K Number
K200251
Date Cleared
2020-04-02

(59 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in Fetal, Abdominal, Pediatric, Musculo-skeletal (conventional, superficial) , Peripheral Vascular, Trans-rectal, Trans-vaginal, Small organ (breast, thyroid and testes) , Cephalic (neonatal and adult) , Cardiac (adult and pediatric) and Urology exams.

Device Description

The Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System is a general purpose, portable, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, Color-Mode, Power/Dirpower Mode, THI, Smart3D, 4D, iScape, Biopsy Guidance or the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color). This system is a Track 3 device that employs an array of probes that include Linear array, Convex array probe.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving performance against specific acceptance criteria through novel clinical studies.

Therefore, many of the requested details about specific acceptance criteria and a study proving those criteria are not present in this document because it's not a de novo or PMA submission that would typically require such detailed performance data from a new study. The document primarily focuses on demonstrating that the new device is as safe and effective as a previously cleared device.

However, I can extract the information available from the document that is relevant to your request:

1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria or report specific device performance metrics in the way a clinical study would for a novel device. Instead, it relies on demonstrating that the new device has "the same intended uses, technological characteristics, is comparable in key safety and effectiveness features, and has the same basic operating modes as the predicate devices."

The key performance aspect mentioned is that "The acoustic power levels of Z5/Z50/Z50T/Z50S/Z50 Pro are below the limits of FDA, which is the same as the predicated device Z5 (K130695)". This implicitly means the acoustic output meets FDA limits as specified by relevant standards.

Acceptance Criteria (Implied)Reported Device Performance
Acoustic Power LevelsBelow FDA limits (same as predicate device Z5)
Compliance with Safety StandardsDesigned in compliance with FDA recognized electrical and physical safety standards (same as predicate device Z5)
Intended Uses and Basic Operating ModesSame as predicate devices
Technological CharacteristicsSame as predicate devices
Safety and Effectiveness FeaturesComparable to predicate devices

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "Not applicable. The subject of this submission, Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence." This means there was no specific "test set" of patient data or images used in a clinical study to evaluate the device's performance as would be done for a new device. The evaluation is based on technical comparisons to predicate devices and adherence to recognized standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical studies with a test set requiring ground truth establishment were conducted or reported in this submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical studies with a test set requiring adjudication were conducted or reported in this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The document is for a diagnostic ultrasound system and does not mention any AI components or MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The device is a diagnostic ultrasound system, not an algorithm, and the submission does not describe standalone algorithm performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable, as no clinical studies requiring ground truth were conducted or reported in this submission.

8. The sample size for the training set

Not applicable, as the document is not about an AI/ML algorithm that would require a training set. The evaluation is based on compliance with standards and comparison to predicate devices.

9. How the ground truth for the training set was established

Not applicable, as the document is not about an AI/ML algorithm that would require a training set or its associated ground truth.Conclusion from the Document:

The Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System received 510(k) clearance based on demonstrating substantial equivalence to existing predicate devices (Mindray Z5 (K130695), Z6 (K182603), and DC-40 (K183377)). This equivalence was supported by:

  • Having the same intended uses and basic operating modes.
  • Comparable technological characteristics.
  • Acoustic power levels remaining below FDA limits, similar to the predicate device.
  • Compliance with recognized electrical, physical, and software safety standards (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, IEC 62366, IEC 60601-1-6, ISO14971, ISO 10993-1).
  • No clinical studies were deemed necessary to support this substantial equivalence claim.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

ShenZhen Mindray Bio-Medical Electronics Co., Ltd. % Shi Jufang Engineer of Technical Regulation Mindray Building, Keji 12th Road South Hi-Tech Industrial Park Shenzhen, Guangdong 518057 CHINA

April 2, 2020

Re: K200251

Trade/Device Name: Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 20, 2020 Received: February 3, 2020

Dear Shi Jufang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K200251

Device Name Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System

Indications for Use (Describe)

Z5/Z50/Z50S/Z50 Pro Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in Fetal, Abdominal, Pediatric, Musculo-sketal (conventional, superficial) , Peripheral Vascular, Trans-rectal, Trans-vaginal, Small organ (breast, thyroid and testes) , Cephalic (neonatal and adult) , Cardiac (adult and pediatric) and Urology exams.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Diagnostic Ultrasound Indications for Use Format
System:Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System
Transducer:N/A
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1,2,3,6,7
AbdominalPPPPPPNote 1,2,3,6,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,3,6
Small Organ (Specify**)PPPPPPNote 1,2,3,6
Neonatal CephalicPPPPPPNote 1,2,3,6
Adult CephalicPPPPPPNote 1,2,3,6
Trans-rectalPPPPPPNote 1,2,3,6
Trans-vaginalPPPPPPNote 1,2,3,6
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPPPPNote 1,2,3,6
Musculo-skeletal (Superficial)PPPPPPNote 1,2,3,6
Intravascular
CardiacCardiac AdultPPPPPPNote 1,2,3,6
Cardiac PediatricPPPPPPNote 1,2,3,6
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2,3,6
Other (Specify***)PPPPPPNote 1,2,3,6
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color+B、Power+B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note 5: Color M
Note 6: Smart3D
Note 7:4D(Real-time 3D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription USE (Per 21 CFR 801.109)
Diagnostic Ultrasound Indications for Use Format
System:Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System
Transducer:35C50EA
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1,2,3,6
AbdominalPPPPPPNote 1,2,3,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,3,6
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2,3,6
Musculo-skeletal (Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2,3,6
Other (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color+B、Power + B、PW+Color+B、Power + PW +B.
* Intraoperative includes abdominal, thoracic, and vascular.
** Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note 5: Color M
Note 6: Smart3D
Note 7:4D(Real-time 3D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription USE (Per 21 CFR 801.109)
Diagnostic Ultrasound Indications for Use Format
System:Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System
Transducer:35C20EA
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationSpecific (Track 1 & 3)Mode of Operation
General (Track1 Only)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPNote 1,2,3,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,3,6
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
CardiacCardiac AdultPPPPPPNote 1,2,3,6
Cardiac PediatricPPPPPPNote 1,2,3,6
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note 5: Color M
Note 6: Smart3D
Note 7:4D(Real-time 3D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription USE (Per 21 CFR 801.109)
Diagnostic Ultrasound Indications for Use Format
System:Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System
Transducer:65C15EA
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPNote 1,2,3,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,3,6
Small Organ (Specify**)
Neonatal CephalicPPPPPPNote 1,2,3,6
Adult CephalicPPPPPPNote 1,2,3,6
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color+B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note 5: Color M
Note 6: Smart3D
Note 7:4D(Real-time 3D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Diagnostic Ultrasound Indications for Use Format
System:Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System
Transducer:65EC10EA
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (Specify)
OphthalmicOphthalmic
Fetal Imaging &OtherFetalPPPPPPNote 1,2,3,6
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1,2,3,6
Trans-vaginalPPPPPPNote 1,2,3,6
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vessel
Other (Specify***)PPPPPPNote 1,2,3,6
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M, PW+B, Color+B, Power+B, PW+Color+B, Power + PW +B.
* Intraoperative includes abdominal, thoracic, and vascular.
** Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note 5: Color M
Note 6: Smart3D
Note 7:4D(Real-time 3D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription USE (Per 21 CFR 801.109)
Diagnostic Ultrasound Indications for Use Format
System:Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System
Transducer:10L24EA
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (Specify)
OphthalmicOphthalmic
Fetal Imaging &OtherFetal
AbdominalPPPPPPNote 1,2,3,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,3,6
Small Organ (Specify**)PPPPPPNote 1,2,3,6
Neonatal CephalicPPPPPPNote 1,2,3,6
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2,3,6
Musculo-skeletal (Superficial)PPPPPPNote 1,2,3,6
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vesselPPPPPPNote 1,2,3,6
Other (Specify***)
N=new indication; P=previously cleared by FDA;E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color+B、Power + B、PW+Color+B、Power + PW +B.
* Intraoperative includes abdominal, thoracic, and vascular.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note 5: Color M
Note 6: Smart3D
Note 7:4D(Real-time 3D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Diagnostic Ultrasound Indications for Use Format
System:Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System
Transducer:75L38EA
IntendedUse:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (Specify)
OphthalmicOphthalmic
FetalImaging &OtherFetal
AbdominalPPPPPPNote 1,2,3,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2,3,6
Small Organ (Specify**)PPPPPPNote 1,2,3,6
Neonatal CephalicPPPPPPNote 1,2,3,6
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2,3,6
Musculo-skeletal (Superficial)PPPPPPNote 1,2,3,6
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2,3,6
Other (Specify***)
N=new indication; P=previously cleared by FDA;E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
* Intraoperative includes abdominal, thoracic, and vascular.
** Small organ-breast, thyroid, testes.
*** Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note 5: Color M
Note 6: Smart3D
Note 7:4D(Real-time 3D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription USE (Per 21 CFR 801.109)
Diagnostic Ultrasound Indications for Use Format
System:Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System
Transducer:D6-2EA
IntendedUse:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (Specify)
OphthalmicOphthalmic
FetalImaging &OtherFetalNNNNNNNote 1,2,7
AbdominalNNNNNNNote 1,2,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
(Conventional)
Musculo-skeletal (Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
* Intraoperative includes abdominal, thoracic, and vascular.
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging.
Note 2: Biopsy Guidance
Note 3: iScape
Note 4: TDI
Note 5: Color M
Note 6: Smart3D
Note 7:4D(Real-time 3D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription USE (Per 21 CFR 801.109)

{4}------------------------------------------------

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c).

The assigned 510(k) number is: K200251

1. Submitter:

Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 8188 6293 Fax: +86 755 2658 2680

Contact Person:

Shi Jufang Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China

Date Prepared: January 20, 2020

2. Device Name: Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System Classification

Regulatory Class: II Review Category: Tier II 21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN) 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO) 21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)

3. Device Description:

{12}------------------------------------------------

The Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System is a general purpose, portable, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, Color-Mode, Power/Dirpower Mode, THI, Smart3D, 4D, iScape, Biopsy Guidance or the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color).

This system is a Track 3 device that employs an array of probes that include Linear array, Convex array probe.

4. Intended Use:

The Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in Fetal, Abdominal, Pediatric, Musculo-skeletal(conventional, superficial), Peripheral Vascular, Trans-rectal, Trans-vaginal, Small organ (breast, thyroid and testes) , Cephalic (neonatal and adult) , Cardiac (adult and pediatric) and Urology exams.

5. Summary of Modifications

  • Newly Added Models: . Z50, Z50T, Z50S, Z50 Pro;
  • Newly Added Transducers: D6-2EA:

. Main Added Features and Modifications:

    1. Appearance change;
    1. Add host with three probe board;
    1. Add 4D and Smart 3D;
    1. Add Free Xros M;
    1. Add iLive;
    1. Add iWorks;
    1. Add iNeedle;
    1. Add Smart Face;
    1. Add Smart OB;
    1. Add mobile trolley UMT-170.

{13}------------------------------------------------

6. Comparison with Predicate Devices:

The Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System is comparable with and substantially equivalent to these predicate devices:

Predicate DeviceManufacturerModel510(k) ControlNumber
1. Primary predicatedeviceMindrayZ5K130695
2. Reference deviceMindrayZ6K182603
3. Reference deviceMindrayDC-40K183377

The Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System has the same technological characteristics, is comparable in key safety and effectiveness features, and has the same intended uses and basic operating modes as the predicate devices. All systems transmit ultrasonic energy into patients and perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, and calculations.

  • . Subject device
    The Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System has the same intended uses as the predicated device Z5 (K130695)
Subject DevicePredicate device
The Z5/Z50/Z50T/Z50S/Z50 ProZ5 (K130695)
The Diagnostic Ultrasound SystemThe Diagnostic Ultrasound System
is applicable for adults, pregnant women,is applicable for adults, pregnant women,
pediatric patients and neonates. It ispediatric and neonates. It is intended for
intended for use in Fetal, Abdominal,use in Fetal, Abdominal, Pediatric,
Pediatric, Musculo-skeletalMusculo-skeletal (conventional,
(conventional, superficial), Peripheralsuperficial), Peripheral Vascular,
Vascular, Trans-rectal, Trans-vaginal,Trans-rectal, Trans-vaginal, Small organ

{14}------------------------------------------------

(breast, thyroid and testes), Cephalic Small organ (breast, thyroid and testes ) , Cephalic(neonatal and adult) , Cardiac (neonatal and adult), Cardiac (adult and (adult and pediatric) and Urology pediatric) and Urology exams. exams.

  • . The acoustic power levels of Z5/Z50/Z50T/Z50S/Z50 Pro are below the limits of FDA. which is the same as the predicated device Z5 (K130695)
  • . The Z5/Z50/Z50T/Z50S/Z50 Pro is designed in compliance with the FDA recognized electrical and physical safety standard, which is the same as the predicated device Z5 (K130695)

The Z5/Z50/Z50T/Z50S/Z50 Pro has similar probes as the predicated device.

7. Non-clinical Tests:

The Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been designed to conform with applicable medical safety standards. This device has been tested and evaluated under the following standards:

  • AAMI/ANSI ES60601-1: Medical electrical equipment Part 1: General requirements for basic safety and essential performance
  • . IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3)
  • IEC 60601-2-37: Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
  • IEC 62304: Medical device software - Software life cycle processes
  • IEC 62366:Medical devices - application of usability engineering to medical devices
  • IEC 60601-1-6: medical electrical equipment - part 1-6: general requirements for basic safety and essential performance - collateral standard: usability.
  • ISO14971: Medical devices - Application of risk management to medical devices
  • ISO 10993-1: Biological evaluation of medical devices -- Part 1: Evaluation and

{15}------------------------------------------------

testing within a risk management process

These non-clinical tests relied on in this premarket notification submission can support the determination of substantial equivalence of the subject device.

8. Clinical Studies

Not applicable. The subject of this submission, Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence.

Conclusion:

Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the Z5/Z50/Z50T/Z50S/Z50 Pro Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.