K Number
K201103

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-09-25

(154 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The system is intended for use by Nuclear Medicine (NM) or Radiology practitioners and referring physicians for display, processing, archiving, printing, reporting and networking of NMI data, including planar scans (Static, Whole Body, Dynamic, Multi-Gated) and tomographic scans (SPECT, Gated SPECT, dedicated PET or Camera-Based-PET) acquired by gamma cameras or PET scanners. The system can run on dedicated workstation or in a server-client configuration. The NM or PET data can be coupled with registered and/or fused CT or MR scans, and with physiological signals in order to depict, localize, and/or quantify the distribution of radionuclical structures in scanned body tissue for clinical diagnostic purposes.

The DaTQUANT optional application enables visual evaluation of 1231-ioflupane (DaTscanTM) images. DaTQUANT Normal Database option enables quantification relative to normal population databases of 1231ioflupane (DaTscanTM) images. These applications may assist in detection of loss of functional dopaminergic neuron terminals in the striatum, which is correlated with Parkinson disease.

The O.Lung application may aid physicians in:

· Diagnosis of Pulmonary Embolism (PE), Chronic Obstructive Pulmonary Disease (COPD), Emphysema and other lung deficiencies.

• Assess the fraction of total lung function provided by a lobe or whole lung for Lung cancer resection requiring removal of an entire lobe, bilobectomy, or pneumonectomy.

The Q.Brain application allows the user to visualize and quantify relative changes in the brain's metabolic function or blood flow activity between a subject's images and controls, which may be resulting from brain functions in: · Epileptic seizures

· Dementia. Such as Alzheimer's disease, Lewy body dementia, Parkinson's disease with dementia, and frontotemporal dementia.

  • · Inflammation
  • · Brain death
  • · Cerebrovascular disease such as Acute stroke, Chronic and acute ischemia

· Traumatic Brain Injury (TBI)

When integrated with the patient's clinical and diagnostic information may aid the physician in the interpretation of cognitive complaints, neuro-degenerative disease processes and brain injuries.

The Alcyone CFR application allows for the quantification of coronary vascular function by deriving Myocardial Blood Flow (MBF) and then calculating Coronary Flow Reserve (CFR) indices on data acquired on PET scanners and on stationary SPECT scanners with the capacity for dynamic SPECT imaging. These indices may add information to physicians using Myocardial Perfusion Imaging for the diagnosis of Coronary Artery Disease (CAD).

The Exini Bone application is intended to be used with NM bone scans for the evaluation of adult male patients with bone metastases from prostate cancer. Exini Bone quantifies the selected lesions and provides a Bone Scan Index value as adjunct information related to the progression of disease.

The Q.Liver application provides processing, quantification, and multidimensional review of Liver SPECT/PET and CT images for display, segmentation, and a calculation of the SPECT 'liver to lung' shunt value and the patient's Body Surface Area (BSA) for use in calculating a therapeutic dose for Selective Internal Radiation Therapy (SIRT) treatment using a user defined formula.

Device Description

The Xeleris V Processing and Review Workstation is a Nuclear Medicine Software system that is designed for general nuclear medicine processing and review procedures for detection of radioisotope tracer uptake in the patient's body, using a variety of individual processing applications orientated to specific clinical applications.

Xeleris V is a modification to the predicate device, Xeleris 4.0 Processing and Review Workstation. It includes all the clinical applications and features in the current production version Xeleris 4.0 and introduces new clinical applications and enhancements to previously cleared clinical applications.

With Xeleris V, the customer now has an option where they can purchase the Xeleris V software, and have it installed on a remote server.

New Clinical Applications

Exini Bone: The Exini Bone application is intended to be used with NM bone scans for the evaluation of adult male patients with bone metastases from prostate cancer. The application automatically segments bone lesions based on 2D whole-body planar bone scans and requires the user adjust and/or accept the final segmentation before proceeding. Exini Bone quantifies the selected lesions and provides a Bone Scan Index value as adjunct information related to the progression of disease.

Q.Liver: The Q.Liver application is a comprehensive application that provides processing, quantification, and multidimensional review of liver SPECT/CT exams for display and segmentation. The application provides the user with tools to calculate a therapeutic dose for Selective Internal Radiation Therapy (SIRT) treatment.

Enhancements to Clinical Applications

Q.Lung: The Q.Lung application provides processing, quantification, and multidimensional review for pulmonary scintigraphy for display and quantification of global and regional ventilation and perfusion on SPECT and SPECT/CT studies. The change introduced in Xeleris V is the introduction of an optional lung fissure automatic detection using a Deep Learning based model.

Myovation: The Myovation application is used for the evaluation of patients with suspected or known coronary artery disease. Myovation reconstructs and reformats the SPECT raw data and helps to analyze the myocardial perfusion and function in rest, stress, and viability studies. The change for Xeleris V is the introduction of a new optional, non-Al Automatic Recognition of Cardiac Structures (ARCS) algorithm for improved heart detection and orientation.

Q.Volumetrix MI: The Q.Volumetrix MI application is a compressive tool for processing and reading non-cardiac volumetric data, including NM SPECT and hybrid SPECT-CT, PET-CT, external CT/MR (i.e. CT/MR from a separate nonhybrid scan). The enhancement for Xeleris V is a new, optional, DL-based automatic kidney segmentation algorithm.

Q.Brain: The Q.Brain application is used to process, quantify, and review brain studies obtained that include SPECT , SPECT/CT , PET/CT, and MR datasets. Q.Brain processing includes reconstruction, reorientation, rigid registration, and non-rigid registration of single or sequential tomographic brain data. The enhancement introduced with Xeleris V is that the display of the CT images registered to the template is now included along with other currently available image types.

The system is intended for use by Nuclear Medicine (NM) or Radiology practitioners and referring physicians. The intended use of the system is to provide digital processing, review and reporting of medical images, including data display, quality control, image manipulation and quantification analysis, transfer, storage and printing capabilities.

The system operates in a variety of configurations. The hardware components may include computer workstations, communications devices, video monitors, data storage and hardcopy devices.

Software components provide functions for performing operations related to image display, manipulation, enhancements, analysis and quantification and can operate on dedicated workstations and client-server architectures.

AI/ML Overview

The provided document describes the Xeleris V Processing and Review System, which includes new deep learning (DL) based algorithms for lung fissure segmentation and kidney segmentation. While the document states that no clinical studies were required to support the determination of equivalence for the overall system, it does detail performance testing conducted specifically for these DL-based algorithms.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Stated Acceptance Criteria:

The document doesn't explicitly list quantitative acceptance criteria in the format of a table with thresholds (e.g., "Accuracy > 90%"). Instead, the acceptance criteria are implicitly defined by the rigor of the design control and performance testing, aiming to demonstrate substantial equivalence to a predicate device and ensure that the new features do not raise new safety or effectiveness questions.

The key overarching acceptance criteria are:

  • Successful completion of all design control testing per GE Healthcare's quality system and compliance with relevant standards (NEMA PS3.1 - 3.20, IEC62304).
  • No new safety questions or risks identified by testing.
  • Algorithm performance demonstrated through testing with representative clinical exams.
  • Acceptable scientific methods used to evaluate effectiveness.
  • Substantial equivalence demonstrated to the predicate device, meaning the proposed device is as safe and effective for its intended use, with no new intended use introduced.

Table of Acceptance Criteria and Reported Device Performance (Implicit):

Acceptance Criterion TypeSpecific Criteria Implicit in TextReported Device Performance (Summary from Text)
Overall System Safety & Effectiveness- No new safety questions/risks compared to predicate. - As safe and effective as predicate. - No new Intended Use.- "The testing did not raise any new safety questions or identify any new risks." - "Xeleris V is designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485." - "GE believes Xeleris V is of comparable type and substantially equivalent to the predicate and reference devices, and hence is safe and effective for its intended use."- "Xeleris V's Indications for Use do not create a new Intended Use."
Design Control & QA- Successful completion of all design control testing. - Compliance with relevant standards (NEMA PS3.1 - 3.20, IEC62304).- "Xeleris V and its clinical applications successfully completed all design control testing per GE Healthcare's quality system and also verified compliance with the relevant standards..." - "The software was developed, verified, and validated under GE Healthcare's QMS including software development lifecycle."
DL Algorithm Performance (Lung Fissure & Kidney)- "Algorithm clinical performance testing using test datasets of representative clinical exams where the ground truth and evaluations were performed by clinicians." - "The scientific methods used to evaluate the effectiveness of Xeleris V are acceptable and support the determination of substantial equivalence."- Performance testing for DL-based algorithms was "successfully completed."- "The test datasets were comprised of representative clinical exams that were manually segmented by experienced NM Physicists and Physicians and were used as ground truth." - NM Physicists and Physicians reviewed segmentation results and "scored them using a 5-point Likert scale." (No specific quantitative results from Likert scale are provided in this summary, but the general statement implies positive outcomes.)

Study Details (Performance Testing for DL-based Algorithms):

  1. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: The document refers to "test datasets of representative clinical exams" but "does not specify the exact number of cases or images" in these test sets for either the lung fissure or kidney segmentation algorithms.
    • Data Provenance: The document does not specify the country of origin of the data. It states the data comprised "representative clinical exams," implying real patient data. It does not explicitly state whether the data was retrospective or prospective, but "representative clinical exams" typically implies retrospective use of existing data for testing.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: The document states that the ground truth was "manually segmented by experienced NM Physicists and Physicians." It uses plural ("Physicists and Physicians") but does not specify the exact number of individuals.
    • Qualifications of Experts: They are described as "experienced NM Physicists and Physicians." No further details on their specific years of experience or board certifications are provided.
  3. Adjudication method for the test set:

    • The document states that the test datasets were "manually segmented by experienced NM Physicists and Physicians and were used as ground truth." It also mentions that these experts "reviewed the segmentation results and scored them using a 5-point Likert scale."
    • This implies a consensus ground truth was established by these experts through manual segmentation. However, it does not explicitly describe a formal adjudication method (e.g., 2+1, 3+1, or how disagreements were resolved if multiple experts were involved in the initial ground truth creation) beyond stating they created the ground truth and then evaluated the algorithm's output.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was explicitly stated. The testing described is performance testing of the algorithm itself, with human experts creating ground truth and then evaluating the algorithm's output. The document explicitly states: "Xeleris V did not require clinical studies to support the determination of equivalence."
    • Therefore, no effect size on human reader improvement with AI assistance is provided.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, performance testing of the DL algorithms appears to be a form of standalone evaluation. The algorithms generated segmentations, and these were then compared to the expert-generated ground truth and evaluated by experts. The text states: "...evaluates and demonstrates each algorithm's performance and uses test datasets of representative clinical exams." The output was then "reviewed the segmentation results and scored them." This indicates an algorithm-only performance evaluation against established ground truth.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for the DL algorithm performance testing was expert consensus/manual segmentation. Specifically, it was "manually segmented by experienced NM Physicists and Physicians."
  7. The sample size for the training set:

    • The document does not specify the sample size for the training set used for the deep learning models.
  8. How the ground truth for the training set was established:

    • The document does not explicitly describe how the ground truth for the training set was established. It only details the ground truth for the test set. It is common practice for training data ground truth to be established similarly to test data ground truth (e.g., via expert annotation), but this is not stated in the provided text.

{0}------------------------------------------------

September 25, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of a human figure, while the right side features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

GE Healthcare % Alexandra Lifshits Regulatory Affairs Program Manager 4 Hayozma Street Tirat Hacarmel, 30200 ISRAEL

Re: K201103

Trade/Device Name: Xeleris V Processing and Review Systems Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: August 13, 2020 Received: August 18, 2020

Dear Alexandra Lifshits:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K201103

Device Name Xeleris V Processing and Review System

Indications for Use (Describe)

The system is intended for use by Nuclear Medicine (NM) or Radiology practitioners and referring physicians for display, processing, archiving, printing, reporting and networking of NMI data, including planar scans (Static, Whole Body, Dynamic, Multi-Gated) and tomographic scans (SPECT, Gated SPECT, dedicated PET or Camera-Based-PET) acquired by gamma cameras or PET scanners. The system can run on dedicated workstation or in a server-client configuration. The NM or PET data can be coupled with registered and/or fused CT or MR scans, and with physiological signals in order to depict, localize, and/or quantify the distribution of radionuclical structures in scanned body tissue for clinical diagnostic purposes.

The DaTQUANT optional application enables visual evaluation of 1231-ioflupane (DaTscanTM) images. DaTQUANT Normal Database option enables quantification relative to normal population databases of 1231ioflupane (DaTscanTM) images. These applications may assist in detection of loss of functional dopaminergic neuron terminals in the striatum, which is correlated with Parkinson disease.

The O.Lung application may aid physicians in:

· Diagnosis of Pulmonary Embolism (PE), Chronic Obstructive Pulmonary Disease (COPD), Emphysema and other lung deficiencies.

• Assess the fraction of total lung function provided by a lobe or whole lung for Lung cancer resection requiring removal of an entire lobe, bilobectomy, or pneumonectomy.

The Q.Brain application allows the user to visualize and quantify relative changes in the brain's metabolic function or blood flow activity between a subject's images and controls, which may be resulting from brain functions in: · Epileptic seizures

· Dementia. Such as Alzheimer's disease, Lewy body dementia, Parkinson's disease with dementia, and frontotemporal dementia.

  • · Inflammation
  • · Brain death
  • · Cerebrovascular disease such as Acute stroke, Chronic and acute ischemia

· Traumatic Brain Injury (TBI)

When integrated with the patient's clinical and diagnostic information may aid the physician in the interpretation of cognitive complaints, neuro-degenerative disease processes and brain injuries.

The Alcyone CFR application allows for the quantification of coronary vascular function by deriving Myocardial Blood Flow (MBF) and then calculating Coronary Flow Reserve (CFR) indices on data acquired on PET scanners and on stationary SPECT scanners with the capacity for dynamic SPECT imaging. These indices may add information to physicians using Myocardial Perfusion Imaging for the diagnosis of Coronary Artery Disease (CAD).

The Exini Bone application is intended to be used with NM bone scans for the evaluation of adult male patients with bone metastases from prostate cancer. Exini Bone quantifies the selected lesions and provides a Bone Scan Index value as adjunct information related to the progression of disease.

The Q.Liver application provides processing, quantification, and multidimensional review of Liver SPECT/PET and CT images for display, segmentation, and a calculation of the SPECT 'liver to lung' shunt value and the patient's Body Surface Area (BSA) for use in calculating a therapeutic dose for Selective Internal Radiation Therapy (SIRT) treatment

{3}------------------------------------------------

using a user defined formula.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "ge" in a cursive font, enclosed within a blue circle. The circle is surrounded by a decorative, swirling pattern, also in blue. The background is plain white.

K201103

510(k) SUMMARY OF SAFETY AND EFFECTIVNESS

This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):

Date:April 23rd, 2020
Submitter:GE Medical Systems Israel, Functional Imaging (GE Healthcare)4 Hayozma StreetTirat Hacarmel, 30200, Israel
PrimaryContact:Alexandra LifshitsRegulatory Affairs Program ManagerGE Medical Systems Israel, Functional ImagingTel: +972-4-8563666 ext. 805Fax: +972-4-8577662email: alexandra.lifshits@ge.com
SecondaryContacts:John JaeckleChief Regulatory Affairs StrategistGE HealthcareTel: 262-424-9547email: john.jaeckle@ge.com
Device TradeName:Xeleris V Processing and Review System
Common /Usual Name:System, Image Processing, Radiological
ClassificationNames:21CFR 892.2050
ProductCode:LLZ
PredicateDeviceDevice Name510 (K) numberRegulation NºProduct CodeXeleris 4.0 Processing and Review WorkstationK153355892.2050LLZ

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are also some white swirls around the letters. The logo is simple and recognizable.

ReferenceDevice NameEXINI's aBSI
Devices510 (K) numberK191262
Regulation Nº892.2050
Product CodeLLZ
Device NameGE's Hepatic VCAR
510 (K) numberK133649
Regulation Nº892.1750
Product CodeJAK, LLZ
MarketedDevicesThe Xeleris V Processing and Review System is a modification to the predicatedevice, Xeleris 4.0 Processing and Review Workstation. It includes all of theclinical applications and features in the current production version of Xeleris 4.0and, introduces new clinical applications (i.e. Exini Bone, Q.Liver) andenhancements to previously cleared clinical applications (i.e. Q.Lung,Myovation, Q.Volumetrix MI, Q.Brain).
Xeleris V's Indications for Use remain the same as the predicate device's exceptfor additional indicates for the new clinical applications, Exini Bone and Q.Liver.Xeleris V's Indications for Use do not create a new Intended Use.
DeviceDescription:The Xeleris V Processing and Review Workstation is a Nuclear MedicineSoftware system that is designed for general nuclear medicine processing andreview procedures for detection of radioisotope tracer uptake in the patient'sbody, using a variety of individual processing applications orientated to specificclinical applications.
Xeleris V is a modification to the predicate device, Xeleris 4.0 Processing andReview Workstation. It includes all the clinical applications and features in thecurrent production version Xeleris 4.0 and introduces new clinical applicationsand enhancements to previously cleared clinical applications.
With Xeleris V, the customer now has an option where they can purchase theXeleris V software, and have it installed on a remote server.
New Clinical Applications
Exini Bone: The Exini Bone application is intended to be used with NM bonescans for the evaluation of adult male patients with bone metastases fromprostate cancer. The application automatically segments bone lesions based on2D whole-body planar bone scans and requires the user adjust and/or acceptthe final segmentation before proceeding. Exini Bone quantifies the selectedlesions and provides a Bone Scan Index value as adjunct information related tothe progression of disease.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three water droplets surrounding the circle. The logo is simple and recognizable, and it is associated with a well-known and established company.

  • · Q.Liver: The Q.Liver application is a comprehensive application that provides processing, quantification, and multidimensional review of liver SPECT/CT exams for display and segmentation. The application provides the user with tools to calculate a therapeutic dose for Selective Internal Radiation Therapy (SIRT) treatment.

Enhancements to Clinical Applications

  • Q.Lung: The Q.Lung application provides processing, quantification, and multidimensional review for pulmonary scintigraphy for display and quantification of global and regional ventilation and perfusion on SPECT and SPECT/CT studies. The change introduced in Xeleris V is the introduction of an optional lung fissure automatic detection using a Deep Learning based model.
  • Myovation: The Myovation application is used for the evaluation of patients with suspected or known coronary artery disease. Myovation reconstructs and reformats the SPECT raw data and helps to analyze the myocardial perfusion and function in rest, stress, and viability studies. The change for Xeleris V is the introduction of a new optional, non-Al Automatic Recognition of Cardiac Structures (ARCS) algorithm for improved heart detection and orientation.
  • Q.Volumetrix MI: The Q.Volumetrix MI application is a compressive tool for processing and reading non-cardiac volumetric data, including NM SPECT and hybrid SPECT-CT, PET-CT, external CT/MR (i.e. CT/MR from a separate nonhybrid scan). The enhancement for Xeleris V is a new, optional, DL-based automatic kidney segmentation algorithm.
  • Q.Brain: The Q.Brain application is used to process, quantify, and review brain studies obtained that include SPECT , SPECT/CT , PET/CT, and MR datasets. Q.Brain processing includes reconstruction, reorientation, rigid registration, and non-rigid registration of single or sequential tomographic brain data. The enhancement introduced with Xeleris V is that the display of the CT images registered to the template is now included along with other currently available image types.
  • The system is intended for use by Nuclear Medicine (NM) or Radiology Intended practitioners and referring physicians. The intended use of the system is to Use: provide digital processing, review and reporting of medical images, including data display, quality control, image manipulation and quantification analysis, transfer, storage and printing capabilities.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three water droplets surrounding the circle. The logo is simple and recognizable, and it is associated with a well-known company.

The system operates in a variety of configurations. The hardware components may include computer workstations, communications devices, video monitors, data storage and hardcopy devices.

Software components provide functions for performing operations related to image display, manipulation, enhancements, analysis and quantification and can operate on dedicated workstations and client-server architectures.

The system is intended for use by Nuclear Medicine (NM) or Radiology Indications practitioners and referring physicians for display, processing, archiving, for Use: printing, reporting and networking of NMI data, including planar scans (Static, Whole Body, Dynamic, Multi-Gated) and tomographic scans (SPECT, Gated SPECT, dedicated PET or Camera-Based-PET) acquired by gamma cameras or PET scanners. The system can run on dedicated workstation or in a server-client configuration.

The NM or PET data can be coupled with registered and/or fused CT or MR scans, and with physiological signals in order to depict, localize, and/or quantify the distribution of radionuclide tracers and anatomical structures in scanned body tissue for clinical diagnostic purposes.

The DaTQUANT optional application enables visual evaluation and quantification of 1231-ioflupane (DaTscanTM) images. DaTQUANT Normal Database option enables quantification relative to normal population databases of 123I-ioflupane (DaTscanTM) images. These applications may assist in detection of loss of functional dopaminergic neuron terminals in the striatum, which is correlated with Parkinson disease.

The Q.Lung application may aid physicians in:

  • . Diagnosis of Pulmonary Embolism (PE), Chronic Obstructive Pulmonary Disease (COPD), Emphysema and other lung deficiencies.
  • • Assess the fraction of total lung function provided by a lobe or whole lung for Lung cancer resection requiring removal of an entire lobe, bilobectomy, or pneumonectomy.

The Q.Brain application allows the user to visualize and quantify relative changes in the brain's metabolic function or blood flow activity between a subject's images and controls, which may be resulting from brain function alterations in:

  • · Epileptic seizures
  • . Dementia. Such as Alzheimer's disease, Lewy body dementia, Parkinson's disease with dementia, vascular dementia, and frontotemporal dementia.
  • Inflammation .

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. There are three water droplet shapes surrounding the circle. The logo is blue.

  • . Brain death
  • · Cerebrovascular disease such as Acute stroke, Chronic and acute ischemia
  • Traumatic Brain Injury (TBI)

When integrated with the patient's clinical and diagnostic information, Q.Brain application may aid the physician in the interpretation of cognitive complaints, neuro-degenerative disease processes and brain injuries.

The Alcyone CFR application allows for the quantification of coronary vascular function by deriving Myocardial Blood Flow (MBF) and then calculating Coronary Flow Reserve (CFR) indices on data acquired on PET scanners and on stationary SPECT scanners with the capacity for dynamic SPECT imaging. These indices may add information to physicians using Myocardial Perfusion Imaging for the diagnosis of Coronary Artery Disease (CAD).

The Exini Bone application is intended to be used with NM bone scans for the evaluation of adult male patients with bone metastases from prostate cancer. Exini Bone quantifies the selected lesions and provides a Bone Scan Index value as adjunct information related to the progression of disease.

Q.Liver application provides processing, quantification, The and multidimensional review of Liver SPECT/PET and CT images for display, segmentation, and a calculation of the SPECT 'liver to lung' shunt value and the patient's Body Surface Area (BSA) for use in calculating a therapeutic dose for Selective Internal Radiation Therapy (SIRT) treatment using a user defined formula.

  • Xeleris V's basic functionality for processing and reviewing Nuclear Medicine Technology: and their associated CT images is not changed from its predicate. This functionality includes manual and automatic segmentation to depict, localize, and/or quantify the distribution of radionuclide tracers and anatomical structures in patients for diagnostic purposes. Much of this functionality is designed for streamlined user workflows.
    Xeleris V introduces two new applications that are focused on specific clinical imaging scenarios, using this basic functionality Xeleris V also enhances three existing applications to improve their workflow efficiency using both DL-based and traditional algorithms.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script in the center. There are three water droplets or stylized swirls around the letters. The logo is simple and recognizable, and it is associated with a well-known and established company.

AttributePredicate Device: Xeleris 4.0Proposed Device: Xeleris V
GeneralWorkflowsXeleris 4.0's workflows useboth manual and automatedprocesses.Xeleris V's workflows continueuse both manual and automatedprocesses. Xeleris V also includestwo new, DL-based automaticsegmentation algorithms.
Use ofDeepLearningXeleris 4.0 does not containany DL processes/algorithms.Xeleris V introduces two new DL-based algorithms for:• Lung fissure segmentation asan alternative to the manualmethod.• Kidney segmentation as anenhancement to the non-DLautomatic segmentation.
HardwareNeeded toSupportXeleris V'ssoftwareXeleris 4.0 offered with a GEworkstation or as softwarethat can be loaded onto thecustomer's workstation thatmeets specification.In addition to the two options forthe predicate, with Xeleris V thesoftware can also be installed on aremote server.

The Xeleris V Processing and Review System has identical or equivalent technological characteristics as its predicate and reference devices. The changes and the different technological characteristics do not raise new or different questions of safety and effectiveness. The software was developed, verified, and validated under GE Healthcare's QMS including software development lifecycle.

Design Control Testing: Testing

Summary

Design Control Testing:

Xeleris V and its clinical applications successfully completed all design control testing per GE Healthcare's quality system and also verified compliance with the relevant standards (i.e. NEMA PS3.1 - 3.20, IEC62304). The testing did not raise any new safety questions or identify any new risks. Xeleris V is designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. The following quality assurance measures were applied to the development of the system:

  • Risk Analysis
  • · Requirements Reviews
  • · Design Reviews
  • · SW Development Lifecycle
  • Testing on Unit Level (Module Verification)

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a blue circle. There are three water droplets or splashes emanating from the top and sides of the circle, giving the logo a dynamic and fluid appearance. The color scheme is primarily blue and white.

  • · Integration Testing (System Verification)
  • · Performance Testing (Verification)
  • Safety testing (Verification)
  • · Simulated Use Testing (Validation)

The testing and their results do not raise different questions of safety and effectiveness than the predicate device. GE believes Xeleris V is of comparable type and substantially equivalent to the predicate and reference devices, and hence is safe and effective for its intended use. The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.

Summary of Additional Testing:

In addition to the verification and validation testing successfully completed as required by GE Healthcare's quality system, additional engineering and clinical performance testing was performed to provide the requisite data to demonstrate performance and substantiate claims. This testing included:

Engineering testing including clinical datasets and clinician participation.

  • · Demonstration of clinical outputs, workflow, and tools.
  • Scientific testing for algorithm performance and claims substantiation.
  • · Algorithm clinical performance testing using test datasets of representative clinical exams where the ground truth and evaluations were performed by clinicians.

Clinical Evaluation

Xeleris V did not require clinical studies to support the determination of equivalence. However, the performance testing for the DL-based algorithms for kidney segmentation and lung fissure segmentation evaluates and demonstrates each algorithm's performance and uses test datasets of representative clinical exams.

The test datasets were comprised of representative clinical exams that were manually segmented by experienced NM Physicists and Physicians and were used as ground truth. The NM Physicists and Physicians also performed the algorithm evaluation. They reviewed the segmentation results and scored them using a 5-point Likert scale.

The scientific methods used to evaluate the effectiveness of Xeleris V are acceptable and support the determination of substantial equivalence.

Substantial The Substantial Equivalence of the proposed device has been demonstrated by:

  • Equivalence
    • Review of the proposed Indications for Use shows that they are substantially equivalent to the predicate device's, also with consideration of the reference devices. Xeleris V's Indications for Use do not create a new Intended Use.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a swirling pattern, also in blue. The logo is simple and recognizable, and it is often used to represent the company's brand.

  • The device description and the comparison of device characteristics show that Xeleris V's has identical or equivalent technological characteristics as its predicate and reference devices.
  • The testing (engineering and clinical) demonstrates the effectivity of Xeleris V for its intended purpose.
  • . The changes and the different technological characteristics do not raise new or different questions of safety and effectiveness. The proposed device is as safe and effective as the legally marketed predicate device as demonstrated by the:
    • o Software verification and validation for a Moderate level of concern, without unexpected results;
    • o Development under GE's quality management system, design control activities including risk management;
    • o Device labeling; and
    • o Engineering and clinical testing without unexpected results.
  • Xeleris V's Indications for Use do not create a new Intended Use. Xeleris V has Conclusion: identical or equivalent technological characteristics as its predicate device.

Xeleris V was developed under GE Healthcare's quality system. Design verification testing, along with bench and clinical testing demonstrate that Xeleris V is substantially equivalent and hence as safe and as effective as the legally marketed predicate device. GE's quality system's design, verification, and risk management processes did not identify any new hazards, unexpected results, or adverse effects stemming from the changes to the predicate.

Based on development under GE Healthcare's quality system, the successful verification, engineering bench and clinical testing, GE Healthcare believes that Xeleris V is substantially equivalent to the predicate device and hence is safe and effective for its intended use.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).