K Number
K201510

Validate with FDA (Live)

Date Cleared
2022-02-18

(623 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Traufix medical devices are intended for fixation of fractures and osteotomies of various bones including the pelvis, clavicle, scapula, humerus, ulna, radius, femur, tibia, olecranon, and fibula.

Device Description

Traufix osteosynthesis medical devices are metal implants (stainless steel per ASTM F138, or ASTM F139, or titanium-alloy per ASTM F136) fixed adjacent to the bone for the treatment and consolidation of fractures, osteotomies, and/or joint ossification. The devices include different types and sizes of plates and screws.

AI/ML Overview

The provided text is a 510(k) summary for the Traufix osteosynthesis, osteotomy, and arthrodesis devices. It states that the device has been found substantially equivalent to predicate devices and therefore does not include a separate section outlining acceptance criteria or results from a study specifically designed to demonstrate meeting those criteria.

However, it does describe the non-clinical performance testing performed to support substantial equivalence. Based on the information provided, here's a breakdown:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria in a table format with specific pass/fail thresholds. Instead, it describes comparative engineering analysis against predicate devices and adherence to ASTM/ISO standards.

Acceptance Criteria (Implied)Reported Device Performance
Single cycle bend testing according to ASTM F382 (for plates)Performed according to ASTM F382. (Implicitly, the device performed acceptably or comparably to the predicate for substantial equivalence).
Breaking torque testing according to ISO 6475 (for screws)Performed according to ISO 6475. (Implicitly, the device performed acceptably or comparably to the predicate for substantial equivalence).
Structural properties comparison (bending stiffness, bending structural stiffness, and bending strength) per ASTM F382 A1 of subject proximal tibia plate vs. predicate proximal tibia plateEngineering analysis performed. The conclusion of substantial equivalence implies these properties were comparable and did not raise new safety or effectiveness concerns.
Design and dimensions comparison with predicate plates and screwsEngineering analysis and comparison of technology performed. Minor differences were found but did not raise new questions of safety or effectiveness.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a "sample size used for the test set" in terms of number of devices or data points beyond stating that "non-clinical testing was performed in the subject devices." Given that this is a 510(k) summary for a medical device (bone fixation appliances), the "test set" would typically refer to the specific device designs and materials tested in the non-clinical mechanical tests. The provenance is internal testing conducted by or for Fixier, S.A. de C.V. It is retrospective in the sense that the testing has already been completed to support the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This is not applicable to the non-clinical performance testing described. Ground truth is not established by human experts in the context of mechanical performance tests against engineering standards.

4. Adjudication Method for the Test Set

This is not applicable to the non-clinical performance testing described. Adjudication methods are typically used in clinical studies or human expert review scenarios.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

This is not applicable. The device is a physical bone fixation implant, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

This is not applicable. The device is a physical bone fixation implant, not an algorithm.

7. The Type of Ground Truth Used

For the non-clinical testing, the "ground truth" is defined by the industry standards and the properties of the legally marketed predicate devices. The testing aims to demonstrate that the subject device's mechanical properties are equivalent to or better than previously cleared devices, as measured by established ASTM and ISO standards for bend testing and torque testing.

8. The Sample Size for the Training Set

This is not applicable. The device is a physical implant and does not involve AI or machine learning algorithms that require a "training set."

9. How the Ground Truth for the Training Set Was Established

This is not applicable as there is no "training set" for this type of device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date March 18, 2022. The text is written in a clear, sans-serif font. The date is written in the standard US format, with the month first, followed by the day and year. The text is centered and takes up most of the image.

FDA U.S. FOOD & DRUG
ADMINISTRATION

Fixier, S.A. de C.V. % Aurelia Brownridge Regulatory Affairs Consultant Aurelia Brownridge 11594 Cesped Drive San Diego, California 92124

Re: K201510

Trade/Device Name: Traufix osteosynthesis, osteotomy, and arthrodesis devices Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC

Dear Aurelia Brownridge:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated February 18, 2022. Specifically, FDA is updating this SE Letter as an administrative correction, as two copies of the SE letter were sent in error along with the Indications for Use statement for K201510, but not the 510(k) Summary.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Shumaya Ali, M.P.H., OHT6: Office of Orthopedic Devices, (301) 796-2356, shumaya.ali@fda.hhs.gov.

Sincerely,

Shumaya Ali -S

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

February 18, 2022

Image /page/1/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized depiction of a human figure. The FDA logo is on the right and includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Fixier, S.A. de C.V. % Aurelia Brownridge Regulatory Affairs Consultant Aurelia Brownridge 11594 Cesped Drive San Diego, California 92124

Re: K201510

Trade/Device Name: Fixier osteosynthesis, osteotomy, and arthrodesis devices Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 15, 2022 Received: January 20, 2022

Dear Aurelia Brownridge:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumava Ali -

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K201510

Device Name

Traufix osteosynthesis, osteotomy, and arthrodesis devices

Indications for Use (Describe)

Traufix medical devices are intended for fixation of fractures and osteotomies of various bones including the pelvis, clavicle, scapula, humerus, ulna, radius, femur, tibia, olecranon, and fibula.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary - K201510

The following 510(k) summary has been prepared pursuant to requirements specified in 21 CFR 807.92.

510(k) Bundle: This submission includes a range of Fixier's products with the same regulation classification and regulation code name.

Date Summary Prepared: February 15, 2022

Manufacturer and Contact Information:

Submitter'sNameSubmitter'sPhysical AddressSubmitter'sContact Information
Fixier, S.A. deC.V.Carretera Doctor Mora a San Miguelde Allende km 3.4, C.P. 37967Comunidad de San Rafael, DoctorMora, Guanajuato, MéxicoPrimary Official Contact:Aurelia BrownridgeRegulatory Affairs (Consultant)Telephone: 760-497-4354Secondary Contact:Vianey Mariana Perez SuarezBiological and PharmaceuticalChemistryRegulatory Affairs Chief
Telephone: 011-52-419-688-11-91

Device and Classification Information:

Proprietary Name:Traufix osteosynthesis, osteotomy, and arthrodesis devices
Common Name:Plate, Fixation, Bone (primary)Screw, Fixation, Bone
Classification Name:Single/multiple component metallic bone fixation appliances andaccessories (primary)Smooth or threaded metallic bone fixation fastener
Regulatory Class:II
Product Codes:HRS (primary)HWC
Regulation Number:21 CFR 888.3030 (primary)21 CFR 888.3040

{5}------------------------------------------------

Indications for Use: Traufix medical devices are intended for fixation of fractures and osteotomies of various bones including the pelvis, clavicle, scapula, humerus, ulna, radius, femur, tibia, olecranon, and fibula.

Description: Traufix osteosynthesis medical devices are metal implants (stainless steel per ASTM F138, or ASTM F139, or titanium-alloy per ASTM F136) fixed adjacent to the bone for the treatment and consolidation of fractures, osteotomies, and/or joint ossification. The devices include different types and sizes of plates and screws. The product's list is shown below:

PlatetypePlatecodePlatedescriptionScrewcodeScrew descriptionSurgicaltechnique #Surgical techniquename
Minifragments0172.7 DCP Plates0642.7mm cortex screwDCP plates
2072.0 DCP Plates0792.0mm cortex screwTQ-005
202&2032.7mm L-plate0642.7mm cortex screwTQ-032T and L plates for3.5mm, 2.7mm, and2.0mm screws
0163.5mm One-third tubularplate0573.5mm cortex screwTQ-0043.5mm One-thirdtubular plate
0584.0mm cancellous screwpartially threaded
0594.0mm cancellous screwfully threaded
Smallfragments110,111,172,&173ALP titaniumdistal tibiaplates TIDIS &TIDIS-III1123.5mm titanium cortexscrewALP titanium distaltibia plateTIDIS I ANDTIDIS III
1063.5mm titanium cortexlocking screwTQ-011
1073.5mm titaniumcancellous lockingscrew
129ALP titaniumproximalhumerus platePROH-LOCKLARGE1123.5mm titanium cortexscrewTQ-014ALP titaniumhumerus platePROH-LOCK andPROH-LOCKLARGE
1063.5mm titanium cortexlocking screw
1073.5mm titaniumcancellous lockingscrew
144&145ALP titaniumdistal lateralhumerus plateDH-LOCK1123.5mm titanium cortexscrewTQ-017ALP titanium distallateral humerusplate and ALPtitaniumDistal medialhumerus plate DH-LOCK
1063.5mm titanium cortexlocking screw
1073.5mm titaniumcancellous lockingscrew
1702.7mm titanium cortexlocking screw
148,148S,149,&149S.ALP titaniumdistal radiusplate LIONTER154X2.5mm titanium cortexscrew Torx driveTQ-018ALP titanium distalradius plate
155X2.5mm titanium cortexlocking screw Torxdrive

{6}------------------------------------------------

Fixier
Traufix osteosynthesis, osteotomy and arthrodesis devices
PlatetypePlatecodePlatedescriptionScrewcodeScrew descriptionSurgicaltechnique #Surgical techniquename
164&165ALP titaniumdistal tibiaplates TIDIS-III1123.5mm titanium cortexscrewTQ-023ALP titanium distalplate TIDIS II
1063.5mm titanium cortexlocking screw
1073.5mm titaniumcancellous lockingscrew
0234.5mmReconstructionplates0564.5mm cortex screwTQ-009Reconstructionplates
0606.5mm cancellous screw16mm thread
0616.5mm cancellous screw32mm thread
0626.5mm cancellous screwfully threaded
Largefragments150&151ALP titaniumproximal femurplate TLP1264.5mm titanium cortexscrewTQ-019ALP titaniumproximal femurplate TLP
1085.0mm titanium cortexlocking screw
1715.0mm titaniumcannulated cortexlocking screw
1095.5mm titaniumcancellous lockingscrew
1567.3mm titaniumcannulated cancellouslocking screw
152&153ALP titaniumproximal tibiaplate OPTIMUS1264.5mm titanium cortexscrewTQ-020ALP titaniumproximalTibia plateOPTIMUS
1085.0mm titanium cortexlocking screw
1095.5mm titaniumcancellous lockingscrew
SubstantialEquivalence:Traufix osteosynthesis, osteotomy, and arthrodesis devices andpredicate devices have the same indications for use, targetpopulation, patient contact, materials, and sterilization methods.
Traufix subject devices, and predicate devices are intended forfracture fixation and minor differences exist in the physical shapeand dimension characteristics of the plates and screws. However,these minor dimensional differences do not raise any newquestions of safety or effectiveness.
Primary PredicateDevice:AUXEIN BRAND Of Bone Plates and Bone Screws & AUXEINBRAND Of DCS Plate System AUXEIN brand of Locking Platesand Locking Screws, (K141680)

{7}------------------------------------------------

Other PredicateDevices:-Synthes (USA) Modular Mini Fragment LCP System, (K063049)-Synthes (USA) Medial Distal Tibia Plates, (K001945)-Synthes Variable Angle Locking Compression Plate (VA-LCP),(K071184)-Synthes One-Third Tubular DCL Plate, (K011335)-Synthes 4.5mm LCP Straight Reconstruction Plates, (K051986)-Synthes (USA) LCP Proxima Humerus Plates, Long, (K041860)
Non-Clinical(Performance Testing):-Non-clinical testing was performed in the subject devices asfollows:-Single cycle bend testing was performed according to ASTMF382.-Breaking torque testing was performed according to ISO 6475.-Engineering analysis was performed to compare thestructural properties per ASTM F382 A1 (bending stiffness,bending structural stiffness, and bending strength) of the proximaltibia plate (subject device) and proximal tibia plate (predicatedevice one).-Engineering analysis and comparison of technology was alsoprovided to compare design and dimensions of the subject andpredicate plates and screws.
Clinical Testing:Clinical testing is not a prerequisite, as osteosynthesis, osteotomy,and arthrodesis implants conform to recognized standards and havewell established safety and performance characteristics. Therefore,clinical information is not necessary to demonstrate safety andeffectiveness or substantial equivalence.
Conclusion:The Traufix osteosynthesis, osteotomy, and arthrodesis deviceshave been found to be substantially equivalent to the predicatedevices. Differences between the subject devices and the predicatedevices do not raise new questions of safety or effectiveness.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.