K Number
K200190

Validate with FDA (Live)

Device Name
ScalpelCric
Date Cleared
2020-10-30

(277 days)

Product Code
Regulation Number
868.5090
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Emergency airway access via the cricothyroid membrane. Life-threatening dyspnoea that cannot be controlled in any other way.

Device Description

Surgical Cricothyrotomy set for securing the airway in case of upper airway obstruction, respectively as "Ultima Ratio" if all other attempts to ventilate the patient failed. The set consists of scalpel, bougie, tube, extension tube with swivel connector, syringe and necktape.

AI/ML Overview

The provided text is a 510(k) Premarket Notification for the ScalpelCric device, a cricothyrotomy set. It details the device's characteristics, intended use, and comparison to a predicate device to establish substantial equivalence.

However, this document does not contain information about a study that assesses the device's performance against specific acceptance criteria in the context of an AI/human-in-the-loop system, or a multi-reader multi-case (MRMC) study. The document focuses on the physical and biological characteristics of the medical device itself, its sterilization, biocompatibility, and bench testing, rather than a performance study involving diagnostic accuracy, expert review, or AI assistance.

Therefore, I cannot fulfill the request to describe the acceptance criteria and the study proving the device meets them in the context of an AI-assisted diagnostic system, as that information is not present in the provided text.

The closest information related to "acceptance criteria" is under section 9.4 Bench Testing, which lists features to be tested and the result (all "Passed"). This refers to the physical and functional performance of the device components, not a diagnostic or AI-related performance.

Here's what I can extract from the provided text regarding device testing, but it's important to note this is not a study proving AI performance as requested:

1. A table of acceptance criteria and the reported device performance:

The document provides a table under "9.4 Bench Testing" (page 10), which lists various tests performed on the components of the ScalpelCric set. The "Features to be tested" can be considered the acceptance criteria for those specific physical/functional aspects, and the "Result" indicates "Passed" for all.

Object under TestFeatures to be tested (Acceptance Criteria)Reported Device Performance
Bonding ConnectorBonding between connector and tubePassed
Bonding Inflation LineBonding between inflation line and tubePassed
15mm ConnectorDimensions of 15mm ConnectorPassed
Luer Pilot BalloonDimensions of the female luer connector in the pilot balloonPassed
CuffFunction of the CuffPassed
ScalpelFunction and design of the scalpelPassed
BougieBougie which shall fulfill all relevant specification according [DI] and [MS_Bougie].Passed
TubeFunction and dimensions of the tubePassed
TubeTube is recognizable in X-Ray imagesPassed
Extension TubeFunction and dimension of the extension tubePassed
Neck TapeNeck tape which shall fulfill all relevant specification according [DI] and [MS_NT] after 4 times sterilizationPassed
SyringeFunction of the syringePassed
Tube with inflation line and cuff,Function and features of the tube with attached inflation line and pilot balloon and with cuffPassed
ScalpelCricBiological compatibilityPassed
MaterialFunction and ingredient of the ScalpelCric materialsPassed
Tube and extension TubeOxygen resistance of Tube and Extension tubePassed
LabelDevice LabelPassed
BougieDevice Label BougiePassed
IFUIFU which shall contain all relevant information regarding [DI] and [ER].Passed
ScalpelCricScalpelCric Set can be used in Operating conditionPassed
ScalpelCricFunctionality after simulated transportPassed
Transport ValidationFunctionality after simulated transportPassed
SafetyVerify implementation and effectiveness of all Risk Control Measures defined in the Risk AnalysisPassed
Suction CapConnecting tube with open suction cap during ventilationPassed
User NeedsUser Needs for ScalpelCric, defined in [SoP]Passed
ScalpelCricUsabilityPassed
HeaderbagRequirements of DIN EN ISO 11607-1 and DIN EN ISO 11607-2.Passed
SterilizationSterilization Requirements for ScalpelCricPassed
Business NeedsBusiness Needs for ScalpelCricPassed
Label Sterile PackagingSterile packaging label which shall contain all relevant information regarding the [SFL], [DI] and [ER].Passed
Label Outer PackagingOuter packaging label which shall contain all relevant information regarding the [SFL], [DL] and [ER]Passed
PackagingPackaging Requirements for ScalpelCricPassed
Accelerated Aging SyringeQuality of the syringe after 5 years accelerated agingPassed
Accelerated aging of PouchFunction of the sterile packaging after 5 years accelerated agingPassed
BougieBougie shall fulfill all relevant specifications after 1 year accelerated agingPassed
ScalpelFunction of the scalpel after accelerated aging 1 yearPassed
ScalpelCricScalpel Cric Set application oxygen resistance after 1 year accelerated agingPassed
SurgicricFunction and dimension after 5 years accelerated aging of the Surgicric and the function of the sterile packaging.Passed
Surgicricare identical to the devices in ScalpelCric.Passed
SurgicricFunction and dimension after 3 years real time aging of the Surgicric including Tubus and the function of the sterile packaging. The tested tube and sterile packaging are identical to the devices in ScalpelCric.Passed
SyringeQuality of the syringe after 3 years real time agingPassed
BougieBougie shall fulfill all relevant specifications after 5 years accelerated agingPassed
ScalpelFunction of the scalpel after 5 years accelerated agingPassed
ScalpelCric SetApplication oxygen resistance after 5 years accelerated agingPassed
Cuff Tube Collapse included in the ScalpelCric after 5 years of accelerated aging.Test required by DIN EN ISO 5361:2013-03, Annex C: Resistance of the VBM tracheal tube ventilation lumen collapse with cuff inflated inside a cylinder.Passed

The remaining requested information (sample size, data provenance, experts, adjudication, MRMC, standalone, ground truth type, training set details) is NOT available in the provided document, as it pertains to a different type of device evaluation (e.g., AI/diagnostic performance) than what is described here. This document is focused on the substantial equivalence of a physical medical device (a cricothyrotomy set) to a predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

VBM Medizintechnik GmbH Jannika Jaeger Regulatory Affairs Manager Einsteinstrasse 1 Sulz am Neckar, 72172 De

Re: K200190 Trade/Device Name: ScalpelCric Regulation Number: 21 CFR 868.5090 Regulation Name: Emergency Airway Needle

Regulatory Class: Class II Product Code: BWC, JOH, OGP Dated: September 25, 2020 Received: September 28, 2020

Dear Jannika Jaeger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

Device Name

ScalpelCric Crycothyrotomy set - Scalpel technique

Indications for Use (Describe)

Emergency airway access via the cricothyroid membrane. Life-threatening dyspnoea that cannot be controlled in any other way.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the letters VBM in a sans-serif font. The letters are a bright blue color. The letters are large and take up most of the image.

K200190

VOLUME 003

510(k) Summary

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for VBM, which is written in a sans-serif font. The letters are large and bold, and they are all in the same blue color. The logo is simple and modern, and it is easy to read.

DATE OF APPLICATION:10/26/2020
APPLICANT:VBM Medizintechnik GmbHEinsteinstrasse 172172 Sulz a. N.GermanyTel: + 49 (0) 7454 9596 0Fax: +49 (0) 7454 9596 33E-Mail: info@vbm-medical.de
CONTACT PERSON:Jannika JaegerRegulatory Affairs ManagerTel.: + 49 (0) 7454 9596 660

E-Mail: jannika.jaeger@vbm-medical.de

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the letters VBM in a sans-serif font. The letters are a bright blue color. The background is white.

1 Device Name

Trade Name:ScalpelCric
Common Name:Cricothyrotomy SetNeedle, Emergency Airway
Device Classification Name:Cricothyrotomy KitTube tracheostomy and tube cuff

2 Classification / Product Code

ScalpelCric can be classified according to following device name and product code:

Primary Product Code

DeviceRegulationDescriptionRegulationMedicalSpecialtyReview PanelProductCodeRegulationNumberDeviceClassification
Needle,EmergencyAirwayEmergencyairwayneedle.AnesthesiologyAnesthesiologyBWC868.50902

Additional Product ode

DeviceRegulationDescriptionRegulationMedicalSpecialtyReview PanelProductCodeRegulationNumberDeviceClassification
Tubetracheostomyand tube cuffTracheostomy tube andtube cuff.AnesthesiologyAnesthesiologyJOH868.58002

Predicate Device / Reference Device ന

DevicePredicate DeviceReferenceDevice510(k)Number510(k) Holder
Needle, EmergencyAirwayMelker Cuffed EmergencyCricothyrotomy CatheterSet -Percutaneous, MelkerCuffed EmergencyCricothyrotomy CatheterSet - Surgical, MelkerUniversal CuffedEmergency CricothyrotomyCatheter Set---K160200COOKINCORPORATED750 DANIELS WAYBloomington,Indiana 47404
---AirwayConnectorwith Flex TubeK942392Covidien LlcformerlyMALLINCKRODTDAR S.R.L.15 HAMPSHIRESTREET

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the letters "VBM" in a bold, sans-serif font. The letters are a bright, light blue color. The letters are large and take up most of the frame.

Mansfield, MA US 02048
--------------------------------

4 Device Description

Surgical Cricothyrotomy set for securing the airway in case of upper airway obstruction, respectively as "Ultima Ratio" if all other attempts to ventilate the patient failed.

The set consists of scalpel, bougie, tube, extension tube with swivel connector, syringe and necktape.

5 Intended Use

Surgical cricothyrotomy set for securing the airway in case of upper airway obstruction or as ultima ratio if all other attempts to ventilate the patient have failed. ScalpelCric can be used with resuscitation bags or other standard ventilation systems.

Clinical benefit: ScalpelCric allows the ventilation of patients as ultima ratio.

Patient target group: Adults

Location of use: Hospital and pre-hospital conditions including military use

6 Indication for Use

Emergency airway access via the cricothyroid membrane. Life-threatening dyspnoea that cannot be controlled in any other way.

7 Mechanism of Action

The device is used for securing the airway in case of upper airway obstruction. After incision of the cricothyroid membrane a bouqie is inserted to allow an easy insertion of the tube over the bougie into the trachea. After bougie removal, inflation of the cuff and fixation of the tube the patient can be ventilated.

8 Technological Characteristics

The technological characteristics of ScalpelCric are substantially equivalent to the technological characteristics of the predicate device or the reference device.

CompanyVBM---ScalpelCric,Cricothyrotomy Set(New Device)CookIncorporated---Melker CuffedEmergencyCricothyrotomyCatheter Set -Surgical(PredicateDevice)Covidien IIc---AirwayConnector withFlex Tube(ReferenceDevice)Result
Device NameNeedle, EmergencyAirwayCricothyrotomy kitTube tracheostomy andtube cuffNeedle, EmergencyAirwayCricothyrotomy kitTube tracheostomyand tube cuffAirway Connectorwith Flex Tube-
CompanyVBMScalpelCric,Cricothyrotomy Set(New Device)CookIncorporatedMelker CuffedEmergencyCricothyrotomyCatheter Set –Surgical(PredicateDevice)Covidien IIcAirwayConnector withFlex Tube(ReferenceDevice)Result
Regulation Number868.5090868.5800868.5800868.5090868.5800868.5800868.5810SubstantiallyEquivalent(A)
Class221SubstantiallyEquivalent(A)
CodeBWCOGPJOHBWCOGPJOHBZASubstantiallyEquivalent(A)
510(k) number---K160200K942392
Tube /AirwayCatheterInnerDiameter6 mm5 mmN.A.SubstantiallyEquivalent(A)
Tube /AirwayCatheterOuterDiameter9 mm22 French (7,2mm)N.A.SubstantiallyEquivalent(A)
Tube /AirwayCatheterLength10cm9 cmN.A.SubstantiallyEquivalent(A)
Bougie/LoadingdilatorOuterDiameter14 French14 FrenchN.A.SubstantiallyEquivalent
Bougie/LoadingdilatorLength40 cm11 cmN.A.SubstantiallyEquivalent(A)
Bougie/LoadingdilatorTip ShapeBluntBluntN.A.SubstantiallyEquivalent
SyringeVolume10 ml6 mlN.A.SubstantiallyEquivalent(A)
ScalpelType#10#11N.A.SubstantiallyEquivalent(A)
NecktapeAvailabilityIncludedno NecktapeincludedN.A.SubstantiallyEquivalent(A)
ExtensionTube withSwivelConnectorInnerDiameter -patientside15 mm15 mmSubstantiallyEquivalent
ExtensionTube withSwivelConnectorInnerDiameter -machineside22 mmno Extension Tubeincluded22 mmSubstantiallyEquivalent
ExtensionTube withSwivelConnectorLengthextensiontube13 cm13 cmSubstantiallyEquivalent

8.1 Device Characteristics Table

{8}------------------------------------------------

VBM Medizintechnik GmbH

ScalpelCric, Cricothyrotomy Set 510(k) Premarket Notification

Image /page/8/Picture/2 description: The image shows the letters VBM in a bold, sans-serif font. The letters are all capitalized and are a bright blue color. The letters are arranged horizontally, with the V on the left, the B in the middle, and the M on the right. The background is white.

A. Differences in relevant characteristics have been separately evaluated during Substantial equivalence discussion. They do not raise new questions regarding safety and efficacy.

8.2 Summary of Technological Characteristics

The proposed devices are substantially equivalent in terms of design, operating principles and intended use and have similar technological characteristics to the predicate devices.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for VBM, which is written in a sans-serif font. The letters are large and bold, and they are all in the same blue color. The logo is simple and modern, and it is easy to read.

の Performance Data

9.1 Sterilization and Shelf Life

ScalpelCric Set is supplied in sterile condition. EO-sterilizations is used for the sterilization of the ScalpelCric Set. ScalpelCric Set has a shelf life of 5 years.

9.2 Biocompatibility

Devices of the ScalpelCric Set were evaluated for their biological safety according to ISO 10993-1. All relevant endpoints have been considered. Tests regarding cytotoxicity, irritation, sensitization, Pyrogen, material mediated pyrogenicity and acute systemic toxicity were performed during the biological safety evaluation. All conducted tests have been successful. The evaluation shows the biological safety of the ScalpelCric Set.

ComponentType of physical ContactDuration of Contact
CategoryContact
ScalpelExternal CommunicatingdeviceTissue/ Bone/ Dentine< 24 hours
BougieExternal CommunicatingdeviceTissue/ Bone/ Dentine< 24 hours
Cuffed TubeExternal CommunicatingdeviceTissue/ Bone/ Dentine< 24 hours
Extension TubeSurface deviceMucosal membrane< 24 hours
Neck tapeSurface deviceIntact Skin< 24 hours
SyringeWithout Patient ContactWithout PatientContactN. A.

9.2.1 Patient Contacting Components

9.2.2 Biocompatibility tests

Tested componentConducted TestResult
ScalpelCytotoxicityNon-cytotoxic
ScalpelMaterial Mediated PyrogenicityNo traces of material mediatedpyrogenicity
ScalpelAcute Systemic ToxicityNo acute systemic toxiccharacteristics
ScalpelPyrogenEndotoxin free
BougieIrritationNo intracutaneous reactivity
BougieSensitizationNon-sensitizer
BougieCytotoxicityNon-cytotoxic
BougieMaterial Mediated PyrogenicityNo traces of material mediatedpyrogenicity
BougieAcute Systemic ToxicityNo acute systemic toxiccharacteristics
Cuffed TubeCytotoxicityNon-cytotoxic

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the letters VBM in a large, sans-serif font. The letters are a bright, light blue color. The letters are arranged horizontally, with the V on the left, the B in the middle, and the M on the right. The background is white.

Material Mediated PyrogenicityNo traces of material mediated pyrogenicity
Acute Systemic ToxicityNo acute systemic toxic characteristics
SensitizationNon-sensitizer
IrritationNo intracutaneous reactivity
PyrogenEndotoxin free
Extension TubeCytotoxicityNon-cytotoxic
IrritationNo intracutaneous reactivity
SensitizationNon-sensitizer
PyrogenEndotoxin free
Material Mediated PyrogenicityNo traces of material mediated pyrogenicity
Acute Systemic ToxicityNo acute systemic toxic characteristics
Neck TapeCytotoxicityNon-cytotoxic
SyringeN.A.N.A.

9.3 Sterilization and Shelf Life

ScalpelCric is supplied in sterile condition and intended to be single-use. Validated EOsterilization procedure is used for sterilization. Residuals do not exceed the maximum amount defined by ISO 10993-7. The maximum shelf life of the Set is 5 years.

9.4 Bench Testing

The components of the ScalpelCric Set are extensively tested to prove the fulfillment of requirements defined during development. Tests could verify the suitability of all tested features.

Object under TestFeatures to be testedResult
Bonding ConnectorBonding between connector and tubePassed
Bonding Inflation LineBonding between inflation line andtubePassed
15mm ConnectorDimensions of 15mm ConnectorPassed
Luer Pilot BalloonDimensions of the female luerconnector in the pilot balloonPassed
CuffFunction of the CuffPassed
ScalpelFunction and design of the scalpelPassed
BougieBougie which shall fulfill all relevantspecification according [DI] and[MS_Bougie].Passed
TubeFunction and dimensions of the tubePassed
TubeTube is recognizable in X-Ray imagesPassed
extension TubeFunction and dimension of theextension tubePassed
Neck TapeNeck tape which shall fulfill allrelevant specification according [DI]Passed
and [MS_NT] after 4 timessterilization
SyringeFunction of the syringePassed
Tube with inflation line and cuff,Function and features of the tube withattached inflation line and pilotballoon and with cuffPassed
ScalpelCricBiological compatibilityPassed
MaterialFunction and ingredient of theScalpelCric materialsPassed
Tube and extension TubeOxygen resistance of Tube andExtension tubePassed
LabelDevice LabelPassed
BougieDevice Label BougiePassed
IFUIFU which shall contain all relevantinformation regarding [DI] and [ER].Passed
ScalpelCricScalpelCric Set can be used inOperating conditionPassed
ScalpelCricFunctionality after simulated transportPassed
Transport ValidationFunctionality after simulated transportPassed
SafetyVerify implementation andeffectiveness of all Risk ControlMeasures defined in the Risk AnalysisPassed
Suction CapConnecting tube with open suction capduring ventilationPassed
User NeedsUser Needs for ScalpelCric, defined in[SoP]Passed
ScalpelCricUsabilityPassed
HeaderbagRequirements of DIN EN ISO 11607-1and DIN EN ISO 11607-2.Passed
SterilizationSterilization Requirements forScalpelCricPassed
Business NeedsBusiness Needs for ScalpelCricPassed
Label Sterile PackagingSterile packaging label which shallcontain all relevant informationregarding the [SFL], [DI] and [ER].Passed
Label Outer PackagingOuter packaging label which shallcontain all relevant informationregarding the [SFL], [DL] and [ER]Passed
PackagingPackaging Requirements forScalpelCricPassed
Accelerated Aging SyringeQuality of the syringe after 5 yearsaccelerated agingPassed
Accelerated aging of PouchFunction of the sterile packaging after5 years accelerated agingPassed
BougieBougie shall fulfill all relevantspecifications after 1 year acceleratedagingPassed
ScalpelFunction of the scalpel afteraccelerated aging 1 yearPassed
ScalpelCricScalpel Cric Set application oxygenresistance after 1 year acceleratedagingPassed
SurgicricFunction and dimension after 5 yearsaccelerated aging of the Surgicric andthe function of the sterile packaging.Passed
Surgicricare identical to the devices inScalpelCric.Passed
SurgicricFunction and dimension after 3 yearsreal time aging of the Surgicricincluding Tubus and the function ofthe sterile packaging. The tested tubeand sterile packaging are identical tothe devices in ScalpelCric.Passed
SyringeQuality of the syringe after 3 yearsreal time agingPassed
BougieBougie shall fulfill all relevantspecifications after 5 yearsaccelerated agingPassed
ScalpelFunction of the scalpel after 5 yearsaccelerated agingPassed
ScalpelCric SetApplication oxygen resistance after 5years accelerated agingPassed
Cuff Tube Collapse included in theScalpelCric after 5 years ofaccelerated aging.Test required by DIN EN ISO5361:2013-03, Annex C: Resistanceof the VBM tracheal tube ventilationlumen collapse with cuff inflated insidea cylinder.Passed

{11}------------------------------------------------

VBM Medizintechnik GmbH ScalpelCric, Cricothyrotomy Set 510(k) Premarket Notification

Image /page/11/Picture/1 description: The image shows the logo for VBM in a bright blue color. The letters are large and bold, taking up most of the frame. The background is white, which makes the blue letters stand out. The logo is simple and modern.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the letters VBM in a bold, sans-serif font. The letters are all capitalized and are a bright blue color. The letters are arranged horizontally, with the V on the left, the B in the middle, and the M on the right. The background is white.

10 Substantial Equivalence Summary / Conclusion

Based on available 510(k) information herein provided, ScalpelCric, CricothyrotomySet is considered substantially equivalent to the predicate device "Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical" in terms of intended use, technology and performance specifications.

The VBM component "Extension Tube with Swivel Connector" has been compared separately to the reference device "Airway Connector with Flex Tube". The substantially equivalence has been demonstrated in terms of clinical and biological characteristics as well as technical features.

The subject device and the predicate device have the same intended use and the differences in technological features do not raise different questions of safety and effectiveness.

§ 868.5090 Emergency airway needle.

(a)
Identification. An emergency airway needle is a device intended to puncture a patient's cricothyroid membrane to provide an emergency airway during upper airway obstruction.(b)
Classification. Class II (performance standards).