K Number
K200984
Date Cleared
2020-05-27

(43 days)

Product Code
Regulation Number
882.1350
Panel
NE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Rhythmlink International Subdermal Needle Electrodes are intended for use with recording, monitoring and stimulation equipment for the purpose of recording of biopotential signals. Examples include: Electromyography (EMG), Electroencephalography (EEG), and Nerve potential signals. The electrodes are sterile and for single patient use.

Device Description

Guardian Needle™ Electrodes are a sterile, single-use device. The electrodes are applied during the study of biopotentials such as electromyography (EMG), electroencephalography (EEG), nerve conduction and stimulation/response. The electrodes are invasive as they are placed subcutaneously or in contact with nerve or muscle tissue. The needle is housed inside a sheath until application and secured to the patient during use with adhesive tape.

AI/ML Overview

The Rhythmlink International Guardian Needle™ Electrode (subject device) is compared to the predicate device, Subdermal Needle Electrodes (K022914), to demonstrate substantial equivalence. The document primarily focuses on non-clinical testing.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaSubject Device (Guardian Needle™ Electrode) Performance
Sterilization & Residuals- Compliance with TIR 28:2009 guidance.Subject device determined to fall under previously validated EtO sterilization cycle.
- EtO and Ethylene Chlorohydrin (ECH) levels after 24-hour aeration below maximum limits described in ANSI/AAMI/ISO 10993-7:2008/(R) 2012.Reevaluation confirmed low levels of EtO and ECH.
Biocompatibility- Biological safety for intact skin contact (≤24 hours) as per ISO 10993 standards and risk management framework.Biocompatibility evaluation establishes biological safety for intact skin contact for ≤24 hours.
Functional Performance- Electrical ContinuityPassed predetermined acceptance criteria.
- Adhesion Testing (for adhesive tape)Passed predetermined acceptance criteria.
Technological Equivalence to Predicate- Intended UseIdentical to predicate.
- Anatomical Site(s)Identical to predicate.
- Environment UsageIdentical to predicate.
- Electrode MaterialIdentical to predicate.
- Electrode LengthIdentical to predicate.
- Electrode DiameterIdentical to predicate.
- Leadwire MaterialIdentical to predicate.
- Leadwire LengthIdentical to predicate.
- ConnectorIdentical to predicate.
- Operation of Applying the DeviceIdentical to predicate.
- Prescription UseIdentical to predicate.
- Targeted ProceduresIdentical to predicate.
- Compatibility with other devicesIdentical to predicate.
- PackagingIdentical to predicate.
- Sterilization MethodIdentical to predicate.
- Single Patient UseIdentical to predicate.
- Electrical SafetyIdentical to predicate (Connectors comply with IEC 60601-1 (1988) sub clause 56.3(c) per CFR 898.12).
- Mechanical SafetyIdentical to predicate (Leadwires soldered to electrode using Tin/Silver solder and covered with heat shrink).
- Duration of useIdentical to predicate (≤24 hours).
- StandardsIdentical to predicate (IEC 60601-1-1: 1988/a1: 1991/A2 :1995 § 56.3(c) per CFR 898.12, DIN 42802, ISO 14971, ISO 15223-1).

Note on Differences: The subject device includes a needle sheath (thermoplastic polyester elastomer sheath) and adhesive tape, which differ from the predicate device (PVC tube for sheath, no adhesive tape). These differences were assessed through testing and determined not to adversely impact safety or effectiveness.

2. Sample size used for the test set and the data provenance

The document does not specify exact sample sizes for the functional performance benchtop tests (Electrical Continuity and Adhesion Testing). However, it states that "The test methods were identical to those used to assess the predicate device."

For sterilization, reevaluation of EtO residuals was performed, but the sample size is not explicitly stated.

The data provenance is from non-clinical benchtop testing conducted by Rhythmlink International, LLC. The document does not indicate any country of origin for the data that would suggest external sources. All testing appears to be retrospective relative to the submission date, as it's part of a 510(k) premarket notification.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The document describes non-clinical benchtop tests and compliance with recognized standards. There isn't an explicit "test set" in the context of expert review for establishing ground truth, as the criteria are based on engineering and biocompatibility standards.

4. Adjudication method for the test set

This information is not applicable as the document does not describe a test set requiring expert adjudication for ground truth (e.g., image interpretation). The evaluation relies on established engineering and biocompatibility testing methodologies.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission explicitly states "No Clinical Tests were conducted as referenced in 21 CFR 807.92(b)(2)" and "No Clinical Tests were conducted as referenced in 21 CFR 807.92(b)(3)."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable as the device is a physical invasive medical electrode, not a software algorithm or AI system.

7. The type of ground truth used

The "ground truth" for demonstrating the device meets acceptance criteria is primarily based on:

  • Compliance with recognized industry standards: ANSI/AAMI/ISO 10993-7:2008/(R) 2012 for EtO residuals, ISO 10993 for biocompatibility, IEC 60601-1-1: 1988/a1: 1991/A2 :1995 § 56.3(c) per CFR 898.12, DIN 42802, ISO 14971, ISO 15223-1, and TIR 28:2009 Guidance for sterilization.
  • Predicate device characteristics: Substantial equivalence is established by demonstrating the subject device has identical technological characteristics, intended use, and performance to a legally marketed predicate device, with differences (needle sheath, adhesive tape) assessed not to adversely impact safety or effectiveness.
  • Benchtop test results: Electrical continuity and adhesion testing against predetermined acceptance criteria.

8. The sample size for the training set

This question is not applicable as the device is not an AI/ML product that undergoes training.

9. How the ground truth for the training set was established

This question is not applicable as the device is not an AI/ML product that undergoes training.

§ 882.1350 Needle electrode.

(a)
Identification. A needle electrode is a device which is placed subcutaneously to stimulate or to record electrical signals.(b)
Classification. Class II (performance standards).