K Number
K201074
Date Cleared
2020-09-03

(134 days)

Product Code
Regulation Number
876.5990
Panel
GU
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Delta III Lithotripter is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.

Device Description

The Delta III Lithotripter is a modular urological work station designed for extracorporeal shock wave lithotripsy ("ESWL") and for diagnostic and therapeutic procedures usual in urology. The Delta III is composed of the following modules: Basic Unit with integrated X-ray C-arm and Therapy Arm with camera for Shockwave Treatment; Patient Table; Control Desk/Image Storage (UIMS).

AI/ML Overview

This document is a 510(k) summary for the Delta III Lithotripter, detailing minor changes to an existing device. As such, it focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with acceptance criteria and a human-in-the-loop performance evaluation. The information provided primarily consists of engineering and software validation.

Here's an analysis based on the provided text, addressing your questions to the extent possible:

1. Table of acceptance criteria and the reported device performance

The document does not present a formal table of acceptance criteria for clinical performance in the way one might expect for a new diagnostic or AI device. Instead, the "acceptance criteria" are implied by compliance with standards and functional validation tests, with the reported performance being "does not impact the performance" and "equal to or better than the predicate."

Acceptance Criteria (Implied)Reported Device Performance
Standards Compliance:
IEC 60601-1 (Electrical safety)Compliant
IEC 60601-1-2 (EMC)Compliant
IEC 60601-1-3 (Radiation protection)Compliant
IEC 60601-2-36 (Lithotripsy safety)Compliant
IEC 60601-2-28 (X-ray tube assemblies)Compliant
IEC 60601-2-54 (X-ray equipment safety)Compliant
IEC 62366-1 (Usability engineering)Compliant
IEC/TR 62366-2 (Usability guidance)Compliant
Bench Testing (Functional Validation):
New ultrasound device provides visual and accurate images for stone visualizationPerformance assured; provides visual and accurate images
New camera image quality is equal to or better than the predicateImage quality equal to or better than the predicate
UIMS software with AGFA package does not impact equipment functioning and processes X-ray imagesDoes not impact functioning; provides ability to further process X-ray images
Electrical safety, EMC, functional usability are fully addressedDemonstrated compliance with standards and equivalent performance to predicate
No unanticipated new risks identifiedNo unanticipated new risks identified

2. Sample size used for the test set and the data provenance

The document describes bench testing and standards compliance, not a clinical test set with patient data. Therefore, there is no mention of a "sample size for the test set" in the context of patient data, nor its provenance (country of origin, retrospective/prospective). The tests mentioned are engineering and software validation tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This type of information is not applicable to the submission described. The described tests are about technical performance, safety, and functionality, not about expert interpretation of medical images or patient outcomes. Ground truth in this context would refer to engineering specifications or validated functional requirements.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable as the submission describes engineering and software validation, not a clinical study requiring adjudication of diagnoses or findings.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done or is mentioned. The submission is for minor changes to an existing lithotripter, focusing on hardware component replacements and software updates, not AI integration for improved human reading.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device (Delta III Lithotripter) is a medical device for shock wave lithotripsy. While it includes image processing software (UIMS with AGFA MUSICA), the evaluation described is for the functional performance of these components within the overall system, not a standalone algorithm being evaluated for diagnostic accuracy without human involvement. The AGFA MUSICA software enhances X-Ray image processing, but it's part of the lithotripter system for visual guidance during procedures, not a diagnostic AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the described tests, the "ground truth" would be established by:

  • Engineering specifications and regulatory standards: For electrical safety, EMC, radiation protection, and usability.
  • Predicate device performance: For the ultrasound image quality and camera image quality, the ground truth for comparison is the performance of the components being replaced.
  • Functional requirements: For the UIMS software validation, the ground truth is that the software should perform its intended function (e.g., image processing, PACS connection) without negatively impacting other equipment functions.

There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for these specific validation tests.

8. The sample size for the training set

Not applicable. This submission does not involve a machine learning or AI model that requires a training set in the conventional sense. The UIMS software update involves incorporating the AGFA MUSICA package, which is likely a pre-existing, validated image processing software, not a custom-trained model for this specific application.

9. How the ground truth for the training set was established

Not applicable, as there is no mention of a training set.

§ 876.5990 Extracorporeal shock wave lithotripter.

(a)
Identification. An extracorporeal shock wave lithotripter is a device that focuses ultrasonic shock waves into the body to noninvasively fragment urinary calculi within the kidney or ureter. The primary components of the device are a shock wave generator, high voltage generator, control console, imaging/localization system, and patient table. Prior to treatment, the urinary stone is targeted using either an integral or stand-alone localization/imaging system. Shock waves are typically generated using electrostatic spark discharge (spark gap), electromagnetically repelled membranes, or piezoelectric crystal arrays, and focused onto the stone with either a specially designed reflector, dish, or acoustic lens. The shock waves are created under water within the shock wave generator, and are transferred to the patient's body using an appropriate acoustic interface. After the stone has been fragmented by the focused shock waves, the fragments pass out of the body with the patient's urine.(b)
Classification. Class II (special controls) (FDA guidance document: “Guidance for the Content of Premarket Notifications (510(k)'s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi.”)