K Number
K200236
Date Cleared
2020-02-25

(26 days)

Product Code
Regulation Number
866.3215
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in-vitro diagnostic use only.

For the quantitative measurement of procalcitonin (PCT) in human serum and plasma (lithium heparin and EDTA) using the VITROS 3600 Immunodiagnostic System.

Used in conjunction with other laboratory findings and clinical assessments, the VITROS B R A PA - M S PCT test is intended for use as follows:

· to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock,

· to aid in assessing the cumulative 28-day risk of all-cause mortality for patients diagnosed with severe sepsis or septic shock in the ICU or when obtained in the emergency department or other medical wards prior to ICU admission, using a change in PCT level over time,

· to aid in decision making on antibiotic therapy for patients with suspected or confirmed lower respiratory tract infections (LRTI) defined as community-acquired pneumonia (CAP), acute bronchitis, and acute exacerbation of chronic obstructive pulmonary disease (AECOPD) - in an inpatient setting or an emergency department,

· to aid in decision making on antibiotic discontinuation for patients with suspected or confirmed sepsis.

Device Description

The VITROS B·R·A·H·M·S PCT test is performed using the VITROS B·R·A·H·M·S PCT Reagent Pack and the VITROS B·R·A·H·M·S PCT Calibrators on the VITROS Systems.

Reagent Pack Contents
1 reagent pack containing:

  • 100 coated wells (rat monoclonal anti-procalcitonin antibody, 1.0 µg/mL) ●
  • 10.20 mL assay reagent (buffer containing bovine gamma globulin, bovine serum ● albumin and antimicrobial agent)
  • . 13.10 mL conjugate reagent (HRP-conjugated mouse monoclonal procalcitonin antibody. 1.65 ug/mL in buffer with bovine serum albumin and antimicrobial agent)

Calibrator Contents

  • . 3 sets of VITROS B•R•A•H•M•S PCT Calibrators 1 and 2, 1.0 mL, procalcitonin in buffer with antimicrobial agent, nominal values 0.080 and 75.0 ng/mL (ug/L)
  • . Lot calibration card
  • Protocol card ●
    • 16 calibrator bar code labels (8 for each calibrator) ●
AI/ML Overview

The provided text is a 510(k) Summary for the VITROS B·R·A·H·M·S PCT Reagent Pack and Calibrators, intended for quantitative measurement of procalcitonin (PCT). This document describes the device's performance characteristics and clinical studies to establish substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a 510(k) summary for an in vitro diagnostic device, the "acceptance criteria" are implied by the comparison to the predicate device and established analytical performance metrics. The key criterion is demonstrating substantial equivalence to the predicate device, B·R·A·H·M·S PCT sensitive KRYPTOR (K171338), especially at critical clinical decision points.

Feature / Acceptance Criteria CategorySpecific Criteria (Implied/Stated)Reported Device Performance (VITROS B·R·A·H·M·S PCT)Supporting Study / Section
Predicate Device Comparison (Substantial Equivalence)Correlation with predicate (B·R·A·H·M·S PCT sensitive KRYPTOR)Correlation Coefficient (r): 0.995 (all samples), 0.994 (within measuring range)Method Comparison with Predicate Device
Mean Percent Bias1.31% (all samples), 1.88% (within measuring range)Method Comparison with Predicate Device
Passing & Bablok Slope (95% CI)1.001 (0.9773 to 1.027) (all samples), 1.025 (1.002 to 1.054) (within measuring range)Method Comparison with Predicate Device
Passing & Bablok Intercept (95% CI)0.01041 (0.001562 to 0.03272) (all samples), 0.004237 (-0.004908 to 0.01905) (within measuring range)Method Comparison with Predicate Device
Weighted Deming Slope (95% CI)1.046 (1.025 to 1.066) (all samples), 1.057 (1.035 to 1.078) (within measuring range)Method Comparison with Predicate Device
Weighted Deming Intercept (95% CI)-0.009264 (-0.01498 to -0.003549) (all samples), -0.009994 (-0.01585 to -0.004140) (within measuring range)Method Comparison with Predicate Device
Clinical Agreement at Decision PointsTotal clinical agreement > 97% at 0.100, 0.250, 0.500, 2.00 ng/mL98.5% (0.100 ng/mL), 98.0% (0.250 ng/mL), 97.4% (0.500 ng/mL), 97.8% (2.00 ng/mL)Clinical Concordance
Cohen's Kappa for agreement0.772 (0.100 ng/mL), >0.910 (0.250, 0.500, 2.00 ng/mL)Clinical Concordance
Precision (Within-lab %CV)Acceptable precision specified by CLSI EP05-A3Ranges from 3.1% to 6.4% across various concentrations (VITROS 3600)Precision/Reproducibility
Multi-Site Precision (Reproducibility %CV)Acceptable precision across sitesRanges from 4.6% to 15.9% across various concentrations (VITROS 3600)Multi-Site Precision
Accuracy (Linearity)Linear over the measuring rangeLinear from 0.030 to 100 ng/mL; Bias criteria of ±10% metLinearity/Assay Measuring Range
Limit of Detection (LoD)Determined consistent with CLSI EP17-A20.007 ng/mLDetection Limits
Limit of Quantitation (LoQ)Determined consistent with CLSI EP17-A20.030 ng/mL (at 20% CV)Detection Limits
Analytical Specificity (Interference)No bias >10% at specified concentrations for various compoundsNone of tested compounds caused >10% biasAnalytical Specificity
Matrix EquivalenceLess than 10% bias from serum (slope 0.90-1.10) for plasmaPassing & Bablok Slope: 0.980 (Lithium Heparin), 0.992 (EDTA)Matrix Comparison
High Dose HookNo hook effect up to 5,000 ng/mLNo evidence of high dose hook up to 5,000 ng/mLHigh Dose Hook

2. Sample Sizes and Data Provenance

  • Test Set (Method Comparison with Predicate):

    • Sample Size: 266 patient samples were used for the regression analysis comparing the VITROS B·R·A·H·M·S PCT test to the B·R·A·H·M·S PCT sensitive KRYPTOR assay. 246 samples were within the measuring range.
    • Data Provenance: Not explicitly stated for analytical studies, but for the clinical performance study, samples were "retrospective samples from a study of 858 adult patients diagnosed with severe sepsis or septic shock recruited across 13 investigational sites in the United States." The original MOSES Study (DEN150009) collected samples from patients admitted to ICU from emergency departments, other wards, or directly from out of the hospital. EDTA samples were used.
  • Test Set (Clinical Concordance):

    • Sample Size: 2168 samples (evaluating concordance at various PCT decision points).
    • Data Provenance: "serial sample sets obtained from the Multicenter Procalcitonin MOnitoring SEpsis (MOSES) Study collection, a well-characterized sample collection in 13 sites across the United States." This indicates a prospective collection used retrospectively for this specific device's evaluation.
  • Test Set (28-Day Mortality Prediction):

    • Sample Size: 858 adult patients in the study, with an analysis population of 598 subjects.
    • Data Provenance: Retrospective samples from the Multicenter Procalcitonin MOnitoring SEpsis (MOSES) Study collection in 13 sites across the United States.

3. Number of Experts and Qualifications for Ground Truth

  • For an in vitro diagnostic device measuring an analyte (procalcitonin), the "ground truth" for analytical studies is typically established by reference methods or gravimetric preparation with known concentrations. Experts are not typically involved in establishing ground truth for analytical performance like precision, linearity, LoD/LoQ, and interference.
  • For clinical studies (concordance and mortality prediction), the primary "ground truth" for the device's performance against clinical decision points is the predicate device's measurement (B.R.A.H.M.S PCT sensitive KRYPTOR assay). Physician discretion and clinical assessments were used for patient diagnosis and mortality outcomes in the MOSES study, which provided the samples. While physicians made clinical diagnoses, they were not experts establishing a "ground truth score" for the device; rather, clinical outcomes (mortality, diagnosis of sepsis/septic shock) were endpoints. No individual "experts" were formally used to establish a ground truth for the test set, but rather the clinical outcomes and the predicate device's results served as the reference.

4. Adjudication Method

  • Not applicable in the context of this 510(k) summary for an in vitro diagnostic device assessing an analyte level. Adjudication methods like 2+1 or 3+1 are typically used for subjective evaluations (e.g., imaging interpretation) where expert consensus is needed to determine the correct ground truth for a given case. Here, the ground truth for analytical performance is quantitative, and for clinical performance, it is the predicate device's measurement and observed clinical outcomes (e.g., 28-day mortality).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not performed. This device is an in vitro diagnostic assay, not an AI-assisted diagnostic tool for human readers. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" does not apply.

6. Standalone Performance Study

  • Yes, the performance described is a standalone performance study of the VITROS B·R·A·H·M·S PCT Reagent Pack and Calibrators. The "algorithm only" in this context refers to the assay's ability to quantitatively measure PCT levels. All analytical performance studies (precision, linearity, detection limits, analytical specificity, matrix comparison, high-dose hook, sample auto-dilution, carry-over) were conducted to demonstrate the standalone performance of the VITROS B·R·A·H·M·S PCT test. The clinical concordance with the predicate device also serves as a standalone performance benchmark.

7. Type of Ground Truth Used

  • Analytical Ground Truth: For analytical performance studies (precision, linearity, LoD/LoQ, interference, matrix equivalence, high dose hook, carry over), the ground truth was established by:
    • Reference calibrators: Traceable to in-house reference calibrators, which were value-assigned to correlate to B·R·A·H·M·S PCT sensitive KRYPTOR.
    • Known concentrations: Prepared by gravimetric spiking with recombinant PCT.
    • Validated analytical methods: Following CLSI guidelines (e.g., EP05-A3 for precision, EP17-A2 for detection limits, EP06-A for linearity, EP07-A3 for interference).
  • Clinical Ground Truth:
    • Predicate device results: For clinical concordance, the measurements from the B·R·A·H·M·S PCT sensitive KRYPTOR assay were considered the reference for comparison.
    • Outcomes data: For the 28-day mortality prediction claim, the vital status (mortality) of patients at Day 28 was the ground truth. This outcome data was collected during the original MOSES study.

8. Sample Size for the Training Set

  • This 510(k) document describes a traditional in vitro diagnostic device clearance, not an AI/ML software submission that often explicitly details "training sets."
  • However, the device's calibration is "traceable to in-house reference calibrators, which have been value-assigned to correlate to B·R·A·H·M·S PCT sensitive KRYPTOR." The development of these in-house reference calibrators and the assay itself would have involved some form of "training" or optimization using an internal dataset, but its size and specific characteristics are not provided in this regulatory summary.
  • The clinical study samples from the MOSES study were used for validation (evaluation) of the device's clinical performance, not as a training set for the assay's underlying methodology.

9. How the Ground Truth for the Training Set Was Established

  • As noted above, a formal "training set" in the AI/ML sense is not described. The ground truth for the development of the assay (e.g., for calibrator assignment) would have been established through a combination of:
    • Reference materials: Highly characterized procalcitonin standards.
    • Defined analytical methods: Using established laboratory practices and potentially comparing to existing, well-regarded PCT assays during the R&D phase to create the "in-house reference calibrators."
    • Value assignment: A process where the concentration of an analyte in a calibrator is accurately determined. This often involves multiple measurements using reference methods or by gravimetric preparation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 25, 2020

Ortho Clinical Diagnostics Ann Quinn Director, Regulatory Affairs 100 Indigo Creek Drive Rochester, New York 14626

Re: K200236

Trade/Device Name: VITROS BRAHMS PCT Reagent Pack and Calibrators Regulation Number: 21 CFR 866.3215 Regulation Name: Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis Regulatory Class: Class II Product Code: PRI, NTM, PMT Dated: January 29, 2020 Received: January 30, 2020

Dear Ann Quinn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kristian Roth, Ph.D. Branch Chief Bacterial Multiplex and Medical Counter Measures Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K200236

Device Name VITROS B.R.A.H.M.S PCT Reagent Pack

Indications for Use (Describe) Intended Use For in-vitro diagnostic use only.

For the quantitative measurement of procalcitonin (PCT) in human serum and plasma (lithium heparin and EDTA) using the VITROS 3600 Immunodiagnostic System.

Used in conjunction with other laboratory findings and clinical assessments, the VITROS B R A PA - M S PCT test is intended for use as follows:

· to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock,

· to aid in assessing the cumulative 28-day risk of all-cause mortality for patients diagnosed with severe sepsis or septic shock in the ICU or when obtained in the emergency department or other medical wards prior to ICU admission, using a change in PCT level over time,

· to aid in decision making on antibiotic therapy for patients with suspected or confirmed lower respiratory tract infections (LRTI) defined as community-acquired pneumonia (CAP), acute bronchitis, and acute exacerbation of chronic obstructive pulmonary disease (AECOPD) - in an inpatient setting or an emergency department,

· to aid in decision making on antibiotic discontinuation for patients with suspected or confirmed sepsis.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)×
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

1. Submittername,address,contactOrtho-Clinical Diagnostics, Inc100 Indigo Creek DriveRochester, NY 14626P: (585) 453-4152;F: (585) 453-3368Contact Person: Ann M. Quinn
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
  • January 29, 2020 2. Preparation Date
  • VITROS B·R·A·H·M·S PCT 3. Common Name
  • A. 510(k) Number: K200236
  • B. Purpose for Submission: Procalcitonin assay for use on VITROS Systems
  • C. Measurand(s): Procalcitonin (PCT)
  • D. Types of Test: Quantitative
  • E. Applicant:

Ortho-Clinical Diagnostics, Inc.

F. Proprietary and Established Name:

VITROS® Immunodiagnostic Products B·R·A·M·S PCT Reagent Pack and Calibrators

G. Regulatory Information:

ProductCodeClassRegulation SectionPanel
PRI,PMT,NTMII21 CFR 866.3215, device to detectand measure non-microbialanalyte(s) in human clinicalspecimens to aid in assessment ofpatients with suspected sepsis.Microbiology (83)

Ortho-Clinical Diagnostics, Inc.

{4}------------------------------------------------

H. Intended Use:

    1. Intended use:
      For in vitro diagnostic use only.

For the quantitative measurement of procalcitonin (PCT) in human serum and plasma (lithium heparin and EDTA) using the VITROS 3600 Immunodiagnostic System.

Used in conjunction with other laboratory findings and clinical assessments, the VITROS B·R·A·H·M·S PCT test is intended for use as follows:

  • to aid in the risk assessment of critically ill patients on their first day of ICU ● admission for progression to severe sepsis and septic shock,
  • . to aid in assessing the cumulative 28-day risk of all-cause mortality for patients diagnosed with severe sepsis or septic shock in the ICU or when obtained in the emergency department or other medical wards prior to ICU admission, using a change in PCT level over time,
  • . to aid in decision making on antibiotic therapy for patients with suspected or confirmed lower respiratory tract infections (LRTI) defined as communityacquired pneumonia (CAP), acute bronchitis, and acute exacerbation of chronic obstructive pulmonary disease (AECOPD) - in an inpatient setting or an emergency department,
  • . to aid in decision making on antibiotic discontinuation for patients with suspected or confirmed sepsis.

2. Indications for use:

See intended use above.

3. Special conditions for use statement:

For prescription use only.

Warnings and Precautions - Test Interpretation

  • · VITROS B•R•A•H•M•S PCT is not indicated to be used as a stand-alone diagnostic test and should be used in conjunction with clinical signs and symptoms of infection and other diagnostic evidence. In cases where the laboratory results do not agree with the clinical picture or history, additional tests should be performed.
  • · Decisions regarding antibiotic therapy should NOT be based solely on procalcitonin concentrations.
  • · PCT results should always be interpreted in the context of the clinical status of the patient and other laboratory results.
  • · Changes in PCT levels for the prediction of mortality, and overall mortality, are strongly dependent on many factors, including pre-existing patient risk factors and clinical course.
  • · The need to continue Intensive Care Unit (ICU) care at Day 4 and other covariates (e.g., age and Sequential Organ Failure Assessment (SOFA) score) are also significant predictors of 28-day cumulative mortality risk.

{5}------------------------------------------------

  • · The safety and performance of PCT-guided therapy for individuals younger than age 18 years, pregnant women, immunocompromised individuals or those on immunomodulatory agents, was not formally analyzed in the supportive clinical trials.
  • · Severity of renal failure or insufficiency, may influence procalcitonin values and should be considered as potentially confounding clinical factors when interpreting PCT values.
  • · PCT levels may not be elevated in patients infected by certain atypical pathogens, such as Chlamydophila pneumoniae and Mycoplasma pneumoniae.

Increased PCT levels may not always be related to systemic infection. Patients with increased PCT levels due to other conditions include, but are not limited to:

  • · Patients experiencing major trauma and/or recent surgical procedure including extracorporeal circulation or burns;
  • · Patients under treatment with OKT3 antibodies, OK-432, interleukins, TNF-alpha and other drugs stimulating the release of pro-inflammatory cytokines or resulting in anaphylaxis;
  • · Patients diagnosed with active medullary C-cell carcinoma, small cell lung carcinoma, or bronchial carcinoid;
  • · Patients with acute or chronic viral hepatitis and/or decompensated severe liver cirrhosis (Child-Pugh Class C);
  • Patients with prolonged or severe cardiogenic shock, prolonged severe organ . perfusion anomalies or after resuscitation from cardiac arrest;
  • . Patients receiving peritoneal dialysis or hemodialysis treatment;
  • · Patients with biliary pancreatitis, chemical pneumonitis or heat stroke;
  • · Patients with invasive fungal infections (e.g. candidiasis, aspergillosis) or acute attacks of plasmodium falciparum malaria; and
  • · Neonates during the first 2 days of life.

4. Special instrument requirements:

The VITROS B·R·A·H·M·S PCT Reagent Pack and the VITROS B·R·A·H·M·S PCT Calibrators were validated on the VITROS 3600 Immunodiagnostic system only.

I. Device Description:

The VITROS B·R·A·H·M·S PCT test is performed using the VITROS B·R·A·H·M·S PCT Reagent Pack and the VITROS B·R·A·H·M·S PCT Calibrators on the VITROS Systems.

Reagent Pack Contents

1 reagent pack containing:

  • 100 coated wells (rat monoclonal anti-procalcitonin antibody, 1.0 µg/mL) ●
  • 10.20 mL assay reagent (buffer containing bovine gamma globulin, bovine serum ● albumin and antimicrobial agent)
  • . 13.10 mL conjugate reagent (HRP-conjugated mouse monoclonal procalcitonin antibody. 1.65 ug/mL in buffer with bovine serum albumin and antimicrobial agent)

Calibrator Contents

Ortho-Clinical Diagnostics, Inc. Page 3 of 26

{6}------------------------------------------------

  • . 3 sets of VITROS B•R•A•H•M•S PCT Calibrators 1 and 2, 1.0 mL, procalcitonin in buffer with antimicrobial agent, nominal values 0.080 and 75.0 ng/mL (ug/L)
  • . Lot calibration card
  • Protocol card ●
    • 16 calibrator bar code labels (8 for each calibrator) ●

Ouality Control

Controls containing suitable levels of procalcitonin are recommended for use with the VITROS Immunodiagnostic System. The performance of commercial control fluids should be evaluated for compatibility with this test before they are used for quality control. Control materials may show a difference when compared with other procalcitonin methods if they contain high concentrations of preservatives, stabilizers, or other non-physiological additives, or otherwise depart from a true human sample matrix. Appropriate quality control value ranges must be established for all quality control materials used with the VITROS B.R.A.H.M.S PCT test.

For analytical and clinical studies performed, each day of testing required that VITROS B·R·A·H·M·S PCT controls (levels 0.5, 2.0, 60) met specifications on all reagent lots.

Materials Required but Not Provided

  • VITROS Immunodiagnostic Products Signal Reagent ●
  • . VITROS Immunodiagnostic Products Universal Wash Reagent
  • VITROS Immunodiagnostic Products High Sample Diluent B
  • Quality Controls materials ●

J. Substantial Equivalence Information:

    1. Predicate device name: B.R.A.H.M.S PCT Sensitive KRYPTOR, K171338
    1. Comparison with predicate:
DeviceCharacteristicNew DeviceVITROS ImmunodiagnosticProducts B·R·A·H·M·S PCT TestPredicate DeviceB·R·A·H·M·S PCT SensitiveKRYPTOR, K171338
Similarities
Intended UseRx ONLYFor in vitro diagnostic use only.For the quantitative measurement ofprocalcitonin (PCT) in human serumand plasma (lithium heparin andEDTA) using the VITROS 3600Immunodiagnostic System.Used in conjunction with otherlaboratory findings and clinicalassessments, the VITROSThe B·R·A·H·M·S PCT sensitiveKRYPTOR® is an immunofluorescentassay using Time-Resolved AmplifiedCryptate Emission (TRACE) technologyto determine the concentration of PCT(procalcitonin) in human serum andEDTA or heparin plasma.The B·R·A·H·M·S PCT sensitiveKRYPTOR® is intended to be performed
DeviceCharacteristicNew DeviceVITROS ImmunodiagnosticProducts B-R・A・H・M・S PCT TestPredicate DeviceB-R-A-H-M・S PCT SensitiveKRYPTOR, K171338
B-R・A・H・M・S PCT test is intended foruse as follows: to aid in the risk assessmentof critically ill patients ontheir first day of ICUadmission for progression tosevere sepsis and septicshock, to aid in assessing thecumulative 28-day risk of all-cause mortality for patientsdiagnosed with severe sepsisor septic shock in the ICU orwhen obtained in theemergency department orother medical wards prior toICU admission, using achange in PCT level overtime, to aid in decision making onantibiotic therapy for patientswith suspected or confirmedlower respiratory tractinfections (LRTI) defined ascommunity-acquiredpneumonia (CAP), acutebronchitis, and acuteexacerbation of chronicobstructive pulmonary disease(AECOPD) – in an inpatientsetting or an emergencydepartment, to aid in decision making onantibiotic discontinuation forpatients with suspected orconfirmed sepsis.on the B-R·A·H·M·S KRYPTOR®analyzer family.Used in conjunction with otherlaboratory findings and clinicalassessments, B-R·A·H·M·S PCTsensitive KRYPTOR® is intended foruse as follows: to aid in the risk assessment ofcritically ill patients on their first day ofICU admission for progression to severesepsis and septic shock, to determine the change in PCT levelover time as an aid in assessing thecumulative 28-day risk of all-causemortality for patients diagnosed withsevere sepsis or septic shock in the ICUor when obtained in the emergencydepartment or other medical wards prior to aid in decision making on antibiotictherapy, for inpatients or patients in theemergency department with suspected orconfirmed lower respiratory tractinfections (LRTI) – defined ascommunity acquired pneumonia (CAP),acute bronchitis, and acute exacerbationof chronic obstructive pulmonary disease(AECOPD), to aid in decision making on antibioticdiscontinuation for patients withsuspected or confirmed sepsis.
AnalyteProcalcitoninSame
Sample TypeHuman serum and plasma (EDTA,lithium heparin)Same
AutomatedAutomated assaySame
DeviceCharacteristicNew DeviceVITROS ImmunodiagnosticProducts B·R·A·H·M·S PCT TestPredicate DeviceB·R·A·H·M·S PCT SensitiveKRYPTOR, K171338
Differences
MeasuringRangeDirect measuring range 0.030ng/mL - 100ng/mLMeasuring range with automatic dilution 0.030ng/mL - 1000ng/mLDirect measuring range 0.02µg/L - 50µg/LMeasuring range with automatic dilution 0.02µg/L - 5000µg/L
Sample volume30 µL50 µL
InstrumentVITROS 3600 ImmunodiagnosticSystemKRYPTOR Test System
Basic PrincipleTwo-step dual monoclonalimmunometric assayTwo antibody "sandwich" binding ofProcalcitonin.

{7}------------------------------------------------

{8}------------------------------------------------

K. Standard/Guidance Documents Referenced (if applicable):

  • CLSI. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved ● Guideline. CLSI guideline EP25-A. Wayne, PA: Clinical and Laboratory Standards Institute, 2009.
  • CLSI. Evaluation of Precision Performance of Quantitative Measurement ● Methods; Third Edition. CLSI guideline EP05-A3. Wayne, PA: Clinical and Laboratory Standards Institute, 2014.
  • . CLSI. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Second Edition. CLSI guideline EP17-A2. Wayne, PA: Clinical and Laboratory Standards Institute, 2012.
  • CLSI. Evaluation of the Linearity of Quantitative Measurement Procedures: A ● Statistical Approach. CLSI document EP06-A. Wayne, PA: Clinical and Laboratory Standards Institute, 2003.
  • CLSI. Interference Testing in Clinical Chemistry Third Edition. CLSI guideline ● EP07-A3. Wayne, PA: Clinical and Laboratory Standards Institute, Clinical and Laboratory Standards Institute; 2018
  • . CLSI. Supplemental Tables for Interferent Testing in Clinical Chemistry - First Edition. CLSI supplement EP37. Wayne, PA: Clinical and Laboratory Standards Institute. Clinical and Laboratory Standards Institute: 2018
  • CLSI. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory Approved Guideline - Third Edition. CLSI guideline EP28-A3c.Wayne, PA: Clinical and Laboratory Standards Institute, 2010.
  • CLSI. Measurement Procedure Comparison and Bias Estimation Using Patient ● Samples. 3rd ed. CLSI guideline EP09c. Wayne, PA: Clinical and Laboratory Standards Institute; 2018.

L. Test Principles:

A two-step immunometric technique is used, which involves the reaction of procalcitonin present in the sample with a biotinylated anti-procalcitonin antibody (rat monoclonal antiprocalcitonin) bound to streptavidin coated on a microwell in the first step. Unbound

{9}------------------------------------------------

materials are removed by washing. The second step involves the reaction of antigen-antibody complex with a horseradish peroxidase (HRP)-labeled antibody conjugate (mouse monoclonal anti-procalcitonin). Unbound materials are removed by washing. The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrates (a luminol derivative and a peracid salt) and an electron transfer agent, is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron transfer agent (a substituted acetanilide) increases the level of light produced and prolongs its emission. The light signals are read by the system. The amount of HRP conjugate bound is directly proportional to the concentration of procalcitonin present.

M. Performance Characteristics (if/when applicable):

    1. Analytical performance:
    • Precision/Reproducibility: a.

Precision

Precision was evaluated consistent with CLSI document EP05-A3. Two replicates each of seven patient pools and three controls were tested on two separate occasions per day on at least 20 different test days. The experiment was performed using three reagent lots on one VITROS 3600 Immunodiagnostic System. Representative performance data are shown below.

Units = ng/mL (µg/L)
VITROSSystemMeanVITROSB•R•A•H•M•SPCT Conc.Within-run*Within-cal**Within-lab***No.ObservationsNo.Days
SD%CVSD%CVSD%CV
0.0410.00061.40.00184.30.00256.48020
0.0960.00101.00.00222.30.00293.18020
0.2410.00361.50.00522.10.00823.58020
0.4810.00731.50.01132.30.01743.78020
1.920.0291.50.0472.40.0703.78020
360027.90.421.50.562.00.893.28020
77.41.451.81.832.32.883.88020
0.4860.00701.40.01152.30.01653.48020
1.930.0442.30.0562.90.0764.08020
55.51.132.01.452.62.163.98020

Precision Data Summary: VITROS 3600 Immunodiagnostic System

  • Within-run (repeatability). Between Duplicate precision averaged over all runs.

** Within-calibration. Total precision with weighted components of within-run, between-run, and between-day variation.

*** Within-lab. A measure of the effect of recalibration on total precision, calculated within reagent lot, using data from at least 4 calibrations.

Ortho-Clinical Diagnostics, Inc. Page 7 of 26

{10}------------------------------------------------

Multi-Site Precision

Multi-site precision was evaluated incorporating between site and between day variations. The study was performed at three testing sites (two external and one internal sites) using one VITROS B·R·A·H·M·S PCT test lot. Five replicates each of seven panel members were tested once per day on five different days. The repeatability, between day, between site, and reproducibility precision estimates (CV (%)) were derived from a variance component analysis.

The repeatability (within day), between day, between site, and reproducibility (total) precision estimates (CV (%)) derived from a variance component analysis were summarized as shown in the table below.

Multi-Site Precision Data Summary: VITROS 5600 Immunodiagnostic System
SampleMean PCTResults(ng/mL)NRepeatability *Between Day **Between Site ***Reproducibility****
SDCV (%)SDCV (%)SDCV (%)SDCV (%)
R10.484750.0091.90.0214.40.0091.90.0255.2
R21.854750.0492.60.24413.20.0000.00.24913.4
R352.028751.4612.86.03411.64.0197.77.39514.2
R40.093750.0022.00.01010.80.01111.50.01515.9
R50.223750.0052.10.0115.10.0135.70.0187.9
R60.442750.0071.70.0153.40.0122.70.0214.6
R71.736750.0291.70.0482.80.0694.00.0895.1

Multi-Site Precision Data Summary: VITROS 3600 Immunodiagnostic System

Abbreviation: %CV: Coefficient of variation expressed as a percentage: N: Total number of test results obtained in the study (across site, days and replicate); SD: Standard deviation.

*Repeatability: Variability of the VITROS B·R·A·H·M·S PCT test performance within day replicate, calculated using data across all sites

** Between Day: Variability of the VITROS B-R-A-H·M·S PCT test performance from day to day, calculated using data across all sites.

*** Between site: Variability of the VITROS B·R·A·H·M·S PCT test performance from site to site.

*** Reproducibility: Variability of the test incorporating factors of site, and day (Total).

b. Traceability/Expected Values/Stability

Traceability

Calibration of the VITROS B.R.A.H.M.S PCT test is traceable to in-house reference calibrators, which have been value-assigned to correlate to B•R•A•M•M•S PCT sensitive KRYPTOR.

Expected values/Reference range:

The overall observed 95th percentile Upper Reference Limit (URL) from 150 normal, healthy donor samples is 0.077 ng/mL as shown in the table below.

{11}------------------------------------------------

Sample TypeNumber ofSubjects95th Percentile URLng/mL
Serum, lithium heparinplasma, EDTA plasma1500.077

VITROS B·R·A·H·M·S PCT URL

Sample Stability

Serum, lithium heparin plasma, and EDTA plasma samples can be stored for up to 24 hours at room temperature (15-30 ℃ or 59-86 °F), up to 48 hours at 2-8 °C (36-46 °F), or up to 1 month at -20 °C (-4 °F) and at -20 °C (-4 °F) with up to 4 freeze thaw cycles before analysis.

Reagent Stability

The data collected in support of Long-Term Shelf Life of Reagent Pack and Calibrators demonstrate acceptable performance up to the 20 week timepoint. The shelf life may be extended following collection of acceptable data at additional time points according to the protocol summarized above.

The data collected in support of On-Board storage of Reagent Packs demonstrated acceptable performance up to 12 weeks on board the VITROS System Open off board storage of Calibrators supports the storage of Calibrators for up to 18 weeks stored at <-20°C with up to 3 freeze-thaws or 13 weeks at 2-8°C.

c. Detection Limits:

The Limit of Detection (LoD) for the VITROS B·R·A·H·M·S PCT test is 0.007 ng/mL (0.007 ug/L), determined consistent with CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Second edition (2012). The Limit of Quantitation (LoQ) was determined consistent with CLSI EP17-A2. The observed Limit of Quantitation at 20% CV was determined to be 0.013 ng/mL (0.013 ug/L) and the claimed LoQ was set at 0.030 ng/mL (0.030 ug/L).

LoDLoO
ng/mLug/Lng/mLug/L
0.0070.0070.0300.030

Limit of Detection and Limit of Quantitation

d. Linearity/assay measuring range:

Linearity studies were performed according to CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (2003). The VITROS Immunodiagnostics Products B·R·A·M·S PCT test was tested on the VITROS System. A low and high sample pool was prepared and mixed to give twelve (12) further pools of intermediate concentrations. The low and high linearity pools and the interim dilutions between the low and high linearity pools were assayed in triplicate.

{12}------------------------------------------------

Statistical regression coefficients for the 18, 2nd, and 3rd order regressions using nonweighted concentrations and predictions based on weighted statistical regressions are shown in the following tables.

The VITROS B·R·A·H·M·S PCT test was linear over the measuring range, from 0.030 to 100 ng/mL.

VITROS B.R.A.H:M:S PCT Test Statistical Regression Analysis for 1st, 2nd and 3rd Order Regressions

OrderCoefficientCoefficient ValueCoefficient SEp-valueStd Error Regression (Sy.x)
1stb00.894450.5560.000
1stb11.281830.01230.0001.05693
2ndb00.544960.6880.000
2ndb11.319690.04460.000
2ndb2-0.000500.0005640.4061.07160
3rdb0-0.063350.5620.000
3rdb11.497510.07610.000
3rdb2-0.006690.002420.033
3rdb30.000050.0000200.0410.794911

VITROS B·R·A·H·M·S PCT Test Predictions and Bias Results from Regressions Using Weighted Concentrations

% HighPoolMeanMeasuredConc.(ng/mL)WeightWeighted Predicted Conc.Weighted Bias
1st Order3rd Order3rd_1st%Bias*
801040.231107104-3.437-3.3
7090.20.25493.990.1-3.815-4.2
6076.10.82480.577.1-3.382-4.4
5064.73.57167.164.7-2.430-3.8
4054.02.15853.752.4-1.247-2.3
3039.63.29740.340.1-0.124-0.3
2027.48.10826.927.50.6522.4
1014.175.00013.414.20.7905.6
11.561200.0001.371.500.1298.3
00.026923645.3200.0260.0260.000-0.8

*Acceptance Criteria = ±10% bias

Ortho-Clinical Diagnostics, Inc.

{13}------------------------------------------------

The Percent Total Error of the VITROS Immunodiagnostic Products B.R.A.H.M.S PCT (VITROS B·R·A·H·M·S PCT) test was based on the precision profiles from the Limit of Quantitation study, the Within-Lab %CVs from the precision studies, and the Deming Regression statistics from the method comparison study to the predicate method, the B·R·A·H·M·S PCT sensitive KRYPTOR assay. The results obtained using one VITROS B-R-A-H-M-S PCT Master Lot tested on a VITROS 3600 Immunodiagnostic System (3600) are shown in the table below.

PCT Level (ng/mL)Bias (%)CV (%)Total Error (%)
0.03027.94.134.7
0.1004.23.19.3
0.2501.93.57.7
0.5004.03.710.1
2.005.53.711.6

Percent Total Error

e. Analytical specificity:

The VITROS B.R.A.H.M.S PCT test was evaluated for interference consistent with CLSI document EP07-A3. Of the compounds tested, none were found to cause a bias of >10% with the test at the concentrations indicated at nominal procalcitonin concentrations of 0.250 ng/mL and 2.00 ng/mL.

CompoundConcentrationCompoundConcentration
Acetaminophen200 µg/mL1323 µmol/LHemoglobin600 mg/dL6.00 g/L
Acetylsalicylic Acid65.2 mg/dL3.62 mmol/LHeparin8000 IU/LN/A
Alcohol400 mg/dL86.8 mmol/LIbuprofen50.0 mg/dL2.42 mmol/L
Azithromycin1.15 mg/dL14.6 µmol/LImipenem1.18 mg/mL3.72 mmol/L
Bilirubin, Conjugated30.0 mg/dL513 µmol/LLevofloxacin1.75 mg/dL47.2 µmol/L
Bilirubin, Unconjugated40.0 mg/dL475 µmol/LLoratadine0.030 mg/dL0.784 µmol/L
Biotin3500 ng/mL14.3 µmol/LNicotine0.100 mg/dL6.20 µmol/L
Caffeine5.98 mg/dL308 µmol/LNoradrenaline2.00 µg/mL11.8 µmol/L
Celecoxib24.0 mg/dL629 µmol/LOxymetazoline HCI0.009 mg/dL0.334 µmol/L
Cetirizine HCI0.360 mg/dL7.80 µmol/LPhenylephrine0.018 mg/dL1.10 µmol/L
Dextromethorphan0.140 mg/dL3.80 µmol/LPrednisolone0.300 mg/dL8.31 µmol/L
Dobutamine11.2 µg/mL37.2 µmol/LRheumatoid Factor2000 IU/mLN/A
Dopamine13.0 mg/dL686 µmol/LSalmeterol60.0 ng/mL0.099 µmol/L
Doxycycline50.0 mg/L104 µmol/LTiotropium21.6 ng/mL0.046 µmol/L
Epinephrine0.180 mg/dL8.20 µmol/LTotal Protein11.7 g/dLN/A
Fentanyl10.0 mg/L29.7 µmol/LTriglyceride2160 mg/dL24.4 mmol/L
Furosemide2.00 mg/dL60.5 µmol/L

Interference testing

{14}------------------------------------------------

CompoundConcentrationCompoundConcentration
HAMA (HumanAnti- MouseAntibody)3600 ng/mL0.024 µmol/LVancomycin2.60 mg/mL1.75 mmol/L

The cross-reactivity of the VITROS B.R.A.H.M.S PCT test was evaluated by adding the following substances to one human serum sample pool containing no procalcitonin.

Cross Reactivity Testing -No Procalcitonin

Cross-ReactantCross ReactantConcentrationng/mLMean Result ofControl Poolng/mLMean Result ofCross-Reactant Poolng/mL% Cross-Reactivity
Human Calcitonin3.90 ng/mL***
Human Katacalcin25.6 ng/mL***
Human α-CGRP30.0 ng/mL***
Human β-CGRP30.0 ng/mL***
  • Not Detectable (ND). Concentration was below the measuring range of the test, 0.030-100 ng/mL.

The cross-reactivity of the VITROS B·R·A·H·M·S PCT test was evaluated by adding the following substances to one human serum sample pool containing procalcitonin at a concentration of 0.500 ng/mL.

Cross Reactivity Testing -With Procalcitonin

Cross-ReactantCross ReactantConcentrationMean Result ofControl PoolMean Result ofCross-Reactant Pool% Cross-Reactivity
ng/mLng/mLng/mL
Human Calcitonin3.90 ng/mL0.4910.461-0.8
Human Katacalcin25.6 ng/mL0.4610.4680.0
Human α-CGRP30.0 ng/mL0.4910.460-0.1
Human β-CGRP30.0 ng/mL0.4910.467-0.1

Cross-reactivity was expressed as the mean result obtained for the cross-reactant pool minus the mean result obtained for the control sample divided by the crossreactant concentration in percentage term.

% Cross-reactivity = (Mean Procalcitonin Result Cross-reactant Pool) - (Mean Procalcitonin Result Control Sample) x 100 Concentration of Cross-Reactant

  • f. Assay cut-off:

Risk assessment for progression to severe sepsis and septic shock

PCT >2.00 ng/mL

A PCT level above 2.00 ng/mL on the first day of Intensive Care Unit (ICU) admission is associated with a high risk for progression to severe sepsis and/or septic shock.

{15}------------------------------------------------

PCT <0.500 ng/mL

A PCT level below 0.500 ng/mL on the first day of Intensive Care Unit (ICU) admission is associated with a low risk for progression to severe sepsis and/or septic shock.

Note: PCT levels below 0.500 ng/mL do not exclude an infection, because localized infections (without systemic signs) may also be associated with such low levels. If the PCT measurement is done very early after the systemic infection process has started (usually < 6 hours), these values may still be low.

Various non-infectious conditions are known to induce changes in PCT level. PCT levels between 0.500 ng/mL and 2.00 ng/mL should be interpreted in the context of the specific clinical background and condition(s) of the individual patient. It is recommended to retest PCT within 6-24 hrs if any concentrations <2.00 ng/mL are obtained.

28-Day Mortality

△PCT ≤80%

A decrease of PCT levels below or equal to 80% defines a positive ΔPCT test result representing a higher risk for 28-day all-cause mortality of patients diagnosed with severe sepsis or septic shock. If the PCT level increases over the first 4 days, the change in PCT result (ΔPCT) is interpreted as ΔPCT decline <80% and is defined a positive ΔΡCT test result representing a higher risk for 28-day all-cause mortality of patients diagnosed with severe sepsis or septic shock.

ДРСТ >80%

A decrease of PCT levels of more than 80% defines a negative ΔPCT result representing a lower risk for 28-day all-cause mortality of patients diagnosed with severe sepsis or septic shock.

Use the Change in Procalcitonin Calculator (http://www.B-R-A-H-M-S-PCT-Calculator.com) to determine ΔPCT results from the absolute PCT concentrations of a patient obtained on the day severe sepsis or septic shock was first diagnosed (or 24 hours later) and four days thereafter.

The PCT level on Day 1 (the day after severe sepsis or septic shock is first clinically diagnosed) can be used to calculate the percent change in PCT level at Day 4 if the Day 0 measurement is unavailable.

LRTI Antibiotic Decision Making

Initiation:

{16}------------------------------------------------

PCT Result<0.100 ng/mL0.100-0.250 ng/mL0.251-0.500 ng/mL>0.500 ng/mL
InterpretationAntibiotic therapystrongly discouraged.Antibiotic therapydiscouraged.Antibiotic therapyencouraged.Antibiotic therapystrongly encouraged.
Follow-upAntibiotic therapy should be consideredregardless of PCT result if the patient isclinically unstable, is at high risk foradverse outcome, has strong evidence ofbacterial pathogen, or the clinical contextindicates antibiotic therapy is warranted.If antibiotics are withheld, reassess ifsymptoms persist/worsen and/or repeatPCT measurement within 6-24 hours.In order to assess treatment success and tosupport a decision to discontinue antibiotictherapy, follow up samples should be testedonce every 1-2 days*, based upon physiciandiscretion taking into account patient'sevolution and progress.Antibiotic therapy may be adjusted accordingto the description below:

*Schuetz, P., Birkhahn, R., Sherwin, R., Jones, A., Kline, J. A., & Gaieski, D. F. (2017). Serial Procalcitonin Predicts Mortality in Severe Sepsis Patients: Results from the Multicenter Procalcitonin MOnitoring SEpsis (MOSES) Study. Critical Care Medicine.

Discontinuation:

Antibiotic therapy may be discontinued if the PCT is <0.250 ng/mL or if the ΔPCT is >80% .

Decision making on antibiotic discontinuation for suspected or confirmed septic patients

Antibiotic therapy may be discontinued if the PCT is ≤0.500 ng/mL or if the ΔPCT is >80% .

Antibiotic therapy may be continued based upon other clinical findings, such as failure to control a local infection, or ongoing physiologic instability. If clinical picture has not improved, and PCT remains high, re-evaluate and consider treatment failure or other causes.

  • g. High dose Hook:
    A human serum-based matrix, containing no measurable procalcitonin (PCT), was spiked with recombinant PCT measured gravimetrically to a concentration of 5,000 ng/mL and then diluted to create a panel of six fluids (seven samples total) having concentrations ranging from 25 to 5,000 ng/mL. These high dose hook fluids were assayed in triplicate using one Master Lot. No evidence of high dose hook was observed up to 5,000 ng/mL.

These results support the claim that the VITROS B·R·A·H·M·S PCT test will not hook back into the measuring range with samples that have a concentration up to 5,000 ng/mL (5,000 µg/L).

  • h. Sample Auto Dilution Study:
    Two studies were performed to assess high sample dilution. The first study was designed to verify the recovery of automated (on-board) dilution to manual dilution.

{17}------------------------------------------------

The second study was designed to assess the recovery of automated (on-board) diluted samples compared to the neat sample result.

In the first study, six (6) patient samples with procalcitonin (PCT) concentrations above the expected upper end of the assay measuring range (>100 ng/mL) of the VITROS B·R·A·H·M·S PCT test were automatically and manually diluted 1:5 and 1:10 with VITROS Immunodiagnostic Products High Sample Diluent B (HSDB). Triplicate determinations of both the automated and manual dilution series were made, using one Master Lot. Dilution recovery was assessed by comparing the automated to manual dilution results.

In the second study, five (5) patient samples with procalcitonin concentrations within the assay measuring range were automatically diluted 1:5 and 1:10 with VITROS Immunodiagnostic Products High Sample Diluent B (HSDB). Samples within the assay measuring range were used for this study to obtain an accurate value of the neat values. Triplicate determinations of both the automated dilution series were made, using one Master Lot. Dilution recovery was assessed by comparing the automated dilution result to the neat result.

The study results demonstrate that samples with procalcitonin concentrations above the expected upper measuring range can be automatically (on-board) diluted with VITROS High Sample Diluent B to either 5-fold, or 10-fold with acceptable recovery.

  • i. Carry Over Study:
    Ten (10) replicates of a sample above the expected upper end of the assay measuring range were run interleaved with ten (10) replicates of a known blank sample. These twenty total interleaved determinations were made, using one Master Lot. The study results demonstrate that the VITROS B·R·A·M·S PCT test is not susceptible to high procalcitonin concentration samples affecting the predicted concentrations of low procalcitonin samples through the process of sample carryover.

2. Comparison studies:

  • a. Method comparison with predicate device:
    The quantitative comparison for the VITROS Immunodiagnostic Products B.R.A.H:M:S PCT test (VITROS B·R·A·H·M·S PCT) was established in accordance with CLSI EP09c Measurement Procedure Comparison and Bias Estimation Using Patient Samples, using the B.R.A.H.M.S PCT sensitive KRYPTOR assay as the comparative method. Data presented were generated using one Master Lot of the VITROS B·R·A·H·M·S PCT test on a VITROS 3600 Immunodiagnostic System (3600). The relationship between the methods was determined by Weighted Deming and Passing & Bablok regression analyses and Pearson correlation. The table shows the results of a method comparison study using patient samples analyzed on the

{18}------------------------------------------------

VITROS 3600 Immunodiagnostic System compared with those analyzed using the B·R·A·H·M·S PCT sensitive KRYPTOR test.

Method Comparison
VITROSB·R·A·H·M·S PCTValue (ng/mL)Number ofSamples
0 - ≤ 0.25043
> 0.250 - ≤ 0.50048
> 0.500 - ≤ 2.0054
> 2.00 - ≤ 10.051
> 10.0 - ≤ 20.018
> 20.0 - ≤ 50.019
> 50.0 - ≤ 10014
> 100 - ≤ 25014
> 250 - ≤ 10005
Total266

Sample Distribution for

VITROS B.R.A.H.M.S.PCT Test Regression Results vs. the B.R.A.H.M.S.PCT sensitive KRYPTOR assay including Samples outside the Measuring Range

N266
Range of Samples (KRYPTOR)0.027 to 437 ng/mL
Range of Samples (VITROS)0.031 to 407 ng/mL
Correlation Coefficient ( r )0.995
Mean Percent Bias1.31%
Passing & Bablok RegressionWeighted Deming Regression
Slope (95% CI two-sided)1.001 (0.9773 to 1.027)1.046 (1.025 to 1.066)
Intercept (95% CI two-sided)0.01041 (0.001562 to 0.03272)-0.009264 (-0.01498 to -0.003549)

VITROS B·R·A·H·M·S PCT Test Regression Results vs. the B·R·A·H·M·S PCT sensitive KRYPTOR assay Using Samples Within Measuring Range

N246
Range of Samples (KRYPTOR)0.027 to 94.28 ng/mL
Range of Samples (VITROS)0.031 to 99.80 ng/mL
Correlation Coefficient ( r )0.994
Mean Percent Bias1.88%
Passing & Bablok RegressionWeighted Deming Regression
Slope (95% CI two-sided)1.025 (1.002 to 1.054)1.057 (1.035 to 1.078)
Intercept (95% CI two-sided)0.004237 (-0.004908 to 0.01905)-0.009994 (-0.01585 to -0.004140)

Results presented below are using Weighted Deming analysis

Ortho-Clinical Diagnostics, Inc.

{19}------------------------------------------------

SystemnSlopeCorrelationCoefficientConventional Units (ng/mL) =Alternate Units (µg/L)*
Range ofSamplesIntercept
VITROS 3600 vs. Comparative Method2461.0570.9940.031 - 99.8-0.010

*The alternate units are 1.00 ng/mL = 1.00 µg/L.

Image /page/19/Figure/4 description: This image is a scatter plot comparing two methods of measurement. The x-axis represents the Comparative Method: BRAHMS PCT sensitive KRYPTOR (ng/mL), and the y-axis represents VITROS 3600 Master Lot 8 (ng/mL). The plot includes a "Weighted Deming fit" line, described by the equation y = -0.009994 + 1.057x, along with a gray line that appears to represent a line of equality.

  • b. Matrix comparison:
    A total of 83 samples covering the measuring range of the VITROS Immunodiagnostic Products B-R-A-H·M·S PCT (VITROS B·R·A·H·M·S PCT) test were used to assess differences in sample matrix between serum, lithium heparin, and EDTA.

Eleven (11) of the eighty-three (83) samples were unaltered prior to testing. The remaining seventy-two (72) samples were spiked with recombinant procalcitonin

{20}------------------------------------------------

(PCT) to ensure that sample concentrations covering the full measuring range of the assay were tested.

The results used for this analysis met the acceptance criteria that the comparison between serum, lithium heparin, and EDTA samples spanning the reportable range shall demonstrate less than 10% bias from serum based on regression analysis (slope of 0.90 - 1.10). Based on the analysis, the results indicate that serum and plasma (lithium heparin and EDTA) matrices are suitable for use with the VITROS B·R·A·H·M·S PCT test.

VITROS 3600
SerumLithium HeparinEDTA
Passing and Bablok AnalysisSlope0.9800.992
Corr. Coef (r)1.0001.000
n8383

Comparison to Serum, Passing and Bablok Regression

Comparison to Serum, Weighted Deming Regression

VITROS 3600
SerumLithium HeparinEDTA
Weighted Deming AnalysisSlope0.9790.990
Corr. Coef (r)1.0001.000
n8383
    1. Clinical studies:
      The clinical performance study was conducted to establish the clinical performance characteristics of the VITROS® Immunodiagnostic Products B·R·A·H·M·S PCT Reagent Pack and VITROS® Immunodiagnostic Products B.R.A.H.M.S PCT Calibrators (VITROS B·R·A·H·M·S PCT test) using serial sample sets obtained from the Multicenter Procalcitonin MOnitoring SEpsis (MOSES) Study collection, a well-characterized sample collection in 13 sites across the United States. The study enrolled patients with severe sepsis and septic shock who were admitted to the intensive care unit (ICU) from the emergency department, other wards or directly from out of the hospital. Blood samples were collected (when possible) for each subject on Day 0 (within 12 hours after diagnosis), Day 1, Day 3 and Day 4. To verify vital status, subjects were followed during the hospital stay and contacted by telephone at Day 28. See DEN150009 for more details. Available EDTA samples collected from each consented subject were evaluated in this clinical performance study.

EDTA samples were tested with the VITROS B·R·A·H·M·S PCT test on the VITROS 3600 Immunodiagnostic System at one internal testing site. The technologists performing the VITROS B·R·A·H·M·S PCT test were unaware of the subjects' clinical outcomes. If

{21}------------------------------------------------

the VITROS B·R·A·H·M·S PCT test gave a result above the test measuring range, the sample was diluted as volume allowed and was repeated.

The MOSES Study samples were previously tested with the B.R.A.H.M.S.PCT sensitive KRYPTOR assay, the predicate method, and the existing data was used in the concordance study. No additional testing with the B.R.A.H:M·S PCT sensitive KRYPTOR assay was performed in this study due to sample volume limitations.

  • a. Clinical Sensitivity:
    Clinical Concordance between VITROS B·R·A·H·M·S PCT and B·R·A·H·M·S PCT sensitive KRYPTOR

Clinical concordance of the VITROS B·R·A·H·M·S PCT test to the B·R·A·H·M·S PCT sensitive KRYPTOR test was evaluated at relevant clinical decision points, i.e., 0.100 ng/mL, 0.250 ng/mL, 0.500 ng/mL and 2.00 ng/mL. Results are summarized in the tables below:

3x3 Concordance VITROS B.R.A.H.M.S.PCT versus B.R.A.H.M.S.PCT sensitive KRYPTOR

B·R·A·H·M·S PCT sensitive KRYPTOR
VITROS B·R·A·H·M·S PCT≤ 0.500 ng/mL> 0.500 ng/mL to ≤ 2.00 ng/mL> 2.00 ng/mLTotal
≤ 0.500 ng/mL500121513
> 0.500 ng/mL to ≤ 2.00 ng/mL444056455
> 2.00 ng/mL04011601200
Total54445711672168

5x5 Concordance VITROS B·R·A·H·M·S PCT versus B·R·A·H·M·S PCT sensitive KRYPTOR

VITROSB·R·A·H·M·S PCTB·R·A·H·M·S PCT sensitive KRYPTORTotal
B·R·A·H·M·S PCT≤0.100ng/mL> 0.100 ng/mLto 0.250 ng/mL> 0.250 ng/mLto 0.500 ng/mL> 0.500 ng/mLto 2.00 ng/mL> 2.00ng/mLTotal
≤0.100 ng/mL622300085
> 0.100 ng/mL to0.250 ng/mL72011601225
> 0.250 ng/mL to0.500 ng/mL220169120203
> 0.500 ng/mL to2.00 ng/mL13404056455
> 2.00 ng/mL0004011601200
Total7224722545711672168

Clinical Agreement between VITROS B·R·A·H·M·S PCT and B·R·A·H·M·S PCT Sensitive KRYPTOR at Relevant Clinical Decision Points

{22}------------------------------------------------

ClinicalDecision PointPositive Agreement(95% CI)Negative Agreement(95% CI)TotalAgreementCohen's Kappa
0.100 ng/mL98.9%(98.4 - 99.3%)86.1%(75.9 - 93.1%)98.5%0.772
0.250 ng/mL99.1%(98.5 - 99.5%)91.8%(88.3 - 94.6)98.0%0.917
0.500 ng/mL99.2%(98.6 - 99.6%)91.9%(89.3 - 94.1%)97.4%0.926
2.00 ng/mL99.4%(98.8 - 99.8%)96.0%(94.6 - 97.1%)97.8%0.955

N = 2168 (72 ≤ 0.100 ng/mL; 319 ≤ 0.250 ng/mL; 544 ≤ 0.500 ng/mL; 1001 ≤ 2.00 ng/mL)

Total clinical agreement between the VITROS B.R.A.H.M.S.PCT and the B.R.A.H.M.S.PCT sensitive KRYPTOR tests was greater than 97% at all relevant clinical decision points. The Cohen's Kappa was equal to 0.772 at the clinical decision point of 0.100 ng/mL, indicating substantial agreement between the VITROS B·R·A·H·M·S PCT and the B·R·A·H·M·S PCT sensitive KRYPTOR tests. The Cohen's Kappa was greater than 0.910 at the clinical decision points of 0.250 ng/mL, 0.500 ng/mL and 2.00 ng/mL, indicating almost perfect agreement between the VITROS B·R·A·H·M·S PCT and the B·R·A·H·M·S PCT sensitive KRYPTOR tests.

Association of 2 with 28-day All-Cause Mortality b.

The VITROS B·R·A·H·M·S PCT test was evaluated for the prediction of cumulative 28-day all-cause mortality using retrospective samples from a study of 858 adult patients diagnosed with severe sepsis or septic shock recruited across 13 investigational sites in the United States. The analysis population (598 subjects) included 44% female and 56% male patients with a mean age of 64 years. About half of the patients had severe sepsis (51%) versus septic shock (49%). Infections were mainly community acquired (91%).

The binary test result (APCT decline > 80% or ≤ 80%) was significantly associated with 28-day cumulative mortality (vital status on day 28). The two-sided Fisher's exact test p-value was 0.006. Adjusted for ICU versus non-ICU patient subgroups (based on hospital location at Day 4 after initial diagnosis), the association remained significant (Cochran-Mantel-Haenszel test p-value = 0.026). In each binary △PCT subgroup, the 28-day cumulative mortality rate was stratified by need to continue ICU care on Day 4 and the selection of Day 0 versus Day 1 as the baseline measurement day for the △PCT calculation:

{23}------------------------------------------------

28-Day Mortality Risk Stratified by Patient Location on Day 4:ΔPCT Decline > 80% = Test Negative; ΔPCT Decline ≤ 80% = Test Positive
ΔPCTIntervalDay 4PatientLocation28-Day Mortality (%)Prognostic Accuracy* (%)
ΔPCT Decline> 80%(95% CI)ΔPCT Decline≤ 80%(95% CI)Sensitivity(95% CI)Specificity(95% CI)
Day 0to Day 4ICU21.1(11.6-30.6)29.6(23.0-36.3)77.5(67.4-87.6)31.4(24.5-38.3)
Non-ICU5.4(1.5-9.3)11.0(6.6-15.4)74.6(58.2-91.1)42.3(36.2-48.4)
Day 1to Day 4ICU21.0(11.7-30.3)29.8(23.1-36.4)77.2(67.0-87.3)32.1(25.2-39.0)
Non-ICU6.1(1.7-10.5)10.2(6.1-14.3)74.8(58.5-91.2)37.2(31.3-43.1)
Prediction Performances of Binarv APCT Stratified by ICU Care on Day 4
  • Prognostic accuracy refers to how accurate the ΔΡCT (decline ≤ 80% vs. > 80%) can predict mortality risk.

Additional stratification of patients based on absolute initial PCT levels (> 2.00 ng/mL or ≤ 2.00 ng/mL) at Day 0 (or Day 1) revealed subgroups with particularly reduced or elevated mortality risk considering their patient location on Day 4. Mortality risk and prognostic performance are given for the following subgroups in the tables below:

    1. Patients with PCT > 2.00 ng/mL at Day 0 (or Day 1) receiving ICU care on Day 4
    1. Patients with PCT ≤ 2.00 ng/mL at Day 0 (or Day 1) receiving ICU care on Day 4
    1. Patients with PCT > 2.00 ng/mL at Day 0 (or Day 1) without ICU care on Day 4
    1. Patients with PCT ≤ 2.00 ng/mL at Day 0 (or Day 1) without ICU care on Day 4

Prediction Performances of Binary ΔPCT Stratified by ICU Care on Day 4 and Initial PCT on Day 0

28-Day Mortality Risk Stratified by Patient Location on Day 4, Absolute Initial PCT Value on Day 0:ΔPCT Decline > 80% = Test Negative; ΔPCT Decline ≤ 80% = Test Positive
ΔPCTIntervalDay 4PatientLocationInitial PCTValue atDay 0(ng/mL)ΔPCT Decline> 80%(95% CI)ΔPCT Decline≤ 80%(95% CI)Sensitivity(95% CI)Specificity(95% CI)
Day 0toDay 4ICU≤ 2.005.5(0.0-27.0)23.7(13.9-33.4)97.7(88.5-100.0)10.3(2.0-18.7)
> 2.0022.7(12.6-32.7)33.7(24.9-42.6)70.5(57.9-83.1)42.1(33.2-51.0)
Non-ICU≤ 2.005.0(0.0-14.8)8.3(3.4-13.3)90.9(73.9-100.0)14.7(7.8-21.6)
> 2.005.5(1.2-9.8)15.1(6.9-23.3)64.6(41.4-87.9)62.6(54.7-70.5)
  • Prognostic accuracy refers to how accurate the △PCT (decline ≤ 80%) can predict mortality risk.

{24}------------------------------------------------

28-Day Mortality Risk Stratified by Patient Location on Day 4, Absolute Initial PCT Value on Day 1:ΔPCT Decline > 80% = Test Negative; ΔPCT Decline ≤ 80% = Test Positive
ΔPCTIntervalDay 4PatientLocationInitial PCTValue atDay 1(ng/mL)28-Day Mortality (%)Prognostic Accuracy* (%)
ΔPCT Decline> 80%(95% CI)ΔPCT Decline≤ 80%(95% CI)Sensitivity(95% CI)Specificity(95% CI)
Day 1toDay 4ICU≤ 2.0022.9(0.0-61.0)21.3(11.4-31.1)92.0(77.6-100.0)7.4(0.0-15.1)
> 2.0020.9(11.3-30.4)34.6(26.0-43.3)73.0(60.9- 85.1)42.7(33.9-51.4)
Non-ICU≤ 2.000.0(0.0-20.6**)7.2(2.7-11.8)100.0(66.4**-100.0)11.9(5.8-17.9)
> 2.007.0(2.0-12.0)14.8(7.0-22.6)63.0(40.9-85.1)57.5(49.2-65.8)

Prediction Performances of Binary ΔPCT Stratified by ICU Care on Day 4 and Initial PCT on Day 1

  • Prognostic accuracy refers to how accurate the △PCT decline (≤ 80% ) can predict mortality risk. ** Normality approximation of within-imputation variance not valid, therefore the estimate corresponds to withinimputation variation based on exact confidence intervals

The relative mortality ratios for ΔΡCT positive < 80%) versus ΔΡCT negative (decline > 80%) patient subgroups were:

1.48 for patients with PCT > 2.00 ng/mL at Day 0 receiving ICU care on Day 4 4.31 for patients with PCT ≤ 2.00 ng/mL at Day 0 receiving ICU care on Day 4 2.74 for patients with PCT > 2.00 ng/mL at Day 0 without ICU care on Day 4 1.66 for patients with PCT ≤ 2.00 ng/mL at Day 0 without ICU care on Day 4

Based on relative mortality ratios, a decrease in PCT concentration by ≤ 80% from Day 0 (or Day 1) to Day 4 constitutes a higher risk for mortality within 28 days compared to > 80% decreases in each subgroup.

Time-to-event analyses, illustrated by the Kaplan-Meier curves below, demonstrate that patients had a lower survival probability (higher cumulative mortality risk) from study Day 4 until the end of follow-up time (Day 28) when the APCT test result was positive compared to when the APCT result was negative in all patient subgroups according to patient location on Day 4 and initial PCT value.

{25}------------------------------------------------

Kaplan-Meier Curves Stratified by Binary ΔPCT Results of the Per-Protocol Population separated by Patient Location on Day 4 (ICU versus Non-ICU) and PCT Level on Day 0 (PCT on Day 0 > 2.00 ng/mL versus ≤ 2.00 ng/mL)

Image /page/25/Figure/3 description: This image contains four Kaplan-Meier survival curves, each displaying the proportion of patients surviving over a 28-day period. The curves are stratified by initial Procalcitonin (PCT) levels (≤2.00 ng/mL vs. >2.00 ng/mL) and ICU status on Day 4. Within each plot, survival is further differentiated by whether the change in PCT (ΔPCT) was negative or positive, along with the number of patients (n) and deaths in each group. The plots show survival until Day 28, with the x-axis representing time in days and the y-axis representing the proportion of patients surviving.

For the prediction of absolute mortality risks, patient location on Day 4 and initial PCT value should be considered:

  • An initial PCT value ≤ 2.00 ng/mL on Day 0 followed by a PCT decline of . more than 80% by Day 4 indicates a 4-fold lower cumulative 28-day mortality risk (5.5%) for patients with severe sepsis or septic shock who are still in the ICU by Day 4 compared to those patients with an initial PCT value > 2.00 ng/mL (22.7%). Regardless of the initial PCT value, patients in the ICU on Day 4 that do not have more than an 80% decline in PCT plasma value from Day 0 to Day 4 have an even higher mortality risk of 23.7% - 33.7%.
    Ortho-Clinical Diagnostics, Inc.

{26}------------------------------------------------

  • . An initial PCT value > 2.00 ng/mL that does not decline by more than 80% by Day 4 signals that such patients remain at high mortality risk (15.1%) even when they are no longer receiving ICU care on Day 4. Mortality was otherwise observed between 5.0% to 8.3% for patients discharged from the ICU by Day 4.
    A ΔPCT from Day 0 to Day 4 (decline ≤ 80% versus decline > 80%) as a prognostic for 28-day cumulative risk of mortality was quantified by Cox proportional hazards regression analysis with a hazard ratio of 1.93 (95% CI of 1.19-3.12; p-value = 0.008). The relative risk of cumulative 28-day mortality is about 2-fold higher if an individual tests positive for ΔΡCT (decline ≤ 80%) than if an individual tests negative (decline > 80%).

As a comparison, the table below lists the univariate hazard ratios for other clinical factors evaluated as separate predictors of mortality in the study population.

PredictorsComparisonHazardRatio95% CIp-Value
ΔPCT (Day 0 to Day 4)Decline ≤ 80% vs. > 80%1.931.19-3.120.008
ΔPCT (Day 1 to Day 4)Decline ≤ 80% vs. > 80%1.731.07-2.790.025
APACHE on Day 1Difference of 5 units1.361.22-1.53<0.001
Maximum SOFA of Day 0-Day 4Difference of 3 units1.731.50-2.00<0.001
Antibiotic AdequacyNo vs. Yes1.591.00-2.530.051
Sepsis SeveritySeptic Shock vs. Severe Sepsis1.190.80-1.760.386
Biological Infection TypeGram Positive vs. Gram Negative0.830.48-1.450.522
Biological Infection TypeOther vs. Gram Negative0.990.63-1.540.960
Biological Infection TypeFungal vs. Gram Negative2.440.87-6.840.090
Clinical Infection TypeNosocomial vs. Community Acquired0.760.35-1.640.481
Positive Blood CultureYes vs. No1.050.69-1.580.834
PCT on Day 0> 2.00 ng/mL vs. ≤ 2.00 ng/mL1.390.90-2.150.139
AgeDifference of 5 years1.161.08-1.24<0.001
GenderMale vs. Female0.950.64-1.400.782
ICU Care on Day 4Yes vs. No3.452.24-5.31<0.001

Univariate Hazard Ratios for 28-Day All-Cause Mortality of APCT and Clinical Covariates

A ΔPCT from Day 0 (or Day 1) to Day 4 remains a prognostic parameter for the risk of cumulative 28-day mortality in patients diagnosed with severe sepsis or septic shock even when the hazard ratio is adjusted for other mortality predictors in Cox multiple regression models. The relative mortality risk estimates for APCT and selected predictors are presented below with 95% confidence intervals. For continuous predictors, the hazard ratio (HR) was calculated for one standard deviation (SD) change in the predictor. For binary predictors, the risk estimate compares the hazards for the two binary results.

{27}------------------------------------------------

Hazard Ratios for ΔPCT and Selected Predictors from Multivariate Cox Regression Models
ModelHazard Ratio (HR)(95% Confidence Interval)
Binary PredictorsContinuous Predictors (HR per 1 SD)
ΔPCTIntervalScore +Covariates*ΔPCT Decline(≤ 80% vs. > 80%)Day 4 PatientLocation (ICUvs. Non-ICU)APACHE (1SD = 8.13)Maximum SOFA(1 SD = 3.98)Age (1 SD= 16.18)
Day 0to Day 4APACHE1.75(1.00-3.04)2.63(1.64-4.21)1.24(0.99-1.56)N/A1.58(1.26-1.98)
MaximumSOFA1.59(0.92-2.73)1.68(1.02-2.78)N/A1.96(1.53-2.52)1.68(1.35-2.10)
Day 1to Day 4APACHE1.67(0.99-2.82)2.65(1.65-4.24)1.29(1.03-1.61)N/A1.57(1.25-1.96)
MaximumSOFA1.48(0.88-2.51)1.73(1.05-2.84)N/A1.98(1.54-2.54)1.67(1.34-2.09)

Hazard Ratios for APCT and Selected Predictors from Multivariate Cox Regression Models

*The models also included the following predictors (hazard ratio results not shown): antibiotic adequacy, sepsis severity, biological infection type, clinical infection type, positive blood culture, PCT value on Day 0, gender. In the analysis, missing values for predictors were multiple imputed assuming they were Missing at Random (MAR), with the multiple imputations combined according to Rubin's rules.

The change of PCT over time can also be described by the ratio of PCT values from Day 4 and Day 0 (or Day 1):

PCTratio = __ PCT Day4 PCT Day0 (or Day1)

A decline of △PCT = 80% translates into a PCT ratio of 0.2. The PCT ratio has values larger than 0.2 when the △PCT decline is less than 80%, which is associated with a higher risk for cumulative 28-day all-cause mortality in patients diagnosed with severe sepsis or septic shock. Likewise, a PCT ratio below 0.2 indicates a lower risk for mortality within 28 days. On a continuous scale, the relative mortality risk for such patients is higher the larger the PCT ratio. The following table lists the hazard ratios for an increase by the factor 2 in PCT ratio (i.e., the relative increase in mortality risk for a patient with any given PCT ratio compared to a patient with a 2-fold lower PCT ratio). For the patient location at Day 4, the risk estimate compares the hazards for patients with versus without ICU care on Day 4.

Hazard Ratios for APCT and Selected Predictors from Multivariate Cox Regression Models
------------------------------------------------------------------------------------------
ModelHazard Ratio(95% Confidence Interval)Binary Predictor
Continuous Predictors(HR per 2-fold increase in PCT ratio or per equivalent in SD)
ΔPCTIntervalScore +Covariates*PCT ratio(2-fold increase)APACHE(SD equivalent)**Maximum SOFA(SD equivalent)**Age(SD equivalent) **Day 4PatientLocation(ICU vs. Non-ICU)
Day 0to Day 4APACHE1.29(1.13-1.47)1.08(0.95-1.23)N/A1.32(1.16-1.49)2.52(1.56-4.06)
MaximumSOFA1.21(1.06-1.38)N/A1.40(1.21-1.61)1.35(1.19-1.53)1.68(1.02-2.76)

Ortho-Clinical Diagnostics, Inc.

{28}------------------------------------------------

ModelHazard Ratio (95% Confidence Interval)
Continuous Predictors (HR per 2-fold increase in PCT ratio or per equivalent in SD)Binary Predictor
Day 1 to Day 4APACHE1.40(1.18-1.66)1.20(1.01-1.43)N/A1.44(1.21-1.71)2.60(1.62-4.16)
Maximum SOFA1.33(1.11-1.59)N/A1.65(1.36-2.00)1.51(1.27-1.79)1.75(1.07-2.88)

*The models also included the following predictors (hazard ratio results not shown): antibiotic adequacy, sepsis severity, biological infection type, clinical infection type, positive blood culture, PCT on Day 0, and gender. In the analysis, missing values for predictors were multiple imputed assuming they were Missing at Random (MAR), with the multiple imputations combined according to Rubin's rules.

** A unit change of APCT on log-2-scale corresponded to 0.56 SD of APCT from Day 0 until Day 4 (0.78 SD for APCT from Day 1 until Day 4). Accordingly, the reported APCT hazard ratios refer to an increase of APCT by a factor of 2. For comparability, hazard ratios of the other continuous predictors were estimated for the same fractional SD (i.e., 0.56 or 0.78, respectively).

  • c. Clinical Specificity:
    See clinical sensitivity above.

N. Proposed Labeling:

The labeling is sufficient and satisfies the requirements of 21 CFR Parts 801 and 809, as applicable and the special controls for this device type under 21 CFR 866.3215.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3215 Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis.

(a)
Identification. A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
e.g., age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.30 design controls, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
e.g., collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.