K Number
K201151

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-06-22

(54 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The O-Mec Laparoscopes 690 Series (Models, 90-331000H, 690-300500H, 690-300500H, are intended to be used by surgeons in diagnostic and therapeutic procedures. Laparoscopic minimally invasive procedures are performed in the abdominal cavity by means of small skin punctures that allow the insertion of the laparoscope and laparoscopic instruments.

Device Description

O-Mec laparoscopes 690 Series Models 690- 331000H, 690-331030H, 690-300500H, 690-300530H are rigid endoscopes that are used to view a patient's internal anatomy for examination, diagnosis and therapy during laparoscopic procedures. An endoscope is a slender, tubular optical instrument used as a viewing system for examining an inner part of the body. The inside of the endoscope contains a series of lenses that transmit the endoscopic image, which is illuminated by an external light source.

O-Mec laparoscopes 690 Series Models 690- 331000H, 690-331030H, 690-300500H, 690-300530H are available in two sizes: 5mm or 10mm diameters, with either 0° or 30° directions of view. O-Mec laparoscope provides a high definition image and can correct chromatic aberration in the 400-900nm wavelength range.

O-Mec laparoscopes 690 Series Models 690- 331000H, 690-331030H, 690-300500H, 690-300530H are delivered non-sterile and are reusable and fully autoclavable. O-Mec laparoscopes must be cleaned according to the Instructions of Use before each use.

AI/ML Overview

The provided text describes the O-Mec Laparoscopes 690 Series and its substantial equivalence to predicate devices, but it does not contain information related to AI/ML device performance, acceptance criteria for AI/ML, or studies explicitly designed to prove an AI/ML device meets such criteria.

The document is a 510(k) Pre-Market Notification for a traditional medical device (laparoscopes). The "acceptance criteria" discussed are for the physical and functional performance of the laparoscope itself (e.g., optical performance, sterilization validation, electrical safety) in comparison to predicate devices, not for an AI/ML algorithm's diagnostic or predictive capabilities.

Therefore, I cannot extract the requested information, such as:

  1. A table of acceptance criteria and reported device performance for an AI/ML device.
  2. Sample sizes used for test sets (for AI/ML), data provenance, etc.
  3. Number of experts, qualifications, and adjudication methods for AI/ML ground truth.
  4. MRMC comparative effectiveness study details (AI vs. human).
  5. Standalone performance of an AI algorithm.
  6. Type of ground truth used (for AI/ML).
  7. Sample size for the training set (for AI/ML).
  8. How ground truth for the training set was established (for AI/ML).

The document explicitly states: "Clinical testing was not performed with this device." and the "Summary of Non-Clinical Testing" focuses on hardware performance, cleaning, and safety. There is no mention of an AI/ML component or any associated performance metrics or studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an abstract design. To the right of the seal, the text "FDA U.S. FOOD & DRUG ADMINISTRATION" is displayed in blue. The word "FDA" is in a larger, bolder font, and the words "U.S. FOOD & DRUG" are stacked above the word "ADMINISTRATION".

June 22, 2020

OMEC Medical, Inc. % Ms. Yolanda Smith Sr. Regulatory Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114 Re: K201151 Trade/Device Name: O-Mec Laparoscopes 690 Series (Models 690- 331000H, 690-300500H, 690-300530H)

Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: April 14, 2020 Received: April 29, 2020

Dear Ms. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201151

Device Name

O-Mec Laparoscopes 690 Series (Models, 690-331000H, 690-331030H, 690-300500H, 690-300530H)

Indications for Use (Describe)

The O-Mec Laparoscopes 690 Series (Models, 90-331000H, 690-300500H, 690-300500H, are intended to be used by surgeons in diagnostic and therapeutic procedures. Laparoscopic minimally invasive procedures are performed in the abdominal cavity by means of small skin punctures that allow the insertion of the laparoscope and laparoscopic instruments.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

1. Submitter Information

Company Name:OMEC Medical Inc.
Company Address:2370 Qume Drive, Suite A
San Jose, CA 95131
Telephone:408-667-5307
Fax:408-436-5557
Contact Person:Ms. Claire Nippress
Email:clairen@omectech.us

2. Correspondent Information

Contact:Ms. Yolanda Smith
Sr. Regulatory Consultant
Smith Associates
Address:1468 Harwell Ave
Crofton, MD 21114
Phone:(410)451-0639
Ysmith9746@fdaconsultants.com

3. Date Prepared: June 17, 2020

4. Device Name

Classification Name:Laparoscope, General & Plastic Surgery
Common/Usual Name:Laparoscope
Proprietary Name:O-Mec Laparoscopes 690 Series (Models 690-331000H, 690-331030H, 690-300500H,690-300530H)
Regulation Number:21 CFR 876.1500
Product Code:GCJ
Device Class:Laparoscopes 21 CFR 876.1500
Reviewing Panel:General & Plastic Surgery
Basis for Submission:New device 510k

5. Predicate Device

Legally Marketed Equivalent Device

ManufacturerBrand Name510(k)Number
PrimaryStryker EndoscopyStryker Bariatric LaparoscopeK993045
ReferenceSchoelly Fiberoptics GmbHSchoelly LaparoscopeK143221
ReferenceStryker CorporationStryker LaparoscopeK910132

6. Device Description

{4}------------------------------------------------

O-Mec laparoscopes 690 Series Models 690- 331000H, 690-331030H, 690-300500H, 690-300530H are rigid endoscopes that are used to view a patient's internal anatomy for examination, diagnosis and therapy during laparoscopic procedures. An endoscope is a slender, tubular optical instrument used as a viewing system for examining an inner part of the body. The inside of the endoscope contains a series of lenses that transmit the endoscopic image, which is illuminated by an external light source.

O-Mec laparoscopes 690 Series Models 690- 331000H, 690-331030H, 690-300500H, 690-300530H are available in two sizes: 5mm or 10mm diameters, with either 0° or 30° directions of view. O-Mec laparoscope provides a high definition image and can correct chromatic aberration in the 400-900nm wavelength range.

O-Mec laparoscopes 690 Series Models 690- 331000H, 690-331030H, 690-300500H, 690-300530H are delivered non-sterile and are reusable and fully autoclavable. O-Mec laparoscopes must be cleaned according to the Instructions of Use before each use.

Catalog No.Working LengthMaximum Width ofInserted PortionDirection of View
690-331000H331mm10.1mm
690-331030H334mm10.1mm30°
690-300500H301mm5.5mm
690-300530H303mm5.5mm30°

Table 1: Model Specifications - 690 Series

7.Indications for Use

The O-Mec Laparoscopes 690 Series (Models, 90-331030H, 690-331030H, 690-300530H) are intended to be used by surgeons in diagnostic and therapeutic procedures. Laparoscopic minimally invasive procedures are performed in the abdominal cavity by means of small skin punctures that allow the insertion of the laparoscope and laparoscopic instruments. CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.

8. Comparison of Technical Characteristics

The table below lists the comparison of the indications for use and technological characteristics of the subject and predicate device.

ParameterSubject DevicePrimaryPredicate DeviceReferenceK143221Comments
510(k)NumberK201151K993045K143221
Brand Name690 SeriesStryker BariatricLaparoscopeSchoellyLaparoscope
ParameterSubjectDeviceK201151Predicate DeviceK993045Predicate DeviceK143221Comments
RegulationNumber876.1500876.1500876.1500Same
RegulationNameLaparoscope, general& Plastic SurgeryEndoscope andAccessoriesEndoscope andAccessoriesSame
Product CodeGCJGCJGCJSame
Indications forUseThe O-Mec 690 SeriesLaparoscopes (690-331000H, 690-331030H, 690-300500H, 690-300530H) are intendedto be used by surgeonsin diagnostic andtherapeuticprocedures.Laparoscopic minimallyinvasive proceduresare performed in theabdominal cavity bymeans of small skinpunctures that allowthe insertion of thelaparoscope andlaparoscopicinstruments.Laparoscopes areintended to be usedby surgeons indiagnostic andtherapeuticprocedures.Laparoscopicminimally invasiveprocedures areperformed in theabdominal cavity bymeans of small skinpunctures that allowthe insertion of thelaparoscope andlaparoscopicinstruments. Thisincludes, but is notlimited to such usesas gallbladder andappendix removal,hernia repair, gastricbypass, laparoscopicNissen andexamination of theabdominal cavity,appendix, gallbladderand liver.The SchoellyLaparoscope isindicated forexamination ofbody cavities,hollow organs,and canals, andusing additionalaccessories, toperform variousdiagnostic andtherapeuticprocedures.Similar
Principle ofOperationRigid, tubular, opticalinstrument throughan incision in theabdominal wall, usedto examine organsinside the abdomen.Rigid, tubular, opticalinstrument throughan incision in theabdominal wall, usedto examine organsinside the abdomen.Rigid, tubular,optical instrumentthrough anincision in theabdominal wall,used to examinebody cavities,hollow organs,and canals.Similar
Models Feature Comparison
ParameterSubject DeviceK201151Predicate DeviceK993045Predicate DeviceK143221Comments
Catalog no690-331000H502-537-01011.0031a
WorkingLength331mm300mm344mmSimilar
MaximumWidth ofinsertedPortion10.1mmØ10.0mm10.0mmSimilar
Direction ofviewSame
Catalog no690-331030H502-537-03011.0043a
WorkingLength334mm300mm344mmSimilar
MaximumWidth ofinsertedPortion10.1mmØ10.0mm10.0mmSimilar
Direction ofview30°30°30°Same
Catalog no690-300500H502-937-010N/A
WorkingLength301mm300mm312mmSimilar
MaximumWidth ofinsertedPortion5.5mmØ5.5mm5.0mmSimilar
Direction ofviewSame
Catalog no690-300530H502-937-030N/A
WorkingLength303mm300mm312mmSimilar
MaximumWidth ofinsertedPortion5.5mmØ5.5mm5.0mmSimilar
Direction ofview30°30°30°Same
Optical System
ParameterSubject DeviceK201151Predicate DeviceK993045Predicate DeviceK143221Comments
Depth of Field25mm to 150mmN/AN/AInformationnot available
Field of View75°N/AWide angleOnly wideanglereferenced
CorrectChromaticAberration400-900nmwavelength rangeN/AN/AReferencepredicateK910132
Material Choices
BodyAssemblyStainless Steel 304Stainless Steel 316LOptical GlassSapphireEpoxy GlueStainless SteelOptical glassStainless Steel304Optical GlassSimilar
EyepieceEyepiece isstandard:ISO/TS18339:2015-11Endotherapydevices -Eyepiece capand light guideconnectorPolyetheretherketone(PEEK)YesYesSame
Accessories
Cable AdapterCompatibilityWolf AdapterStorz AdapterACMI AdapterN/AN/A
Non- sterileDelivered innon-sterilecondition.Needs to besterilizedbefore eachuse.YesYesYesSame
AutoclavableYesYesYesSame
Reusable/ReprocessableYesYesYesSame

Table 2: Comparator table for Subject Device

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Reference Predicate

K910132 is included as a reference predicate for the 400-900nm wavelength. The subject and reference device have very minor modifications to enhance near infrared images.

9. Summary of Non-Clinical Testing

The following performance data were provided in support of the substantial equivalence determination.

Table 3: Summary Table of Non-Clinical Testing

CategoryTest TitleEvaluationTest CriteriaResults
FunctionalPerformanceLaparoscopePerformance TestMarkingMet acceptance criteria
Surface & EdgesMet acceptance criteria
ConnectionMet acceptance criteria
DimensionsMet acceptance criteria
CleanlinessMet acceptance criteria
Aperture and CenterFocusMet acceptance criteria
Evenness ofIlluminationMet acceptance criteria
Image Consistencyand Depth of FieldMet acceptance criteria
Image Focus andColorMet acceptance criteria
Reflection and RunoutMet acceptance criteria
FOV/DOVMet acceptance criteria
MTFMet acceptance criteria
Transmission /VignettingMet acceptance criteria
DistortionMet acceptance criteria
RunoutMet acceptance criteria
Sterilization/CleaningValidationStudiesManual/MechanicalCleaning ValidationGLP ReportCleaning validationwas performed forrecommendedprocedures and EnzoldetergentHemoglobin TestMicro BCA ProteinTestMEM Elution TestAAMITIR30:2011The cleaning study wassuccessful in showingthe reduction of thephysical markers.This provides evidencethat laparoscopes canbe cleaned effectivelyusing the instructionsprovided.
Simulated UseCycles Final ReportSix simulated usecycles wereperformed. Each cycleincluded soiling,cleaning andAAMITIR30:2011No soil was observed onany of the articles.
CategoryTest TitleEvaluationTest CriteriaResults
sterilizationproceduresArticle oilingManual CleaningprocedureSterilization
Steam Sterilizationvalidation FinalReportTest article wasevaluated to a sterilityassurance level (SAL)of $≤10-6$ANSI/AAMI/ISO17665-1:2006/®2013,Annex D and thevalidationapproach outlinedin ANSI/AAMI/ISO14937:2009®2013,Annex D(Approach 3)Results validated followingsterilization set points:Sterilizer Type: PrevacuumPreconditioning Pulses: 4Temperature: 132°C FullCycle Time: 4 minutesDry Time: 40 minutes TestArticle Configuration:Individually wrapped intwo layers of 1-plypolypropylene wrap(Halyard Health H200- 510(k)K082554) using sequentialenvelope foldingtechniques LoadConfiguration: One 20pound tray filled withmiscellaneous stainlesssteel dunnage was placedabove the test article inthe sterilizer
ElectricalSafetyMedical electricalequipment Part 1generalrequirements forbasic safety andessentialperformanceClause 5.7 HumidityPreconditioningTreatmentClause 5.9.2Determination ofAccessible partsClause 7.1.2 Legibilityof Marking Clause7.1.3 MarkingDurabilityClause 8.7 LeakagecurrentClause 8.8.3Dielectricstrength testIEC 60601-12005+A1:2012Pass
Medicalelectricalequipment Part 2:particularrequirement forthe basic safetyand essentialperformance ofendoscopicequipmentIEC60601-2-18Pass
CategoryTest TitleEvaluationTest CriteriaResults
Clause 11.1,. 1w=ExcessivetemperaturesClause 15.3.2 PushtestClause 15.3.4.1 DroptestClause 15.3.6Mould stress relieftest
HazardAnalysisMedical devices -Application of riskISO 14971 2ndEdition R 2010

{9}------------------------------------------------

{10}------------------------------------------------

Clinical testing was not performed with this device.

management to medical devices

10. Conclusion

The subject device is similar and or same for technological features to the predicate devices with slight differences for working length and width of inserted portion. The differences introduce not new issues of safety.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.