(111 days)
The Bendit2.7 Steerable Microcatheter is intended for general intravascular use in the peripheral vasculature. The microcatheter can be used for the delivery of diagnostic, embolic materials into the vasculature.
The Bendit2.7 is not intended to be used in intracranial or coronary vessels.
The Bendit2.7™ is a steerable microcatheter with a steerable distal tip. The tip's deflection is controlled using the Steering Slider on the proximal Steering Handle. The tip can be rotated bi-directionally while deflected by turning the Torque Knob on the Steering Handle.
The total working length of the Bendit2.7™ is 130 cm. It is comprised of two Nitinol hypo tubes that are welded together at their distal ends, with proprietary laser-cut patterns along the 28-centimeter distal section. The laser cuts give the Bendit2.7™ its flexibility while maintaining the Nitinol torsional rigidity for a high torque response. The distal 12 mm section is steerable and includes a radiopaque atraumatic tip. The distal 80 cm of the shaft is covered with a hydrophilic coating.
Sliding the Steering Slider forward moves the hypo tubes so that the distal tip deflects. When the Steering Slider is released, the tip shape is locked. The Bendit2.7™ lumen can accommodate compatible guidewires (≤0.018"). A standard Luer lock port for attachment of accessories is located at the proximal end of the Steering Handle.
This document describes the Bendit2.7 Steerable Microcatheter and its substantial equivalence to a previously cleared device (K190126) following design changes.
The focus here is not on an AI/ML device, but rather a medical device with mechanical and material changes. Therefore, much of the requested information regarding AI/ML studies (such as MRMC studies, ground truth establishment for training, expert qualifications for ground truth in test sets, and separate training/test sets) is not applicable to this submission.
Here's the breakdown based on the provided text, addressing only the applicable points:
1. Table of Acceptance Criteria and Reported Device Performance
The submission states that "All test results met the pre-defined test acceptance criteria." However, the specific acceptance criteria values are not explicitly detailed in the provided document, nor are the quantitative results for each test. Instead, the document lists the types of tests performed to verify the design changes:
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Mechanical/Physical Performance: | |
Visual inspections and dimensional verifications | Met pre-defined test acceptance criteria |
Tensile bond strength (tip and luer connections) | Met pre-defined test acceptance criteria |
Torsional bond strength | Met pre-defined test acceptance criteria |
Kink resistance | Met pre-defined test acceptance criteria |
Pushability, Retraction, and Torqueability | Met pre-defined test acceptance criteria |
Simulated use validation testing (including trackability) | Met pre-defined test acceptance criteria |
Material/Functional Performance: | |
Power Injection (for Flowrate and Device Pressure) | Met pre-defined test acceptance criteria |
Coating integrity | Met pre-defined test acceptance criteria |
Corrosion | Met pre-defined test acceptance criteria |
Chemical characterization according to ISO 10993-17-2020 Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachable substances and a toxicological risk assessment | Met pre-defined test acceptance criteria |
2. Sample size used for the test set and the data provenance
The document does not specify the sample sizes used for each of the listed tests. The data provenance is derived from in vitro bench testing performed by the manufacturer (Bend It Technologies, Ltd, Israel). It is not clinical data, nor does it involve human subjects or retrospective/prospective studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable as the evaluation involved in vitro bench testing of a physical medical device, not an AI/ML system requiring expert-derived ground truth.
4. Adjudication method for the test set
This question is not applicable as the evaluation involved in vitro bench testing of a physical medical device, not an AI/ML system requiring adjudication of interpretations.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable as the device is a physical microcatheter, not an AI/ML diagnostic tool. No MRMC study was performed.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This question is not applicable as the device is a physical microcatheter, not an AI/ML algorithm.
7. The type of ground truth used
The "ground truth" for this device's performance is established by engineering specifications and recognized industry standards for medical devices (e.g., ISO 10993-17-2020 for biocompatibility, and internal specifications for physical properties like tensile strength, kink resistance, etc.). The tests verify that the device meets these pre-defined engineering and safety specifications.
8. The sample size for the training set
This question is not applicable as this is not an AI/ML device.
9. How the ground truth for the training set was established
This question is not applicable as this is not an AI/ML device.
§ 870.1210 Continuous flush catheter.
(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).