(118 days)
The AcQMap 3D Imaging and Mapping Catheter is intended to be used in the right and left atrial chambers to collect ultrasound data for visualizing the selected chamber and recording electrical impulses in patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
The AcQMap 3D Imaging and Mapping Catheter ("AcQMap Catheter") is provided sterile and is a singleuse, non-pyrogenic, invasive device that is inserted into the femoral vein and advanced through the venous circulatory system to the inferior vena cava and into the right and/or left atrium of the heart.
The AcQMap Catheter consists of a shaft with a lumen, an integral handle with a deployment mechanism and a flush port, and a connector, as shown in Figure II.I. The AcQMap Catheter is intended to be used with the AcQMap High Resolution Imaging and Mapping System ("AcQMap System"). The AcQMap Catheter will provide ultrasound imaging and non-contact electrograms for Charge heart chamber mapping. The AcQMap Catheter has six (6) splines at the distal end that support a configuration of 48 electrodes and 48 ultrasound transducers. The AcQMap Catheter is placed within the desired heart chamber and the distal end is deployed. There is no requirement for the electrodes or transducers to be in contact with the heart wall. The AcQMap Catheter is capable of over-the-wire delivery and contains a flexible distal segment that allows it to be directed via a steerable sheath to various locations and directions of interest within the heart. Refer to Figure 10.1 for images of the device.
The AcQMap Catheter, Model 900009 is sterilized by EO method, has a reusable interface cable, is compatible with a 0.035" guidewire, and has minor design changes to improve manufacturability.
The provided text describes modifications to an existing device, the AcQMap 3D Imaging and Mapping Catheter, and asserts its substantial equivalence to a predicate device. It does not contain acceptance criteria or a study directly proving the modified device meets them in the traditional sense of a clinical performance study for AI devices. Instead, it relies on substantial equivalence to the predicate device, which had prior clinical testing.
Therefore, the following information will be extracted or inferred to the best of my ability from the provided document, addressing each of your points:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding performance metrics for the modified device in the context of clinical effectiveness. Instead, it relies on the predicate device's clinical performance and non-clinical testing for the modified device to demonstrate substantial equivalence.
The "acceptance criteria" for this submission are implicitly that the modified device does not raise new questions of safety or effectiveness compared to the predicate device. The reported performance relies on the predicate device's clinical study and the modified device's non-clinical testing.
| Acceptance Criteria (Implied for Substantial Equivalence) | Reported Device Performance (Modified Device) |
|---|---|
| Safety: | Biocompatibility: Passed biocompatibility testing, including NAVI testing. Sterilization: Sterilization validation performed (changed from e-beam to EO method). Mechanical Integrity: Met established specifications through mechanical testing (tensile testing). Corrosion Resistance: Passed corrosion testing. Electrode Coating Particulate: Passed electrode coating particulate testing. Acoustic Output: Maintained identical acoustic output (MI 0.06, ISTPA.3 0.08 mW/cm2, ISPPA.3 1.03) as the predicate, which itself would have been shown to be safe within regulatory limits. Usability: Usability testing performed. |
| Effectiveness: | Dimensional Accuracy: Passed dimensional inspection. Functional Performance (Electrical): Passed functional (electrical) and compatibility testing. Accuracy of Mapping/Imaging: Accuracy testing performed. Shelf Life & Packaging: Validated. Clinical Effectiveness: The clinical validation of the original AcQMap 3D Imaging Catheter (K170819) is applicable and demonstrated safety and effectiveness for its intended use in 84 patients based on "Dipole Density Right (and left) Atrial Mapping and Assessment of Therapy In Complex Supraventricular Tachycardia, (DDRAMATIC-SVT)" study. This study demonstrated that the AcQMap 3D Imaging and Mapping Catheter is safe and effective for its intended use, which is identical to the modified device. |
| Identical Indications for Use | The indications for use for the modified device are identical to those of the predicate device. |
| No New Questions of Safety/Effectiveness | The non-clinical testing demonstrates that the modified AcQMap 3D Imaging and Mapping Catheter does not raise different questions of safety or effectiveness for collecting data that enables the creation of 3D anatomic maps that display chamber-wide electrical activation when compared to the predicate device. |
2. Sample Size Used for the Test Set and Data Provenance
For the modified device, no clinical test set is described. The submission relies on non-clinical (bench) testing.
For the predicate device (clinical testing incorporated by reference):
- Sample Size: 84 patients.
- Data Provenance: Prospective, non-randomized, open-label study conducted at eight clinical sites outside the U.S.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
This information is not provided in the document. The document refers to the "DDRAMATIC-SVT" study for the predicate device but does not detail how ground truth was established within that study or the number/qualifications of experts involved. Given it's an invasive medical device for mapping arrhythmias, the "ground truth" would likely involve comparison to established clinical methods and expert interpretation of the complex arrhythmia data, but specifics are absent here.
4. Adjudication Method for the Test Set
This information is not provided in the document.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device (AcQMap 3D Imaging and Mapping Catheter) is an imaging and mapping catheter for collecting ultrasound data and electrical impulses, not an AI-assisted diagnostic or interpretation tool for human readers in the traditional sense of an MRMC study. It collects data that physicians then use. The document does not describe any human-in-the-loop performance evaluation or AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable in the context of an "algorithm only" performance that would typically be seen for AI products. The device itself is designed to function and capture data. The "standalone" performance here refers to the device's ability to "collect ultrasound data for visualizing the selected chamber and recording electrical impulses," which was assessed through non-clinical accuracy and functional testing and clinically by its ability to facilitate identification of complex arrhythmias, as demonstrated by the predicate device's clinical study. The device itself performs the data acquisition, which is then presented to clinicians.
7. The Type of Ground Truth Used
For the predicate device's clinical study (DDRAMATIC-SVT), the document does not explicitly state the type of ground truth used. However, for a device intended to help identify "complex arrhythmias that may be difficult to identify using conventional mapping systems alone," the ground truth would likely involve:
- Clinical Outcomes/Resolution: Efficacy in guiding successful ablation or treatment of arrhythmias.
- Expert Consensus/Diagnosis: Comparison of the mapping system's output to expert electrophysiologist (EP) diagnoses using established clinical practices or other validated mapping techniques.
For the non-clinical testing of the modified device, the "ground truth" would be against engineering specifications and known physical properties and measurements (e.g., dimensional accuracy, electrical performance, acoustic output measurements).
8. The Sample Size for the Training Set
This device is not described as an AI/ML device that requires a training set in the conventional sense. Therefore, information on a "training set" and its size is not applicable/provided.
9. How the Ground Truth for the Training Set was Established
As above, this is not applicable as the device is not presented as an AI/ML product with a training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 15, 2020
Acutus Medical, Inc. Karla Schaffner Principal Regulatory Affairs Specialist 2210 Faraday Ave., Suite 100 Carlsbad, California 92008
Re: K201341
Trade/Device Name: AcOMap 3D Imaging and Mapping Catheter, Model 900009 Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD, ITX Dated: August 5, 2020 Received: August 6, 2020
Dear Karla Schaffner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Fellman Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K201341
Device Name
AcQMap 3D Imaging and Mapping Catheter, Model 900009
Indications for Use (Describe)
The AcQMap 3D Imaging and Mapping Catheter is intended to be used in the right and left atrial chambers to collect ultrasound data for visualizing the selected chamber and recording electrical impulses in patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
| Type of Use (Select one or both, as applicable) |
|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Acutus Medical. The word "ACUTUS" is in gray, with a small orange dot in the "A". Below the word "ACUTUS" is the word "MEDICAL" in orange. A registered trademark symbol is to the right of the word "ACUTUS".
5 | 0(K) Notification K20 | 34 I
General Information [807.92(a)(l)]
Date Prepared: 15 September 2020
Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1-442-232-6080 Fax: 1-442-232-6081
Contact Person:
Applicant:
Karla Schaffner Principal Regulatory Affairs Specialist Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1-442-232-6161 FAX: 1-442-232-6081 Email: karla.schaffner@acutus.com
Device Information [807.92(a)(2)]
Trade Name:
AcQMap® 3D Imaging and Mapping Catheter, Model 900009
Generic/Common Name:
Electrode Recording Catheter or Electrode Recording Probe/ Transducer Ultrasonic
Classification:
Class II / 21 CFR § 870.1220 and Class II / 21 CFR § 892.1570
Product Code(s): MTD/ITX
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Acutus Medical. The logo has the word "ACUTUS" in gray, with a small orange dot in the lower left corner of the "A". Below the word "ACUTUS" is the word "MEDICAL" in orange. The logo is simple and modern.
PREDICATE DEVICE [807.92(a)(3)]
AcQMap 3D Imaging and Mapping Catheter (K170819)
DEVICE DESCRIPTION [807.92(a)(4)]
The AcQMap 3D Imaging and Mapping Catheter ("AcQMap Catheter") is provided sterile and is a singleuse, non-pyrogenic, invasive device that is inserted into the femoral vein and advanced through the venous circulatory system to the inferior vena cava and into the right and/or left atrium of the heart.
The AcQMap Catheter consists of a shaft with a lumen, an integral handle with a deployment mechanism and a flush port, and a connector, as shown in Figure II.I. The AcQMap Catheter is intended to be used with the AcQMap High Resolution Imaging and Mapping System ("AcQMap System"). The AcQMap Catheter will provide ultrasound imaging and non-contact electrograms for Charge heart chamber mapping. The AcQMap Catheter has six (6) splines at the distal end that support a configuration of 48 electrodes and 48 ultrasound transducers. The AcQMap Catheter is placed within the desired heart chamber and the distal end is deployed. There is no requirement for the electrodes or transducers to be in contact with the heart wall. The AcQMap Catheter is capable of over-the-wire delivery and contains a flexible distal segment that allows it to be directed via a steerable sheath to various locations and directions of interest within the heart. Refer to Figure 10.1 for images of the device.
The AcQMap Catheter, Model 900009 is sterilized by EO method, has a reusable interface cable, is compatible with a 0.035" guidewire, and has minor design changes to improve manufacturability.
INDICATIONS FOR USE [807.92(a)(5)]
The AcQMap 3D Imaging and Mapping Catheter is intended to be used in the right and left atrial chambers to collect ultrasound data for visualizing the selected chamber and recording electrical impulses in patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
DEVICE MODIFICATIONS
The following changes were introduced for the AcQMap Catheter Model 900009:
- . Changed the distal basket material from nitinol to a molded plastic
- Updated the handle design from a rotating knob deployment to a sliding knob deployment
- Changed the distal tip design to a PEEK distal core and stainless steel distal ring .
- . Modified distal seal design
- · Introduced a re-usable cable to reduce hospital waste
- · Updated the design to accommodate a more common 0.035" guidewire
- · Changed the sterilization method from e-beam radiation to EO
- Replaced coaxial wires with long flex circuits .
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Acutus Medical. The logo consists of the word "ACUTUS" in gray block letters, with a small orange circle in the lower left corner of the "A". Below the word "ACUTUS" is the word "MEDICAL" in smaller, orange letters. The "®" symbol is located in the upper right corner of the "S".
ACQMAP® 3D IMAGING AND MAPPING CATHETER
Comparison of Technological Characteristics with the Predicate Device [807.92(A)(6)]
| COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE [807.92(A)(6)] | |||
|---|---|---|---|
| Characteristics | SUBJECTModified AcQMap 3D Imaging &Mapping Catheter, Model 900009 | PREDICATEAcQMap 3D Imaging & MappingCatheter, Model 900003 | Rationale forSubstantialEquivalence |
| Regulatory | |||
| 510(k) Number | K201341 | K170819 | -- |
| Classification/Regulation Number/Regulation Name/Product Code | Class II / 21 CFR §870.1220 /Electrode recording catheter orelectrode recording probe / MTDClass II / 21 CFR §892.1570 /Transducer Ultrasonic/ ITX | Class II / 21 CFR §870.1220 /Electrode recording catheter orelectrode recording probe / MTDClass II / 21 CFR §892.1570 /Transducer Ultrasonic/ ITX | Identical |
| Indications for Use | The AcQMap 3D Imaging andMapping Catheter is intended to beused in the right and left atrialchambers to collect ultrasound datafor visualizing the selected chamberand recording electrical impulses inpatients with complex arrhythmiasthat may be difficult to identify usingconventional mapping systems alone. | The AcQMap 3D Imaging andMapping Catheter is intended to beused in the right and left atrialchambers to collect ultrasound datafor visualizing the selected chamberand recording electrical impulses inpatients with complex arrhythmiasthat may be difficult to identifyusing conventional mapping systemsalone. | Identical |
| Physical Characteristics | |||
| Configuration | Electrode/Transducer mounted onsplines | Electrode/Transducer mounted onsplines | Identical |
| Deployed Diameter | 25mm | 25mm | Identical |
| Working Length | 100 cm | 100 cm | Identical |
| Number of Spines | 6 | 6 | Identical |
| Heart Wall Contacting | No | No | Identical |
| Deflection Y/N | No | No | Identical |
| Flush Port | Yes | Yes | Identical |
| Technical Specifications - Electrical | |||
| Voltage | Yes | Yes | Identical |
| Charge Density | Yes | Yes | Identical |
| Number of Electrodes | 48 | 48 | Identical |
| Data Points/heartbeat | 2500 | 2500 | Identical |
| Power source | Electrical source from AcQMapSystem | Electrical source from AcQMapSystem | Identical |
| Technical Specifications – Ultrasound | |||
| Mode | M-mode | M-mode | Identical |
| Number oftransducers | 48 | 48 | Identical |
| Phased Array Y/N | N | N | Identical |
| Center Frequency | 10 MHz | 10 MHz | Identical |
| Characteristics | SUBJECTModified AcQMap 3D Imagingand Mapping Catheter, Model900009 | PREDICATEAcQMap 3D Imaging andMapping Catheter, Model900003 | Rationale forSubstantialEquivalence |
| Acoustic Output | |||
| MI (Mechanical Index) | 0.06 | 0.06 | Identical |
| ISTPA.3 (DeratedSpatial-PeakTemporal-AverageIntensity (milliwattsper squarecentimeter)) | 0.08 (mW/cm2) | 0.08 (mW/cm2) | |
| ISPPA.3 (DeratedSpatial-Peak PulseAverage Intensity(watts per squarecentimeter)) | 1.03 | 1.03 | |
| Accessories | |||
| Compatible Sheath | For Use with Acutus MedicalAcQGuide 12 F Steerable Sheath(K162925) | For Use with Acutus MedicalAcQGuide 12 F Steerable Sheath(K162925) | Identical |
| Compatible Guidewire | 0.035" (0.89 mm) diameter J-tipguidewire | 0.032" (0.81 mm) diameter J-tipguidewire | The difference inaccessory sizecompatibility doesnot raise differentquestions of safetyor effectiveness |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in gray, with a small orange dot above the left side of the "A". Below the word "ACUTUS" is the word "MEDICAL" in orange, with each letter spaced out. The logo is simple and modern, and the colors are muted and professional.
SUBSTANTIAL EQUIVALENCE
The indications for use of the subject device are identical to those of the predicate device. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the modified AcQMap 3D Imaging and Mapping Catheter, is substantially equivalent to the predicate AcQMap 3D Imaging and Mapping Catheter device.
Performance Data [807.92(b)]
All necessary bench testing was conducted on the modified AcQMap 3D Imaging and Mapping Catheter to support a determination of substantial equivalence to the predicate device. The necessary clinical testing was completed for the original AcQMap Catheter (K170819) and is incorporated by reference. No further clinical testing is required to support the subject device.
Nonclinical Testing Summary [807.92(b)(1)]
The nonclinical, bench testing is repeated or included in this submission by reference. A full list of the non-clinical testing is provided below:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in gray, with a small orange dot above the left side of the "A". Below the word "ACUTUS" is the word "MEDICAL" in orange. The logo is simple and modern, and the colors are complementary.
- Design Verification .
- Dimensional Inspection o
- Visual Inspection 0
- Functional (electrical) and Compatibility Testing 0
- Mechanical Testing (tensile testing) O
- Corrosion Testing O
- Electrode Coating Particulate Testing O
- Acoustic Output Testing O
- Accuracy Testing o
- Design Validation ●
- Usability Testing 0
- 0 Animal Testing
- Biocompatibility Testing including NAVI testing ●
In addition, Acutus performed sterilization, shelf life and packaging validations. The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the AcQMap 3D Imaging and Mapping Catheter meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the AcQMap 3D Imaging and Mapping Catheter does not raise different questions of safety or effectiveness for collecting data that enables the creation of 3D anatomic maps that display chamber-wide electrical activation when compared to the predicate device.
Clinical Testing Summary [807.92(b)(2)]
As discussed above, no further clinical testing is required to support the modified AcQMap 3D Imaging and Mapping Catheter. The necessary clinical testing was completed for the original AcQMap 3D Imaging and Mapping Catheter (K170948) and is incorporated by reference. That study, entitled, "Dipole Density Right (and left) Atrial Mapping and Assessment of Therapy In Complex Supraventricular Tachycardia, (DDRAMATIC-SVT)" was a prospective, non-randomized, open-label study conducted at eight clinical sites outside the U.S. The results for 84 patients demonstrated that the AcQMap 3D Imaging and Mapping Catheter is safe and effective for its intended use.
Conclusions [807.92(b)(3)]
Extensive nonclinical performance testing, either repeated for the modified device or incorporated by reference to the original AcQMap 3D Imaging and Mapping Catheter 510(k), was conducted on the modified AcQMap 3D Imaging and Mapping Catheter to evaluate the overall performance of the device. The clinical validation of the original AcQMap 3D Imaging Catheter (K170819) is applicable to the modified device. The collective results demonstrate that the modified AcQMap 3D Imaging and Mapping Catheter is safe and effective for its intended use, and further, is substantially equivalent to the predicate device
§ 870.1220 Electrode recording catheter or electrode recording probe.
(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).