(153 days)
The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation on nonresectable liver tumors.
The Emprint™ Ablation System is not intended for use in cardiac procedures
The Emprint™ Ablation System with Thermosphere™ Technology, Covidien's microwave ablation system, was released for commercial distribution in the United States in April 2014.
The 510(k) Cleared Emprint™ Ablation System consists of the following components:
-
- Emprint™ Ablation Generator (2450 MHz)
-
- Emprint™ Percutaneous Antenna (sterile, single use)
-
- Emprint™ Ablation Reusable Cable
-
- Emprint™ Ablation Pump
The system also includes the following optional equipment/accessories:
-
- Emprint™ Ablation Cart (with Isolation Transformer)
-
- Ablation Footswitch
-
- Remote Temperature Probe (sterile, single use)
The current Emprint™ Ablation System utilizes a 2450 MHz 100W generator (CAGEN1) to deliver power to a dedicated single microwave ablation antenna. The optional cart, Emprint™ Ablation Cart is intended to be used with the Emprint™ Ablation Generator.
The new system component, Emprint™ HP Ablation Generator (CAGENHP) [subject device] is similar to the device, Emprint™ Ablation Generator (CAGEN1) [predicate device] as described in K193232 with two exceptions. The new system component, Emprint™ HP Ablation Generator (CAGENHP), will extend the power setting range from 5 to 100W (Emprint™ Ablation Generator - CAGEN1) to 5 to 150W. The increased power will allow clinicians to create larger ablation zones when ablating soft tissue lesions, compared to use of the system with the existing 100-watt microwave generator (CAGEN1). Further the Emprint™ HP Ablation Generator (CAGENHP) is composed of analog and digital circuits with software whereas the predicate Emprint™ Ablation Generator (CAGEN1), is composed of analog and digital circuits with no software or firmware. The Emprint™ HP Generator will be compatible with all current Emprint™ ablation system components and accessories except the cart. The new optional Emprint™ HP Cart (CARTHP) has been added to the system and is compatible with the Emprint™ HP Ablation Generator (CAGENHP).
The optional cart holds the Emprint™ HP Generator (CAGENHP) and the Emprint™ Ablation Pump (CAPUMP1) securely in one location. The optional footswitch can be placed on the cart's lower shelf. There are hooks on both sides of the cart which are meant to hold either the normal saline bag during a procedure, or to loop (and store) the reusable cable over while not in use.
Emprint™ HP Ablation Cart (CARTHP) is similar to the previously cleared (K193232) Emprint™ Ablation Cart (CART1), which holds the Emprint™ Generator (CAGEN1) and the Emprint™ Ablation Pump (CAPUMP1) securely in one location.
The provided document is a 510(k) summary for the Covidien Emprint™ Ablation System with Thermosphere™ Technology. It describes a submission for new components, specifically the Emprint™ HP Ablation Generator (CAGENHP) and its optional accessory, the Emprint™ HP Ablation Cart (CARTHP).
This document does not describe an AI medical device or a study proving its performance against acceptance criteria for an AI or imaging-based diagnostic device. Instead, it focuses on demonstrating substantial equivalence of a microwave ablation system (a physical medical device used for soft tissue coagulation) to a predicate device.
Therefore, many of the questions asked in the prompt, such as those related to AI-specific performance criteria (e.g., sample size for training/test sets, expert consensus, MRMC studies, effect size of AI assistance), are not applicable to the content of this document.
However, I can extract information relevant to the device's acceptance criteria and the studies performed to prove its substantial equivalence as described in the document.
Analysis of Device Acceptance Criteria and Performance (Based on the provided non-AI device document):
Since the document describes a microwave ablation system and not an AI/imaging diagnostic device, the acceptance criteria and supporting studies are focused on the safety and effectiveness of the physical device and its new components, primarily demonstrating substantial equivalence to a predicate device.
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't present a formal "acceptance criteria" table in the way one might for an AI device's performance metrics (e.g., sensitivity, specificity thresholds). Instead, the "acceptance criteria" are implied by the various tests and verification activities conducted to demonstrate the new components are safe and effective and do not alter the system's fundamental performance compared to the predicate.
| Implied Acceptance Criterion / Test Objective | Reported Device Performance / Outcome (Conclusion) |
|---|---|
| Bench Verification Testing: | |
| 1. System meets all design requirements. | Demonstrated. |
| 2. Ablation performance design requirements met; equivalent ablation zone dimensions at identical power settings (HP Generator vs. Predicate Generator). | Demonstrated that the ablations zone dimensions created with the Emprint™ HP Ablation Generator (subject) and Emprint™ Ablation Generator (predicate) are equivalent for identical power settings. |
| 3. Compatibility with previously approved Emprint™ Ablation System accessories (except cart). | Verified as compatible. (Note: New HP Cart is compatible with HP Generator). |
| 4. Meets IEC 60601-1-2:2014 EMC regulatory requirements. | Met. |
| 5. Meets IEC 60601-1:2005 and IEC 60601-2-6:2016 safety requirements. | Met. |
| 6. Performs as specified within storage and use conditions. | Verified. |
| 7. Usability of Emprint™ HP Ablation Generator. | Verified. |
| 8. Packaging protects generator and cart during simulated transportation. | Verified. |
| Animal Testing (In Vivo Porcine Model): | |
| Equivalent performance in terms of ablation zone creation relative to predicate control device. | Demonstrated equivalent performance and procedural safety of the Emprint™ Ablation System with CAGENHP generator in an in vivo porcine model undergoing soft tissue ablation procedures. |
| Overall Substantial Equivalence: | |
| Differences (extended power range, software control) do not impact safety and effectiveness. | Performance data demonstrate that these differences do not impact the safety and effectiveness of the device. (Implies all tests support this). |
| Addition of optional accessory (HP Ablation Cart) has no impact on safety and effectiveness. | Verification and validation activities conducted have shown that the addition of the optional accessory, Emprint™ HP Ablation Cart, has no impact on the safety and effectiveness of the Emprint™ Ablation System. (Implies all tests support this). |
The following points are not applicable as the document describes a physical medical device, not an AI/imaging diagnostic device. I'll explicitly state "Not Applicable" or "Not Provided/Relevant" for these.
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size for Test Set: Not explicitly specified for discrete tests. For the animal study, it mentions "an in vivo porcine model" but doesn't give a specific number of animals. For bench testing, typical engineering verification involves numerous tests without a "sample size" in the clinical trial sense.
- Data Provenance (Country of origin, retrospective/prospective): Not specified. The animal study is stated as "GLP-compliant pivotal study," suggesting a prospective, controlled experimental design, but location is not mentioned.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not Applicable. This is not an AI/diagnostic device where human expert "ground truth" for image interpretation would be established. The ground truth for this device is based on physical measurements (e.g., ablation zone dimensions), engineering specifications, and safety standards.
4. Adjudication Method for the Test Set:
- Not Applicable. No human interpretation for "ground truth" to be adjudicated.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No. This is not a diagnostic device involving human readers.
- Effect size of human readers improve with AI vs. without AI assistance: Not Applicable.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a hardware device, not an algorithm.
7. The Type of Ground Truth Used:
- For the ablation performance, the "ground truth" appears to be physical measurements of ablation zone dimensions (compared directly between the subject and predicate devices under identical power settings) and compliance with engineering design requirements, international safety and EMC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-6), and preclinical (animal) study results demonstrating equivalent therapeutic effect and safety.
8. The Sample Size for the Training Set:
- Not Applicable. This is not an AI device trained on a dataset.
9. How the Ground Truth for the Training Set Was Established:
- Not Applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 26, 2020
Covidien LLC Liron Bar Yaakov Senior Manager Regulatory Affairs 5920 Longbow Drive Boulder, Colorado 80301
Re: K200796
Trade/Device Name: Emprint Ablation System with Thermosphere Technology Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: NEY Dated: March 25, 2020 Received: July 20, 2020
Dear Liron Bar Yaakov:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|---|
| Food and Drug Administration |
| Indications for Use |
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K200796
Device Name Emprint™ Ablation System with Thermosphere™ Technology
Indications for Use (Describe)
The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation on nonresectable liver tumors.
The Emprint™ Ablation System is not intended for use in cardiac procedures
| Type of Use (Select one or both, as applicable) | XPrescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ------------------------------------------------- | ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
2 . information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (901) 443-6740 EF
{3}------------------------------------------------
Covidien LLC Traditional 510(k) Submission Emprint™ Ablation System with Thermosphere™ Emprint™ HP Ablation Generator and Emprint™ HP Ablation Cart
510(k) Summary
l. Submitter
Medtronic Inc 5920 Longbow Dr. Boulder, CO 80301
| Contact Person: | Liron Bar yaakovSenior Manager, Regulatory Affairs | |
|---|---|---|
| Phone: | 303-581-6668 | |
| Email: | Liron.baryaakov@medtronic.com | |
| Address: | 5920 Longbow Drive,Boulder, CO, 80301 | |
| Date of Summary Preparation: | 25 March 2020 |
II. Devices
Table 5.1 Devices
| Device Trade Names | Emprint™ Ablation System with Thermosphere™Technology |
|---|---|
| Device Common Name | Microwave Ablation System |
| Classification Name | System, Ablation, Microwave and Accessories |
| Regulation Number | 21 CFR 878.4400 |
| Regulatory Class | II |
| Product Code | NEY |
III. Predicate devices
This Traditional 510(k) submission describes the addition of a new system component (Emprint™ HP Ablation Generator - CAGENHP) and its optional accessory (Emprint™ HP Ablation Cart -CARTHP) as well as new and updated IFU's for the Emprint™ Ablation System with Thermosphere™ Technology.
The Emprint™ Ablation System with the addition of the Emprint™ HP Generator - CAGENHP and its optional accessory (Emprint™ HP Ablation Cart - CARTHP) is substantially equivalent to the predicate devices outlined in Table 5.2.
{4}------------------------------------------------
Table 5.2 Predicate Device
| Device Trade Names | Emprint™ Ablation System with Thermosphere™Technology |
|---|---|
| Device Common Name | Microwave Ablation System |
| Classification Name | System, Ablation, Microwave and Accessories |
| Regulation Number | 21 CFR 878.4400 |
| Regulatory Class | II |
| Product Code | NEY |
| 510(k) Clearance No. | K193232 |
Table 5.3 Reference Device
| Device Trade Names | Emprint™ Ablation System with Thermosphere™Technology |
|---|---|
| Device Common Name | Microwave Ablation System |
| Classification Name | System, Ablation, Microwave and Accessories |
| Regulation Number | 21 CFR 878.4400 |
| Regulatory Class | II |
| Product Code | NEY |
| 510(k) Clearance No. | K133821 |
IV. Device description
The Emprint™ Ablation System with Thermosphere™ Technology, Covidien's microwave ablation system, was released for commercial distribution in the United States in April 2014.
The 510(k) Cleared Emprint™ Ablation System consists of the following components:
-
- Emprint™ Ablation Generator (2450 MHz)
-
- Emprint™ Percutaneous Antenna (sterile, single use)
-
- Emprint™ Ablation Reusable Cable
-
- Emprint™ Ablation Pump
The system also includes the following optional equipment/accessories:
-
- Emprint™ Ablation Cart (with Isolation Transformer)
-
- Ablation Footswitch
-
- Remote Temperature Probe (sterile, single use)
The current Emprint™ Ablation System utilizes a 2450 MHz 100W generator (CAGEN1) to deliver power to a dedicated single microwave ablation antenna. The optional cart, Emprint™ Ablation Cart is intended to be used with the Emprint™ Ablation Generator.
{5}------------------------------------------------
The new system component, Emprint™ HP Ablation Generator (CAGENHP) [subject device] is similar to the device, Emprint™ Ablation Generator (CAGEN1) [predicate device] as described in K193232 with two exceptions. The new system component, Emprint™ HP Ablation Generator (CAGENHP), will extend the power setting range from 5 to 100W (Emprint™ Ablation Generator - CAGEN1) to 5 to 150W. The increased power will allow clinicians to create larger ablation zones when ablating soft tissue lesions, compared to use of the system with the existing 100-watt microwave generator (CAGEN1). Further the Emprint™ HP Ablation Generator (CAGENHP) is composed of analog and digital circuits with software whereas the predicate Emprint™ Ablation Generator (CAGEN1), is composed of analog and digital circuits with no software or firmware. The Emprint™ HP Generator will be compatible with all current Emprint™ ablation system components and accessories except the cart. The new optional Emprint™ HP Cart (CARTHP) has been added to the system and is compatible with the Emprint™ HP Ablation Generator (CAGENHP).
The optional cart holds the Emprint™ HP Generator (CAGENHP) and the Emprint™ Ablation Pump (CAPUMP1) securely in one location. The optional footswitch can be placed on the cart's lower shelf. There are hooks on both sides of the cart which are meant to hold either the normal saline bag during a procedure, or to loop (and store) the reusable cable over while not in use.
Emprint™ HP Ablation Cart (CARTHP) is similar to the previously cleared (K193232) Emprint™ Ablation Cart (CART1), which holds the Emprint™ Generator (CAGEN1) and the Emprint™ Ablation Pump (CAPUMP1) securely in one location.
V. Indications for use
The addition of a new system component (Emprint™ HP Ablation Generator - CAGENHP) and its optional accessory (Emprint™ HP Ablation Cart - CARTHP) do not alter the Indications for use, shown below, for the Emprint™ Ablation system.
The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, endoscopic, and intraoperative coagulation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.
The Emprint™ Ablation System is not intended for use in cardiac procedures.
VI. Comparison of technological characteristics with the predicate device
The addition of a new system component (Emprint™ HP Ablation Generator - CAGENHP) and its optional accessory (Emprint™ HP Ablation Cart - CARTHP) to the Emprint™ Ablation System with Thermosphere™ Technology does not alter the principles of operation and fundamental technology.
The Emprint™ Ablation System's performance with addition of a new system component (Emprint™ HP Ablation Generator - CAGENHP) and its optional accessory (Emprint™ HP Ablation Cart - CARTHP) is equivalent to its predicate device with respect to its intended use of soft tissue ablation in open, laparoscopic, endoscopic and percutaneous procedures.
The Emprint™ HP Ablation Generator (CAGENHP) and its optional accessory Emprint™ HP Ablation Cart (CARTHP) share the same underlying technology as the predicate devices,
{6}------------------------------------------------
Covidien LLC Traditional 510(k) Submission Emprint™ Ablation System with Thermosphere™ Emprint™ HP Ablation Generator and Emprint™ HP Ablation Cart
Emprint™ Ablation Generator (CAGEN1) and its optional accessory Emprint™ Ablation Cart (CART1). Both CAGENHP and CAGEN1 are used to deliver microwave energy to soft tissue in a highly controlled way.
The new system component, Emprint™ HP Ablation Generator (CAGENHP) is used to coagulate (ablate) soft tissue in the same way, as the predicate Emprint™ Ablation generator. Although the Emprint™ HP Ablation Generator differs from the predicate device with respect to being able to deliver an extended power setting range from 5 to 150W and being software controlled when compared to the predicate device, performance data demonstrate that these differences do not impact the safety and effectiveness of the device.
The verification and validation activities conducted has shown that the addition of the optional accessory, Emprint™ HP Ablation Cart - CARTHP, has no impact on the safety and effectiveness of the Emprint™ Ablation System.
The addition of a new system component (Emprint™ HP Ablation Generator - CAGENHP) and its optional accessory (Emprint™ HP Ablation Cart - CARTHP) do not affect the design requirements and the essential performance of the system as a whole.
VII. Performance data
Non-clinical performance data was provided in support of the substantial equivalence determination.
a) Bench Verification Testing
Bench testing was conducted at the system and subsystem level to demonstrate the following:
-
- Emprint™ Ablation System with Emprint™ HP Ablation Generator and Emprint™ HP Ablation Cart (CARTHP) meets all design requirements.
-
- Emprint™ Ablation System with Emprint™ HP Ablation Generator meets all ablation performance design requirements and to demonstrate that the ablations zone dimensions created with the Emprint™ HP Ablation Generator (subject) and Emprint™ Ablation Generator (predicate) are equivalent for identical power settings.
-
- Verify that the previously approved Emprint™ Ablation System accessories (except the Emprint™ Ablation Cart - CART1) are compatible with the new Emprint™ HP Ablation Generator. Note: The new optional Emprint™ HP Ablation Cart (CARTHP) has been added to the system and is compatible with the Emprint™ HP Ablation Generator (CAGENHP).
-
- The Emprint™ Ablation System with Emprint™ HP Ablation Generator meets IEC 60601-1-2:2014 EMC requlatory requirements.
-
- Emprint™ Ablation System with Emprint™ HP Ablation Generator meets IEC 60601-1:2005 and IEC 60601-2-6:2016 safety requirements.
-
- Verify the Emprint™ Ablation System with Emprint™ HP Ablation Generator (CAGENHP) and Emprint™ HP Ablation Cart (CARTHP) performs as specified within specified storage and use conditions.
{7}------------------------------------------------
-
- Verify the usability of Emprint™ HP Ablation Generator (CAGENHP) to the Emprint™ Ablation System.
-
- Verify the Emprint™ HP Ablation Generator (CAGENHP) and Emprint™ HP Ablation Cart (CARTHP) packaging can protect the generator and cart respectively during simulated transportation.
b) Animal Testing
The Emprint™ Ablation System with CAGENHP generator demonstrated equivalent performance in terms of ablation zone creation relative to the predicate control device (Emprint™ Ablation System with CAGEN1 generator) using the Emprint™ Percutaneous Antenna in an in vivo porcine model undergoing soft tissue ablation procedures. This GLP-compliant pivotal study verified the performance and procedural safety of the Emprint™ Ablation System with CAGENHP generator in an in vivo porcine model undergoing soft tissue ablation procedures.
c) Clinical Testing
Clinical studies in human subjects were not required to demonstrate the performance and safety of the new system component Emprint™ HP Ablation Generator (CAGENHP) and optional accessory Emprint™ HP Ablation Cart (CARTHP).
VIII. Biocompatibility
The biocompatibility of the Emprint™ Ablation System with Thermosphere™ Technology is not affected by the proposed addition of the new system component. The Emprint™ HP Ablation Generator (CAGENHP) and the optional accessory, and Emprint™ HP Ablation Cart (CARTHP).
The Emprint™ HP Ablation Generator (CAGENHP), and Emprint™ HP Ablation Cart (CARTHP) do not contact the patient, are reusable, not provided sterile and are not used in the sterile field. Cleaning instructions for these components are provided in the applicable user guides.
IX. Sterilization and Shelf Life
The sterilization and shelf life of the Emprint™ Ablation System with Thermosphere™ Technology is not affected by the proposed addition of the new system component. The Emprint™ HP Ablation Generator (CAGENHP) and the optional accessory, Emprint™ HP Ablation Cart (CARTHP). The Emprint™ HP Ablation Generator (CAGENHP) and Emprint™ HP Ablation Cart (CARTHP) are reusable, not provided sterile and are not used in the sterile field. Wipe down instructions for these components are provided in the applicable user guides.
{8}------------------------------------------------
Covidien LLC Traditional 510(k) Submission Emprint™ Ablation System with Thermosphere™ Emprint™ HP Ablation Generator and Emprint™ HP Ablation Cart
X. Conclusions
The new system component, Emprint™ HP Ablation Generator (CAGENHP) is used to coagulate (ablate) soft tissue in the same way as the predicate Emprint™ Ablation Generator. Although the Emprint™ HP Ablation Generator differs from the predicate device with respect to being able to deliver an extended power setting range from 5 to 150W and being software controlled when compared to the predicate device, performance data demonstrate that these differences do not impact the safety and effectiveness of the device.
The verification and validation activities conducted have shown that the addition of the optional accessory, Emprint™ HP Ablation Cart, has no impact on the safety and effectiveness of the Emprint™ Ablation System.
Based on the information provided the Emprint™ HP Ablation Generator - CAGENHP (subject) is substantially equivalent to the Emprint™ Ablation Generator - CAGEN1 (predicate) and the optional accessory Emprint™ HP Ablation Cart - CARTHP (subject) is substantially equivalent to the optional accessory Emprint™ Ablation Cart - CART1 (predicate).
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.