K Number
K200697
Manufacturer
Date Cleared
2020-12-23

(281 days)

Product Code
Regulation Number
882.4250
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For Adult Patients
AtriCure's Cryo2 cryo2 cryoablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway.
The Cryo2 cryolCE cryo-ablation probes are also intended for use to temporarily block pain by ablating peripheral nerves.
For Adolescent Patients
The Cryo2 cryolCE cryo-ablation probes are intended for use to temporarily block pain by ablating intercostal nerves under direct visualization1 in adolescent patients of at least 12 years of age.
1 Direct visualization, in this context, requires that the surgeon is able to see the targeted tissue for cryoablation directly or with assistance from a camera, endoscope or other similar optical technology.
For Adult Patients
AtriCure's cryolCE cryoSPHERE cryoablation probes are sterile, single use devices intended for use performed by freezing target tissues, creating an inflammatory response (cryonecrosis) for blocking pain by temporarily ablating peripheral nerves.
For Adolescent Patients
The cryolCE cryoSPHERE cryo-ablation probes are intended for use to temporarily block pain by ablating intercostal nerves under direct visualization1 in adolescent patients of at least 12 years of age.
1 Direct visualization, in this context, requires that the surgeon is able to see the targeted tissue for cryoablation directly or with assistance from a camera, endoscope or other similar optical technology.

Device Description

The cryosurgical handpiece, or cryo-ablation probe, utilized in the AtriCure Cryosurgical System is a hand held, single use, cryosurgical instrument intended for the cryosurgical treatment of cardiac surgery; it is also intended for use in blocking pain by temporarily ablatinq peripheral nerves. The cryosurgical handpiece uilizes a high-pressure cryogen (nitrous oxide, N20) to freeze target tissues, creating an inflammatory response, and ultimately, cryonerosis. The cryosurgical handpiece provides below -40°C, the temperature at which intracellular ice formation occurs which is considered lethal to cells. When high pressure nitous oxide is supplied to the cryprobe via the AtriCure Cryo Module, rapid cooling is achieved via the Joule-Thompson effect. The end effector, or cryolip, of the probe is malleable to allow access to varying anatomy and anatomical sites.
The cryo-ablation probes are comprised of a cryotip end effector, shaft, handle, thermocouple, inlet tube, and exhaust tube. The cryotip consists of an aluminum boiler and three infices distributed throughout the cryotip internally to provide uniform cooling. The 4-mm diameter, cryotip is malleable throughout its 10-cm length. A supplied forming tool can be cryotip into the desired form. The cryolip is attached to an insulated rigid shaft that allows the surgeon to adjust the exposed cryolip up to 10 cm in therapeutic length. A thermocouple is affixed to the proximal external surface of the temperatures on the console. The handle is attached to the shaft. Inlet and exhaust tubes and thermocouple wire pass through the handle and connect to the AtriCure Cryo Module (ACM).
The cryoICE PROBE is a sterile, single-use cryosurgical instrument.
The form tool facilitates bending of the malleable tip.
This PROBE was designed for treatment of cardieving controlled temperatures down to -50° to -70°C; it can also be used to temporarily block pain by ablating peripheral nerves in adolescent patients of at least 12 years of aqe.
The cryolCE cryoSPHERE cryoablation (CRYOS/CRYOS-L) probe is a single use cryablation of peripheral nerves by allowing a surgeon to insert the probe through an incision to reach the probe tip on the target site, and in conjunction with an AtriCure Cryo Module (ACM), temporarily freeze the tissue in by circulating a cryogenic agent, nitrous oxide, through the device. The CRYOS probe is offered in two probe length configurations: approximately 11" and 18" long. The probe is malleable and is capable of being bent by the end user. At the distal end, the CRYOS device features an 8 mm diameter ball tip shaped probe.
The PROBE is a single-use device offered in two configurations: standard length probe shaft (CRYOS), and extended length probe shaft (CRYOS-L). The probe shaft is malleable and supports forming by the supplied TOOL. The PROBE features a spherical 8mm cryoablation tip.

AI/ML Overview

This document is a 510(k) summary for AtriCure CryoICE cryo-ablation probes (Cryo2, CryoS, CryoS-L). It focuses on the expanded indication for use in adolescent patients aged 12 and above for temporary pain unique to its use in intercostal nerves under direct visualization.

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with numerical targets. Instead, it states that clinical evidence demonstrated the use of the cryoablation probes in adolescent patients (≥ 12 years) for intercostal nerve ablation was "as safe and effective" as its use in adults. The evaluation was based on the following endpoints, implying these were the performance areas assessed:

Acceptance Criteria (Implied)Reported Device Performance
Safety (regarding intraoperative or post-operative complications)Demonstrated to be as safe as adult use.
Effectiveness (regarding intraoperative or post-operative complications)Demonstrated to be as effective as adult use.
Incidence of adverse eventsComparable to adult use.
Incidence of cryoablation associated numbnessComparable to adult use.
Incidence of neuropathic painComparable to adult use.
Incidence of chest wall numbnessComparable to adult use.
Incidence of post-operative pneumothoraxComparable to adult use.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document states that AtriCure partnered with the UCSF Stanford Pediatric Device Consortium (PDC) to evaluate clinical evidence, which comprised clinical publications. The exact sample size from these publications is not specified in this summary. The data provenance (country of origin, retrospective/prospective nature) is also not explicitly stated in this summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the 510(k) summary. The evaluation relies on published clinical evidence rather than a dedicated expert panel for ground truth establishment within this specific submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the 510(k) summary. Given that the evidence is based on clinical publications, any adjudication would have been part of the original studies, and details are not reiterated here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A multi-reader multi-case (MRMC) comparative effectiveness study was not performed as this device is a physical medical instrument (cryoablation probe) and not an AI-assisted diagnostic or interpretative tool. Therefore, the concept of human readers improving with AI assistance is not applicable here.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not applicable as the device is a physical cryoablation probe and not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the comparison of safety and effectiveness was based on outcomes data from the clinical publications. Specifically, it assessed "intraoperative or post-operative complications," "adverse events observed," and specific post-ablation conditions like "cryoablation associated numbness," "neuropathic pain," "chest wall numbness," and "post-operative pneumothorax." The comparison was against the known safety and effectiveness in adult patients.

8. The sample size for the training set

This document describes a premarket notification (510(k)) for a medical device and relies on existing clinical evidence. There is no "training set" in the context of machine learning for this type of device submission. The clinical evidence used for assessment is analogous to a validation set.

9. How the ground truth for the training set was established

As there is no "training set" in the context of machine learning for this device, this question is not applicable. The clinical evidence used for evaluation was derived from existing clinical publications, where the "ground truth" (e.g., presence of complications, pain levels) would have been established within those individual studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 23, 2020

AtriCure Inc. Melissa Smallwood Senior Regulatory Affairs Specialist 7555 Innovation Way Mason, Ohio 45040

Re: K200697

Trade/Device Name: AtriCure CryoICE cryo-ablation probe (Cryo2), AtriCure CryoICE CryoSPHERE cryoablation probe (CryoS, CryoS-L) Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Device Regulatory Class: Class II Product Code: GXH Dated: March 13, 2020 Received: March 17, 2020

Dear Melissa Smallwood:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200697

Device Name

AtriCure® cryoICE® cryoablation probe (Cryo2)

Indications for Use (Describe)

For Adult Patients

AtriCure's Cryo2 cryo2 cryoablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway.

The Cryo2 cryolCE cryo-ablation probes are also intended for use to temporarily block pain by ablating peripheral nerves.

For Adolescent Patients

The Cryo2 cryolCE cryo-ablation probes are intended for use to temporarily block pain by ablating intercostal nerves under direct visualization1 in adolescent patients of at least 12 years of age.

1 Direct visualization, in this context, requires that the surgeon is able to see the targeted tissue for cryoablation directly or with assistance from a camera, endoscope or other similar optical technology.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K200697

Device Name

AtriCure® cryoICE® cryoSPHERE™ cryoablation probe (CryoS, CryoS-L)

Indications for Use (Describe)

For Adult Patients

AtriCure's cryolCE cryoSPHERE cryoablation probes are sterile, single use devices intended for use performed by freezing target tissues, creating an inflammatory response (cryonecrosis) for blocking pain by temporarily ablating peripheral nerves.

For Adolescent Patients

The cryolCE cryoSPHERE cryo-ablation probes are intended for use to temporarily block pain by ablating intercostal nerves under direct visualization1 in adolescent patients of at least 12 years of age.

1 Direct visualization, in this context, requires that the surgeon is able to see the targeted tissue for cryoablation drectly or with assistance from a camera, endoscope or other similar optical technology.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

l. Submitter

Manufacturer:AtriCure, Inc.7555 Innovation WayMason, OH 45040P: 513-644-4736F: 513-755-4108
Contact Person:Melissa Smallwood, MHASenior Regulatory Affairs Specialist
Alternate Contact Person:Mary Galeano, RACSenior Regulatory Affairs Specialist
Date Submitted:11/23/2020
II.Device
Name of Devices:AtriCure® cryolCE® cryoablation probe (Cryo2)AtriCure® cryolCE® cryoSPHERE™ cryoablation probe (CryoS, CryoS-L)
Common Name:Cryosurgical Probe
Classification Name:Cryogenic Surgical Device (21 CFR 878.4250)
Regulatory Class:Class II
Product Code:GXH

lll. Predicate Device

Predicate Device:

K180138 AtriCure® cryolCE® cryoablation probe (Cryo2) K182565 AtriCure® cryolCE® cryoSPHERE™ cryoablation probe (CryoS, CryoS-L)

IV. Device Description

Cryo2

The cryosurgical handpiece, or cryo-ablation probe, utilized in the AtriCure Cryosurgical System is a hand held, single use, cryosurgical instrument intended for the cryosurgical treatment of cardiac surgery; it is also intended for use in blocking pain by temporarily ablatinq peripheral nerves. The cryosurgical handpiece uilizes a high-pressure cryogen (nitrous oxide, N20) to freeze target tissues, creating an inflammatory response, and ultimately, cryonerosis. The cryosurgical handpiece provides below -40°C, the temperature at which intracellular ice formation occurs which is considered lethal to cells. When high pressure nitous oxide is supplied to the cryprobe via the AtriCure Cryo Module, rapid cooling is achieved via the Joule-Thompson effect. The end effector, or cryolip, of the probe is malleable to allow access to varying anatomy and anatomical sites.

The cryo-ablation probes are comprised of a cryotip end effector, shaft, handle, thermocouple, inlet tube, and exhaust tube. The cryotip consists of an aluminum boiler and three infices distributed throughout the cryotip internally to provide uniform cooling. The 4-mm diameter, cryotip is malleable throughout its 10-cm length. A supplied forming tool can be cryotip into the desired form. The cryolip is attached to an insulated rigid shaft that allows the surgeon to adjust the exposed cryolip up to 10 cm in therapeutic length. A thermocouple is affixed to the proximal external surface of the temperatures on the console. The handle is attached to the shaft. Inlet and exhaust tubes and thermocouple wire pass through the handle and connect to the AtriCure Cryo Module (ACM).

The cryoICE PROBE is a sterile, single-use cryosurgical instrument.

The form tool facilitates bending of the malleable tip.

{5}------------------------------------------------

This PROBE was designed for treatment of cardieving controlled temperatures down to -50° to -70°C; it can also be used to temporarily block pain by ablating peripheral nerves in adolescent patients of at least 12 years of aqe.

CryoS, CryoS-L

The cryolCE cryoSPHERE cryoablation (CRYOS/CRYOS-L) probe is a single use cryablation of peripheral nerves by allowing a surgeon to insert the probe through an incision to reach the probe tip on the target site, and in conjunction with an AtriCure Cryo Module (ACM), temporarily freeze the tissue in by circulating a cryogenic agent, nitrous oxide, through the device. The CRYOS probe is offered in two probe length configurations: approximately 11" and 18" long. The probe is malleable and is capable of being bent by the end user. At the distal end, the CRYOS device features an 8 mm diameter ball tip shaped probe.

The PROBE is a single-use device offered in two configurations: standard length probe shaft (CRYOS), and extended length probe shaft (CRYOS-L). The probe shaft is malleable and supports forming by the supplied TOOL. The PROBE features a spherical 8mm cryoablation tip.

PACKAGE CONTENTS

  • 1.One (1) PROBE

2.One (1) TOOL

The PROBE and TOOL are supplied STERILE and NON-PYROGENIC in unopened package. For single use only. Do not resterilize. Do Not Re-Use.

V. Indications for Use

Cryo2

For Adult Patients

AtriCure's Cryo2 cryolCE cryablation probes are steriles intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway. The cryolCE cryo-ablation probes are also intended for use to temporarily block pain by ablating peripheral nerves.

For Adolescent Patients

The Cryol cryolCE cryo-ablation probes are intention intercostal nerves under intercostal nerves under direct visualization! in adolescent patients of at least 12 years of age.

'Direct visualization, in this context, requires that the surgeon is able to see the targeted tissue for cryoablation directly or with assistance from a camera, endoscope or other similar optical technology.

CryoS, CryoS-L

For Adult Patients

AtriCure's cryolCE cryoablation probes are sterile, single use devices intended for use performed by freezing target tissues, creating an inflammatory response (cryonecrosis) for blocking pain by temporarily ablating peripheral nerves.

For Adolescent Patients

The cryolCE cryoSPHERE cryo-ablation probes are intended for use to temporarily block pain by abilating intercostal nerves under direct visualization1 in adolescent patients of at least 12 years of age.

'Direct visualization, in this context, requires that the see the targeted tissue for cryoablation directly or with assistance from a camera, endoscope or other similar optical technology.

VI. Comparison of Technological Characteristics with the Predicate Device

AltiCure's cryolCE cryoablation probes (CryoS-L) have no changes in technological characteristics as compared to the predicate.

Modifications included in the cryolCE crycabation probes included indication for use in addescent patients to temporarily block pain by ablating intercostal nerves under direct visualization.

{6}------------------------------------------------

Summary of Technical Characteristics
CRYO2CRYOS/CRYOS-L
Market Product NamecryolCE cryoablation probecryolCE cryoSPHERE cryoablation probe
Intended UseFor Adult PatientsAtriCure's Cryo2 cryolCE cryoablation probes are sterile,single use devices intended for use in the cryosurgicaltreatment of cardiac arrhythmias by freezing target tissues,creating an inflammatory response (cryonecrosis) that blocksthe electrical conduction pathway.The cryolCE cryo-ablation probes are also intended for useto temporarily block pain by ablating peripheral nerves.For Adolescent PatientsThe Cryo2 cryolCE cryo-ablation probes are intended for useto temporarily block pain by ablating intercostal nerves underdirect visualization¹ in adolescent patients of at least 12 yearsof age.¹Direct visualization, in this context, requires that thesurgeon is able to see the targeted tissue for cryoablationdirectly or with assistance from a camera, endoscope orother similar optical technologyFor Adult PatientsAtriCure's cryolCE cryoSPHERE cryoablation probes aresterile, single use devices intended for use performed byfreezing target tissues, creating an inflammatory response(cryonecrosis) for blocking pain by temporarily ablatingperipheral nerves.For Adolescent PatientsThe cryolCE cryoSPHERE cryo-ablation probes are intendedfor use to temporarily block pain by ablating intercostal nervesunder direct visualization¹ in adolescent patients of at least12 years of age.¹Direct visualization, in this context, requires that the surgeonis able to see the targeted tissue for cryoablation directly orwith assistance from a camera, endoscope or other similaroptical technology.
Operating PrincipleJoule-Thompson EffectJoule-Thompson Effect
TechnologyThe system consists of cryoprobes that are used for freezingtarget tissue. A console is used to control the supply of gasto the cryoprobe.The system consists of cryoprobes that are used for freezingtarget tissue. A console is used to control the supply of gasto the cryoprobe.
Energy UsedNitrous OxideNitrous Oxide
OperatingTemperature-50°C to 70°C-50°C to 70°C
Human FactorsHand-held device containing cryogen with activation buttonon console or footswitch.Hand-held device containing cryogen with activation buttonon console or footswitch.
CryotipMaterial: aluminum alloyConstruction: smooth malleable cylindrical linear probeDiameter: 4 mmMaterial: aluminum alloyConstruction: smooth spherical ball welded to a malleablecylindrical aluminum shaft.Diameter: ball - 8mm; aluminum shaft - 5mm
Length: 10cmLength: reference the "end-effector" length
End-effector Length- Shaft and Cryotip28 cm28cm (CRYOS)46cm (CRYOS-L)
BiocompatibilityBiocompatible patient contacting materials.Biocompatible patient contacting materials.
Probe PackagingSterile - Single Use disposable deviceSterile - Single Use disposable device
SterilizationGamma IrradiationGamma Irradiation
Power SourceMains PoweredMains Powered

{7}------------------------------------------------

VII. Performance Data

Clinical Evidence

AtriCure partnered with UCSF Stanford Pediatic Device Consortium (PDC) to evaluate Clinical Evidence in support of the cryolCE cryoablation probes (Cryo2, CryoS and CryoS-L) Indication for Use. The clinical publications demonstrated that use of the cryol CE cryoablation probes (Cryo2, CryoS and CryoS-L) in intercostal nerve ablation for temporarily pain block was as safe and effective when used in patients aged ≥ 12 years as compared to use of the cryolCE cryo2, CryoS and CryoS-L) for equivalent procedures in adults, based on the following endpoints:

  • Intraoperative or post-operative complications for patients receiving cryoanalgesia during Nuss procedure .
  • . Adverse events observed for patients receiving cryoanalgesia
  • Cryoablation associated numbness
  • Neuropathic pain
  • . Chest wall numbness
  • . Post-operative pneumothorax

VIII. Conclusions

The Clinical Evidence reviewed in the 510(k) submission demonstrates that AtriCure's cryolCE cryos, CryoS-L), as labeled per the Indication for Use do not present any new or additional questions of safety and effectiveness.

§ 882.4250 Cryogenic surgical device.

(a)
Identification. A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.(b)
Classification. Class II (performance standards).