K Number
K200414
Device Name
Pulse Oximeter
Date Cleared
2020-09-23

(217 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pulse Oximeter is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in clinical institution and home environments.

Device Description

The Pulse Oximeter is a battery powered device intended for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Pulse Oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 905 nm, which is Infrared light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. The device mainly composed of PCB board, On/Off button, mode button, OLED&LED screen, battery compartment, Bluetooth® module and plastic shell. The device has wireless connection via Bluetooth®. The device is a spot-check pulse oximeter and does not include alarms. The device is not intended for life-supporting or life-sustaining.

AI/ML Overview

The provided document is a 510(k) premarket notification for a Pulse Oximeter, establishing its substantial equivalence to a previously cleared predicate device. It includes sections on the device description, intended use, comparison to the predicate, and performance data. The following details are extracted to answer your request:

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. A table of acceptance criteria and the reported device performance

Based on the "Comparison to the Predicate Device" table and "Clinical data" section:

ParameterAcceptance Criteria (Predicate / Standard)Reported Device Performance (Subject Device)
SpO2 Measuring Range0%-100%0%-100%
SpO2 Resolution1%1%
SpO2 Accuracy (Arms)70~100%, ±3% (based on ISO 80601-2-61)±3% for saturations from 70% to 100%
SpO2 Accuracy (0-69%)UnspecifiedUnspecified
PR Measuring Range30-250 bpm30-250 bpm
PR Resolution1 bpm1 bpm
PR Accuracy± 2 bpm± 2 bpm
BiocompatibilityIn accordance with ISO 10993-1 (Cytotoxicity, Sensitization, Irritation)Testing conducted and considered surface contacting (exceeds 24 hours but not 30 days)
Electrical Safety/EMCIEC 60601-1, IEC 60601-1-2Compliance via testing
Home Healthcare Env.IEC 60601-1-11Compliance via testing
Software ValidationFDA Guidance for Content of Premarket Submissions for Software Contained in Medical DevicesSoftware validated
CybersecurityFDA Guidance for Content of Premarket Submissions for Management of Cybersecurity in Medical DevicesSoftware validated according to guidance
Wireless CoexistenceANSI C63.27L: 2017, AAMI TIR69: 2017, FDA Guidance Radio Frequency Wireless Technology in Medical DevicesTesting conducted
Bluetooth ComplianceFCC CFR Title 47 Part 15 Subpart CTesting conducted

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Clinical Test Set: 12 healthy adult volunteers (4 men and 8 women).
  • Data Provenance: The document does not explicitly state the country of origin for the clinical data. It generally refers to "laboratory CO-Oximeter," implying a controlled clinical setting. The study was prospective in nature, involving desaturation of subjects to collect data across the specified SpO2 range.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The ground truth for SpO2 accuracy was established by a laboratory CO-Oximeter analysis of arterial blood samples. This does not involve human experts establishing ground truth in the traditional sense of image interpretation. The CO-Oximeter itself serves as the gold standard instrument. The document does not specify the number or qualifications of personnel operating the CO-Oximeter or analyzing blood samples, but it implies a standard laboratory setting.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable for this type of device and study. The ground truth (SpO2 values from CO-Oximeter, PR from reference method implied by the standard) is objective and instrument-based, not based on expert consensus or adjudication. The agreement was analyzed using Bland and Altman statistics.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study done. This device is a standalone measurement device (pulse oximeter) and does not involve AI assistance for human readers/interpreters. The focus is on the accuracy of the device's measurements compared to a recognized standard.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the primary performance evaluation for SpO2 and PR accuracy was a standalone performance study of the device's measurements against a reference method (CO-Oximeter for SpO2).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For SpO2, the ground truth was instrumental measurement (CO-Oximeter) of arterial hemoglobin oxygen (SaO2) from blood samples. This is considered a highly accurate and objective gold standard for oxygen saturation.
For Pulse Rate (PR), while not explicitly detailed for ground truth method, given the context of ISO 80601-2-61, it would typically involve a reference ECG or other validated heart rate measurement.

8. The sample size for the training set

This document describes a clinical validation study for a medical device, not a machine learning model specific training and test set. Therefore, there is no "training set" in the context of AI/ML. The clinical data of 12 healthy adult volunteers and 1440 data points were used for performance validation.

9. How the ground truth for the training set was established

As there is no "training set" for an AI/ML model, this question is not applicable. The device's underlying principle relies on light absorption, not trained algorithms in the sense of deep learning or similar AI methodologies that require large labeled training datasets.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 23, 2020

Shenzhen Aeon Technology Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. No. A415, Block A, NanShan Medical devices Industrial Park Nanshan District Shenzhen. 518067 Cn

Re: K200414

Trade/Device Name: Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: August 25, 2020 Received: September 1, 2020

Dear Kevin Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Expiration Date: 06/30/2020

See PRA Statement below.

Form Approved: OMB No. 0910-0120

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

Device Name Pulse Oximeter

Indications for Use (Describe)

The Pulse Oximeter is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in clinical institution and home environments.

Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

K200414

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: 2020/09/17

Submission sponsor 1.

Name: Shenzhen Aeon Technology Co., Ltd.

Address: RM6H02, Block 27-29, Tianxia IC Industrial Park, Majialong, No.133 of Yiyuan road, Nantou

Street, Nanshan District, Shenzhen, P.R China.

Contact person: Xie Hua

Title: General Manager

E-mail: Raymond@aeon-med.com

Tel: +86-755-86182140

2. Submission correspondent

Name: Chonconn Medical Device Consulting Co., Ltd.

Address: No. A415, Block A, NanShan Medical devices Industrial Park Nanshan District, Shenzhen,

Guangdong, P.R. China 518067

Contact person: Kevin Wang

E-mail: kevin@chonconn.com

Tel: +86-755 33941160

Trade/Device NamePulse Oximeter
ModelA310B, A330B
Common NameFingertip Pulse Oximeter
Regulatory ClassClass II
Classification21CFR 870.2700 / Oximeter / DQA
Submission typeTraditional 510(K)

3. Subiect Device Information

4. Predicate Device

By submission of the Traditional 510(k), Shenzhen Aeon Technology Co., Ltd. is requesting clearance for Pulse Oximeter. It is comparable to the following legally marketed system:

  1. Shenzhen Aeon Technology Co., Ltd. Fingertip Pulse Oximeter under K190869.

The subject device has same intended use, same target patient population, same performance effectiveness, performance safety as the predicate devices and no question to effectiveness and safety. So, the conclusion is that the subject device is substantial equivalent to the predicate.

{4}------------------------------------------------

5. Device Description

The Pulse Oximeter is a battery powered device intended for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).

The Pulse Oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 905 nm, which is Infrared light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.

The device mainly composed of PCB board, On/Off button, mode button, OLED&LED screen, battery compartment, Bluetooth® module and plastic shell. The device has wireless connection via Bluetooth®.

The device is a spot-check pulse oximeter and does not include alarms. The device is not intended for lifesupporting or life-sustaining.

6. Intended use & Indication for use

The Pulse Oximeter is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in clinical institution and home environments.

FeaturesSubject DeviceA310B and A330BPredicate Device K190869Model: A310 and A330Remark
ApplicantShenzhen Aeon Technology Co.,Ltd.Shenzhen Aeon Technology Co.,Ltd.Same
ClassificationRegulation21CRF 870.270021CRF 870.2700Same
Classificationand CodeClass II,DQAClass II,DQASame
CommonnameFingertip Pulse OximeterFingertip Pulse OximeterSame
Intended useThe Pulse Oximeter is a non-invasive device intended for spotchecking of functional oxygensaturation of arterial hemoglobin(SpO2) and pulse rate (PR). Thisportable device is indicated foruse in adult patients in clinicalThe Pulse Oximeter is a non-invasive device intended for spotchecking of functional oxygensaturation of arterial hemoglobin(SpO2) and pulse rate (PR). Thisportable device is indicated foruse in adult patients in clinicalSame
FeaturesSubject DeviceA310B and A330BPredicate Device K190869Model: A310 and A330Remark
institution and homeenvironments.institution and homeenvironments.
PatientPopulationsAdultsAdultsSame
PrincipleThe device displays numericalvalues for functional oxygensaturation of arterial hemoglobin(SpO2) and pulse rate bymeasuring the absorption of redand infrared (IR) light passingthrough perfused tissue. Changesin the absorption caused by thepulsation of blood in the vascularbed are used to determine oxygensaturation and pulse rate.The device displays numericalvalues for functional oxygensaturation of arterial hemoglobin(SpO2) and pulse rate bymeasuring the absorption of redand infrared (IR) light passingthrough perfused tissue. Changesin the absorption caused by thepulsation of blood in the vascularbed are used to determine oxygensaturation and pulse rate.Same
LEDwavelengthRed=660 nm;Infrared=905 nm;Red=660 nm;Infrared=905 nm;Same
Power source2 AAA alkaline batteries2 AAA alkaline batteriesSame
Display dataSpO2%, PRSpO2%, PRSame
SpO2MeasuringRange0%-100%0%-100%Same
SpO2Resolution1%1%Same
SpO2 Accuracy70~100%, ±3%;0-69%, unspecified;70~100%, ±3%;0-69%, unspecified;Same
PR MeasuringRange30-250 bpm30-250 bpmSame
PR Resolution1 bpm1 bpmSame
PR Accuracy± 2 bpm± 2 bpmSame
WirelessconnectionBluetoothNoneDifferent 1)
StandardIEC 60601-1IEC 60601-1-2IEC 60601-1-11ISO 80601-2-61ISO 10993-5ISO 10993-10IEC 60601-1IEC 60601-1-2IEC 60601-1-11ISO 80601-2-61ISO 10993-5ISO 10993-10Same

7. Comparison to the Predicate Device

{5}------------------------------------------------

{6}------------------------------------------------

Justification of difference:

Different 1): The subject device is identical to predicate device except the wireless connection function. The subject device can transmit the measurement of SpO2, PR and Plethysmogram to the mobile application on the cell phone. The mobile application A-OXIMETER is only to display, store and review the measurement from pulse oximeter models A310B and A330B. The software is validated according to FDA's Guidance for Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. The wireless testing is conducted according to FDA's Guidance Radio Frequency Wireless Technology in Medical Devices. So. the different does not raise different questions of safety and effectiveness.

8. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the Pulse Oximeter was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The battery of testing included the following tests:

  • Cytotoxicity .
  • Sensitization .
  • . Irritation

The Pulse Oximeter is considered surface contacting for a duration of exceed 24 hours but not 30 days.

Non-clinical data

The Pulse Oximeter has been tested according to the following standards:

  • IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance
  • IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • ISO 80601-2-61: Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use.
  • IEC 60601-1-11: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
  • FDA Guidance for Pulse Oximeters - Premarket Notification Submissions [510(k)s]
  • FDA Guidance for Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.

{7}------------------------------------------------

Wireless testing:

  • Bluetooth test according to FCC CFR Title 47 Part 15 Subpart C.
  • ANSI C63.27L: 2017: American National Stand for Evaluation of Wireless Coexistence.
  • AAMI TIR69: 2017 Risk management of radio-frequency wireless coexistence for medical devices and systems.
  • Radio Frequency Wireless Technology in Medical Devices: Guidance for Industry and Food and Drug Administration Staff (August 14, 2013)

The test was selected to show substantial equivalence between the subject device and the predicate.

Clinical data

The functional oxygen saturation (SpO2) measurement has been validated in accordance with ISO 80601-2-61. The clinical testing was completed on a total of 12 healthy adult volunteers (4 men and 8 women) with light to dark skin pigmentations in the range of 70% to 100% against a laboratory CO-Oximeter. The subjects include 3 people with medium skin, 6 with light skin, and 3 with dark skin pigmentation. Total 1440 data points were sampled for analysis.

The measured arterial hemoglobin saturation value (SpO2) of the proposed device was compared with arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a CO-oximeter. The accuracy of the device is in comparison with the CO-oximeter samples measured over the SpO2 range 70%-100% .

Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, the result showed that the error is far less than the scope specified in the standard; and the Agreement between Methods of Measurement with Multiple Observations per each subject was analyzed with the Bland and Altman statistics, the analysis demonstrated that the vast majority of data is within ±95% limit of agreement.

The SpO2 accuracy result showed that the root-mean-square (Arms) value of the Fingertip Pulse Oximeter is ±3% with the saturations from 70% to 100%.

9. Conclusion

According to the 510(k) submission, the subject device and the predicate device has the same intended use, and the difference in technological features of the proposed devices and the predicate devices do not raise different questions regarding safety and effectiveness. Performance testing and compliance with voluntary standards demonstrate that the proposed device models are substantially equivalent to the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).