(76 days)
No
The summary describes a standard in vitro diagnostic immunoassay for measuring CEA levels, focusing on reagents, performance characteristics, and comparison to a predicate device. There is no mention of AI or ML in the device description, intended use, or performance studies.
No.
This device is for in vitro diagnostic use to measure CEA levels, which aids in management but does not directly treat or cure disease.
Yes.
The "Intended Use / Indications for Use" section explicitly states "For in vitro diagnostic use in the quantitative measurement of carcinoembryonic antigen (CEA)... to aid in the management of cancer patients".
No
The device description clearly outlines physical reagents and reagent packs, indicating it is a hardware-based in vitro diagnostic assay, not a software-only device.
Yes, this device is an IVD (In Vitro Diagnostic).
The "Intended Use / Indications for Use" section explicitly states: "For in vitro diagnostic use in the quantitative measurement of carcinoembryonic antigen (CEA) in serum and plasma..." This is the defining characteristic of an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
For in vitro diagnostic use in the quantitative measurement of carcinoembryomic antigen (CEA) in serum and plasma (EDTA and lithium heparin) to aid in the management of cancer patients in whom changing concentrations of CEA are observed using the ADVIA Centaur® XP and ADVIA Centaur® XPT systems.
Product codes
DHX
Device Description
The assay reagents come in the following configurations:
5 ReadyPack primary reagent packs containing ADVIA Centaur CEA Lite Reagent and Solid Phase ADVIA Centaur CEA Master Curve card (500 tests)
1 ReadyPack primary reagent pack containing ADVIA Centaur CEA Lite Reagent and Solid Phase ADVIA Centaur CEA Master Curve card (100 tests)
The ReadyPack consists of the following:
ADVIA Centaur® CEA ReadyPack® primary reagent pack; Lite Reagent: 5.0 mL/reagent pack polyclonal rabbit anti-CEA antibody (~400 ng/mL) labeled with acridinium ester in phosphate buffered saline with protein stabilizers, sodium azide (0.12%), and preservatives
ADVIA Centaur® CEA ReadyPack® primary reagent pack; Solid Phase Reagent: 25.0 mL/reagent pack monoclonal mouse anti-CEA antibody (~120 µg/mL) covalently coupled to paramagnetic particles in phosphate buffered saline with protein stabilizers, sodium azide (0.11%), and preservatives
ADVIA Centaur® CEA ReadyPack® ancillary reagent pack; CEA Diluent: 5.0 mL/reagent pack bicine buffer, gelatin, and BSA with preservatives and sodium azide (0.1%)
ADVIA Centaur® CEA Diluent: 10.0 mL/reagent vial bicine buffer, gelatin, and BSA with preservatives and sodium azide (0.1%)
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Performance Characteristics Data
Detection Capability determined in accordance with CLSI Document EP17-A2.
Limit of Blank (LoB): 0.50 ng/mL (µg/L)
Limit of Detection (LoD): 1.00 ng/mL (µg/L)
Limit of Quantitation (LoQ): 2.00 ng/mL (µg/L)
Precision
Cleared under premarket submission K981478
Method Comparison
Method comparison data and regression analysis was revised in K200215.
Accuracy / Method Comparison: For 201 samples in the range of 2.00 to 78.93 ng/mL (µg/L), the relationship between the ADVIA Centaur CEA assay and the ACS:180 CEA assay is described by the equation: ADVIA Centaur CEA = 0.97 (ACS:180 CEA) + 0.11 ng/mL
Specimen Equivalence
Determined using the Deming linear regression model in accordance with CLSI Document EP09-A3.
Dipotassium EDTA plasma (y) vs. Serum (x): Na=64, Sample Interval=2.08-97.10 ng/mL (µg/L), Slope=0.95, Intercept=0.20 ng/mL (µg/L), rb=1.00
Lithium heparin plasma (y) vs. Serum (x): Na=46, Sample Interval=2.11-97.10 ng/mL (µg/L), Slope=0.99, Intercept=0.19 ng/mL (µg/L), rb=1.00
The assay is designed to have a slope of 0.90–1.10 for alternate tube types versus serum.
Interferences
Interference testing was performed in accordance with CLSI Document EP07-ed3.
Dipotassium EDTA (9.0 mg/mL): Analyte Concentration=5.78 ng/mL (µg/L) with Bias=-0.3%; Analyte Concentration=55.62 ng/mL (µg/L) with Bias=4.6%
Heparin (75 U/mL): Analyte Concentration=5.83 ng/mL (µg/L) with Bias=-0.6%; Analyte Concentration=61.23 ng/mL (µg/L) with Bias=0.9%
Expected Values
Cleared under premarket submission K981478
Key Metrics
Correlation coefficient (r) = 1.00
Slope for Dipotassium EDTA plasma vs. Serum: 0.95
Intercept for Dipotassium EDTA plasma vs. Serum: 0.20 ng/mL (µg/L)
r for Dipotassium EDTA plasma vs. Serum: 1.00
Slope for Lithium heparin plasma vs. Serum: 0.99
Intercept for Lithium heparin plasma vs. Serum: 0.19 ng/mL (µg/L)
r for Lithium heparin plasma vs. Serum: 1.00
Bias for Dipotassium EDTA: -0.3% and 4.6%
Bias for Heparin: -0.6% and 0.9%
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 866.6010 Tumor-associated antigen immunological test system.
(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
April 13, 2020
Siemens Healthcare Diagnostics Inc. Mey Lyn Vasquez Regulatory Clinical Affairs Specialist 511 Benedict Ave. Tarrytown, NY 10591
Re: K200215
Trade/Device Name: ADVIA Centaur CEA Assay Regulation Number: 21 CFR 866.6010 Regulation Name: Tumor-associated antigen immunological test system Regulatory Class: Class II Product Code: DHX Dated: January 27, 2020 Received: January 28, 2020
Dear Mey Lyn Vasquez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
1
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carolina Kagan Acting Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K200215
Device Name ADVIA Centaur® CEA assay
Indications for Use (Describe)
For in vitro diagnostic use in the quantitative measurement of carcinoembryomic antigen (CEA) in serum and plasma (EDTA and lithium heparin) to aid in the management of cancer patients in whom changing concentrations of CEA are observed using the ADVIA Centaur® XP and ADVIA Centaur® XPT systems.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
---|---|
---------------------------------------------- | --------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is an orange graphic of several circles of varying sizes.
510(k) Summary of Safety and Effectiveness
Introduction: According to the requirements of SMDA 1990 and 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: k200215_______________
APPLICANT 1.
Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue, Tarrytown, NY 10591 USA
Contact: | Mey Vasquez |
---|---|
Regulatory Clinical Affairs Specialist | |
Phone: | (914) 524-2458 |
Fax: | (914) 524-3579 |
E-mail: | mey.vasquez@siemens-healthineers.com |
Date Prepared: April 09, 2020
Regulatory Information 2.
Assay
Trade Name | ADVIA Centaur® CEA assay |
---|---|
Device | system, test, carcinoembryonic antigen |
Regulation Description | Tumor-associated antigen immunological test system |
FDA Classification | Class II |
Review Panel | Immunology |
Product Code | DHX |
Regulation Number | 21 CFR 866.6010 |
4
Image /page/4/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
3. PREDICATE DEVICE
Assay
Name of Device: ADVIA Centaur® CEA assay
510 (k): K981478
4. DEVICE DESCRIPTION
The assay reagents come in the following configurations:
| Contents | Number
of Tests |
|---------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| 5 ReadyPack primary reagent packs containing ADVIA Centaur
CEA Lite Reagent and Solid Phase
ADVIA Centaur CEA Master Curve card | 500 |
| 1 ReadyPack primary reagent pack containing ADVIA Centaur
CEA Lite Reagent and Solid Phase
ADVIA Centaur CEA Master Curve card | 100 |
The ReadyPack consists of the following:
ADVIA Centaur® CEA ReadyPack® primary reagent pack; Lite Reagent
5.0 mL/reagent pack polyclonal rabbit anti-CEA antibody (~400 ng/mL) labeled with acridinium ester in phosphate buffered saline with protein stabilizers, sodium azide (0.12%), and preservatives
ADVIA Centaur® CEA ReadyPack® primary reagent pack; Solid Phase Reagent
25.0 mL/reagent pack monoclonal mouse anti-CEA antibody (~120 µg/mL) covalently coupled to paramagnetic particles in phosphate buffered saline with protein stabilizers, sodium azide (0.11%), and preservatives
ADVIA Centaur® CEA ReadyPack® ancillary reagent pack; CEA Diluent
5.0 mL/reagent pack bicine buffer, gelatin, and BSA with preservatives and sodium azide (0.1%)
ADVIA Centaur® CEA Diluent
10.0 mL/reagent vial bicine buffer, gelatin, and BSA with preservatives and sodium azide (0.1%)
5
Image /page/5/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are several orange dots arranged in a circular pattern.
5. INDICATIONS FOR USE
For in vitro diagnostic use in the quantitative measurement of carcinoembryonic antigen (CEA) in serum and plasma (EDTA and lithium heparin) to aid in the management of cancer patients in whom changing concentrations of CEA are observed using the ADVIA Centaur® XP and ADVIA Centaur® XPT systems.
6. INTENDED USE
Same as Indications for Use
Purpose of the Submission 7.
The purpose of this submission is for the addition of plasma (EDTA and lithium heparin) sample claim for the ADVIA Centaur® CEA assay.
COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH THE 8. PREDICATE DEVICE
The following table demonstrates substantial equivalence between the ADVIA Centaur® CEA assay (Candidate Device) that has modified Instructions for Use (Package Inserts) with the addition of the plasma (EDTA and lithium) sample claim and the currently marketed ADVIA Centaur® CEA assay (Predicate Device) that was cleared under 510(k) K981478.
6
Image /page/6/Picture/0 description: The image contains the logos for Siemens and Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
Assay | ||||
---|---|---|---|---|
ltem | Predicate Device | Candidate Device | ||
ADVIA Centaur® CEA assay | ADVIA Centaur® CEA assay | |||
Intended Use | For in vitro diagnostic use in the quantitative | |||
measurement of carcinoembryonic antigen (CEA) | ||||
in serum to aid in the management of cancer | ||||
patients in whom changing concentrations of CEA | ||||
are observed using the ADVIA Centaur®, ADVIA | ||||
Centaur XP, and ADVIA Centaur XPT systems. | For in vitro diagnostic use in the quantitative measurement | |||
of carcinoembryonic antigen (CEA) in serum and plasma | ||||
(EDTA and lithium heparin) to aid in the management of | ||||
cancer patients in whom changing concentrations of CEA | ||||
are observed using the ADVIA Centaur® XP and ADVIA | ||||
Centaur® XPT systems. | ||||
Measurement | Quantitative | Same | ||
Assay Range | 0.5-100 ng/mL (µg/L) | 2.00-100 ng/mL (µg/L) | ||
Assay Principle | Sandwich immunoassay | Same | ||
Technology | Direct chemiluminescent | Same | ||
Sample Type | Serum | Serum or plasma (EDTA or | ||
lithium heparin) | ||||
Sample Volume | 50 µL | Same | ||
Reagent | ||||
Volume | 50 µL of Lite Reagent and 250 µL of Solid Phase | Same | ||
Incubation | ||||
Time | 7.5 minutes at 37°C. | Same | ||
Standardization | The ADVIA Centaur CEA assay is traceable to an | |||
internal standard manufactured using highly | ||||
purified material. Assigned values for calibrators | ||||
are traceable to this standardization. | Same | |||
Calibration | 2-point | Same | ||
Calibrators | ADVIA Centaur Calibrator D | Same |
Table 1: Substantial Equivalence Comparison
Siemens Healthcare Diagnostics Inc.
7
Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
Assay | ||
---|---|---|
Item | Predicate Device | Candidate Device |
ADVIA Centaur® CEA assay | ADVIA Centaur® CEA assay | |
Number of | ||
Calibrator | ||
Levels | Two levels | Same |
Controls | Commercial Controls | Same |
Number of | ||
Control Levels | 2 | Same |
8
Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
PERFORMANCE CHARACTERISTICS DATA 9.
Detection Capability
Detection capability was determined in accordance with CLSI Document EP17-A2. 11
Limit of Blank (LoB) | 0.50 ng/mL (µg/L) |
---|---|
Limit of Detection (LoD) | 1.00 ng/mL (µg/L) |
Limit of Quantitation (LoQ) | 2.00 ng/mL (µg/L) |
The LoB corresponds to the highest measurement likely to be observed for a blank sample with a probability of 95%.
The LoD corresponds to the lowest concentration of carcinoembryonic antigen that can be detected with a probability of 95%.
The LoQ corresponds to the lowest amount of carcinoembryonic antigen in a sample at which the within laboratory CV is ≤ 20%.
Precision
Cleared under premarket submission K981478
Method Comparison
Method comparison data and regression analysis was revised in K200215.
Accuracy / Method Comparison
For 201 samples in the range of 2.00 to 78.93 ng/mL (µg/L), the relationship between the ADVIA Centaur CEA assay and the ACS:180 CEA assay is described by the equation:
ADVIA Centaur CEA = 0.97 (ACS:180 CEA) + 0.11 ng/mL Correlation coefficient (r) = 1.00
Specimen Equivalence
Specimen equivalency was determined using the Deming linear regression model in accordance with CLSI Document EP09-A3. The following results were obtained:
Tube (y) vs. Serum (x) | Na | Sample Interval | Slope | Intercept | rb |
---|---|---|---|---|---|
Dipotassium EDTA plasma | 64 | 2.08-97.10 ng/mL (µg/L) | 0.95 | 0.20 ng/mL (µg/L) | 1.00 |
Lithium heparin plasma | 46 | 2.11-97.10 ng/mL (µg/L) | 0.99 | 0.19 ng/mL (µg/L) | 1.00 |
9
Image /page/9/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a cluster of orange dots arranged in a circular pattern.
The assay is designed to have a slope of 0.90–1.10 for alternate tube types versus serum.
Agreement of the specimen types may vary depending on the study design and sample population used. Assay results obtained at individual laboratories may vary from the data presented.
Interferences
Interference testing was performed in accordance with CLSI Document EP07-ed3. The following results were obtained:
| Substance | Substance Test Concentration | Analyte Concentration
ng/mL (µg/L) | Bias (%) |
|------------------|------------------------------|---------------------------------------|----------|
| Dipotassium EDTA | 9.0 mg/mL | 5.78 | -0.3 |
| | | 55.62 | 4.6 |
| Heparin | 75 U/mL | 5.83 | -0.6 |
| | | 61.23 | 0.9 |
Assay results obtained at individual laboratories may vary from the data presented.
Expected Values
Cleared under premarket submission K981478
X. CONCLUSION
Comparative testing of the ADVIA Centaur® CEA assay is substantially equivalent in principle and performance to the Predicate Device – ADVIA Centaur® CEA assay cleared under 510(k) K981478.