K Number
K201280
Date Cleared
2020-07-27

(75 days)

Product Code
Regulation Number
878.4750
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ethicon™ Circular Staplers have application throughout the alimentary tract for end, end-to-soside anastomoses.

Device Description

The Ethicon™ Circular Staplers are sterile, single use devices that simultaneously staple and cut tissue to create an anastomosis. The devices deliver 2 concentric rows of staples on the outside of the cut line. The Ethicon™ Circular Staplers are available in two device configurations: CDH and ECS and two shaft lengths; a standard 26 cm shaft and an XL 37 cm shaft. Each configuration is available in 4 endeffector sizes: 21 mm, 29 mm, 33 mm. These configurations function in the same manner - to compress tissue and to produce the same closed staple on tissue. Each device has a detachable anvil that allows a surgeon to place the anvil in the desired location.

AI/ML Overview

This 510(k) summary describes a device modification, rather than a completely new device. Therefore, the "acceptance criteria" and "device performance" discussed here refer to the device's ability to maintain its previous performance after the modification, not to establish initial performance benchmarks.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Design Requirements)Reported Device Performance
Tissue cutting reliabilityMaintained reliable tissue cutting (as demonstrated by testing)
Force-to-fire (test skin)Maintained appropriate force-to-fire (as demonstrated by testing)
Device stapling performanceMaintained stapling performance (as demonstrated by testing)
Overall device functionPerformed as intended after modification
Meeting design specificationsMet design specifications after modification

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes used for each of the "ex-vivo (bench) testing" aspects (tissue cutting reliability, force-to-fire, stapling performance). It only mentions that testing was performed.

The data provenance is retrospective in the sense that the testing was performed on the modified device to verify that the changes did not negatively impact performance, essentially re-validating against existing design requirements. The data is from bench testing, meaning it was conducted in a laboratory setting, not in a clinical environment. No country of origin for the data is specified, but it can be inferred to be associated with the manufacturer, Ethicon Endo-Surgery, LLC.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This was bench testing against design specifications, not a study requiring expert interpretation of results or establishing ground truth in a clinical or diagnostic context.

4. Adjudication Method for the Test Set

Not applicable. As this was bench testing, results likely involved objective measurements against predefined engineering specifications. There would be no need for adjudication by multiple human observers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission form explicitly states: "This submission does not include data from Clinical Studies."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. The device is a mechanical surgical stapler. It does not involve any algorithms or AI that would require standalone performance evaluation.

7. The Type of Ground Truth Used

The "ground truth" for this submission was the existing design requirements and specifications for the Ethicon™ Circular Staplers. The testing aimed to demonstrate that the modified device continued to perform to these established specifications.

8. The Sample Size for the Training Set

Not applicable. There is no "training set" as this device is a mechanical stapler and does not involve machine learning or AI.

9. How the Ground Truth for the Training Set was Established

Not applicable. As there is no training set for a mechanical device. The original design requirements and performance characteristics would have been established during the initial development and clearance processes for the predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 27, 2020

Ethicon Endo-Surgery LLC % Rubina Dosani Manager, Regulatory Affairs Ethicon Endo-Surgery, LLC 4545 Creek Road Cincinnati, Ohio 45242

Re: K201280

Trade/Device Name: Ethicon Circular Stapler, Ethicon Circular Stapler -XL Sealed Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: May 9, 2020 Received: June 29, 2020

Dear Rubina Dosani:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201280

Device Name Ethicon™ Circular Stapler

Indications for Use (Describe)

The Ethicon™ Circular Staplers have application throughout the alimentary tract for end, end-to-soside anastomoses.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

|X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter Information: Ethicon Endo-Surgery, LLC 475 Calle Street Guaynabo, PR 00969

Application Correspondent

Rubina Dosani Manager, Regulatory Affairs Ethicon Endo-Surgery, LLC. Telephone: (513) 236-2841 (513) 337-2314 Fax: Email: rdosani@its.jnj.com

Date Prepared May 9, 2020

Device Trade Name:Ethicon™ Circular Staplers,Ethicon™ Circular Staplers XL Sealed
Device Common Name:Circular Stapler
Classification Regulation:21 CFR 878.4750; Implantable Staple
Device Class:II
Panel:79, General & Plastic Surgery
Classification (Product) Code:GDW
Legally Marketed Predicate Device:Ethicon™ Circular Staplers, Ethicon™ Circular Staplers

Device Description

The Ethicon™ Circular Staplers are sterile, single use devices that simultaneously staple and cut tissue to create an anastomosis. The devices deliver 2 concentric rows of staples on the outside of the cut line. The Ethicon™ Circular Staplers are available in two device configurations: CDH and ECS and two shaft lengths; a standard 26 cm shaft and an XL 37 cm shaft. Each configuration is available in 4 endeffector sizes: 21 mm, 29 mm, 33 mm. These configurations function in the same manner - to compress tissue and to produce the same closed staple on tissue. Each device has a detachable anvil that allows a surgeon to place the anvil in the desired location.

XL. (cleared under K181653)

Indications for Use

The Ethicon™ Circular Staplers have applications throughout the alimentary tract for end-toend, end-to side, and side-to-side anastomoses. The predicate device also has the same indications for use. There is no change to the indications for use.

Technological Characteristics

{4}------------------------------------------------

The Ethicon™ Circular Stapler is substantially equivalent to the predicate Ethicon Endo-Surgery® Curved Intraluminal Staplers with respect to the device function and design. The subject device has a similar design as the predicate except for a change in the knife grind angle and the washer thickness in the end-effector of the device. Similar to the predicate, a rotatable adjustment knob enables the compression of tissue and selection of a target staple height based on the tissue compression within the green zone. The device is manually powered; it is not powered by an outside energy source. Neither the subject device nor predicate device uses software.

Performance Data

Ex-vivo (bench) testing was performed to ensure that the device performed as intended and met design specifications. Device performance was assessed against the design requirements. Risk analysis and design verification testing were conducted for the changes described in this submission to ensure that the performance of the device was not affected by the device modifications. Verification testing included tissue cutting reliability testing, test skin force-tofire, and device stapling performance testing. The results of the testing support the conclusion that the performance of the device has not been affected by the changes.

This submission does not include data from Clinical Studies.

Conclusion

The modification described in this submission does not affect the intended use of the device or alter the fundamental scientific technology of the device. Summary information from the design control process serves as the basis for this submission. The Ethicon™ Circular Stapler is substantially equivalent to the legally marketed predicate device based upon intended use, technological characteristics and performance testing.

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.