(195 days)
IPS CaseDesigner is indicated for use as a software and image segmentation system for the transfer of imaging information from a scanner such as a CT scanner. It is also indicated to support the diagnostic and treatment planning process of craniomaxillofacial procedures. IPS CaseDesigner facilitates the service offering of individualized surgical aids.
IPS CaseDesigner has specific functionalities to visualize the diagnostic information, e.g. from CT-imaging, to perform specific measurements in the image data and to plan surgical actions in order to support the diagnostic and treatment planning process. Based on the diagnostic and planning data, the IPS design service can offer individualized surgical aids.
The IPS CaseDesigner 2.0 is a software and image segmentation system for the transfer of imaging information from a CT scanner, indicated to support the diagnostic and treatment planning process of craniomaxillofacial procedures and facilitate the service offering of individualized surgical aids.
Here's an analysis of its acceptance criteria and the supporting study:
1. Table of Acceptance Criteria and Reported Device Performance
The provided documentation does not explicitly list quantitative acceptance criteria in a dedicated table format. The "Differences" section, however, outlines new functionalities and improvements compared to the predicate device, implying that the performance of these new features needed to be validated. The "Performance Data" section states that the device was verified and validated and that "requirements for the features have been met."
Based on the text, the key performance aspects that were likely evaluated for the new features are:
| Feature/Functionality | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Additional Segmental Maxillary Osteotomies | Accurate simulation and planning of segmental maxillary osteotomies (Split, Y-cut, H-cut). | Allows planning of segmental maxillary osteotomies, enabling different types of virtual cuts and bone fragment movement. |
| Intraoral Surface Scan Data Import | Ability to accurately import and utilize intraoral surface scan data (dental casts) for occlusal information. | Can use intraoral surface scans for detailed occlusal information. |
| Osteosynthesis Plates Selection/Ordering | Correct display and selection of osteosynthesis plates, and accurate generation of a list for ordering. | Possible to export a list of specific plates selected from an available list. |
| 3D Cephalometry | Accurate setting of 3D landmarks and planes, correct calculation of measurements, and automatic updates based on planning. | Supports 3D Cephalometry with setting landmarks, measurements, and automatic updates. Virtual lateral and frontal cephalograms are calculated. |
| New Algorithm for Virtual Occlusion | Accuracy and efficiency comparable to the manual workflow of placing and digitizing dental casts. | Concluded to be as efficient and accurate as the manual workflow; provides the same level of accuracy and reliability. |
| Splint Visualization | Accurate generation and visualization of the surgical splint directly within the 3D workspace. | Allows generation and visualization of surgical splint directly in the 3D workspace. |
| 3D Photo Mapping (Soft Tissue Simulation) | Improved visualization of soft tissue simulation with the ability to add "real-world" textures. | Improved visualization of soft tissue simulation. |
| Operating System Compatibility | Compatibility and validated performance on specified operating systems (Windows 7, 10; Mac OS Catalina, Mojave, High Sierra). | Tested and validated on all specified systems. |
2. Sample Size for the Test Set and Data Provenance
The document does not specify the sample size used for the test set for any of the performance evaluations. It also does not mention the data provenance (e.g., country of origin, retrospective or prospective nature) for any specific testing data.
3. Number of Experts and Qualifications for Ground Truth
The document does not specify the number of experts or their qualifications used to establish ground truth for the test set. For the "New algorithm for virtual occlusion and occlusion alignment," it states "This algorithm was validated and it was concluded that it is as efficient and as accurate as the manual workflow..." but does not detail who performed this validation or how the "manual workflow" ground truth was established by experts.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method used for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not describe a multi-reader multi-case (MRMC) comparative effectiveness study. There is no mention of human readers improving with AI assistance or without. The device functions as a planning tool for clinicians, and the focus of the validation is on the accuracy and functionality of the software's new features.
6. Standalone Performance
Yes, the study primarily describes the standalone performance of the algorithm and software features. The validation of the "New algorithm for virtual occlusion and occlusion alignment" specifically states that it was concluded to be "as efficient and as accurate as the manual workflow," which implies a standalone assessment of the algorithm's output against a defined standard. The other listed features also focus on the core functionality of the software itself.
7. Type of Ground Truth Used
The type of ground truth used is not explicitly stated in detail for each feature. However, based on the description, it can be inferred that:
- Expert Consensus/Manual Workflow: For the "New algorithm for virtual occlusion and occlusion alignment," the ground truth likely involved a manual workflow of placing dental casts in occlusion and digitizing them, implicitly representing an expert-derived or standard clinical practice ground truth. The algorithm's output was compared against this.
- Engineering/Design Specifications: For functionalities like "Additional segmental maxillary osteotomies," "Osteosynthesis plates," "3D Cephalometry," and "Splint visualization," the ground truth for validation likely involved verifying the software's output against pre-defined engineering specifications, anatomical accuracy, and expected physiological movements/measurements as determined by medical and software experts during the design and development phases.
8. Sample Size for the Training Set
The document does not specify the sample size for the training set. As this device is a planning and visualization tool, rather than a deep learning model requiring extensive training data, explicit mention of a "training set" might not be applicable in the same way as for diagnostic AI algorithms. However, if any machine learning components were used (e.g., for occlusion alignment), this information is not provided.
9. How the Ground Truth for the Training Set was Established
Since the document does not specify a training set or its sample size, it does not describe how ground truth for a training set was established.
{0}------------------------------------------------
October 8, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nobel Biocare AB % Vladislavs Mihailovs Regulatory Affairs Manager Medicim NV Stationsstraat 102 Mechelen, Antwerp 2800 BELGIUM
Re: K200810
Trade/Device Name: IPS CaseDesigner 2.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: August 26, 2020 Received: September 11, 2020
Dear Vladislavs Mihailovs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K200810
Device Name IPS CaseDesigner 2.0
Indications for Use (Describe)
IPS CaseDesigner is indicated for use as a software and image segmentation system for the transfer of imaging information from a scanner such as a CT scanner. It is also indicated to support the diagnostic and treatment planning process of craniomaxillofacial procedures. IPS CaseDesigner facilitates the service offering of individualized surgical aids.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
5. 510(k) Summary
I. Submitter
Submitted by: Nobel Biocare c/o Medicim NV Stationsstraat 102 Mechelen, Belgium 2800 Establishment Registration No. 3005739655 Telephone: +32 15 44 32 00 +32 15 44 32 09 Facsimile: Email: vladislavs.mihailovs@nobelbiocare.com
Submitted for: Nobel Biocare AB Vastra Hamngatan 1 Göteborg, SE-411 17, Sweden Establishment Registration No. 9611992
Date Prepared: 2 March 2020
II. Device
Name of Device: IPS CaseDesigner 2.0 Common or Usual Name: Picture Archiving and Communications System Classification Name: System, Image Processing, Radiological (21 CFR 892.2050) Regulatory Class: II Product Code: LLZ
III.Predicate Device
Substantial equivalence is claimed to the following devices:
Predicate Device: IPS CaseDesigner Predicate 510(k): K161634 Compay: NobelBiocare AB
Reference device: KLS Martin Individual Patient Solutions (IPS) Planning System Reference device 510(k): K181241 Company: KLS-Martin L.P.
{4}------------------------------------------------
IV. Device Description
IPS CaseDesigner has specific functionalities to visualize the diagnostic information, e.g. from CT-imaging, to perform specific measurements in the image data and to plan surgical actions in order to support the diagnostic and treatment planning process.
Based on the diagnostic and planning data, the IPS design service can offer individualized surgical aids.
V. Indications for Use
IPS CaseDesigner is indicated for use as a software and image segmentation system for the transfer of imaging information from a scanner such as a CT scanner. It is also indicated to support the diagnostic and treatment planning process of craniomaxillofacial procedures. IPS CaseDesigner facilitates the service offering of individualized surgical aids.
VI. Comparison of Technological Characteristics
The subject and predicate devices are software based surgical planning tools which allow for transfer of medical images, and creation of 3D models.
Summary comparison of technological similarities and differences
IPS CaseDesigner 2.0 and the previous cleared IPS CaseDesigner 1.1 share following characteristics:
- Intended Use/Indications for Use ●
- Clinical use intended to support the diagnostic and treatment planning process of . craniomaxillofacial procedures
- Clinical image data import, visualization, distance and angular measurements ●
- 3D Surgical Model creation from CBCT scan data ●
- Soft tissue simulation
- Surgical splint calculation for a planned dental occlusion and export for centralized ● manufacturing at the legal manufacturer - KLS Martin
- Supported osteotomies: Le Fort I, Sagittal Split, Ramus, Chin ●
IPS CaseDesigner 2.0 is different from the previously cleared IPS CaseDesigner 1.1 version as follows:
- . 3D model generation based on imported surface scan data (dental casts)
- Newly supported osteotomies: Segmental (Split, Y-cut, H-cut) ●
- Osteosynthesis Plates users can select and order desired plates for their planning from ● the legal manufacturer – KLS Martin
- 3D cephalometry and cephalometric analysis based on user-picked points
{5}------------------------------------------------
- New algorithm for virtual occlusion and occlusion alignment an alternative to manual . workflow, where the software derives dental occlusion automatically, based on the points indicated by users. It is possible to edit the occlusion after automatic alignment based on a distance map if a user desires so.
A comparison of the subject and predicate devices is provided in the table below.
{6}------------------------------------------------
| Criteria | IPS CaseDesigner 2.0 (subject device) | IPS CaseDesigner 1.1 (predicate device) | KLS Martin Individual Patient Solutions (IPS) Planning System | Comments |
|---|---|---|---|---|
| Indications for Use Statement | IPS CaseDesigner is indicated for use as a software and image segmentation system for the transfer of imaging information from a scanner such as a CT scanner. It is also indicated to support the diagnostic and treatment planning process of craniomaxillofacial procedures. IPS CaseDesigner facilitates the service offering of individualized surgical aids. | IPS CaseDesigner is indicated for use as a software and image segmentation system for the transfer of imaging information from a scanner such as a CT scanner. It is also indicated to support the diagnostic and treatment planning process of craniomaxillofacial procedures. IPS CaseDesigner facilitates the service offering of individualized surgical aids. | The KLS Martin Individual Patient Solutions (IPS) Planning System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the IPS Planning System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, guides, splints, and case reports for use in maxillofacial surgery. The IPS Planning System is also intended as a pre-operative software tool for simulating/ evaluating surgical treatment options | Same |
| Clinical Use | Cranio-maxillofacial and orthognathic treatment | Cranio-maxillofacial and orthognathic treatment | Cranio-maxillofacial and orthognathic treatment | Same |
| Classification code | LLZ | LLZ | DZJ, LLZ | Same |
| Image Import | DICOM data format from CT/ CBCT scanner3D models and surface scan data in generic open file format (STL) | DICOM data format from CT/ CBCT scanner3D models in generic open file format (STL) | DICOM data format from CT/ CBCT scanner3D models in generic open file format (STL) | Different - see discussion below |
| Software output | Encrypted STL file with design of thefinal splint for centralizedmanufacturing at KLS Martin | Encrypted STL file with design of thefinal splint for centralizedmanufacturing at KLS Martin | N.A. | Same |
| Software Functions | 2D and 3D Visualization | 2D and 3D Visualization | 2D and 3D Visualization | Same |
| Distance and angle measurements | Distance and angle measurements | Distance and angle measurements | Same | |
| Bone fragments can be moved in the 3Dspace (translations and rotations) inorder to plan the ideal post-operativeposition of each fragment. | Bone fragments can be moved in the3D space (translations and rotations) inorder to plan the ideal post-operativeposition of each fragment. | Bone fragments can be moved inthe 3D space (translations androtations) in order to plan theideal post-operative position ofeach fragment. | Different –seediscussionbelow | |
| Supported osteotomies :Le Fort I, Sagittal Split, Ramus, Chin,Segmental (Split, Y-cut, H-cut) | Supported osteotomies :Le Fort I, Sagittal Split, Ramus, Chin | Supported osteotomies :Le Fort I, Sagittal Split, Ramus,Chin | ||
| 3D Surgical Model creation fromCBCT scan data and STL file | 3D Surgical Model creation fromCBCT scan data | 3D Surgical Model creation fromCBCT scan data | Different –seediscussionbelow | |
| Soft tissue simulation, 3D photomapping | Soft tissue simulation | Soft tissue simulation | Different -seediscussionbelow | |
| Based on the created orthognathic planthe user can export surgical splint orderfiles. These files can be used tocalculate and produce orthognathicsplints in the production backend | Based on the created orthognathic planthe user can export surgical splintorder files. These files can be used tocalculate and produce orthognathicsplints in the production backend | Based on the created orthognathicplan the user can export surgicalsplint order files. These files canbe used to calculate and produceorthognathic splints in theproduction backend | Same | |
| Osteosynthesis Plates & IPS Gate: theuser can create a list of requiredosteosynthesis plates based on a createdsurgical plan. | N.A. | N.A. | Different –seediscussionbelow | |
| 3D CephalometryDifferent types of 2D/3Dmeasurements (distances, angles,coordinates, etc.) 2D/3D visualization Take snapshots to clipboard Create/export reports Perform cephalometric analysis onsurgery models.Remap landmarks Visualize a set of keymeasurements Check pre-operative and planningvalues | N.A. | N.A. | Different –seediscussionbelow | |
| New algorithm for occlusionalignment New algorithm for virtualocclusion | N.A. | N.A. | Different –seediscussionbelow | |
| Operating systemrequirements | Windows 64-bit (Windows 7, 10)Mac OS Catalina, Mojave, High Sierra | Windows 64-bit (Windows 7, 8, 10)Mac OS X (Yosemite, El Capitan) | Windows 64-bit (Windows 7, 8,10)Mac OS X (Yosemite, ElCapitan) | Different –seediscussionbelow |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Discussion
Similarities:
The Indications for Use statement between the subject and predicate devices are equivalent. Both the subject and predicate devices are standalone software medical devices intended for the transfer of medical images and to support diagnostics and treatment planning of craniomaxillofacial procedures.
Both softwares utilize a graphic user interface with a large 3D based main window, several visualization and image processing tools and implement a system of wizards that guide the user through all steps of workflows.
Both software devices are capable of simulating surgery planning and the desired post-operative situation of bone fragments, as well as calculating of final splints in an STL format for the desired dental occlusion and exporting these files for backend production at the legal manufacturer – KLS Martin.
The output of subject device as well as the predicate device is in form of an encrypted STL file which is transferred via IPS Gate to ensure that only KLS Martin is able to receive this file and un-encrypt it for final quality checks and production at their designated centralized manufacturing facilities. The subject device allows export only to legal manufacturer of final devices based on their 510(k) clearance.
Differences:
Additional osteotomies
This new version of IPS CaseDesigner allows to plan the orthognathic surgery including additional segmental maxillary osteotomies:
Segmental maxillary osteotomies can be simulated, and orthognathic surgery can be . virtually performed. In v1.1 only standard LeFort I osteotomies for the maxilla could be virtually planned for orthognathic surgery simulation. The new version of the IPS CaseDesigner supports planning of segmental maxillary osteotomies, which means allowing different types of virtual cuts to the maxilla, and simulate the movement of these bone fragments prior to surgery
Intraoral surface scans can be used as well for obtaining detailed occlusal information. In . v1.1only CBCT scan data of the plaster models could be used for obtaining this detailed occlusal information.
Osteosynthesis plates are used to secure bone fragments after orthognathic surgery. Now . it is possible to export the list of specific plates that the user selects from the available osteosynthesis plate list.
These differences between the predicate device and subject device are existing functionality extensions.
Cephalometry
This new version of IPS CaseDesigner allows support of Cephalometry in 3D planning, with possibility to set relevant 3D landmarks and planes, and measurements that are updating based on the planning. A number of cephalometric measurements tools have been included in the
{10}------------------------------------------------
software, allowing the user to perform a cephalometric analysis to gauge the size and spatial relationships of the teeth, jaws, and cranium. The workflow consists of two main steps: first a user needs to indicate all required cephalometric landmarks, next a set of pre-defined measurements which are dependent on these landmarks, are automatically calculated. For indicating the landmarks, special views were created.
Moreover, a virtual lateral and frontal cephalogram are calculated based on the provided CBCT data. A user can select from a list of predefined measurements based on literature, which measurements should be included in the analysis. Based on this measurement selection, the user will be asked for a set of linked landmarks. When any landmark position is updated (example: by performing a surgical movement), measurement values are updated, and the delta is shown.
These differences between the predicate device and subject device are existing visualization extensions for analysis of aesthetic outcome, which do not affect safety and efficacy of software.
Splint visualization
This software version focuses on an improvement of existing splint calculation algorithm, allowing to visualize the user to generate and visualize surgical splint directly in the 3D workspace (scene).
These differences between the predicate device and subject device are existing visualization extensions for already existing calculations, which do not affect safety and efficacy of software.
Virtual occlusion and occlusion alignment
In the version 2.0 of IPS CaseDesigner (the subject of this 510(k) pre-market application) an alternative workflow of virtual occlusion was introduced. This is a more guided / automated approach by which corresponding points between upper and lower arch can be placed by the user on the imported STL type file
Based on these correspondences the software automatically calculates the target dental occlusion. To evaluate the occlusion an occlusiogram (distance map) can be visualized.
Finally, a user can virtually grind the dental model to eliminary contacts. It is possible to edit the desired dental occlusion at any stage of treatment planning.
Next to the virtual occlusion, the user can also derive a target occlusion by manually placing physical models in occlusion and digitizing the occlusal position of these models. This workflow was further improved to ensure that the occlusal position can be calculated in almost any case. Due to new algorithm introduced for virtual occlusion which is an alternative workflow for the existing indications for use, but an automated one, this triggered new 510(k) application for a major change in software.
This algorithm was validated and it was concluded that it is as efficient and as accurate as the manual workflow of the user placing dental casts in occlusion and digitizing them prior to import to IPS CaseDesigner 2.0. This algorithm provides the same level of accuracy and reliability as the manual workflow.
{11}------------------------------------------------
These differences between the predicate device and subject device are an introduction of an alternative workflow for the same intended clinical outcome - placing 3D model into dental occlusion for the purpose of treatment planning and surgical splint calculation.
Minor differences:
Another difference is 3D photo mapping, which is related only to an improved visualization of soft tissue simulation - an ability to add "real-world" textures to the model.
Different operating systems are currently supported by the IPS CaseDesigner, due to release of new versions of these operating systems during the lifecycle of IPS CaseDesigner. IPS CaseDesigner was tested and validated on all of these systems as per Appendix 10 and 11 of this submission.
Reference Device:
The reference device KLS Martin Individual Patient Solutions (IPS) Planning System is added to the comparison of technological similarities and differences table for the purpose of reflecting that non-clinical testing for the final physical devices (splints) was performed under K181241 by the legal manufacturer KLS Martin and this testing is not being included in current 510(k) premarket notification.
VII. Performance Data
IPS CaseDesigner is designed and manufactured under the Quality System Regulations as outline in 21 CFR § 820 and ISO 13485:2016 Standards. This device is in conformance with the applicable parts of EN IEC 62304:2006 standards. Design Control Activities, including risk management following the ISO 14971:2012, verification/validation testing, were conducted and are included in this submission.
The performance of the subject device was verified and validated following the guidance provided in FDA Guidance General Principles of Software Validation. This documentation includes testing which demonstrates that the requirements for the features have been met. Software documentation for Moderate Level of Concern and description of respective V&V activities, per the FDA guidance document. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005, is also included as part of this submission.
Software Validation
Software verification and validation testing was conducted on the subject device and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
{12}------------------------------------------------
VIII. Conclusion
Based on the comparison of the intended use, the features and workflows, the user interface, the technical characteristics, and based on the software verification activities described in this submission, IPS CaseDesigner 2.0 is found to be substantially equivalent to the identified Predicate Device.
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).