trade Name:Primelase Excellence
Device Common Name:PRIMELASE Excellence System
Classification Name:Powered Laser Surgical Instrument
Device Class:Class II
Regulation Number:21 CFR 878.4810
Product Code:GEX

5.3. Predicate Devices:

Device Name510(k) Number
Primelase ExcellenceK191321
The Modified Alma Lasers XL TM Familyof Multi-Application and Multi-Technology Platforms [SopranoXLSopranoXL and SopranoICE]K172193

{5}------------------------------------------------

Device Name510(k) Number
Lightsheer Desire; Lightsheer DesireLight; Lightsheer Duet; LightsheerInfinityK170179

5.4. Device Description

PRIMELASE Excellence is a diode laser device which consists of a central unit and external applicators.

The PRIMELASE equipment emits laser radiation (beam infrared light with a wavelength range of 755 nm to 1060 nm, typically and most used 810 nm.), pulsed through the laser aperture situated at the tip of the applicator.

The device applicator contains the diode which emits the laser energy whereas the power delivered, and the working frequency being controlled by the machine's central unit.

The emission of energy is activated in the form of continuous pulses when pressing the applicator button. The applicator sapphire tip is cooled to a constant temperature to cool the skin, so that it partially anaesthetizes the tissue reducing the risk of damage to the epidermis during treatment.

Intended Use & Indications for Use 5.5.

Intended Use

PRIMELASE Excellence is intended for use in dermatologic and general surgical procedures.

Indications for Use

PRIMELASE Excellence System with 810nm Laser applicator is intended for:

  • Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
  • Treatment of Pseudofolliculitis barbae (PFB).
  • Use on all skin types (Fitzpatrick I-VI). ●
  • . Treatment of benign pigmented lesions

PRIMELASE Excellence System with 755nm Laser applicator is intended for:

  • Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

{6}------------------------------------------------

  • Treatment of Pseudofolliculitis barbae (PFB).
  • o Use on all skin types (Fitzpatrick I-VI).
  • Treatment of benign pigmented lesions o

PRIMELASE Excellence System with 1060nm Laser applicator is intended for:

  • Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
  • Treatment of Pseudofolliculitis barbae (PFB). ●
  • Use on all skin types (Fitzpatrick I-VI). ●
  • Treatment of benign vascular lesions, including angiomas, hemangiomas, ● telangiectasia, and other benign vascular lesions and leg veins.
  • Treatment of wrinkles .

PRIMELASE Excellence System with 810-1060mm Laser applicator is intended for:

  • Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
  • Treatment of Pseudofolliculitis barbae (PFB). 0
  • Use on all skin types (Fitzpatrick I-VI). ●

5.6. Comparison of Technological Characteristics

The fundamental scientific technology, materials of construction and mechanism of operation are identical between the subject PRIMELASE equipment and the predicate devices. Table 2 summarizes the comparison of technological characteristics between the subject and predicate devices.

S.NoParametersSoprano ICE(K172193)LightsheerInfinity(K170179)PrimelaseExcellence(K191321)PrimelaseExcellence(Subjectdevice)
1.ManufacturerAlma Lasers,Ltd,Lumenis LtdHighTechnologyProductsS.L.U.HighTechnologyProducts,S.L.U.
2.Product CodeGEXGEXGEXGEX
3.RegulationNumber21 CFR878.481021 CFR878.481021 CFR878.481021 CFR878.4810
Table 2 - Substantial Equivalence Table

{7}------------------------------------------------

High Technology Products SLU, Spain

S.NoParametersSoprano ICE(K172193)LightsheerInfinity(K170179)PrimelaseExcellence(K191321)PrimelaseExcellence(Subjectdevice)
4.Principle ofOperationAlGaAsLaser diodearrayAlGaAsLaser diodearrayAlGaAs LaserdiodearrayAlGaAs Laserdiodearray
5.LaserClassificationClass IVClass IVClass IVClass IV
6.UserInterfaceLCD touchscreenLCD touchscreenLCD touchscreenLCD touchscreen
7.PulsingControlFinger switchFinger switchFinger switchFinger switch
8.ConfigurationMain unit,Handpieceand FootcontrolMain unitandHandpieceMain unit,Handpieceand Footcontrol(optional)Main unit,Handpieceand Footcontrol(optional)
9.LaserWavelength755 nm810 nm1064nm805 nm1060 nm755nm810nm810-1060nm755nm810nm1060nm810-1060nm
10.Spot Size(CM × CM)12x10,15x10,20x10,10x109x9, 27x9,22x3520x9, 30x9,30x1720x9, 30x9,30x1710x10
11.Laser ContactSapphirewindowPuresapphire,A1203Pure sapphire,A12O3Pure sapphire,A1203
12.FrequencyUp to 3Hz(HR)5 - 10Hz(SHR)Up to 3HzUp to 3Hz(static)5 — 10Hz(dynamic)Up to 3Hz(static)5 — 10Hz(dynamic)
13.Pulseduration3.3-200ms5-400ms3-400ms3-400ms
14.Fluence120J/cm²100J/cm²80J/cm²120J/cm²
15.TissuecoolingContactcontinuous,Thermo-electricalChilltipcontactcoolingContactcontinuous,Thermo-electricalContactcontinuous,Thermo-electrical
16.CoolingTemperature4 °C2 °C -12 °C5 °C5 °C
17.Power supply120VAC,11A, 50/60100-240VAC +/-10%Single phase,100-240V50-60 HzSingle phase,100-240V50-60 Hz
S.NoParametersSopranoICE(K172193)LightsheerInfinity(K170179)PrimelaseExcellence(K191321)PrimelaseExcellence(Subjectdevice)
Hz, singlephase220/230VAC,6A, 50/60Hz, singlephase15 A max. 50/60 Hz.

{8}------------------------------------------------

High Technology Products SLU, Spain

5.7. Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed device complies with the following standards:

    1. IEC 60601-1:2005 + AMD1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General 2. requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 62304:2006/AMD:2015 Medical device software software life cycle 3. processes
    1. IEC 60601-2-22 Edition 3.1: Medical electrical equipment Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
    1. IEC60601-1-6 Edition 3.1: Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • IEC 60825 1 Edition 2: Safety of laser products Part 1: Equipment 6. classification and requirements.
    1. ISO 14971:2007 Medical devices Application of risk management to medical devices.
    1. ISO 10993-5:2009 Biological Evaluation of Medical Device, Part 5-Tests for In Vitro cytotoxicity
    1. ISO 10993-10:2010 Biological Evaluation of Medical Device, Part 10-Test for irritation and skin sensitization

5.8. Clinical Testing

No clinical study is included in this submission.

{9}------------------------------------------------

High Technology Products SLU, Spain

5.9. Conclusion

Based on the comparison and analysis above, the PRIMELASE Excellence device's intended use and technological characteristics do not raise new types of questions regarding safety and efficacy when compared to the predicates. Based on its technical characteristics, indications for use, and performance data, the PRIMELASE Excellence device is considered to be substantially equivalent to the predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.